Em132-133 0
Em132-133 0
Em132-133 0
1. Ensure that all incoming AC power and other power sources are turned OFF before performing
any work on the instrument. Protect the measurement AC Inputs voltage (V1, V2, V3) with 2A
external overcurrent protection device and the power supply source inputs with 5A external
overcurrent protection device, located close to the equipment.
2. Before connecting the instrument to the power source, check the labels on the back of the
instrument to ensure that your instrument is equipped with the appropriate power supply
voltage, input voltages and currents. Failure to do so may result in serious or even fatal injury
and/or equipment damage.
3. Under no circumstances should the instrument be connected to a power source if it is damaged.
4. To prevent potential fire or shock hazard, do not expose the instrument to rain or moisture.
5. The secondary of an external current transformer must never be allowed to be open circuit
when the primary is energized. An open circuit can cause high voltages, possibly resulting in
equipment damage, fire and even serious or fatal injury. Ensure that the current transformer
wiring is secured using an external strain relief to reduce mechanical strain on the screw
terminals, if necessary.
6. Only qualified personnel familiar with the instrument and its associated electrical equipment
must perform setup procedures.
7. Do not open the instrument under any circumstances when it is connected to a power source.
8. Do not use the instrument for primary protection functions where failure of the device can cause
fire, injury or death. The instrument can only be used for secondary protection if needed.
Read this manual thoroughly before connecting the device to the current carrying
circuits. During operation of the device, hazardous voltages are present on input
terminals. Failure to observe precautions can result in serious or even fatal injury or
damage to equipment.
Table of Contents
Chapter 1 General Information ..........................................................................11
1.1 Features.......................................................................................................................13
1.2 Available Options .......................................................................................................14
Digital I/O ............................................................................................................ 15
Analog Output ...................................................................................................... 15
Additional Communication Port– COM2 ................................................................... 15
1.3 Customized Options ..................................................................................................16
Device Resolution.................................................................................................. 16
Display Options..................................................................................................... 16
1.4 Measured Parameters ..............................................................................................16
1.5 Labelling ......................................................................................................................18
Chapter 2 Installation ..........................................................................................19
2.1 Site Requirements .....................................................................................................19
2.2 Package Contents......................................................................................................19
2.3 Mechanical Installation.............................................................................................19
Wall Mounting....................................................................................................... 21
DIN Rail Mounting ................................................................................................. 22
Add-On Module Mounting....................................................................................... 23
2.4 Electrical Installation.................................................................................................23
Connecting the wires ............................................................................................. 24
Typical Installation ................................................................................................ 25
Terminals ............................................................................................................. 30
Power Source Connection....................................................................................... 31
Voltage Input connection ....................................................................................... 33
Current Input Connection....................................................................................... 34
Wiring Diagrams ................................................................................................... 35
2.5 Battery Replacement ................................................................................................42
2.6 I/O Connections ........................................................................................................43
2DI/1RO Standard ................................................................................................ 43
4DI/2RO Module ................................................................................................... 44
8 DI module ......................................................................................................... 45
12DI/4RO Module ................................................................................................. 45
4AO Module - Analog Outputs................................................................................. 46
2G/3G+2 AI Module – 2 x Analog Inputs ................................................................. 48
2.7 Communications Connections.................................................................................49
RS-485 Connection ............................................................................................... 50
ETH module – COM2 Ethernet Connection................................................................ 51
PRO module – COM2 PROFIBUS Connection............................................................. 53
CANopen module – COM2 CANopen Connection ....................................................... 53
RS-232/422-485 module – COM2 Connection .......................................................... 54
Connecting a Cellular module ................................................................................. 55
Connecting an RF module ...................................................................................... 57
Connecting a WiFi module ...................................................................................... 57
Chapter 3 Using Front Display ............................................................................58
Energy Pulse LED .................................................................................................. 58
COMPort Activity LEDs ........................................................................................... 58
VOLTAGES LEDs ................................................................................................... 58
Navigation Buttons ................................................................................................ 58
The EM13X Series measuring and power supply inputs comply with Measuring
Category III
1
Imax= 63A at 60°C max. amb. temperature, Imax=100A at 55°C max. amb. temperature
Two digital inputs with 10-ms scan time and one relay output with 1-
cycle update time
1
Imax= 63A at 60°C max. amb. temperature, Imax=100A at 55°C max. amb. temperature
1.1 Features
Multifunctional 3-phase Power Meter
3 voltage inputs and 3 current transformer-isolated AC inputs for
direct connection to power line or via potential and current
transformers
True RMS, volts, amps, power, power factor, neutral current, voltage
and current unbalance, frequency
Automatic daily energy and maximum demand profile log for total
and tariff registers
Digital I/O
2 digital inputs (standard) and four to twelve digital inputs (optional)
with 1/2 cycle scan time; automatic recording of last five digital input
1
25Hz and 400Hz options are available only with internal CT.
Display
Easy to read 2x16 characters LCD display, adjustable update time
Communications
Standard 2-wire RS-485 communication port; MODBUS RTU, DNP3,
SATEC ASCII and IEC 60870-5-101 communication protocols
Meter Security
Password security for protecting meter setups and accumulated data
from unauthorized changes
Upgradeable Firmware
Easy upgrading device firmware through a serial or Ethernet port.
Software Support
PAS™ – meter configuration and data acquisition SATEC Software
tool
Analog outputs
RF communication port
Digital I/O
The EM13X Series digital I/O expansion module provides:
4DI/2RO Module
4dry contact digital inputs (DI) for monitoring external contacts and
receiving pulses from energy, water, and gas meters; Programmable
de-bounce time; 1-ms scan time.
8DI module
8 dry contact digital inputs (DI) for monitoring external contacts and
receiving pulses from energy, water, and gas meters
12DI/4RO Module
12 dry contact digital inputs (DI) for monitoring external contacts
and receiving pulses from energy, water, and gas meters;
Programmable de-bounce time; 1-ms scan time.
Analog Output
The EM13X Series analog output (AO) expansion module provides:
4 optically isolated analog outputs with an internal power supply;
PROFIBUS DP port
CANopen port
Device Resolution
A low or high-resolution option can be selected for the presentation of
voltage, current, and power for use in high and low power applications.
