Appendix 1 Alignment&Calibration-Pano/CT

Table of Contents
1. Introduction ..................................................................................................................................................... 3
1.1 General term ....................................................................................................................................................................................................................3
1.2 Preparation material.......................................................................................................................................................................................................3
1.2.1 Utility tool .............................................................................................................................................................................................................3
1.2.2 JIGs and Phantoms ..............................................................................................................................................................................................4

2. Device connection ......................................................................................................................................... 6


2.1 PAPAYA_REC program and how to connect with machine ................................................................................................................................6

3. Preparation process ...................................................................................................................................... 9


3.1 How to run the program .............................................................................................................................................................................................9
3.2 Preparing ...........................................................................................................................................................................................................................9

4. Alignment & Calibration ........................................................................................................................... 11


4.1 Step1 : Set up the gantry reference positioning................................................................................................................................................ 11
4.2 Step2 : Set up the angle of gantry rotation reference positioning .............................................................................................................. 12
4.3 Step3 : Detector Alignment ...................................................................................................................................................................................... 16
4.4 Step4 : Laser Alignment............................................................................................................................................................................................. 19
4.5 Designation of detector rotation and chinrest height...................................................................................................................................... 20
4.7 Step6 : Set the height and width of Collimator ................................................................................................................................................. 24
4.8 Collimator setting ........................................................................................................................................................................................................ 25
4.8.1 Step7 : Pano Collimator position setting.................................................................................................................................................... 25
4.9 Step8 :Pano Detector Calibration (Detector model : XID-C15DP)................................................................................................................. 28
4.9.1 Detector Model : I3 .......................................................................................................................................................................................... 28
4.9.2 Detector Model :Extor or Dualray s (uPano) ............................................................................................................................................. 29
4.10 Step9 : Auto Shading Calibration ......................................................................................................................................................................... 31
4.11 Step10 : How to check Pano Alignment ............................................................................................................................................................ 32
4.11.1 Step11 : CT Collimator position setting ................................................................................................................................................... 33
4.12 Step12 : CT Detector Calibration .......................................................................................................................................................................... 35
4.13 Step13 :How to take an image of Half GCP Phantom................................................................................................................................... 37
4.14 Half GCP steps ........................................................................................................................................................................................................... 40
4.15 Step13 :How to take an image of Full GCP Phantom.................................................................................................................................... 42
4.16 Full GCP Steps ............................................................................................................................................................................................................ 44
4.17 Step14 : HU Calibration ........................................................................................................................................................................................... 49

PAPAYA 3D Technical Manual (ver 1.5) 2 / 86


Appendix 1 Alignment&Calibration-Pano/CT

1. Introduction
1.1 General term

This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.

1.2 Preparation material


1.2.1 Utility tool

[Driver]

[Wrench Set]

3 / 86 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

1.2.2 JIGs and Phantoms

[Ball Phantom] [HU Phantom]

[Full_GCP Phantom] [Half_GCPPhantom]

[Pin Jig] [Collimator Jig]

PAPAYA 3D Technical Manual (ver 1.5) 4 / 86


Appendix 1 Alignment&Calibration-Pano/CT

[SD-Line Jig] [Theta Jig]

[Detector Base] [Detector Align Jig]

[Laser Jig] [Base Plate]

5 / 86 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

2. Device connection
2.1 PAPAYA_REC program and how to connect with machine

1) “C:/GENORAY/PAPAYA”. Inside of the folder, find PAPAYA_REC.exe and execute the file.
2) Select About->Model Option

PAPAYA 3D Technical Manual (ver 1.5) 6 / 86


Appendix 1 Alignment&Calibration-Pano/CT

3) Select machine

'

Please make sure to check Panoramaic detector option DRS for uPano detector.
Please make sure to check Equipment option PAPAYA 3Ds ENT for PAPAYA 3D ENT

4) After execution of the machine alignment program, push [Connect]


button where it is located at middle upper side and connecting program with machine.

7 / 86 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

5) Click About-> Version Info to , verify the version of the Detector and each Board.

PAPAYA 3D Technical Manual (ver 1.5) 8 / 87


Appendix 1 Alignment&Calibration-Pano/CT

3. Preparation process
3.1 How to run the program

1) Click “Calibration Wizard” to run it.

3.2 Preparing

1) When Calibration Wizard is activated the following screen will appear.


2) Remove Driving, Tank, Gantry, and Detector Cover before machine alignment and modification.

9 / 86 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

3) Select “Preparing2”tab and it shows how to remove “Chinrest cover”

4) As shown below, adjust the equipment level by selecting Preparing3

PAPAYA 3D Technical Manual (ver 1.5) 10 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4. Alignment & Calibration


4.1 Step1 : Set up the gantry reference positioning

1) After select Step1, you should align the side of gantry moving block and driving block.
2) Input kVp offer value in No.3(exposure condition offset) with refer to No.2 shown in the below picture.
3) Click “Set current position” for setting up Y-axis starting point.
4) After complete “5” setting, its value will appear on the bottom of the screen.
5) The value will be Y-axis starting point.

4 5

11 / 86 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.2 Step2 : Set up the angle of gantry rotation reference positioning

These tools use for machine’s alignment of rotating degree to starting point.

Theta Jig SD-Line Jig Ball Phantom

1) Select “Step2” tab.


2) Fix “Theta Jig” to driving side.
3) Adjust the gantry part meets “Theta Jig” and parallel each other.
4) Click “ Set current position” to save the Gantry reference position to Main B/D.
5) After the data saves, the value will be written on the red circle marked 2.

