GA-180 Manual 032519
GA-180 Manual 032519
GA-180 Manual 032519
The information contained in this manual was current at the time of printing. The most current versions of all
Hydro Instruments manuals can be found on our website: www.hydroinstruments.com
I. Installation
A. Monitor Installation ..................................................................................3
B. Sensor Installation.....................................................................................3
C. Battery Backup Installation.......................................................................4
II. Operation
A. Navigating the Controller..........................................................................9
B. Operating Screens and Settings ................................................................9
C. Configuring a Sensor Channel ................................................................10
D. Installing and Configuring Temperature .................................................16
E. Alarms and Outputs ................................................................................16
F. Modbus Communication .........................................................................21
G. Response Checks (Bump Testing) ..........................................................21
H. Data Logger ............................................................................................22
III. Troubleshooting
A. Installation Check ...................................................................................23
B. Symptoms, Likely Causes, and Suggested Responses............................23
C. Explanation of Responses .......................................................................24
Figures
1a. Sensor Installation (heavy gases) ............................................................3
1b. Sensor Installation (light gases) ..............................................................3
2. Remove Calibration Cap .........................................................................4
3. Wiring Diagram (inside of door) .............................................................5
4. GA-180 Controller Electronics ...............................................................6
5. Pinout Diagrams for MB108 and MB123 ...............................................7
6. Pinout Diagrams for MB122, MB141, and MB101 ................................8
7. Calibration Cap .....................................................................................12
8. Sensor and Calibration Kit ....................................................................12
9. GA-180 Operating Screens ...................................................................13
10. GA-180 Configuration Screens .............................................................14
11. GA-180 Modbus Configuration and Node Enable Screens...................15
12. External Alarm Light and Horn.............................................................20
13. Bump Testing (Chlorine Gas Example) ................................................21
Tables
1. Example Controller Configurations ........................................................6
2. Standard Ranges For Hydro Instruments Gas Sensors..........................11
3. Relay Options for Common Relays (33 & 34)......................................17
4. Relay Options for Sensor Relays (1-32) ...............................................17
5. Circuit Board Descriptions and Node Numbers ....................................18
2
I. INSTALLATION AND OPERATION
A. Monitor Installation
All monitors are able to accept either 120 VAC or 240 VAC single phase power at 50-60 Hz. When
connecting A/C power it is imperative that the power source is well grounded. Improper grounding
will disrupt proper operation of the unit.
Warning: Ensure source power is disconnected from main power, prior to making the instrument
connection.
Monitors are NEMA 4X rated and should be installed at/near eye level and protected from exposure
to direct sunlight and rain. The monitor should be installed near, but outside of the chemical storage
room it is monitoring (see Figure 1a and 1b). Four 5⁄16" DIA through holes are provided on the
enclosure for mounting the monitor against the wall.
FIGURE 1a FIGURE 1b
For gases heavier than air, sensor should be For gases lighter than air, sensor should be
mounted low. mounted high.
12" - 24"
GAS SENSOR
GAS SENSOR
12" - 24"
B. Sensor Installation
Refer to Figures 1a and 1b for more information.
1. Determine a suitable mounting location. For monitoring gases heavier than air, the sensor should
be placed approximately 12-24 inches from the floor. For gases lighter than air, the sensor
should be placed 12-24 inches from the ceiling. Do not place sensor in any location where it
could become wet. This will damage the sensor.
2. Attach the enclosure to the wall using two 1⁄4"-20 mounting screws (recommended).
3. Remove sensor calibration cap. Store this cap in a known location as it is necessary for
calibration when using a span gas. See Figure 2.
3
4. If necessary, open the sensor enclosure front cover and reconnect the signal wires. Connect
Red to Red and Black to Black on the terminal strip inside the sensor enclosure. Replace
enclosure front cover and be sure to check that the gasket is evenly sealed around the
enclosure. Also tighten the liquid tight fitting.
5. If necessary, connect the signal wire to the appropriate terminals. Refer to Table 1 and Figure 4
so that you can connect the sensor to the right channel. The black wire will connect to AI1 or
AI2 and the red wire will connect to V+ depending on the required channel number.
