Strain Indicator
And Recorder
Instruction Manual
Micro-Measurements
Division of Vishay Precision Group
-1-
Model P3 Strain Indicator and Recorder
Instruction Manual
1 DESCRIPTION
1.1 GENERAL
The Model P3 Strain Indicator and Recorder is a portable, bat-
tery powered precision instrument for use with resistive strain
gages and strain-gage-based transducers.
-2-
Model P3 Strain Indicator and Recorder
Instruction Manual
1.2 FUNCTIONAL FEATURES
The Model P3 Strain Indicator and Recorder has the fol-
lowing features:
Four input channels.
Direct-reading, backlit LCD display.
Data storage using MMC/SD cards.
Hardware and software support for quarter-, half-
and full-bridge circuits.
Built-in precision bridge completion for 120-, 350-
and 1000-ohm half and quarter bridges.
Automatic and manual zero-balance and calibration.
Intuitive, menu-driven operations.
Full control of all functions via USB Interface.
Selectable analog output.
Portable, lightweight and rugged design.
Battery, USB, or line-voltage power.
Intuitive, user-friendly software.
-3-
Model P3 Strain Indicator and Recorder
Instruction Manual
2 SPECIFICATIONS
Note: Performance may be degraded at high levels of repetitive electrostatic
discharge; however, no damage to the unit will occur.
-4-
Model P3 Strain Indicator and Recorder
Instruction Manual
-5-
Model P3 Strain Indicator and Recorder
Instruction Manual
-6-
Model P3 Strain Indicator and Recorder
Instruction Manual
-7-
Model P3 Strain Indicator and Recorder
Instruction Manual
3 FRONT PANEL
USB Memory Input
Interface Card Slot Terminals
-8-
Model P3 Strain Indicator and Recorder
Instruction Manual
-9-
Model P3 Strain Indicator and Recorder
Instruction Manual
- 10 -
Model P3 Strain Indicator and Recorder
Instruction Manual
4 OPERATION
4.1 Getting Started
The Model P3 is designed for ease of use. This section de-
scribes the operation of the Model P3 Strain Indicator and Re-
corder.
4.2 Power Up
The Model P3 is ready to use as received. The unit is shipped
with two D cells pre-installed.
When the unit is powered up, the firmware is loaded into the
DSP. During this time an audible beep will be heard, indicating
the unit is booting. After approximately two seconds, the beep
tone will cease, unit will display the opening screen.
- 11 -
Model P3 Strain Indicator and Recorder
Instruction Manual
P+ S- D D D P- S+
120 350 1K
- 12 -
Model P3 Strain Indicator and Recorder
Instruction Manual
P+ S- D D D P- S+
120 350 1K
T C
P+ S- D D D P- S+
120 350 1K
C T
T C
E XT E RN A L CA L C IR C UIT
- 13 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 14 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 15 -
Model P3 Strain Indicator and Recorder
Instruction Manual
a .046 inch (1.17 mm) diameter solder well. Each channel uses
a separate (and interchangeable) input plug. Four loose plugs
are supplied with each P3 (one per channel). Wiring connec-
tions are defined as follows:
Amphenol PT06A-12-10P(SR)
- 16 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 17 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Ch1 με Ch2 με
+01000 +00216
Ch3 με Ch4 με
+00524 +00914
REC OFF CAL OFF
- 18 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Ch1 με Ch2 με
+01000 +00216
Paused ->Resume
Ch3 με Ch4 με
+00524 +00914
REC OFF CAL OFF
Pressing the key will update the display if a value larger than
the currently displayed value is measured.
Ch1 με Ch2 με
+01000 +00216
Max Peak ↑=Reset
Ch3 με Ch4 με
- 19 -
Model P3 Strain Indicator and Recorder
Instruction Manual
+00524 +00914
REC OFF CAL OFF
Ch1 με Ch2 με
+01000 +00216
Min Peak ↓=Reset
Ch3 με Ch4 με
+00524 +00914
REC OFF CAL OFF
- 20 -
Model P3 Strain Indicator and Recorder
Instruction Manual
The lower right of the screen indicates whether the calibration
switch is active (CAL ON) or inactive (CAL OFF).