Display Options
Different display options are available for customization to be used in
dark or non-safe locations, or in places that are hardly accessible for
observation.
Harmonic MeasurementsEH
Voltage THD per phase
Current THD per phase
Current TDD per phase
K-factor per phase
Voltage harmonics per phase up to order 40
Current harmonics per phase up to order 40
Voltage harmonic angles up to order 40
Current harmonic angles up to order 40
Fundamental Component
Voltage and Current per phase
kW, PF per phase
kvar, KVA per phase
Total kW, PF
Total kvar, KVA
Min/Max Logging
Min/Max A, V, total kW, kvar, kVA, PF
Min/Max Frequency, Neutral current
Phase Rotation
Voltage and Current Phase Angles
Day and Time
1.5 Labelling
Chapter 2 Installation
This chapter discusses the following types of physical installations for
the EM13X Series SMART MULTIFUNCTION METER:
Mechanical Installation
Electrical Installation
I/O Connections
Technical Documentation CD
Wall Mounting
To mount the meter on the wall:
Position the meter on the wall according to hole locations as shown in
figure 2-2
Affix the meter using washers and nuts.
Turn off the EM13X Series auxiliary power supply or voltage supply
inputs (SE model).
Remove the module connector cover from the EM13X Series to access
the module connector as shown in figure 2-4below
Conductor size
Torque
Terminals EM133 model Minimum Maximum Notes
[Nm]
AWG (mm²) AWG (mm²)
Aux. Power N/A in EM133- 22 (0.5) 12 (4.0) 0.5-0.7 Use 600V insulated conductors
Supply Inputs SE models Required crimped ferrule:
L/+, N/- Panduit (22AWG) F75-10-M
Panduit (12AWG) F81-15-M
Voltages Inputs N/A in EM133- 22 (0.5) 12 (4.0) 0.5-0.7 Use 600V insulated conductors
V1, V2, V3, Vn SE-100A Required crimped ferrule:
Panduit (22AWG) F75-10-M
Panduit (12AWG) F81-15-M
Current Inputs EM133/SE- 22 (0.5) 10 (6.0) 1.8-2 Use 600V insulated conductors
I1, I2, I3, N 1A, 5A, HACS Required crimped ferrule:
Panduit (22AWG) F75-10-M
Panduit (10AWG) F82-15-M
Current Inputs EM133/SE- 6 (16) 4 (25) 2-2.7 Use 600V insulated conductors
I1, I2, I3, N 63A or 100A Required crimped ferrule:
Panduit (6AWG) F84-15-M
Panduit (4AWG) F85-15-M
COM1, I/O All Models 22 (0.5) 12 (4.0) 0.5-0.7 Use 600V insulated conductors
connections Required crimped ferrule:
Panduit (22AWG) F75-10-M
Panduit (12AWG) F81-15-M
Typical Installation
*
1. SATEC RF chokes to be installed between CBR and Meter
Voltage Inputs in residential area or other places require ANCI
C12.20 compliance. Otherwise it may be bypassed.
2. AC0242 shall be routed with the voltage lines (bundled),
separately from communication & I/O lines
Figure 2-6b Typical Installation Diagram with Self Energized Power Supply – direct connection with HACS
Figure 2-6c Typical Installation Diagram with Self Energized Power Supply – direct connection
*
1. SATEC RF chokes to be installed between CBR and Meter
Voltage Inputs in residential area or other places require ANCI
C12.20 compliance. Otherwise it may be bypassed.
2. AC0242 shall be routed with the voltage lines (bundled),
separately from communication & I/O lines
Figure 2-6d Typical Installation Diagram with Self Energized Power Supply – Transformer Operated
(indirect connection)
Figure 2-6e Typical Installation Diagram with 12/24 VDC Auxiliary Power Supply
Terminals
The EM132/133/133-SE terminals are constructed as a built-in part of
the meter equipment. The terminals are divided in two different
terminal categories:
Figure 2-9EM13X Series (with Aux. Power Supply) measured Voltage connections
690V inputs are usually used with direct connection. Use any of the
seven wiring configurations shown in Figures 2-8 through 2-15.
1
Imax= 63A at 60°C max. amb. temperature, Imax=100A at 55°C max. amb. temperature
Wiring Diagrams
For AC input ratings, see Technical Specifications in Appendix A for
more details.
Figure 2-12 3-Wire 2-Element Delta Direct Connection Using 2 CTs (Wiring Mode = 3dir2)
Figure 2-13 4-Wire Wye 3-Element Direct Connection Using 3 CTs (Wiring Mode = 4LL3 or 4Ln3)
Figure 2-14 4-Wire Wye 3-Element Connection Using 3 PTs, 3 CTs (Wiring Mode = 4LL3 or 4Ln3)
Figure 2-15a 3-Wire 2-Element Open Delta Connection Using 2 PTs, 2 CTs (Wiring Mode = 3OP2)
Figure 2-15b 3-Wire 2-Element Open Delta Connection Using 3 PTs, 2 CTs (Wiring Mode = 3OP2)
This configuration provides accurate power measurements only if the voltages are balanced.
Figure 2-16 4-Wire Wye 2½-Element Connection Using 2 PTs, 3 CTs (Wiring Mode = 3LL3 or 3Ln3)
Figure 2-17 3-Wire 2½-Element Open Delta Connection Using 2 PTs, 3 CTs (Wiring Mode = 3OP3)
Figure 2-18 4-Wire 3-Element Delta Direct Connection Using 3 CTs (Wiring Mode = 4LL3 or 4Ln3)
Figure 2-19 3-Wire 2½-Element Broken Delta Connection Using 2 PTs, 3 CTs (Wiring Mode = 3bLn3 or
3bLL3)
2DI/1RO Standard
Relay Output
The EM13X Series basic provides one dry contact relay (SSR FORM A)
for energy pulsing, alarms or remote control
Digital Inputs
Two optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time
synchronization.
4DI/2RO Module
Before I/O Module installation ensure that all incoming power sources
are shut OFF. Failure to observe this practice can result in serious or
even fatal injury and damage to equipment.
Relay Outputs
There are two relay outputs provided for energy pulsing, alarms, or
remote control.
