DB GB NC56
DB GB NC56
DB GB NC56
Principles of Operation
Two parallel metal rods with a fixed di-
stance between them form a capacitor
when charged with an A.C. voltage.
When these rods are immersed in a li-
quid medium, the capacitance value is a
function of the immersed length. With
the probes fixed vertically downward, Features
the immersed length is proportional to • Rugged design and construction, IP67
the liquid level. The capacitance of the
• Built-in electronic signal conversion
probes is therefore a function of the le-
• Very easy level setting
vel. An electronic circuit module in NC56
Level Sensor converts the capacitance • EC type approval
value to a level measurement and trans-
Applications
mits as a linear standard electrical si-
gnal: 0-20 mA, 4-20 mA, 0-10 V DC, The Model NC56 Capacitive Level Sensor can be
0-5 V DC, 1-5 V DC or 2-10 V DC. used in a variety of applications areas, for example:
DB_GB_NC56.fm
• Marine applications
Specifications
General Connection Diagram
Sensing technique Capacitance sensing
Level measuring range 400 - 2000 mm (other ranges available against special order)
Operating pressure 10 bar, max.
Temperature Max. 80°C (ambient and media)
Number of electrodes 2 (3 for diesel applications)
Process connection Threaded G1¼" (M). With protective tube: G 2" (M)
Degree of protection IP67
Electrical
Operating voltage 9 - 32 V DC 9 - 32 V DC 12 - 32 V DC 12 - 32 V DC 12 - 32 V DC
Supply current (without approx. 30 mA approx. 30 mA approx. 30 mA approx. 30 mA approx. 30 mA
signal)
Output signal 0 - 20 mA 4 - 20 mA 0 - 10 V DC 0/1 - 5 V DC 2 - 10 V DC
Output load impedance (UB - 9 V) / 20mA (UB - 9 V) / 20mA > 5 k Ω >5kΩ >5kΩ
UB = Operating voltage
Electrical Connection Male M12 type round shell connector
Materials
Housing Plastic
Media: contact Stainless steel 1.4404, ECTFE, shrinking plastic tube (polyolefine)
Approval EC type-approval acc. to 72/245/EEC, 95/54/EC
Type-approval No. e13*72/245*95/54*2182*00
Installation
The NC56 Capacitive Level Sensor is installed vertically downward at the top of the
vessel in which liquid level is to be measured. It is fitted into a G1¼-F threaded socket,
located as close as possible to the center of the tank. The lower ends of the rods must
be at least 10 mm above the tank bottom to avoid contact with any sludge that may be
present. The electrode rods can be cut at site to a shorter length, if required, but must
be of equal length.
every 500 mm
every 500 mm
M12 connection
adjusted to electrode length
every 500 mm
(4,8)
Ordering Code
Function
Water / waste water
(1 electrode coated with shrinking plastic tube /
1 uncoated electrode, stainless steel 1.4404)..................... > 2
Diesel
(3 uncoated electrodes, stainless steel 1.4404)................... > 3
Sewage
(1 electrode ECTFE coated /
1 uncoated electrode, stainless steel 1.4404)..................... > 4
Firefighting foam
(1 electrode ECTFE coated /
1 uncoated electrode, stainless steel 1.4404)..................... > 5
Signal Output
Supply Voltage
Technische Änderungen vorbehalten • Subject to change without notice • Changements techniques sous réserve
Fischer Mess- und Regeltechnik GmbH • Bielefelder Str. 37a • D-32107 Bad Salzuflen • Tel. +49 5222 9740 • Fax +49 5222 7170 • eMail: [email protected] • www.fischermesstechnik.de