Starrett 3810A
Starrett 3810A
Starrett 3810A
Operation Manual
1
6.5.6 Hardness/бb Setting ..................................................................................................... 16
6.6 Print Function ................................................................................................................. 16
6.6.1 Print Current ................................................................................................................ 16
6.6.2 Print Memory ............................................................................................................... 17
6.6.3 Print All Memory ......................................................................................................... 17
6.6.4 Paper Feeding .............................................................................................................. 17
6.7. Memory Manager .......................................................................................................... 17
6.7.1 View from No.1 Group/View from Ending Group ...................................................... 17
6.7.2 View from Selected No. Group .................................................................................... 18
6.7.3 Data Transfer ............................................................................................................... 18
6.7.4 Delete by Group No. .................................................................................................... 18
6.7.5 Delete All Data............................................................................................................. 18
6.7.6 Deletion Confirmation ................................................................................................. 18
6.8 Browsing Memory Data Groups ..................................................................................... 19
6.9 System Set ...................................................................................................................... 20
6.9.1 LCD Brightness Set ..................................................................................................... 21
6.9.2 Time Date Set .............................................................................................................. 21
6.10 Software Information .................................................................................................... 21
6.11 System Calibration ........................................................................................................ 22
6.12 EL Background Light ................................................................................................... 22
6.13 Auto Power Off ............................................................................................................. 22
6.14 Paper Loading ............................................................................................................... 23
6.15 Battery Charge .............................................................................................................. 23
6.16 Battery Replacement ..................................................................................................... 23
6.17 Connection of Data Transmission Cable....................................................................... 24
7 Fault Analysis & Evacuation .................................................................................... 24
8 Servicing & Maintenance ......................................................................................... 24
8.1 Impact Device Servicing ................................................................................................. 24
8.2 Normal Maintenance Program ........................................................................................ 24
9 Calibration................................................................................................................. 24
10 Notice of Transport and Storage Conditions........................................................... 24
APPENDIX .................................................................................................................. 25
Table 1 .................................................................................................................................. 25
Table 2 .................................................................................................................................. 26
Table 3 .................................................................................................................................. 27
Table 4 .................................................................................. Error! Bookmark not defined.
2
1 Overview
1.1 Advantages
Wide measuring range. Based on the principle of Leeb hardness testing theory. It
can measure the Leeb hardness of all metallic materials.
Large screen(128×64 dot matrix LCD), showing all functions and parameters.
Test at any angle, even upside down.
Direct display of hardness scales HRB(Rockwell B), HRC(Rockwell C),
HRA(Rockwell A), HV(Vickers), HB(Brinell), HS(Shore), HL(Leeb).
Seven impact devices are available for special application. Automatically identify
the type of impact devices.
Large capacity memory could store 500 groups (Relative to average times32~1 )
information including single measured value, mean value, testing date, impact
direction, impact times, material and hardness scale etc.
Upper and lower limit can be preset. It will alarm automatically when the result
value exceeds the preset limit.
Battery symbol indicates the remaining capacity of the battery and the charge
status.
User calibration function.
Software to connect with PC via USB port.
With EL background light.
Integrated thermal printer, convenient for in field printing.
NI-MH rechargeable battery as the power source. Continuous working period of
no less than 150 hours (Backlight and Printer off).
Auto power off to save energy.
Overall dimensions:212mm×80mm×32mm
3
1.3 Configuration
Table 1-1
No. Item Quantity Remarks
Standard 1 Main Unit 1
Configuration 2 D type impact device 1 With cable
3 Standard test block 1
4 Cleaning brush (I) 1
5 Small support ring 1
6 Battery Charger 1 9V 500mA
7 Paper for printing 1
8 Manual 1
9 Instrument case 1
Optional 11 Cleaning brush (II) 1 For use with G
Configuration type impact device
12 Other type of impact Refer to Table 3
devices and support and Table 4 in the
rings appendix.
13 DataPro software 1
14 Communication cable 1
15
16
4
2 Structure Feature &Testing Principle
1. Main unit
2. Impact device
4
9
3 10 HL0108070811
5
PHASEⅡ
HARDNESS TESTER
1 7
MODEL: PHT-3500
BATTERY: 6V NI-MH
CHARGER: 9V DC
CAUTION:
2 8
1 Do not disassemble
2 Do not dispose of in fire or water
OFF---ON
6
MAT SCALE
5
2.1.3 D Type Impact Device
1 2 3 4 5 6 7
DC C D D+15 DL G
6
3 Technical Specifications
Error and repeatability of displayed value see Table 3-1.