Digital Inputs
Four optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time
synchronization.
8 DI module
Eight optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time
synchronization
12DI/4RO Module
The 12DI/4RO module can be equipped with optional communication
port COM2 – ETHERNET or RS-422/485.
Relay Outputs
Digital Inputs
The 4AO module TERMINAL is for use only with equipment which has no
live parts which are ACCESSIBLE
The RATING of the insulation of the external equipment for use with the
4AO module, shall comply according to Installation Category III for
insulation to be suitable for SINGLE FAULT CONDITION
The external equipment TERMINAL connection type is normally terminal
block for wire size 14 AWG (up to 1.5 mm2)
The type of equipment that might be connected to the TERMINAL is:
Programmable Logic Controller for automation – PLC
Digital or Analog meter
Ethernet 10/100BaseT
PROFIBUS DP
CANopen
RS-232 or RS-422/485
2G/3G/4G
2G/3G+2AI
RF
COM3 (standard): IR
RS-485 Connection
RS-485 COM1
The PRO module TERMINAL is for use only with equipment which has no
live parts which are ACCESSIBLE
The RATING of the insulation of the external equipment for use with the
PRO module, shall comply according to Installation Category III for
insulation to be suitable for SINGLE FAULT CONDITION
The external equipment TERMINAL connection type is DB9
The type of equipment that might be connected to the TERMINAL is:
Programmable Logic Controller for automation – PLC
1) 2G/3G module, mainly for GSM network (Regular or DIN rail form
factor)
The EM13x Cellular module can be used in two different form factors:
The Cellular SIM must not have any incoming voice call. The customer
must require from the Service Provider for DATA services only
The Cellular module can be equipped with two different antennas:
internal Antenna for installation into plastic closet or no-metallic
environment. For metallic installation use external antenna
Connecting an RF module
The energy test mode can be enabled through the Device Options
setup. When in test mode, the energy and demand accumulators do
not account for consumed energy.
VOLTAGES LEDs
The meter has three green LEDs “L1, "L2" and "L3”, which indicate
voltage measurement input presence. When LEDs are "ON", that
means that the voltage measurement inputs exist in the meter
Navigation Buttons
The EM13X Series is provided with two push buttons that are normally
used to navigate between different measurement displays.
Navigation Buttons
The EM13X Series is provided with two navigation buttons,
Display Features
The EM13X Series display has a number of programmable features that
can be disabled, enabled and adjusted via the meter Display Setup
(see Display Setup in Chapter 5).
Backlight
A short press on any button while the display backlight is off switches
the backlight on.
The backlight stays on as long as you selected in the display setup and
then dims to conserve power. The backlight time is factory set to 1
minute and can be programmed from 1 to 10 minutes. You can
temporarily set the backlight to continuous operation if you need to
work in dark for more time.
Auto-Return
To stop auto scrolling, press briefly any button if the backlight is on;
else press briefly any button twice since the first press only sets the
backlight on and does not affect auto scrolling.
Measurement Units
The following table shows the display resolution for common displayed
quantities. All measured data is displayed in primary units.
Measured Voltage Connection Units and Display Resolution
Quantity
Energy kWh, kvarh, kVAh with one decimal
place. The number of digits is
programmable (see Device Options
and Mode Control in Chapter 5).
Power Direct kW, kvar, kVA with three decimal
(PT = 1.0) places
Transformer MW, Mvar, MVA with three decimal
(PT>1.0) places
Voltage Direct Volts with one decimal place
(PT = 1.0)
Transformer Kilovolts with three decimal places
(PT>1.0)
Current Amperes with two decimal places
Only registers you selected in the billing/TOU register setup and tariff
rates listed in the TOU daily profiles are included (see Configuring
Billing/Tariff Registers and Configuring the Daily Tariff Schedule in
Chapter 5).
Reg 2 Imp
Total Import Reactive energy data.
221 kvarh
1
Only from S/W V12.2.1
DISPLAY Description
DISPLAY Description
Total Import Max. DMD active power data
(Reg1 for TOU/Max. DMD actve power,
Reg 1 Imp MAX Reg2 for TOU/Max. DMD reactve power
predefined TOU/Register factory setup,
0.008 kW can be changed by user, see Configuring
Billing/Tariff Registers and Configuring the
Daily Tariff Schedule in Chapter 5)
Reg 1/T1 Imp MAX Tariff 1 Import max. active power dmd
0.008 kW data.
Reg 1/T2 Imp MAX Tariff 2 Import max. active power dmd
0.008 kW data
Reg 1/T3 Imp MAX Tariff 3 Import max. active power dmd
0.002 kW data
Reg 2/T1 Imp MAX Tariff 1 Import max. reactive power dmd
0.003 kvar data.
Reg 2/T2 Imp MAX Tariff 2 Import max. reactive power dmd
0.003 kvar data.
Reg 2/T3 Imp MAX Tariff 3 Import max. reactive power dmd
0.001 kvar data.
Reg 3/T1 Imp MAX Tariff 1 Import max. apparent power dmd
0 kVA data.
Reg 3/T2 Imp MAX Tariff 2 Import max. apparent power dmd
0 kVA data.
Reg 3/T3 Imp MAX Tariff 3 Import max. apparent power dmd
0 kVA data.
DISPLAY Description
MAX. DMD
Total Import Max. DMD active power data
P Imp 0.008 kW
MAX. DMD
Total Export Max. DMD active power data
P Exp 0 kW
MAX. DMD
Max. DMD Phase 1 current data
I1 0 A
MAX. DMD
Max. DMD Phase 2 current data
I2 0 A
MAX. DMD
Max. DMD Phase 3 current data
I3 0 A
MAX. DMD
Max. DMD neutral current data
In 0 A
MAX. DMD
Max. DMD Phase 1 voltage data
V1 0 V
MAX. DMD
Max. DMD Phase 2 voltage data
V2 0 V
MAX. DMD
Max. DMD Phase 3 voltage data
V3 0 V
Instrumentation Measurement
Instrumentation Measurements represent general instrumentation data
you can use while installation and inspecting the meter. Use phase
angles displays to check the order of phases when connecting wires to
the meter terminals.
