Technical Manual-Appendix
2017.9.13 Document Ver.1.5
Table of Contents
1. Introduction ..................................................................................................................................................... 3
1.1 General term ....................................................................................................................................................................................................................3
1.2 Preparation material.......................................................................................................................................................................................................3
1.2.1 Utility tool .............................................................................................................................................................................................................3
1.2.2 JIGs and Phantoms ..............................................................................................................................................................................................4
1. Introduction
1.1 General term
This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.
[Driver]
[Wrench Set]
2. Device connection
2.1 PAPAYA_REC program and how to connect with machine
1) “C:/GENORAY/PAPAYA”. Inside of the folder, find PAPAYA_REC.exe and execute the file.
2) Select About->Model Option
3) Select machine
'
Please make sure to check Panoramaic detector option DRS for uPano detector.
Please make sure to check Equipment option PAPAYA 3Ds ENT for PAPAYA 3D ENT
5) Click About-> Version Info to , verify the version of the Detector and each Board.
3. Preparation process
3.1 How to run the program
3.2 Preparing
1) After select Step1, you should align the side of gantry moving block and driving block.
2) Input kVp offer value in No.3(exposure condition offset) with refer to No.2 shown in the below picture.
3) Click “Set current position” for setting up Y-axis starting point.
4) After complete “5” setting, its value will appear on the bottom of the screen.
5) The value will be Y-axis starting point.
4 5
These tools use for machine’s alignment of rotating degree to starting point.
Gantry 90°
1 2
6) According to above picture, “SD-Line Jig” and “Ball Phantom” are installed on machine. And loose bolts for
Chinrest in which are fixing chinrest.
7) Set up the height of Ball Phantom to be aligned by moving SD-Line Jig ruler.
8) By pushing “0°” button, making perpendicular between rotating part and driving part.
9) SD-Line Jig line and Ball Phantom’s reference line need to be matched by turning ‘Ball phantom’ stage’s knob
and move ‘Ball phantom’s location. After then, tight the bolts of Chinrest part for fixing.
10) Click “180°” button. Then the rotating part move to 180°.
11) If, at 0° location and 180° location, “SD-Line Jig” is not on same location, on adjusted “Ball phantom’s
reference line and “SD-line jig’s line that needs to be located by turning each knob and locating them to the
middle of “SD-line jig’s locations of 0° and 180°.
12) By repeating the above steps to reduce the errors.
Gantry 90°
13) Tight the bolts in the red circle below for fixing “Chinrest part”.
14) Make “Gantry” position at 90° and then “SD-Line Jig” line meet the center of “Ball Phantom” by using
Left/Right chinrest button.
15) Click “Set X-Axis Center” button for setting the center of X-axis.
16) After completing data, the value will appear on the bottom of the screen.
17) This value is X-axis reference line.
18) After checking “Ball Phantom” stage ruler’s value, put this value into the blank marked 1 in red circle.
19) Click “Set Chinrest Y-Axis offset”
When you click “Help (Guide Image)”, the instruction will appear to teachhow to check the
ruler value from Ball Phantom stage.
7) Put “20” in the blank marked 1 and click “Move Detector V”.
8) After clicking, thedetector move to 20mm upper side
9) Select marked 3, Pin offset 20mm, in the following picture, and fix “Pin jig” to align Detector Base’s hole and
Moving block’s hole.
10) If the holes are not to be aligned, adjust the detector couple ring for matching them.
11) After removing “Pin jig”, click “Set current position” according to No.4. The value will be saved as you can see
in the blanks marked 5.
Laser Jig
1) Select “Step4”
2) Click “Move init position” button to set up the machine for laser alignment.
3) Install “Ball Phantom” on the Chinrest part. After then, adjust Midsagittal and Canine’s laser positions to meet
the guide line in the Laser jig according to marked “A” below.
4) Set F/R laser behind the top of detector and click “Set current position”.
5) After completing, the value will be set in the blank No. 4.