Gantry 90°

1 2

PAPAYA 3D Technical Manual (ver 1.5) 12 / 87


Appendix 1 Alignment&Calibration-Pano/CT

6) According to above picture, “SD-Line Jig” and “Ball Phantom” are installed on machine. And loose bolts for
Chinrest in which are fixing chinrest.
7) Set up the height of Ball Phantom to be aligned by moving SD-Line Jig ruler.
8) By pushing “0°” button, making perpendicular between rotating part and driving part.
9) SD-Line Jig line and Ball Phantom’s reference line need to be matched by turning ‘Ball phantom’ stage’s knob
and move ‘Ball phantom’s location. After then, tight the bolts of Chinrest part for fixing.
10) Click “180°” button. Then the rotating part move to 180°.
11) If, at 0° location and 180° location, “SD-Line Jig” is not on same location, on adjusted “Ball phantom’s
reference line and “SD-line jig’s line that needs to be located by turning each knob and locating them to the
middle of “SD-line jig’s locations of 0° and 180°.
12) By repeating the above steps to reduce the errors.

Gantry 90°

13 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

13) Tight the bolts in the red circle below for fixing “Chinrest part”.

Tighten bolts in this part

14) Make “Gantry” position at 90° and then “SD-Line Jig” line meet the center of “Ball Phantom” by using
Left/Right chinrest button.
15) Click “Set X-Axis Center” button for setting the center of X-axis.
16) After completing data, the value will appear on the bottom of the screen.
17) This value is X-axis reference line.

PAPAYA 3D Technical Manual (ver 1.5) 14 / 87


Appendix 1 Alignment&Calibration-Pano/CT

18) After checking “Ball Phantom” stage ruler’s value, put this value into the blank marked 1 in red circle.
19) Click “Set Chinrest Y-Axis offset”

When you click “Help (Guide Image)”, the instruction will appear to teachhow to check the
ruler value from Ball Phantom stage.

15 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.3 Step3 : Detector Alignment

These tools used for the detector Alignment.

Detector Plate Detector Align Jig Pin Jig

1) Select “Step3” tab


2) Click “Move Gantry Link 0° position”

PAPAYA 3D Technical Manual (ver 1.5) 16 / 87


Appendix 1 Alignment&Calibration-Pano/CT

3) Fix “Detector Align Jig” according to the picture .


4) Adjust the stage of Detector Align Jig as  to meet the detector Base part.
5) If the detector Align Jig and Detector Base are not parallel, loose some bolts and adjust the detector base part
aspicture.
6) After fixing the part, remove the detector plate in the Yellow line. Please refer to the following picture.

17 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

7) Put “20” in the blank marked 1 and click “Move Detector V”.
8) After clicking, thedetector move to 20mm upper side
9) Select marked 3, Pin offset 20mm, in the following picture, and fix “Pin jig” to align Detector Base’s hole and
Moving block’s hole.
10) If the holes are not to be aligned, adjust the detector couple ring for matching them.
11) After removing “Pin jig”, click “Set current position” according to No.4. The value will be saved as you can see
in the blanks marked 5.

PAPAYA 3D Technical Manual (ver 1.5) 18 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.4 Step4 : Laser Alignment

This tool uses for “Laser Alignment”

Laser Jig

1) Select “Step4”
2) Click “Move init position” button to set up the machine for laser alignment.
3) Install “Ball Phantom” on the Chinrest part. After then, adjust Midsagittal and Canine’s laser positions to meet
the guide line in the Laser jig according to marked “A” below.
4) Set F/R laser behind the top of detector and click “Set current position”.
5) After completing, the value will be set in the blank No. 4.

3 4
A 2

19 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.5 Designation of detector rotation and chinrest height.

These tools use for setting the chinrest of height

Ball Phantom Pin Jig

1) Select “Step5” tab.


2) Like the below picture’s Figure1, attach Ball Phantom in Chinrest, place Canine laser in the marked part, click
No.2 shown in the picture.
3) Click No.3 shown in the picture and take an exposure, Check if the center ball in Ball Phantom is correspond
with red circle.
4) When clicking No.4 shown in the picture, the values of detector rotation and chinrest are inputted.

Move Cu Filter to right side → If the value is higher than initial value.
Move Cu Filter to left side → If the value is lower than initial value.

The up position of Ball Phantom ->It is higher than standard value.


The down position of Ball Phantom ->It is lower than standard value.

Please be caution when adjusting the height of chinrest.

PAPAYA 3D Technical Manual (ver 1.5) 20 / 87


Appendix 1 Alignment&Calibration-Pano/CT

5
1
3

21 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

5) As like the Packing Position Set in the picture, adjust chinrest U/D Block and insert Pin Jig and mate with the
hole off Block, after that remove Pin Jig.
6) Input the value of chinrest packing position after removing and clicking No.10 in the picture.
7 The inputted value is applied to No.11.whether it is correct for Cu Filter(0°) like the below picture Figure2.
3) If it is not correct, change the value of No.1 and adapt as like aCu Filter(0°) form. With that value, please input
in No.30°pulse.

Right side movement of Cu Filter -> It is higher than standard value.


Left side movement of Cu Filter ->It is lower than standard value.

2
1

PAPAYA 3D Technical Manual (ver 1.5) 22 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4) ‘1400’ in No.1 of the picture, move Cu Filter by clicking No.2 (move filter position)
5) Check whether it is correct for Cu Filter(90°) like the below picture Figure2.
6) If it is not correct, change the value of No.1 and adapt as like aCu Filter(90°) form.
7) Click No.3(move filter position & get image) in the picture, with dragging the right mouse button check if the
panorama Cu Filter is in straight line. If it is not straight line, please adjust it.
8) With that value, input in No.4(90°pulse) of picture, click No.5 (Calc. Coef. & Set Filter), input the value of Cu
Filter.
9) The Cu Filter value of No.6 in the picture is inputted.