Table 3-1
Type of impact Hardness value of Leeb Error of displayed
No.
device standard hardness block value
760±30HLD ±6 HLD
1 D
530±40HLD ±10 HLD
760±30HLDC ±6 HLDC
2 DC
530±40HLDC ±10 HLDC
878±30HLDL
3 DL ±12 HLDL
736±40HLDL
766±30HLD+15
4 D+15 ±12 HLD+15
544±40HLD+15
590±40HLG
5 G ±12 HLG
500±40HLG
822±30HLC
6 C ±12 HLC
590±40HLC
Measuring range:HLD(170~960)HLD
Measuring direction:360°
Hardness Scale:HL、HB、HRB、HRC、HRA、HV、HS
Display:dot matrix LCD,128×64 dots
Data memory: 500 groups max.(relative to impact times 32~1)
Printing paper: width is (57.5±0.5)mm, diameter is 30mm
Battery pack: 6V NI-MH
Battery charger: 9V/500mA
Continuous working period:approx. 150 hours(With backlight off, no printing)
Communication interface:USB1.1
7
4 Preparation & Testing
The sample being tested must be close to 1” thick of solid material and should
conform to Table 3.
For the sample with hardened layer on surface, the depth of hardened layer
should conform to Table 3.
Coupling. Light-weight sample must be firmly coupled with a heavy base plate.
Both coupled surface must be flat and smooth. The impact direction must be
vertical to the coupled surface. When the sample is a large plate or long rod it can
be deformed and become unstable, even though its weight and thickness meets
minimum standards, and accordingly, the test value may not be accurate. So the
sample should be reinforced or supported at its back.
Magnetism of the sample itself should be avoided.
8
4.2 Testing Program
Verification of the 3810A is performed by testing on the supplied standard test block.
The error and repeatability of displayed value should be within the noted tolerance
shown in Appendix table 2.
4.2.1 Start-Up
Insert the plug of the impact device into the socket of impact device on the tester.
Press【 】key,to power the unit on. The instrument is in testing condition.
4.2.2 Loading
Pushing the loading-tube downwards until contact is felt. Then allow it to slowly
return to the starting position or using other method locking the impact body.
4.2.3 Placement
Stand the impact device supporting ring on the surface of the sample firmly, the
impact direction should be vertical to the testing surface.
4.2.4 Testing
Gently press the release button on the top of the impact device to take a test. The
sample and the impact device as well as the operator are all required to be stable
at this point. The action direction should pass the axis of the impact device.
Part should be tested a minimum of 5 times. The results should not more than
mean value±15HL.
The distance between any two impact points or from the center of any impact
point to the edge of testing sample should conform to the regulation of Table 4-1.
9
Table 4-1
D、DC 3 5
DL 3 5
D+15 3 5
G 4 8
C 2 4
5: Notes
Replacing the impact device must be done while the unit is powered off.
Otherwise the main body cannot identify the type of the impact device.
You could not save the current test value if the test times are less than the
presetting times value.
(ie: average set at 5 times. You must take minimum of 5 tests to save the values)
Only type D and type DC of impact device have the function of strength measure
option. You cannot modify the 【Set hardness orбb】setting when using other
types of impact device. The 【Set hardness orбb】setting would be set to
【Hardness】 automatically after replacing the impact device whether the setting
is 【Hardness】 or not before.
Not all materials could convert to all hardness scales. The hardness scale is reset
to HL automatically after changing the material. So select material first before
changing the hardness scale.
10
6 Operation in Details
6.1 Power On
3810A
HARDNESS TESTER
STARRETT
Probe Type:D
The system will automatically detect the type of the impact device during power up,
and would display this information on the screen. Users should pay attention to the
probe type displayed on the screen. After pausing for several second, the screen will
exit and enter the main display interface.
Note: If the instrument is powered off, it will turn on automatically after the
charger is connected.
6.3 Testing
Below is the main display interface:
Impact direction
Times count
Material
11
6.3.1 Instruction of the Main Display Interface
Battery symbol:Display the information of the remaining capacity of the battery, and
the charging status.
Impact direction:The present impact direction.
