Internship Report
Internship Report
BY
ODONG SAM
REG. NO: 19/U/28364
Signature…………………………………………………
i
APPROVAL
This is to clarify that this report was done under supervision and is ready for the submission
as partial fulfillment of the award of Bachelor of Science in Biomedical Engineering at
Makerere University.
ii
ACKNOWLEDGEMENT
First, I sincerely thank the Almighty God for helping me to successfully prepare this report. I
am thankful to ARRAY MEDICAL SUPPLIERS AND CONSULTS LIMITED that gave me
the opportunity to carry out internship in their organization and giving me a lot of
information and hands on practical skills in the filed which has helped me relate the
knowledge studied at university and the professional working environment. I also pay my
heart rendering respect and appreciation for the consideration and co-operation which is
given to complete my industrial training work assigned.
I express my profound thanks to the Managing Director Mr. Philip Ddungu Yeko of Array
Medical Suppliers and Consults Limited and the biomedical engineers i.e., Eng. Jonah
Muyanja, Eng. Jimmy Togura, Eng. Phillip Lukwago, Eng. Moses Ochen and Eng. Gabriel
Ategeka whose encouragement and hands on training was very instrumental in accomplishing
this task successfully. lastly, I would like to thank my university supervisor Ms. Racheal
Nalwoga for her encouragement and advise during the supervision process, this was key in
appreciation the importance of the entire internship process.
iii
ABSTRACT
This report provides details of activities carried out during the internship training at ARRAY
MEDICAL SUPPLIERS AND CONSULTS LIMITED from 6 th September to 13th October
2022 (6 weeks). It consists of four chapters. Chapter one has the introduction, background of
the field attachment, objectives of internship, background of the field attachment
organization, organization structure and main activities of the organization. The second
chapter is made up of the experience acquired in the field i.e. position occupied, duties and
responsibilities, authority etc. The main activities carried out by the engineering department
of the company were; troubleshooting, maintenance, repair, calibration, procurement and
installation of all sorts of medical equipment and also electrical fault repairs. These activities
were done mainly in the clinics, laboratories and equipment maintenance workshop and the
hospitals. Chapter three is made up of the evaluation of the field attachment which includes
interesting experiences, challenges faced and how they were overcome and benefits derived
from field attachment. Chapter four is made up of the conclusions and recommendations
made for both the organization and university.
iv
TABLE OF CONTENTS
DECLARATION-----------------------------------------------------------------------------------------i
APPROVAL----------------------------------------------------------------------------------------------ii
ACKNOWLEDGEMENT----------------------------------------------------------------------------iii
ABSTRACT----------------------------------------------------------------------------------------------iv
LIST OF FIGURES-----------------------------------------------------------------------------------vii
LIST OF TABLES------------------------------------------------------------------------------------viii
LIST OF ABBREVIATIONS------------------------------------------------------------------------ix
1.2 Objectives-------------------------------------------------------------------------------------------1
1.4.1 Vision------------------------------------------------------------------------------------------2
1.4.2 Mission-----------------------------------------------------------------------------------------2
2.3 Methodology---------------------------------------------------------------------------------------4
2.3.1 Demonstration---------------------------------------------------------------------------------4
2.3.2 Observation------------------------------------------------------------------------------------5
v
2.4.2 Installation-------------------------------------------------------------------------------------5
2.9 Microscope----------------------------------------------------------------------------------------22
2.12 Coagulometer-----------------------------------------------------------------------------------27
2.15 Centrifuge----------------------------------------------------------------------------------------30
3.2 Challenges-----------------------------------------------------------------------------------------31
4.1 Recommendations--------------------------------------------------------------------------------33
4.2 Conclusion----------------------------------------------------------------------------------------33
REFERENCES-----------------------------------------------------------------------------------------34
APPENDIX----------------------------------------------------------------------------------------------36
vi
LIST OF FIGURES
Figure 15: WBC differential bath before service(left) and after service (right)____________36
vii
LIST OF TABLES
viii
LIST OF ABBREVIATIONS
Eng. Engineer
HGB Hemoglobin
PCT Plateletcrit
ix
CHAPTER ONE: INTRODUCTION
1.1 Background industrial training
Internship is an opportunity given to a student to have a hands-on experience in a real
working world setting under the guidance and supervision of the experienced supervisor. This
exercise gives a student chance to put into practice what he or she has learnt in the class
environment and it involves wide range of tasks, responsibility, achievements, challenges,
opportunities, observations and experiences.