Be careful that the x-ray is released during the Cu Filter Alignment process.

2 3
1
A
4 5

23 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.7 Step6 : Set the height and width of Collimator

These tools use for setting the height and width of Collimator.

(-) Driver ‘
Collimator Jig

1) Select “Step6” tab.


2) Click “Move init position” for adjusting the position of blade of collimator.
3) Put “Collimator jig” into the collimator.
4) The top of blade motor to meet the top of collimator jig by using (-) Driver. Make sure to have a little gap
between jig and blade for removing the jig later.
5) The left side of blade motor meets the left side of collimator jig by using (-) Driver. Make sure to have a little
gap between jig and blade for removing the jig later.
6) After fixing the blade motors, remove the collimator Jig.
7) Click “Calc. Panoramic Collimator Position” and put “Pano Collimator” value.
8) After putting the value, it will appear on the blanks marked 4.

4
1

3
`

Be careful! When you adjust blade motor.

PAPAYA 3D Technical Manual (ver 1.5) 24 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.8 Collimator setting


4.8.1 Step7 : Pano Collimator position setting

For Panoramic Collimator process, and if the Equipment Option is for ‘PAPAYA 3Ds ENT’, it is not necessary
process.

1) Select“Step7” tab.
2) Click“Find Collimator position” to set up the location of the collimator.
3) After complete the process, the Panorama collimator value appears as “A”in the picture.
4) Click “OK” button and then click “Set Position & Get Image”.

X-ray exposure should be necessary for this step.

1 2

25 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

5) An image will be displayed as same as“P” below.


6) In this condition, if the collimator’s left or right position is not well found, adjust LR Posvalue to correct it.
7) If the collimator’s up side or down side position is not well found, adjust UD Posto correct it.
8) After finding appropriate value, click “Set Position & Get Image” to save “Pano Collimator” value.

LR Pos → Adjust X-ray’s right / left exposure direction. ((+) is for moving right side and (-) is for moving left
side.)
UD Pos → Adjust X-ray’s up / down exposure direction. ((+) is for moving upper side, (-) is for moving down
side)

3 4

PAPAYA 3D Technical Manual (ver 1.5) 26 / 87


Appendix 1 Alignment&Calibration-Pano/CT

In case the location of


collimator is incorrect.

Well found the position of In case the location of


collimator collimator is incorrect.

27 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.9 Step8 :Pano Detector Calibration (Detector model : XID-C15DP)


4.9.1 Detector Model : I3

1) Select “Step8”
2) Click “Auto” for processing “Pano Calibration”
3) After complete “Pano Calibration” step, the message will appear.
4) Click “OK”

PAPAYA 3D Technical Manual (ver 1.5) 28 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.9.2 Detector Model :Extor or Dualray s (uPano)

1) Select Step8
2) Select the article related to Preset Option like the No.1 in the picture.
2-1) For UPano, select [PANO_CT_ROI_BANK_0].
2-2) For Extor Pano, select [PANO_PANORAMA].
3) For exposure, please click No.2 (Check All) and No.3 (Exposure).
4) When finishing the exposure, Avg.Level for whole image and Avg for piece of images are shown. In addition,
when appearing Saturation pop-up like No.7 after appearing the Avg value, please retake an exposure by
adjusting condition.
5) After finish of all exposure and Avg. value marking, compare it with the range of Ref. Level(No.4)

2 6

29 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.8(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’, adjust for
correct range.
8) After adapt to Ref.Level standard range, click No.8(create) and complete it.
9) When finish the process, the content of ‘Parameters are generated’ is shown.
10) For UPano, after selecting “PANO_CT_ROI_BANK_1”, and “PANO_CT_ROI_BANK_2” in No.1 , go on to the next
step like above method.

10

Do not place objects on the Chinrest Block when process for Panoramic Detector Calibration.

If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value

PAPAYA 3D Technical Manual (ver 1.5) 30 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.10 Step9 : Auto Shading Calibration

Auto shading Calibration is not required if the equipment option is PAPAYA 3Ds ENT

1) Select “Step9”
2) Click “Auto shading calibration”
3) After complete “Shading calibration” step, the message will appear. Then click “OK”
4) Totally three message appear after this step. Click “OK” for next step.

31 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.11 Step10 : How to check Pano Alignment

These tools use for checking panoramic alignment.


Ball Phantom Collimator Jig

1) Select “Step10”
2) Fix “Ball Phantom” on the chinrest and click “Confirm”.
3) After Confirming, align canine laser to the frond reference side of ball Phantom by moving Y-axis.
4) Click “Ready & Shot” to take an image of Ball Phantom.
5) After getting an image of ball Phantom, the screen shows the image as below.
6) Select the third line and draw four lines on the image.
7) After drawing four lines, If the message, marked 4, shows “PASS” three time, the panoramic alignment and
calibration will be finished.
8) If any “FAIL” message occurs, back to STEP2, STEP3, and STEP7 with instruction given.

PAPAYA 3D Technical Manual (ver 1.5) 32 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.11.1 Step11 : CT Collimator position setting

1) Select "Step11” tab


2) Select “D35 and H40” and click “Set Position & Get Image on CT”. The image should appear over the red line
as same as below picture.
3) If the image is smaller than the red line, adjust value from the tabs marked 8.
4) All the F.O.V sizes, D40XH50, D60XH60, D70XH70, D80XH80 D140XH80, and D140XH140, D160XH80,
D160XH140 should be adjusted as same as above step.

