SGPL CMP 001
SGPL CMP 001
SGPL-CMP-001
SILICA ANALYZER CALIBRATION
Revision No. 00 Revision Date : 06-07-2017 Issue Date : 06-07-2017 No. of Pages: 1 of 5
SGPL-CMP-001
Prepared By Reviewed By Approved By
Name: M.Rajesh Name: V. Venkata Krishna Sign:
Dt: 29-06-2017 Dt: 01-07-2017 Dt: 04-07-2017
Designation: (Engineer) Designation: Manager-C&I Designation: General Manager
Page 1 of 5
05
SGPL-CMP-001
SILICA ANALYZER CALIBRATION
Revision No. 00 Revision Date : 06-07-2017 Issue Date : 06-07-2017 No. of Pages: 2 of 5
TABLE OF CONTENTS:
1.0 PURPOSE 3
2.0 SCOPE 3
3.0 RESPONSIBILITY 3
4.0 TOOLS REQUIREMENT 3
5.0 SAFETY &ENVIRONMENTAL MEASURES 4
6.0 PROCEDURE 4
7.0 REFERENCE DOCUMENTS 5
8.0 RECORD FORMATS 5
Page 2 of 5
05
SGPL-CMP-001
SILICA ANALYZER CALIBRATION
Revision No. 00 Revision Date : 06-07-2017 Issue Date : 06-07-2017 No. of Pages: 3 of 5
1.0 Purpose:
This Procedure is intended to provide guidance and instructions for Carrying out the
Calibration Of Silica Analyzer.
2.0 Scope:
This procedure is applicable to Silica Analyzer 9610Sc.
3.0 Responsibility:
1. All the C&I personnel involved in carrying out /supervising the above activities are
required to ensure implementation of this procedure and may consult any other relevant
documents indicated under Reference section below.
2. In the event of any changes required in the procedure such changes may be requested
with HOD
Page 3 of 5
05
SGPL-CMP-001
SILICA ANALYZER CALIBRATION
Revision No. 00 Revision Date : 06-07-2017 Issue Date : 06-07-2017 No. of Pages: 4 of 5
6.0 Procedure:
2. Calibration: The automatic calibration uses the installed known standards to calibrate
the analyzer. Calibration can be done manually or scheduled to occur automatically. Use the
calibration menu to view calibration data, start or cancel the auto calibration, override the
auto cal settings, calibrate manually or to reset to the default calibration.
3. Push cal to display the calibration menu. Option Description START MANUAL CAL Refer to
start a calibration manually.
4. SET AUTO CALIBRATION Refer to Schedule automatic calibration.
5. CALIBRATION DATA Displays data from the last calibration and the due date and time of the
next scheduled calibration.
6. OVERRIDE CALIBRATION Enter new slope and zero values and disables automatic calibration.
7. OUTPUT CALIBRATION Select a 4-20 mA output and enter the output values to be sent.
8. RESET DEFAULT CAL Reset calibration data to default values and disables automatic
calibration. On completion, perform a new calibration. Schedule automatic calibration.
1) Push cal, then select SET AUTO CALIBRATION.
2) Select ACTIVATE AUTO CAL>YES.
3) Select STD SOLUTION and enter the standard value in ppb.
4) Select a schedule option for calibration.
Page 4 of 5
05
SGPL-CMP-001
SILICA ANALYZER CALIBRATION
Revision No. 00 Revision Date : 06-07-2017 Issue Date : 06-07-2017 No. of Pages: 5 of 5
7.0 Reference:
Hach 9610Sc Silica Manual. (SGPL-CMP-001-F01)
HIRA: SGPL/MAINT/C&I/BTG-SWAS-006
Page 5 of 5