An internee should be ready to take up assignments and specific tasks which often come with
no financial facilitations, these tasks basically come from your field supervisor and also the
other staff at large. Internship is carried out within a set period of time and by the end of it all,
a student should be able to appreciate the skills gained, experience, challenges and
opportunities of a real working world environment.
1.2 Objectives
1.2.1 General objectives
To understand practical applications, develop skills and advance knowledge necessary for a
biomedical engineer by participating in various equipment management principles and
protocols of the Hospital.
1
procurement, project management and administration of the company. The management team
proactively gathers feedback, identifies changes in business environment, review work
processes and communicates keys learning points and company policies to all staff at regular
meetings.
The field work teams are crucial in the delivery of their services and solution to clients, in
order to ensure that everyone is equipped with the right skills, knowledge and attitude, a
comprehensive training program is put in place to constantly upgrade people in technical and
management skills, the company believe firmly in providing the right training, accredited
certificate and practical knowledge for its engineers, in order for them to execute their duties
and responsibilities confidently. The aim of the company is to stay relevant and reliable to the
ever-changing market place and client's requirement.
1.4 Company profile
1.4.1 Vision
To be a respectable Biomedical Engineering company delivering beyond expectation, always.
1.4.2 Mission
ARRAY MEDICAL SUPPLIERS AND CONSULTS LIMITED delivers quality and reliable
medical suppliers and hospital equipment, installations and support to meet all your
biomedical engineering needs.
We offer services to the private and public medical facilities. ARRAY MEDICAL
SUPPLIERS AND CONSULTS LIMITED is committed to implementing the latest health
technologies, skills and practices to improve our clients with the solutions needed for success
in today's health sector.
We provide clients with experience, quality, dedication, and professionalism and as a result.
2
1.4.3 Company organization structure
Managing Director
(CEO)
Technical manager
Biomedical engineer
(technician)
3
CHAPTER TWO: INTERNSHIP EXPERIENCE
2.1 Position Occupied
During the internship period, I worked as a Biomedical Engineering intern. I mainly
performed activities such as equipment service, repair, planned preventive maintenance,
installation, user training among others in the various sites i.e., hospitals, medical centers,
clinics etc.
2.3 Methodology
There were a number of methods of training that were used by my supervisors that were all
aimed at gathering information about different medical equipment during the industrial
training. The information could be how the equipment works, the possible errors, the cause of
breakdown and so many others. Some of the methods employed to gather information
include;
2.3.1 Demonstration
This method basically involved physical interaction with the machines. In involved the field
service engineers practically demonstrating to us the operation of the devices physically as
we observed. The FSE also demonstrated how the various parts of the machines operates as
they performed their various tasks. The main objective of demonstration was to enable us
interact physically with the equipment we were learning about in class and also be able to
identify them and troubleshoot them accordingly.
4
2.3.2 Observation
This method involves one observing or making use of naked eyes to see what the equipment
is doing, how the equipment is operating, how the equipment is installed, and how it is
deviating from the normal operation.
Types of maintenance
Corrective/breakdown
This is work done to take corrective action in the event of a breakdown of equipment.
Planned/scheduled/preventative maintenance
This is work done in a planned way before repair is required and scheduled time for work
circulated well in advance. It involves cleaning, safety tests, and calibration and makes sure
that any problems are picked up while they are small.
2.4.2 Installation
The first step of proper medical equipment setup is placing the equipment item(s) in the
specified area of use. The equipment was located in the area where it is to be used. The next
step is inspection. Inspection is done for each piece of equipment for the following; missing
parts, damaged parts and mismatched parts. After assembling the medical equipment, testing
for functionality is the next step. This is because sometimes everything appears to be fine yet
5
there may be a glitch that cannot be seen unless the device is powered and tested for its use.
Additionally, testing uncovers assembly mistakes before any permanent damage occurs.
Then all packing materials and other debris are removed from the medical equipment area.
This is to help ensure a clean and effective workspace for equipment users. It also prevents
contamination and other mishaps during work. Finally, after all inspections have been done
and everything is working properly, it is crucial that each applicable member of staff is
instructed on the proper use of the medical equipment since these are new models that had
not been used by the hospital before. Clear, detailed instructions were to prevent misuse,
which could lead to poor patient care or damaging of the equipment.
