ProteusXRaSM19 PDF
ProteusXRaSM19 PDF
Publications
Direction 2273022-100
Revision 19
Service Manual
Copyright © 2000, 2001, 2002, 2003, 2004, 2005, 2006 by General Electric Co.
Service Documentation
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
ii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Direction 2273022-100
Revision 19
i
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
If you have any comments, suggestions or corrections to the information in this document,
please write them down, include the document title and document number, and send them to:
ii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
REGULATORY REQUIREMENTS
This product conforms with the requirements of Council Directive 93/42/EEC concerning medical devices
when it bears the following CE marking of conformity:
0459
iii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
This product conforms with IEC60601-1-2:2001 EMC standard for medical devices.
Note: This equipment generates, uses, and can radiate radio frequency energy. The equipment
may cause or subject to radio frequency interference with other medical and non–medical
devices and radio communications. To provide reasonable protection against such
interference, the Proteus XR/a System (32, 50, 65, 80kW) complies with emissions limits for
a Group 1, Class A Medical Devices and has applicable immunity level as stated in EN
60601-1-2:2001.
However, there is no guarantee that interference will not occur in a particular installation.
Special precautions and other information regarding EMC provided in the accompanying
documents of this equipment shall be observed during installation and operation of this
equipment.
Note: If this equipment is found to cause interference (which may be determined by switching the
equipment on and off), the user (or qualified service personnel) should attempt to correct the
problem by one or more of the following measure(s):
Note: To comply with the regulations applicable to an electromagnetic interface for a Group 1,
Class A Medical Device, and to minimize interference risks, the following requirements shall
apply:
Note: Do not use devices which intentionally transmit RF signals (Cellular Phones, Transceivers, or Radio
Controlled Products) in the vicinity of this equipment as it may cause performance outside the
published specifications. Keep the power to these type devices turned off when near the equipment.
The medical staff in charge of this equipment is required to instruct technicians, patients, and
others.
iv
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an elec-
tromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
The Proteus XR/a system uses RF energy only for
RF Emissions Group1 its internal function. Therefore, its RF emissions are
CISPR11 very low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions Class A The Proteus XR/a system is suitable for use in all
CISPR11 establishments other than domestic and those
Harmonic emissions Not directly connected to the public low-voltage power
IEC 61000-3-2 applicable supply network that supplies buildings used for
Voltage fluctuations/ Not domestic
flicker emissions applicable purposes.
IEC 61000-3-3
v
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an electromagnetic
environment as described below:
Immunity Test IEC 60601-1-2 Compliance Electromagnetic Environment
Test Level Level
Electrostatic ±6 kV contact ±6 kV contact Floors are wood, concrete, or ceramic
discharge (ESD) ±8 kV air ±8 kV air tile, or floors are covered with synthetic
IEC 61000-4-2 material and the relative humidity is at
least 30 %.
±2 kV for power ±2 kV for
Electrical fast supply lines power supply
transient/burst lines Mains power quality is that of a typical
IEC 61000-4-4 ±1 kV for commercial and/or hospital environment
input/output ±1 kV for
lines input/output
lines
±1 kV differential ±1 kV
Surge mode differential Mains power quality is that of a typical
IEC 61000-4-5 ±2 kV common mode commercial and/or hospital environment.
mode ±2 kV common
mode
Voltage dips, Mains power quality is that of a typical
short < 5 % UT 0 % UT for 5 commercial and/or hospital environment.
interruptions (> 95 % dip in UT) sec If the user of the Proteus XR/a system
and voltage for 0.5 cycle requires continued operation during
variations on power mains interruptions, it is
power supply 40 % UT recommended that the Proteus XR/a
input lines (60 % dip in UT) system be powered from an
IEC 61000-4-11 for 5 cycles uninterruptible power supply or a battery.
70 % UT
(30 % dip in UT)
< 5 % UT
(> 95 % dip in UT)
for 5 s
Power Power frequency magnetic fields are at
frequency 3 A/m 3 A/m levels characteristic of a typical location
(50/60 Hz) in a typical commercial and/or hospital
magnetic field environment.
IEC 61000-4-8
vi
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an electromagnetic
environment as described below:
Immunity IEC 60601-1-2 Compliance Electromagnetic Environment
Test Test Level Level
Portable and mobile RF communications
equipment are used no closer to any part of the
[EQUIPMENT and/or SYSTEM], including cables,
than the recommended separation distance
calculated from the equation appropriate for the
frequency of the transmitter.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If
the measured field strength exceeds the RF compliance level above, observe the Proteus XR/a system to
verify normal operation in each use location. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the [EQUIPMENT and/or SYSTEM].
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
The Recommended Separation Distances are listed in the next table.
Note: These are guidelines. Actual conditions may vary.
vii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Recommended Separation Distances for Portable and Mobile RF Communications Equipment and the
Proteus XR/a system
Frequency of
Transmitter
150KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Equation
d= 1.2 d= 1.2 d= 2.3
Rated Power of
Transmitter DISTANCE DISTANCE DISTANCE
(W) (meters) (meters) (meters)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
viii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
ix
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
x
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
REVISION HISTORY
REVISION DATE OF ISSUE REASON FOR CHANGE
0 8, 24, 2000 FIRST RELEASE
1 9, 25, 2000 • Add table lubricant boxes replacement procedure.
• Add new table control board schematics and table control board adjusting
procedure.
• Update System Schematics and Tomo-link Schematics.
• Supplement tomo (option) installation procedure.
• Add collimator bulb replacement procedure.
• Modify touch screen calibration procedure.
• Update FRU parts list.
2 11,10,2000 • Update page i, 4-6, 4-7, 10-12 and 10-14
3 11,20,2000 • Add collimator mounting screws’ torque requirement. Chapter 2.
• Update renewal parts list and illustration. Chapter 9.
• Add collimator bulb replacement notes. Chapter 3.
• Update Schematics page numbers. Chapter 10.
• Add a note on removing vinyl protective films. Chapter 2.
• Add a note on ESC key in Service Software. Chapter 3.
• Add a trolley stud picture. Chapter 2.
• Add a torque requirement to Trunion screws. Chapter 2.
• Add a Trunion installation warning. Chapter 2.
4 02,05,2001 • Update renewal parts list. Chapter 9
• Update system console mounting Job Card. Chapter 2.
• Add printer mounting Job Card. Chapter 2.
• Add system console replacement Job Card. Chapter 3.
• Add notes. Chapter 2, 3, 4, 8.
• Add OTS console schematics. Chapter 10.
• Add DOSE Calibration Job Card. Chapter 4.
• Add collimator alignment Job Card. Chapter 4.
• Add printer cables connection. Chapter 2.
• Add collimator update Job Card. Chapter 3.
• Add a maintenance item about ‘LAT bar’. Chapter 8.
5 09,28,2001 • Update room dimension in Chapter 2
• Add collimator D1&D2 trouble shooting and replacement in Chapter 3
• Transfer the Job Card SW JEDI 7-4,7-3,7-2 from Chapter 3 to Chapter 4
• Update touch screen calibration and AEC calibration in Chapter4
• Add new error code in chapter 7
• Add new Job Card PM010 in Chapter 8
• Update renewal parts list in Chapter 9
• Add schematics in Chapter 10
6 11,16,2001 • Update collimator mounting procedure in Chapter 2
• Update PM010 in Chapter 8
7 06,21,2002 • Update renewal parts in Chapter 9
• Update 90WS SW installation in Chapter 4,Grip adjust/EMC in Chapter2
8 11,03,2002 • Update renewal parts in Chapter 9
xi
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
xii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
xiii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
xiv
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
xv
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
TABLE OF CONTENTS
Chapter Page
xvi
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Chapter Page
6 JEDI GENERATOR THEORY AND FUNCTIONAL DESCRIPTION 6-1
SECTION 1 – THEORY OF OPERATION 6-1
SECTION 2 – THEORY PRESENTATION 6-3
10 SCHEMATICS 10-1
SECTION 1 – PROTEUS XR/a SYSTEM MIS MAP 10-1
SECTION 2 – JEDI GENERATOR ILLUSTRATIONS 10-25
SECTION 3 – JEDI GENERATOR CENTRAL LISTING 10-50
SECTION 4 – OTS CONSOLE ILLUSTRATIONS 10-62
xvii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
xviii
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The goal of this document is to guide the service representative through an entire
X-ray room installation.
The following On-site Requirements must be accounted for prior to beginning
installation of the system.
This direction will lead you in a correct sequence through the manuals needed to
install the equipment. Following the suggested sequence will result in an orderly
installation. However, extenuating circumstances may dictate a deviation from the
recommended sequence. Feel free to deviate, but keep track of your progress by
checking off the tasks in the flowchart as you complete them.
Check equipment according to the packing list (Product Delivery Instructing) for
your system against the main catalog numbers in the “System Listing Chart.” Also,
verify that all manuals are present in the binders according to the Table of
Contents.
All interconnection cables are furnished in variable, preselected lengths from the
“Cable Data Packet.” Before starting to install your system, identify all the cables
and check their length to verify that they are long enough. If you find a short cable,
contact us immediately so we can send the required length cable to you and
thereby, not delay the installation.
To obtain a good overall view of how these objectives are to be achieved review
the Table of Contents for this document before beginning the installation work.
Then, if it is necessary to deviate from the stated plan, you will be aware of the
essential interface areas. This will give you the information required to make
another effective plan which applies to our particular situation.
2-2 Interconnection
Interconnection of products by furnished cables or contractor supplied wiring is
indicated by numbered MIS runs in the installation sections. Run interconnect
charts are available which describe the connection and function of each wire.
2-3 Presentation
The procedures in this steering manual are presented in job card form.
1-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
There are additional tools and test equipment besides the standard service
representative tool kit which must be available before installing the Proteus XR/a
System. Refer to each subsystem service manual for details.
3-2 Documentation
Directions are necessary for correct installation of the Proteus XR/a system.
Procedures from each document will be used and reflect the latest information
available at this time.
Not all procedures in this manual apply to all sites. Before you begin installation,
cross out the procedures that do not apply to your site.
The proper system cable interconnect map is necessary for correct installation of
a Proteus XR/a system. Refer to:
1-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7
8
9 / 10
1 2
1-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Each system is installed in the sequence that will insure the safest and most time
efficient installation. The installation sequence performed at most sites is as
follows:
DONE
RECEPTION
STEP CHECK, PERFORM IST 001 JOB CARD
1 UNPACKING
ASSEMBLE
STEP OTS PERFORM IST 002 JOB CARD
2 CARRIAGE
CONNECT
STEP PERFORM IST 003 JOB CARD
OTS CABLES
3
POSITIONING
STEP PERFORM IST 004 JOB CARD
THE TABLE
4
MOUNT
STEP PERFORM IST 005 JOB CARD
GRID
5
INSTALL
STEP PERFORM IST 006 JOB CARD
TABLE TOP
6
2-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
MOUNT
STEP SYSTEM PERFORM IST 010 JOB CARD
8 CONSOLE &
PRINTER
WARNING
STEP SIGNS AND PERFORM IST 011 JOB CARD
9 LABELS
STEP
OR
11
ELECTRICAL
STEP PERFORM IST 020 JOB CARD
CONNECTIONS
13
2-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
ALIGN TABLE
STEP BUCKY WITH PERFORM IST 021 JOB CARD
14 COLLIMATOR
ALIGN WALL
STEP BUCKY WITH PERFORM IST 022 JOB CARD
15 COLLIMATOR
MOUNT
STEP ANCHOR PERFORM IST 023 JOB CARD
16 BOLTS
TOMO-LINK
STEP PERFORM IST 024 JOB CARD
ASSEMBLY
17
INSTALLATION
STEP PERFORM IST 025 JOB CARD
CHECK
18
APPLY
STEP SPANISH PERFORM IST 033 JOB CARD
19 LABELS
Note: If your country requires the product rating plate label to be in the
language of that country, follow the instructions on document
#45474336. Paste the appropriate language label over the English
rating plate. The rating plate labels are: 2379361 (Spanish),
2379361-2 (German), 2379361-3 (French), 2379361-4 (Italian).
Note: Spanish Label Kit (2382506), Focus Label (46-23972P2), and Console Label
(610-3089) are in the OPEN ME FIRST box in Crate #4-6.
2-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1) Verify all required cable ductwork, conduits, and raceways are installed according to Site Plans. Ensure
Customer supplied cables/wires are in place for connection to the system.
• Ensure that High Voltage and power cables are separated from signal cables as
much as possible.
• Verify that 3-phase power and ground lines can be placed in the same trough or
conduit.
2) At this time, the “Sales/Service Pre-installation Checklist,” provided in the applicable planning direction,
should be reviewed and signed off by the salesman and the Service Department representative.
2. Unpacking
The PROTEUS XR/a system is shipped in four crates to facilitate transport and installation. Crates
dimensions and weights are shown in Table 1-1 below.
2-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• Do not discard any packing material such as envelopes, boxes, bags until all parts
are accounted for against the packing list.
• Check the mechanical condition and external appearance of all parts for possible
Note: Do not use the capacitors in the Accessories crate #4 for installation. Please pay attention to pick
out and dispose the capacitors in the Accessories crate #4. Ensure the eco-friendly disposal of
the capacitors.
2-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Unpacking
Remove Cardboard box and wood shipping support.
Cardboard &
Wood shipping
support
2-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
On both sides
2-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
On both
2-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
b) Place the front crosstie into position on bridge and bolt the
crosstie to the bridge.
2-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2258381/2378475
2-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Mount the trolley wheels on the bracket and fasten the four
screws.
4. Take off the wood stoppers. Then take out the shipping base.
2-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3600 3600
2-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. Remove right side two metal tabs and keep left side metal Metal
tabs loose enough. tabs
2-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Slide the table top assembly onto the bearing assemblies from one end.
Anti-collision
Note: Pay attention to the anti-collision
switches. Table top frame may collide
switch
with the switch.
To obtain the smoothest and most quiet ride of the table top, adjustments may be made if necessary to the
table top bearings. These adjustments are made to precise specifications at the factory prior to shipping, but
shipping and handling may make minor “fine tuning” adjustments necessary after the table is assembled.
Note: Ensure that there is no lubricant, dust or other feculence on the slide rod.
2-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Before adjusting, remove table top, then turn the table top around to opposite side. If the table
top still does not ride smoothly, proceed as follows:
While viewing the underside of the table top, move the table top back and forth. Check the bearings that ride
horizontally (eccentric bearings) for looseness or tightness. If eccentric bearings are loose, adjust eccentric
bearings out. If bearing is too tight, adjust eccentric bearings in.
2-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: The content with a star is only for the wall stand
with a knee spacer. *
* Fix the carriage at a lower position.
2-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: The floor where the Wall Stand installed must be flat and level to prevent the Wall Stand from
falling down.
Note: Before installing the SG100 Wall Stand, the X-ray tube supports and table must firstly be
installed.
Note: The procedures in this section are concerted with the revision 6 of SG100 Wall Stand Service
Manual (GPCP Number 2150297-100). For the reason of the update and change in that
manual, it is recommended to see also the latest Service Manual of SG100 Wall Stand for
reference.
1. Move the SG100, in the shipping container, to the installation area. (Refer to page 5-2, 5-3 in Proteus XR/a
Pre-installation manual for Room Layout). Leave the container in the horizontal position. Unpack the SG100
from the shipping container.
2-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3. Stand the SG100 against the wall and attach to the wall using the locally provided 6 wall anchors.
4. Align the SG100 laterally across the wall, to the X-ray Tube Suspension beam or table centerline using the
centerline marks on the wall and floor or the Laser Alignment Tool.
Note: The SG100 must be level on the wall with no more than 1/16 in. (1.5mm) deflection to
ensure that the anchors holding the SG100 to the wall does not come loose over time.
Use the shims provided with SG100 to ensure this requirement.
5. Verify the SG100 level across wall to ensure that no more than 1/16 in. (1.5mm) deflection exist, laterally or
vertically. See Illustration for recommended placement of the levels.
6. Insert supplied shims, as required, between the SG100 and the wall.
7. Tighten the six SG100 anchor bolts according to the torque specifications for the anchors.
8. Check the Counterweight and Bucky Carriage Counterbalance, and add or remove Counterweight when
required. (See SG100 Service Manual, GPCP Number 2150297-100)
2-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
THE SG100 WALL STAND MAY BE FACTORY SET FOR 110VAC OR 220VAC. THE
WARNING MISMATCH BETWEEN THE VOLTAGE SETTING OF SG100 AND THE POWER SUPPLY
VOLTAGE FROM SYSTEM PDU CAN RESULTS IN THE BROKEN POWER FUSE AND
LEAD TO THE FAILURE OF SG100 BUCKY. PAY ATTENTION TO CHECK AND
ENSURE THE SG100 WALL STAND LINE VOLTAGE AND BUCKY VOLTAGE TO BE SET
220VAC AND 230VAC RESPECTIVELY (SEE STEP 9,10 BELOW) ON THE INSTALL
SITES BEFORE POWER ON.
9. Check the lock board jumper according to the table below to ensure the line voltage setting is 220VAC. Set
the TS-6 jumpers to 220VAC when necessary.
To access the lock board, remove the angulation support trim cover (behind the Bucky panel, at a 45 angle
facing the floor.). Remove the four screws. (See following Illustrations.)
The SG100 Wall Stand line voltage may be factory set for 110VAC or 220 VAC. Pay
CAUTION attention to check and ensure the line voltage to be set 220VAC on the install site via
jumper settings on its lock board.
Check and set the jumpers when necessary according to the following table:
LINE VOLTAGE Board: 45433412
TS-6 JUMPERS
220 VAC 2-3 & 4-5
110 VAC 3-5 & 4-6
2-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10.Set the Bucky voltage jumpers to the appropriate position for 230V. The Bucky voltage jumpers are
located next to the signal transformer in the electronic compartment behind the front panel (see
following illustrations).
The SG100 Wall Stand Bucky voltage may be factory set for 115VAC or 230 VAC.
CAUTION Pay attention to check and ensure the Bucky voltage to be set 230VAC on the install
site via Bucky voltage jumper.
To access this electronic compartment, turn the large knurled screws located underneath the front panel
counterclockwise. Tilt the front panel up and secure it with the prop on the right hand side. Then loosen,
without removing, the three flathead screws on the bottom of the lower panel, remove the two top roundhead
screws and slide the black cover plate up and out to access the jumpers.
Knurled
Screws-Loosen
to open Bucky
front panel
2-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
11. Route and connect system cables to the I/O panel at the top of the SG100 (see table below).
Cable Name Cable SG100 end (stick to connectors on Other end
Part the I/O panel at the top of the
Number SG100) (See illustration “SG100
connectors” below)
SG100 Ion Chamber Cable 2145642-2 J40 Jedi connector AEC2 (refer
to page 2-92 in Job Card
IST 020 “Electrical
Connections”)
SG100 Auto Tray Cable 2145644-2 J41 System I/F Board
connector J124 (See
illustration “System I/F
Board connector J124” in
next page)
SG100 Bucky Cable 2145641-2 J42 Jedi Wall Bucky (J6)
connector (refer to page
2-92 in Job Card IST 020
“Electrical Connections”)
SG100 Power Supply 2151729-2 Power (See illustration “SG100 PDU connector TM4
Cable connectors” below) SG100 Power (See
illustration “PDU connector
TM4 SG100 Power” in next
page)
SG100 Ground Cable 2140362-2 Ground PDU ground
2-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
J124
12. Attach the self-adhesive height correlation scale to the SG100 Wall Stand so it corresponds with the height
scale on the x-ray tube hanger or stand.
2-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Vertical Bucky Wall Stand not equipped with the automatic collimator option do not have
positioning microswitches.
Check and ensure the parameter value in System Console to be the same with the actual distance between
Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low POS. distance”,
Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High Auto
CAUTION Exposure position set in System Console (see parameter 9 “High, low POS. distance”,
Job Card CAL 002) is 40cm, which is different from the default distance set on SG100
Wall Stand (44cm). So it’s a must to check and ensure the parameter value in System
Console to be the same with the actual distance between Low Auto Exposure position
and High Auto Exposure position when performing the SG100 SID calibration (refer to
Job Card CAL 006 in Chapter 4), even if the positions of the Microswitch S4 Activator
Cam and Microswitch S5 Activator Cam are not modified and be on the default position
in the upper step.
2-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. Unpacking
SG120 is shipped in one crate to facilitate transport and installation.
1) Move SG120, in the shipping container, to the installation area. Leave the container in the horizontal
position.
2) Remove the banding and cardboard box that is covering the equipment and clamped to the wooden pallet.
3) Remove the shipping wooden crosspieces holding SG120 to the pallet.
4) Rest the bottom end of SG120 on soft support material outside the shipping container.
2. Standing
1) Before standing, remove the top cover to ensure the pulley chains are on their sprockets. If they are not,
try to set the chains on the sprockets. USE CAUTION.
If the problem cannot be easily corrected, SG120 must be returned as DOA.
2) Using two people, carefully raise SG120 to upright position.
INJURY HAZARD! AT LEAST TWO PEOPLE ARE REQUIRED TO TURN OR TILT SG120.
WARNING SG120 WEIGHS APPROXIMATELY 180 KG (330 LBS) AND CAN EASILY BE TIPPED OR
UNBALANCED.
3. Floor mounting
Tools required:
- Standard service person tool kit
- Drill template
- Wrench set
- Laser alignment tool
2-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Procedure:
1) Using the drill template supplied with the equipment (included in installation kit), set the position of SG120
taking into account.
Levelling screws
2-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: SG120 Vertical Bucky Stand can be either LEFT or RIGHT loaded. The equipment is
factory LEFT handed configured. The procedure to change the configuration follows.
MAKE SURE THE EQUIPMENT IS SWITCHED OFF DURING THIS OPERATION.
WARNING DURING THIS OPERATION, THE BUCKY CARRIAGE AND COUNTERWEIGHT
CARRIAGE CAN MOVE.
4. Bucky Reversal Procedure
1) Remove covers and panel
a. Remove the back covers from the bucky support assembly.
2-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
a. Remove the plate below the bucky support assembly to access to the interface board.
Interface Board
e. To assembly the handle on the new side, first remove the white plug for cable insertion and
the four screws.
2-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
f. Pass the switch cable through the hole on the new side and fix the handle with the 4 screws.
g. Connect the switch cable to J8 on interface board and assembly the plate removed in step a
in 2) above.
3) Bucky assembly
To change the configuration of the bucky for the cassette tray insertion, follow the next procedure:
Note: There is some difference between with patient support and without patient support.
Please confirm whether you’ve installed patient support.
a. Extract the cassette tray with CAUTION. THE BUCKY CARRIAGE WILL MOVE UP.
b. Remove the bucky cover protecting the bucky electronic devices.
c. Remove the 8 white plugs (4 on each side) located on the back of the bucky support.
Note: Remove the 6 white plugs (3 on each side) located on the back of the bucky support,
for the other 2 on the top (1 on each side) are covered if the patient support option is
installed.
Four more
plugs on the
other side.
2-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
d. Remove the 7 screws that hold the lateral cover of the bucky: 3 of them are situated on the
bucky frame and the other 4 are accessible from the back of the bucky support through the
holes left uncovered in the previous step.
Note: When the patient support option is installed, one of the 4 holes located on the back of
the bucky support (the one of the top on either side) is partially obstructed by the
magnets that are assembled with the patient support option. To access the four screws
from the back of the bucky support referred in step d, it is first needed to disassembly
the magnets that obstruct the holes.
e. Once the lateral cover is free, move it to the opposite side of the bucky to fix it. To enable the
displacement of the lateral cover, It is needed you cut the flanges that hold the cable to the
bucky frame.
2-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Since the ion chamber is assembled to the bucky, you can temporary displace it to let
the lateral cover pass through the area between the bucky structure and the ion
chamber. For that, just loosen the screws that hold the ion chamber and move it in
order to facilitate the operation.
f. Guide the cable, making sure that you do not interfere with the internal bucky mechanism
and fix it with flanges to the bucky sheet metal frame.
2-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
g. Make sure about the guiding of the cable. See below bucky for LEFT handed loading
(factory) and RIGHT handed loading.
MEDYS Bucky
cable
LEFT
cable
RIGHT
h. Fix the lateral cover to the opposite side with the 7 screws removed in step c in 3) above.
Note: When the patient support is installed, remember to fix the magnets removed in order to
be able to access the screws.
i. Insert again the 8 white plugs removed in step c.
Note: When the patient support is installed, there are 6 white plugs need to be insert.
j. Assembly the bucky cover removed in step b.
k. In the front panel, change the position of the cover to the opposite side.
Assembly this
cover on the
opposite side.
l. To finish, assembly the front panel and the back covers removed in 4. 1) (Remove covers
and panel) above.
m. Insert the tray using the new configuration.
2-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Check the Counterweight and Bucky Carriage Counterbalance, and add or remove
Counterweight as following when required.
Cassette tray
insertion
Pulley chains
5) Check that the counterweight and bucky carriage are well balanced:
- WITH POWER ON: Release the electromechanical brake and check the bucky carriage can
move easily in vertical direction.
- WITH POWER OFF: Check that the bucky carriage can move freely and smoothly in vertical
direction.
2-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Metal cover
Metal cover
Counterweights
7. Electrical Installation
1) Power cable connection
Power and ground cables are supplied optionally with the equipment.
a. Connect power supply cable from J1 VAC on the bucky control board.
b. Connect the ground cable to the ground stud on the bucky control board.
2) System cable connection
System Cable Connection (Code: 2145641-2, 2145642-2 and 2145644-2) is supplied optionally with SG120.
2-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
a. Connect Bucky cable (Code: 2145641-2) to J42 on the bucky control board.
ENSURE THAT THE BUCKY IS SET CORRECTLY FOR THE GENERATOR VOLTAGE
WARNING (115VAC OR 230VAC). THE BUCKY IS SET FOR 115VAC OR 230VAC IN FACTORY TO
AVOID EQUIPMENT DAMAGED. PAY ATTENTION TO CHECK THE VOLTAGE, AND SET
JUMPERS BY REFERRING TO THE VOLTAGE CHANGE INSTRUCTIONS IN THE BUCKY
MANUAL.
b. Connect Ion Chamber cable (Code: 2145642-2) to J40 on the bucky control board.
c. Connect Sensing Tray cable (Code: 2145644-2) to J41 on the bucky control board.
2-38
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
J124
2-39
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Vertical Bucky Wall Stand not equipped with the automatic collimator option do not have
positioning microswitches.
2-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Check and ensure the parameter value in System Console to be the same with the actual distance between
Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low POS. distance”,
Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High Auto
CAUTION Exposure position set in System Console (see parameter 9 “High, low POS. distance”,
Job Card CAL 002) is 40cm, which is different from the default distance set on SG100
Wall Stand (44cm). So it’s a must to check and ensure the parameter value in System
Console to be the same with the actual distance between Low Auto Exposure position
and High Auto Exposure position when performing the SG120 SID calibration (refer to
Job Card CAL 006 in Chapter 4), even if the positions of the Microswitch S4 Activator
Cam and Microswitch S5 Activator Cam are not modified and be on the default position
in the upper step.