DISPLAY Description
V3 0 V
V2 Phase-to-neutral voltage
V12 0 V
Phase-to-Phase voltages.
V23 0 V
V31 0 V
Phase-to-Phase voltages.
I1 0 A
Phase currents.
I2 0 A
P 0 kW
Total powers.
Q 0 kvar
L1 P 0 kW
Phase 1 active and reactive power
L1 Q 0 kvar
DISPLAY Description
L1 S 0 kVA
Phase 1 apparent power and power factor
L1 PF 0
L2 P 0 kW
Phase 2 active and reactive power
L2 Q 0 kvar
L2 S 0 kVA
Phase 2 apparent power and power factor
L2 PF 0
L3 P 0 kW
Phase 3 active and reactive power
L3 Q 0 kvar
L3 S 0 kVA
Phase 3 apparent power and power factor
L3 PF 0
H1 P 0 kW
First Harmonic Total powers.
H1 Q 0 kvar
DISPLAY Description
V1 THD 0 %
Phase 1 voltage and current THD
I1 THD 0 %
V2 THD 0 %
Phase 2 voltage and current THD
I2 THD 0 %
V3 THD 0 %
Phase 3 voltage and current THD
I3 THD 0 %
I1 TDD 0 %
Total Demand current Distortion
I2 TDD 0 %
I3 TDD 0 %
Total Demand current Distortion
V Unb 0 %
Voltage and Current unbalance
I Unb 0 %
Freq 0 Hz
Frequency
DISPLAY Description
DISPLAY Description
DI: 123456
Digital Inputs status
000000
Relays: 123
Relays control status
000
Counter 1:
Status/Event Counter 1
0
Counter 2:
Status/Event Counter 2
0
Counter 3:
Status/Event Counter 3
0
Counter 4:
Status/Event Counter 4
0
The diagnostic icon can be disabled or enabled via the Display Setup
menu.
DISPLAY Description
Diagnostics:
Power Down message
8
DISPLAY Description
Boot: V1.1.1
Boot SW version and Modem setup
Modem: N/A
COM1: a221,b115.2
Communication port COM1 info
Modbus RTU
COM3: a1,b19.2
Communication port COM3 info
Modbus RTU
Navigation Buttons
The following table gives a summary of the button operations in
programming mode.
Button Press Operations
SCROLL Short press Scroll through a menu item list in a
highlighted window or increment a
highlighted digit in a numeric field
SELECT/ENTER Short press (less than Highlight a menu window or a digit in a
1 second) = SELECT numeric field
SELECT/ENTER Long press (1 to 2 Store the changed item or perform an
seconds) = ENTER action indicated in a highlighted window
Basic PT Ratio Whenever a numeric value is to be changed, use a short press on the
100.0 SELECT/ENTER button to highlight a desired digit, and then use the
SCROLL button to change the value of the highlighted digit. A
SHORT PRESS
highlighted digit appears in inversed color. If you missed a digit, just
Basic PT Ratio continue moving through the rest of digits until you reach the desired
100.0 place.
Once the number is set to the desired value, press and hold the
Basic PT Ratio SELECT/ENTER button for 1-2 seconds to save your new setting.
150.0
LONG PRESS
To reject your changes and restore the previous value, use a short
press on the SELECT/ENTER button to return to the higher-level
window.
Password Security
The setup menus are secured by 8-digit user passwords. Every time
Enter Password you enter programming mode, you are prompted for a correct
00000000 password. The meter is primarily shipped with all passwords preset to
9 at the factory.
SHORT PRESS
Enter Password It is recommended that you change the factory set passwords as fast
00000009 as possible to protect your setups and accumulated data from
LONG PRESS unauthorized changes. See Configuring Meter Passwords in Chapter 5
on how to change passwords in your meter.
Enter the password as you enter numeric values. As you move to the
next place, the digit entered is saved and then zeroed. If you missed a
digit, you should re-type all preceding digits before you reach the
missed place again.
Once the password is set to the desired value, press and hold the
SELECT/ENTER button for more than 1 second. If the password you
entered is correct, you move to the main device menu, otherwise you
return back to the data display.
If your security level does not allow access to a menu, it will not be
listed in the main menu list, and you will not be able to highlight menu
items that you are not allowed to change, but you can still view their
present settings.
Disp Exit
Use the SCROLL button to scroll through the menu list until the desired
menu entry appears
LONG PRESS
Press the SELECT/ENTER button for more than 1 second to enter the
Disp Autoscroll
5
selected submenu.
SHORT PRESS Once you entered a submenu, the left window is still showing the menu
name, while the upper-right window represent a submenu options list,
Disp Autoscroll
5 and the lower-right window indicates the present option value.
This chapter gives information on how to install and run PAS on your
computer, and how to prepare information for your meter using PAS.
For general information on how to work with PAS, see the “PAS Getting
Started” guide supplied on the installation CD.
From the Look in box, select the directory where a new database will
be stored. By default, it is the Sites directory.
Type a site name for your device in the File name box, click New, and
then click OK.
On the Instrument Setup tab, select EM13X Series in the Model
box. PAS automatically selects the appropriate instrument options for
your meter.
Select a correct CT secondary current (5A or 1A) for your meter.
If you wish to add any comments for your meter, type the comments
in the Comment box.
Select Communications Setup from the Meter Setup menu, and then
click on the Serial Ports Setup tab. In the Port box, select the desired
device port.
Figure 5-1: Communication Setup Dialog Box – Serial Ports Setup Tab
1
GPRS – also covers 3G & 4G network
Select Net from the main menu. See Viewing and Changing Setup
Net IP Address
Options in Chapter 3 for information on navigating in menus.
192.168.000.203
Using PAS
Select Communications Setup from the Meter Setup menu, and then
click on the Network Setup tab.
NOTES
The meter provides the permanent MODBUS TCP server on port 502.
Selecting the DNP3 TCP service port launches the second DNP3 TCP
server allowing simultaneous connections on both TCP ports. Selecting
the Modbus TCP port disables the DNP3 TCP server.
The TCP service port can also be changed trough the COM2 serial port setup.
Changing the communication protocol for the port automatically changes the
TCP port for the Ethernet.