- 21 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Save Settings?
[REC] = Save
[MENU] = Don’t save
- 22 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Ch1 με Ch2 με
+01000 +00216
Ch3 με Ch4 με
+00524 +00914
MAN REC 1 CAL OFF
4.4.1.6 Backlight
- 23 -
Model P3 Strain Indicator and Recorder
Instruction Manual
MENU
Pressing the Menu key will display the Main Menu (see
section 4.5.1).
1 2
** MAIN MENU **
Select Channels
Bridge Type
Gage Factor/Scaling
Recording
Balance
Shunt Calibration
Options
The following table briefly describes the menu items and the
section of this manual associated with the item.
- 24 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 25 -
Model P3 Strain Indicator and Recorder
Instruction Manual
This menu can be directly accessed while in the run mode or the
1 2
* BRIDGE TYPE *
Ch 1: Undef HB/QB
Ch 2: Undef HB/QB
Ch 3: Undef HB/QB
Ch 4: Undef HB/QB
←→ Toggle Menu=Exit
- 26 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 27 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Two active gages in
uniaxial stress field--
one aligned with max-
imum principal strain,
HB adj ,
the other with trans-
verse "Poisson" strain.
Default Poisson’s Ratio
= 0.3.
- 28 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Poisson’s Ratio = 0.3.
Undefined full bridge
circuit. No correction
Undef FB for Wheatstone Bridge
nonlinearity or active
arms.
- 29 -
Model P3 Strain Indicator and Recorder
Instruction Manual
*GAGE FACTOR/SCALING*
Chan : 1
Units : με
Gage Factor : 2.000
*GAGE FACTOR/SCALING*
Chan : 1
Units : mV/V
- 30 -
Model P3 Strain Indicator and Recorder
Instruction Manual
- 31 -
Model P3 Strain Indicator and Recorder
Instruction Manual
the selected engineering units) and the mV/V output at the full
scale value.
*GAGE FACTOR/SCALING*
Chan : 1
Units : psi
Full Scale : 1
F.S. mV/V : 2.000
Dec. Places : Auto
To change the full scale value, use the and keys to se-
lect the desired digit to modify. The digit is changed by using
the and keys. The default full scale value is 1.
The full scale mV/V value must be between +15.625 mV/V and
-15.625 mV/V. A value of 0 is not permitted. The default value
is 2.000.
Note that during manual calibration, the full scale value is fixed,
and the full scale mV/V output is changed. It is therefore rec-
ommended to set this value at or near the expected full scale
value when manually scaling the channel.
- 32 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* RECORDING *
Mode : Disabled
←→ Toggle Menu=Exit
- 33 -
Model P3 Strain Indicator and Recorder
Instruction Manual
4.5.5.2 Manual Recording
In the Manual recording mode, selected channels are recorded
whenever the key is pressed.
- 34 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* RECORDING *
Mode : Manual
Chan 1 : ON
Chan 2 : OFF
Chan 3 : OFF
Chan 4 : OFF
←→ Toggle Menu=Exit
* RECORDING *
Mode : Auto
Chan 1 : 1 sec
Chan 2 : OFF
Chan 3 : OFF
Chan 4 : OFF
←→ Toggle Menu=Exit
- 35 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* BALANCE MODE *
Chan: 1
Mode: Auto
←→ Toggle Menu=Exit
* BALANCE MODE *
Chan: 1
Mode: Manual
Adjust ←→
- 36 -
Model P3 Strain Indicator and Recorder
Instruction Manual
+00010
←→ Toggle Menu=Exit
←→ Toggle Menu=Exit
- 37 -
Model P3 Strain Indicator and Recorder
Instruction Manual
+05000
* OPTIONS MENU *
Display
Outputs
Save Setup
Clock
Advanced
Version
Exit
- 38 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* DISPLAY OPTIONS *
Backlight: 5 sec
Contrast:
←→ Toggle Menu=Exit
- 39 -
Model P3 Strain Indicator and Recorder
Instruction Manual
4.5.9.1 Backlight
Adjusts the duration of the backlight while in the run mode. The
backlight key activates the backlight. Valid options are:
4.5.9.2 Contrast
The contrast control adjusts the contrast of the LCD display.