Average value indicator:It appears to show the mean value of the samples when
reaching the preset test times.
Hardness scale:Hardness scale of the present measured value.
Measured value:Display present single time measured value(without mean value
indicator ), or display the present mean value (with average value indicator
prompting). means over conversion or measure range. means lower than
conversion or measure range.
Material:The present preset material.
Impact times count: Amount of tests taken.
12
Press key 【MAT】to change the material set. Presetting hardness scale defaults to
HL automatically after material presetting changed.
13
6.5 Test Set
At the main display interface, press key【MENU】 to enter the main menu.
Press key 【ENTER】 to enter Test Set
Test Set Menu.
The symbol ↓ at the left side of
Print Set underside menu indicates that the menu
Memory Manager has not ended. Press key 【 】 could
continuously glance downward. The
System Set
symbol ↑ at the left side of the upside
menu indicates that the menu has not
ended. Press【 】could continuously
glance upward.
Press key【 】or【 】to move the
Impact Direc. cursor to the line you want to set, and
press key【ENTER】to confirm it.
Average Note:1. When 【Hard/бb】is switched
Material to бb, the hardness scale could not be
selected. The cursor will skip over
Hardness Scale 【Hardness Scale】while moves the
Tolerance Limit cursor.
2. Only D type of impact device has
Hard/б :Hard b the function of бb measure. So the
cursor could not move to 【Hard/бb】
while use other impact device.
6.5.1 Impact Direction Setting
Press key【 】 or 【 】to move
Impact Direction the cursor to the impact direction that
--------------- you will preset.
Press key【ENTER】to confirm it.
Press key【ESC】to cancel it.
14
6.5.3 Material Setting
When 【Hard/бb】is preset to hardness, it will display the following material: Steel
and Cast Steel、Cold Work Tool Steel、Stainless Steel、Gray Cast Iron、Nodular Cast
Iron、Cast Aluminum Alloys、Copper-Zinc Alloys、Copper-Aluminum Alloys、
Wrought Copper and Wrought Steel.
Press key【 】or【 】to move the
Copper-Zinc cursor to the material you want to
preset.
Copper-Alumin Press key【ENTER】to confirm it.
Wrought Copper Press key【ESC】to cancel it.
Wrought Steel Note 1. Presetting hardness scale
defaults to HL automatically after
material presetting is changed.
2. Please select material first, then
select hardness scale.
When 【Hard/ бb】is preset to бb, it will display the following material: Mild Steel、
High-Carbon Steel、Cr Steel、Cr-V Steel、Cr-Ni Steel、Cr-Mo Steel、Cr-Ni-Mo Steel、
Cr-Mn-Si Steel、Super Strength Steel and Stainless Steel.
Press key【 】or 【 】 to move the
Mild Steel cursor to the material to want to preset.
Press key【ENTER】to confirm it.
High-C Steel Press key【ESC】to cancel it.
Cr Steel
Cr-V Steel
15
material is changed.
6.5.5 Tolerance Limit Setting
【 】【 】to move the cursor to print menu and press 【ENTER】to enter the print
menu.
Note:
Printing function is unavailable while charging.
Printing can be stopped by pressing the 【ESC】key.
Do not open the cover of the paper compartment during printing. Otherwise the
instrument may not print normally.
Over high ambient humidity (above 85% of relative humidity) or over low
ambient humidity (below 20% of relative humidity) may reduce the print
quality.
Printing with paper that has been stored for over long period of time or of poor
quality may reduce the print quality or even damage the printer.
16
6.6.2 Print Memory
Print out the selected group of measured value stored inside the instrument.
View From No.1 Press key【 】or key【 】to move the
cursor to the function wanted, then
View From End press key【ENTER】to confirm.
View From No.
Transfer
Delete By No.
Delete All
6.7.1 View from No.1 Group/View from Ending Group
【 View from No.1】Start display values in the memory from the first group.
【View from End】Start display values in the memory from the ending group.
17
6.7.2 View from Selected No. Group
0 01
displaying memory data from the
selected beginning group. Press key
【ESC】to cancel current operation.
18
is.
19
6.9 System Set
At the main display interface, press key【MENU】 enter the main menu.
System Set
20
6.9.1 LCD Brightness Set
Software Info
21
6.11 System Calibration
The tester and impact device must be calibrated using the supplied hardness block
before use as the first time, or having not been used for a long time, or having reset
the system.