When you change the device network settings through the Ethernet port,
the device port restarts so communication will be temporarily lost. You
may need to wait some additional time until PAS restores a connection
with your device.Setting-Up Dial-Up GPRS Network
You can check the status of the GPRS communications from the front
panel via the Status Display or via the Device Control dialog in PAS
(see Viewing Communication Status and Statistics).
Security Key – the pass phrase to authenticate the meter with your
wireless network.
If you connect the meter to a SATEC WiFi network, use same network
authentication parameters as you configured for the network access
point/router (see Section 3 below).
If you connect the meter to a foreign WiFi network, you can identify
the wireless network parameters via the Windows Network and Sharing
Center on your PC connected to the WiFi network:
On Windows 7 and 8.1, click the Start button, and then open the
Control Panel. Select Network and Internet -> Network and Sharing
Center.
On Windows 10, click the Start button, then select Settings -> Network
& Internet -> Status -> Network and Sharing Center.
WiFi Protocol – the WiFi physical layer protocol that defines the
supported WiFi network technology and network bandwidth. If you find
it difficult to choose the right one to cover all technologies supported
by your network devices, it’s recommended that you leave the default
802.11b/g/n protocol setting.
10.0.0.1 – 10.255.255.254
172.16.0.1 – 172.31.255.254
192.168.0.1 – 192.168.255.254
You can configure the network address in your meter either from the
meter front display (see the meter operation manual for more
information), or via PAS.
If you connect the meter to a foreign WiFi network, you can identify
the default gateway address and subnet mask via the Windows
Network and Sharing Center on your PC connected to the WiFi network
as shown above in Section 2. In Network and Sharing Center, next to
Connections, select your Wi-Fi network name. In Wi-Fi Status, select
Details. Consult with the network administrator for the
applicable host IP addresses, or select one within the network
address segment that has not yet been assigned to any host across
the WiFi network.
You can monitor the connection status and signal quality via the meter
front display or via PAS.
In the EM133, move to the Diagnostics display and then scroll to the
“WiFi” page where you can see the network connection status and
received signal quality. The following page indicates the meter network
address.
In the PM130, move to the STA display and then scroll to the “rSSi”
page where you can see the received signal quality in percent followed
by the network connection status.
In the PM135, move to the Device Info display and then scroll to the
“WiFi” page.
To check a WiFi connection via PAS, select Device Control from the
Monitor menu and click on the Communications tab. The connection
status and received signal quality are indicated under the WiFi Wireless
Communications section.
NOTES
Do not enable the eXpertPower client in your meter if you do not use the
eXpertPowerTM service.
Do not change the connection time setting. It is for information only. The
eXpertPower server updates it automatically.
See the PM130 PLUS Modbus Reference guide for more information on
operation of the notification client and the notification message
structure.
Before operating your meter, provide the device with basic information
about your electrical network.
Select the Basic entry from the main menu. See Viewing and Changing
Basic Exit
Setup Options in Chapter 3 for information on configuring parameters
via the front display.
LONG PRESS
See the table below for the Basic Setup options.
Basic Conf
4Ln3
SCROLL until
CT menu
Basic CT
50
Using PAS
Select General Setup from the Meter Setup menu. See the table
below for the Basic Setup options.
1. Always specify the wiring mode and transformer ratings prior to setting up
setpoints and analog outputs.
2. The maximum value for the product of the phase CT primary current and PT
ratio is 57,500,000. If the product is greater, power readings are zeroed.
In 4LN3, 3LN3 and 3BLN3 wiring modes, Min/Max volts, volt demands and
voltage harmonics represent line-to-neutral voltages; otherwise, they will be
line-to-line voltages.
Device Options
The Device Options setup allows changing user-configurable device
options or putting the meter into energy test mode.
Select Options from the main menu. See Viewing and Changing Setup
Options TestMode
Options in Chapter 3 for information on configuring parameters via the
OFF front display.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Device Options tab.
The power calculation mode option allows you to change the method
for calculating reactive and apparent powers in presence of high
harmonics. The options work as follows:
When the reactive power calculation mode is selected, active and
reactive powers are measured directly and apparent power is
calculated as:
S P2 Q2
N S2 P2
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Digital Inputs tab.
Figure 5-8: General Setup Dialog Box – Digital Inputs Dialog Box
Parameter
T Options Default Description
Input
h Polarity NORMAL NORMAL For the normal polarity, the open
e INVERTING to closed transition is considered
closed. For the inverting polarity,
the closed to open transition is
d
considered a closed
e
Pulse
b Input PULSE MODE PULSE In pulse mode, either leading, or
Mode KYZ MODE MODE trailing edge of the input pulse is
o
recognized as an event. In KYZ
u mode, both leading and trailing
n edges of the input pulse are
c recognized as separate events.
e Polarity
Pulse NORMAL (N.O.), NORMAL For the normal polarity, the open
INVERTING (N.C.) to closed transition is considered
t a pulse. For the inverting polarity,
i the closed to open transition is
m considered a pulse.
e It has no meaning in KYZ mode
where both transitions are used.
Debounce
i 1-100 ms 10 ms The amount of time while the
Time
s state of the digital input should
not change to be recognized as a
new state. Too low debounce time
a could produce multiple events on
p the input change.
a
The debounce time is applied the same for all digital inputs. If you
change the debounce time for a digital input, the same debounce time
is automatically assigned to the others.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Relay Outputs tab.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Analog Outputs tab.
When you select an output parameter for the analog output channel,
the default engineering scales are set automatically. They correspond
to the maximum available scales. If the parameter actually covers a
lower range, you can change the scales to provide a better resolution
on an analog output.
The engineering scale for the signed power factor emulates analog
power factor meters.
For an unsigned output value, you can change both zero and full
engineering scales.
The meter does not allow access to the low scale setting if the
parameter is directional. Whenever the sign of the output parameter is
changed to negative, the meter automatically uses the full engineering
scale setting for +1 mA with a negative sign.
Scaling Analog Outputs for 0-2 mA and ±2 mA
The 0-1mA and ±1mA current outputs provide a 100% overload, and
actually output currents up to 2 mA and ±2mA whenever the output
value exceeds the engineering scale you set for the 1 mA or ±1mA.