3 4
A 2
Move Cu Filter to right side → If the value is higher than initial value.
Move Cu Filter to left side → If the value is lower than initial value.
5
1
3
5) As like the Packing Position Set in the picture, adjust chinrest U/D Block and insert Pin Jig and mate with the
hole off Block, after that remove Pin Jig.
6) Input the value of chinrest packing position after removing and clicking No.10 in the picture.
7 The inputted value is applied to No.11.whether it is correct for Cu Filter(0°) like the below picture Figure2.
3) If it is not correct, change the value of No.1 and adapt as like aCu Filter(0°) form. With that value, please input
in No.30°pulse.
2
1
4) ‘1400’ in No.1 of the picture, move Cu Filter by clicking No.2 (move filter position)
5) Check whether it is correct for Cu Filter(90°) like the below picture Figure2.
6) If it is not correct, change the value of No.1 and adapt as like aCu Filter(90°) form.
7) Click No.3(move filter position & get image) in the picture, with dragging the right mouse button check if the
panorama Cu Filter is in straight line. If it is not straight line, please adjust it.
8) With that value, input in No.4(90°pulse) of picture, click No.5 (Calc. Coef. & Set Filter), input the value of Cu
Filter.
9) The Cu Filter value of No.6 in the picture is inputted.
Be careful that the x-ray is released during the Cu Filter Alignment process.
2 3
1
A
4 5
These tools use for setting the height and width of Collimator.
(-) Driver ‘
Collimator Jig
4
1
3
`
For Panoramic Collimator process, and if the Equipment Option is for ‘PAPAYA 3Ds ENT’, it is not necessary
process.
1) Select“Step7” tab.
2) Click“Find Collimator position” to set up the location of the collimator.
3) After complete the process, the Panorama collimator value appears as “A”in the picture.
4) Click “OK” button and then click “Set Position & Get Image”.
1 2
LR Pos → Adjust X-ray’s right / left exposure direction. ((+) is for moving right side and (-) is for moving left
side.)
UD Pos → Adjust X-ray’s up / down exposure direction. ((+) is for moving upper side, (-) is for moving down
side)
3 4
1) Select “Step8”
2) Click “Auto” for processing “Pano Calibration”
3) After complete “Pano Calibration” step, the message will appear.
4) Click “OK”
1) Select Step8
2) Select the article related to Preset Option like the No.1 in the picture.
2-1) For UPano, select [PANO_CT_ROI_BANK_0].
2-2) For Extor Pano, select [PANO_PANORAMA].
3) For exposure, please click No.2 (Check All) and No.3 (Exposure).
4) When finishing the exposure, Avg.Level for whole image and Avg for piece of images are shown. In addition,
when appearing Saturation pop-up like No.7 after appearing the Avg value, please retake an exposure by
adjusting condition.
5) After finish of all exposure and Avg. value marking, compare it with the range of Ref. Level(No.4)
2 6
6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.8(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’, adjust for
correct range.
8) After adapt to Ref.Level standard range, click No.8(create) and complete it.
9) When finish the process, the content of ‘Parameters are generated’ is shown.
10) For UPano, after selecting “PANO_CT_ROI_BANK_1”, and “PANO_CT_ROI_BANK_2” in No.1 , go on to the next
step like above method.
10
Do not place objects on the Chinrest Block when process for Panoramic Detector Calibration.
If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value
Auto shading Calibration is not required if the equipment option is PAPAYA 3Ds ENT
1) Select “Step9”
2) Click “Auto shading calibration”
3) After complete “Shading calibration” step, the message will appear. Then click “OK”
4) Totally three message appear after this step. Click “OK” for next step.
‘
Ball Phantom Collimator Jig
1) Select “Step10”
2) Fix “Ball Phantom” on the chinrest and click “Confirm”.
3) After Confirming, align canine laser to the frond reference side of ball Phantom by moving Y-axis.
4) Click “Ready & Shot” to take an image of Ball Phantom.