Power Unit OFF. Plug in the impact device.
Press key【 】, while pressing【ENTER】to power on the system. Then the software
calibration screen shows as below.
key operation within 5 minutes. Except key【 】, press any key could stop the
blinking of LCD screen and stop the operation of power off at the moment.
While the voltage of the battery is too low, the screen will show < Battery
Empty!>, then power off automatically.
When the instrument is being charged, the Auto Power Off will not function.
22
6.14 Paper Loading
Hold Both ends of the paper
compartment with fingers and open
the compartment cover with
moderate strength.
According to the illustration, put the
paper into the paper compartment
with attention to the paper direction.
If the paper is misplaced, the
instrument will fail to print.
Allow about 1” of paper to stick out
of printer head.
Make sure that the paper is well in
place and close the paper
compartment cover.
23
6.17 Connection of Data Transmission Cable
Insert one connection plug of the transmission cable into the USB socket on the right
side of main body, and insert the other end into the USB port in your computer. Refer
to the manual of the DataPro software for detailed information.
9 Calibration
Calibration should be performed yearly or more often depending upon usage and
quality inspections
10 Storage Conditions
Always avoid vibration, strong magnetic fields, corrosive medium, viscous fluids,
metal chips and dust. Store in room temperature.
Keep 3810A in it’s carry case when not in use.
24
APPENDIX
Table 1
Impact device
Material Scale
D/DC D+15 C G DL
19.3~ 20.0~
HRC 20~68.5 20.6~68.2
67.9 69.5
38.4~
HRB 47.7~99.9 37.0~99.9
99.6
Steel and cast steel 59.1~
HRA
85.8
HB 127~651 80~638 80~683 90~646 81~646
HV 83~976 80~937 80~996 80~950
32.2~ 33.3~ 31.8~
HS 30.6~96.8
99.5 99.3 102.1
20.4~ 19.8~ 20.7~
Cold work HRC DL
67.1 68.2 68.2
tool steel
HV 80~898 80~935 100~941
46.5~
HRB
Stainless steel 101.7
HB 85~655
HV 85~802
HRC
Grey cast iron
HB 93~334 92~326
HV
HRC
Nodular cast iron HB 131~387 127~364
HV
HB 19~164 23~210 32~168
Cast aluminum
23.8~ 22.7~
alloys HRB 23.8~85.5
84.6 85.0
BRASS HB 40~173
(copper-zinc alloys) 13.5~
HRB
95.3
BRONZE(copper-alu
HB 60~290
minum/tin alloys)
Wrought copper
HB 45~315
alloys
25
Table 2
26
Table 3
Type of impact device DC(D)/DL D+15 C G
Impacting energy 11mJ 11mJ 2.7mJ 90mJ
Mass of impact body 5.5g/7.2g 7.8g 3.0g 20.0g
Test tip hardness: 1600HV 1600HV 1600HV 1600HV
Dia. Test tip: 3mm 3mm 3mm 5mm
Material of test tip: Tungsten Tungsten Tungsten Tungsten
carbide carbide carbide carbide
Impact device diameter: 20mm 20mm 20mm 30mm
Impact device length: 86(147)/ 162mm 141mm 254mm
Impact device weight: 75mm 50g 80g 75g 250g
Max. hardness of sample 940HV 940HV 1000HV 650HB
Mean roughness value of 1.6μm 1.6μm 0.4μm 6.3μm
sample surface Ra:
Min. weight of sample:
Measure directly >5kg >5kg >1.5kg >15kg
Need support firmly 2~5kg 2~5kg 0.5~1.5kg 5~15kg
Need coupling tightly 0.05~2kg 0.05~2kg 0.02~0.5kg 0.5~5kg
Min. thickness of sample
Coupling tightly 5mm 5mm 1mm 10mm
Min. layer thickness for
surface hardening ≥0.8mm ≥0.8mm ≥0.2mm ≥1.2mm
Hardness Indentation 0.54mm 0.54mm 0.38mm
300HV diameter
Depth of 24μm 24μm 12μm
indentation
Hardness Indentation 0.54mm 0.54mm 0.32mm
600HV diameter
Depth of 17μm 17μm 8μm
indentation
27
Optional 12pc Support Ring Set
Part No. S38R
28
3810A Data Output Software Installation:
29