LR Pos→ Adjust Right / Left ((+) to right, (-) to left)


UD Pos→ Adjust Up / Down ((+) to up, (-) to down)
W → Adjust the width of Right / Left ((+) to be wider, (-) to be smaller)
H → Adjust the height of X-ray exposure ((+) to be taller,(-) to be smaller)

X-ray exposure should be necessary for this step.

2
1

D35 X H40

33 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

Reference images to each F.O.V size

PAPAYA 3D Technical Manual (ver 1.5) 34 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.12 Step12 : CT Detector Calibration

1) Select Step12
2) Select Full Scan CT in No.1 of the picture.
3) Click ‘Check All’ in No.2 of picture, retake an exposure by clicking No.3(exposure).
4) When finishing the exposure, Avg.Level for whole image and Avg for piece of images are shown No.5 & No.6
in the graph. In addition, when appearing Saturation pop-up after appearing the Avg value, please retake an
exposure by adjusting condition.
5) After finish of all exposure and Avg. value marking, compare it with the range of Ref. Level(No.4)
6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.7(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’ in No.7,
adjust for correct range, and complete FullScan CT Calibration by clicking (create).

Both Full Scan_CT and Full Scan Stitch_CT in CT Detector Calibration are not required if the
equipment option is papaya 3Ds ENT

During CT detector calibration, any object should not be on the chinrest Block.

If the gap between output value and initial value is lower than 50→ change mA value
If the gap between output value and initial valueis higher than 50→ change kVpvalue

7 5

4 3

2
6

35 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

8) After complete the process, the message “Parameters are generated” appears.
9) Select “HalfScan_CT” and process “HalfScan_CT Calibration” with same method above mentioned.
10) Select “FullScanStitch_CT” and process “FullScanStitch_CT Calibration” with same method above mentioned.
11) Select no.1 [FullScanEX_CT] as shown below. Then process FullScanEX_CT Calibration with described above.
12) Select no.1 [FullScanEXStitch_CT] as shown below. Then process FullScanEXStitch_CT with described above.

PAPAYA 3D Technical Manual (ver 1.5) 36 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.13 Step13 :How to take an image of Half GCP Phantom

These tools use for Half GCP process.

Base Plate Half_GCP Phantom

1) Select “Step13”tab
2) Attach the Base Plate in the Chinrest like ‘A’ of picture, Use the bolts located under the base plate after
adjusting horizontally, and align the line for Base Plate to the standard of Half GCP Phantom line.

37 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

3) Click “0° Move & Get Image” to start to get an image.


4) After getting an image, theimage of Half_GCP Phantom appears.
5) Make the same length of “D” and “E” which is the distance from green line to Bead. Change “Gantry Y Axis
Start Offset” value to find the appropriate position.

Gantry Y Axis Start Offset → Put Plus(+) value to move left side.
Gantry Y Axis Start Offset → Put Minus(-) value to move right side.

1
D E
2

PAPAYA 3D Technical Manual (ver 1.5) 38 / 87


Appendix 1 Alignment&Calibration-Pano/CT

6) After the adjustment, check the box of “Half Scan Stitch” and click “Confirm”.
7) After confirming it, click “Ready & Shot” to take an image of “Half Scan_GCP Phantom”
8) When the shooting is complete, click “Calc, GCP-EV” for running“evGCP” program.

D E

3 4
5

X-ray exposure should be necessary for Cu Filter Alignment.

39 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.14 Half GCP steps

1) Run“evGCP” and click “Bead Detecting Test”


2) Select “2st for seperating layer”
3) According to the red box marked 3, set the mouse curser between two layers and left click.
4) After clicking, a green line will appear between two layers.
5) Click “Process” to start the GCP steps.

PAPAYA 3D Technical Manual (ver 1.5) 40 / 87


Appendix 1 Alignment&Calibration-Pano/CT

6) After finishing GCP process, the screen will appear as below. Then click “Save” and “OK” to complete “Half_GCP”
process.

41 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.15 Step13 :How to take an image of Full GCP Phantom

These tools use for taking an image of full GCP Phantom.

Base Plate Full_GCP Phantom

1) Attach the Base Plate in the Chinrest like ‘A’ of picture. After adjusting horizontally, attach the full GCP
Phantom in Base Plate correspondently for the line.
2) Click on No.1 [0 ° Move and Get Image] shown in the below picture and take an exposure.

PAPAYA 3D Technical Manual (ver 1.5) 42 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4) After getting an image, the image of Full_GCP Phantom appears.


5) The distance of Both sides for Bead & green line is needed to be same. For this, change the value of
No.2 [Gantry Y Axis Start Offset], and take an exposure.

Gantry Y Axis Start Offset → Put Plus(+) value to move left side.
Gantry Y Axis Start Offset → Put Minus(-) value to move right side.

5) After the adjustment, check the box of “Full Scan Stitch” and click “Confirm”.

Process Full Scan Stitch by Selecting FullScanEX Stitch if the equipment option is papaya 3Ds
ENT

6) After confirming it, click “Ready & Shot” to take an image of “Full Scan_GCP Phantom”
7) When the shooting is completed, click “Calc, GCP-EV” for running“evGCP” program.

4
3 5

43 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

4.16 Full GCP Steps

1) When “evGCP” program is activated, Click “Bead Detecting Test”.


2) Select “1st for seperating layer” from the workselect bar, make a line between two layers as the marked 3 on
the picture by left click.
3) The red line will appear on the screen.

PAPAYA 3D Technical Manual (ver 1.5) 44 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4) Select “2st for seperating layer “ fromthe work select bar in below picture.
5) Move the mouse curser between two layers as the marked 4 on the picture by left click.
6) After then the green line will appear.

45 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

7) Select a line“3st for seperating layer” and then left click.