5) After getting an image of ball Phantom, the screen shows the image as below.
6) Select the third line and draw four lines on the image.
7) After drawing four lines, If the message, marked 4, shows “PASS” three time, the panoramic alignment and
calibration will be finished.
8) If any “FAIL” message occurs, back to STEP2, STEP3, and STEP7 with instruction given.
2
1
D35 X H40
1) Select Step12
2) Select Full Scan CT in No.1 of the picture.
3) Click ‘Check All’ in No.2 of picture, retake an exposure by clicking No.3(exposure).
4) When finishing the exposure, Avg.Level for whole image and Avg for piece of images are shown No.5 & No.6
in the graph. In addition, when appearing Saturation pop-up after appearing the Avg value, please retake an
exposure by adjusting condition.
5) After finish of all exposure and Avg. value marking, compare it with the range of Ref. Level(No.4)
6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.7(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’ in No.7,
adjust for correct range, and complete FullScan CT Calibration by clicking (create).
Both Full Scan_CT and Full Scan Stitch_CT in CT Detector Calibration are not required if the
equipment option is papaya 3Ds ENT
During CT detector calibration, any object should not be on the chinrest Block.
If the gap between output value and initial value is lower than 50→ change mA value
If the gap between output value and initial valueis higher than 50→ change kVpvalue
7 5
4 3
2
6
8) After complete the process, the message “Parameters are generated” appears.
9) Select “HalfScan_CT” and process “HalfScan_CT Calibration” with same method above mentioned.
10) Select “FullScanStitch_CT” and process “FullScanStitch_CT Calibration” with same method above mentioned.
11) Select no.1 [FullScanEX_CT] as shown below. Then process FullScanEX_CT Calibration with described above.
12) Select no.1 [FullScanEXStitch_CT] as shown below. Then process FullScanEXStitch_CT with described above.
1) Select “Step13”tab
2) Attach the Base Plate in the Chinrest like ‘A’ of picture, Use the bolts located under the base plate after
adjusting horizontally, and align the line for Base Plate to the standard of Half GCP Phantom line.
Gantry Y Axis Start Offset → Put Plus(+) value to move left side.
Gantry Y Axis Start Offset → Put Minus(-) value to move right side.
1
D E
2
6) After the adjustment, check the box of “Half Scan Stitch” and click “Confirm”.
7) After confirming it, click “Ready & Shot” to take an image of “Half Scan_GCP Phantom”
8) When the shooting is complete, click “Calc, GCP-EV” for running“evGCP” program.
D E
3 4
5
6) After finishing GCP process, the screen will appear as below. Then click “Save” and “OK” to complete “Half_GCP”
process.
1) Attach the Base Plate in the Chinrest like ‘A’ of picture. After adjusting horizontally, attach the full GCP
Phantom in Base Plate correspondently for the line.
2) Click on No.1 [0 ° Move and Get Image] shown in the below picture and take an exposure.
Gantry Y Axis Start Offset → Put Plus(+) value to move left side.
Gantry Y Axis Start Offset → Put Minus(-) value to move right side.
5) After the adjustment, check the box of “Full Scan Stitch” and click “Confirm”.
Process Full Scan Stitch by Selecting FullScanEX Stitch if the equipment option is papaya 3Ds
ENT
6) After confirming it, click “Ready & Shot” to take an image of “Full Scan_GCP Phantom”
7) When the shooting is completed, click “Calc, GCP-EV” for running“evGCP” program.
4
3 5
4) Select “2st for seperating layer “ fromthe work select bar in below picture.
5) Move the mouse curser between two layers as the marked 4 on the picture by left click.
6) After then the green line will appear.
10) After complete GCP, click no.1 [save] as shown below and then click no2. [ok] to exit Full GCP.
11) As above, click[Second projection dataset] as shown as below and then repeatedly execute.
1) After selecting Step 14, attach the base plate to the Chinrest and adjust it horizontally.
2) After selecting No.1(Half Scan) /Select No.1 (Half Scan), click No.2 (Confirm), the path is set automatically.