Configuring Counters
The EM13X Series has four six-digit general counters that can count
pulses delivered through the device digital inputs with a programmable
scale factor. Each counter can also be incremented in response to any
internal or external event, checked and cleared through the Control
Setpoints.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Pulse/Event Counters tab.
The logical controller provides very fast response to events. The scan
time for all setpoints is 1 cycle time (16.6 ms at 60Hz and 20 ms at
50/400 Hz).
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Control/Alarm Setpoints tab.
For numeric (analog) triggers, you can specify two thresholds for each
trigger to provide hysteresis (dead band) for setpoint operations.
The Operate Limit defines the operating threshold, and the second
Release Limit defines the release threshold for the trigger. The trigger
thresholds are specified in primary units.
If you do not want to use hysteresis for the trigger, set the Release
Limit to the same value as the Operate Limit.
Using Binary Triggers
Binary (digital) triggers like digital inputs and relays are tested for
ON/CLOSED or OFF/OPEN status.
A relay output is operated when one of the setpoints linked to the relay
is activated and remains operated until all of these setpoints are
released (except for latched relays that require a separate release
command to be deactivated).
Select the Disp entry from the main menu. See Viewing and Changing
Setup Options in Chapter 3 for information on configuring parameters
Disp AutoScroll via the front display.
5
See Table 16 for the available options.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Display Setup tab.
Select Local from the main menu. See Viewing and Changing Setup
Locak Exit Options in Chapter 3 for information on configuring parameters via the
front display.
Using PAS
Select General Setup from the Meter Setup menu, and then click on
the Local Settings tab.
Select the Access entry from the main menu. See Viewing and
Changing Setup Options in Chapter 3 for information on configuring
parameters via the front display.
Access Passwrd 1 Apply LONG PRESS on ENTER button to select the Passwrd 1 entry
9 Apply SHORT PRESS on ENTER button to select the password digit to
change.
LONG PRESS
Apply LONG PRESS on ENTER button to exit the Passwrd 1 digit field
Access Passwrd 1
00000009 Apply LONG PRESS on ENTER button to exit the Passwrd 1 field to
move to Access field
SHORT PRESS
Using PAS
Ensure that the On-line button on the PAS toolbar is checked, select
Administration from the Monitor menu, and then select Change
Password -> Password 1.
The EM13X Series can provide automatic recording of the daily energy
and maximum demand profile to a data log file. Maximum demand
profiling can be individually configured for every register.
See Factory Preset Data Log Files in Chapter 5 and Billing Profile Log
File in Appendix E for more information on the file layout and contents.
To configure the billing registers and the tariff system in your meter:
Link the billing registers to the respective energy sources.
Configure the options for the registers to whether the only totalization or
both total and tariff registers would be used, and whether daily profiling
should be enabled for the energy usage and maximum demand
registers.
Configure the daily tariff schedule using the TOU daily profiles for all
types of days and seasons.
Configure the season tariff schedule using the TOU calendar.
Figure 5-17: Energy/TOU Setup Dialog Box – TOU Daily Profiles Tab
The daily profile setup allows you to specify the daily tariff change
points with a 15-minute resolution for 4 seasons using 4 different daily
schedules for each season.
To configure your daily profiles:
Select the desired season and day type.
Select the start time for each tariff change point and the corresponding
active tariff number.
Repeat the setup for all active profiles.
The first tariff change point is fixed at 00:00 hours, and the last tariff
change you specified will be in use until 00:00 hours on the next day.
NOTE
The billing daily profile log file is automatically configured for the
number of active tariffs you defined in the meter TOU daily profiles.
Figure 5-18: TOU Calendar Setup Dialog Box – TOU Calendar Tab
The meter TOU calendar allows you to configure any tariff schedule
based on any possible utility regulation. The calendar provides 32
entries that allow you to specify profiles for working days and holidays
through all seasons in any order that is convenient for you, based on
simple intuitive rules.
To configure your season tariff schedule:
In the Season/Period box, select the season, and in the Day Type
box, select a day type for this calendar entry.
Define the time interval when this daily tariff schedule is effective,
based on the start and end weekdays and, for a multi-season schedule,
on the start and end month for the selected season. It does not matter
which order of weekdays or months you select: the meter recognizes
the correct order.
For exception days like weekends and designated holidays, define a
specific day either by a month and a month day, or by selecting a
month, a week and a weekday within the month.
Event log
The two data log files are pre-configured at the factory for recording a
15-minute energy and demand profile and for the daily billing energy
data profile. If you wish to change the factory settings, follow the
guidelines in the next section.
For your reference, the record size and the number of records available
for your file are reported in the dialog box.
To delete an existing file partition:
Click on Delete.
Click OK.
NOTES
Memory is allocated for a file statically when you set up your files and
will not change unless you re-organize files.
The meter automatically performs de-fragmentation of the memory each
time you re-organize your files. This prevents possible leakage of
memory caused by fragmentation. It may take a couple of seconds.
The following table can help you calculate an estimated file size when
planning your memory allocation.
Click OK, and then send your new setup to the meter, or save to the
device database.
Data log #16 is configurable to store TOU daily profile log records on a
daily basis.
Double click on the Data Log#16partition with the left mouse button.
Select the number of season tariffs in your TOU schedule. Add one
additional parameter if you selected to record the Summary (TOU
total) registers as well.
Click OK and send your setup to the meter or save to the meter
database.
Data log #16 is pre-configured for daily billing energy and maximum
demand recording for the last 90 days. It is automatically updated
once a day.
See Billing Profile Log File in Appendix E for the file record structure.
Configuring Modbus
Modbus Point Mapping
Type in the actual addresses you want to read from or write to via the
assignable registers. Refer to the PM130 PLUS Modbus Reference Guide
for a list of the available registers. Note that 32-bit Modbus registers
should always start at an even register address.
Click Send to download your setup to the meter.
The EM13X Series allows you to read 32-bit Modbus analog registers,
energy counters and binary counters either in integer format, or in
IEEE single precision floating point format.
Configuring DNP3
Refer to the EM13X Series DNP3 Reference guide for information on
the DNP3 protocol implementation and a list of the available data
points.
DNP Options
Select Protocol Setup from the Meter Setup menu and click on the
DNP Options tab.