2-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
There are cutting operations to be made on some pre-cut areas in most of the types of
CAUTION SG120 covers.
The cutting operation has to be carried out with CAUTION to avoid personal injury or
damage to the cover.
It is not recommended to carry out the cutting work in only one step, but to repeat the
cutting operation carefully and as many times as necessary until being able to remove
easily the cover part to reject.
a. Cut with an appropriate tool the pre-cut slot located on one of the sides of the front bucky
cover for cassette tray insertion. If you wish LEFT hand load, cut the left side of the cover; if
you wish RIGHT hand load, cut the right side of the cover.
Note: The equipment is factory LEFT handed configured, so the LEFT pre-cut side of the cover
should normally be cut, unless you want to inverse the tray insertion. If you wish to change
the hand load configuration, follow the procedure detailed in 4 above.
2-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: To disassembly the different bucky covers, loosen the screws shown below.
b. Connect the ground cable to the ground stud on the bucky control board.
2-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2) Column covers
The column frontal cover is assembled at factory. If you need to disassembly it, follow the next procedure:
a. Remove the screws (4 in total, 2 on each side) fixing the cover to the squares at the bottom
of the column.
b. Remove the 2 screws fixing the cover to the top of the column.
c. Move the bucky carriage assembly to the lowest position and remove the front cover with
care.
The column lateral and top covers are supplied disassembled into the packing box.
2-44
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
c. Assembly one of the lateral covers (2 screws) to the plate located on the column top.
d. Fix the lateral cover to the square located on the foot of the column (1 screw).
2-45
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
e. Take the second cover and make it fits with the cover already assembled.
f. Fix the second lateral cover to the top and foot of the column (steps b and c).
g. Connect the ON/OFF led cable supplied with the equipment and assembly the top cover
(4 screws) to the plate on the top of the column.
Note: Some of the holes to fix the covers have some safe clearance enabling to adjust the
assembly in the best conditions.
2-46
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Not to change this screw will cause that cassette tray falls when rotating.
2-47
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
WARNING THE LIMIT ROTATION SCREW SHOULD BE AT S2-1 WHEN LEFT HAND
CONFIGURATION, AND AT S2-2 WHEN RIGHT HAND. IN OTHER CASE THE
ROTATION (SEE FIGURE LEFT CONFIGURATION AND FIGURE RIGHT CONFIGURATION IN
SECTION 5-3-3-3) WILL BE DONE IN THE WRONG WAY AND THE CASSETTE TRAY WILL
SLIP WHEN ROTATING. HEEDING THIS WARNING MAY CAUSE INJURY TO THE
OPERATOR.
2-48
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
c. Take the upper rear cover included in patient support option kit.
d. Cut the pre-cut windows of the upper rear cover (the one chosen in previous step) lodging the patient
handgrips. Replace the old upper rear cover by this new cover and leave it ready for further assembly.
e. Discard the old upper rear cover.
f. Follow the procedure described in IST025 of this chapter, referring to installation of patient support option.
g. Assembly the new upper and lower rear covers.
Figure: Look of rear bucky covers with Patient Support OPTION installed
2-49
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: There are three types of mounting available for each system. You should choose one according
to the system configuration and customer’s requirement.
Note: Connect the cables according to the job card IST 020.
610-3081
2-50
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
610-3067
2-51
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
610-3068 610-3091
2-52
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-53
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
610-3091
3-5 Hang the bracket and console onto the bolts on the
wall.
2-54
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-55
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-56
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: There four types of warning labels are delivered as accessories in “OPEN
ME FIRST” box.
2-57
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-58
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Upper
Clamp
Screws
2. Loosen the four screws of the upper clamp of tube
bracket.
3. Remove the upper clamp of tube bracket.
Screws
TO PREVENT THE CLAMPS FROM ANY IMPACT.
WARNING
IN CASE IMPACT OCCURRED, IT SHOULD BE
VERIFIED THAT THE CLAMP HAS ENOUGH
STRENGTH BEFORE INSTALLATION.
2-59
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: This Job Card is only for MX100 X-ray tube. For the assembling of RAD14, 21 X-ray tube, please
refer to Job Card IST 012 in this Chapter.
1. Assemble the lead cup (GPCP No.: 2379458) to the MX100 X-ray tube output window if the lead cup is not
attached before arrival at the install site.
BECAUSE OF THE POSSIBILITY OF IMPAIRMENT TO HUMAN HEALTH, PLEASE PAY
WARNING ATTENTION TO WEAR GLOVES AND PREVENT THE DIRECT CONTACT WITH HUMAN
BODY WHEN CARRYING OR ASSEMBLING THE LEAD CUP.
Note: The lead cup should only be applied with MX100 X-ray tube (GPCP No.: D2301R) and Auto
collimator GPCP No. 2259298-54 (or Manual collimator GPCP No. 2259989). Do not use Lead cup
with X-ray tubes or collimators of any other types, such as RAD14,21 X-ray tubes.
1) Clean the output window cover of MX100 tube.
Output
window
2) Remove the self-adhesive tape circle cover from the lead cup and attach the lead cup onto the output
window cover of MX100 tube. Apply pressure to attach tightly and ensure the fixation.
Remove the
self-adhesive Lead
tape circle cup
cover
2-60
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Attach the spacer to the tube as shown in illustration below with four hexagon socket head bolts (GPCP No.:
46-170498P11, see illustration). Apply locktite to each bolt. The flat side must be away from the tube output
window.
Tube
output
window
Note: Apply locktite to mount screws to ensure the firmly tube assembling.
Pay attention to the mounting direction of spacer. The flat side must be away from
CAUTION the tube output window.
3. Position the tube with spacer attached as shown in illustration below and secure them to the gooseneck with
four hexagon head bolts (GPCP No.: 001-4060-03, see illustration.). Apply locktite to each bolt.
2-61
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: This Job Card is only for RAD 14,21 X-ray tube. For the mounting of Proteus XR/a Collimator to
MX100 X-ray tube, please refer to Job Card IST 015 in this Chapter.
Note: For Manual Collimator, the transformer on OTS also needs replaced for proper lamp power supply,
please refer to Job Card IST002 and OTS service manual, after the collimator installation, please refer
to CAL 015 to calibrate lamp voltage.
1. Remove the three safety screws with red heads from the collimator (marked with white ellipses).
3. Loosen the three collimator mounting screws to their maximum permissible limit.
Remove all 3 transportation safety screws (marked with white ellipses), see
illustration A and B, before applying power to the collimator.
CAUTION
Before transportation tighten the transport safety screws A fully (they have to screw
in easily). Therefore the smallest x-ray field has to be set. For manual setting of the
collimator blades replace the left cover and adjust an opening of the collimator blades
of approx. 4.5 mm.
Turn off voltage before fitting or removing the cover.
Mind installation instructions and ESD regulations.
2-62
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6. Be careful not to damage the collimator fingers. Position the collimator on the X-ray tube. Be sure the detent
lever is not preventing the collimator from seating fully.
The collimator rotation lever must be fully depressed when mounting/installing the
CAUTION collimator to the mounting flange. Failure to do so could cause the collimator to mount
improperly and fall from the unit.
Note: Apply grease supplied to collimator mount flange and
collimator adapter for easy movement.
2-63
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7. Tighten the 3 collimator mounting screws in equal number of turns until they lock securely and can not be
turned anymore. Then back out each screw equally by maximum 1/2 turn, so that collimator can rotate freely
Verify that when the collimator is not in a detent location, the minimum rotate force is 20N.
9. Loosen the X-ray tube. Use level gauge to check that the
tube and collimator are in level position, and then fasten
the tube using trunion handle securely.
Note: For RAD 21 X-ray tube, there is no trunion handle.
Just need to fasten clamp screws.
Apply locktite to mount screws.
10. Mount the OTS Console.
2-64
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. Remove the three safety screws with red heads from the
collimator (marked with white ellipses).
Transportation safety screw B
2. Check the collimator aluminum plate quantity, if there is only one plate in the collimator-tube interface, please
order one plate 2383309. Make sure there are two plates installed on the collimator-tube interface.
Note: This action in step 2 is only for the MX100 Tube.
For RAD14,21 Tube, only reserve one piece of Aluminium plate.
(Refer to Job Card IST 014)
Aluminium
plate
3. Loosen the three collimator mounting screws to their maximum permissible limit.
Remove all 3 transportation safety screws (marked with white ellipses), see illustration
A and B, before applying power to the collimator.
CAUTION Before transportation tighten the transport safety screws A fully (they have to screw in
easily). Therefore the smallest x-ray field has to be set. For manual setting of the
collimator blades replace the left cover and adjust an opening of the collimator blades
of approx. 4.5 mm.
Turn off voltage before fitting or removing the cover.
Mind installation instructions and ESD regulations.
2-65
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Apply grease on
the inside surface
of adapter
2-66
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
8. Tighten the 3 collimator mounting screws in equal number of turns until they lock securely and can not be
turned anymore. Then back out each screw equally by maximum 1/2 turn, so that collimator can rotate
freely. Verify that when the collimator is not in a detent location, the minimum rotate force is 20N.
CHECK THAT THE MOUNTING IS CORRECT BEFORE TURNING THE TUBE BACK TO
WARNING NORMAL POSITION.
BE SURE THAT THREE THRUST BUSHINGS ARE SYMMETRIC TO THE CENTER AND HOLD
THE FLANGE FIRMLY.
10. Loosen the X-ray tube. Use level gauge to check that the tube and collimator are in level position, then fasten
the tube using trunion handle securely.
11. Connect the Collimator Power & CAN cable (GPCP No., 2258383).
Note: For MX100 X-ray tube, if the X-ray field and light field of Collimator are not aligned after mounting,
the performing of Job Card IST 032 “MX100 X-ray tube-Collimator Interface Plate Alignment” in
Chapter 3 can calibrate and ensure the alignment of X-ray field and light field of Collimator.
2-67
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Mount the collimator mounting flange and the OTS console bracket with four provided screws.
Note: If the flange has been mounted in the factory, skip this step.
USE SOME MATERIAL TO PROTECT TUBE
WARNING WINDOW WHEN FASTENING SCREWS.
PAY ATTENTION TO THE DIRECTION OF
FLANGE. THE SLOT SHOULD BE ALIGNED
WITH OTS CONSOLE BRACKET.
2-68
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3. Loosen the screw (1), put the two clamps outwards to the max opening position. Put the locking Block (3) inside
the holder.
Illustration: Eclipse Proteus Collimator Top View
2
3
2-69
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6. Loosen the locking screw, rotate the collimator to the normal position,
and then, tighten the locking screw.
Note: The collimator is in normal position only after a click is heard.
CHECK THAT THE MOUNTING IS CORRECT BEFORE TURNING THE TUBE BACK TO
WARNING
NORMAL POSITION.
8. Loosen the X-ray tube. Use level gauge to check that the tube and collimator are in level position, and then
fasten the tube using trunion handle securely.
Note: For RAD 21 X-ray tube, there is no trunion handle.
Just need to fasten clamp screws.
Apply locktite to mount screws.
9. Mount the OTS Console.
10. Connect the collimator cable (GPCP No., 2258383).
11. Fix the cable on the strain relief screw with M5 nut.
12. Mount the cable cover.
2-70
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-71
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4. Loosen the screw (1), put the two clamps outwards to the max opening position. Put the locking Block (3)
inside the holder.
Illustration: Eclipse Proteus Collimator Top View 2
3
2-72
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7. Loosen the locking screw, rotate the collimator to the normal position,
and then, tighten the locking screw.
Note: The collimator is in normal position only after a click is heard.
CHECK THAT THE MOUNTING IS CORRECT BEFORE TURNING THE TUBE BACK TO
WARNING NORMAL POSITION.
9. Loosen the X-ray tube. Use level gauge to check that the tube and Eclipse Proteus Collimator are in level
position, then fasten the tube using trunion handle securely.
10. Connect the Collimator Power & CAN cable (GPCP No., 2258383).
11. Fix the cable on the strain relief screw with M5 nut.
12. Mount the cable cover.
Note: For MX100 X-ray tube, if the X-ray field and light field of Eclipse Proteus Collimator are not aligned
after mounting, the performing of Job Card IST 032 “MX100 X-ray tube-Collimator Interface Plate
Alignment” in Chapter 3 can calibrate and ensure the alignment of X-ray field and light field of
Eclipse Proteus Collimator.
2-73
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-74
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Collimator Cable
2-75
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-76
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
13. Lower the tube and collimator to the table top until the
bottom of the collimator rests on the table top. The bottom
of the collimator should remain flush with the table when the
locks are engaged.
14. Make a reference measurement from the table top to the
center of the focal spot label and record this dimension
(See Figure 1 and Figure 2). It is recommended to use
a steel scale (as shown) with 0.5mm divisions.
2-77
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Steel Strap
2-78
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
21. Apply the focal spot label (46-239720P2) such that the
center of the focal spot label coincides with
the dimension recorded in step 13.
2-79
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Tube
fan
cover
2-80
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
space
8. Connect OTS Console CAN (2378473), OTS Console Lock PWR (2378475) to the back of OTS Console.
2-81
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Refer to Illustration “LOWER TUBE COVER INSTALLATION” on the next page while performing
the following steps.
13. Install the end cover mounting brackets (1&2) on both ends of the tube using three (3) screws for each bracket.
14. Install the bottom cover (4) using 2 screws and washers (5).
2-82
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Frontal of Rear of
indicator indicator
2-83
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
pin
Insert
16. Install one screw on each side of top cover. Engage each screw only a few turns to allow the tab on bottom
cover to slide under the screw head.
2-84
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
18. Squeeze the top and bottom cover together and tighten the two screws (8) that were started in step 16. When
complete, the cover gap should look like Illustration below.
19. Install 2 end cover mounting screws (9) in each end cover mounting bracket. One from the bottom and one
from the back. Engage each screw only a few turns leaving room for the end cover internal bracket to slide
between the screw head and the end cover mounting bracket (1&2).
20. Position the end covers (10&11) by sliding the slots on the internal bracket under the screw heads on the end
cover mounting brackets.
21. Tighten the four screws (9) that were only started in step 17. The bottom screw is accessed through a hole in
the end cover.
22. Secure the end covers (10&11) by installing two screws (12&13) through each end cap and into the top cover.
2-85
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
24. Using the same metal scale and the dimension recorded earlier, determine the location of the focal spot and
apply the new focal spot decal in the center of the anode X-ray tube cover.
2-86
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
THE PDU CONTAINS TWO POWER TRANSFORMERS. THE FACTORY DEFAULT INPUT
WARNING IS SET TO 380V. THE FIELD ENGINEER MUST ADJUST INPUT VOLTAGE ACCORDING
TO LOCAL POWER.
Jumper VB
Jumper VA
6. After checking components grounding and cable connections, power on the system.
2-87
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-88
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-89
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-90
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PDU HAS 380-480V POWER EVEN WHEN THE CONSOLE TURNED OFF BUT CIRCUIT
WARNING BREAKER ON.
2-91
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-92
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: For MX100 X-ray tube, pay attention to ground the tube grounding cable (Green-Yellow
group) and tube fan shield grounding cable (labeled with “F3”) to the tube casing. Refer to
the illustration on prior page.
Note: For Rad 14,21 X-ray tube, to avoid possible slipping, clamps must not be put over the plastic
labels wrapped around the end of the stator cable. Remove the plastic labels when
necessary. (See figures below)
Note: For Rad 14,21 X-ray tube, when replacing the old tube with new one, the stator cable strain
relief parts and Cable clamps used to fix the cables should be retained and re-installed to
the new X-ray tube. (See figures below)
1xM4 screw,
2xM4 flat washers.
Thermal cable Apply Loctite
242(46-170686P1)
clamp
mounting
block
Stator cable
Attach P-clamp Copper
using M4 hex. nut strap
2-93
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Please use the P-clamp (on the top of Jedi) to grounding the shield layer of the tube rotor
cable and thermal cable.
Make sure that cables should be placed at inner sides of the P-clamps, or else it will be
difficult to fasten the bolts.
Make sure the P-clamps are tight enough so that they can contact with cable shield well.
16 15
5 4 3 2 1
Bus DC A C M
- +
Note: Make sure the brown wire from inner shield is connected to J3 (4) DC BUS (+)"
2-94
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Table
Bucky (J5)
PIN1
Service
We suggest to use the shield Terminal (J8)
NOTICE
cables as the Safety Door and
Light Switch cables, and their
shield layer should grounded Inside Comm-
to the system ground. unication (J9)
2-95
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Cable hose
3. Hang the cable hose along stationary rail and make sure
OTS bridge can move freely.
2-96
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Cable 221-2872
2. Unplug the connector on the socket J15 of the PDU, plug the other connector of the cable 221-2872.
3. Connect the cable 221-2872 with the cable 221-2871. Cable 221-2872
Connect the cable 221-2871 with the printer.
2-97
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Connector # 1
Connector # 9
Connector # 2
Connector # 10
Connector # 3, 4
Connector # 5 Connector # 11
Connector # 6
Connector # 12
Connector # 7
Connector # 13
Connector # 8 Connector # 14
2-98
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Subassembly: Align Table Bucky with Proteus XR/a Collimator / Eclipse Proteus Collimator
Purpose: Installation Procedure Version no. 0
Date: 20/12/1999
NF No. Production date Serial No. Personnel: 1
From / From Time: 1h
/ To
To /
/
Table Bucky and Proteus XR/a Collimator/Eclipse Proteus Collimator Alignment
Note: The front panel of the wall stand is removable, so before this step, perform wall stand
panel alignment calibration first.
Note: The following procedures for Table Bucky/Collimator Alignment may require slight
repositioning of the table base. This movement may affect the position or arrangement of
the levelling plates, requiring an additional check prior to the final table mounting.
Note: Collimator calibration, OTS/collimator alignments and collimator light field to x-ray field
adjustments, as described in their respective manuals, must be performed prior to the
following procedure.
MAKE SURE ALL CABLE CONNECTIONS ARE CORRECT AND ALL GROUND WIRES
WARNING ARE FASTENED. DAMAGE TO EQUIPMENT COULD OCCUR IF COMPONENTS ARE
NOT PROPERLY SET BEFORE INITIAL UP.
1. If the table top is installed, remove the table top (keep
table top frame on) by removing the two screws at each
end of the table top underside. Then take the table top
out of the frame.
Note: Since the bucky has previously been “squared” to the bucky rails, any deviation from
alignment may be due to inaccurate alignment of either the OTS or collimator. If necessary,
repeat alignment procedures.
Note: Light field to x-ray field alignment should be performed before this step.
2-99
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Subassembly: Align Wall Bucky with Proteus XR/a Collimator (for Wall Stand (GPCP No. 2260354) and
SG100 Wall Stand (GPCP No. 2225849)), Eclipse Proteus Collimator (for SG120 Wall
Stand (GPCP No. 2402562))
Purpose: Installation Procedure Version no. 0
Date: 20/12/99
NF No. Production date Serial No. Personnel: 1
From / From Time: 30 min
/ To
To /
/
Note: Proteus XR/a Collimator, Eclipse Proteus Collimator, SG100 Wall stand and SG120 Wall
Stand should follow the below contents.
Note: The following procedures for Wall Bucky/Collimator Alignment may require slight
repositioning of the wall stand. This movement may affect the position or arrangement of
the levelling plates, requiring an additional check prior to the final wall stand mounting.
Collimator calibration, OTS/collimator alignments and collimator light field to x-ray field
adjustments, as described in their respective manuals, must be performed prior to the
following procedure.
Note: For SG100 and SG120 Wall Stand, ensure the Bucky to be in vertical position before
performing the following steps.
Note: Light field to x-ray field alignment should be performed before this step.
2-100
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-101
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: For SG 100 Wall Stand, the Anchor bolt mounting should follow the steps from 11 to 16,
instead of step 5 to 10.
Note: The procedures in following steps are concerted with the revision 6 of SG 100 Wall Stand
Service Manual. For the reason of the update and change in that manual, it is
recommended to see also the latest Service Manual of SG 100 Wall Stand for reference.
11. Using two people, carefully raise the SG 100 to an upright position.
12. Stand the SG100 against the wall and attach to the wall using the locally provided (6) wall anchors.
13. Align the SG100 laterally across the wall, to the X-Ray Tube Suspension beam or table centerline
using the centerline marks on the wall and floor or the Laser Alignment Tool.
Note: The SG100 must be level on the wall with no more than 1/16 in. (1.5 mm) deflection to
ensure that the anchors holding the SG100 to the wall does not come loose over time.
Use the shims provided with the SG100 to ensure this requirement.
14. Verify the SG100 level across wall to ensure that no more than 1/16 in.(1.5mm) deflection exists,
laterally or vertically. See following Illustration for recommended placement of the levels.
15. Insert supplied shims, as required, between the SG100 and the wall.
16. Tighten the six SG100 anchor bolts according to the torque specifications for the anchors.
2-102
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: For SG120 Wall Stand, the Anchor bolt mounting should follow the steps from 17 to 23,
instead of step 5 to 10.
Note: The procedures in following steps are concerted with SG120 Wall Stand Service Manual
(P/N: S0012532). For the reason of the update and change in that manual, it is
recommended to see also the latest Service Manual of SG120 Wall Stand for reference.
17. Place SG120 at the distance focal dependant to the anti-diffusion grid.
18. Align SG120 laterally using the collimator light or the laser alignment tool.
19. Using the base template of SG120, mark the holes.
20. Move SG120 and drill the holes. Place the anchors.
21. Replace SG120 and place the anchor bolts.
22. Level the equipment using the 3 levelling screws and tight the bolts.
Levelling screws
2-103
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: For Tomo-Link installed with the system, refer to Job card IST 025 “Tomo-Link
installation” in Chaper3.
2-104
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Upper
Cover Lower
Cover
2-105
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6. Paste the console label provided with the equipment according to local regulation.
Note: After completing installation, remove vinyl protective films from table top and
pedals.
7. Abdomen Compression Band Installation
Adapting screw
Safety
Lock Lock
Lock
2-106
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9. Adjusting
Adapting
screw
10. Collimator Cone
Note: The following contents fit to both Proteus XR/a Collimator and Eclipse Proteus Collimator.
Note: The purpose of the collimator cone is to collimate the beam and image to a cylinder shape
when doing certain procedures. Mainly used for sinuses, head, hip etc.
Note: Adjust the alignment of the collimator cone with the collimator to ensure the concentricity of
the collimator cone and collimator window.
When taking exposure using the collimator cone, pay attention to choose correct
CAUTION cassette size to avoid the X-ray field to beyond the film size.
Accessory Rail
2-107
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Collimator Cone
ii. For removal, push in the Accessory Locking Spring and remove the collimator cone.
Accessory
Locking Spring
Note: If choose SG120 Wall Stand, please refer to the following contents.
Installation
Tools required:
- Standard service person tool kit.
Procedure
1) Do not discard any packing material such as envelopes, boxes, bags until all parts are compared
against the packing list.
2) Check the general condition and external appearance of all parts for possible damage or missing
items. Notify the distribution center immediately of any damage or missing parts.
2-108
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-109
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
c. Take the lateral bar steel weldment assembly and the Teflon strip that is supplied with the
kit and apply it with adhesive along the contact surface between the steel weldment
assembly and the slide track.
d. Fit the steel weldment assembly to the slide track and make sure it can be displaced along
the whole track.
2-110
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
e. Assembly the two magnets (one on each side) included in the kit to the bucky support. The
magnets will lock the inertial sliding movement when the bucky assembly is rotated, in the
case of SG120.
i. To remove the lateral bar, press the key on the lateral bar assembly and extract it by pulling
up on the right of the figure above.
6) Add the needed counterweight included in the kit and proceed to counterbalance the bucky and bucky
carriage as described in step 5 in Jobcard IST 009.
2-111
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
LOCATION OF SPAIN
LANGUAGE LABELS
2-112
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
LOCATION OF SPAIN
LANGUAGE LABELS
LOCATION OF SPAIN
LANGUAGE LABELS
2-113
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
DESIGNATION
Eclipse Proteus Collimator
SPANISH 5116700
LANGUAGE LABELS
PART NO
LOCATION OF NAME
PLATE
LOCATION OF SPAIN
LANGUAGE LABELS
2-114
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, be sure to wear the antistatic bracelet.
4 PREREQUISITES
None.
3-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Interface Unit
3-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
AEC board
J5
J1
J2
Rad I/F board
TP1
TP2
6 REASSEMBLY
Remount the Front cover. Reverse the disassembly procedure.
3-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2 SPECIAL TOOLS
Hex wrench, 10 mm
Hex wrench, 11 mm
3 ORDERING INFORMATION
Refer to Chapter 9
3-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D1-Board:
Verify the setting of the DIP switch on the D1-Board. The DIP switch S2 on the D1-board must be set as follows:
S2.1 = ON
S2.2 = OFF
S2.3 = ON
S2.4 = OFF
S2.5 = OFF
S2.6 = ON
S2.7 = ON
S2.8 = ON
Start the collimator calibration mode to get the calibration data stored in the EEPROM.
Note: Without calibration data in the EEPROM of the board D1 the collimator startup may fail.
For collimator calibration carry out the following steps:
a) Verify the DIP switch S2 on D1 (see above). It’s important that switch S2.4 is set to OFF.
b) Close the lamellas in both levels (longitudinal and transversal) manually.
c) Connect the collimator to the power supply.
d) Set dip switch S2.4 to ON
e) The collimator will start a calibration run. The lamellas will fully open to the photoelectric barriers and then
close to the minimum opening.
The calibration data will be stored in the EEPROM.
f) Verify the setting of the DIP switch on the D1-Board again. The new settings are
S2.1 = ON
S2.2 = OFF
S2.3 = ON
S2.4 = ON
S2.5 = OFF
S2.6 = ON
S2.7 = ON
S2.8 = ON
g) Now the collimator is full working.
h) Disconnect the power supply before closing the collimator.
3-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Close the front panel and fix it with 2 screws (2.5 mm allen wrench).
Add the two parts of the cover to the collimator and fix it with 9 screws
(2 mm allen wrench).
Pull out the marker parts (2/Fig.1) towards the front and remove them.
Using an Allen wrench (3/Fig.1), loosen the 1.5 mm clamping screws on the knobs as shown and remove the
knobs (1/Fig.1).
Using an 11 mm hex wrench, remove the securing nuts on the potentiometer shafts and take off the snap rings.
Fig.1
3-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D2-Board: 1
Fig.2
a) Pull the following plugs from the D2-Board; when doing so, make sure that all connectors are labeled. If labels
are lost, label these connectors with appropriate tags.
b) Remove the panel mounting (1/Fig.2) by removing the two 2.5 mm Allen screws (2/Fig.2), unhook the front
panel and put it down.
c) Remove the 5 cross-head screws, which are used to fix the board.
e) Remove the securing nuts and snap rings on the new D2 board.