FIGURE 2
Sensor Enclosure
Front
Bottom
Calibration
Cap
Note: Battery 1 will need to be enabled to back up sensors 1-4, Battery 2 will need to be enabled
to back up sensors 5-8, Battery 3 will need to be enabled to back up sensors 9-12, and Battery 4
will need to be enabled to backup sensors 13-16. For example, if a system is ordered with 8
sensors then two batteries will need to be installed (into their respective enclosures). Battery 1
and Battery 2 will need to be enabled. If a unit was purchased at the factory to have battery
backup this will have already been done.
4
0LFUR6'+& 0-FLUFXLWERDUG
FDUGVORW IRUGDWDORJJLQJ
IRUGDWD RSWLRQDO
ORJJHU
7R%DFN3ODWH*URXQG/XJ
)RU02'%86
&RPPXQLFDWLRQ
7R7HUPLQDO%ORFN
*1'JUHHQ
/,1(EODFN
FIGURE 3: Wiring Diagram (inside of door)
1(87ZKLWH
7R&LUFXLW%RDUG
9EODFN
9UHG
5
FIGURE 4: GA-180 Controller Electronics
MB101 (CB-BAT)
Battery Battery Board
(CAUTION:
REMOVE BATTERY MB108 (CB-8RELAY)
MB220 Eight Relay Board
BEFORE SHIPPING) AC Power
Display Board
Terminal Block
MB123 (CB-2RELAY)
Power Supply Board MB122 (CB-2X2MA) Two Relay Board
Two In, Two Out Board
Photo of a GA-180 unit configured for two sensors with a battery backup.
Some other configuration examples are summarized below.
TABLE 1: Example
ou TH ard
r )
In CB d )
oa AT d
B )
oa AY
o 2 ( oar LAY
m CB t B )
r
e M
Controller Configurations
u A
oa
s
o
M ela B-2 d
pl ER
or
Tw 12 y B EL
Th 14 o O 2M
rd )
B E
e?
ns
a R
y R
B B
el -8
pl
M , Tw -2
Se
oc -
-
t R CB
ou
ry B
E
R C
um
te (C
PL
oc
?
o 3(
er 1 (
ry
gh 8
at 1
im
m
M
Ei 10
B 10
Tw 2
te
1
A
er
ax
B
at
B
B
EX
Th
M
M
Date: 2018-07-11-v1
GA-180 CONTROLLER
ELECTRONICS Dwg. No. GA-180-CONTROLLER
6
FIGURE 5: Pinout Diagrams for MB108 and MB123
MB108 (CB-8RELAY)
EIGHT RELAY BOARD
Normally Closed 8
Common 8
Normally Open 8
} RELAY 8
Normally Closed 7
Common 7
Normally Open 7
} RELAY 7
Normally Closed 6
Common 6
Normally Open 6
} RELAY 6
Normally Closed 5
Common 5
Normally Open 5
} RELAY 5
Normally Closed 4
Common 4
Normally Open 4
} RELAY 4
Normally Closed 3
Common 3
Normally Open 3
} RELAY 3
Normally Closed 2
Common 2
Normally Open 2
} RELAY 2
Normally Closed 1
Common 1
Normally Open 1
} RELAY 1
V+
V–
MB123 (CB-2RELAY)
TWO RELAY BOARD
Normally Closed 2
Common 2
Normally Open 2
} RELAY 34
Normally Closed 1
Common 1
Normally Open 1
} RELAY 33
7
FIGURE 6: Pinout Diagrams for MB122, MB141, and MB101
MB122 (CB-2X2MA)
Analog Output
TWO IN, TWO OUT BOARD AO1
Analog Output
AO2
} SENSOR
4-20mA
OUTPUTS
to Horn
}
RED Analog Input
BLACK AI2 SENSOR
SHIELD
BLACK [connect to metal back plate] INPUTS
RED
Analog Input
AI1
SHIELD
[connect to metal back plate]
MB141 (CB-THERM)
THERMOCOUPLE BOARD
Normally Closed 1
Common 1
Normally Open 1
} HIGH
TEMPERATURE
ALARM
YELLOW
Thermocouple Input
RED
MB101 (CB-BAT)
BATTERY BOARD
RED
To Power Supply
BLACK
RED
To Battery
BLACK
8
II. OPERATION
A. Navigating the Controller
The GA-180 is provided with a Nema 4x, 2 line alphanumeric display controller that will display
and output all important features and conditions. Navigating the controller is done by the use of four
push button keys. The push button functions are described below.
key: Cycles to the previous screen.
key: Cycles to the next screen.
key: Increases/changes value, also used to enter screens.
key: Decreases/changes value.