6
HGB measurement
HGB is determined by the colorimetric method. The WBC/HGB dilution is delivered to the
bath where it is bubble mixed with a certain amount of lyse, which converts hemoglobin to a
hemoglobin complex that is measurable at 525 nm. An LED is mounted on one side of the
bath and emits a beam of monochromatic light, whose central wavelength is 525nm, and then
is measured by a photo-sensor that is mounted on the opposite side. The signal is then
amplified and the voltage is measured and compared to the blank reference reading (readings
taken when there is only diluent in the bath). The HGB is calculated per the following
equation and expressed in g/L.
RBCs/PLTs are counted and sized by the impedance method. This method is based on the
measurement of changes in electrical resistance produced by a particle, which in this case is a
blood cell, suspended in a conductive diluent as it passes through an aperture of known
dimensions. An electrode is submerged in the liquid on both sides of the aperture to create an
electrical pathway. As each particle passes through the aperture, a transitory change in the
resistance between the electrodes is produced. This change produces a measurable electrical
pulse. The number of pulses generated indicates the number of particles that passed through
the aperture. The amplitude of each pulse is proportional to the volume of each particle. Each
pulse is amplified and compared to the internal reference voltage channels, which only
accepts the pulses of a certain amplitude. If the pulse generated is above the RBC/PLT lower
threshold, it is counted as an RBC/PLT.
Manufacturer Dymind
Model DH36
7
Activity: General service and maintenance
Work done;
General PPM
Cleaning of the counting chamber was done using probe cleanser solution
Cleaning of the exhaust valves
Fastening of loose tubing
Cleaning of the general interior surface of the machine
Running of the calibrators and validation using quality control solutions.
Writing and signing of the equipment service report.
8
Kairos Medical Centre (Namuwongo)
Table 2: Showing details of Zybio CBC at Kairos Medical Centre
Manufacturer Zybio
Model Z3
Activity: Installation
Work done:
The unit was unpacked from the box and placed on a firm flat surface.
The reagents i.e. diluent and lyse where connected to the machine using the reagents
connection tubes.
The machine was powered and the startup initiation process done.
Testing of the machine using a fresh control sample was done to confirm full
functionally.
User training for the laboratory staff was done and an installation report written and
signed.
Manufacturer Dymind
Model DH56
Activity: Repair
Fault description; machine failure to connect to the computer (network connection loss)
9
Work done:
It was discovered that the IP address of the machine from the computer set up was
correct but still could not establish a connection machine software.
The manufacturer was contacted using team viewer to try and rectify this issue but he
concluded that the entire main board of the machine needed to be replaced as the
damage caused was non-repairable.
The facility was recommended to purchase a new main board in order to rectify this
issue.
The cause of this according to the manufacture was poor electrical wiring of the
building (no earth protection)
Kampala hospital
Table 4: Showing details of Dymind 5-part CBC at Kampala hospital
Manufacturer Dymind
Model DH56
10
Activity: Repair
Fault description; machine unable to differentiate WBCs (No results for WBC differentials)
Work done
The counting chambers i.e. WBC and RBC were removed and cleaned manually
using clep solution and probe detergent.
The differential bath was also removed and air blown inside to remove any hard blood
clots within its channels.
The removed parts were assembled and the machine retested.
On testing the problem was persistent and after consultations with the manufacturers,
the recommended that the entire flow cell unit be replaced as it is the one primarily
responsible for differentiation of the WBCs.
11
Friends Poly Clinic
Table 5: Showing details of Zybio CBC at Friends Poly Clinic
Manufacturer Zybio
Model Z3-1
Work done:
12
Nsangi Medical Centre
Table 6:Showing details of Zybio CBC at Nsangi Medical Centre
Manufacturer Zybio
Model Z3
Activity: Repair
Cause: poor installation of the printer rolls into the printer. This cause damage to the printer
gear rollers hence preventing it from printing the test results.
Work done;
The thermal printer of the machine was extracted and checked.it was discovered that
the paper drive gears were broken.
The thermal printer unit was replaced with a new thermal printer as the broken gears
could not be repaired.
The laboratory staff were trained on the proper procedure of installing the thermal
paper.