Install the new D2 board. When doing so, make sure that the X11 pins (1/Fig.3) on the front panel are guided
exactly into the female connector on the D2. If the board is exactly installed, all the tips of the pins must be visible
from the component side of the D2X11.
3-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Fig.3
f) Secure the D2 with the 5 mounting screws and replug the connectors similar to a). Install the snap rings and
securing nuts (11 mm hex).
g) Place the knobs on the potentiometer shafts and secure them using the 1.5 Allen wrench. When the knobs
are turned, they may not scratch against the front panel. Insert the marker parts.
h) Install the front panel and install the panel mounting (1/Fig.2)
Fuse
1.6A 250V
Fuse
1.6A 250V
Tape
Measure
3-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Replace the whole collimator, collimator control board, user interface board, collimator
lamp assembly, bucky centering light, LCD display, tape
purpose: version no. 1
date: 15/12/06
NF No. Production date Serial No. Personnel:1
from / / from Time: 0h 30 min
to / / to
3-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• Loosen the 4 screws which are used to fix the control board
3-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
ONLY THE LAMP OFFERED BY GEHL WITH THE HOLDER CAN BE USED AS
WARNING REPLACED PART. OTHERWISE IT MAY CAUSE ISSUE OF LIGHT ILLUMINANCE,
EDGE CONTRAST OR LIGHT/X-RAY FIELD ALIGNMENT.
• Loosen 2 mounting screws of the lamp assembly from the hole on read side of the collimator and
remove the defective lamp assembly from the right side of the collimator.
3-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
To avoid damage the lamp, do not touch the bulb or inside of the reflector with bare
CAUTION hands.
3-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
LOCKOUT/TAGOUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
3-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Please get different length by using different blocks according to the table’s height by the below table.
Note: Table must not be driven up and down when the blocks are put over the rod.
4 3 2 1
3-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Example 2: The blocks 5137108-3 and 5137108-4 are over the rod.
4 3
3-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Example 4: The blocks 5137108-2 and 5137108-3 are over the rod.
3 2
Example 5: The blocks 5137108-1 and 5137108-2 are over the rod.
1 2
3-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4) Remove the old motor together with the motor chassis from the table.
3-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5) Demount the screws which fixing the motor and motor chassis.
Note: The new motor should be sent together with a gear. If it is, please go to step 2) to assemble
the motor chassis. Otherwise, please proceed to step 1).
1) Mount the gear onto the motor shaft with rubber or wood hammer. The gear should be hammered
very close to the motor. There may be a little adjustment while fixing the chain with the gear.
Note: The width of the two sides of the gear is not equal.
The narrow side of the gear should be placed outside.
3-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
b) Fix the motor onto the motor chassis with the rubber piece between the motor and
the motor chassis.
3-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
a) Place the rubber piece on the interface between the motor chassis and the table.
3-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
d) Partly mount the screws which fixing the motor chassis and the table.
Note: Don’t mount the screws fully. And make the motor could be shaken slightly on the table.
e) Modify the angle and position of the motor to fasten the chain by shaking.
3-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Bucky
Cable
Rail Seats
2. Remove the four bucky rail seats. Replace the old cassette
cabinet with new bucky.
3-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Insert the Ion chamber into bucky and locate the Ion
chamber on the mounting bracket.
Mounting Bracket
3. Fasten the four lock screws
Lock Screws
3-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5. Route the Ion-chamber cable with bucky cable through the Ion-
hole in the middle of the table. Use ty-rap to fasten the chamber
cables with adhesive disc and make sure the bucky move cable
freely without any cable scratching.
3-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The following procedures in addition to the Tomo-link Service Manual (Direction: 2133821-100) describe how to
install the Tomo-Link option.
Note: The Tomo-Link assembly should be installed with the
Proteus table and XT suspension only. The system
software should be Rev1.84, or above.
1 Drive unit installation (Inboard bridge), refer to Tomo-Link
Service Manual (Direction: 2133821-100).
Note: The driver unit can only be installed on right side of
bridge.
Cut a slot
2 Cut a slot on the top of cable hose. Then insert two driver
cables (2124056 and 2132261) into hose.
Plug in these two cables and route in hose properly.
4 For system with RAD 14 or RAD 21 X-ray tube, transfer the accessories (rubber spacer, top shim and all
screws) factory-mounted on the Tomo-link Yoke 2109740-8 to the Extended Tomo-link Yoke 2352540.
Note: For systems with MX100 tube, skip this step.
Note: The Proteus XR/a systems with Tomo are shipped with both the Tomo- link Yoke 2109740-8 and
the extended Tomo-link Yoke 2352540. But only the Tomo- link Yoke 2109740-8 is shipped with
the accessories (rubber spacer, top shim and all screws) mounted.
For systems equipped with RAD 14/ RAD 21 tube, apply the extended Tomo-link Yoke 2352540.
Remove the accessories from the Tomo-link Yoke 2109740-8 and mount the accessories to the
extended Tomo-link Yoke 2352540 before installing the Tomo-link Yoke. (See steps following)
3-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
a) Demount the screws and remove the rubber spacer from the Tomo-link Yoke 2109740-8.
top shims
b) Fit the rubber spacer to the Extended Tomo-link Yoke and mount the screws without fully fasten .
NOTE: Pay attention to leave loose the screws for Fulcrum arm alignment.
3-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
d) After aligned with the Fulcrum Arm (installed in step 14 of this Job Card), fasten fully the screws holding the
rubber spacer.
6 Take off the table back covers and the lower right side
cover.
Note: For Tomo-link upgrade only.
Cut out two mounting slots on upper back cover.
Mount slots
3-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7 Mount Tomo control box into table base using four M4x8
screws, washers and lock washers.
Note: For Tomo-link upgrade only.
Tomo Cables
3-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: If the X-ray beam and collimator light field are not aligned to the image receptor, you can
refer to Step 16 to adjust when necessary.
3-38
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Auto collimator
1. Power off the system.
2. Open the side window of collimator.
3. Replace the collimator bulb.
Note: Don’t touch bulb surface with fingers.
Use plastic cover on bulb to install it.
4. Open the left side cover of collimator, then replace the bulb
fan.
Note: This step is just used for Auto-collimator.
Manual collimator
1. Power off the system.
2. Open the rear cover of collimator.
3. Replace the collimator bulb.
Note: Don’t touch bulb surface with fingers.
Use plastic cover on bulb to install it.
3-39
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Introduction
This procedure is used when APR disk being lost or broken.
Procedure
1. Prepare a new formatted (Windows 95 or DOS6.22) data floppy disk.
Note: Don’t copy system files
2. Power on the system.
3. Press Procedure Edit button.
4. Insert the system console SW disk into the floppy Disk Drive.
5. Press OK to enter Procedure edit mode. See page 5-9 in Proteus XR/a Operator
Manual (Direction 2259724-100).
6. Press Save/Retrieve button to enter Save/Retrieve menu.
7. Take out the system console disk, then insert the new formatted disk.
8. Press Save Backup button.
9. Wait until “Retrieving APR data complete” is displayed.
10. Press EXIT button to exit this menu. APR disk recovery is done.
3-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Procedure
1. Remove both side covers and top cover on X-ray tube.
2. Disconnect two OTS Console cables (2259867 and 2258381 for RAD 14, 21 tube,
2378473 and 2378475 for MX100 tube, respectively).
3. Replace the OTS Console.
4. Connect two OTS Console cables.
5. Power on the system.
6. Check the display on OTS Console match the System Console.
Note: Ensure the coherent reading is same between the System Console and the
OTS Console (Check should be performed with incrementing/decrementing
technique factors from the OTS Console).
7. Power off the system.
8. Put on Tube covers.
3-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Procedure
1. Remove and disconnect the covers and cables. This procedure is a reversed procedure of the installation.
2. Replace the System Console.
3. Repeat the installation procedure. Reference to the Job Card IST 010 and IST 011.
4. Power on the system.
5. Check the display on OTS Console match the System Console.
Note: Ensure the coherent reading is same between the System Console and
the OTS Console.
The Rev 2.14 of the System Console Software 2286799-16 can be used together with
CAUTION either the System I/F board 2375230 (Item #1 of Cabinet 1) or the System I/F Board
2259298-45 and CAN I/F Board 2259298-46.
WARNING LABELS SHOULD BE INCLUDED IN “OPEN ME FIRST” BOX. IF THEY
WARNING ARE MISSING, PLEASE ORDER A NEW CONSOLE (2259976) IN CONFORMITY WITH
REGULATION REQUIREMENTS.
3-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D1-Board:
3-44
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Pull the Eprom from socket J10 and plug the new one which is labeled J10-D1.
Note: The Eprom fits only in one position (look for the slant sides of the Eprom and the socket).
3-45
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
S2.1 = ON
S2.2 = OFF
S2.3 = ON
S2.4 = ON
S2.5 = ON for AL01C
S2.5 = OFF for AL01F
S2.6 = ON
S2.7 = ON
S2.8 = ON
3-46
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Pull the Eprom from socket J9 und plug the new one which is labeled J9-D2.
Note: The Eprom fits only in one position
(Look for the slant sides of the Eprom and the socket).
Close the front panel and fix it with 2 screws (2.5 mm allen wrench).
Add the two parts of the cover to the collimator and fix it with 9 screws
(2 mm allen wrench).
3-48
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3. Replace the defective kV control board with the new one (Refer to Job Card D/R JEDI 17-48, Chapter 9).
Note: Pay attention to Write the existing number on the failed kV control board to the blank label, and
stick it on the new kV control board.
Note: Pay attention to verify the software version and ensure its compliance with the KV control board
and X-ray tube type (refer to Job Card IST JEDI 5-2, Chapter 4).
Alternately, you may try to swap the NVRAM chip (U32) from the former to the new KV Control board to keep all
the parameters (Refer to Job Card PM JEDI 15-2, Chapter 8). If this chip is not involved in the failure or the cause
for replacement, it will allow to keep up-to-date data.
5. Calibrate:
- Perform several exposures on both focuses in several workstations, with exposure time longer than 12.5
ms
If default database was restored, it is necessary to resume room set up (refer to Job Card CAL JEDI 8-4, Chapter
4) and AEC calibration (if present) (refer to Job Card CAL JEDI 8-2, Chapter 4). Then perform the checksum
(refer to Job Card CAL JEDI 8-5,Chapter 4).
6. Save Database on the service floppy delivered with the equipment (refer to Job Card CAL JEDI 8-6, Chapter
4).
3-49
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: OTS
purpose: MX100 X-ray tube-Collimator Interface Plate Alignment version no. 0
date: 7/03/03
NF No. Production date Serial No. Personnel:1
from / / from Time: 0h 30 min
to / / to
1. Proceed Alignment between tube and tube/collimator
interface plate.
1) Obtain the following focal spot alignment tools:
- A Collimator Alignment Tool 46-166390G1
- Hex Wrench, 46-165137P1
- An alignment tool adapter 2296120.
2) Place the Alignment tool adapter (2296120) on the tube/collimator interface plate.
3) Place the Collimator Alignment Fixture (46-166390G1) into the tool adapter ring.
4) Rotate the Collimator Alignment Fixture (46-166390G1) until the grid axis (y or x) is parallel to
the axis of the tube’s cathode/anode.
3-50
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6) Adjusts the position of the x-ray tube relative to the tube/collimator interface plate.
7) To make tube position adjustments, loosen the four socket head cap screws (See illustration on Page 2-
63 or Page 2-68.) 1/4 turn counterclockwise. The screws retain the x-ray tube to tube/collimator interface
plate.
8) Move the x-ray tube until the cross hairs in the alignment tool window are midway between the
two focal spots displayed.
9) Make certain the concentricity of tube/collimator interface plate, OTS console bracket and tube window.
10) After adjustment, torque each of the four mounting screws to the same force.
11) Remove the Alignment Fixtures.
Note: Apply grease supplied underneath the edge of tube/collimator interface plate and the
inside surface of the adapter on collimator for easy movement.
3-51
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-52
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Introduction
purpose: Calibration and Configuration version no. 0
date: 20/12/99
NF No. Production date Serial No. Personnel:1
from / / from Time: 10 min
to / / to
Sequence
Procedures
System Power On
Document Modifications
Do not attempt to run Jedi Service Mode without first marking your Cal Users Guide manual to reflect
changes covered in the Proteus XR/a Release Notes. Failure to do so may cause some Service Mode
Units to operate incorrectly.
The majority of the system calibration was performed at the factory. Therefore, only the unchecked
procedures need to be run. Should there be any problems with any of the factory settings, we recommend
that only those units required to correct the problem be run.
4-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. The initial parameters were set at factory. The field engineer needs to re-calibrate the system
configuration according to requirements.
Note: The System Console Software version is 2.14.
The parameters shown in the illustration below are not the default configuration in
factory. The illustration is used only to show the interface.
4-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
the wall center and table center detent code is same: the wall center and table center detent code is different:
0 1 2 3 NOTE 0 1 2 3 NOTE
0 X 0 X X Table & Wall 0 X 0 X X Table center
0 0 X 0 X Wall-XT-Switch A 0 0 X X X Wall center
X 0 X 0 X Wall-XT-Switch B 0 0 X 0 X Wall-XT-Switch A
0 X X 0 X Wall-XT-Switch C X 0 X 0 X Wall-XT-Switch B
X X X 0 X Wall-XT-Switch D 0 X X 0 X Wall-XT-Switch C
X X 0 0 X Wall-XT-Switch E X X X 0 X Wall-XT-Switch D
X X 0 0 X Wall-XT-Switch E
Note: The magnet red face upwards.
4-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Introduction
This Jobcard describes how to adjust initial setting of table inhibition button and anti-collision buzzer
function. (This function is available on Rev 01 or later version PCB)
When power on the system, the default setting of table inhibition button is on. If you want to operate the
table, you will need to deselect this button.
The following procedures describe how to adjust the initial setting of table inhibition button.
4-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
JP2
JP1
4-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Malfunction Analyze
1. Power on the elevating table.
2. Lower the table height, at the same time, use a force gauge to hold back table motion under each
corner of table top frame.
3. When the force exceed 45lbs, the table should stop motion, otherwise, that anti-collision switch isn’t
working correctly.
4-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Anti-
collision
Supportive
screw
Bracket
screws
Start
Yes
No No
Yes
No Table moves Yes anti-collision
downward switches can work No
Fasten two
bracket screws.
Anti-collision
switch adjust-
ment complete
4-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Procedures
1. Open the front cover of table base.
Safety
There are two sets of switches. One set is table height
switch Limitation
limitation switch, the other set is safety switch for
switch
limitation switch in case limitation switch can’t work.
The lowest
2. Analyze whether is the highest position or the lowest position detector
position not correct, then adjust the corresponding
detector’s position of the pedal I/F board.
The highest
position detector
4. Adjust detector again. Then check whether table height The highest
meets the specification. If not, do step 2 again. safety switch
4-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. When the table is in its’ highest position, verify that the mark
on table height potentiometer is in alignment with mark on
the rack. System
2. Open the front cabinet cover by loosing two screws I/F Board
underneath the cover.
3. Disconnect the connector J110 in system I/F board and measure
the resistance between pin1 and pin2.
When the table is in the highest position, it should be 40±10Ω.
When the table is in the lowest position, it should be 965±10Ω.
4. Check the cables are connected correctly. Refer to the
system MIS map in page 10-1.
5. Power on and check the +5VDC±5%, +/-12VDC±5% on the
system IF board.
+12V
-12V GND
+5V
+9V
TP2
TP1
4-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PA3
100K TP2
TP1
Note: Ensure the switch 1, 2 of the DIP Switch in red color to be on.
DIP Switch
4-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7. With the table in its lowest position drive to the highest position, verify the SID display
on the OTS console is changing during entire travel range without rolling over
(continuous pot).
Adjust PA1 if the SID doesn’t change.
Note: Raise the table to its highest position. Then adjust PA1 until TP1 is about 0.93V.
8. Verify the SID display on the OTS console is changing while driving the OTS up and
down.
Adjust PA3 if the SID doesn’t change.
Note: Raise the OTS to its highest position. Then adjust PA3 until TP2 is about 0.89V.
9. Drive the table to the lowest position and write down the SID value shown on OTS console. Drive the
table to the highest position. Adjust the PA1 so that the OTS console SID changes 30cm(Table height
is from 50cm to 80cm).
10. Repeat step 9 until SID changing is the same as actual moving distance.
11. Drive the OTS to the 100cm SID (from table top) using the measure tape on collimator and write
down the SID value shown on OTS console. Lower the OTS to 70cmSID (from table top) using
measure tape. Adjust the PA3 so that the OTS console SID changes 30cm.
12. Repeat step 11 until SID changing is same as real moving distance.
13. Repeat step 9 to 12 until both of table and OTS SID values changing reflect the real moving distance.
14. Adjust P5 until the right SID is displayed on the OTS console.
FOR SG100 WALL STAND SID CALIBRATION, CONTINUE WITH FOLLOWING STEPS. IF SG100 IS
NOT SELECTED, IT’S ENOUGH TO PERFORM THE STEP 1 TO 14 ONLY FOR SID CALIBRATION.
FOR SG120 WALL STAND SID CALIBRATION, SKIP TO STEP 20 DIRECTLY. IF SG120 IS NOT
SELECTED, IT’S ENOUGH TO PERFORM THE STEP 1 TO 14 ONLY FOR SID CALIBRATION.
4-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
16. Modify the Microswitch S4 Activator Cam and Microswitch S5 Activator Cam to decide the distance
between the Low Auto Exposure position and the High Auto Exposure position, write down the value
of the distance (Distance1). (When the Bucky in horizontal, there are only two positions of Bucky on
the column at which the Auto exposure can be made. Of these two positions, the lower one is
defined as the “Low Auto Exposure position”, and the higher position is defined as the “High Auto
Exposure position”. And these two positions can be modified by adjusting the Microswitch S4
Activator Cam and Microswitch S5 Activator Cam. The default distance between these two positions
is 44cm.). (Refer to Job Card IST 008, Chapter2.)
Check and ensure the parameter value in System Console to be the same with the actual distance
between Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low
POS. distance”, Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High
CAUTION Auto Exposure position set in System Console (see parameter 9 “High, low POS.
distance”, Job Card CAL 002) is 40cm, which is different from the default distance
set on SG100 Wall Stand (44cm). So it’s a must to check and change the
parameter value in System Console to ensure it to be the same with the actual
distance between Low Auto Exposure position and High Auto Exposure position
before performing the SG100 SID calibration according to the following steps, even
if the positions of the Microswitch S4 Activator Cam and Microswitch S5 Activator
Cam are not modified before and be on the default position.
17. Set Wall Stand mode using OTS console or System console. Drive the OTS console, and make it
vertically over the Bucky of SG100.
4-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
18. Insert a cassette, lower the Bucky to the Low Auto Exposure position, pull out the tray and measure
the distance from OTS to the cassette top (actual SID value) using measure tape, then insert the
tray all the way into the Bucky and adjust P4 to make the display of SID on the OTS is the same
with the actual SID. (If there is no change of SID value displayed on OTS when the Bucky moving
up and down with the tray inserted all the way into the Bucky, adjust the P4 to make the SID value
displayed on the OTS.)
Note: When the tray is pulled out, it is preset to be no SID display on OTS, not a wrong.
To see the SID display on OTS, pay attention to insert the tray all the way into the Bucky
after measured the distance from OTS to the tray.
PA2
100K
P4
100K
FOR SG120 WALL STAND SID CALIBRATION, CONTINUE WITH FOLLOWING STEPS. IF SG120 IS
NOT SELECTED, IT’S ENOUGH TO PERFORM THE STEP 1 TO 14 ONLY FOR SID CALIBRATION.
4-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
21. Modify the Microswitch S4 Activator Cam and Microswitch S5 Activator Cam to decide the distance
between the Low Auto Exposure position and the High Auto Exposure position, write down the value
of the distance (Distance1). (When the Bucky in horizontal, there are only two positions of Bucky on
the column at which the Auto exposure can be made. Of these two positions, the lower one is
defined as the “Low Auto Exposure position”, and the higher position is defined as the “High Auto
Exposure position”. And these two positions can be modified by adjusting the Microswitch S4
Activator Cam and Microswitch S5 Activator Cam. The default distance between these two positions
is 44cm.). (Refer to Job Card IST 008, Chapter2.)
Check and ensure the parameter value in System Console to be the same with the actual distance
between Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low
POS. distance”, Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High
CAUTION Auto Exposure position set in System Console (see parameter 9 “High, low POS.
distance”, Job Card CAL 002) is 40cm, which is different from the default distance
set on SG100 Wall Stand (44cm). So it’s a must to check and change the
parameter value in System Console to ensure it to be the same with the actual
distance between Low Auto Exposure position and High Auto Exposure position
before performing the SG120 SID calibration according to the following steps, even
if the positions of the Microswitch S4 Activator Cam and Microswitch S5 Activator
Cam are not modified before and be on the default position.
4-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
22. Set Wall Stand mode using OTS console or System console. Drive the OTS console, and make it
vertically over the Bucky of SG120.
23. Insert a cassette, lower the Bucky to the Low Auto Exposure position, pull out the tray and measure
the distance from OTS to the cassette top (actual SID value) using measure tape, then insert the
tray all the way into the Bucky and adjust P4 to make the display of SID on the OTS is the same
with the actual SID. (If there is no change of SID value displayed on OTS when the Bucky moving
up and down with the tray inserted all the way into the Bucky, adjust the P4 to make the SID value
displayed on the OTS.)
PA2
100K
P4
100K
Section 2 SID calibration procedure: (For systems with SG100 or SG120 Wall Stand, without Table)
1. Move the SG100 Wall Stand Bucky to the High Auto Exposure position, Rotate the Bucky to 0°
and angulate it to 90°(horizontal).
4-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2. Modify the Microswitch S4 Activator Cam and Microswitch S5 Activator Cam to decide the distance
between the Low Auto Exposure position and the High Auto Exposure position, write down the
value of the distance (Distance1). (When the Bucky in horizontal, there are only two positions of
Bucky on the column at which the Auto exposure can be made. Of these two positions, the lower
one is defined as the “Low Auto Exposure position”, and the higher position is defined as the “High
Auto Exposure position”. And these two positions can be modified by adjusting the Microswitch S4
Activator Cam and Microswitch S5 Activator Cam. The default distance between these two
positions is 44cm.). (Refer to Job Card IST 008, Chapter2.)
Check and ensure the parameter value in System Console to be the same with the actual distance
between Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low
POS. distance”, Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High Auto
CAUTION Exposure position set in System Console (see parameter 9 “High, low POS. distance”,
Job Card CAL 002) is 40cm, which is different from the default distance set on SG100
Wall Stand (44cm). So it’s a must to check and ensure the parameter value in System
Console to be the same with the actual distance between Low Auto Exposure position
and High Auto Exposure position before performing the SG100 SID calibration according
to the following steps, even if the positions of the Microswitch S4 Activator Cam and
Microswitch S5 Activator Cam are not modified and be on the default position in the
upper step.
3. Insert the tray with a cassette all the way into the Bucky.
4. Set Wall Stand mode using OTS console or System console. Drive the OTS console, and make it
vertically over the Bucky of SG100.
4-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5. Verify the SID display on the OTS console is changing while driving the OTS up and down.
Adjust P4 if the SID doesn’t change.
6. Lower the Bucky to the Low Auto Exposure position, drive the OTS to 100cm to the
cassette top (pull out the tray and measure the distance from OTS to the cassette
top using the measure tape on collimator) and write down the SID value shown on
OTS console. Lower the OTS to 70cm to the cassette top (pull out the tray and
measure the distance from OTS to the cassette top using the measure tape on
collimator). Adjust the PA3 so that the OTS console SID changes 30cm.
Repeat step 5 until SID changing is the same as actual moving distance.
Note: When the tray is pulled out, it is preset to be no SID display on OTS, not a wrong.
To see the SID display on OTS, pay attention to insert the tray all the way into the Bucky
after measured the distance from OTS to the tray.
7. Ensure the Bucky to be on the Low Auto Exposure position, pull out the tray and measure the
distance from OTS to the cassette top (actual SID value) using measure tape, then insert the tray
all the way into the Bucky and adjust P4 to make the display of SID on the OTS is the same with
the actual SID. (If there is no change of SID value displayed on OTS when the Bucky moving up
and down with the tray inserted all the way into the Bucky, adjust the P4 to make the SID value
displayed on the OTS.)
8. Use PA2 for auxiliary adjustment.
PA2
100K
P4
100K
PA3
100K
4-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9. Move the SG120 Wall Stand Bucky to the High Auto Exposure position, Rotate the Bucky to 0°
and angulate it to 90°(horizontal).
10. Modify the Microswitch S4 Activator Cam and Microswitch S5 Activator Cam to decide the distance
between the Low Auto Exposure position and the High Auto Exposure position, write down the
value of the distance (Distance1). (When the Bucky in horizontal, there are only two positions of
Bucky on the column at which the Auto exposure can be made. Of these two positions, the lower
one is defined as the “Low Auto Exposure position”, and the higher position is defined as the “High
Auto Exposure position”. And these two positions can be modified by adjusting the Microswitch S4
Activator Cam and Microswitch S5 Activator Cam. The default distance between these two
positions is 44cm.). (Refer to Job Card IST 008, Chapter2.)
Check and ensure the parameter value in System Console to be the same with the actual distance
between Low Auto Exposure position and High Auto Exposure position. (see parameter 9 “High, low
POS. distance”, Job Card CAL 002)
The default value of the distance between Low Auto Exposure position and High Auto
CAUTION Exposure position set in System Console (see parameter 9 “High, low POS. distance”,
Job Card CAL 002) is 40cm, which is different from the default distance set on SG100
Wall Stand (44cm). So it’s a must to check and ensure the parameter value in System
Console to be the same with the actual distance between Low Auto Exposure position
and High Auto Exposure position before performing the SG100 SID calibration according
to the following steps, even if the positions of the Microswitch S4 Activator Cam and
Microswitch S5 Activator Cam are not modified and be on the default position in the
upper step.
4-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
11. Insert the tray with a cassette all the way into the Bucky.
12. Set Wall Stand mode using OTS console or System console. Drive the OTS console, and make it
vertically over the Bucky of SG120.
13. Verify the SID display on the OTS console is changing while driving the OTS up and down.
Adjust P4 if the SID doesn’t change.
14. Lower the Bucky to the Low Auto Exposure position, drive the OTS to 100cm to the cassette top
(pull out the tray and measure the distance from OTS to the cassette top using the measure tape
on collimator) and write down the SID value shown on OTS console. Lower the OTS to 70cm to the
cassette top (pull out the tray and measure the distance from OTS to the cassette top using the
measure tape on collimator). Adjust the PA3 so that the OTS console SID changes 30cm.