Note: When adjusting parameter values, the number displayed is automatically saved upon leaving
the screen. Thus no “enter” button is needed.
Password: All editable features of the GA-180 are password protected to prevent the unwanted
tampering of the gas detector settings. To access these screens the correct password must be entered
to proceed. The password for the GA-180 is “180”.
3. Temperature: This screen provides a live reading of the room temperature. This screen will
only appear if the thermocouple channel has been enabled.
4. Test Operation: This screen allows the user to test the alarm horn and all enabled relays on the
GA-180. Pressing the key will activate the horn. Pressing the key will trip the relays.
5. Channel Alarm Status: These screens will notify the user of what alarm state a particular
sensor channel is in, if any. Typical statuses would be “Normal”, “Danger”, “Alarm”, or “Error”.
Refer to Sections II.E and III for alarm warning explanations and corrective actions.
9
6. Communication Status: Whenever a channel is activated, or a different circuit board is enabled
(relays, thermocouples, battery backups) the main display CPU will request information from
the relevant circuit boards to display and operate properly. The GA-180 is able to detect if
any boards on this system are not communicating properly and address the user for immediate
correction. Communication statuses will be “Normal” or “Error”. Refer to Sections II.F for
more details.
7. Temperature Alarm: This screen will indicate if the air temperature in the room has exceeded
the high temperature alarm setting or fallen below the low temperature alarm setting. Status will
indicate one of the following: Normal, Alarm High, Alarm Low or Fail. Refer to Section II.D for
more details.
8. Password: This screen allows access to the adjustable/configuration settings of the GA-180.
Using the and keys, the password may be entered. The password for this unit is “180”.
10
5. Zero Calibration: After installation, or throughout the sensor’s lifetime, the sensor signal may
periodically drift from 0.0 ppm even though no gas is present. To reset this press the key to
increase the value and press the key to decrease the value on this screen. The keys must be
pressed one at a time to move the value, pressing and holding will have no effect. Wait at least
10 seconds after a new zero has been entered to confirm signal stability. A bump test should be
performed after calibration.
Warning: The zero calibration should only be adjusted if the sensor is offset by 5% or less from
zero. Confirm that no gas is present prior to recalibration and always ensure adequate safety
procedures are enacted to confirm there is no gas present. If the value is offset by more than 5%
do not recalibrate and consult Hydro Instruments for more information.
6. Channel Gas Type and Units: This screen allows the user to enable the sensor channel by
changing the gas type from “OFF” to one of the following options “NH3”, “O2”, “O3”, “SO2”,
“Cl2”, “ClO2”, “CO”, “H2”, and H2S”. The gas type must match that which the sensor is designed
to detect.
7. Channel Full Scale: This screen allows the user to adjust the full scale of the monitor channel,
so that it will match the full scale of the sensor output connected to the channel. Should the two
outputs be misaligned, improper operation will result. Standard full scales for sensors supplied
by Hydro Instruments can be seen in Table 1. If unsure about the full scale output of a sensor,
please contact Hydro Instruments.
8. Channel Span Calibration: The gas detector system is factory calibrated and does not require
calibration upon installation setup. Span calibration is rarely required, however, it may be
required or desired to perform span calibrations periodically over the life of the sensor. If
calibration is to be carried out, then the appropriate span gas calibration kit must be purchased
(Figures 7 and 8).
8a. The calibration cap must be installed on the sensor and connected as indicated in Figures 7
and 8. Span gas must be allowed to flow at 500 cc/min for at least 1 or 2 minutes until the
displayed reading stabilizes. The reading on this screen should be adjusted to match the
ppm value of the span gas being used. Use the key to increase the reading of the key
to decrease the reading. Press and release the keys one at a time. Do not press and hold the
keys. After adjusting, wait 10 seconds to confirm that the reading is stable before proceeding
to the next step.
