RELATIVE HUMIDITY AND AIR TEMPERATURE

Psychrometer sensor
Wet&Dry forced ventilation temperatures sensor.
Design made according to ISO7726 standard
Direct calculation of the Relative Humidity and Dew
Point using LSI-LASTEM’s data logger according to
ISO7726 standard
2% accuracy assured over the whole range (0-100%)
and wide operative temperature (0...60°C)
High accuracy Pt100 1/3 DIN B sensitive elements
Forced ventilation on the sensitive elements: >2 m/s
Long term stability and robust principle. Applications
in polluted/salty air even near the saturation with high
temperatures

Internal ISO17025 accreditated calibration laboratory

Psychrometer sensor with wet&dry bulbs and forced ventilation. Wet bulb is closed inside a wet cotton sock from a distilled
water tank. This sensor has a very good stability because it is based on a direct measurement of temperature using 1/3 DIN
element and it is not sensitive to the time drift of commonly used capacitive hygrometer elements. LSI-LASTEM data loggers
can calculate Relative Humidity starting from wet&dry temperatures.

Technical Specifications

PN ESU102
Temperature Principle Wet&Dry bulbs with forced ventilation Pt100 1/3 DIN B
Measuring range -5...60°C
Accuracy 0.10°C (@ 0°C)
Output 2xPt100 DIN-IEC 751 table (EN 60751)
Resolution 0.01°C (M/R/E-Log)
Response time (T90 Air) 1.5 min
Relative Humidity % Principle Psychrometric
Measuring range 0...100%

Accuracy 2% (5...45°C)
Resolution 1%
Response time (T90 Air) 1.5 min
General Information Protection grade IP54
Ventilation on the sensitive > 2 m/s
elements
Power consumption 30 mA (forced ventilation on)
Operative temperature 0...60°C
Cable L=1 m
Connector Double min-din
Input type on E/M-Log N. 2 analog
E/M-Log derived quantities • Pyshrometric Relative humidity
obtained
• Dew point
Mounting On BVA305-315 stands or on table

MW9001-ENG-07-13/02/2023
1
Psychrometer sensor

Accessories

SVICA1102 Calibration certificate. ISO9001 type (Relative Humidity)

SVACA1105 Calibration certificate. ISO17025-ACCREDIA type (Relative Humidity)

MM3101 Spare cotton sock

There are several reasons to select psychrometer (wet&dry bulbs) or capacitive sensors to measure Relative Humidity. Here
above the list of advantage and disadvantage for both technologies:

Capacitive Wet&dry bulb

Water consumption No water Water must always be inside the water tank

Measurement below 0°C Can be used below 0°C Cannot be used below 0°C (because of the
water presence in the tank)
Measurement at high temperature Cannot discriminate when RH is above Can discriminate even above RH 95%
and RH values 95%
Temperature accuracy and re- Good temperature accuracy (0.1°C) Very good temperature accuracy (0.1°C). The
sponse time forced ventilation across the element facili-
tate the convective exchanges reducing the
response time
Measurements in severe conditions More fragile. When corrupted, sensor Even in severe conditions the wet&dry bulb
(polluted, salty, corrosive air) needs to be fully replaced is very rugged. Simple maintenance is re-
quired when the cotton sock becomes dirty
Sensor life time In fix applications, life time is about 2- In fix applications the sensor life time is only
3 years depending by the mechanical conditions.
(stand alone applications)
Sensitive elements are very rugged

Power comsumption Low power consumption High power consumption because of pres-
ence of the forced ventilation by the fan
Application Can be used indoor and outdoor Can be used indoor. Outdoor, only in above
(using a radiant screen) freezing temperatures. In outdoor, it needs
to be closed in a Stevenson screen

M-Log (ELO009) reads wet&dry bulb temperatures and produces Rela-


tive Humidity and Dew Point values according to the ISO7726 standard
formulas.

LSI LASTEM Srl Tel. +39 02 954141


Via Ex SP. 161 Dosso, 9 Fax +39 02 95770594
20049 Settala (MI) Email [email protected]
Italy www.lsi-lastem.com

2
2
RELATIVE HUMIDITY AND AIR TEMPERATURE

Thermohygrometer

Small dimensions and outstanding performances allow this


sensor to be used in a wide range of environmental applica-
tions

Measurement of both air temperature and Relative Humidity

Fast return after saturation

Standard dust filter to ensure good protection against


pollutants

Air temperature and RH% sensor with Pt100 output for temperature and 0...1 Vdc output for RH%.
It is a precise and reliable sensor, suitable for a continuous measurement even in severe environments, in presence of pol-
lutants, deep thermal and hygrometric ranges in long-term monitoring.

