ACUvent
ICU & EMERGENCY CARE VENTILATOR
Features of ACUvent
Designed for ICU & Emergency Care, ACUvent
offers Invasive and Non Invasive treatment to
Adult and Paediatric patients
Latest “Micro-controller technology” for
accurate control of parameters and monitored
values.
Easily movable to any ward with mobile trolley.
Needs only Oxygen supply to run the ventilator. Rechargeable Battery backup for up to 4 hrs.
FiO2 Blending using Venturi system to adjust High Visibility LED displays for breath rate,
from 100 to 50%. Pmax, PEEP, Pressure trigger and Tidal Volume.
OPERATING MODES: Standby mode to enable easy cleaning routine
INVASIVE : VCV,Assist control and SIMV-VC
for the patient. Mains on & Battery use indicator
NON-INVASIVE and INVASIVE : CPAP
NON-INVASIVE : HFNC Therapy
All alarms with audio visual indication.
ACUvent provides an Apnea Alarm in CPAP
modes
Part Number: 1680-10004 Rev 7
Technical Specification
VENTILATION MODES VENTILATION PARAMETERS
INVASIVE Tidal Volume 50 to 999ml
Respiratory rate 0 - 30 bpm
VCV Volume Controlled Ventilation
Oxygen concentration 50 - 100%
VC-AC Assist Control Ventilation Pressure Trigger +20 to -20cm H2O
V-SIMV Synchronized Intermittent Mandatory Ventilation Flow rate (HFNC) 0 to 80 lpm
CPAP Continuous Positive Airway Pressure
NON INVASIVE
MONITORED PARAMETERS
Continuous Positive Airway Pressure
CPAP Maximum Inspiratory Pressure 1 - 80 cm of H2O
High Flow Nasal Cannula Therapy PEEP 2- 20 cm of H2O
HFNC
(Displayed under Pmax window) (Adjustable with external PEEP valve)
Tidal Volume Delivered 999 ml
Flow rate (HFNC) 0 to 80 lpm
SPECIAL FEATURES
• Built in O2 Pressure regulating valve
• Provision for Nebulizer and Humidifier
• Venturi system to blend O2 with atmospheric air
• Mute key for 2mts silence of alarm
• Standby key
PHYSICAL FEATURES
Weight : 4.8 Kgs Dimensions : L 270 x W 245 x H200 mm
INDICATOR LEDS ALARMS
• Operating mode • Disconnection
/ Low Airway Pressure
• Power on
• High Air Way Pressure
• Battery on
• O2 Supply Fail
• Standby on
• Low Battery
• Spontaneous effort in Asst
• AC Failure
Control / SIMV modes Designed by Manufactured by
• Apnea Kriti Kare India Private Limited AiDen Medical Private Limited
Plot No.1, 3/141, Foundry Road, Thuraipakkam,
GAS SUPPLY Chennai - 600 097.
Oxygen 50 - 100 psi (345 - 690 kPa)
ELECTRICAL FEATURES
13485 : 2016 172766
Supply (100 - 240 V AC, 50 / 60 Hz)
Fuse 2A glass fuse
Rechargable Battery Sealed, Lead/acid, 12 V DC, 1.3 Ah
Battery Backup Approx. 4 Hrs at full charge ACCESSORIES :
Mobile trolley; Adult Breathing Circuit with Y Connector
Hoses ; Power cord; Test lung; PEEP valve; HME Filter.
Optional ( Compressor and Humidifier)
Part Number: 1680-10004 Rev 7