11
FIGURE 7: Calibration Cap FIGURE 8: Sensor and Calibration Kit
Ordering Information
Product Note: The sensor is shipped with the calibration
Number Description cap already installed. After calibration the
GA-CK-CL2-10 Calibration Kit – 10 PPM Chlorine Gas calibration cap should be removed for normal
with Regulator use. Do not dispose of the calibration cap as
GA-CRS-CL2-10 Replacement Sensor – 0-10 PPM it will have to be reinstalled and used for any
Chlorine further sensor calibrations.
9. Channel 4 mA Output Calibration: Each sensor channel has its own isolated 4-20 mA output
used for remote monitoring/indication of gas residual. Although each channel is calibrated using
NIST certified equipment, it may be necessary to adjust the output of from the controller to
match the input of the remote system. This screen allows the user to adjust the A/D count that
outputs the 4 mA signal. Pressing the key will increase the current output and pressing the
key will decrease the output.
10. Channel 20 mA output Calibration: Similar to screen #9 (4 mA Output Calibration), this
screen allows the user to make minor adjustments to the A/D value that controls the 20 mA
output. Pressing the key will increase the current output and pressing the key will decrease
the output.
11. Channel Filter Time: This filter time is used to eliminate unwanted signal disturbances from
the sensor. The filter time can be set anywhere from 0-60 seconds using the and keys.
The recommended factory default value is 5 seconds.
12. Calibration Diagnostic Screen: This screen allows the user to view the most recent calibration
points and their corresponding A/D values for diagnostic purposes.
13. Live diagnostic Screen: This screen allows the user to see the current sensor value, A/D value
and current output for a particular sensor.
12
FIGURE 9: GA-180 Operating Screens
* ...
* 15 Status: Error
16 Status: Error
Normal
Comm Status Node Error
Normal *
...
Enter Password
180
13
14
Screens shown with grey border
are hidden screens, Enter Password
accessed by holding - 180
at the “Alarm Delay” screen.
Setup: Sensor Temp Setup: Sensor Temp Setup: Sensor Temp Setup: Sensor Temp
Display DL Display DL Display DL Display DL
Configure Channel Ch1 Temp: Units = F F Data Log
Ch Ch2 Alm Delay = 5s C On
.
.
. Temp: Units = F Data Log Frequency
Alm Delay = 5s On 60 secs
Ch15
Off Set Time and Date?
Ch16 Temp: Alm Low = 40F
Alm High = 120F No Yes
Temp: Alm Low = 40F Set Time and Date?
Alm High = 120F No Yes
Change Time and Date
Cl2 Tue Dec 19 '17 11:13
ClO2
Alarm Delay CO Time and Date
5 Secs H2 Was Changed!
H2S
OFF
HOLD
NH3
Gas Type= Cl2 O2
Units= ppm O3
SO2
Gas Type= Cl2
Units= ppm ppm
%
FIGURE 10: GA-180 Configuration Screens
FIGURE 11: GA-180 Modbus Configuration and Node Enable Screens
Enter Password
** 180
Low and High Alarms for specific
channels can be accessed 2400
Setup: Sensor Temp
according to the table below.