3) Process the exposure by clicking No.3(Ready & Shot) after confirming.
When finishing the exposure, reconstruction will proceed automatically, and the reconstructed slice will be
displayed on the screen.
4) Input ‘40’ in Edit text of No.4 (Slice index), click No.5 (Rec).
5) Click No.6 (Insert HU), drag the mouse like No.7, make the range of HU.
6) After making the range of HU from A to D like above way, click No.8 (Calibrate HU), and the result values of
HU in the left under part are shown.
7) After selecting Full Scan mode, process for same way from 2) to 6).
1. Introduction
1.1 Regular condition
This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.
Before reviewing this Appendix 2, read “Appendix 1 Align & Calibration –Pano/CT” and familiar with all calibration
steps first.
1.2 Preparation
1.2.1 Required Tools
[Wrench Set]
1.2.2 JIG
[Pin Jig]
2. Device connection
2.1 PAPAYA_REC program and how to connect with the machine.
3) When the machine is connected with the software, “Connect” button changes for “Disconnect”.
3. Preparation process
3.1 How to run the program
3.2 Preparing
2, Pin jig
Pin Jig
1) Select “Step1” and match Detector Moving Block’s hole with the reference point of Cephalo Base.
2) After then, make them to be aligned by the Pin Jig and then remove the Pin Jig. Select “Set current position”.
3) When you click “Set current position”, Result detector offset value will be appeared.
4) This value is a reference position of the detector.
2 3
During removing “Pin Jig” from the part, make sure the detector not to move.
LR Pos → Add up value from the current value to move right side, take away value from the
current value to move to left side.
UD Pos → Add up value from the current value to move upside, take away value from the
current value to move to down side.
The Pano collimator setting is perfectly complete before starting 1st Collimator’s reference
positioning process.
1 4
3 5
B C
1) Select “Step3”
2) Click “Find 2nd Collimator position” to find “2nd Collimator” value.
3) After searching, the result offset value will appear. Click “OK” and save the value of 2nd Collimator.
4) When you complete to save data, each value will be updated in 2nd collimator position blank.
5) Click “Set position & Get Image” and check the 2nd Collimator is in a good position as the sample picture
below;
1
3 4
6) If the image is positioned lower than correct position, loosen four bolts, marked 5 on the following picture,
and adjust the position of Lead.
7) If the image position is in the straight position, loosen three bolts, marked 6 on the following picture, and
adjust the position of Lead.
1) Click Step4
2) Select [CEPH_CEPHALO] in No.1 of the below picture.
3) Click [Check All] in No.2, take an exposure by clicking No.3 [Exposure].
4) When finishing the exposure, Avg.Level for whole image in No.4 and Avg for piece of images in ‘status’ are
shown. In addition, when appearing Saturation pop-up like No.5 after appearing the Avg value, please retake
an exposure by adjusting condition.
5) After all exposure and marking of Avg.value are finished, compare them to the value of Ref.Level.
6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.7(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’ in No.8,
adjust for correct range.
8) After adapt to Ref.Level standard range, click No.7(create) and complete it.
9) When finish the process, the content of ‘Parameters are generated’ is shown.
If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value
2
4
8 6
3
1) Select “Step5”
2) Fold Nasal post parts.
3) Make the rotating part to AP position and remove Ear posts.
4) Click “Auto scan calibration” to start the detector calibration process.
5) After getting an image, click “OK” on the message windows.
6) For taking the Lateral image, click “Auto scan FFC calibration” to get another image.
7) After complete the shooting, click “OK” to complete the calibration process.
When selecting the center of circle, place the cross form within the range of the center for circle.
3
4
6 1
2
6) If the ‘PASS’ message is not shown, cancel the articles for ‘A’(Ear rod), Nasal positioner, rotation part cover with
referring to Help button.
7) After removing, loosen a screw ‘B’ a bit, move rotation part side to side, attach Ear rod and retake an exposure.
