Uesr manual
Contents
Safety................................................................3
Introduction………………………………..………9
Specifications………………..……..……..………24
2
A100D Fetal Doppler
Section 1
Safety
1.1 Instructions for the Safe Operation and Use of the
A100D Doppler
Examine the doppler and any accessories
periodically to ensure that the cables, probes, and
instruments do not have visible evidence of
damage that may affect patient safety or
performance. The recommended inspection
interval is once per week or less. Do not use the
doppler if there is any visible sign of damage.
Do not attempt to service the A100D doppler. Only
qualified service personnel should attempt any
needed internal servicing.
The A100D is not specified or intended for
operation in the presence of electrosurgical
equipment.
The A100D is not specified or intended for
operation in conjunction with any other type of
monitoring equipment except the specific devices
that have been identified for use in this uesr
manual.
Perform periodic safety testing to insure proper
patient safety. This should include leakage current
measurement and function testing. The
3
A100D Fetal Doppler
4
A100D Fetal Doppler
1.3 Cautions
5
A100D Fetal Doppler
6
A100D Fetal Doppler
Type B Equipment
Batch code
Date of manufacture
Information of manufacture,
including name and address
Information of authorized
representative in the
European community
7
A100D Fetal Doppler
8
A100D Fetal Doppler
Section 2
Introduction
2.1 General
This chapter provides a general description of the A100D
fatal doppler including:
Brief device description
Product features
Options and accessories
2.2 Brief Device Description
A100D Ultrasonic fetal doppler is a high performance fetal
heart beat detector with dual color TFT display, which can
meet routine examination requirements of obstetricians. It is
applicable for midwives, private obstetrician's office,
obstetrical department of hospital and remote monitoring
situations.
It has audio output, and can be connected with
earphone or recorder with audio input. A100D use
rechargeable battery and charger as standard accessories,
and support standard 3.7 V Li-ion as optional accessories.
2.3 Product Features
Lightweight and Easy-To-Use.
0.96 Inch tft Color LCD screen.
Detection and Indication of probe off.
Change to power saving mode automatically
9
A100D Fetal Doppler
Continuous
Portable FHR
wave Doppler
probe
probe
Rechargeable
battery
(600mAH GP,21-64015
3.7V Li-ion)
10
A100D Fetal Doppler
Input:100-240V
AC
Output:12VDC
USB port
500mA
charger
Connecting
method is as
shown on the
right figure
Optional Accessories:
Ultrasound
Ultrasound transmission gel
Gel (optional
accessory)
11
A100D Fetal Doppler
Section 3
Installation, Setup, and Operation
3.1. Description of the Front Panel &Rear Panel (as
figure 3.1.1)
12
A100D Fetal Doppler
3.2 Display
This device is equipped with the TFT. There are two types
display mode for TFT screen which can be changed if short
13
A100D Fetal Doppler
15
A100D Fetal Doppler
3.5 Operation
3.5.1 Turn the Doppler on
Press button for 2 seconds to turn the doppler on
after connecting FHR probe. The doppler is on FHR
mode after self-testing.
3.5.2 Monitoring Fetal Heart Rate
Step 1: Preparing works
Check the portable ultrasound probe to verify proper
attachment to the doppler. Adjust heart rate channel
one speaker volume to mid level. Apply ultrasound gel
to the face of the probe
17
A100D Fetal Doppler
Figure 3.5.6
18
A100D Fetal Doppler
Section 4
Cleaning and Disinfection
4.1 Cleaning
Switch off the power and take out the batteries before
cleaning,
Keep the exterior surface of the device clean and free of
dust and dirt. Cleaning exterior surface (TFT display screen
included) of the unit with a dry and soft cloth. If necessary,
clean it with a soft cloth soaked in a solution of soap, or
water and wipe dry with a clean cloth immediately.
Wipe the probe with soft cloth to remove any remaining
coupling gel and clean with soap and water only.
CAUTION: Don’t use strong solvent. For example,
acetone.
4.2 Disinfecting
After Cleaning the equipment exterior surface and probe,
Immerse the probe into the solutions of benzalkonium
Bromide, 0.5% Chlorhexidine, 2% Glutaraldehyde or 75%
ethanol to disinfect the water proof probe. Wipe the probe
with a clean and dry cloth to remove any remaining
moisture.
Note: Please pay attention to the height when the probe is
immersed in order to prevent the sterilant from entering the
probe socket.
Section 5
20
A100D Fetal Doppler
22
A100D Fetal Doppler
Section 6
Specifications
Power
Internal: 3.7V DC alkaline battery support 300
minutes working time.
Rechargeable battery (cycle life>500 times)
4 hours to full recharge at 160mA charge
current
Power
3 watts, maximum
Dissipation:
Environmental
Operating 5°C to 40°C
Temperature:
Storage –10°C to 50°C
Temperature:
Relative Humidity: 20% to 85% non-condensing
25
A100D Fetal Doppler
face:
Ultrasonic frequency: 2.0 MHz ±10%
BPM Range: 60-210 BPM
Accuracy: ±2 BPM
Others
Potable FHR probe Detect the FHR and make fetal
heart sound in early pregnancy.
Speaker 0.5W speaker, support volume
adjusted
Probe line length Less than 1.5m
IEC 61000-3-3
± 2 kV for
Electrostatic
power Mains power quality
transient /
supply lines should be that of a
burst
N/A typical commercial or
± 1 kV for hospital
IEC
input/output environment.
61000-4-4
lines
± 1 kV
differential Mains power quality
Surge
mode should be that of a
N/A typical commercial or
IEC
± 2 kV hospital
61000-4-5
common environment.
mode
Voltage dips, 0 % UT; 0,5 Mains power quality
N/A
short cycle g) should be that of a
28
A100D Fetal Doppler
and d is the
recommended
separation distance in
metres (m).b
31
A100D Fetal Doppler
a Field strengths from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in
which the A100D fetal doppler is used exceeds the
applicable RF compliance level above, the A100D fetal doppler
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be
necessary, such as reorienting or relocating the A100D fetal
doppler .
b Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than 3V/m.
33
A100D Fetal Doppler
Product Guarantee
Product Name Portable Fetal Doppler
Model Name A100D
Warranty Period 1 Years (Probe excluded)
Date of Purchase
Hospital:
Address:
Customer
Name:
Telephone:
Sales Agency
Manufacture
Company logo:
34
A100D Fetal Doppler
35