This document outlines the procedure and calculations for determining the uncertainty in thermocouple measurements. It measures the temperature using a thermocouple five times at a setpoint of 1200°C and standard deviation of the readings is calculated. Various sources of uncertainty are identified and calculated including the uncertainty of the standard, indicator, dry block stability, drift of the master, and inhomogeneity of the thermocouple. The combined uncertainty is calculated to be 0.80°C and with a coverage factor of 2, the expanded and final uncertainty is determined to be +/- 1.60°C.

Uploaded by

A kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
132 views1 page

cmc7 PDF

This document outlines the procedure and calculations for determining the uncertainty in thermocouple measurements. It measures the temperature using a thermocouple five times at a setpoint of 1200°C and standard deviation of the readings is calculated. Various sources of uncertainty are identified and calculated including the uncertainty of the standard, indicator, dry block stability, drift of the master, and inhomogeneity of the thermocouple. The combined uncertainty is calculated to be 0.80°C and with a coverage factor of 2, the expanded and final uncertainty is determined to be +/- 1.60°C.

Uploaded by

A kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Thermocouple

Uncertainty Procedure
UNCERTAINITY MEASUREMENT Range of UUC/set point 1200
(The values having ( ) right to them are variables Least count of UUC 0.01
TYPE - A Uncertainity of STD. 1.45
0
C Uncertainity of indicator 0.06
1st Reading 1201.65 1196.44 -5.21
2nd Reading 1201.71 1196.47 -5.24
3rd Reading 1201.62 1196.46 -5.16
4th Reading 1201.68 1196.43 -5.25
5th Reading 1201.59 1196.41 -5.18

Average -5.21
Standard Deviation 0.038340579
Total readings (n) 5
Under Root (n) 2.236067977
0
Uncertainity Ua( Under Test ) 0.01715 C
TYPE - B
0
Uncertainity due to master 1.45 C
Normal Distr. K=2 2 R-type
0
U1= 0.725 C
0
Uncertainity due to master 0.00 C Master Indicator
Normal Distr. K=2 2
0
U2= 0.0000 C
0
Uncertainty due to 0.3 C Dry Block-Stability
Stability of Dry Block 1.73205
Rect.Dist. K=1.73205 0.173205162
0
Uncertainty due to 0.36 C Uniformity
Stability of Dry Block 1.73205
Rect.Dist. K=1.73205
0
U3= 0.173205 C
0
Uncertainty due to 0.04 C Drift of Master
Drift in Master 1.73205
Rect.Dist. K=1.73205
0
U4= 0.023094 C
0
Uncertainty due to 0.06 C Uncertainity of indicator
Indicator 1.73205
Rect.Dist. K=1.73205
0
U5= 0.034641 C
0
Uncertainty due to LC of uuc 0.01 C Least Count of Indicator
Rect.Dist. K=1.73205 1.73205*2
0
U6= 0.00289 C
0
Uncertainty due to Variation Error 0.10 C Variation error in the Ice
Rect.Dist. K=1.73205 1.73205 Point
0
U7= 0.05774 C
0
Uncertainty due to In homogeneity 0.240 C In Homogeneity of
Rect.Dist. K=1.73205 1.73205 Thermocouple
0
U8= 0.13856 C

0
Combin Uncertainity Uc 0.80 C

Effective Degrees of freedom(v) 18966564.6115

Coverage Factor 2

0
Expanded Uncertainity (+) 1.60 C

0
Final Uncertinity (+/-) 1.60 C