The following table describes available options. Refer to the DNP3 Data
Object Library document available from the DNP User’s Group on the
DNP3 object types.
Refer to the PM130 DNP3 Reference Guide for information on the data
point scales and a reverse conversion that should be applied to the
received scaled values.
1
The Select Before Operate command causes the meter to start a timer. The
following Operate command must be sent before the specified timeout value
expires.
2
The meter requests time synchronization by bit 4 in the first octet of the
internal indication word being set to 1 when the time interval specified by the
Time Sync Period elapses. The master should synchronize the time in the meter
by sending the Time and Date object to clear this bit. The meter does not send
time synchronization requests if the Time Sync Period is set to 0.
If the scaling unit is greater than 1, the counter value is reported being
divided by the selected scaling unit from 10 to 1000. To get the actual
value, multiply the counter reading by the scaling unit.
The most common method of getting static object information from the
meter via DNP is to issue a read Class 0 request. The EM13X Series
allows you to configure the Class 0 response by assigning ranges of
points to be polled via Class 0 requests.
To view or build a DNP Class 0 response message:
Select Protocol Setup from the Meter Setup menu and click on the
DNP Class 0 Points tab.
Select the object and variation type for a point range.
Specify the start point index and the number of points in the range.
Refer to the EM13X Series DNP3 Reference Guide for available data
points.
Repeat these steps for all point ranges you want to be included into the
Class 0 response.
Click Send to download your setup to the meter.
The factory-set Class 0 point ranges are shown in the picture below.
Figure 5-23: Protocol Setup Dialog Box – DNP Class 0 Points Tab
Select the RESET entry from the main menu. See Viewing and
Changing Setup Options in Chapter 3 for information on configuring
parameters via the front display.
To reset the desired values:
Highlight the middle window by pressing briefly the SELECT button.
Select a reset entry by scrolling through the list with the UP and
DOWN arrow buttons.
Press the SELECT button briefly to highlight the lower item.
Press and hold the ENTER button for 5 seconds.
Release the button.
The do entry is replaced with done showing the operation is complete.
Table 21 shows the reset options available from the front display.
Using PAS
Ensure that the On-line button on the PAS toolbar is checked, and
then select Reset from the Monitor menu.
Using PAS
Ensure that the On-line button on the PAS toolbar is checked, and
then select RTC from the Monitor menu or click on the Real-Time
Clock button on the PAS toolbar.
The RTC dialog box displays the current PC time and the time in your
meter.
Using PAS
Ensure that the On-line button on the PAS toolbar is checked, select
Device Control from the Monitor menu, and then click on the Device
Diagnostics tab.
You can also clear the communications counters via the PAS Reset
dialog (see Resetting Accumulators, Maximum Values and Files).
Figure 6-6: Device Control Dialog Box – Remote Relay Control Tab
You are asked for the password regardless of the password protection
setting in your meter. Type the meter password, and click OK. If you
did not change the password in the meter, enter the default password
0.
Wait until PAS completes upgrading your device. It takes about 3-4
minutes at 115,200 bps to download the file to the meter.
Polling Devices
You can open as many monitor windows as you wish, either for
different sites, or for the same site using different data sets. An open
data monitor window is linked to the current site and does not change
if you select another site in the site list.
Some data sets are preset for your convenience and others are empty.
You can freely modify data sets.
Polling Options
To change the polling options, click on the Data Monitor window with
the right mouse button and select Options.
Click the window with the right mouse button, select Options, adjust
the number of records you want to see in the window, and then click
OK. When the number of retrieved records exceeds the number of rows
in the window, the window scrolls up so that older records are erased.
To change the time range for your graph, click on the button on the
local toolbar, and then select the desired date and time range.
To save retrieved data to a file for later analysis, click on the Save
button , select an existing database or type the name for a new
database, and then click Save.
To avoid confusion, do not store data files into the Sites directory
where site databases are located.
Printing Data
To check the report, as it will look when printed, select Print Preview
from the File menu.
When you run real-time data polling, PAS automatically saves retrieved
records to a database at the rate you specified.
PAS supports 9 programmable data sets that you can organize as you
wish. To build your data sets, select MinMax Data Sets from the
Monitor menu or click on the button on the local toolbar.
Use the Poll button for a single-step poll or the Continuous poll
button for continuous polling.
To select the channels you want to view, click with the right mouse
button on the waveform window, select Channels..., check the
channels for the phase you want displayed, and then click OK.
When you open a new file, PAS shows you a waveform graph with non-
overlapped waveforms as shown in the picture above.
The phasor diagrams show you relative magnitudes and angles of the
three-phase voltage and current fundamental component. All angles
are shown relative to the reference voltage channel.
PAS can give you indication on whether harmonic levels in the sampled
waveforms exceed compliance limits defined by the power quality
standards or local regulations.
To review or change harmonic limits:
Click on the spectrum window with the right mouse button and select
Limits….
Select a known harmonics standard, or select Custom and specify
your own harmonic limits.
Check the Enabled box to visualize harmonic faults on the spectrum
graph and in harmonic tables.
The spectrum table shows voltage, current, active power and reactive
power harmonic components both in percent of the fundamental and in
natural units, and harmonic phase angles.
To change a phase, click on the window with the right mouse button,
select Options..., check the phase you want displayed, and then click
OK.
Historical data can be uploaded on demand any time you need it, or
periodically through the Upload Scheduler that retrieves data
automatically on a predefined schedule, for example, daily, weekly or
monthly.
Select a database, or type the name for a new database, and select a
directory where you want to save it.
Click on the Select Logs button and check boxes for logs you want to
be retrieved from the meter.
If you wish to retrieve data starting with a known date, check the
“From” box and select the start date for retrieving data.
If you wish to retrieve data recorded before a known date, check the
“To” box and select the last date for retrieving data.
Click OK.
Click Add Site, point to the site database for which you want to
organize the schedule, and then click OK.
Click Browse and select a database for storing retrieved data, or type
the name for a new database, select a directory where you want to
save it, and then click OK.
Click Configure or double click on the site row.
Select a daily, weekly or monthly schedule, and adjust the start time.
If you wish to upload data periodically in predefined intervals, click on
Periodic and define the time period in hours and minutes.