A job card was written and signed by the laboratory in charge.
Manufacturer Zybio
Model Z3-1
13
Activity: General service and maintenance
Work done;
Manufacturer Edan
Model H30
Work done;
14
Figure 5: Assembling the RBC and WBC counting chambers
2.6 Suction Machine
A suction machine is a portable apparatus used in patient wards and operating theaters for
aspiration of fluids from the mouth, airways, or surgical sites by sucking the material through
a catheter into a reservoir bottle.[6] Suctioning is a procedure that removes excess secretions
from the mouth and thorax, from the nose and the throat and from the wind pipe (trachea)
using mechanical aspiration device.
15
Sterile patient tubing: Patient tubing attaches to the suctioning tip and carries the
patient’s secretions into the collection canister. Sterile patient tubing should be
properly disposed of after each suctioning session.
Disposable canister: the disposable canister holds the patient’s secretions and often
provides overflow protection capabilities in case too much fluid is suctioned out of
the patient. This canister should be disposable to ensure all parts of the suction
machine remain sterile.
Power cord: Portable suction machines come with a power cord that can be used to
charge the machine when you’re close to a power outlet.
Filters: Ideally, a disposable canister should support the use of bacteria/viral filters to
prevent contamination within the aspirator’s inner components. Certain filters may
also be used to protect against dust and dangerous gases that can damage the machine.
Faults description
Repairing a Suction Machine and troubleshooting. This involved the use of a multi
meter, screw driver and our naked eyes to compare a normal working Suction
Machine to the Machines that were encountered during the training.
Among the Suction machines, one had its tube that connects to the Vacuum
misplaced.
Work done.
Replacing the tube on the Suction Machine whose tube that connects to the Vacuum
was misplaced.
Troubleshooting the machine by checking the power source, continuity and we
realized the fault was with the fuse which was replaced with the new one.
16
Checking if there is power at the power supply using a tester or multimeter.
Checking if the fuse in the top plug is working using a multimeter.
Checking for continuity in the wire of the machine using a multimeter.
Ensuring that there are no any blockages in the tubing and filters in the machine.
Checking for proper continuity in the switch.
Checking if the pump is in good condition.
Checking if the winding in the pump is not burnt.
Principle of operation
Atmospheric air consists of approximately 80% nitrogen and 20% oxygen.[8] An oxygen
concentrator uses air as a source of oxygen by separating these two components. It utilizes
the property of zeolite granules to selectively absorb nitrogen from compressed air.[9]
Atmospheric air is gathered, filtered and raised to a pressure of 20 pounds per square inch
(psi) by a compressor. The compressed air is then introduced into one of the canisters
17
containing zeolite granules where nitrogen is selectively absorbed leaving the residual
oxygen available for patient use. After about 20 seconds the supply of compressed air is
automatically diverted to the second canister where the process is repeated enabling the
output of oxygen to continue uninterrupted. While the pressure in the second canister is at 20
psi the pressure in the first canister is reduced to zero. This allows nitrogen to be released
from the zeolite and returned into the atmosphere. The zeolite is then regenerated and ready
for the next cycle.[10]
By alternating the pressure between the two canisters, a constant supply of oxygen is
produced and the zeolite is continually being regenerated. Individual units have an output of
up to five liters per minute with an oxygen concentration of up to 95%.
Maintenance
Always run the machine for few hours after maintenance before patient use.
Check if the zeolite canister has expired. Granules become grey when not effective.
Clean the humidifier and tubes wash in warm soapy water and ensure filter is dry
before replacing.
Remove any dust with a damp cloth and dry off.
Remove any dust with a damp cloth and dry off.
Check alarm system battery, an audible alarm, replace battery if necessary.
Check all screws, connections and parts are tightly fitted.
Check that the flow rate varies with flow control.
18
If the unit is not operating and there is no power failure alarm, possibly the alarm
needs to be replaced.
No oxygen flow which can be due loose tube connections, blocked flow meter. This
can be rectified by tightening the loose tubing, replace flow meter or flushing the
tubes with water and make them dry.
Activities done
Tightening of the loose connection tubing.
Servicing of the concentrator.
User training and how to troubleshoot.