8) A white line will appear as below picture.
9) Click “ Process” start “GCP” calibration.

PAPAYA 3D Technical Manual (ver 1.5) 46 / 87


Appendix 1 Alignment&Calibration-Pano/CT

10) After complete GCP, click no.1 [save] as shown below and then click no2. [ok] to exit Full GCP.

47 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 1 Alignment&Calibration-Pano/CT

11) As above, click[Second projection dataset] as shown as below and then repeatedly execute.

PAPAYA 3D Technical Manual (ver 1.5) 48 / 87


Appendix 1 Alignment&Calibration-Pano/CT

4.17 Step14 : HU Calibration

When taking an exposure HU Phantom, The below plate and

Phantom are needed.

Base Plate HU GCP Phantom

1) After selecting Step 14, attach the base plate to the Chinrest and adjust it horizontally.
2) After selecting No.1(Half Scan) /Select No.1 (Half Scan), click No.2 (Confirm), the path is set automatically.
3) Process the exposure by clicking No.3(Ready & Shot) after confirming.
When finishing the exposure, reconstruction will proceed automatically, and the reconstructed slice will be
displayed on the screen.
4) Input ‘40’ in Edit text of No.4 (Slice index), click No.5 (Rec).
5) Click No.6 (Insert HU), drag the mouse like No.7, make the range of HU.
6) After making the range of HU from A to D like above way, click No.8 (Calibrate HU), and the result values of
HU in the left under part are shown.
7) After selecting Full Scan mode, process for same way from 2) to 6).

49 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

Appendix 2 Alignment & Calibration –Cephalo ..............오류! 책갈피가 정의되어 있지 않습니다.


1. Introduction ................................................................................................................................................... 51
1.1 Regular condition ............................................................................................................................................................................................................. 51
1.2 Preparation ....................................................................................................................................................................................................................... 51
1.2.1 Required Tools ........................................................................................................................................................................................................ 51
1.2.2 JIG............................................................................................................................................................................................................................ 51

2. Device connection ....................................................................................................................................... 52


2.1 PAPAYA_REC program and how to connect with the machine. .................................................................................................................... 52

3. Preparation process .................................................................................................................................... 53


3.1 How to run the program.................................................................................................................................................................................................. 53
3.2 Preparing........................................................................................................................................................................................................................... 54

4. Alignment & Calibration ........................................................................................................................... 55


4.1 Step1 :Set up the detector’s reference positioning ....................................................................................................................................................... 55
4.2 Step2 :1st Collimator’s reference positioning ................................................................................................................................................................ 56
4.3 Step3 : Set standard position for 2ndCollimator............................................................................................................................................................ 57
4.4 Step4 : Detector Calibration ............................................................................................................................................................................................ 59
4.4.1 Detector Model : I3 ................................................................................................................................................................................................ 59
4.4.2 Detector Model : Extor .......................................................................................................................................................................................... 60
4.6 Step5 : Cephalo Scan Calibration .................................................................................................................................................................................... 61
4.7 Step6 : Ear rod and Cephalo Calibration reference positioning .................................................................................................................................... 62

PAPAYA 3D Technical Manual (ver 1.5) 50 / 86


Appendix 2 Alignment&Calibration-Ceph

1. Introduction
1.1 Regular condition

This manual includes necessary contents, especially machine alignment and calibration, of when production and
installation of machine. The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that can cause danger.

Before reviewing this Appendix 2, read “Appendix 1 Align & Calibration –Pano/CT” and familiar with all calibration
steps first.

1.2 Preparation
1.2.1 Required Tools

[Wrench Set]

1.2.2 JIG

[Pin Jig]

51 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

2. Device connection
2.1 PAPAYA_REC program and how to connect with the machine.

1) Turn on the machine and run PAPAYA_REC.exe from C:\GENORAY\PAPAYA.


2) Click “Connect” on the screen to connect with the machine.

3) When the machine is connected with the software, “Connect” button changes for “Disconnect”.

PAPAYA 3D Technical Manual (ver 1.5) 52 / 87


Appendix 2 Alignment&Calibration-Ceph

3. Preparation process
3.1 How to run the program

1) Select “Cephalometric” tab and then click “Calibration Wizard”

53 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

3.2 Preparing

1) After activating calibration Wizard, go to “Preparing” tab.


2) Prepare a Pin jig and remove upper cover.

2, Pin jig

PAPAYA 3D Technical Manual (ver 1.5) 54 / 87


Appendix 2 Alignment&Calibration-Ceph

4. Alignment & Calibration


4.1 Step1 :Set up the detector’s reference positioning

This tool uses for processing this stage.

Pin Jig

1) Select “Step1” and match Detector Moving Block’s hole with the reference point of Cephalo Base.
2) After then, make them to be aligned by the Pin Jig and then remove the Pin Jig. Select “Set current position”.
3) When you click “Set current position”, Result detector offset value will be appeared.
4) This value is a reference position of the detector.

2 3

During removing “Pin Jig” from the part, make sure the detector not to move.

55 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

4.2 Step2 :1st Collimator’s reference positioning

1) Select “Step2” tab


2) Click “Find collimator position” to find the ceph collimator’s values automatically.
3) The message will appear and click “OK” to save values.
4) The value of LR Pos will be updated.
5) Click “Set Position & Get Image” to get an image. Then check the collimator is in a good position as “A”
sample on the picture below.
6) If the image comes up as “B” sample on the picture, add up some values on the current value of UD Pos. If
the images comes up as “C” ,take away some values from the current value of UD Pos.

LR Pos → Add up value from the current value to move right side, take away value from the
current value to move to left side.
UD Pos → Add up value from the current value to move upside, take away value from the
current value to move to down side.