Repeat step 5 until SID changing is the same as actual moving distance.
15. Ensure the Bucky to be on the Low Auto Exposure position, pull out the tray and measure the
distance from OTS to the cassette top (actual SID value) using measure tape, then insert the tray
all the way into the Bucky and adjust P4 to make the display of SID on the OTS is the same with
the actual SID. (If there is no change of SID value displayed on OTS when the Bucky moving up
and down with the tray inserted all the way into the Bucky, adjust the P4 to make the SID value
displayed on the OTS.)
4-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PA2
100K
P4
100K
PA3
100K
4-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
R54 10K
10K
P13
R50 10K 100K
P12
R55 10K 100K
R51
P11
100K
P10
P9 100K
100K
P8
100K TP12
P7
100K TP11
P6
100K
TP10
TP9
18. Pull out the cassette, then re-insert it. Make sure the film size display is correct. If not, redo step 16 and
17. Adjust the Volt slightly, then try it again.
Try other size cassettes. If the film size displayed is not correct, redo step 16 and 17. Adjust the Volt
slightly, then try other cassettes.
4-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Wall Stand (GPCP No. 2260354), SG100 Wall Stand (GPCP No. 2225849), and SG120 Wall
Stand (GPCP No. 2402562)
purpose: Calibration and Configuration version no. 1
date: 20/12/99
NF No. Production date Serial No. Personnel:1
from / / from Time: 20 min
to / / to
Wall Stand counterweight has been calibrated at factory. The following procedure is provided in case re-
calibration is needed.
Counterweight Calibration Procedure:
Note: The content with a star is only for the wall stand with a knee spacer.*
4. Check the bucky up and down motion is smooth and operating force is less than 5kg when inserting a 14”x17”
cassette.
5. Mount the counterweight container cover and wall stand cover, fasten all screws.
WARNING Keep the film cabinet and counterweight pieces under control, while you are taking the
calibration procedure. Any unexpected movement may result in body hurts.
4-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: The information in this page is concerted with the revision 6 of SG 100 Wall Stand Service Manual
(GPCP Number 2150297-100). For the reason of the updates and changes in that manual, it is
recommended to see also the latest Service Manual of SG 100 Wall Stand for reference.
220/110v Caution Label
1. Raise the Bucky assembly to its top position.
2. Remove the two lower trim covers (3) screws.
3. Remove the lower (charcoal grey)
counterweight cover (6 screws).
Knurled Screws –
loosen to open
Bucky Front Panel
4. Add or remove counterweights until the Bucky carriage is balanced. Lower Trim Covers
Note: Extra counterweights are furnished as part of the Ion Chamber option.
Iron Counterweights
4-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Metal cover
Metal cover
Counterweights
5) Reinstall the two sheet metal covers and the column lateral covers.
4-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Touch Screen calibration has been performed at the factory. The following procedure is provided in case the
touch screen needs to be recalibarated.
1. Tools
- Proteus System Software Disk
- A PS2 type of computer keyboard
- A communication cable
Note: All systems are shipped with a keyboard cable. If the cable is missing, FE should make a
communication cable according to illustration.
2. Procedure
1. Power off.
2. Demount the printer if have.
3. Use the communication cable connecting the keyboard and the console connector CN1.
4. Insert the system software disk into the console.
5. Power on.
6. At the c:/ prompt type “Tbdriver.exe”
7. Type “Tbcal.exe”; enter the main menu of Touch Screen Calibration Program.
4-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Installers must indicate that the combination of installed HHS Certified Components is compatible on Form
F3382 provided in Direction 46-013894, System Field-Test For HHS.
4-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
4-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
The service software can be used either from the calibration diskette or from the Hard Disk Drive. We recommend
to launch it from the Hard Disk Drive for speed performance.
It can operate under DOS or Windows 95; but for software download or database backup, it is required to
run asian.exe under DOS to avoid communication problem (with the mouse, click on START, SHUTDOWN,
RESTART COMPUTER UNDER DOS MODE; see page 4-40 ”Software Download / Upgrade” Job Card).
• Power ON Service laptop. Under Windows, run the File Manager (Explore)
• Create a specific directory (e.g.: JEDI) under C:\
• Insert into the floppy drive the ”Service Software” floppy delivered with the equipment in the Service Tool set.
• Copy the file SPX_X.EXE from the floppy disk to that directory
(Double click on A:\, drag and drop on C:\ JEDI or whatever name of your directory)
The floppy drive supplied with the equipment (or the upgrade kit) contains “Auto-zipped” files.
It is necessary to load those files on the hard disk drive and run them; they will unzip and create other files that
you will use.
• Double click on the SPX_X.EXE file, and verify that the new files are created:
appl.mx , boot.mx, asian.exe ,Asiane.msg and data base files.mx
4-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The software will load the help screen information contained in Asiane .msg file.
4-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
FROM PROCEDURE TO
APPLICATION MODE Type <ALT> - <S> (Together) SETUP/MAINTENANCE MODE
MENU 3
APPLICATION MODE Type <ALT> - <D> (Together) SETUP/MAINTENANCE MODE
LAST MENU OR SUB-MENU
selected.
SETUP/MAINTENANCE MODE From any menu, type <R> (several APPLICATION MODE
types if necessary) or <Q>
4-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
MENU 1
[F1] [F3] [F5] [F7]
HEAT ANOD HIGH AEC
AUTO AUTO,
3,4,5 3k, 9k
NoLD GATE ShC
CD INV
Tst Tst
MENU 2
[F3] [F5] [F7]
LIST NEXP DATE
MENU 3
[F1] [F5]
TECH MOD
MENU 4
[F5] [F7]
AEC CAL
MENU 5
[F1] [F7]
RAMS Vx Px.x DSPL
[F1] [F3]
CKSM MCL
4-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. Configuration - MENU3
TECH: programming workstation, technique parameters.
MODE: Select Renard or Normal scale for kV selection.
(See Page 4-52 “Workstation Parameter” Job Card ).
2. Calibration - MENU4
AEC: To set parameters if known, or modify them manually.
CAL: To calibrate Automatic Exposure Control, for each AEC chamber and Screen-film combination.
(See page 4-60 Job Card “Automated AEC calibration”)
3. Ram - MENU5
RAMS: NVRAm operation & display:
- CKSM: To update Checksum after configuration, calibration, parameter change. (See page 4-83, Job
Card “Checksum Validation”)
- MCLR: Clear NVRam memory. Normally not to be used, unless specifically instructed. (Do not use if
Data Base backup has not been performed previously)
Vx Px.x: Display generator software version (e.g. P3.4)
DSPL: In Maintenance Mode, to display parameter after exposure using key [Z]
4. Utilities - MENU2
LIST: To display last recorded error codes (up to 64)
- Last error is visualized with Date and Time
- Function keys allow to scroll down and up
- [Z] to get back to the last recorded error
- Same consecutive error are counted Up in the same line
- [^H ], and [W] to clear error log
NEXP: To visualize/clear tube and generator recorded tracking Data. Clear must be performed only when
replacing X-Ray tube. Meaning of each parameter is found in page 4-44, ”Tube Replacement Software
Procedure” Job Card.
DATE: To enter / modify the date and time.
4-38
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Requires manual set up. Read related information in diagnostics section.
- ShC INV Tst: To verify operation of inverter with inverter in short circuit. Test is run during 100 msec.
At 0 kV.
Note: Requires manual set up. Read related information in diagnostics section.
AEC: To verify operation of AEC board. Refer to diagnostics section.
4-39
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
Database backup must have been performed.
4-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The floppy drive supplied with the equipment (or the upgrade kit) contains “Auto-zipped” files.
It is necessary to load those files on the hard disk drive and run them; they will unzip and create other files that
you will use.
• Double click on the SPX_X.EXE file, and verify that the new files are created. (appl.mx and boot.mx)
4-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10. Select menu 2 with the arrows, select LIST, and clear generator error log. (^H, then W)
11. To exit program, press Q, or R to return then ESC, or ESC twice.
4-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
Knowledge of Service Software (refer to page 4-32, ”Service Software User Guide” Job Card).
4-44
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
Refer to page 4-86 Job Card ”Backup & Restore Database Setup Parameters”.
1. Connect Service laptop as follows:
• Use the 9 pin Serial cable, part number 2207478, to
connect serial line of Service laptop to connector XJ8 on Safety Door
the I/F board. and Light
Switch (J7)
2. Restart the computer in DOS mode (for software download
or database backup, it is required to run asian.exe under
DOS to avoid communication problem). Service
Terminal (J8)
• With the mouse, click on START, SHUTDOWN,
RESTART COMPUTER UNDER DOS MODE.
3. Run the program: Inside Comm-
• select Disk Drive from where you want to run the unication (J9)
program: C:\JEDI [enter] or A: [enter]
• key in : “asian” [enter]
4. Save memory
• From the laptop software main screen, type “F6: save
memory”.
4-45
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Before resetting all the memory, it is advised to record them on table ”Generator Counter
Reading Record” (Illustration 1).
4-46
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-47
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-48
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
The service laptop must be configured and software installed (refer to page 4-32 “Service Software User Guide”
Job Card).
4-49
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
Two, four or six database are available on the JEDI Service Software supplied in Service Tools depending of the
configuration:
- one for 32 kW tube: type RAD 12,(Low speed) File = F32RAD12.mx
- one for 32 kW tube: type RAD 14,(Low speed) File = F32RAD14.mx
- one for 50 kW tube: type RAD 14,(High speed) File = F50RAD14.mx
- one for 65 kW tube: type RAD 14,(High speed) File = F65RAD14.mx
- one for 65 kW tube: type RAD 21,(High speed) File = F65RAD21.mx
- one for 80 kW tube: type RAD 21,(High speed) File = F80RAD21.mx
Four database for the system equipped with tube fan:
- one for 65 kW tube: type RAD 21,(High speed) File = F65RA21F.mx
- one for 80 kW tube: type RAD 21,(High speed) File = F80RA21F.mx
- one for 65 kW tube: type MX 100,(High speed) File = F65KOUSB.mx
- one for 80 kW tube: type MX 100,(High speed) File = F80KOUSB.mx
Load from Service laptop the file corresponding to the tube present in the installation and the customer
configuration Refer to Job Card “Back-up and Restore Database Setup Parameters” CAL Jedi 8-6.
Note: When replacing the database file of RAD12, 14, 21 with database file of MX100 or in reverse,
pay attention to replace the Rotation Capacitor Set (refer to Job Card D/R JEDI 17-38 in
Section 3 of Chapter 9) and ensure its correct match with the X-ray tube type.
5-2 Verify
4-50
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-51
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
The JEDI is preprogrammed. Therefore, even if the program seems to correspond to the room parameters, it
must be double checked, for example, for AEC enable, wall Bucky enable.
4 PREREQUISITES
Service laptop must be configured and software installed (refer to page 4-32 “Service Software User Guide” Job
Card).
4-52
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
5-1 Introduction
Workstation programming uses function keys indicated on screen for the following :
• workstation select,
• tube select,
• define nominal number of exposures for 2-point mode operation,
• AEC : types of chamber and number of pickup fields in each,
• Film-screen combination select, (if only one Select A).
<-
R: Q:
TECH<tok_Tab>:<tok_Tab>Se
MODE<tok_Tab>:<tok
4-53
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Any new parameters created must be stored in the installation report (same as Illustration 3).
1. Workstation Selection
OFF No selection
SPE * Anatomic (for RF room)
DIR* Anatomic (for RAD room)
POMU Wall-mounted Bucky (i.e. SG50 or SG70)
SETE Spot-film device technique on remote-controlled table (i.e. Prestilix 1600)
TOPO Tomography Bucky
TOSE Tomography spot-film device technique
POTA Table-mounted Bucky
PULM Pulmorapid
CHAN Film changer
TOCO Special tomographic applications
SECO Spot-film device technique on conventional table
NUME Digital mode
TONU Digital tomography
Note: For systems equipped with no Table configuration, TECH1 should be set DIR, TECH2 and
TECH4 should be set OFF.
4-54
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Nominal
Workstation Tube Tomo tech & time number of Type of AEC
exposures In
2-point mode
2. Tube Selection
T1 Tube 1 only
3. Select the number of nominal exposures for 2-point mode operation.
DIR N = 2 POTA N = 2
POMU N = 2
Select AEC in or out using function keys [F7] et [F8]. Each chamber corresponds to a specific technique, i.e.
POMU and POTA.
4-55
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Function keys are assigned using the alphanumeric keyboard ([1] corresponds to function key 1, etc., up to 8).
Use the shift key for numbers 0 to 9.
Keys [A], [B] and [C] are used to select the first, second and third screen-film combinations, respectively.
One combination is always selected. Select Key A for the RAD used screen film combination.
7. Selection of Number of Pickup Fields
The number of pickup fields of the exposure chamber selected for AEC Mode with ionizing chamber or II must
be keyed in:
Use [W] to validate all values for a given function appearing on the screen.
1. Go to Service Mode, and call Menu 3 using the arrows on the keyboard.
2. Call MODE by pressing F5.
3. Select NORMAL or RENARD Scale by pressing F5.
4. Press W to confirm.
4-56
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-57
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
WORKSTATION SELECT
FUNCTION 1 2 3 4 5 6 7 8
WORKSTATION DIR POTA POMU TOPO OFF OFF OFF OFF
TUBE T1 T1 T1 TI
NUMBER OF
EXPOSURES IN 2 2 2 2
2-POINT MODE
TOMOGRAPHIC
UNIT USED:
Y1 T1
Y2 T2 N N N N
Y3 T3
Y4 T4
Y5 T5
TYPE
OF OUT IN1 IN2 IN1
CHAMBER
mA PLATEAU
TIME FOR - - - - -
1-POINT MODE
EXTERNAL -
WORKSTATION NO NO NO NO
COMMAND*
SCREEN-FILM
COMBINATIONS
1
A A A A
2
3
NUMBER OF
PICK-UP - 3 3 3 Number
FIELDS of cells
used
in AEC
MODE RENARD OR NORMAL kV SELECTION
ERGO FULL OR MINI ERGONOMY
Illustration3 - Workstation Parameters (Example for RAD with Wall Bucky and two Ion-chambers)
4-58
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
None.
3 SAFETY PRECAUTIONS
4 PROCEDURE
1. Power on the system.
2. Close collimator blades or block x-ray output.
3. Take 2 exposures (30 seconds apart) with the following technique Parameters:
• Table top receptor
• Large focal spot
• 70 kV
• 200 mA at 1 sec
4. Once exposures are taken the system is ready for use.
4-59
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-60
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PURPOSE
Application mode
Service mode
NO YES
MENU4/AEC/F7 MENU4/AEC/F5
7-4 Automatic: 7-6 Manual parameters
Fast calibration entry
Select a ionization
chamber
7-8Lateral check
YES
Other ionization chamber?
END
Illustration - OVERVIEW
4-61
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
JOB CARD CAL JEDI 8-2 - Automated AEC Calibration has three purposes:
• Fast AEC film calibration menu (MENU4/AEC/F7)
The main purpose of this job card is to perform a quick AEC film calibration.
This job card explains in particular how to use the “Fast calibration” Service menu of JEDI software.
• Manual entering AEC parameters menu (MENU4/AEC/F5)
For already known screen-film combinations and/or ionization chambers, the corresponding parameters, (KVEF,
KTIME, KF) and/or (KG), could be entered directly without performing the Fast Calibration.
• AEC response check
Check the good response of ionization chambers and measuring circuitry.
6 OPERATING CONDITIONS
base fog
4-62
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6-2-2 Incidence:
Orthogonal incidence.
Lateral
Phantom centered over the ionization central measurement cell.
6-2-4 Collimation
4-63
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7 PROCEDURE
The goal of this section is to identify the different film/screen combinations, and classify them by sensitivity: Slow,
Medium and Fast.
• Ask the customer which combinations of Screen / Film he is using, and their relative sensitivities: which one is
the fastest, which one is the slowest.
• Limit to 3 the number of combinations to be exposed with the Automatic Exposure Control.
Warn the customer to operate without AEC for the other combinations.
• Fill the following table:
4-64
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SENSITOMETRIC CURVE
3.6 3.6
3.4 3.4
3.2 3.2
3.0 3.0
2.8 2.8
2.6 2.6
2.4 2.4
2.2 2.2
2.0 2.0
1.8 1.8
1.6 1.6
1.4 1.4
1.2 1.2
1.0 1.0
0.8 0.8
0.6 0.6
0.4 0.4
0.2 0.2
0 0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
4-65
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-66
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-2-2 Preselections:
Enter Service mod MENU4/AEC/F7
• Screen-film
YES
7-4-4 Run KTIME calibration (if needed)
Run KTIME calibration (if needed) For each of the 4 time value
For each of the 4 time values
NO
All exposure times performed?
NO
All exposure times
YES
YES
7-4-5
YES
YES
Other screen-film? Select screen-film
(2 more possible B, C) combination
NO
7-4-6-1
7-4-6-3
Other ionization chamber? YES
(3 more possible) Select ionization chamber
7-4-6-4
NO
Run KG calibration
END
4-67
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• If an error occurs
- To redo an exposure after an exposure, reset by pressing Z key.
- To redo an exposure after entering the optical density, enter “0” for optical density and redo the
exposure.
- To clear optical density, press CTRL + H keys and enter the new optical density.
• Loops
- In case of one ionization chamber:
Run the KVEF and KTIME calibration for each film screen combination for the ionization
chamber.
- In case of multiple ionization chambers:
Run the KVEF and KTIME calibration for each film screen combination for one ionization
chamber.
Run only the KG calibration for the other ionization chambers using the same screen film
combination.
7-4-1 Preliminary for KTIME
N 0 1 2 3 4 5
7-4-2 Initialization
4-68
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-2-2 Preselection
• A, B or C key to select screen film type (slow, medium, fast) (if only one film-screen, select A key) The
selected film-screen key flickers.
• Run the first exposure proposed by the software with the 1st first value (80kV) of the kV series
• Run the next exposure with the new parameters calculated by the software
• Validate the value with the W key - Redo the next exposure and enter the density until the V button
flicker
• When the V button flickers, record in table para. 7–5 the mAs0 displayed and press V key
• Select the next kV step in the kV series (80kV, 100 kV, max kV, 40 kV and 60 kV). and repeat
operation
• Start the KT time calibration by pressing V key after V button lighted at the end of the first screen-film
KVEF calibration
4-69
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• Enter the associated read optical density. Redo the exposure for other exposure time, 2s, 3.2s, 10ms
• Validate the value with the W key, after entered density value
• Select the next screen/film combination if existing and start again the KVef + KTime setting procedure
from para 7–4–1
• Press the V key again to enter the KF calibration (bypass the KF calibration, if not required by hitting V
key again)
4-70
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-4-2 Warning:
In case of film too dark, after calibration, for exposure time below 10ms, you must insert an additional step
between 0 and 10ms and remove the step at 2s.
How to proceed:
Write down manually the Kt values you obtained during automatic calibration for the step 10ms, 0.1s, 0.5s, 1s, 2s
and 3.2s in table below (see procedure to read the value in para 7.6).
1/ Use the procedure in para 7.6.1.2 to change manually the T0 to T5 time values according to table below.
2/ enter in the table below the values already known (from auto calibration, table above). Note that the 2s value is
removed.
3/ Enter for T0=2ms a Kt new value which is lower than the one for T1=10ms.
4/ Run a manual exposure in application mode with an exposure time between 2ms and 10ms (ex: 120kV 10cm
of plexiglass or water).
5/ Check the film darkness and adjust the Ktnew value until the darkness result is good.
4-71
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
This procedure has to be run for only 1 ionization chamber, only for one screen / film combination only in case of
Automatic collimation.
The phantom must cover the entire field for 36x43 format cassette.
• Format coding
X-ray field depends on the cassette size and the eventual division.
It is sensed and coded by the cassette tray and sent to JEDI.
Table top
Cassette l = lateral
L=
4-72
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-5-1 Procedure
4-73
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-6-1 Applicability
This procedure has to be performed for 2nd ionization chamber, if it is present., and with only one screen / film
combination (the slowest for example).
In case of only one ionization chamber: Press R key to by-pass KG calibration and go to section 7–4–6–5.
7-4-6-2 Procedure
7-4-6-3 STEP 1
7-4-6-4 STEP 2
4-74
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-4-6-5 STEP 3
4-75
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
mAs0 records at the end of each kV step of KVEF calibration. Fill the table below.
mAs0 reporting table:
mAs0 AEC1
Sensitivity 40 kV 60 kV 80 kV 100 kV Max kV
A
B
C
mAs0 AEC2
Sensitivity 40 kV 60 kV 80 kV 100 kV Max kV
A
B
C
Access to the AEC parameters, select the Service Mode, MENU4 / AEC / F5 and fill the table below.
AEC parameters reporting table:
4-76
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
KG [F3] key
Image receptor #1 #2
Value
To continue the Automatic Fast Calibration, go to section 7–7, and then 7–8.
4-77
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-6-1 Procedure
4-78
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-6-1-3 Set KG
4-79
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-6-1-4 Set KF
Before exiting service mode, the checksum must be validated after settings have been completed; refer to
Calibration Job Card “Checksum Validation” CAL JEDI 8-5.
7-8-1 Goal
7-8-2 Principle
7-8-4 Procedure
• Plexiglass thickness = 10 cm
• Select one image receptor
• X-ray field should be large enough to cover the 3 cells
• Application mode
• AEC on
• density correction = 0
• film screen /film key = fastest
• 3 point mode
4-80
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• kV = 60
• mA = 100
• mAs = 32
• Select the cell(s) and make exposure according the reporting table
• Compute the mAs variation
• Repeat the procedure for each image receptor equipped with ionization chamber.
7-8-5 If failure
• Replace the ionization chamber and restart the complete calibration process this ionization chamber.
4-81
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-82
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
- Service laptop with Floppy Drive
- Service tools Part
- JEDI service software
- 9-wire serial cable Part No. 2207478.
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
Service laptop must be configured and software installed (refer to ”Service Software User Guide” Job Card).
4-83
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
In the checksum validation process, the processor computes a number which is a function of all the setup
parameters. When the generator is powered up, this number is computed, i.e. the sum is ”checked”.
If the two numbers differ, the generator displays error an code, indicating that one or more parameters have
been changed since the last time the sums were checked, or that a RAM-related incident has occurred.
Note: The display Va . b on menu 5 is for the software version in the CPU board EPROMS.
4-84
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Resulting display :
MCLR is used to erase all setup parameters. Function key [F3] corresponding to this
NOTICE function, should never be used.
4-85
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
- Service laptop
- Service tools
- JEDI service software
- 9-wire serial cable Part No. 2207478.
3 SAFETY PRECAUTIONS
Not applicable.
4 PREREQUISITES
1. The service laptop must be configured and the software installed (refer to ”Service Software User Guide”
Job Card)
2. All calibrations must have been completed.
4-86
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 APPLICABILITY
Setup parameters must be logged after the setup procedure has been completed, or whenever theses
parameters are changed. This enable service engineers to key in parameters rapidly, i.e. when changing the KV
Control Board, in the event of a battery outage.
6 PROCEDURE
The generator must be in application mode. The control console push buttons will be inactive during the transfer.
- Connect service laptop to XJ8 on the I/F Board.
- Power up.
- Insert the calibration floppy disk without write protection.
- From Laptop Dos, A: (It can also be run from Service Laptop Hard Disk Drive but in DOS mode).
- Type JEDI on the terminal.
- Hit [F6] to save memory.
The system ask for a file name. Default file name is A:\sv_var.mx.
- Type a file name or hit [ENTER].
During transfer, the number of sent trames is displayed.
When transfer ends, the message “Transfer successful” is displayed.
If any error is detected, the transfer stops and an error message is displayed.
See next page for list of error messages.
4-87
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
All parameters stored in CPU NVRAM memory have been saved on calibration floppy disk. Therefore, in case of
change of KV Control board you can initialize CPU NVRAM by transferring data from this floppy disk.
The procedure needs to Unplug the cable from system I/F board Connector J118. Then hit [F7] instead of [F6]
in service mode to restore memory.
This procedure is also used to down load new software (File type for Database or Software is automatically
recognized).
Note: Error during memory restore may cause unpredictable result. A new transfer is required.
If generator is locked, power OFF and begin again the entire memory restore procedure.
MESSAGE ERROR
4-88
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3 Select the figure you want to change, and press the Up or Down arrow button to change it.
4 Press OK or CANCEL to save or not save the new configuration and return to the upper interface.
4-89
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Light field – X ray field alignment (For Proteus XR/a Collimator Only)
purpose: Calibration version no. 1
date: 18/4/2001
NF No. Production date Serial No. Personnel:1
from / / from Time: 0h 10min
to / / to
4-90
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Introduction
The collimator had been calibrated in the factory before the shipment. The situation in shipment maybe will result
in some small variation. Therefore, if necessary, the collimator should be realigned.
Note: After adjustment of the light field, it may be necessary to readjust the line laser.
Caution: The line laser is already adjusted at the factory. Only under the specific situation that the
deviation between the light field center and line laser could not be limited within 3mm through
adjusting the light field should the line laser be readjusted at the install sites.
For the line laser readjustments, please refer to the procedures below.
• Line laser readjustment.
1) Open the collimator front panel with the 2 screws on the right side (2.5 mm allen wrench.)
Note: Observe ESD guidelines!
2) Adjust the laser line with the two screws (2.5 mm Allen wrench) (Illustration 3 resp.
Illustration 4) to the middle of the cross hair. The laser line must not differ for more than +/-
3mm from the middle of the crosshair.
4-91
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Line Laser
1
1
Line Laser
3) Close the front panel and fix it with 2 screws (2.5 mm Allen wrench).
4-92
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Procedure
2b
1a
1b
2a
WARNING
DO NOT TOUCH THE BULB SOCKET CONTACTS TO AVOID SHORT CIRCUIT. TIGHTEN
THE SCREW 2B WHILE THE BULB IS SWITCHED OFF TO AVOID SHORT CIRCUIT.
4-93
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Calibrate Proteus XR/a Manual Collimator Bulb and its Voltage
purpose: Calibration version no. 0
date: 12/12/2005
NF No. Production date Serial No. Personnel:1
from / / from Time: 0h 10min
to / / to
Introduction
Lamp voltage on Manual collimator is provided by OTS transformer, The main power situation on site maybe
result in improper voltage to Lamp on Collimator. The appropriate tap should be seleted on XT control board by
measuring voltage.
Tools
RMS Multimeter
Procedure
measure voltage between ground and J8 , J9, J10, J11, on XT control board, select the appropriate tap for
voltage is between 22.5VAC and 24VAC.
4-94
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: If you have chosen SG120 Wall Stand as an option, please refer to the following contents.
Even if the calibration of the vertical locks is performed when the equipment leaves the factory, they can be re-
calibrated when needed.