Principle of operation
Analyzers are highly automated to maximize throughput, to improve user safety from
biohazards, and to diminish the risk of cross-contamination.[12] Samples are loaded into the
19
tray and a pipette aspirates a precisely measured aliquot of sample and discharges it into the
reaction vessel; a measured volume of diluent rinses the pipette. Reagents are dispensed into
the reaction vessel. After the solution is mixed (and incubated, if necessary), it is either
passed through a colorimeter, which measures its absorbance while it is still in its reaction
vessel, or aspirated into a flow cell, where its absorbance is measured by a flow-through
colorimeter. The analyzer then calculates the analyte’s chemical concentrations. Tests are
programmed by the user via a keypad or bar-code scanner and the results are displayed on
screen or sent to a printer or computer.[13]
The types of problems that signaled us to check the flow-cell included: noisy and drifting
baselines, poor reproducibility, and poor peak shape. These problems were resolved by either
cleaning or replacing the glass flow-cell windows.
Procedure
In order to closely inspect the window simply remove the setscrews on each side of the flow-
cell with the proper size Allen wrench. Remove the O-rings on each side carefully with
tweezers. The O-rings should be replaced if they are cracked, torn, or compressed since they
will not properly press the glass window into the flow-cell housing. Next, remove the glass
windows, which should easily fall out by gently tapping; sometimes they may need to be
tapped more aggressively if there is considerable buildup or much time has gone since it was
last rebuilt.
The glass window should be inspected for physical damage and replaced if any condition
exists that may interfere with the light paths such as stains, buildup/coatings, breaks or
cracks. If everything looks good, clean everything thoroughly with a solvent (alcohol) and
dry thoroughly. Be careful not to contaminate or smudge the glass window when returning it
into the slot. Next, carefully push the O-ring back into place (all the way to the bottom) so as
20
not to damage the rubber with the threads of the setscrew. Finally, tighten the setscrew
securely on each side.
Be sure to leak test the flow-cell before placing it back onto the detector so that no damage
will occur if there is indeed a leak. Test the flow signal first with water, then with reagents,
and finally, with standards.
Manufacturer BIOBASE
Model B 200
Procedure
Setting up of the machine on a form concrete surface was done. The distilled water
and waste tubes were connected to the machine.
The test reagents were added into the reagent’s containers into the machine i.e.
reagent R1 for single reagent and reagents R1 and R2 for double reagents.
The machine was connected onto the computer using an ethernet cable through the
machine’s software and programming of the reagents was done.
Machine testing using QC samples was to ascertain functionality, repeatability,
accuracy and consistency in the way the machine produces results. User training was
done.
An installation job card and user training report were written and signed by the
laboratory manager.
21
Figure 8: Reagent programming of a chemistry analyzer
2.9 Microscope
The eye piece magnifies the primary image and brings it into focus with the retina of the eye.
The objective and eyepiece are systems of lenses where the distance from each other is fixed
22
by the mechanical length of the tube. To adjust to form a clear image, either the specimen
stage or tube as a whole has to be moved by means of course and fine adjustment knobs.
Total magnification can be got from a product of the magnification of the ocular lens and the
objective lens.
Microscope troubleshooting and repair at the workshop.
Faults of the microscope
No light from the bulb.
Potentiometer faulty.
Due for servicing.
Work done
Replacement of the faulty halogen lamp with a new 6V 20W one was done.
Replacement of the faulty potentiometer (variable resistor) 500K ohms with a new
one was done.
Cleaning of the eye pieces
Cleaning of the lens using lens paper.
23
2.10 Vital Signs Monitor (Patient Monitor)
Vital signs monitor usually provide data on body temperature, pulse rate, respirations, and
blood pressure, though some clinicians also count other measurements along with the basic
vital signs, including pulse oximetry and end tidal carbon dioxide.