Temperature variations may make it necessary to adjust the con-
trast of the LCD for the best viewing experience. Change the
selection by using the and keys. A bar graph indicates
the amount of contrast.
- 40 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* OUTPUT OPTIONS *
Excitation: ON
Analog Out: Off
Rejection : 60Hz
←→ Toggle Menu=Exit
4.5.10.1 Excitation
ON: Bridge Excitation is applied during the measurement.
OFF: Bridge excitation is not applied during the measurement.
This provides the ability to identify self-generating noise
sources.
- 41 -
Model P3 Strain Indicator and Recorder
Instruction Manual
4.5.10.3 Rejection
The digital filters in the Model P3 can be tuned to optimize the
noise rejection from power line frequencies. Choose between
50Hz and 60Hz
* SET DATE/TIME *
MM/DD/YY HH:MM:SS
10/17/02 16:47:43
←→ Change Field
↑↓ Change Value
- 42 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Menu = Set
- 43 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* ADVANCED OPTIONS *
Calibrate System
Erase MMC
Poisson Ratio: 0.30
Factory Defaults
Exit
* ADVANCED OPTIONS *
Calibrate System
Erase MMC
Poisson Ratio: 0.30
Factory Defaults
Exit
- 44 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Install calibrator
On channel 1 now
CAL = Calibrate
MENU = Exit
Set calibrator to
0 microstrain
CAL = Calibrate
- 45 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Zero measurement
Exceeds tolerance
Calibration aborted
Menu = Exit
The system will then revert back to the previous calibration val-
ues. If the readings are within tolerance, the system initiates the
next step:
SYSTEM CALIBRATION
Set calibrator to
+30000 microstrain
CAL = Calibrate
- 46 -
Model P3 Strain Indicator and Recorder
Instruction Manual
SYSTEM CALIBRATION
Set calibrator to
-30000 microstrain
CAL = Calibrate
SYSTEM CALIBRATION
Set calibrator to
S+/S- short
CAL = Calibrate
- 47 -
Model P3 Strain Indicator and Recorder
Instruction Manual
REC=Yes Menu=No
MENU
Press the key to erase the card. Press the key to abort
without erasing.
- 48 -
Model P3 Strain Indicator and Recorder
Instruction Manual
* ADVANCED OPTIONS *
Calibrate System
Upgrade Firmware
Erase MMC
Poisson Ratio: 0.30
Factory Defaults
Exit
4.5.18 Version
This option displays the splash screen and the version infor-
mation. Normal operation resumes after approximately 3 sec-
onds.
- 49 -
Model P3 Strain Indicator and Recorder
Instruction Manual
5 Battery Replacement
When the P3 is operated under battery power, battery condition
is indicated by the battery icon, located on the lower right hand
corner of the screen in the run mode.
Ch1 με Ch2 με
+01000 +00216
Ch3 με Ch4 με
+00524 +00914
When the battery icon indicates that the battery is low (by re-
placing the battery icon with "LO" on the lower right corner of
the display), the batteries should be changed. To replace the bat-
teries, follow these steps:
- 50 -
Model P3 Strain Indicator and Recorder
Instruction Manual
Older Units:
- 51 -
Model P3 Data Acquisition Conditioner
Instruction Manual
6 Software Operation
6.1 Overview
The Model P3 is supplied with application software to config-
ure the P3 and collect data via the USB Interface. The software
supports a combination of up to six Model P3 or Model D4 de-
vices, differentiated by serial number. The firmware version of
your P3 (or D4) must be 2.50 or greater to work with this ap-
plication. Use the upgrade firmware feature of your previous
application to update the firmware to the latest level.