Procedure
1. Make sure the equipment is switched OFF.
2. To access the vertical lock assembly, remove the lateral and frontal covers from the column assembly.
3. The vertical lock assembly is located between the vertical carriage and the column stand and is
accessible from the bottom behind the bucky support assembly.
4. Set the bucky carriage at the appropriate height and adjust the brakes actuating on the 2 (1 for each
brake) adjustment screws marked on the right of the figures above.
1) Rotate the screws CCW (counter-clockwise) direction to approximate the brake to the brake
surface (column stand).
2) Rotate the screw CW (clockwise) direction to separate the brake from the brake surface.
Note: A distance of 0.5mm between the brake surface and the brake can be set as a start point
for the calibration.
Note: The calibration has to be made considering there is no friction between the brakes and the
brake surface (column stand) and the noise factor.
Check there is no collision between the brakes and the brake surface and the bucky
carriage can be displaced smoothly all along the vertical whole track. Just a slight friction
between the brakes and the brake surface at any point along the vertical track could be
accepted in order to minimize the noise of brake actuation.
5. Assembly the covers removed in step 2.
4-95
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-96
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 1
INTRODUCTION
Functional Checks describes the procedures necessary to determine whether the
Proteus XR/a system (Refer to Illustration 5-1) is operating correctly and
performing within specifications.
All the Proteus XR/a system functions should be conform to safety regulations,
local codes or other applicable standards.
• Operational Checks.
• Performance Checks.
Illustration 5-1
PROTEUS XR/a SYSTEM
5-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
TOOLS AND TEST EQUIPMENT
In addition to the standard serviceman tool kit the following tools and test
equipment will be necessary to perform the functional checks described in this
chapter:
• DVM
• Metric Tape Measure
• KEITHLEY 35080A kV peak meter (or equivalent)
• RADCAL MDH2025 dosimeter (or equivalent) with 20x5-60 probe
SECTION 3
SYSTEM POWER ON
Note: Before this step, all electrical cable must be connected and examined
Power on the Proteus XR/a system, verify that system console’s booting
correctly.
5-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 4
TABLE
4-1 Operational Check
When the assembly and electrical connections have been completed, perform
the following tests to verify the table's travel and vertical operations. Tests for the
table's elevating function is required only for elevating table.
Depressing the bucky switch should release the bucky magnet, and the bucky
should move freely in the longitudinal.
If the table can’t move when double click and depress the Up or Down foot pedal,
please check the switch backside in the sensor.
There should
leave a gap
Switch backside
Sensor
Adjust each switch backside; insert it into the notch of the sensor. Make sure the
switch backside won’t touch the sensor when release the foot pedal. There
should leave a gap between the switch backside and the sensor.
5-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
800+/-10mm
max
500mm
min
If a modification in the height limits (lower and upper) is required to meet these
specifications, it is necessary to adjust the Up/Down Limit Board located at the
front of the base.
5-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 5
WALL STAND, SG100 WALL STAND, AND SG120 WALL STAND
5-1 Wall Stand (GPCP No. 2260354)
1. Depress the handle button, Wall Bucky should be able to move up and down
freely and locks in place when released.
2. Check the bucky up and down motion is smooth and operating force is less
than 5kg when inserting a 14”x17” cassette. If not, go to the Job Card CAL 008-
Wall Stand (GPCP No. 2260354), SG100 Wall Stand (GPCP No. 2225849), and
SG120 Wall Stand (GPCP No. 2402562) counterweight calibration.
Illustration 5-1
Note: The information in this section is concerted with the revision 6 of SG100 Wall Stand Service
Manual (GPCP Number 2150297-100). For the reason of the update and change in that manual, it is
recommended to see also the latest Service Manual of SG 100 Wall Stand for reference.
With power on: using the vertical lock handle, release the mechanical and
electro-mechanical vertical locks, then verify that the carriage can move freely
and smoothly over its entire range of travel with minimum side play.
If the carriage remains locked when the mechanical lock is released, check the
lock board feeding the electro-mechanical lock (Refer to Form 1).
5-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
If carriage motion is not smooth and free, ensure that counterweights and trim
covers are installed correctly. Add or remove counterweights until
counterbalance is achieved. (Refer to Job card CAL 008, Wall Stand (GPCP No.
2260354), SG100 Wall Stand (GPCP No. 2225849), and SG120 Wall Stand
(GPCP No.: 2402562) in Chapter 4.)
Note: The carriage is counterbalanced when leaving the factory, however, if necessary,
Apply further counterweights to achieve the counterbalance (Refer to Job Card
CAL 008, Chapter 4).
Release the lock handle partially to engage the electro-mechanical locks. Verify
that the electro-mechanical locks engage before the mechanical lock
(approximately 1 mm before the mechanical lock engages). If the locks do not
engage, check the lock board. (Refer to Form1).
Release the lock handle all the way and verify that the mechanical lock’s teeth
fully engage. If not, check for misalignment, wear etc., and perform the
appropriate corrective action.
Push one of the angulation lock pushbuttons on either side of the Bucky support
(and slightly rock the Bucky assembly if necessary) to release the angulation lock.
Verify that the angulation lock releases and that the Bucky can tilt forward 20o and
backward 90o freely and smoothly.
Illustration 5-2
Bucky Vertical (0o) Position Bucky Horizontal (90o) Position
If angulation motion is not satisfactory, check the condition of the bearings at the
Bucky pivot point in the Bucky support assembly.
If the angulation lock does not release, check the lock board. (Refer to Form 1).
Release the angulation lock pushbutton and verify that the angulation lock
engages, preventing further angulation of the Bucky. If the lock does not hold or it
binds, check the lock hardware for misalignment, wear etc., and perform the
appropriate corrective action.
5-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Vertical Bucky Stands not equipped with an automatic collimator do not have
microswitches S1 and S3.
If the lock mechanism slips, it allows the bucky to rotate. Use the following
procedure to adjust the arm assembly. If the brake pad is too worn and the lock
5-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
mechanism still moves after making several adjustment attempts, the whole arm
assembly must be replaced.
The adjustment of the rotation lock mechanism is performed by adjusting the two
threaded rods located inside the rotation lock system. These two rods push
against the lock cam enabling the desired adjustment.
Remove the two screws that hold the lock arm assembly using a 2.5 mm Allen
key.
Illustration 5-5
DO NOT ROTATE ARM ASSEMBLY. REMOVE THE BRAKE PAD CAREFULLY AS IT CAN
WARNING FALL INTO THE SG RATATION BOX AND WOULD BE IMPOSSIBLE TO RECOVER.
Illustration 5-6
5-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3. Adjust the threaded rods so that there is only enough movement between the brake pad and the arm to be able
to lock the brake assembly. Use a 1.5 mm Allen key.
Illustration 5-7
4. Re-insert the brake pad and arm assembly in position and check to ensure there is no rotation movement. This
adjustment may have to be done several times until there is no movement.
5. Once adjusted, apply loctite to threaded rods and re-insert brake pad and arm assembly.
Illustration 5-8
5-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: Vertical Bucky Stands not equipped with an automatic collimator do not have microswitches
S2-1 and S2-2.
Illustration 5-9
Raise the Bucky and verify that microswitch S4 is actuated when the Bucky reaches the near vertical SID point
(normally the point at which the Bucky front panel surface is at the same height as the table).
Note: Vertical Bucky Stands not equipped with an automatic collimator do not have microswitches S4
and S5.
5-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 5-10
5-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: The carriage is counterbalanced on leaving the factory, however, if necessary add further
counterweights to achieve the counterbalance.
Illustration 5-11
5-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: To realize a Right hand configuration and correct rotation way, please refer to IST009 of
Bucky Reversal Procedure.
5-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Rotate the Bucky to 0 degrees and angulate it to 90 degrees (horizontal). Verify that micro switch S5 is actuated
when the Bucky is at the far vertical SID point.
Raise the Bucky and verify that micro switch S4 is actuated when the Bucky reaches the near vertical SID point
(normally the point at which the Bucky front panel surface is at the same height as the table).
Check the control board only in the case of failure of the electromechanical locks. See Table 1 below for checks.
TABLE 1
LOCK BOARD FUNCTIONAL CHECKS
LOCK FAILURE CHECK
Locks do not engage when 24 VDC in D3 or Q2
equipment is switched on
Vertical Locks remain engaged when Handle switch
operating lock handle
Locks do not engage when Handle switch
releasing lock handle
Angulation Locks remain engaged when 24 VDC in D3 or Q1/Q3
operating lock handle
5-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 6
CASSETTE TRAY
1. Confirm the pull-out and pull –in movement is smooth.
2. Pull out the cassette tray, insert a film cassette, then fasten the cassette lock.
Check the cassette size displayed on OTS console and collimator is correct.
3. If not, see Chapter 3-“cassette size calibration”.
SECTION 7
OTS SUSPENSION
As part of the functional check of the Proteus XR/a system, also perform the
functional checks for the OTS Suspension as found in the OTS Suspension
Service Manual 2277285-100 chapter 3 Functional Checks.
5-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 8
JEDI GENERATOR
PROTEUS XR/a (JEDI 50Plus R 1T & JEDI 80R1T) 1/12 CHK JEDI 38-2
1 SUPPLIES
Not applicable.
2 TOOLS
Service tool.
KEITHLEY 35080A kV peak meter.
RADCAL MDH 2025 dosimeter with 20x5-60 probe.
3 SAFETY PRECAUTIONS
4 PREREQUISITES
5-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The precision of setting of all parameters (kVp, mA, exposure times, mAs) is checked, at variable settings of kVp,
first in Three Point mode and then in Two Point mode.
Note: kVp values are measured using kV peak meter Keithley 35080A.
Values for mA, exposure times, and mAs are displayed on the CONSOLE after each exposure.
5-2-1 Summary
Test exposures are made at specified settings, first in Three Point mode, then in Two Point mode, (without the
use of AEC, Automatic exposure Control). The values of parameters read during the exposure are noted and
compared with specified limits.
In each mode, exposures are made first with the small focal spot, then with the large focal spot, at kVp values of
60, 80, and 120 kVp.
In Three Point mode, each exposure is made at 20 mAs, and exposures are made with different values of mA and
exposure time at each kVp value.
In Two Point mode, exposures are made with different values of mAs at each kVp value.
The parameters to be set for each exposure are specified in Illustrations 1, 2, 3, and 4 of this document.
5-2-2 Procedure
Refer to Illustrations 1, 2, 3, 4.
Make a series of exposures corresponding to the values specified in the tables, and note the actual parameters
displayed for each exposure in the spaces provided.
For each parameter and each exposure, the tables specify acceptable high and low values.
CAUTION The measurements of mA, mAs and ms must be made using the post display
of the Service laptop. For kVp values, use the Keithley 35080A.
5-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5-3-1 Summary
Reproducibility and linearity are checked by making a series of exposures, measuring the actual dose produced in
each case. The results are used to calculate coefficients of variation (C) and linearity (L).
Record values in Illustration 5.
Measurements are made using a RADCAL MDH 2025 dosimeter, with the probe (20x5-60) positioned at 100 cm
(39 inches) from the focal spot.
CAUTION The measurements of mAs must be made using the post display of the Console
Note: Take care not to saturate the ionization cell. The maximum output dose permitted with the 20x5-
60 probe is 3.3 mR/ms.
3. For each exposure, note the measured dose (mR), and the mAs value shown on the POST DISPLAY in the
space provided in Illustration 5.
Note: Take care not to saturate the ionization cell. The maximum output dose permitted with the 20x5-
60 probe is 3.3 mR/ms.
3. For each exposure, note the measured dose (mR), and the mAs value shown on the POST DISPLAY in the
space provided in Illustration 5.
5-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 n
(Ri − Rm ) 2
C=
Rm
∑
i =1 n −1
where:
Ri = dose measured during exposure i
where:
Rm1 = average dose measured over ten successive exposures taken
under the same conditions (in particular with the same value of mA,
here 100 mA), in less than an hour
A1 = the mAs value used during the exposures used to measure Rm1
(here 20 mAs)
A2 = the mAs value used during the exposures used to measure Rm2
(here 25 mAs)
Make the calculation and enter it into Illustration 5.
5-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6-1 Reproducibility
Reproducibility is checked in Single Point, Two Point, and Three Point modes of operation. In each case, ten
exposures are made and the doses are recorded. A coefficient of variation is calculated, as defined in section 5–
4.
Note: Take care not to saturate the ionization cell. The maximum output dose permitted with the 20x5-60
probe is 3.3 mR/ms.
Repeat the complete procedure specified above for Two Point mode, but use Three Point mode at 70 kVp, 250
mA, and 1000 ms.
5-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
This procedure checks the limitation on maximum mAs. When kVp is equal to or greater than 50 kVp, the mAs
value should not exceed 600 mAs.
1. Select a workstation which uses AEC.
7. Press the key on Service laptop to display the exposure parameters; check that the value for mAs is 600
mAs or less.
Pass Fail
This procedure checks that the minimum exposure time is less than 16 ms.
1. Select a workstation which uses AEC.
- Press the key on Service laptop to display the exposure parameters; check that the value for
exposure time is less than 16 ms.
Pass Fail
5-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1 - JEDI 50Plus R 1T Accuracy - Small focal spot - Radiography - 3 points - Tube:
20
--- --- --- --- --- --- --- --- ---
24.12 1055 24.12 1055 24.12 1055
50
--- --- --- --- --- --- --- --- ---
56.44 423 56.44 423 56.44 423
100
--- --- --- --- --- --- --- --- ---
111.86 211.82 111.86 211.82 111.86 211.82
*160
--- --- --- --- --- --- --- --- ---
179.36 132.69 179.36 132.69 179.36 132.69
2- mA, time and mAs measurements must be made using the post display of Console.
5-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 2 - JEDI 50Plus R 1T Accuracy - Large focal spot - Radiography - 3 points - Tube:
20
24.12 1055 24.12 1055 24.12 1055
50
56.44 423 56.44 423 56.44 423
100
111.86 211.82 111.86 211.82 111.86 211.82
200
222.93 106.31 222.93 106.31 222.93 106.31
* 320
For 50/65/80 356.56 68.34 356.56 68.34 356.56 68.34
kW only
5-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 3 - JEDI 50Plus R 1T Accuracy Small Focal spot - Radiography 2 points - Tube:
0.5
20
46 46 46
50
54 54 54
5-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 4 - JEDI 50Plus R 1T Accuracy Large focal spot - Radiography 2 points - tube:
0.5
0.74 0.74 0.74
20
21.72 21.72 21.72
80
86.28 86.28 86.28
5-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
70 kVp - 200 ms
10
Rm
Xm
C
(< 0.05)
L
(< 0.1)
5-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1
2
3
4
5
6
7
8
9
10
Rm
C
(< 0.05)
5-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
Not applicable.
2 TOOLS
3 SAFETY PRECAUTIONS
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
5-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
Inverter
Unit
LEFT
ACCESS RIGHT
ACCESS
1
1
5-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
46
sh3 sh4
10 ohms 10 ohms To KV Control Board
1% 1%
3 45
2 1
KV- measure
HV- 1
Measurement tool
(multimeter or
mA= V/(sh3//sh4)= V/5ohms
2. Put a probe inside the generator on the shunt SH3 (right side of the resistor) of KV Measure board on HV
Tank 1 .
1
Take great care not to let screws fall into
NOTICE the HV connector receptacles. Cover
them with plastic caps.
5-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
46
sh3 sh4
10 ohms 10 ohms To KV Control Board
1% 1%
3 45
KV- measure
HV- 1
Measurement tool
(multimeter or
mA= V/(sh3//sh4)= V/5ohms
2. Put a probe inside the generator on the shunt SH3 (right side of the resistor) of KV Measure board on HV
Tank 2 .
3. Connect a second probe on the mechanical ground 3 .
4. Check the resistor value with a multimeter: it should be 5 ohms (+/- 0.05).
6 mA MEASUREMENT
1. Power on the generator
2. Use a multimeter for mA <= 50 mA with a exposure time set to 1s minimum.
3. Use an oscilloscope for other mA, the signal must be filtered at f <= 10 kHz.
5-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 1
THEORY OF OPERATION
1-1 GLOSSARY
Glossary of terms used in this document:
Term Definition
ABC Automatic Brightness Control. Regulation loop which makes the measured
brightness equal to brightness demand
AEC Automatic Exposure Control. Exposure cut off technique which uses the
brightness signal to cut the exposure
CAN Controller Area Network. A network used for localized control.
CPU Control Processor Unit. Microprocessor and peripherals which run the
software/firmware
EPLD Erasable Programmable Logic Device.
EMC Electro Magnetic Compatibility. The EMC function prevents the generator
from polluting the power source.
FPGA Field Programmable Gate Array. It is programmed by the CPU core after
the reset and handles all the exposure control logic including the system
interface real–time lines.
HV Ripple High voltage variations due to inverter current pulses. Typically a few
percent.
State Machine Software or hardware function which handles the state of a system and
authorize to go to the next state upon reception of specific events.
IGBT Insulated gate bipolar transistor. A type of power switch
Ilp HV power inverter parallel resonant current; current in the parallel inductor
Ilr HV power inverter serial resonant current; current in the serial inductor.
6-1
1-2
REV 19
JEDI GENERATOR / FUNCTIONAL ARCHITECTURE
Illustration 1 -
System Control Bus
System Interface
GE MEDICAL SYSTEMS
Low Voltage
heater supply
kV Control Heater Power Rotation
bus
Supply
DC Bus
JEDI HIGH LEVEL BLOCK DIAGRAM
AC/DC
6-2
Jedi Generator / Functional Architecture
rotation phases
X*Ray Tube 1
1 phase
3 phase power
tube cooling
input
input
DIRECTION 2273022-100
Service Manual
PROTEUS XR/a
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
THEORY PRESENTATION
2-1 INTRODUCTION
Jedi is the engineering name for a family of compact high frequency X-Ray generators. This generator family
covers a wide range of applications from mobile equipment up to vascular systems:
• JEDI 12-25 kW: Mobile applications
• JEDI 24-48 kW: CT applications
• JEDI 32-50 kW: RAD applications
• JEDI 50-65-80 kW: RF applications
• JEDI 100 kW: VASCULAR applications
Jedi is a family of 150 kV generators operating from 12 kW up to 100 kW for all the major radiological,
fluoroscopic and CT applications. The family handles 1 ms to continuous exposures with tube currents ranging
from 0mA up to 1000 mA.
The generators feature the very latest technology available:
• Constant potential independent of line voltage variations
• Power generation by a high-frequency converter (High voltage ripple: 40 kHz-140 kHz)
• Distributed micro-processor controlled functions (CAN bus)
Other features include:
• Single phase, three phase or battery power source
• Very low kV and mA ripple, excellent accuracies and dose reproducibility
• Compatible with a wide range of tubes, high speed or low speed, can supply up to 3 different tubes. Thermal
load interactive integrator ensuring optimum use of the heat protection curve of the x-ray tube
• Available in various packaging configurations: gantry, under-table, cabinet
• Serviceability: high reliability, fast installation (no generator calibration), application error codes ensure fast
troubleshooting
• Meets CE marking (and in particular EMC), IEC, UL, CSA, MHW regulations (if required)
• Optional pulsed fluoroscopy
6-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-3 APPLICATIONS
Legend:
- AEC - Automatic Exposure Control
- AET - Automatic Tomographic Exposure
- ABC - Automatic Brightness Control in fluoroscopy
6-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-4 ARCHITECTURE
2-4-1 A Kernel
6-5
REV 19
JEDI GENERATOR / FUNCTIONAL ARCHITECTURE
Illustration 2 -
System Control Bus
GE MEDICAL SYSTEMS
System
Interface
6-6
AC/DC
Jedi Generator / Functional Architecture
fil. drives
Management
HV measures
inverter controls
High Voltage High Voltage HV Cables
rotation
rotation
EMC Inverter
+safeties+f
Tank
Filter
+safeties+fans
The packaging architecture consists in a set of boxes which can be put together in several ways to make Jedi fit
either in a cabinet, or a console foot, or a table foot. The boxes can also be split in 2 units distant of several
meters (example: CT gantry). Refer to Illustration 3.
The boxes normally consist of the following:
• Rotation function
• Heater function
• Low voltage power supply (which can also be in the AC/DC box)
This box is always present.
• HV tank
• HV power inverter
• kV control
• System interface (for the less complex system interface)
This box is always present.
• MC filter (optional)
• AC/DC function
• Low voltage power supply (which can also be in the auxiliaries box)
This box is always present.
• RAD interface
• AEC interface
This box is present in the RAD product.
• Tube selection
• Grid/Bias control
6-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Inverter Assembly
Dual Snub LVPS Board
(for TIGER) AC/DC
Tube Board
Gate Command
HV Board
+-
CT
Inter KV Control Board
face
EMC Filter (Optional)
Heater Board
Tube Grid/Bias
Selection Control
Rotation Rotation
Capacitor Capacitor
(optional)
Auxiliaries Box Optional Boxes
6-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Four electro-magnets (24 VDC) are positioned extremely close to the rails that
the bearings ride on. These magnets “grab” the steel rails securely, locking the
longitudinal travel of the table. Four solenoid-operated fail-safe tilt locks are used
to grab the transverse shafts, locking the transverse travel of the table.
Depressing the four-way float pedal de-energizes the magnets and energizes the
solenoids, allowing the table top to float freely.
7-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
U7 and its related circuitry form a Time Safety Protection circuitry (12 seconds,
adjustable) for the motor.
7-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
• Power Supply
1. +24VDC is supplied by a switching power supply with UL and EMC approved
and can be adjustable in the range from 21VDC to 28VDC.
2. +5VDC is supplied by the MAX726 and its related circuitry.
3. +45VDC supplier circuitry consists of the transformer T1, bridge CR2, etc.
7-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
JEDI EXCEPTION HANDLING
JEDI software performs auto-test at power up and continuously monitors the
correct operation of its functions during application. Any malfunction is stored in
the JEDI error log and reported to the system through a protocol that transports
error code. Errors found can only be reported if the generator is powered on and
live.
2-1 DIAGNOSTICS
At power-up the kV control performs its own initialisation, checks its memory
integrity (checksum of program and NVRAM) and starts the communication with
its peripherals as well as the system. Communication is permanently checked
afterwards. It then initialises the Rotation board and Heater board with their
respective database parameters and loads kV control FPGA.
Eight LEDs (S7..S0) on kV the control board show the software status.
During power-on, the Heater board and the Rotation board CPUs are initialised
and check their memory integrity and hardware. If a problem is encountered, a
PRD error is reported to the kV control.
• heating function
• rotation function
3. Manual Diagnostics
7-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The error code structure described in this chapter applies to JEDI error detection
and logging. The JEDI error log file can be accessed from the system through the
system console or a laptop (JEDI error log upload functionality).
When an error is detected, it is sent to the system and is logged in parallel in the
JEDI error log file. The file contains a maximum of sixty four logs.
Each log shows the following structure:
The simplified error code is a grouping of the JEDI error codes. This field gives a
rapid understanding of which part of JEDI is faulty.
Simplified Description
Error Code
30 Tube spits errors
40 Rotation errors
50 Heater errors
60 Exposure errors (HV inverter + mA measure + exposure control
70 Power supply errors (low voltage + DC bus)
80 Hardware errors (internal communications + cables)
90 Application errors (saved RAM + software)
100 External communication errors
110 Thermal errors
10 Rotation warnings (engineering use)
20 Heater warnings (engineering use)
25 LVPS warnings (engineering use)
27 Application warnings (mainly saved RAM battery change)
7-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The generator phase field contains the state of the generator when the error
occurred.
Generator Description
Phase
0 idle: entered in diagnostic mode
1 powered up; waiting for configuration
2 Stand by: configuration completed, waiting for a preparation
3 Preparation in progress: JEDI gets ready to take X-rays
4 Ready for exposures (rotation at speed; filament; HV inverter drive
5 High voltage on
6 Error detected and not yet cleared
There are five classes of errors that correspond to different levels of impact to the
system. The class of errors correspond to the seriousness of the error and the
system software will manage operations
Class Description
1 Errors which have no impact on the system operation.
Errors detected in background diagnostics during application. They are referred
as ”Warning” errors, to monitor drifts and used for engineering tracking.
The generator phase remains unchanged.
Errors are stored in the JEDI error log.
2 Errors which are detected by JEDI and that are recoverable automatically without
noticeable effect on the system, such as error related to recovered tube spits.
The generator phase remains unchanged.
Errors which usually occur during exposure.
3 Errors detected by JEDI during exposure. They stop the exposure and revert the
generator into a safe state.
Error will be reset on Exposure Command release. It will require another
Exposure Command to restart the sequence.
The generator phase is set to ”error” until the error is cleared.
4 Errors which are related to any hardware failures, software application or
communication errors.
JEDI will revert to a safe state.
If preparation is in progress, it is stopped.
Errors are cleared either by a reset error action from the system (for system
having a reset error mechanism) or by a prep release or by a new prep command.
The generator phase is set to ”error” until ther error is cleared.
Application cannot work if errors are persistent.
5 These codes may appear when the generator or tube temperature limits are
reached.
The application waits until the thermal information disappears. The error
information is temporary.
The generator phase is set to ”error” until the error is cleared.
7-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Each error code comprises two fields (which cannot be generated and used
separately);
• the first field describes the JEDI function which is faulty (referred to as the
function code)
• the second field describes the error detected.
Example: error code 0306 means:
• 03: high voltage generation function;
• 06: no kV feedback on anode.
Function Codes List
Function Description
Code
01 Rotation
02 heater
03 High voltage generation
04 mA control
05 Power supplies
06 System interface
07 Software
08 Application
09 Tube control
10 Tube management
11 Grid
12 Bias
13 AEC
14 Hardware
7-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 3
ERROR CODES, DIAGNOSTICS & TROUBLESHOOTING
3-1 INTRODUCTION
This diagnostics section is to introduce the JEDI generator with the following
information:
Power-On diagnostics LED indication.
List all the potential error codes that can be issued by JEDI Generator.
Provide error code explanation, potential cause and recommended action.
List of diagnostics aids and explanation of diagnostics.
This section also includes the Collimator error codes.
_ ⊕ ~~........
S0.........................S7
One out of two is lit: Application (Flash) problem. Reloaded the APPL. mx file.
- When an application error occurs (Not PRD)
P
P
e.g: No communication
The simplified error code is displayed on the Leds. They blink; when the error is
cleared ( by a return to the standby mode for example ), the 8 LEDs are lit
successively.
Heater board LED status: (See central listing)
After the power on diagnostics, heater board LEDs DS1 and DS2 are lit
successively. Any different status correspond to an abnormal situation. An error
code is logged. Refer to error code description.
Rotation board LED status: (See central listing)
After the power on diagnostics, rotation board LED DS5 is blinking. Any different
status correspond to an abnormal situation. An error code is logged. Refer to
error code description.
7-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
The error code list and associated short description is presented below.