Principle of operation
The vital signs have the SPO2 (Oxygen Concentration) measuring probe which also has the
CO2 sensor, the Cuff that measures (NIBP) Human Non-invasive Blood Pressure. To
measure blood pressure/ NIBP, the cuff is placed around the bare and stretched out upper
arm, and inflated until no blood can flow through the brachial artery. Then the air is slowly
let out of the cuff. As soon as the air pressure in the cuff falls below the systolic blood
pressure in the brachial artery, blood will start to flow through the arm once again.[19] As the
blood moves around the body, it puts pressure on the walls of the vessels. Blood pressure
readings are made up of two values:
Systolic blood pressure is the pressure when the heart beats – while the heart muscle is
contracting (squeezing) and pumping oxygen-rich blood into the blood vessels. Diastolic
blood pressure is the pressure on the blood vessels when the heart muscle relaxes. The
diastolic pressure is always lower than the systolic pressure. Blood pressure is measured in
units of millimeters of mercury (mmHg). The readings are always given in pairs, with the
upper (systolic) value first, and followed by the lower (diastolic) value.[20] So, someone who
has a reading of 132/88 mmHg (often spoken “132 over 88”) has a Systolic blood pressure of
132 mmHg, and a diastolic blood pressure of 88 mmHg
1. Oxyhemoglobin and deoxyhemoglobin, which differ in their absorption of red and infra-
red light (that is, Spectrometry)
2. The volume of arterial blood in the tissue (and hence, light absorption by that blood)
changes during the pulse.
The SpO2 probe has a pulse oximeter which determines SpO2 by passing red and infrared
light into an arteriolar bed and measuring changes in the light absorption during the pulsatile
cycle. The Red and Infrared low-voltage light emitting diodes (LEDs) in the oximetry probe
serve as light sources; a photodiode serves as the photo detector. This percentage should be
24
between 94 percent and 100 percent, which indicates a healthy level of hemoglobin carrying
oxygen through the blood. If it is less than 90 percent, one must see a doctor
Tasks done
Checking for all the accessories of the vital signs monitor because it was a new one and I had
to test it for the client and do user training on how the machine works and how different
accessories are connected to the machine.
25
2.11 Tables (Operating Theaters and Delivery)
Tables are required to hold the patient in a position that is comfortable for both the patient
and for medical staff during procedures.[21] They can include dedicated supports for head,
arms and legs and often have movable sections to position the patient appropriately. They are
made both with wheels and on static platforms and can have movements powered by electric
motors, hydraulics or simply manual effort. They can be found in emergency departments,
operating theatres and delivery suites.
Mode of operation
Where the table has movement, this will be enabled by unlocking a catch or brake to allow
positioning. Wheels have brakes on the rim or axle of the wheel, which locks for moving
sections will normally be levers on the main table frame. Care should be taken that the user
knows which lever applies to the movement required, as injury to the patient or user may
otherwise result. The table will be set at the correct height for patient transfer from a trolley
then adjusted for best access for the procedure.
Work done
Disinfection of the tables in the different theatres i.e. maternity theatre and main
operating theatre.
Lubrication of the moving parts so that the joints could freely operate. This was all
done during servicing or preventive maintenance at Kampala hospital.
Two operating tables were assembled in company’s workshop
26
2.12 Coagulometer
Working principle.
Special cuvettes with a steel ball inside are placed on the measuring positions in instrument
related racks. As the measuring block is sloping slightly the ball always remains due to
gravity at the deepest point of the cuvette. In the height of this point there is a magnetic
sensor. At first the sample is pipetted into a measuring cuvette, then if required the first
reagent is added and the incubation is started. The instrument turns the cuvette with the
adjusted speed around the longitudinal axis. When the incubation is finished, the start reagent
is added and the measurement is start simultaneously. When the coagulation begins the
growing clot pulls the ball out of the basic position and the magnetic sensor detects a
magnetic impulse which causes the end of the measurement.[22]
Activity: service
Work done;
Working principle
The light's absorption and reflection form the basis of the urine analyzer test principle.[23]
The multi-linked reagent strip with various solidified reagents already present receives the
liquid sample right away. The color of the module containing several special reagents on the
multi-linked reagent strip changes in response to the corresponding chemical composition in
the urine. The depth of color is in line with the particular chemical of the urine sample. The
multi-link test strip is inserted into the colorimetric injection tank of the urine analyzer, where
each module is exposed to the instrument's light source and generates various amounts of
reflected light. The component concentration is proportionate. The device receives a light
signal with a variety of intensities and transforms it into the matching signal. The electrical
27
signal is processed by the microprocessor to determine the reflectance of each test item. The
result is then automatically written out in a qualitative or semi-quantitative format after being
compared with the standard curve and corrected to the observed value.
Activity: Repair.
Fault description; failure of the sample/urine strip driver motor to take samples into the
analyzer.
