Yhtt6 Scale Calibration Document

The YH-T6 technology provides high resolution weighing through a 1,000,000 count A/D converter sampling at 10 times per second. It can connect to a single 350Ω load cell using a 3V excitation voltage and operates from 85-265V power at 50-60Hz with an internal battery. Calibration is performed by connecting the load cell, powering on, and following a 7 step process to set divisions, decimal places, full capacity, zero point, and load calibration values.

Uploaded by

nieyaazb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
749 views1 page

Yhtt6 Scale Calibration Document

The YH-T6 technology provides high resolution weighing through a 1,000,000 count A/D converter sampling at 10 times per second. It can connect to a single 350Ω load cell using a 3V excitation voltage and operates from 85-265V power at 50-60Hz with an internal battery. Calibration is performed by connecting the load cell, powering on, and following a 7 step process to set divisions, decimal places, full capacity, zero point, and load calibration values.

Uploaded by

nieyaazb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

YH-T6 Technology Introduction

1. Main parameters:
1. A/D conversion:∑-Δ,resolution is 1,000,000,10 times/second。Signal input range:-19mV ~ +19mV

2. Load cell excitation voltage:DC +3V,capable of connecting 1 load cell of 350Ω, 4-wire.

3. Power supply:AC 85V~265V;50Hz~60Hz, with built-in rechargeable battery 4V/4 AH.

4. Working temperature:0~40℃;humidity≤85%RH;storage temperature:-25℃~55℃.

2. Connection between load cell & weighing indicator:


-E Shield End of Indicator End of load cell

Stimulating supply Power+


① ② ③ ④ ⑤ ○
6 +E _+++=+++Supply +
⑥ ⑦ ⑧ ⑨ ○- E
1 Stimulating supply Power -

9 +IN _______ Supply
Signal Output +
Signal Output -
+

+E -IN +IN ○
8 +IN

5 Shield
3.Calibration & inspection
Correctly connecting weighing indicator with load cell, power on the indicator, it begins self-inspection. When finishes, it enters
weighing mode (first power-on, it’s better to pre-heat weighing indicator and load cell for 15~30 minutes, then begin to calibrate).
Turn the calibration head up, then it enters calibration mode, as per the steps listed below you can finish the calibration:

Step Operation Display Marks

1 Press & together [d 001] It enters calibration mode

Select the division by pressing Setup of division:press the key again and again to choose the division among
2 the key , then press press 1、2、5、10、20、50、100, 200, If confirmed, press then
[d 2]
to confirm
It enters to next step. I.e.: 2
Press to choose Setup of decimal point:Press again and again to choose the bit of decimal
the decimal point, then press point among 0、1、2、3、4. I.e.: 2
3 to confirm
[d P 2]
(Note:If add the bit of decimal point 5,it can stop the indicator entering the saving power

mode)
Set full capacity by Setup of full capacity:
pressing , Pressing to make flash bit move to right side, and pressing
4 [ F030.00]
To make flash bit move to left side. The number on flash bit can be added by pressing ,
Press to confirm then input full capacity in this way, later press to enter to next step. I.e.: 3000

After it gets stable, press Setup of zero point:Make sure there is no load on the scale。When the stable light
5 to confirm [ noLoAd ] is on,press to confirm the zero point,the indicator displays“ ---- ”about two seconds, then

entering load calibration.

Press Setup of linearity:


To input the real weight of the Add a certain load(better to close to its full capacity).
[AdLoAd ]
6 loads,after it gets stable To input the real weight of the loads(I.e:
3000),
the input way is same as step 4. when the stable
[ A030.00]
press to confirm light is on, press to confirm, the indicator displays “ ---- ”,entering next step.

Calibration ends, it returns to The calibration ends.


7 [ ****** ]
weighing mode Note:In the calibration, press can interrupt it,then back to the weighing mode.

4.Review inner code: Under weighing mode,press the two keys & together ,can enter or exist inner mode。(1d=20 inner codes)