7-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
25 Low Voltage 1 0550H No more warn +160V too low or too gigh
Power supply
Warnings
1 0551H Detected +160V too high
1 0555H Detected +160V too low
1 0560H No more warn +15V too low or too high
1 0561H Detected +15V too high
1 0565H Detected +15V too low
1 0570H No more warn -15V too low or too high
1 0571H Detected -15V too strong
1 0575H Detected -15V too weak
1 0599H Unknown LVPS warning
7-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
If the lamp cannot be switched on when push the lamp button on front panel, refer to the following trouble
shooting.
1. Open the collimator front cover, assure stable cable connection on J4, J7 and J8 of user interface board.
2. Push ‘lamp’ button to light on lamp. Check the LED indicator status (DS7, DS10, DS13), as shown in following
Figure
Indicator
Error Recommended Reactions
DS13 DS10 DS7
on - - Lamp failure Exchange lamp
off on on UIF board failure Exchange lamp
off off - System lamp power failure Check power supply unit on system side
Exchange UIF board; if failure remains, change whole
off - off UIF board or UIF panel failure
collimator
7-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
For Collimator error code, refer to diagnostic section afterwards for potential
cause and recommended action on Page 7-29
7-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
25 Low Voltage 0550H No more warn +160V too low or too high
Power Supply
Warnings
0555H Detected +160V too low
0560H No more warn +15V too low or too high
0561H Detected +15V too high
0565H Detected +15V too low
0570H No more warn -15V too low or too high
0571H Detected -15V too strong
0575H Detected -15V too weak
0599H Unknown LVPS warning
7-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-6 DIAGNOSTICS
This chapter describes diagnostics tools based on error codes and specific
diagnostics.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS
WARNING OFF. APPLY LOCK OUT-TAG OUT PROCEDURE FOR YOUR OWN SAFETY
WHEN MANIPULATING INSIDE THE EQUIPMENT IS REQUIRED.
3-6-1 Detailed Error Description & Troubleshooting Guide
The first part details errors, while the warning errors are listed in a second part of
this chapter.
The table below provides guidelines to troubleshoot Generator and Collimator
problems based on error code.
For each code, there is:
-Associated message and additional explanation related to the error
occurrence.
-List potential cause, in the order of expected probability.
Codes are sorted by ascending order both for simplified code and error code.
Refer to the theory of operation for error code structure for Generator error code.
Information about associated data structure is located at the end of each error
code subset whenever it applies.
Whenever wiring, cabling, LED check is mentioned in the recommended actions,
refer to the central listing.
3-6-1-1 Errors
Tube spits detection Errors (Code 30)
7-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
component failure:
0001H=RAM
0002H=RAM stack
0200H=EPLD
8000H=program checksum
7-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
other errors:
rotation state:
0=inverter OFF
1=acceleration 0 to low speed
2=acceleration 0 to high speed
3=acceleration low speed to high
4=low speed run
5=high speed run
6=high speed to low speed brake
7=brake reverse
8=brake DC
7-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Not Proteus
100- AEC communication error 1. AEC board
1301H KV Control has detected a 2. I/F board
communication problem 3. kV control
with AEC board. 4. Wiring
7-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Z14 Collimator Homing Error Reset collimator; if fail again, change collimator
7-30
REV 19
Neon's
*
LEDs
*
GE MEDICAL SYSTEMS
DS 1
DS 3
+ M -
1
F
*
INVERTER. & LLC. Circuit DS 1 DC bus, 160V
- 15V
(120 to 200V)
+ 15V
Gate Command Board
DS 300
*
* LVPS
I/F AC/DC NE 1 DS 2
Console PUPT-
7-31
or
DS7-6
DS 1, 2
Illustration1 - GENERATOR Visual Power Supply Distribution
-
- 15 V
+
-15V, +15V DS 3, 2, 1
AEC + 15 V Rotor +5, -15, +15
kV ctrl Board
*
- 15 V "VCC": +5V DS 7
N1 (230 V)
*
+ 15 V
EMC Flt.
3ph, 380 - 480 VAC
CB1
PDU K1
DIRECTION 2273022-100
Service Manual
PROTEUS XR/a
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Pre-requisites:
• kV control board alive and running: S0-S7 Leds are lit successively or a
combination of them blink
• heater board alive and running: DS1 and DS2 Leds are lit successively
• heater DC supply present: DS3 Led is lit
Sequence:
Select the tube to run the test on, start the diagnostic.
The following sequence runs on the small focus and then on the large focus:
• 5 seconds preheat
• 0,4 s boost
• 5 s heat (focal spot max current - 1Amp)
There is 10 s stop time between each focal spot run
During the test, the heater safeties are checked the same way than in application
mode.
7-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Error reporting:
Error Conclusion
DS1 neon lit Check that DC bus cables have been removed
DS300 neon off Check the gate_cmd supply cable between AC/DC and gate_cmd
board
0301/0302/0303/0304/0309/ Check that DC bus cables have been removed.
0310/0311/0312/0313/0314/ If yes, replace kV control board
0319/0323 (H)
0320 if problem persists, replace Kv control board
0501 kV control or inverter fault, replace inverter first
0503 kV control or inverter fault, replace inverter first
0504 kV control or inverter fault, replace inverter first
one of Replace inverter
DS101/DS102/DS201/DS202
LEDs off while no error
reported
DS100 and/or DS200 LEDs Check cabling between kV control, HV Tank and gate_cmd board.
off If cabling is right, kV control or inverter fault, replace inverter first
7-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Error reporting:
error Conclusion
DS1/DS2 neon off Check that DC bus cables between AC/DC and inverter
DS300 neon off Check the gate_cmd supply cable between AC/DC and gate_cmd
board
0301/0302/0303/0304/0309/ Check that HV Tank primary cables have been removed.
0310/0319 (H) If yes, replace kV control board
0311 No Ilp current detected. See note 2
0312 No Ilr current detected. See note 3
0313 replace kV control board
0314 Ilr current resonant frequency is lower than expected. See note 4
0320 if problem persists, replace kV control board
0323 Both Ilr and Ilp currents not detected. See note 1
0501 kV control or inverter can be faulty
0503 kV control or inverter can be faulty
0504 kV control or inverter can be faulty
0505 Isolation fault between inverter components and ground. Check inverter
inductors. If no faulty component, kV control or inverter can be faulty
NOTE 1: Ilp and Ilr currents not detected
1. Check the -15 V (Led DS1) on kV control board (see central listing).
If it is not lit, refer to “other failures” section. Else:
2. Power off the Generator. Wait until all neons are off
3. Check that the currents transformers (capacitor set) to gate_cmd board cable
is correctly connected.
If yes:
4. Check that the inverter inductors are correctly connected. If yes:
5. Check that the gate_cmd to HV Tank cable and HV Tank to kV control cables
are correctly connected.
If yes, replace the inverter
6. Reconnect all the cables
NOTE 2: Ilp current not detected
1. Power off the Generator. Wait until all neons are off
2. Check that the parallel inductor is correctly connected. If yes:
3. Check that the parallel inductor impedance is 0 Ohms. If no: replace inverter.
If yes:
4. Check that inverter capacitors (capacitors set) are not broken. If yes, replace
the capacitor set. Else:
5. Disconnect the currents transformers to gate_cmd board cable. Check that
the parallel current transformer impedance is 0. If no: replace the capacitor
set. Else:
6. Check that the gate_cmd to HV Tank cable and HV Tank to kV control cables
are correctly connected. If yes:
7. Disconnect the HV Tank to kV control cable. Check that the impedance
between pin 20 and pin 21 of J2 of HV Tank is 3,3 Ohms. If no replace the
inverter. If yes replace kV control board.
8. Reconnect all the cables.
7-39
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Pre-requisites:
• application mode without AEC mode is running properly
• kV Control board alive and running: S0-S7 LEDs are lit successively or a
combination of them blink
• cabling between IF board, AEC board and ionization chambers checked.
Test type:
Automated, with the use of Service Laptop
Sequence:
Select AEC function in menu 1 on Service Laptop.
Press V to run the test.
AEC function is verified through a series of 7 tests as follows:
• AEC return measurement test:
A voltage reference is switched to one input of the AEC select multiplex.
Nominal value of return voltage = 5.24 V
Test 1 is OK if: 4 V < measured value < 6 V
This test is also performed at lower frequency count up:
Test 2 is OK if: 4 V < measured value < 6 V.
• Switching Gain test (x10):
A voltage reference is switched to one input of the AEC select multiplex and
count up gain is applied.
Nominal value of return volatge = 4.76 V
Test 3 is OK if: 4 V < measured value < 6 V.
• AEC ion chamber HV power supply test:
A portion of HV power supply is switched to one input of the AEC select
multiplex:
- No cell selected, Test 4 is OK if return voltage measurement is 0 V.
- Select respectively one of the three cells, left, right, middle.
Test 5, 6, 7 are OK if: 1 V < measured value < 3 V.
7-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
(*) except for AEC board with serial number: 2184569-X: the resulting error code
must be 0001111 (test is OK on this board the HV power supply test is not
performed).
For the other error codes, refer to the troubleshooting table.
7-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
7-44
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: The planned maintenance to the OTS suspension see Direction 2277285-100, OTS Suspension
Service Manual.
8-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
OPERATION FREQUENCY JOB DESCRIPTION
CARD
ACCESSORIES
Compression band 12 months PM007 Check condition
Patient hand grips 12 months Check condition
GENERATOR
Check generator functions 12 months PM008 Test exposure, including AEC if resent
Record last error codes
Record result on counters
Check technique factors
Tighten terminals and remove dust
Replace NonVolatile RAM Every 8 Years PM Jedi 1. Upload database
on the kV Control Board 15-2 2. Battery replacement
3. Download database
WALL STAND (GPCP NO. 2260354) (OPTIONAL COMPONENT)
General cleaning and As required PM001 Keep all surfaces clean and touch-up paint chips
painting
Visual inspection 12 months PM002 Check for debris that would indicate abnormal wear
Functional checks 12 months PM003 Perform the tasks described in Chapter 5,
Functional Checks
Steel cables 12 months PM009 Inspect and grease length of steel cables; replace if
worn out
Carriage roller bearing 12 months Clean roller bearings and tracks
inspection
Vertical lock 12 months Clean and inspect
LAT bar 12 months Replace grease between the flanges
SG100 WALL STAND (GPCP NO. 2225849) (OPTIONAL COMPONENT)
General cleaning and As required PM001 Keep all surfaces clean and touch-up paint chips
painting
Visual inspection 12 months PM002 Check for debris that would indicate abnormal wear
Functional checks 12 months PM003 Perform the tasks described in Chapter 5,
Functional Checks
Steel cables 12 months PM010 Inspect and grease length of steel cables; replace if
worn out
Carriage roller bearing 12 months Clean roller bearings and tracks
inspection
Counterweight Chain 12 months Inspect and grease length of chains. Replace if
Maintenance excessively worn or rusty.
Vertical lock 12 months Clean and inspect
Angulation Components 12 months Inspect and Replace if necessary
(lock & cable)
Bucky Rotation Lock Check 12 months Verify lock mechanism does not move
8-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
OPERATION FREQUENCY JOB DESCRIPTION
CARD
Electrical Checks 12 months Inspect and replace any damaged cables or
connectors.
Check grounding.
JEDI PREVENTIVE
MAINTENANCE
Replace Nonvolatile RAM Every 8 Years 1. Upload database
on the kV Control board 2. Battery replacement
3. Download database
COLLIMATOR
Check the collimator 12 months PM012 1. Check the flange.
mounting screws 2. If the collimator mounting screws are loose,
apply LOCTITE onto the screws.
X-RAY TUBE CABLES
CHECK
Check the fixing and 12 months PM013 1. Check the fixing and intactness of Anode
intactness of Anode Rotation cable and Thermal Protection
Rotation cable and Thermal cable.
Protection cable 2. Check the stator cable strain relief.
8-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
PREVENTIVE MAINTENANCE JOB CARDS
The following pages list the preventive maintenance Job Card detailed activities.
SUPPLIES
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
8-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
8-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
No specific precautions.
PREREQUISITES
None.
PROCEDURE
8-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
Lubricant.
TOOLS
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
During the maintenances, before removing the motor from under the table, or untie the
CAUTION motor chain, to avoid the injury by the potential dropping table, put the blocks kit
(5137108) over the rod.
2)Emergency buttons
a) Confirm emergency button removes table power when depressed and applies table power when
released.
3)Electrical cables.
a) Remove power from system.
b) Check condition of all cables including cable sheathing, connectors and overall condition; if any cable
damage is noted, replace or repair affected cable.
c) Confirm that the cable is correctly routed.
8-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
8-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
C. When the rod is pulled out of rod end support, take out the rod.
8-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROTEUS XR/a 4/4 PM 004
h) Remove the old brake plates and replace them with new ones (renewal part number is 612-3440).
i) Use a soft cloth to clean the lubricant oil on the surface of a new rod (renewal part number is 613-3003)
j) Use a soft cloth to clean the lubricant oil in the groove of the two bearings on each side.
k) Insert the new rod.
l) Adjust the screws to make sure the plate travel distance is between 3-5mm, and lock it with locking nut.
m) Repeat the similar steps from e to l as above, to replace the rod and couple of brake plates on the other
end of the table.
8-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
Grease.
Light Machine Oil.
TOOLS
None.
SAFETY PRECAUTIONS
PREREQUISITES
None.
8-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROCEDURE
1)Surface finish
Remove the dust from the table top and table topside rails.
2)Longitudinal bearings
Remove the dust.
8-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
Grease.
Light Machine Oil.
TOOLS
None.
SAFETY PRECAUTIONS
PREREQUISITES
None.
8-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROCEDURE
6)Tray
Confirm the pull-in and pull-out movement is smooth.
8-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
None.
TOOLS
None.
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
1)Compression band
Mount and remove the compression band on the table top.
2)Hand grip
Mount and remove the hand grip on the table top.
8-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
Film.
Error–code recording form.
X–ray counter recording form.
TOOLS
Laptop.
SAFETY PRECAUTIONS
No specific precautions.
PREREQUISITES
8-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROCEDURE
Technique factors (mA, mAs and ms) are displayed after exposure on the console.
1.2 mA accuracy
Run the exposures in table below and check if the mA value on console after exposure is between the min
and max value.
If not , run the same exposure 10 times with exposure time set to 12.5s and then check again.
8-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
8-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
8-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROTEUS XR/a 1/2 PM JEDI 15-2
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
4 PREREQUISITES
Upload the database that is stored in the RAM before starting the replacement procedure (refer to Job Card CAL
JEDI 8-6 ”Backup & Restore Database Setup Parameters”).
8-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
On the kV Control board, note the direction that the RAM faces. Insert the tip of a small screw driver underneath
left side of U32 and lever gently, then grasp it between your thumb and forefinger and pull it directly away from its
support on the board.
Insert the new Nonvolatile Parameters RAM in its place. Be careful to insert it facing in the same direction.
Reapply power.
Download the database (refer to Job Card CAL JEDI 8-6 ”Backup & Restore Database Setup Parameters” in
Chapter 4).
8-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROTEUS XR/a 1/1 PM 009
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
8-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROTEUS XR/a 1/1 PM 010
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
Note: The procedures in this Job Card are concerted with the revision 6 of SG 100 Wall Stand Service
Manual (GPCP Number 2150297-100). For the reason of the update and change in that manual, it is
recommended to see also the latest Service Manual of SG 100 Wall Stand for reference.
8-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
PROTEUS XR/a 1/1 PM 011
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
8-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Collimator
purpose: Maintenance Procedure version no. 0
date: 27/09/01
NF No. Production date Serial No. Personnel:1
from / / from Time: 15 min
to / / to
SUPPLIES
None.
TOOLS
Allen Key
Calliper
SAFETY PRECAUTIONS
PROCEDURE
8-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SUPPLIES
None.
TOOLS
None
SAFETY PRECAUTIONS
PREREQUISITES
None.
PROCEDURE
1xM4 screw,
2xM4 flat washers.
Thermal cable Apply Loctite
242(46-170686P1)
clamp
mounting
block
Stator cable
Attach P-clamp
Copper strap
using M4 hex. nut
8-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
*1: Item 68 will replace item15. These two kinds of Ion-chamber have same installation procedures and
functions.
Note: If JEDI’s model number on it’s label (see the figure) is 2268970, you should follow the table
Cabinet 1. If the number is 2244165-2, you should follow the table Cabinet 2.
9-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: For SG100 Wall Stand (GPCP No. 2225849) FRU, refer to SG100 Service Manual (2150297-100).
Note: For SG120 Wall Stand (GPCP No. 2402562) FRU, refer to SG120 Service Manual (Supplier Part No.
S0012532).
9-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Note: If the site needs to install the tube fan kit, Please use S3918LZ to order item #25,
item #28 and item #32 together.
9-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-1
Table Illustrated Parts (1)
Item# 1
Item# 7
Item# 9
Item#48,49
Item# 53 (the
spring leaf under
the button)
Item# 11
Item# 52
Item# 51
Item# 12
Item# 13
9-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-2
Table Illustrated Parts (2)
Item# 15
Item# 14
Item# 22
Item# 21
Item# 23
9-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-3
Table Illustrated Parts (3)
Item# 27
Item# 28
9-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-4
Table Illustrated Parts (4)
Item# 29
Item# 33
Item# 3
Item# 32 Item# 30
Item# 48
Item# 47
Item# 31
Item# 34
Item# 35
Item# 38
Item# 36
Item# 71
Item#37
Item# 70
9-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-5
Table Illustrated Parts (5)
Item# 40
Item# 39
Item# 42
Item# 41
Item# 44
Item# 43
Item# 72 Item # 58
9-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-6
Table Illustrated Parts (6)
Item# 45
Item# 46
Item # 56
Item # 68
Item # 55
Item # 59
Item # 69
Item # 57
9-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Item # 65
Illustration 1-7
Table Illustrated Parts (7)
Item # 62
Item # 60 Item # 61
Item # 66/67
Item # 63
Item # 66/67
Item # 64
9-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-8
Cabinet 1 Illustrated Parts (1)
Item# 1
Item# 6 Item# 5
Item#4
Item#7
Item# 2
Item# 3
Item# 13
Item# 8
Item# 14
Item #32
Item# 10 Item# 12
Item# 9 Item# 11
9-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-9
Cabinet 1 Illustrated Parts (2)
Item# 16
Item# 15 Item# 30
Item# 17
Item# 18
Item# 19
Item# 20
Item# 24
Item# 23
Item# 25
Item# 26
Item# 21 Item# 22
Item# 27
Item#31
Item#28
Item#29
Item# 33
9-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-10
Cabinet 2 Illustrated Parts (1)
Item# 2
Item# 1
Item# 3
Item# 11
Item# 12
Item# 4 Item# 8
9-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-11
Cabinet 2 Illustrated Parts (2)
Item# 15
Item# 18
Item# 19
Item# 16
Item# 20
Item# 17 Item# 22
Item# 23
Item# 21
9-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-12
Wall Stand Illustrated Parts (1)
Item # 3
Item# 1 Item # 2
Item # 12 Item # 13
Item # 7, 8, 9, 10 Item # 15
Item # 14
Item # 19
Item# 19
Item # 16 Item # 20
Item # 18 Item # 17
9-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Item # 23
Item # 30
Item # 21 Item # 22
Item # 27/28
Item # 24
Item # 27/28
Item # 25
9-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-14
OTS Illustrated Parts (1)
Item# 1
Item# 2
Item# 3
Item# 5
Item# 54
Item#7
Item# 4 Item# 6
Item# 53
Item# 8
Item# 52
Item# 9
9-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-15
OTS Illustrated Parts (2)
13
10
14
10
11
Item#19
Item# 60
Item# 61
12
Item# 63
Item# 59
Item# 64 Item# 65
9-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-16
OTS Illustrated Parts (3)
Item 20
Item 33
Item# 21
Item# 22
Item# 24
Item# 23
Item# 34
Item# 29 Item# 55
Item# 35
9-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-17
OTS Illustrated Parts (4)
Item #27
Item #25
Item #67
Item #26
Item #38
Item #39
Item #40
Item #37
9-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-18
OTS Illustrated Parts (5)
Item #43
Item #44
Item #45
Item #48
Item #47
Item #51
Item #49
Item #46
Item #50
9-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration 1-19
OTS Illustrated Parts (6)
Item #58
Item #57
Illustration 1-20
Console Illustrated Parts
Item #6
Item #1
Item #2
Item #4
9-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
JEDI GENERATOR RENEWAL PARTS
2-1 Introduction
This section provides information for the identification of Renewal Parts for the
Jedi 50Plus R1T generator
BEFORE ORDERING A SPARE PART, BE SURE TO KNOW THE JEDI
WARNING
GENERATOR YOU HAVE ON SITE.
The model number of the Jedi generator is localized on rating plate on Auxiliary
module (see page 10-24 Jedi Generator Illustration: labelling)
If the model number is 2244165 you have a JEDI 50plus R 1T.
If the model number is 2268970 you have a JEDI 80 R 1T
Order the spare part according to the JEDI generator you have.
If only one part number is written order this one, otherwise order the spare part
according to your generator
When replacing the Jedi Generator, pay attention to check the GPCP No. of the Rotation
CAUTION
Capacitor Set of the Auxiliary Module of Jedi Generator (refer to Job Card D/R JEDI 1, D/R
JEDI 17-38) and ensure its correct match with the X-ray tube type. Pay attention to
download and upgrade the Jedi software (Refer to Job Card SW JEDI 7-2 in Chapter 4.).
For Rad 14,21 X-ray tube, the GPCP No. should be 2250002.
9-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Wherever possible, the FRU or its position is illustrated for easy indentification.
The Rev 2.14 of the System Console Software 2286799-16 can be used together with
either the System I/F board 2375230 (Item #1 of Cabinet 1) or the System I/F Board
2259298-45 and CAN I/F Board 2259298-46.
9-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration – JEDI 50Plus R 1T and JEDI 80R 1T Generator Location of Main Assemble FRU2: 2244165.
Miscellaneous Items:
First aid spares: (FRU: 1) GPCP No. 2217057 Services tools (FRU: 2)
1 PLASTIC ZIP LOCK BAG 160X220 10 TY-WRAPS D22/L99X2.5 Service software
10 SCREW HEX FLANGE HEAD M4X8 1 SHORT CIRCUIT CABLE 9-wire serial cable, No. 2207478
SCREW HEX FLANGE HEAD 4MM 8MM 1 CONNECTORS KIT FOR FC2 For JEDI 50Plus R 1T: 2244165-2
SCREW HEX FLANGE HEAD 5 MM 12 2 PLASTIC ZIP LOCK BAGS For JEDI 80R 1T: 2268970
MM 40X60
1 FUS FF 14X51 15A 700VDC C 50KA 3 PLASTIC ZIP LOCK BAGS
ULR 80X120
1 FUS FF 14X51 2A 700VDC C 50KA 1 RECT CONNEC 9C FEM
ULR AWG14 SCREW
1 SLOW BLOW FUSE HCP 6X32 10A 10 SCREW HEX FLANGE
250V UL HEAD 5 MM 12 MM
1 PADDED BAG W/ ADH CLOSURE
260X370
9-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration – JEDI 50Plus R 1T and JEDI 80R 1T AC/DC Unit, Interface Unit & Auxiliary Unit
Bucky Connector
91651653 FRU: 2
AEC Board V2
2226456-3 2222253-3
FRU: 1 FRU: 1
Job Card D/R JEDI 17-47 Job Card D/R JEDI 17-46
Rotor Capacitor
2250002 (2215674 for MX
100 X-ray tube)
FRU: 2
Job Card D/R JEDI 17-38
9-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2265276
-2
9-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-4 Cables
9-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 3
JEDI GENERATOR DISASSEMBLY/REASSEMBLY
3-1 Introduction
3-2 TOOLS
K
Digital multimeter (A)
Philips screwdriver (B)
Allen key (C) J
Screwdriver with interchangeable socket head (D)
Long extension for ratchet wrench (E) I
F
Short extension for ratchet wrench (F)
Cutting pliers (G)
High Voltage wrench (H)
Ratchet wrench (I) C
Socket wrench (J) E
A
Hexagonal head socket set (K).
D G
B H
9-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 INTRODUCTION
Jedi Generator are consist of four modules. Each module can be disassembled from the system.
AC/DC Auxiliary
Unit Unit
Power Interface
Unit Unit
9-38
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Screw
Communication KV Control
Cable Cable
9-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
Check with the Multimeter in DC mode that the voltage is below 10 V between “+” and “-”
CAUTION terminals on the Power Unit (see below). IF NOT, WAIT UNTIL VOLTAGE BECOME O V
BEFORE REMOVING CABLES.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
9-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4 PREREQUISITES
THE WHOLE INVERTER UNIT WEIGHTS 30 KILOGRAMS (66 POUNDS). TWO PEOPLE
WARNING
ARE REQUIRED TO CARRY IT.
5 DISASSEMBLY
5 4
1
3
9-44
INVERTER UNIT
TOP VIEW
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3 J1 4
9-45
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
9-46
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10. Take off the assembly constituted by: EMC covers, Coils Inverter Assy, Capa Inverter Assy, and KV Control
Board. The HV Tank 9 and the Inverter Power Assembly 10 are now free .
10
In case of change the Inverter Power Assembly, you must recover the KV control Board
NOTICE 11 and mount it in the new Inverter Power Assembly.
12
13
11
9-47
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
2. Mount the flat cable J1 on the KV Measure Board of the HV Tank (refer to section 5–3 4 ).
3. Mount the eight 3 mm Allen screws to secure the heat sink of the Inverter Power Assembly to the HV Tank.
(refer to section 5–3 8 ).
Tighten to torque 8 N.m (80 kg.cm).
4. Put back the 3 front 5.5 mm hex head screws (refer to section 5–3 6 ), one at the left side, two at the right
side.
Tighten to torque between 1.5 to 2 N.m (15 to 20 kg.cm).
5. Reconnect the ribbon cable from J1 on the KV Measure Board to J3 on the Gate Command Board (refer to
section 5–3 4 ).
6. Using the 13 mm socket wrench, connect the primary coil flat cables to the four capacitors 1 .
Respect the position of the two flat cables with white bands on them (refer to illustration
CAUTION above).
Make sure the presence of the IGBT Middle point 2 .
White
BAND
White
BAND
9-48
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
8. Reconnect the wider short ribbon cable from J2 on the KV Control Board to J2 on the kV Measure Board
(refer to section 5–3 3 ).
9. Mount the 5.5 hex head screws securing the two EMC covers to the HV Tank on the HV outlet side (refer to
section 5–3 1 ) and the three screws on the other side (refer to section 5–3 7 ).
Tighten the screws to torque between 1.5 to 2 N.m (15 to 20 kg.cm).
10. If Inverter Power Assembly has been changed, reverse the disassembly procedure to reinstall the KV Control
Board (refer to section 5–3 11 ).