Display DL 4800
In general, odd numbered relays HOLD
can access Low Alarms, and
9600
Modbus Baud=19200 19200
even numbered relays can Node= 1 Data=8/N/1
access High Alarms. 8/N/1 38400
Modbus Baud=19200 8/N/2 57600
Node= 1 Data=8/N/1 8/E/1 115200
Channel Relay Alarm Relay Alarm
1 1 Ch1 Low Alarm 2 Ch1 High Alarm
8/O/1 250000
2 3 Ch2 Low Alarm 4 Ch2 High Alarm
Modbus Baud=19200
Node= 1 Data=8/N/1 Any Low Alarm
3 5 Ch3 Low Alarm 6 Ch3 High Alarm
4 7 Ch4 Low Alarm 8 Ch4 High Alarm Any High Alarm
Relay 33 Sensor/DL/Temp Fail
5 9 Ch5 Low Alarm 10 Ch5 High Alarm
6 11 Ch6 Low Alarm 12 Ch6 High Alarm
Any Low Alarm A/C Power Fail
7 13 Ch7 Low Alarm 14 Ch7 High Alarm Relay 34 Battery Low
8 15 Ch8 Low Alarm 16 Ch8 High Alarm Temp Low Alarm
HOLD Any High Alarm
9 17 Ch9 Low Alarm 18 Ch9 High Alarm
Temp High Alarm
10 19 Ch10 Low Alarm 20 Ch10 High Alarm Relay 1
11 21 Ch11 Low Alarm 22 Ch11 High Alarm
Ch1 Low Alarm Ch1 <Low/High> Alarm **
12 23 Ch12 Low Alarm 24 Ch12 High Alarm
Off
13 25 Ch13 Low Alarm 26 Ch13 High Alarm Relay 2
14 27 Ch14 Low Alarm 28 Ch14 High Alarm Any Low Alarm
Ch1 High Alarm
15 29 Ch15 Low Alarm 30 Ch15 High Alarm Any High Alarm
16 31 Ch16 Low Alarm 32 Ch16 High Alarm Relay <odd #> ** Sensor/DL/Temp Fail
Ch Low Alarm A/C Power Fail
Battery Low
Relay <even #>
** Temp Low Alarm
Ch High Alarm
Temp High Alarm
Relay 31
Ch16 Low Alarm Ch <Low/High> Alarm **
Off
Relay 32 Any Low Alarm
Ch16 High Alarm Any High Alarm
Sensor/DL/Temp Fail
A/C Power Fail
Battery Low
Temp Low Alarm
Temp High Alarm
Relay 33-34 Enable Ch16 <Low/High> Alarm
Yes **
Off
Relay 1-8 Enable Any Low Alarm
Yes Any High Alarm
Sensor/DL/Temp Fail
Battery 1 Enable A/C Power Fail
Yes Battery Low
Relay 9-16 Enable Temp Low Alarm
No Temp High Alarm
Battery 2 Enable
No
Relay 17-24 Enable
No
Battery 3 Enable
No
Relay 25-32 Enable
No
Battery 4 Enable
No
15
D. Installing and Configuring Temperature
1. Determine adequate mounting location for thermocouple and support bracket. Mount the support
bracket to a wall or fixture using appropriate hardware (hardware not supplied with GA-180).
2. If necessary, connect the thermocouple to the thermocouple circuit board using the TI + and TI-
terminals. Connect the Red wire to TI- and the Yellow wire to TI+. Only type K thermocouples
can be used on the GA-180.
3. If necessary, connect the thermocouple board (MB141) to the display through the Modbus connector.
4. If necessary, enable the thermocouple communication. Follow the screen tree (Figure 11) to
navigate to the screen “thermocouple enable”. Change to yes by pressing the key and cycle
the power. Unless you turn the power off and back on, the change will not take effect.
5. Check communication by confirming that the appropriate live temperature screen and
temperature alarm screen have appeared in the correct area (Figure 9).
6. Using the screen tree (Figure 10), navigate to the “Temperature Units” screen. Units can be
adjusted between “F” for Fahrenheit and “C” for Celsius. Select the appropriate unit using the
and keys.
7. Press and select the high and low temperature alarm points. Values can be selected anywhere
from 0-255 F/C. Hydro Instruments recommends the high temperature alarm setting be at least
20ºF lower than the fusible plug melting point of gas cylinders and/or ton containers. Contact
your chemical supplier to confirm the temperature for the fusible plug melting point. Other
equipment or local restrictions may apply. Each value is adjusted using the and keys.
8. Press and select the alarm delay time. This setting sets the amount of time the temperature
must remain above the high temp. alarm setting or below the low temp. alarm setting before the
alarm will activate. Values can be set in the range of 0-60 seconds. Values are adjusted using the
and keys. Press to exit the thermocouple configuration.
16
TABLE 3: Relay Options for Common Relays (33 & 34)
Relay Options
Relay Options
17
TABLE 5: Circuit Board Descriptions and Node Numbers
Node Number
Circuit Board Board Description Application
(Comm Error)
1 MB122 Two In, Two Out Board Sensor 1, Sensor 2
18
2. Acknowledgement of Alarms: If an alarm condition occurs, the alarm (red) LED will
illuminate and the relay will be activated. To acknowledge an alarm (and deactivate the relay) for
all sensors in the “Alarm” condition press the key. Pressing the key once will inactivate
the horn, pressing the key twice will inactivate the relay. Alarms can also be acknowledged
remotely through the use of a SCADA system / PLC. To do this, connect a contact input to the
DI1 input terminal on the 2 relay board (MB123) and to a relevant ground. Connecting the input
once will inactivate the horn, connecting the input twice will inactivate the relay.