8) Select Cross (1, 2) to select the result value for PASS, Secure the screws of the B shown in the figure below.
9) After pressing the laser light button, loosen a screw ‘C’ a bit, adjust the laser level horizontally, place the height
in the middle part of the Ear rod hall for the laser, and fix the screw.
Take care not to bend the photo sensor and bracket when moving the rotating rotation.
6) If the position of the Ear rod is indicated by PASS, check if the quality of the printed image is shown as the
image shown ‘good’ below. If similar, the Cephalo calibration is completed without defect.
7) If the image is abnormal, proceed to step 5 again.
When checking the position of Ear rod and Cephalo Calibration, the x-ray could be exposed, please be
careful.
This manual contains general usage descriptions and other information about Seasoning.
You have to finish the process (Align & Calibration) first and keep power off for a long time to install.
1) After turning on the appliance, run the PAPAYA_REC.exe file in C:\GENORAY\PAPAYA folder.
2) When the program runs as shown below, click the [Connect] button to connect to the equipment.
2. Running Seasoning
3. Completion of Seasoning
4. Notes
This manual contains information about the confirming and management of RAW files after exposure.
Ensure that this manual is interlocked to the database, please keep fully in mind with the contents of this manual.
The Auto Save Raw Manage tab is divided by Panorama / Cephalo / CT like below.
Set the number of exposure list for the management in [Auto Save management] of PAPAYA OP Setting.
The button instruction is like a below.
1) After connecting DB, renew the list.
2) Open the folder saved for Raw file.
3) After connecting DB, remove the item.
1 2 3
This manual includes information and method to use remote update of PAPAYA Series(unify word "PAPAYA").
Keep informed this manual to use equipment effectively.
1. Remote update
2. Run LisaTools
2) As shown in the below, the equipment picture is shown in the center, and the equipment name is displayed in
the upper right corner.
- Check picture and name whether it is same with the equipment you want to update.
In the case of PAPAYA PLUS or SLIM, Check equipment name surely because it is equal to equipment
picture.
[Checking possible B/D for remote update – Impossible for main FPGA update]
- Once you complete the process, you can run indivisual/selective update.
[Select binary]
3) Once you select the file, the current version and the version for update are shown.
4) Once you press Step3 [Upload files on boards], the update is run.
5) The checking marks beside each B/D, like below, is shown when finishing update.
6) If you want to do another update, please do step2 [Select Files to uploard] (P.69) first.
- The checking marks, like below, is shown when finishing main B/D update after chinrest B/D update.
2) Select binary files of B/D for update( ex. Chinrest & LAN Switch).
[Select twice]
3) Once you finish selecting file, current version and the version for update are shown.
- Like below, all versions for B/D and update are shown if you select all binary of B/D.
4) Once you press step3 [Upload files on boards], the update is started.
5) After finishing update, checking marks of all status lights are shown.
Below is the chart for management system. Please refer to this chart.
PPY : PAPAYA
Product name
P3D : PAPAYA 3D
[3 letters with english]
SLM : PAPAYA SLIM
MAIN : MAIN B/D
COLL : COLLIMATOR B/D
B/D name CHIN : CHINEREST B/D
[3 letters with english] LIFT : LIFTER B/D
LANS : LAN SWITCH B/D
FPGA : MAIN B/D FPGA
MAIN : 01
CHIN : 02
COLL : 03
CAN_ID
LIFT : 04
LANS : 05
FPGA :06
BOOT : Bootloader
function APPL : Application Firmware
[4letters with english] FACT : FPGA Factory Image
UFDA : FPGA Application
1 : PAPAYA
Number for A-1
2: PAPAYA 3D
[Product classification]
3 : PAPAYA SLIM
Number for C -2
0~99
[Function modification]
The below chart shows the firmware versioncompatibilityof equipment for remote update.
LAN SW B/D
- - 2_2_5_001 or higher
(LING LED MAIN)
LAN SW B/D
- - 2_2_5_001 or higher