The Pano collimator setting is perfectly complete before starting 1st Collimator’s reference
positioning process.

X-ray exposure should be necessary for 1st Collimator’s reference positioning.

1 4
3 5

B C

PAPAYA 3D Technical Manual (ver 1.5) 56 / 87


Appendix 2 Alignment&Calibration-Ceph

4.3 Step3 : Set standard position for 2ndCollimator

1) Select “Step3”
2) Click “Find 2nd Collimator position” to find “2nd Collimator” value.
3) After searching, the result offset value will appear. Click “OK” and save the value of 2nd Collimator.
4) When you complete to save data, each value will be updated in 2nd collimator position blank.
5) Click “Set position & Get Image” and check the 2nd Collimator is in a good position as the sample picture
below;

In case the location of


collimator is incorrect.

In case the position of collimator In case the location of


is perfectly correct. collimator is incorrect.

1
3 4

57 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

6) If the image is positioned lower than correct position, loosen four bolts, marked 5 on the following picture,
and adjust the position of Lead.
7) If the image position is in the straight position, loosen three bolts, marked 6 on the following picture, and
adjust the position of Lead.

X-ray exposure should be necessary for this process.

PAPAYA 3D Technical Manual (ver 1.5) 58 / 87


Appendix 2 Alignment&Calibration-Ceph

4.4 Step4 : Detector Calibration


4.4.1 Detector Model : I3

1) Select “Step4” tab


2) Click “Auto” to start “Detector Calibration”
3) When the ceph calibration is complete, click “OK”.

X-ray exposure should be necessary for this process.

59 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

4.4.2 Detector Model : Extor

1) Click Step4
2) Select [CEPH_CEPHALO] in No.1 of the below picture.
3) Click [Check All] in No.2, take an exposure by clicking No.3 [Exposure].
4) When finishing the exposure, Avg.Level for whole image in No.4 and Avg for piece of images in ‘status’ are
shown. In addition, when appearing Saturation pop-up like No.5 after appearing the Avg value, please retake
an exposure by adjusting condition.
5) After all exposure and marking of Avg.value are finished, compare them to the value of Ref.Level.
6) If Air Set 0~4 Avg.Level value is within the range of Ref. level. Verify the storage path and click No.7(create).
7) If Avg.Level value is higher or lower than the range of Ref.Level ,change the value of ‘kVp’ and ‘mA’ in No.8,
adjust for correct range.
8) After adapt to Ref.Level standard range, click No.7(create) and complete it.
9) When finish the process, the content of ‘Parameters are generated’ is shown.

If the gap between output value and initial value is lower than 50 → change mA value
If the gap between output value and initial value is higher than 50→ change kVp value

2
4

8 6
3

PAPAYA 3D Technical Manual (ver 1.5) 60 / 87


Appendix 2 Alignment&Calibration-Ceph

4.6 Step5 : Cephalo Scan Calibration

1) Select “Step5”
2) Fold Nasal post parts.
3) Make the rotating part to AP position and remove Ear posts.
4) Click “Auto scan calibration” to start the detector calibration process.
5) After getting an image, click “OK” on the message windows.
6) For taking the Lateral image, click “Auto scan FFC calibration” to get another image.
7) After complete the shooting, click “OK” to complete the calibration process.

X-ray exposure should be necessary for this process.

61 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 2 Alignment&Calibration-Ceph

4.7 Step6 : Ear rod and Cephalo Calibration reference positioning

1) Select “Step6” tab.


2) Attach the Ear rod.
2) After change Cephalo rotation part to Lateral direction, click No.1(confirm) of the picture.
3) After confirming, click No.2(ready & shot) and get the images.
4) After exposure, select No.3 (Cross1(Blue)), select the one center of two circles like No.4, select (Cross2(Yello)),
select the other center of circle.
5) After selecting Cross(1,2), the ‘PASS’ message is shown like No.5. If the ‘FAIL’ message is shown, retake an
exposure by clicking No.6 (Help) with a guidance.

Move Ear rod to Lateral direction and take an exposure.

When selecting the center of circle, place the cross form within the range of the center for circle.

X-ray exposure should be necessary for this process.

3
4
6 1
2

PAPAYA 3D Technical Manual (ver 1.5) 62 / 87


Appendix 2 Alignment&Calibration-Ceph

6) If the ‘PASS’ message is not shown, cancel the articles for ‘A’(Ear rod), Nasal positioner, rotation part cover with
referring to Help button.
7) After removing, loosen a screw ‘B’ a bit, move rotation part side to side, attach Ear rod and retake an exposure.
8) Select Cross (1, 2) to select the result value for PASS, Secure the screws of the B shown in the figure below.
9) After pressing the laser light button, loosen a screw ‘C’ a bit, adjust the laser level horizontally, place the height
in the middle part of the Ear rod hall for the laser, and fix the screw.

Take care not to bend the photo sensor and bracket when moving the rotating rotation.

6) If the position of the Ear rod is indicated by PASS, check if the quality of the printed image is shown as the
image shown ‘good’ below. If similar, the Cephalo calibration is completed without defect.
7) If the image is abnormal, proceed to step 5 again.

When checking the position of Ear rod and Cephalo Calibration, the x-ray could be exposed, please be
careful.

63 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appeendix 3 Seasonging

Appendix 3 Seasonging ..........................................................오류! 책갈피가 정의되어 있지 않습니다.


1. Ready for seasoning ................................................................................................................................... 66
2. Running Seasoning ..................................................................................................................................... 67
3. Completion of Seasoning .......................................................................................................................... 68
4. Notes ............................................................................................................................................................... 68

64 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 3 Seasonging

This manual contains general usage descriptions and other information about Seasoning.
You have to finish the process (Align & Calibration) first and keep power off for a long time to install.