Work done;
28
micro-organisms. This helps to maintain quality control of the material being worked with as
it filters both the inflow and exhaust air.
Principle of operation
The BSC provides personnel and environmental protection, but no product protection. Air
movement is similar to a chemical fume hood, but it has a HEPA (High Efficiency Particulate
Air) filter in the exhaust system to protect the environment from bioaerosols. Unfiltered room
air is drawn across the work surface by a fun. Personnel protection is provided by this inward
airflow as long as a minimum velocity of 75 linear feet per minute (lfpm) is maintained
through the front opening and appropriate work practices are observed. BSC is hard-ducted
i.e., directly connected to the building exhaust system, and the building exhaust fan provides
the negative pressure necessary to draw room air into the cabinet.
Work done: The old HEPA filter was removed and replaced by a new on.
29
2.15 Centrifuge
The centrifuge works on the principle of gravity and the generation of the centripetal force to
sediment different fractions.[24] The rate of sedimentation depends on the applied centrifugal
field (G) being directed radially outwards G depends on Angular velocity (ω in radians/sec)
Radial distance (r in cm) of the particle from the axis of rotation (G = ω2r)
Activity: repair
Fault description; failure of the machine door to open automatically after processing samples.
Work done;
The centrifuge was opened and the locking system was checked. It was discovered
that opening switch system had rusted due to blood spills on it.
WD 40 was used to clean and loosen the rusted parts of the locking and opening
system. The device was assembled and the problem was solved.
30
CHAPTER THREE: EVALUATION OF FILED ATTACHMENT
3.1 Experience gained and Skills Attained
For my technical skill development as intern trainee, I was assigned a number of duties to
carry out in the workshop and outside the workshop in the field at different hospitals and
medical centers.
3.2 Challenges
Filed work was minimal as most of the machines were being carried to the workshop
for repair so no chance to visit other departments in the hospitals.
Limited access to different tools used as some were limited to only the experienced
engineers.
31
Access to spare parts such as fuses, transformers, motors, among others was limited
for some of the medical equipment and this delayed hands-on works on such
equipment that could have been done to rectify of current faults.
Reaching work place late due to bad weather condition.
Limited time for the industrial training.
Some machines did not have manuals.
32
CHAPTER FOUR: RECOMMENDATIONS AND CONCLUSION
4.1 Recommendations
4.1.1 Recommendations to the organization:
Equipment manuals could be acquired early from the manufactures to avoid delays
during equipment troubleshooting
Complicated flow diagrams should be translated into English by the manufacturers in
order to ease their interpretation.
The workshop should be given more funds in order to admit more internees and
purchase more tools to be used in the workshop
The organization should work hand in hand with the hospitals, medical centers, clinics
etc. to make proper schedules for service and maintenance of their equipment.
The organization should facilitate the trainees when going out to the field.
4.2 Conclusion
In a nutshell, this internship has been an excellent and rewarding experience. I can conclude
that there has been a lot I have learnt from my work at Array Medical Suppliers and Consults
Limited. Needless to say, the technical aspects of the work I have done are not flawless and
could be improved provided enough time. As someone with no prior experience with some
medical equipment whatsoever, I can say that most of the objectives of this undertaking were
archived. Two main things that I have learned outside engineering are time management and
self-motivation as key to being a proper engineer.
33
REFERENCES
[1] G. Shu, H. Lu, H. Du, J. Shi, and G. Wu, “Evaluation of Mindray BC ‐3600 hematology
analyzer in a university hospital,” Int J Lab Hematol, vol. 35, no. 1, pp. 61–69, 2013.
[2] A. Ojaghi et al., “Label-free hematology analysis using deep-ultraviolet microscopy,”
Proceedings of the National Academy of Sciences, vol. 117, no. 26, pp. 14779–14789, 2020.
[3] T. Sun and H. Morgan, “Single-cell microfluidic impedance cytometry: a review,” Microfluid
Nanofluidics, vol. 8, no. 4, pp. 423–443, 2010.
[4] H. P. Mansberg, A. M. SAUNDERS, and W. Groner, “The Hemalog D white cell differential
system,” Journal of Histochemistry & Cytochemistry, vol. 22, no. 7, pp. 711–724, 1974.