- 52 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 53 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 54 -
Model P3 Data Acquisition Conditioner
Instruction Manual
6.6.1 Identify a P3
The serial number of each P3 displays in the title bar of the
window. If you need to identify which P3 is associated with a
serial number, you may select Hardware>>Identify and then
select that serial number. The front panel will blink on
the P3. You may also use the icon on the display area
to illuminate the front panel.
- 55 -
Model P3 Data Acquisition Conditioner
Instruction Manual
No peak values
displayed
Peak values
displayed
Peak values
reset
- 56 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 57 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 58 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 59 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 60 -
Model P3 Data Acquisition Conditioner
Instruction Manual
Value Description
0 Display no decimal places; e.g. 123
1 Display 1 decimal place; e.g. 123.1
2 Display 2 decimal places; e.g. 123.12
n Where n is the number of decimal places.
6.7.4 Scaling
The engineering units for a channel may be microstrain (ue),
millivolts per volt (mV/V), or may be user defined. Scaling
options on the P3 are determined by the units selected.
Millivolts per volt: If the selected units are millivolts per volt
(mV/V), no scaling is necessary.
- 61 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 62 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 63 -
Model P3 Data Acquisition Conditioner
Instruction Manual
The slide bar has the range selections of coarse, medium, fine
and ultra-fine. The coarse selection has the widest range of
balance settings and a coarser step size. Ultra-fine has the nar-
rowest range of balance settings and a small step size. The
medium and fine selections fall in between.
- 64 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 65 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 66 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 67 -
Model P3 Data Acquisition Conditioner
Instruction Manual
6.13 Charting
Charting maps the real time display readings. To begin chart-
ing, select Channels>>Chart. The chart window will display
and charting will begin.
- 68 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 69 -
Model P3 Data Acquisition Conditioner
Instruction Manual
The first line is the data header. All columns are separated by
the ASCII tab character 09 (09h). The line is terminated with a
carriage return and line feed.
- 70 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 71 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 72 -
Model P3 Data Acquisition Conditioner
Instruction Manual
Manual and auto recording modes stop when the stop recording
button is pressed.
In order to make sure your MMC card doesn’t fill up, it’s rec-
ommended that you erase the MMC card after your downloads
have been verified.
6.17 Calibration
6.17.1 Shunt Calibration
Shunt Calibration functionality allows you to adjust the sensi-
tivity of the selected channel (gage factor or full-scale value)
while a calibration signal is applied. The shunt calibration fac-
tor is primarily used to remove any offsets due to lead-wire re-
sistance.
- 74 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 75 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 76 -
Model P3 Data Acquisition Conditioner
Instruction Manual
The default file name for the calibration report, whether created
before or after calibration, is as follows:
SSSSSS_YYYY_MM_DD_M-M_Device_Cal_Report.txt
where:
SSSSSS is the serial number of the P3,
YYYY_MM_DD is the date, and
M-M_Device_Cal_Report.txt is the terminator.
- 77 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 78 -
Model P3 Data Acquisition Conditioner
Instruction Manual
Note: After the first channel has been completed, the Abort
Calibration is available.
- 79 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 80 -
Model P3 Data Acquisition Conditioner
Instruction Manual
SSSSSS_YYYY_MM_DD_M-M_Device_Cal_Report.txt, where:
SSSSSS is the serial number of the P3,
YYYY_MM_DD is the date, and
M-M_Device_Cal_Report.txt is the terminator.
- 81 -
Model P3 Data Acquisition Conditioner
Instruction Manual
A file browser appears that lets you select the firmware up-
grade file. By default this file is named “158000056_P3.hex”.