Tighten 5.5 mm screws to torque between 1.5 to 2 N.m (15 to 20 kg.cm).
11. Remount the Interface Unit following the inverse procedure indicated in section 5–2.
12. Remount the Module in the Cabinet following the inverse procedure indicated in section 5–1.
Be sure to place the Flat ribbon cable 3 J3 KV Control Board at the right position, see
NOTICE
illustration below.
9-49
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
CAUTION Be sure to place the “+” and “-” DC BUS in the right connector.
13. Put 5 cl of oil into the HV receptacles. Reconnect Console, buckies and AEC cables.
7 VERIFY
Connect the service laptop (refer to page 4-26 ”Service Software User Guide” JOB CARD) and proceed to ”HV no
load diagnostics” to ensure correct operation.
Skip steps ”Gate Command and Inverter in Short Circuit” diagnostics.
9-50
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
subassembly: Rotation Capacitors Set 23/56 (Rotation Capacitor Set 23/23 for MX100 X-ray tube)
purpose: version no. 0
date: 20/12/99
NF No. Production date Serial No. Personnel:1
from / / from Time: 0h 15min
to / / to
Note: When installing the system, do not use the capacitors in accessories crate #4 (refer to form “Crate
dimensions & weight” on page 2-5 in Chapter 2) for installation. Pay attention to pick out and
dispose the capacitors in accessories crate #4. Ensure the eco-friendly disposal of the capacitors.
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
The Rotation Capacitors are in the Jedi Generator Auxiliary Unit (10 kg).
On the Auxiliary Unit, disconnect internal CAN cable, DC BUS, heater cable, and Tube cable.
9-51
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 PROCEDURE
Rotation
Capacitors 2
1. Cut the tie-wraps 1 1
1
2. Disconnect both capacitors 2 2
9-52
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
1 2
2
- connect the right capacitor 6.4 µF to the board with cable 2259372
connect the two cable terminals marked “C6.4 UF” to the upper pins of the right capacitor 6.4 µF
connect the two cable terminals marked “CF 11” and “CF 12” to CF11 and CF12 on the board,
respectively.
For MX 100 tube:
1
- connect the left capacitor 12.6 µF to the board with cable 2215658
connect the two cable terminals marked “C12.6 UF” to the below pins of the left capacitor 12.6 µF
connect the two cable terminals marked “CF 9” and “CF 10” to CF9 and CF10 on the board, respectively.
2
- connect the right capacitor 10.5 µF to the board with cable 2215659
connect the two cable terminals marked “C10.5 UF” to the upper pins of the right capacitor 10.5 µF
connect the two cable terminals marked “CF 11” and “CF 12” to CF11 and CF12 on the board, respectively.
9-53
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
CAUTION Pay attention to check the GPCP No. of the Rotation Capacitor Set and ensure its correct
match with the different X-ray tube type.
For Rad 14,21 X-ray tube, the GPCP No. should be 2250002.
7 VERIFY
9-54
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
THE WHOLE INVERTER UNIT WEIGHTS 30 KILOGRAMS (66 POUNDS). TWO PEOPLE
WARNING
ARE REQUIRED TO CARRY IT.
9-55
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
Interface
Unit
9-56
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
1. Reverse the disassembly procedure to reassemble the unit. Tighten screws to torque between 1.5 and 2 N.m
(15 to 20 kg.cm).
2. Power the system on and verify that the system functions properly.
9-57
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
THE WHOLE INVERTER UNIT WEIGHTS 30 KILOGRAMS (66 POUNDS). TWO PEOPLE
WARNING
ARE REQUIRED TO CARRY IT.
The KV Control board contains a saved memory Database with essentially the following parameters:
- Tube configuration parameters, AEC calibration parameters, and Room configuration (workstation, AEC,
screen film pairs).
- Filament aging variable
- Error Log
- Recorded parameter data
If it is possible, backup database first. Else, the diagram hereafter helps you to determine the action to take for
your situation.
9-58
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
First
DB Is
Upload Yes Upload No original DB No
Database Possible? saved?
(*)
Yes
Yes No No
Download Download
Software Software
Download Configura-
Download Original
Database tion if appli-
Database cable
(*)
Calibrate if
applicable
Reset
Database:
* Date
* Time
*Clear Error
log
Operational
Checks
Save Non
Volatile RAM
Database
9-59
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
Interface
Unit
9-60
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3 J1 4
1
2. Remove the seven 5.5 mm hex head screws 2 which
secure the kV Control Board; remove the board .
2
9-61
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
Reverse the disassembly procedure to reinstall the KV Control board and the Interface Unit.
Tighten 5.5 mm screws to torque between 1.5 and 2 N.m (15 to 20 kg.cm).
7 VERIFY
1. Connect the Service Laptop (refer to page 4-32 ”Service Software User Guide” JOB CARD)
2. Reapply power and verify that the 8 LEDs on KV Control board are scrolling.
4. Restore Database (refer to page 4-86”Backup & Restore Database Setup Parameters” Calibration Job Card)
- Up-to-date database if available
- Most recent or originally saved Data Base if available
- Default one from the floppy delivered with the equipment in the Service Tool set.
Alternately, you may try to swap the NVRAM chip (U32) from the former to the new KV Control board to keep all
the parameters. Refer to page 8-18”Replace Non Volatile Parameters RAM” Job Card. If this chip is not involved
in the failure or the cause for replacement, it will allow to keep up-to-date data.
5. Calibrate:
- Perform several exposures on both focus in several workstation, with exposure time longer than 12.5 ms
If default database was restored, it is necessary to resume room set up (refer to page 4-52”Workstation
parameter” Calibration Job Card) and AEC calibration (if present) (refer to page 4-60” Automated AEC
Calibration” Job Card”). Then perform the checksum (refer to page 4-83”Checksum Validation” Job Card).
6. Save Database on the service floppy delivered with the equipment (refer to page 4-80”Backup & Restore
Database Setup Parameters” Job Card).
9-62
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
None.
9-63
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
3
4
1 2
9-64
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 1
5 4
2
9-65
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
Ensure that U1 Standoff is tighten to the torque between 1.5 and 2 N.m (15 to 20 kg.cm).
CAUTION
Refer to section 5–3 4 .
Remount the AC/DC and Auxiliary Units Assembly inside the Cabinet.
CAUTION Be sure to place the “+” and “-” DC BUS in the right connector.
7 VERIFY
9-66
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
5.5 mm socket, with ratchet if possible (to remove cover and board)
10 mm socket with ratchet (to remove DC and rectifier cables)
Large Philips screw driver (to remove cables from the rectifier block).
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
1. Insure that the neon (orange) indicator light on the board has gone out.
2. Be careful touching components on this board; some become very hot.
3. To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
None.
9-67
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
Note: The AC/DC 3 Phases Power Assembly is made up of two main elements: AC/DC BOARD 3
Phases and RECTIFIER BLOCK 3 Phases.
3
4
1 2
9-68
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 2
9-69
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
1. Reverse the disassembly procedure to reinstall the AC/DC 3 PH Mid Power Assembly and the EMC Board.
2. Tighten the 10 mm cable retaining nuts to 3 N.m (30 kg.cm), and the 5.5 mm hex head screws to 1.5 to 2 N.m
(15 to 20 kg.cm).
Verify that the + and - DC Bus cables are not reversed to prevent damage to the Inverter
CAUTION 1
(Refer to section 5–3 ).
The Rectifier Block has to be placed with the two poles in regard of the AC/DC Board
CAUTION 1
(Refer to section 5–4 ).
Be careful to preserve the special grease that is spread between the Rectifier Block and
CAUTION the chassis. This grease, called: Conductive Surface Compound, provides both thermal
and electrical conductivity that is important in reducing failure of the rectifier.
3. Remount the AC/DC and Auxiliary Units Assy inside the Cabinet.
7 VERIFY
Reapply power and verify that the system functions correctly.
9-70
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
1. Insure that the Neon N1 (orange) indicator light on the LVPS Board has gone out.
2. Be careful touching components on the LVPS Board. Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
None.
9-71
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
1 2 3
1. Disconnect the 3 cables , ,
1 from the LV Power Supply 3 Phases
4
board .
4
2. Unscrew the six 5.5 mm hex head
screws and remove the LV Power
4
Supply 3 Phase Board .
3 2 9-72
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
Reverse the disassembly procedures to reinstall the LV Power Supply 3 Phases Board.
Verify that the following cables are not reversed before power ON (refer to section 5–1):
CAUTION
2
1. Cables CF2 (+DC BUS) and CF1 (- DC BUS) .
3
2. Cables CF4 (160 V) and CF3 (0 V) .
Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Remount the cover. Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
7 VERIFY
9-73
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
1. Insure that the Neon DS3 (green) indicator light on the Filament Board has gone out.
2. Be careful touching components on the Filament Board . Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
None.
9-74
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
1
1. Disconnect the ribbon cable
4
1 between the Filament Board and
5 2
the LVPS Board . Disconnect
3
and the connector J2 .
5
2 3
9-75
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 REASSEMBLY
2
Verify that cables CF4 (160 V) and CF3 (0 V) are not reversed before power ON (refer
CAUTION to section 5–2).
Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Remount the cover. Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
7 VERIFY
9-76
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
1. Insure that the Neon DS7 (orange) indicator light on the Rotor Board has gone out.
2. Be careful touching components on the Rotor Board. Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
None.
9-77
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
9-78
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1. The Auxiliary Unit being free, unscrew ten 5.5 hex head
screws and remove its cover.
1
2. Disconnect the ribbon cable between the Rotor
4 5
Board High Speed and the LVPS Board .
2
3
Disconnect cables CF1 and CF2 .
3. Disconnect Capacitor cables CF9, CF10, CF11, CF12
6 4
.
1
4. Unscrew two 5.5 mm hex head screws that are in holes
3
below the surface of the board and that secure an
IGBT.
5. Unscrew the eight 5.5 mm hex head screws and
4 5
remove the Rotor Board High Speed .
6 REASSEMBLY
Reverse the disassembly procedure to reinstall the Rotor Board High Speed.
Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
CAUTION Ensure that cables are correctly connected. Ensure that marks on the cables match the
marks on the board.
Remount the cover. Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
7 VERIFY
9-79
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 SUPPLIES
None.
2 TOOLS
3 SAFETY PRECAUTIONS
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
4 PREREQUISITES
THE WHOLE INVERTER UNIT WEIGHTS 30 KILOGRAMS (66 POUNDS). TWO PEOPLE
WARNING ARE REQUIRED TO CARRY IT.
9-80
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
5 DISASSEMBLY
Interface
Unit
9-81
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4
2. Disconnect the flat ribbon cable from J5 between
2 3
the I/F Board RAD and the AEC Board . 4
2
3
6 7
3. Disconnect the connectors J2 and J1 , and Test
5 5
Points TP1 and TP2 on the AEC Board.
Board 3 .
7
6 REASSEMBLY
Reverse the disassembly procedure. Tighten screws to torque between 1.5 and 2 N.m (15 to 20 kg.cm).
7 VERIFY
Reapply power.
Proceed to ”AEC diagnostics” to ensure correct operation.
9-82
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
CHAPTER 10 SCHEMATICS
SECTION 1 PROTEUS XR/a SYSTEM MIS MAP
1-1 System Schematics (1) (For systems with Wall Stand 2260354)
10-1
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
System Schematics (2) (For systems with SG100 Wall Stand 2225849)
X-ray
2365706
10-2
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-2 PDU Schematics
10-3
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-3 Wall stand Schematics
10-4
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10-5
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SG100 Wall stand Schematics (sheet 2 of 2)
10-6
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
S0005408
BRAZO - PCB
D D
L1 VERT. BRAKE 1
S4 3
2 B. UPPER
INTERFACE BOARD
GND[1..1]
1
CONTROL BOARD L2 VERT. BRAKE 2
S0005058 S5 3
J6[1..4] 2
1
2410108-* (MIS 27225) S0008838 S0007675 B. LOWER
J1 VAC[1..3] J3 [1..16] J11 [1..16] J7[1..4]
SW1 3
2
1 VERT. MOV.
S0008837
J8[1..2]
SW2-1 3
J13 [1..2]
2
1
L3 TILT. BRAKE 3 S1 3
2
S0004982 1 TILT. VERT.
J12[1..2]
S3 3
2
1
S0007674 TILT. HOR.
J10[1..6]
S0004978
S2-1 3
J14 [1..6]
2
S0007602 1
S2-2 3
2 BUCKY ROT.
1
B B
S0007651 S0007652
JP13 J13 D1 INDIC. ALIMENTACION
J13 [1..2]
1 1
2 2
BUCKY ASSY.
2145641-* (MIS 27221) S0007648
J42 [1..26] J42 [1..26] J4 BUCKY[1..6]
A A
2145642-* (MIS 27222)
J40 [1..14] S0007026
J40 [1..14] PL1 ION CH.[1..14]
S0005193 BUCKY
Title
CABLEADO W.S. SG 120
OPCION: J12 ALIMENTACION EN VDC.
Size Document Number Rev
* SEGÚN LONGITUD DE CABLE A3 S0009143SCH
1
Date: Monday, August 02, 2004 Sheet 1 of 1
5 4 3 2 1
10-7
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
J6
VERT. BRAKE COIL
1
2
+24V1 3
D 4 D
AMP280371-2
J7
BUCKY UPPER SW.
1
BUCKY LOWER SW. 2
3
4
AMP280371-2
J11
J8
1 VERT. MOVE SW.
2 1
3 0V1 2
4 AMP280370-2
5
6
7
C C
8
9
J9
10 TILT MOVE SW.
11 1
12 +24V1 2
13 3
14 4
15 +24V1
0V1 AMP280371-2
16
AMP280385-2 J12
TILT BRAKE COIL 1
+24V1 2
J13 AMP280370-2
1 +24V1 J10
2 0V1
BUCKY VERT. SW.
1
B B
AMP280370-2 BUCKY HOR. SW. 2
3
4
5
6
AMP280372-2
J14
BUCKY ROT. SW.
1
2
H1 24VDC EXT. 3
4
1 5
H2 6
1 AMP280372-2
H3
A 1 A
H4 SUINSA
1 Title
PWA INTERCONEXIÓN COLORADO
Size Document Number R ev
A4 S0007602SCH 1
Date: Wednesday, May 26, 2004 Sheet 1 of 1
5 4 3 2 1
10-8
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
J2 * 230Vca 0V1
3
AMP350210-1 J3
BUCKY UPPER 1
1
2
3
4
AMP350430-1
BUCKY LOW ER 2
VERT. MOVE 3
D1
TILT MOVE 4
D BUCKY VERT. D
FB1002L 5
F2 BUCKY HORIZ. 6
1 T1 5 +24V1 BUCKY ROTATED
BLUE/AZUL BLACK/NEGRO 7
J1 24VDC EXT. 8
1
0,5 A SB C6
VERT. BRAKE 9
470pF 400Vac RV3 F1 MF-R400
L GRAY/GRIS TILT. BRAKE 10
T
V130L10 2 6 RED/ROJO 11
N
DS1 4 8 4 - + 2 D3 + C3 + C4
12
LAMP NEON C8 VIOLET ORANGE/NARANJA
+24V1
2
47nF 250Vca 13
IEC 320 RV2 R5 SMCJ30A 2200uF 50V 2200uF 50V 14
BROWN/MARRON 3 15
V130L10 7 YELLOW/AMARILLO 0V1
3
230Vac +/- 10% 470 ohm 10W 16
R15 C7
50/60 Hz 0V1
270K 470pF 400Vac TRAFO TOROID 50VA AMP280385-2
R6
0 OHM
R7 2K2
J13
+24V1
1 1
2 2 0V1
AMP280370-2
J6
C 1 C
2 +24V1 TILT. BRAKE
3
4 R1
5 PUENTE R2 R3 R4
6 CR2
7 100 ohm 10W
8 S1J
9 C2 100uF 50V
10
+
11
4
Q3 Q1
12 MTD20N06HDL MTD20N06HDL
13 D2 C1 100uF 50V
14
+
15 TILT MOVE 1 1
16 R14 10K R9
3
C9 BAS85 R12 R13 4K7
AMP280385-2 JP1 100nF 2K2 D4 4K7
B BAS85
3 0V1
2
1 A
J41
PUENTE
8 R8 10K
15
7
14 24VDC EXT.
6
13 BUCKY LOW ER
B B
5
12 BUCKY UPPER
4
11 BUCKY ROTATED
3
10 BUCKY HORIZ.
2 +24V1
9 C5
BUCKY VERT.
1 1uF 250v
VERT. BRAKE
AMP 747845-3
CR1
R11 S1J
10K
4
RV1
J45 V130L10
5 Q2
9 1 FQD9N25
VERT. MOVE
4
8 R10
3
3 C10
7 100nF 4K7
2 0V1
6
1
A 21 A
11
AMP 747840-4 1 1
31 41
1 1
51 61
1 1 Title
10-9
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D D
8 TAB_SID_REF_RTN SW-1
R1 L7D 7.5K
3
10K LF444
7 TAB_SID 12 1-1 +12
2 14
6 TAB_SID_REF R5 13
C28
4
1 100K R15
10u 25V P5
10K L7A
+12 5K R18 3 LF444
RA10 1 R7 AI1
C 4.7K C
7.5K PA1 TP2 2 10K
5K R17
R16
10K
10K
11
+12
RA12 L7C
R10
10 LF444
120 PA3 C36 100K
8 -12
0.1u 50V
2K 9 2-1
SW-2
OCS_SID_REF
R3
10K
OCS_SID +12
2-2
OCS_SID_REF_RTN R8
C29
100K R4
10u 25V R85
120
TP3 3-2
0
+12
RA13
SW-3
B 120 L8B B
4
P4 L8A
5 3-1
7 100K R12 3
6 1 R19 AI0
R11 4.7K
2 10K
10K R14
10K
11
R13
C35 10K
-12
0.1u 50V
A A
Date 2004-10-29
Revision 3
Number 1/6
1 2 3 4
10-10
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D D
J111 +5
DB25
13
25 +12
12 SID_A
24 SID_B
4
6
8
11 SID_C
(J101)
23 SID_D
10 SID_E
22 DETENT VALID VERT_DETENT_IN D9 RLY
9 DS2E-M
1N4148
21 VERT_DETENT_1
8
2
3
4
5
6
7
8
20 VERT_DETENT_2
16
13
11
9
7 RES5
OCS CONTROL BOARD
19 ORDO_LOCK_RELEASE
C C
6
18 P21V R36 4.7K T1
5 7*10K
17 XT_OK_FOR_TOMO 2N708
1
4
16 OCS_SID_REF_RTN C25
3 10u 25V
15 OCS_SID
2
14 OCS_SID_REF
1
+5
EMI1
104Z
1 3
Vin Vout
GND
RA1 CTA1 CTA2
B B
10M/0.5W 470u/25v 470u/25v
2
+12
D101N4148
J114 VR11
6PIN KA7809A +12
1 3 +9V
Vin +9V
PDU (TM2)
C38
5 C26 100u/25V 0.1u/50V 100u/25V
DC0V 0.1u/50V
4 C27 C10
+12V
3 10u/25V 0.1u 50V
2 0.33u 50V
2
-12V
1
+5V
A A
-12
Date 2004-10-29
Revision 3
Number 2/6
1 2 3 4
10-11
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4
DB9 9 L5D
L3A L5A R66
5 TABLE_LAT_SIZE RES7-2 12 LF444
TABLE CASSETTE TRAY
R37 10K LF444 R39 LF444 R57 P10 10K
9 510 3 3 14 AI2
10K 10K 100K
4 1 1 13
8 TABLE_LONG_SIZE 2 2 P11
C13
3 100K
10u 25V R41
7 T_C R61
10K
2 TABLE_VOLT_REF 2K
11
11
R94 100K L3C
6
LF444 R44 L5B
1 TP5 10
-12 +12 10K LF444
-12 -12 8 5
R42
L3B 9 7
P6 R38 10K 10K
5 LF444 6
R32 R40 10K R93
VR9 R31 100K 7 C15 R62 R58
120 100K
9.1V C14 6 2K 10K
120 0.1u 50V 10u 25V TP6
R46
L3D 10K
VR7 VR6 R43 10K LF444 R45
9.1V 12
9.1V 10K
C P7 14 C
R33 100K 13
120 C16
0.1u 50V
TP11
L6C
+12
-12 LF444 R67
+12 10 AI5 TP12
10K
8
4
J113 +5 9 L6D
R68
4
DB9 L6A 12 LF444
R47 L4A R59 10K
WALL CASSETTE TRAY
C17 P13
8 WALL_LONG_SIZE 2
10u 25V R51 100K
3
10K
7 R91 R63
11
L4C
2 WALL_VOLT_REF 100k 2K
11
LF444 R54 L6B
6 10
10K LF444
1 TP9 -12 8 5
+12 R52
-12 9 7
R48 10K
L4B 6
10K R50 R60
P8 5 LF444 R92 R64
R34 10K C19 10K
100K 7 100K 2K
120
B C18 6 TP10 B
10u 25V R56
0.1u 50V
J124 L4D 10K
R53 10K LF444 R55
E N 12
P9 10K
R VR8 14
D A 9.1V 100K C20 13
P 0.1u 50V
C M
L
B K
H
F
BUCKY_VERT BUCKY_HORIZ BUCKY_ROTATE BUCKY_UPPER BUCKY_LOWER WALL_CASS_IN
A A
Date 2004-10-29
Revision 3
Number 3/6
1 2 3 4 5 6 7 8
10-12
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
D D
J115
DB25
CAN_H 1 J118
CAN_L 14 DB9
2 CAN_H 1
15 CAN_L 6
3 2
TOMO_STROKE_START 16 7
JEDI(J9)
4 3
ORDER_RX 17 PREP_SW 8
5 EXP_SW 4
ORDER_RX_RTN 18 9
6 5
P12V 19
TOMO
7
XT_OK_FOR_TOMO 20
C 8 C
ORDO_LOCK_RELEASE 21
9 J119
P21V 22 CAN_H 1
10 CAN_L 9
23 2
11 10
24 J120 PSW 3
12 1 PREP_SW 11
25 2 EXP_SW 4
13 PSW 3 12
DC12V 4 5
5 DC12V 13
6 6
J116
DC12V 14
DB9
6PIN 7
CAN_H 1
DC12V 15
CAN_L 6
8
2
CC2 CC1 7
B 3 DB15 B
22p 50V 22p 50V
8
XT
+9V 4
9
5
J123
DB9 ORDER_RX_RTN
1 ORDER_RX
6 P12V
CAN_L 2
CAN_H 7 TOMO_STROKE_START
3
8
OPTION
4
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
9
5
J117
A A
26PIN
JEDI(J1)
Date 2004-10-29
Revision 3
Number 4/6
1 2 3 4
10-13
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1 2 3 4
D D
2 7
C5 V5 CAP
0.1u 3 6 C34
V2.5 VBG R86
0.1u
A 4 5 0 A
COM STRB
Date 2004-10-29
Revision 3
Number 5/6
1 2 3 4
10-14
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
+5V
D D
1
RES3
+5V +5V
510*7
D1 D3
+5 RES3-8
2
3
4
5
6
7
8
R74 2K R73 2K
1
AI0 ADC0 AI1 ADC1
SID_A
RES1
U20 TLP521-4 D2 D4
R75 1K E1 C2 E6
SID_A 1 16 C1
SID_B
0.1u
2 15 1K*7 0.1u 10u/25V 10u/25V
SID_B R76 1K 3 14 U21
SID_E
SID_C
4 13 74LS373
2
3
4
5
6
7
8
C SID_C R77 1K 5 12 3 2 P40 C
P21V
SID_D
D0 Q0
6 11 4 5 P41
R78 1K D1 Q1
SID_D 7 10 7 6 P42 +5V +5V
D2 Q2
8 9 8 9 P43
D3 Q3
13 12 P44
D4 Q4
14 15 P45
D5 Q5 D16 D5
U19 TLP521-4 17 16 P46
R30 1K D6 Q6
SID_E 1 16 18 19 P47
D7 Q7 R72 2K R70 2K
2 15 AI4 ADC4 AI2 ADC2
TABLE_CASS_IN R21 1K 3 14 1
OE
4 13 11 D15 D6
R22 1K +5 LE E2 C4 E3
WALL_CASS_IN 5 12 C3
0.1u
6 11 0.1u 10u/25V 10u/25V
P21V R23 2K 7 10
XT_OK_FOR_TOMO 8 9
+5
+5V +5V
1
B B
RES2
U18 TLP521-4 D14 D7
BUCKY_VERT R24 1K 1 16 1K*7
2 15 R71 2K R69 2K
R25 1K AI5 ADC5 AI3 ADC3
BUCKY_HORIZ 3 14 U16
2
3
4
5
6
7
8
4 13 +5 74LS373
D13 D8
5 12 3 2 P10 C6 E7 C7 E8
D0 Q0
P12 6 11 VERT_DETENT_IN 4 5 P11 0.1u 0.1u
R26 1K D1 Q1 10u/25V 10u/25V
BUCKY_ROTATE 7 10 7 6 P13
D2 Q2
8 9 8 9 P14
D3 Q3
13 12 P15
D4 Q4
14 15
D5 Q5
U17 TLP521-4 17 16
R27 1K D6 Q6
BUCKY_UPPER 1 16 18 19
D7 Q7
2 15
R28 1K
BUCKY_LOWER 3 14 1
OE
4 13 11
R29 1K +5 LE
5 12 P21V
+5V
P17 6 11 DETENT_VALID
A A
7 10
8 9
Date 2004-10-29
Revision 3
Number 6/6
1 2 3 4
10-15
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10-16
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Table Logic Control Board (1)
10-17
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Table Logic Control Board (221-3010 for Rev0)
10-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Table Logic Control Board (221-3010 for Rev1 & Rev2)
10-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-8 Pedal I/F Board
10-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-9 Tomo-link System Schematics
a Cables Illustration
221-3010
10-21
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
b Interconnection.