Note: Even after acknowledging the alarm, both LEDs will remain illuminated until the alarm
condition has been removed.
3. Alarm Explanation: Rising and Falling Alarms
a. Rising: If the Danger (low level alarm) is set to a lower value than the Alarm (high
level alarm), then the GA-180 will automatically configure the channel as a rising alarm.
Therefore, if the sensor reading is higher than the Danger or Alarm setting the GA-180 will
activate the appropriate alarm(s). To configure a sensor rising/falling alarm, follow the steps
outlined in Section II-C.
b. Falling: If the Danger (low level alarm) is set to a higher value than the Alarm (high
level alarm), then the GA-180 will automatically configure the channel as a falling alarm.
Therefore if the sensor reading is lower than the Danger or Alarm setting the GA-180 will
activate the appropriate alarm(s). To configure a sensor rising/falling alarm, follow the steps
outlined in Section II-C.
c. Failsafe Alarms: If a sensor channel is set to failsafe, then the corresponding Alarm (high
level alarm) relay will become normally energized. This will cause a reversal of the NC/
NO connections. Therefore, the normally closed connection will be open unless an alarm
condition is present or power is lost. To configure a sensor failsafe alarm, follow the steps
outlined in Section II-C.
4. 4-20 mA Outputs: The GA-180 also includes isolated 4-20 mA outputs for each channel
ordered. This output signal will represent the current gas residual value. See Figure 6 for
4-20mA output wiring locations.
19
FIGURE 12
Features
Single compact unit
Wall mounting
Rotating strobe light with red lens
90 dB Audible horn
Weather resistant
Available Models
GA-AL-110 (110VAC)
GA-AL-220 (220VAC)
The external alarm light & horn is an optional accessory for use with all Hydro
Instruments gas leak detection equipment. The alarm light connects to a relay
inside the gas detector monitor. This can be a sensor specific relay or a common
relay.
Alarm Light Wiring—Normally Open Relay Circuit
Relay
NO
CO
NC
L L = Line
N = Neutral
PWR
N Light NO = Normally Open
CO = Common
NC = Normally Closed
N L
20
F. Modbus Communication:
Modbus RS-485 communication. The GA-180 is equipped for remote display and communication
using the modbus RS-485 standard. To do this, you must define the node, baud rate and parity
of the system. For more information on how to setup modbus refer to the Modbus Installation
and Instruction manual. The Modbus Installation manual can be downloaded from the Hydro
Instruments website (www.hydroinstruments.com). Printed copies are available upon request.
G. Bump Testing
To verify responsiveness, the gas sensors can be bump tested (exposed to a small amount of the
target gas) in order to test the reaction of the sensor. A plastic squeeze bottle is provided with each
gas detector for this purpose (Figure 13). It is suggested that bump testing be done at quarterly
intervals, however required frequency is determined by environment, conditions, number of and
severity of leaks. Proper bump testing will not substantially degrade the sensor or shorten sensor
life. Figure 13 diagrams a bump testing procedure for chlorine gas. Contact Hydro Instruments for
other bump testing procedures should the sensor be for something other than chlorine gas.
GAS SENSOR
1 (888) 38-HYDRO
www.hydroinstruments.com
GAS SENSOR
Gas
Fumes
Liquid
21
H. Data Logger
1. Description: If installed and enabled, the data logger records each gas sensors reading and
temperature sensor reading. The frequency of recording is adjustable. The data is recorded on
a MicroSDHC memory card located inside the gas alarm enclosure on the MB220 board. The
MicroSDHC memory card is installed in the slot (See Figure 5). Data can be read using a text-
reader program or imported into a spreadsheet program. To use the data logger the controller
must be provided with the MJ500 Real Time Clock board (which mounts directly on the MB220
board as shown on Figure 5).
2. Operation: To enable the data logger, enter the configuration menu and select the option “DL”.
The first menu option that appears will be the On/Off menu to enable/disable the data logger.