- 82 -
Model P3 Data Acquisition Conditioner
Instruction Manual
- 83 -
Model P3 Data Acquisition Conditioner
Instruction Manual
WARRANTY
Vishay Measurements Group, Inc., warrants all instruments it manufactures to be free from defect in materials and factory
workmanship, and agrees to repair or replace any instrument that fails to perform as specified within three years after date of
shipment. Coverage of computers, cameras, rechargeable batteries, and similar items, sold in conjunction with equipment
manufactured by Vishay Measurements Group, Inc. and bearing the identifying name of another company, is limited under this
warranty to one year after the date of shipment. The warranty on no rechargeable batteries and similar consumable items is
limited to the delivery of goods free from defects in materials and factory workmanship. This warranty shall not apply to any
instrument that has been:
i) repaired, worked on or altered by persons unauthorized by the Vishay Measurements Group, Inc. in such
a manner as to injure, in our sole judgment, the performance, stability, or reliability of the instrument;
ii) subjected to misuse, negligence, or accident; or
iii) connected, installed, adjusted, or used otherwise than in accordance with the instructions furnished by us.
At no charge, we will repair, at our plant, or an authorized repair station, or at our option, replace any of our products found to
be defective under this warranty.
This warranty is in lieu of any other warranties, expressed or implied, including any implied warranties of merchantabil-
ity or fitness for a particular purpose. There are no warranties which extend beyond the description on the face hereof.
Purchaser acknowledges that all goods purchased from Vishay Measurements Group, Inc. are purchased as is, and
buyer states that no salesman, agent, employee or other person has made any such representations or warranties or
otherwise assumed for Vishay Measurements Group, Inc. any liability in connection with the sale of any goods to the
purchaser. Buyer hereby waives all rights buyer may have arising out of any breach of contract or breach of warranty
on the part of Vishay Measurements Group, Inc., to any incidental or consequential damages, including but not limited
to damages to property, damages for injury to the person, damages for loss of use, loss of time, loss of profits or income,
or loss resulting from personal injury.
Some states do not allow the exclusion or limitation of incidental or consequential damages for consumer products, so the above
limitations or exclusions may not apply to you.
The Purchaser agrees that the Purchaser is responsible for notifying any subsequent buyer of goods manufactured by Vishay
Measurements Group, Inc. of the warranty provisions, limitations, exclusions and disclaimers stated herein, prior to the time any
such goods are purchased by such buyer, and the Purchaser hereby agrees to indemnify and hold Vishay Measurements Group,
Inc. harmless from any claim asserted against or liability imposed on Vishay Measurements Group, Inc. occasioned by the
failure of the Purchaser to so notify such buyer. This provision is not intended to afford subsequent purchasers any warranties or
rights not expressly granted to such subsequent purchasers under the law.
Vishay Measurements Group, Inc. reserves the right to make any changes in the design or construction of its instruments at any
time, without incurring any obligation to make any change whatever in units previously delivered. Micro-Measurement’s sole
liabilities, and buyer’s sole remedies, under this agreement shall be limited to the purchase price, or at our sole discretion, to the
repair or replacement of any instrument that proves, upon examination, to be defective, when returned to our factory, transporta-
tion prepaid by the buyer, within the applicable period of time from the date of original shipment.
Return transportation charges of repaired or replacement instruments under warranty will be prepaid by Vishay Measurements
Group, Inc..
Vishay Measurements Group, Inc. is solely a manufacturer and assumes no responsibility of any form for the accuracy or
adequacy of any test results, data, or conclusions which may result from the use of its equipment.
The manner in which the equipment is employed and the use to which the data and test results may be put are completely in the
hands of the Purchaser. Vishay Measurements Group, Inc. shall in no way be liable for damages consequential or incidental to
defects in any of its products.
This warranty constitutes the full understanding between the manufacturer and buyer, and no terms, conditions, understanding,
or agreement purporting to modify or vary the terms hereof shall be binding unless hereafter made in writing and signed by an
authorized official of Vishay Measurements Group, Inc..