[5] T. Malla, K. K. Malla, B. Sathian, P. Chettri, S. Singh, and A. Ghimire, “Simple febrile
convulsion and iron deficiency anemia a co-relation in Nepalese children,” Am J Public
Health Res, vol. 3, no. 5A, pp. p11-16, 2015.
[6] B. Lamb et al., “The Principles of Vacuum And Clinical Application in the Hospital
Environment (ISO)”.
[7] W. H. Organization, Technical Specifications for Oxygen Concentrators: WHO Medical
Device Technical Series. World Health Organization, 2016.
[8] P. Bělohradský, P. Skryja, and I. Hudák, “Experimental study on the influence of oxygen
content in the combustion air on the combustion characteristics,” Energy, vol. 75, pp. 116–
126, 2014.
[9] M. Pan, “Experimental and Simulation Studies of an oxygen concentration system using
pressure/vacuum swing adsorption technique: system miniaturization and prototype design,”
2017.
[10] J. Vetrovec, “Chemical oxygen-iodine laser with cryosorption vacuum pump,” in Gas,
Chemical, and Electrical Lasers and Intense Beam Control and Applications, 2000, vol.
3931, pp. 60–70.
[11] E. Schleicher, “The clinical chemistry laboratory: current status, problems and diagnostic
prospects,” Anal Bioanal Chem, vol. 384, no. 1, pp. 124–131, 2006.
[12] K. Obata, H. Tajima, M. Yohda, and T. Matsunaga, “Recent developments in laboratory
automation using magnetic particles for genome analysis,” Pharmacogenomics, vol. 3, no. 5,
pp. 697–708, 2002.
[13] S. Haeberle and R. Zengerle, “Microfluidic platforms for lab-on-a-chip applications,” Lab
Chip, vol. 7, no. 9, pp. 1094–1110, 2007.
[14] J. J. Verweij and C. R. Stensvold, “Molecular testing for clinical diagnosis and
epidemiological investigations of intestinal parasitic infections,” Clin Microbiol Rev, vol. 27,
no. 2, pp. 371–418, 2014.
[15] F. Jentzsch, “I.—The Binocular Microscope,” Journal of the Royal Microscopical Society,
vol. 34, no. 1, pp. 1–16, 1914.
[16] C. Beck, “Ii.—The Binocular Microscope Of The Past, And A New Form Of Instrument.,”
Journal of the Royal Microscopical Society, vol. 34, no. 1, pp. 17–23, 1914.
34
[17] J. Bellare, H. T. Davis, L. E. Scriven, and Y. Talmon, “Controlled environment vitrification
system: an improved sample preparation technique,” J Electron Microsc Tech, vol. 10, no. 1,
pp. 87–111, 1988.
[18] D. N. Rampley, “The Scanning Electron Microscope and its Application,” Measurement and
Control, vol. 8, no. 8, pp. 324–333, 1975.
[19] M. Ward and J. A. Langton, “Blood pressure measurement,” Continuing Education in
Anaesthesia, Critical care and Pain, vol. 7, no. 4, pp. 122–126, 2007.
[20] Y. Ostchega, Blood pressure randomized methodology study comparing automatic
oscillometric and mercury sphygmomanometer devices: National Health and Nutrition
Examination Survey, 2009-2010, no. 59. US Department of Health and Human Services,
Centers for Disease Control and …, 2012.
[21] M. F. Story, E. Schwier, and J. I. Kailes, “Perspectives of patients with disabilities on the
accessibility of medical equipment: Examination tables, imaging equipment, medical chairs,
and weight scales,” Disabil Health J, vol. 2, no. 4, pp. 169–179, 2009.
[22] D. McKay, J. Lipa, and P. Binhammer, “Principles of Microsurgery: Technique, Physiology
& Pharmacology,” 2007.
[23] P. Wobrauschek, “Total reflection x‐ray fluorescence analysis—a review,” X‐Ray
Spectrometry: An International Journal, vol. 36, no. 5, pp. 289–300, 2007.
[24] J. W. Gao and C. Y. Wang, “Modeling the solidification of functionally graded materials by
centrifugal casting,” Materials Science and Engineering: A, vol. 292, no. 2, pp. 207–215,
2000.
APPENDIX
35
Figure 14: assembled operation bed
Figure 15: WBC differential bath before service(left) and after service (right)
36
Figure 16: hydraulic diagram of the Dymind DH 56
37