The menus that follow allow for adjustment of data recording frequency and for changes to the
clock (date and time). See Figure 10.
a. Frequency: The frequency is the time interval between data recordings. The frequency is
adjustable in seconds, with a minimum setting of 5 seconds.
b. Data Logger Clock: The clock is factory-set before shipment. However, because the clock
is set on Eastern Standard time it may be necessary to change the date and time upon start-up.
3. Stored Data Files: The data will be written to text files on the MicroSDHC memory card. The
formatting and handling of these files is as described below:
a. File Format: The following is an example data file to illustrate the format used. As you can
see, there is a three line header for each file. The fourth and fifth lines are headers for the
data. You will see that each header and data entry is delimited by a comma.
b. File Name: Each data file will be named according to the date on which it was created. For
example if created on May 24, 2017, the file name would be May24_17.txt
i. If the MicroSDHC memory card already has a file started earlier on the same day, then
data will be written onto the existing file.
ii. The text files are limited to 5 MB. Once this limit has been reached, a new file will
automatically be created to allow data to continue to be written.
c. Importing data into a spreadsheet: The data files can be imported into a spreadsheet
program. The data format is ‘comma separated values’.
22
III. TROUBLESHOOTING
A. Installation Check – Review each of the following points first.
1. Sensor Installation: Check the following points regarding the sensor installation.
a. Sensor enclosure cover must be securely fastened to protect against corrosion of the transmit-
ter board etc.
b. Sensor must be mounted at a height that is according to Figures 1a and 1b.
c. Sensor must be mounted so that (rain) water cannot come into contact with the sensor element.
Water coming into contact with the sensor element will damage the sensor and cause the need
for sensor replacement. Generally, water damage will cause the sensor to have an above zero
reading that will not return to zero.
d. Ensure that the sensor calibration cap has been removed completely. See Figure 2.
2. Monitor Installation: Check the following points regarding monitor installation.
a. Monitor should be installed at eye level in a location that is suitable for personnel to check
the sensor status before entering the chemical storage room.
b. Monitor should be mounted in a location that is protected from rain and it is recommended
that it should not be mounted under direct sunlight.
c. Monitor enclosure bolts must be securely fastened and wiring seal tights must be plugged if
not used in order to protect against corrosion of the circuit boards etc.
d. Ensure that the alarm relay output and/or 4-20mA outputs are wired according to Section II.E
and Figures 5 and 6.
23
C. Explanation of Responses
1. Zero Calibration: If the display is not reading 0.0 PPM in air, then adjust the zero calibration.
Refer to Section II.C.5 and Figure 10.
2. Calibration Cap: The calibration cap is installed for protection of the sensor during shipping
and storage, but must be removed upon installation. If the sensor cap is not removed, then there
will be no response or a very slow response. Refer to Section I.B.3 and Figure 2.
3. Span Calibration: If the span calibration is performed incorrectly (usually accidentally done in
air with zero target gas) then this will cause the readings to be inaccurate. Unless you intend to
perform the span calibration and have a span gas calibration kit, do not touch the and keys
if you enter the span calibration screen. See Section II.C.8 and Figures 7 and 8.
4. Sensor Wiring: If the display is reading a negative value and giving an “Alarm Status: Error”
message, then the sensor may not be connected to the monitor. Check the wiring from the circuit
board in the monitor to inside the sensor enclosure. See Figures 5 and 6.
5. Sensor Replacement: Repeated or excessive exposure to the target gas and/or cross sensitive
gases will eventually cause failure of the sensor. If water is allowed to contact the sensor element
this will also eventually cause failure of the sensor. Under normal circumstances a sensor life is
typically 2 years or more. However, lightning, other power surges, chemical leaks, and contact
with water can all cause sensor failure. Replacement sensors are easily installed with the quick
disconnect fitting.
6. Damaged Circuit Board: The circuit boards can be damaged if high voltage is connected to the
wrong terminals, by lightning, other power surges, or by corrosion. If you believe that the circuit
board is damaged, then contact the factory and your local sales representative. Refer to Figures
3, 4, 5, and 6.
7. Power Supply: If the display board has no power, then always check to see if the power supply
board is damaged and replace if necessary.
24