65 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 3 Seasonging

1. Ready for seasoning

1) After turning on the appliance, run the PAPAYA_REC.exe file in C:\GENORAY\PAPAYA folder.

2) When the program runs as shown below, click the [Connect] button to connect to the equipment.

3) When connected to the equipment, the [Connect] button changes to [Disconnect]

4) Run Seasoning tab.

PAPAYA 3D Technical Manual (ver 1.5) 66 / 87


Appendix 3 Seasonging

2. Running Seasoning

1) Select [START] button in the figure.

2) It proceed like the below.


- 1 : Displaying the number and steps for exposure.
- 2 : Displaying the progress of exposure.
- 3 : Displaying status of waiting time.

67 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 3 Seasonging

3. Completion of Seasoning

After finishing Seasoning, Pop-up is shown. Click ‘OK’.

4. Notes

The below table shows Seasoning steps, please be informed.

The number of Interval of


kV mA Exposure time
exposure exposure
5 10 60 4 0.5
2min for waiting
5 10 75 8 1
2min for waiting
5 10 90 12 1

PAPAYA 3D Technical Manual (ver 1.5) 68 / 87


Appendix 4 Auto Save Raw Manage(DB)

Appendix 4. Auto SaveRawManage(DB)............................오류! 책갈피가 정의되어 있지 않습니다.


1. The screen configuration........................................................................................................................... 70

69 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 4 Auto Save Raw Manage(DB)

This manual contains information about the confirming and management of RAW files after exposure.
Ensure that this manual is interlocked to the database, please keep fully in mind with the contents of this manual.

1. The screen configuration

The Auto Save Raw Manage tab is divided by Panorama / Cephalo / CT like below.
Set the number of exposure list for the management in [Auto Save management] of PAPAYA OP Setting.
The button instruction is like a below.
1) After connecting DB, renew the list.
2) Open the folder saved for Raw file.
3) After connecting DB, remove the item.

1 2 3

PAPAYA 3D Technical Manual (ver 1.5) 70 / 87


Appendix 5 Remote Update

Appendix 5. Remote Update .................................................오류! 책갈피가 정의되어 있지 않습니다.


1. Remote update ............................................................................................................................................. 72
2. Run LisaTools................................................................................................................................................. 72
3. Process of Remote Update ....................................................................................................................... 73
3.1 Indivisual Remote update................................................................................................................................................................................................ 76
3.2 Selecting Remote update ................................................................................................................................................................................................ 81

4. List of reference for firmware .................................................................................................................. 85

71 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

This manual includes information and method to use remote update of PAPAYA Series(unify word "PAPAYA").
Keep informed this manual to use equipment effectively.

1. Remote update

1) Find update file.


- Find binary file of remote update.
2) Check network connection.
3) Run remote control program after checking pc network connection for PAPAYA.

2. Run LisaTools

1) Move ‘LisaTools.exe’ to pc wallpaper of equipment’s pc.


2) Run ‘LisaTools’.

[LisaTools ‘exe file’ icon]

PAPAYA 3D Technical Manual (ver 1.5) 72 / 87


Appendix 5 Remote Update

3. Process of Remote Update

1) Press Step1 [Get system Info].

[Execution screen of LisaTools]

73 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

2) As shown in the below, the equipment picture is shown in the center, and the equipment name is displayed in
the upper right corner.
- Check picture and name whether it is same with the equipment you want to update.

[Check equipment name]

In the case of PAPAYA PLUS or SLIM, Check equipment name surely because it is equal to equipment
picture.

PAPAYA 3D Technical Manual (ver 1.5) 74 / 87


Appendix 5 Remote Update

3) Like below, current firmware version is shown per B/D.


- When the light beside each B/D is green, The B/D is possible for remote update.
When the light beside main B/D(FPGA) is grey, it is impossible for remote update, and check if main B/D
update harness(JP02A030) is connected.

[Checking possible B/D for remote update – Impossible for main FPGA update]

[CheckB/D for remote update – It is possible for main FPGA update]

- Once you complete the process, you can run indivisual/selective update.

75 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

3.1 Indivisual Remote update

1) Press Step2 [Select File to upload]

[Step2 Run selecting screen for binary file]

2) Select binary file of B/D for update.


ex) Chinrest B/D Update

[Select binary]

PAPAYA 3D Technical Manual (ver 1.5) 76 / 87


Appendix 5 Remote Update

3) Once you select the file, the current version and the version for update are shown.

[Check current version and update version.]

- Compatible : Compatible: firmware compatibility.


If the number of each binary is different, it is impossible for update.
- Major : Version of function.
- Minor : Version of bug revision.

* Refer to "(P.15) Firmware management system"to find binary file name

[Check firmware compatibility]

77 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

4) Once you press Step3 [Upload files on boards], the update is run.

[Step3 Upload files on boards - Update]

PAPAYA 3D Technical Manual (ver 1.5) 78 / 87


Appendix 5 Remote Update

5) The checking marks beside each B/D, like below, is shown when finishing update.

[Check finish of Chinrest B/D update]

6) If you want to do another update, please do step2 [Select Files to uploard] (P.69) first.
- The checking marks, like below, is shown when finishing main B/D update after chinrest B/D update.

[Check Chinrest B/D & Main B/D Update]

79 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

7) Press step 4 [Reset system] if you no longer proceed with update.


Check if the equipment set initial value after reset.

[Step4Reset(Reboot) system after finishing update.]

8) After finishing Inirializing, turn on and off the power of equipment.