System Console
TOMO-LINK SYSTEM INTERCONNECTION ILLUSTRATION
REA00057 TOWER
Tomo-Link I/F Board Tomo-Link Main
P1 J105
CAN
I/O PORTS
40DEG
22 22 40 DEG J2
J104
Sys. I/F
FAST FAST
1,2,14 10 10 J102
SLOW SLOW
*CAN-bus message content: 23 23 OPERATORS
Port 4
• Sweep angle:8,20,40degree; TOMO ON PANEL
•Sweep speed: fast, slow; 2 2
•Tomo ON 3 TOMO RDY
VCC (+5V) 3
•Tomo ready,Tomo start ; +24 TOMO
•Prep. 80C592 6 6 DRIVE UNIT
Isolated
CPU
DC-DC
+24 TOMO
with CAN 7 7
TOMO ON RTN
J102
on chip
15 15
J11 P2 TOMO RDY RTN
J100 J101
1,2,14 I/F Circuitry 16 16
5
Generator
3 I/
+12V 6 6 STROKESTART STROKE START
14 14
ORDO RX
4 4 ORDERRX RX EXP
11 11
ORDO RX RTN 5 5 ORDERRXR RX COM
24 24
ORDO LOCKR TOMO LOCK REL Bucky Lock
J101 25 25
Suspensio
Table Controller
13 13 1 4
16 +24V SEL TOMO
Sys. I/F 20 20 2 5
Board J3
MOTOR+ 1 1
MOTOR- P3
3 2
10-22 Disable table movement
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
1-10 Tomo I/F Board (1)
10-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Tomo I/F Board (2)
10-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 2
JEDI GENERATOR ILLUSTRATIONS
2-1 JEDI 50PLUS R 1T CABLING
Illustration - Jedi 50Plus R 1T Cabling (1/2)
10-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration - Jedi 50Plus R 1T Cabling (2/2)
10-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
2-2 JEDI LABELING PLAN
1 6
2 7
2244165
10-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
10-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration - I/F Board RAD V2 Electrical Schematics, Sheet 1
10-15
10-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration - I/F Board RAD V2 Electrical Schematics, Sheet 2
10-30
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration - I/F Board RAD V2 Electrical Schematics, Sheet 3
10-31
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration - I/F Board RAD V2 Electrical Schematics, Sheet 4
10-18
10-32
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 1
10-33
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 2
10-34
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 3
10-35
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 4
10-36
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 5
10-37
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 6
10-38
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 7
10-39
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 8
10-40
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet 9
10-41
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet10
10-42
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet11
10-43
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet12
10-44
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet13
10-45
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet14
10-46
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet15
10-47
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet16
10-48
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
Illustration- KV control board _V2C Sheet17
10-49
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 3
JEDI GENERATOR CENTRAL LISTING
CAUTION This board forms part of the oil seal of the High Voltage Tank. It can only be removed at
the factory. The Field Replaceable Unit (FRU) is the complete HV Tank.
LF
C
SF
10-50
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-2 FILAMENT BOARD 1 INV.
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS3 GOES OUT.
DANGER
DC
160v LF Common
SF
10-51
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-3 ROTATION BOARD HIGH SPEED
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS6 AND DS7
DANGER
(NEON-ORANGE) GO OUT.
Auxiliary
DC bus Common
Main
DS6
DC bus
10-52
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-3-1 Indicators
3-3-2 J2 Wiring
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
115V
sense
Fan, 115V AC
Line input
Fan*, 115V AC
Line X-ray tube
Thermal safety
10-53
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
DANGER
HIGH VOLTAGE AND HOT SURFACE: DO NOT TOUCH BOARD UNTIL INDICATOR NE1
(NEON - ORANGE) GOES OUT.
+ 400
s
- 400
160V
DC
0
10-54
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-5 GATE COMMAND BOARD
DANGER
HIGH VOLTAGE: DO NOT TOUCH BOARD UNTIL DS300 ON THIS BOARD AND DS1 ON
DUAL SNUB BOARD ARE OUT.
D
S3
00
Inverter ILP
current
feedback
ILR
* Insulated Gate
Bipolar Transistor
10-55
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
6 kV CONTROL BOARD
10-56
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SWITCH OR FUNCTION
JUMPER
RST RESET PUSH BUTTON
3-6-2 Indicators
REFERENCE DESIGNATION
U32 NON VOLATILE PARAMETERS RAM (BATTERY
INCLUDED)
10-57
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
DANGER
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL DS1 (NEON - ORANGE) GOES
OUT.
F1 - Fuse -
DS1 - Neon Protects (on DC Bus):
(Orange) - Indi LV Power Supply 3 Phase Board
cates presence Rotor Board High Speed
of voltage on the Gate Command Board
DC Bus. Type: 15 A, 600 VDC
10-58
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-8 I/F BOARD RAD V2
Table Bucky
Pin 1
Pin 9
Wall Bucky
Pin 1
To AEC board
Pin 9
Room Door & Light
Pin 1
DS4:CAN TRANSMIT
(yellow)
Service Laptop
DS5:EXT.CAN
TRANSMIT (yellow)
DS6: Receive Data
from console (yellow)
Console
OPTIONAL
Presence of strap -
TUBE 1
10-59
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-9 AEC BOARD V2
TESTPOINT MEASUREMENT
SIGNAL RANGE
TP1 230 VDC 230 VDC _5%
TP2 230 VDC 230 VDC _5%
TP3 230 VDC 230 VDC _5%
TP4 230 VDC 230 VDC _5%
TP5 AEC assignment reference 0 to 10 V
TP6 10 VDC (reference) 10 VDC _1%
TP7 AEC return voltage
TP8 AEC1 return voltage
TP9 AEC2 return voltage
TP10 AEC3 return voltage
TP11 AEC4 return voltage
10-60
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
3-10 QUAD SNUB BOARD
DANGER
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS1 (NEON -
ORANGE) GOES OUT.
10-61
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
SECTION 4 OTS CONSOLE ILLUSTRATIONS
4-1 Release Handle Schematics
SW1
2
1.3
ALL_IN 4
8
5.7
6
SW DPDT
SW2
2
1.3 ALL_OUT
4
8
5.7
ALL_VERT_IN 6
ALL_VERT_OUT
SW DPDT
J6
ALL_IN 1
ALL_VERT_IN 2
ALL_OUT 3
ALL_VERT_OUT 4
CONN_4
Title
ALL LOCK/RELEASE HANDLE
Size Document Number Rev
A
WDS-105-C-001
Date: Saturday, May 27, 2000 Sheet 1 of 1
10-62
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 VCC
DIRECTION 2273022-100
4-2 OTS Console Processor Board -1 R88
10K1%0.1W
U7
1 8
2 NC.1 CATHODE VCC VCC VCC VCC VCC VCC VCC VCC
VCC 3 NC.2
4 NC.3 C34 R132
5 NC.4
6 NC.5 SPARE_rc0805
R149 7 NC.6 4 R61 R60 R59 R58 R57 R56 R55 R46
NC.7 ANODE
10K1%0.1W 0.1uF50V SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805
U1D LM285-2.5
R5 R22 100 1%0.1W
9 8
SWITCH16
100K1%0.1W C1 VCC
74HC14
R150
0.1uF50V 10K1%0.1W
U1C R35
R19
VCC 5 6
SWITCH17
R49 R48 R47 R52 R51 R50 R54 R53
100K1%0.1W
R151 1001%0.1W SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805
74HC14
10K1%0.1W C4
0.1uF50V
U1A
R14 R34
1 2
SWITCH18 these pulldowns should not exceed 8k
100K1%0.1W 100 1%0.1W
74HC14 VCC
C9
0.1uF50V
R152
10K1%0.1W
U3E R182
R6 R32
11 10 U6
SWITCH21
VCC 40 R165
100K150.1W 100 1%0.1W VDD 8 100 1%0.1W 10K150.1W
VAREF 1 P3.4/T3EUD 9
74HC14 LED_I U3F
R155 C6 28 Varef P3.6/T3IN 10
VPP VPP/_EA P3.6/T3IN _ENABLE_LED R28 R11
10K150.1W 0.1uF50V 11
P3.9/MTSR P3.9
MUX_SIG0 37 12 KEYSW_DB 12 13
POH.0/AD8 P3.10/TXD0 TXD KEYSW
U3B R24 MUX_SIG1 38 13
POH.1/AD9 P3.11/RXD0 RXD
R8 MUX_SIG2 39 14 100K1%0.1W
POH.2/AD10 P3.12/_BHE/_WRH P3.12
3 4 MUX_SIG3 42 15 100 150.1W
SWITCH23 POH.3/AD11 P3.13/SCLK P3.13
MUX_SIG4 43 16 74HC14 C26
POH.4/AD12 P3.15/CLKOUT P3.15 0.1uF50V
100K1%0.1W 100 1%0.1W MUX_SIG5 44 VCC VCC
45 POH.5/AD13 17 VCC
74HC14 MUX_SIG6 P4.0
C11 VCC 46 POH.6/AD14 P4.0/A16_CS3 18 R42
DE10_DB MUX_SIG7 POH.7/AD15 P4.1/A17/_CS2 P4.1 CAN_RXD
0.1uF50V 19 SPARE
SW9_DB P4.2/A18/_CS1 P4.2 CAN_TXD
SW8_DB MUS_SELECT0 29 22 R39 FB2 R166 R97
POL.0/AD0 P4.3/A19/_CS0 P4.3 4.99K150.1W 4.99K1%0.1W
R156 SW22_DB MUS_SELECT1 30 23 U4 C19
10K150.1W 31 POL.1/AD1 P4.5/A20/CAN_RXD 24 1 3 0.1uF50V
SW23_DB MUS_SELECT2 POL.2/AD2 P4.6/A21/CAN_TXD TXD VCC CAN+
SW21_DB MUS_SELECT3 32 _NMI VPP
33 POL.3/AD3 76 100 1%0.1W 8 7 BLM21A601S
U3C R33 SW20_DB P0L.4 V5REF
VCC R17 34 POL.4/AD4 P5.0/AN0 77 RS CANH L1
SW19_DB P0L.5 POL.5/AD5 P5.1/AN1 V21REF FB1
5 6 SW18_DB P0L.6 35 78 R41 4 6
SWITCH22 POL.6/AD6 P5.2/AN2 P5.2 RXD CANL PLM3216K281SJ2
SW17_DB P0L.7 36 79 475150.1W
POL.7/AD7 P5.3/AN3 P5.3 _ST CAN-
R157 100K150.1W 100 1%0.1W 2 2 5 VCC
SW16_DB P5.4/AN4/T2EUD P5.4 _RSTIN GND VREF
10K1%0.1W 74HC14 59 3 BLM21A601S
P1H.0/_CC6POS0/EX0IN P5.5/AN5/T4EUD P5.5
C12 62 4 PCA82C250
0.1uF50V P1H.1/_CC6POS1/EX1IN P5.6/AN6/T2IN P5.6
63 5 R40 R44
P1H.2/_CC6POS2/EX2IN P5.7/AN7/T4IN P5.7
U2C R25 64 R184 SPARE
R7 65 P1H.3/EX3IN/T7IN 72
5 6 66 P1H.4/CC24IO P8.0/CC16IO 73
SWITCH8 P1H.5/CC25IO P8.1/CC17IO LED_ELASH
67 74 100 1%0.1W 100 150.1W C50 C29 C31 C28 C30 +21V
P1H.6/CC26IO P8.2/CC18IO PWM_INC 0.01uF50V 0.01uF50V
100K150.1W 100 1%0.1W 68 75 0.01uF50V 0.01uF50V 0.01uf50V VCC
P1H.7/CC27IO P8.3/CC19IO TXEN_INC
74HC14 VCC DN1
C3 47 71 BAT54S
0.1uF50V P1L.0/CC60 _NMI _NMI
48 69
R158 49 P1L.1/COUT60 _RSTIN 70 R91 1
P1L.2/CC61 _RSTOUT _RSTOUT
10K1%0.1W 50 25 VCC 100K1%0.1W R90
P1L.3/COUT61 _RD _RD ? 10K1%0.1W
51 26 U5
P1L.4/CC62 _WR_WRL _WR
U2D R29 52 27 _PBRST 2 15
P1L.5/COUT62 ALE ALE _RESET VCC
R16 57 13
VCC 9 8 58 P1L.6/COUT63 VCC 6 _ST 11 D7
SWITCH9 P1L.7/_CTRAP TOL _RST V21REF V5REF
SW15_DB 55 VCC 4 9 C25 R89
XTAL1 XTAL1 TD RST RST_WD 0.1uF50V UTP5 5V
100K150.1W 100 1%0.1W SW14_DB VSS (6,20,41,56,60) 80 10K1%0.1W
R159 VAGND R4 VCC 16 R98 C36
74HC14 SW13_DB
10K1%0.1W C2 60 54 R173 1 NC-16 14 C35 4.99K150.1W 0.1uF50V
SW12_DB VSS XTAL2 NC-1 NC-14
0.1uF50V R43 3 12 0.1uF50V
SW11_DB 1K1%0.1W NC-3 NC-12
U2B 1.00Meg1% SPARE 5 10
R13 R30 R45 7 NC-5 NC-10
C164CI SPARE NC-7
3 4 8
SWITCH10 U9 GND
Y1
100K1%0.1W 100 150.1W R1 10 MHz DS1232
74HC14 VCC XTAL2 3 1
C8 VCC OSC EN R171
0.1uF50V 0 1K1%0.1W
10 MHz R170
R160 C32 C33 1K150.1W
10K150.1W + C27 C51 C52 C53 C54 18pf50V 10pf50V
47uF16V 0.01uF50V 0.01uF50V 0.01uF50V 0.01uF50V
U2A R26
R21 VCC
VCC 1 2
SWITCH11
100K150.1W 100 1%0.1W place caps near microcontroller
R161 74HC14
10K1%0.1W C7
0.1uF50V R77 R175 R79 R66 R67 R68 R69 R70 R71 R176 R177 R82 R83 R72 R73 R74 R75 R178 R179 R180
U2E R36
R18 10K1%0.1W SPARE_rc0805 10K150.1W 10K1%0.1W 10K150.1W 10K150.1W 10K150.1W 10K150.1W 10K150.1W SPARE_rc0805 SPARE_rc0805 10K1%0.1W 10K1%0.1W 10K1%0.1W 10K1%0.1W 10K1%0.1W 10K1%0.1W SPARE_rc0805 SPARE_rc0805 SPARE_rc0805
11 10
SWITCH12
100K1%0.1W 100 150.1W P3.15
P3.8
74HC14
C17 VCC P3.9
0.1uF50V P3.12
P3.13
R162 P4.0
10K1%0.1W P4.1
P4.2
P4.3
U2F
R10 R27 P5.0
13 12 P5.1
SWITCH13 P5.2
VCC
100K1%0.1W 100 150.1W P5.3
P5.4
74HC14
R163 C18 P5.5
10K1%0.1W 0.1uF50V P5.6
P5.7
P8.1
U1E R37
R9 P8.2
11 10 P8.3
SWITCH14
100K1%0.1W R112 R113 R114 R115 R116 R117 R118 R119 R120 R121 R122 R123 R124 R125 R126 R127 R128 R129 R130 R131
74HC14 100 1%0.1W
C23 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805
0.1uF50V VCC
R164
10K150.1W
VCC VCC VCC VCC VCC VCC VCC VCC
U1F R38
R12
13 12
SWITCH15
100K1%0.1W 100 150.1W R137 R136 R135 R134 R133 R148 R147 R138
C24 74HC14
0.1uF50V SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805
MUX_SELECT0
MUX_SELECT1
MUX_SELECT2
MUX_SELECT3
POL.4
VCC POL.5
U3A POL.6
R172 POL.7
1 2
TP17
R141 R140 R139 R144 R143 R142 R146 R145
1K1%0.1W C20 C21 C22
0.1uF50V 0.1uF50V 0.1uF50V SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 SPARE_rc0805 Title
74HC14
spare GE MEDICAL
Size Document Number Rev
D 1.0
place caps near schmitt triggers OCS CONSOLE PROCRSSOR BOARD
these pulldowns should not exceed 8k Date: Tuesday, February 06, 2001 Sheet 1 of 2
10-63
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-3 OTS Console Processor Board -2 VCC
C45
VCC 0.1uF50V
+21V
ALL LOCK/RELEASE LONG TRANS
J4 FB3 FB4
SW3 SW4 SW5
SWIICH8 1 2
3 4 SWIICH9 ALL_INH LONG_SIG TRANS_SIG
SWIICH10 ALL_OUTH
5 6 SWIICH11 ALL_OUTH
SWIICH12 BLM21A601S
7 8 SWIICH13 VERT_SIG
SWIICH14 VCC VCC VCC BLM21A601S
9 10 SWIICH15
SWIICH16
11 12 SWIICH17
SWIICH18
13 14 SWIICH19
SWIICH20
15 16 SWIICH21
SWIICH23
17 18 SWIICH22
SW DPDT SW DPDT SW DPDT
19 20
CONN_20
R101 R102 R104
474 1% 0.1W 475 1% 0.1W 475 15 0.1W
VCC J2
LONG_SIG
R93 R95 TRANS_SIG 1
C38 C39 100K 1% 0.1W C41 100K 1% 0.1W DET_SIG 2
C44 VCC 0.1uF50V 0.1uF50V 0.1uF50V V_SIG 3
0.1uF50V D1 D2 D4 ANGLE_SIG 4
5
HLMP-15-1540 HLMP_1540 HLMP_1540 TP1 6
TP8 7
J5
TP9 8
RST_WD 1 2 TP10 9
MUX_SELECT0 3 4 _ENABLE_LED TP11 10
MUX_SELECT2 5 6 MUX_SELECT1 TP12 11
7 8 MUX_SELECT3 12
MUX_SIG0
MUX_SIG2 9 10 MUX_SIG1 THESE COMPONENTS WILL BE 13
MUX_SIG4 11
13
12
14
MUX_SIG3
MUX_SIG5
LOCATED ON A SEPARATE PCB 14
15
MUX_SIG6
LED_FLASA 15 16 MUX_SIG7 IN THE CONSOLE"S HANDLE" 16
17 18 LED_I 17
19 20 18
TP2 19
CONN_20
TP3 20
SW1
TP4 21
+21V TP5 22
TP6 23
24
25
C47 CONN_25
ALL_OUT R181
0.1uF50V
ALL_IN SW DPDT
4.7uH HI18127800R
J6 R174
ALL_IN 1
ALL_VERT_IN 2
ALL_OUT 3 SPARE_RC0805
ALL_VERT_OUT 4
J1
CONN_4 CAN+
CAN- 1
2
TP7 3
U6-2
VAREF TP13 4
P5.5 1 2 P5.4 TP14 5
3 4 P5.6 R167 6
P5.7
VDD 5 6 VSS 7
LED_I 7 8 KEYSW_DB 0_OHM 8
P3.9 9 10 _ENABLE_LED 9
RXD 11 12 TXD 10
P3.13 13 14 P3.12 TP15 11
P4.0 15 16 P3.15 TP16 12
P4.2 17 18 P4.1 13
19 20 VSS 14
15
CONN_20
PD5 CONN_15
PD7
PD6
U6-3 PD9
VDD
CAN_RXD 1 2 P4.3 +21V
VERT +21V
DETENT +21V
ANGLE PD8
PD2
3 4 CAN_TXD VERT_SIG
RD PD1 +21V
ALE 5 6 WR PD4
SW6 SW7 SW8
MUX_SELECI0 7 8 VPP PD3
9 10 MUX_SELECI1 FB5 FB6 FB7
MUX_SELECI2
POL.4 11 12 MUX_SELECI3
13 14 POL.5 V_SIG DET_SIG ANGLE_SIG
POL.6 VCC VCC VCC C46
MUX_SIG0 15 16 POL.7 0.1uF50V
MUX-SIG2 17 18 MUX_SIG1 BLM21A601S BLM21A601S BLM21A601S
19 20 VDD
CONN_20 J3
SW DPDT SW DPDT SW DPDT ALL_INH
1
ALL_OUTH 2
R103 R106 R105 VERT_SIG 3
U6-4
VSS 475 1% 0.1W 475 1% 0.1W 475 1% 0.1W 4
MUX_SIG4 1 2 MUX_SIG3
CONN_4
MUX_SUG6 3 4 MUX_SIG5
SW8_DB 5 6 MUX_SIG7 R92 R94 R96
SW10_DB 7 8 SW9_DB C40 100K 1% 0.1W C43 100K 1% 0.1W C42 100K 1% 0.1W
SW12_DB 9 10 SW11_DB 0.1uF50V 0.1uF50V 0.1uF50V
VDD 11 12 SW13_DB D3 D6 D5
XTAL1 13 14 XTAL2
J7
SW14_DB 15 16 VSS HLMP-1540 HLMP-1540 HLMP-1540 _PBRST
SW16_DB 17 18 SW15_DB 1
19 20 VSS 2
CONN_20 CONN_2
JU6-5
VDD
SW18_DB 1 2 SW17_DB
J8
SW20_DB 3 4 SW19_DB KEYSW
SW23_DB 5 6 SW21_DB 1
_RSTIN 7 8 SW22_DB 2
_NMI 9 10 _RSTOUT
CONN_2
LED_FLASH 11 12 _ST
TXEN_INC 13 14 PWM_INC
V2IREF 15 16 V5REF
P5.3 17 18 P5.2
19 20 VSS
CONN_20
Title
GE MEDICAL
Size Document Number Rev
C OCS CONSOLE PROCESSOR BOARD 1.0
10-64
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-4 OTS Console Display Board -1
VCC
VCC VCC VCC VCC
VCC
R5 R9 R13 R17
Q31 C11 Q30 C16 Q27 C15 Q29 C13
FDV304P 0.1uF50V FDV304P 0.1uF50V FDV304P 0.1uF50V FDV304P 0.1uF50V
R4 100 1%0.1W R8 100 1%0.1W R12 100 1%0.1W R16 100 1%0.1W
Q4 Q6 Q8 Q10
MUX_SIG7 FDV303N MUX_SIG5 FDV303N MUX_SIG3 FDV303N MUX_SIG1 FDV303N R63 R64 R65 R66 R67 R68 R69 R70
100 1%0.1W 100 1%0.1W 100 1%0.1W 100 1%0.1W 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1%
+ C3 + C2 + C4 + C5 + C6
100uF16V 100uF16V 100uF16V 100uF16V 100uF16V
R1
Q1
LED_GROUP1 FDV303N
100
R32
100K
PROT
R91 R90 R89 R88 R87 R86 R85 R84 R107 R106 R105 R104 R103 R102 R101 R100
20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% KV DISPLAY 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% U2B
U4 ANGLE DISPLAY
11 12 11 13
7 A CATHODE_1 9 10 A CATHODE
4 B CATHOED_2 8 8 B
2 C CATHODE_3 R20 R21 R22 6 C
1 D Q12 Q13 Q14 5 D
E KV_LS FDV303N KV_MID FDV303N KV_MS FDV303N E
10 12
5 F 100 100 100 7 F
G R51 R52 R53 G
3 100K 9
DP 100K 100K DP R71 R72 R73 R74 R75 R76 R77 R78
LTC-571G LTD-6440G 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1%
PROT PROT PROT
R24
Q16
ANGLE_LS FDV303N D39 D40 D38 D37 D15 D14 D13 D12
100
R55 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540
100K
PROT
R99 R98 R97 R96 R95 R94 R93 R92 R115 R114 R113 R112 R111 R110 R109 R108
20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% mAs DISPLAY 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% U2A
U5 ANGLE DISPLAY
11 12 16 14 R2
7 A CATHODE_1 9 15 A CATHODE
4 B CATHOED_2 8 3 B Q2
C CATHODE_3 R23 C LED_GROUP2
2 R27 R26 2 FDV303N
1 D Q18 Q19 Q15 1 D 100
E MA_LS FDV303N MA_MID FDV303N MA_MS FDV303N E
10 18 R34
5 F 100 100 17 F 100K
G R57 R58 100 R54 G
3 100K 100K 4
DP 100K DP PROT
U9A R143
LTC-571G LTD-6440G 1
PROT PROT PROT LED_GROUP2
3 Q34
2 FDV303N
LED_FLASH
R25 100
14081 R148
Q17 100K
ANGLE_MID FDV303N
PROT
100 R56
R123 R122 R121 R120 R119 R118 R117 R116 R128 R127 R126 R125 R124 100K
20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% 20 1% SID DISPLAY 20 1% 20 1% 20 1% 20 1% 20 1%
PROT
U6 U3
11 12 6 4
7 A CATHODE_1 9 3 B CAT_B/C/DP 7
4 B CATHOED_2 8 9 C CAT_B/C/DP
2 C CATHODE_3 R28 R29 R30 1 +
D -
1
E SID_LS
Q20
SID_MID
Q21
SID_MS
Q22
CAT_+/-
2 ANGLE DISPLAY
10 FDV303N FDV303N FDV303N 5 8
5 F 100 100 100 DP CAT_+/- R79 R81 R83 R82 R80
G R59 R60 R61 20 1% 20 1% 20 1% 20 1% 20 1%
LTS-6495G
3 100K 100K 100K
DP
LTC-571G R31
PROT PROT PROT
Q23
ANGLE_OVER FDV303N D18 D19 D20 D17 D16
100
R62 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540 HLMP-1540
100K
PROT
VCC VCC VCC
R146 100K
PROT R3
R149 R145 Q3
VCC LED_GROUP3 FDV303N
100K 100K VCC 100
R140 R33
C24 VCC
8
FET_OFF
R151 R147
10K 1%0.1W
U8 100K
1 8
2 NC.1 CATHODE
3 NC.2 Title
NC.3
4
5 NC.4 C8
R152
GE MEDICAL
6 NC.5 0.1uF50V SPARE Size Document Number Rev
NC.6
7
NC.7 ANODE
4 D
OCS CONSOLE DISPLAY BOARD 1.0
10-65
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 19 DIRECTION 2273022-100
4-5 OTS Console Display Board –2
VCC
C17
0.1uF50V
VCC
SW8 SW16 J4
1 2
SWITCH8 3 4 SWITCH9
KSIOM410 VCC KSIOM410
SWITCH10 5 6 SWITCH11
SWITCH12 7 8 SWITCH13
SWITCH14 9 10 SWITCH15
SWITCH16 11 12 SWITCH17
SWITCH8 SWITCH16 SWITCH18 13 14 SWITCH19
C19
0.1uF50V SWITCH20 15 16 SWITCH21
SWITCH23 17 18 SWITCH22
19 20
SW9 SW17 CONN_20
KSIOM410 KSIOM410
SWITCH9 SWITCH17
VCC
C18
KSIOM410 VCC KSIOM410 0.1uF50V
SWITCH10 SWITCH18 J5
C20
0.1uF50V 1 2
RST_WD 3 4 ENABLE_LED
MUX0 5 6 MUX1
MUX2 7 8 MUX3
SW11 SW19
MUX_SIG0 9 10 MUX_SIG1
MUX_SIG2 11 12 MUX_SIG3
MUX_SIG4 13 14 MUX_SIG5
KSIOM410 KSIOM410 MUX_SIG6 15 16 MUX_SIG7
LED_FLASH 17 18 LED_I
19 20
SW12 SW20
VCC
KSIOM410 KSIOM410
C21 J3
SWITCH12 0.1uF50V SWITCH20
1
FET_OFF 1
2
2
SW13 SW21 CONN2
KSIOM410 KSIOM410
SWITCH13 SWITCH21
SW14 SW23
KSIOM410 KSIOM410
VCC
SWITCH14 SWITCH23
C22
0.1uF50V
SW15 SW22
KSIOM410 KSIOM410
Title
SWITCH15 SWITCH22
GE MEDICAL
Size Document Number Rev
C
OCS CONSOLE DISPLAY BOARD 1.0
10-66