Technical Specifications

PN DMA672.1 ESU403.1 ESU403.2

Output RH%: 0...1 Vdc


°C: Pt100 DIN-IEC 751 table

Power supply 5…24 Vdc / 5…16 Vac

Connector L= 3 m free wires (8 wires) L= 1 m + N.2 Mini-Din L= 10 m + N.2 Mini-Din


connectors connectors

Data logger M-Log (ELO008)


compatibility M-Log (ELO009)
E-Log, A-Log

Common Technical Specifications

Temperature Principle RTD Pt100 1/3 DIN B (Class AA EN60751)

Measuring range -50...100°C

Accuracy 0.1°C (@ 0°C)

Output Pt100 DIN-IEC 751 table (EN 60751)

Resolution 0.01°C

Response time (T90) 4 s (1 m/s air flow)

Long term stability <0.1°C/year

3
Thermohygrometer (direct output)

Relative humidity Principle Capacitive

Measuring range 0...100%

Accuracy ±1% RH (5-95%)

Long term stability <±1%/year

Response time (T90) 10 s (1 m/s air flow)

Hysteresis <1%

Resolution 0,1%

General Information Protection grade IP66

Operative temperature -50...100°C

Accessories
SVICA0003 Calibration certificate ISO9001 (Temperature)

SVACA0006 Calibration certificate ISO17025-ACCREDIA (Temperature)

SVICA1003 Calibration certificate ISO9001 (Relative Humidity)

SVACA1005.1 Calibration certificate ISO17025-ACCREDIA (Relative Humidity)

BVA305 Arm for fixing sensors on BVA304 tripod

BVA315 Arm for fixing sensors on BVA304 tripod

ESU403.1 thermo-hygrometric sensor is


connected to M-Log data logger/reader sys-
tem (ELO009) which reads Temperature and
Relative Humidity and produces Dew Point
values according to the ISO7726 standard
formula.

This system can be used as portable hand-


held solution or for middle term portable
monitoring solutions, in this case ESU403.1
can be fixed on the tripod by means of the
BVA305 stand.

Monitoring solution, tripod Hand-held solution


mounting using BVA305 stand

LSI LASTEM Srl Tel. +39 02 954141


Via Ex SP. 161 Dosso, 9 Fax +39 02 95770594
20049 Settala (MI) Email [email protected]
Italy www.lsi-lastem.com

4
4
RELATIVE HUMIDITY AND AIR TEMPERATURE

Thermohygrometer
(analogue, digital and radio outputs )

4...20 mA, RS485-Modbus RTU and Radio outputs


version

Easy installation of the sensitive part, even in small


spaces or pipes

Cable lenght from 5 to 100 m

Dew Point calculation and output (replace RH%


output)

Simple replacement of the sensitive element

Outstanding accuracy (1.5%) for RH%

Sensor for measuring the air temperature and relative humidity in indoor environments or inside pipes.
In applications with dirty air, it is possible to mount the sensor inside the DYA225 shelter in order to prevent dirt from
accumulating directly on the sensitive element.

Technical Specifications

PN EXP815.1 DMA975.1 DMA875.1

Measurements °C/RH% °C/RH% °C/RH%

Output Radio RS485 2x0-4...20 mA

Frequency 868 MHz - -

Radio transmission 25 ± 3 mW - -
power

Radio transmission 600 m - -


distance (line-of-sight)

Transmission rate 10’ - -

Battery life >2 years - -

Protocol - Modbus RTU®, -


TTY-ASCII

Configuration - Hyperterminal -

RS485 protection - Galvanic insulation -


(3 kV, UL1577)

RS485 speed - 1200...115 kbps -

5
Thermohygrometer
(analogue, digital and radio outputs )

PN EXP815.1 DMA975.1 DMA875.1

Power supply Battery (AA 3,6 V) 10...30 Vac/dc 10...30 Vac/dc

Power consumption <10 µW stand-by 120 mW in 1W 1W


trasmissione

Electric protections NO (electrically insulated Tranzorb and Emifilter Tranzorb and Emifilter
system)

Common Technical Specifications

Temperature Principle RTD Pt100 1/3 DIN B (Class AA EN60751)


Measuring range Programmable: -40...60°C, -50...60°C, -50...70°C, -30...100°C
Accuracy 0.1°C (@0°C)
Resolution 0.01°C
Response time (T90) Typical 4 s (1 m/sec air flow)
Long term stability <0.1°C/year
Relative Humidity Principle Capacitive

Measuring range 0...100%

Accuracy ±1% (@5...95%)

Output Programmable: RH% or Dew Point

Long term stability <±1% / year

Response time (T90) Typical 10 s (1m/s air flow)

Hysteresis <1%
Resolution 0.1%

General Information Protection rate IP65


Operative temperature -50...100°C

6
6
Thermohygrometer
(analogue, digital and radio outputs )

Accessories

SVICA0003 ISO9001 type calibration certificate (Temperature)

SVACA0006 ISO17025-ACCREDIA type calibration certificate (Temperature)

SVICA1003 ISO9001 type calibration certificate (Relartive Humidity)

SVACA1005.1 ISO17025-ACCREDIA type calibration certificate (Relative Humidity)

DWA505A Cable L= 5 m

DWA510A Cable L= 10 m

DWA525A Cable L= 25 m

DWA526A Cable L= 50 m

DWA527A Cable L= 100 m

MG2251.R 7 pin free female connector

ML3015 Sensitive element (spare part) for DMA815.1-875.1. EXP815.1

EXP301 Radio signal receiver from EXP815.1 radio sensor.


Output compatible with data loggers (M/E-Log)
Maximum number of receivable sensors: 200
Battery: NiCd 9 V
Power supply: 12 Vdc
Connection cable to data logger: DWA601

DYA225 Open well for T+RH% sensors when mounted in ducts in presence
of dirty air flow.
Material: Iron
Dimensions: L= 200 mm Ø 92 mm
Support plate: 120x120 mm

LSI LASTEM Srl Tel. +39 02 954141


Via Ex SP. 161 Dosso, 9 Fax +39 02 95770594
20049 Settala (MI) Email [email protected]
Italy www.lsi-lastem.com