PAPAYA 3D Technical Manual (ver 1.5) 80 / 87


Appendix 5 Remote Update

3.2 Selecting Remote update

1) Press step2 [Select File to updaoad]

[Step2 Select File to upload - Binary File open]

2) Select binary files of B/D for update( ex. Chinrest & LAN Switch).

[Select twice]

81 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

3) Once you finish selecting file, current version and the version for update are shown.

[Versions for select binary B/D]

- Like below, all versions for B/D and update are shown if you select all binary of B/D.

[Select all B/D update binary.]

PAPAYA 3D Technical Manual (ver 1.5) 82 / 87


Appendix 5 Remote Update

4) Once you press step3 [Upload files on boards], the update is started.

[Step3 Upload files on boards - Run update]

- The update process screen is shown from no.1 B/D in order.

[FromMain(MCU) to Main(FPGA), the update process screen is shown.]

83 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

5) After finishing update, checking marks of all status lights are shown.

[After updating B/D, check status light.]

6) Press step4 [Reset (Reboot) system].


Check if equipment is initialized after reset

[After finishing update, do step4Reset(Reboot) system]

8) After finishing Initializing, turn on and off the power of equipment

PAPAYA 3D Technical Manual (ver 1.5) 84 / 87


Appendix 5 Remote Update

4. List of reference for firmware

Below is the chart for management system. Please refer to this chart.

File name filename_B/D name_CANID_function_version_system.bin

PPY : PAPAYA
Product name
P3D : PAPAYA 3D
[3 letters with english]
SLM : PAPAYA SLIM
MAIN : MAIN B/D
COLL : COLLIMATOR B/D
B/D name CHIN : CHINEREST B/D
[3 letters with english] LIFT : LIFTER B/D
LANS : LAN SWITCH B/D
FPGA : MAIN B/D FPGA
MAIN : 01
CHIN : 02
COLL : 03
CAN_ID
LIFT : 04
LANS : 05
FPGA :06
BOOT : Bootloader
function APPL : Application Firmware
[4letters with english] FACT : FPGA Factory Image
UFDA : FPGA Application

Version system A_B_C_D

1 : PAPAYA
Number for A-1
2: PAPAYA 3D
[Product classification]
3 : PAPAYA SLIM

Number for B -1 1 : CURRENT VERSION [CAN]


[F/WCompatibility] 2 : REMOTE UPGRADE

Number for C -2
0~99
[Function modification]

In a quality test : B00~B99


Number for D -3
Mass production distribution : 000~999
[Bug modification]
B : For beta version test

85 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Appendix 5 Remote Update

Example to Firmware file name. Refer to chart for management system.

1) In case of Papaya3D Main B/D Application Firmwar (Version : 2_2_5_001)


→Filename : P3D_MAIN_01_APPL_2_2_5_001.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
01. Main(MCU)
Compatible : 2
Major : 5
Minor : 001

2) In case of Papaya3D Main B/D FPGA Application Image (Version : 2_2_9_010)


→Filename : P3D_ FPGA_06_UPDA_2_2_9_010.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
06. Main(FPGA)
Compatible : 2
Major :9
Minor : 010

3) In case of Papaya Slim Lifter B/D Application Firmware (Version : 3_2_7_003)


→Filename :SLM_LIFT_04_APPL_3_2_7_003.bin
- Partial marking ofLisaToolsUpdate(Refer to P.7)
04. Lifter
Compatible : 2
Major : 7
Minor : 003

PAPAYA 3D Technical Manual (ver 1.5) 86 / 87


Appendix 5 Remote Update

The below chart shows the firmware versioncompatibilityof equipment for remote update.

F/W version for


PAPAYA PLUS PAPAYA SLIM PAPAYA 3D(PLUS)
remote update

MAIN B/D(FPGA) 1_2_9_001 or higher 3_2_9_001 or higher 2_2_9_001 or higher

MAIN B/D(STM) 1_2_0_001 or higher 3_2_0_001 or higher 2_2_5_001 or higher

1_2_2_001(new model B/D) 3_2_2_001(new model B/D) 2_2_2_001(new model B/D)


or higher or higher or higher
LIFT B/D
1_2_1_001(old model B/D) 3_2_1_001(old model B/D) 2_2_1_001(old model B/D)
or higher or higher or higher

CHINEREST B/D - - 2_2_5_001 or higher

COLLIMATOR - - 2_2_2_001 or higher

LAN SW B/D
- - 2_2_5_001 or higher
(LING LED MAIN)

LAN SW B/D
- - 2_2_5_001 or higher
(LIN LED LA)

87 / 87 PAPAYA 3D Technical Manual (ver 1.5)


Genoray Co., Ltd. develops equipments continually. Manual and specifications subject to change
without notice

Genoray Co., Ltd.


512, 560, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea
Tel: +82-31-740-4100 Fax: +82-31-737-8018 [email protected] www.genoray.com

Genoray America Inc.


147 E. Bristol Lane, Orange, CA 92865 USA
Tel: +1-855-436-6729 FAX: +1-714-786-8919 [email protected] www.genorayamerica.com

Genoray EU GmbH
Westhafenstr. 1 13353 Berlin, Germany
Tel: +49-30-509-694-98 Fax: +49-30-530-198-08 [email protected]

Genoray Japan
2F Ishibashi-Bldg, 1-4-15 Shinyokohama,Kouhoku-ku, Yokohama-city, kanagawa, 222-0033 Japan
Tel: + 81-45-620-4971 Fax: +81-45-620-4972 [email protected] www.genorayjapan.co.jp

Boulevard Général Wahis 53, 1030 Brussels, BELGIUM


Tel: +32-2-732-59-54 Fax: +32-2-732-60-03 [email protected]

You might also like