Select the number of attempts to upload data in the event of
temporary communication problems or unavailability of your device,
and the delay between attempts in minutes and seconds.
If you wish to use the schedule to synchronize the device clock with
your PC, check the RTC Synchronization Enable box. If your device
is password protected by a communications password, type in the
password you set in the device to allow PAS to update the clock.
Click on the Select Logs button, check the boxes for logs you want to
upload on a schedule, and then click OK.
Check the Enabled box at left to activate a schedule for the device.
Click Close to store your schedule.
You can run the scheduler at any time outside the schedule by
checking the Start Now box at right. This is a one-time action. After
uploading is completed, the Upload Scheduler un-checks this box
automatically.
When the Upload Scheduler fails to retrieve data from the device, or
some data is missing, or another problem occurs, it puts an error
message to the log file. To review this file, select System Log from
the View menu.
To view the log data on-line, check the On-line button on the PAS
toolbar, select the log you want to retrieve in the Logs menu, and then
click on the Pollbutton . Only new log records are retrieved from
the device. If you want to review the entire log from the beginning,
click on the Restore logbutton , and then click on the Poll button
.
NOTE
Manual Converting
Select the database and a data log table you want to export, and then
click Open.
Select a folder where you want to store your exported files, type a file
name that identifies your files, select a file output format, and then
click on the Save button.
The PQDIF files are commonly recorded in compressed format. If you
do not want your files to be compressed, uncheck the Compress box
before saving the file.
12KVSUB_20040928T133038.pqd.
Automatic Converting
PAS allows you to automatically convert data logs into PQDIF format at
the time you upload data from your devices via the Upload Scheduler.
The first row of the Excel table lists data names (see Appendix D) and
the second row provides data codes, which identify recorded data
points (see Modbus communications guide for data codes) that may be
useful for automated table processing.
Each table row is provided with the device identifier that you can
define in the meter database (see Creating a New Site for your Meter).
An archive file keeps the original database name to which the date of
the oldest database record is added, so you can easily identify your
archives and work with them as you work with a regular database.
Check the Enable box and select a periodic schedule for archiving your
files for this site.
Click OK.
Names of the last 16 files you opened are stored in the File menu, so
you can select them directly from the menu.
To avoid confusion, do not store data files into the Sites directory
where site databases are located.
To change units, click on the table with the right mouse button, select
Options, select the desired units for voltages and currents, and then
click OK.
Copying a Table
To copy the entire table, or its part, into the Clipboard or into another
application such as Microsoft Excel or Word:
Click on the data window with the right mouse button and choose
Select All, or click on the upper-left corner of the table (where the
“No.” label is commonly displayed).
Click with the right mouse button on the window again and choose
Copy, or click on the Copy button on the PAS toolbar.
Run the application to which you want to copy data, position the cursor
at the correct place.
Printing a Table
To select the channels you want to view on the screen, click on the
graph window with the right mouse button, select Channels..., check
the channels you want displayed, and then click OK.
Checkboxes for channels that are not available in the present view are
dimmed.
To change units, click on the table with the right mouse button, select
Options, select the desired units for voltages and currents, and then
click OK.
To change the time units, click on the waveform window with the right
mouse button, select Options..., click on the Axes tab, select the
desired units, and then click OK.
To change the colors or line styles, click on the graph window with the
right mouse button, select Options..., click on the Display tab, adjust
colors and styles, and then click OK.
Click on the graph window with the right mouse button, select
Options..., and click on the Display tab
To change the color or style of the grid lines, click on the Grid line on
the left pane, and then select the color and style for the grid. To
disable the grid, uncheck the Grid Visible box.
To change the window frame color to white, check the White Frame
box at right.
The waveform and trend windows have two blue dashed marker lines.
The left marker indicates the starting position and the right marker
indicates the end position for calculating the average and peak values.
The minimum distance between the two markers is exactly one cycle.
You can also drag both markers with the mouse, or use the right and
left arrow keys on your keyboard to change the marker position. Click
on the graph pane to allow the keyboard to get your input before using
the keyboard.
Delta Measurements
The first reference point is still frozen until you uncheck and check the
Delta button again, while the second point can be placed anywhere
within the graph line by clicking on the graph to the left or right from
the reference point.
Using a Zoom
You can use a horizontal and, for waveforms, also a vertical, zoom to
change size of your graph.
Copying a Graph
Printing a Graph
You can use filtering to find and work with a subset of events that
meet the criteria you specify.
Click on the Filter button , or click on the report window with the
right mouse button and select Filter.... Check the causes of events you
want to display, and then click OK. PAS temporary hides rows you do
not want displayed.
To change the default sorting order based on the date and time, click
on the Sortbutton , or click on the report window with the right
mouse button and select Sort..., check the desired sort order, and
then click OK.
To change the time range for your graph, click on the Time Range
button on the local toolbar, and then select the desired date and
time range.
A.2 Construction
Case enclosure: Reinforced Polycarbonate
Packaging case: Carton and Stratocell® (Polyethylene Foam) brackets
Labels: Polyester film (UL94-V0)
Measurement Specifications
If - fundamental current
1
@ 80% to 120% of voltage FS, 1% to 200% of current FS at (In = 5 and 1 A), 2% to 200% of
current FS 50 A and frequency 50/60 Hz
NOTES:
Accuracy is expressed as (percentage of reading + percentage of full
scale) 1 digit. This does not include inaccuracies introduced by the
user's potential and current transformers. Accuracy calculated at
1second average.
Specifications assume: voltage and current waveforms with THD 5%
for kvar, kVA and PF, and reference operating temperature 20C - 26C.
Measurement error is typically less than the maximum error indicated.
Designation Description
Demands E, EH
1In4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-to-
neutral; for any other wiring mode, they will be line-to-line voltages.
Designation Description
LO I3 AVR Low I3 current
1-Sec Values on any Phase
HI VOLT AVR 1 High voltage
LO VOLT AVR 1 Low voltage
HI AMPS AVR High current
LO AMPS AVR Low current
1-Sec Total Values
HI kW IMP AVR High total kW import
HI kW EXP AVR High total kW export
HI kvar IMP AVR High total kvar import
HI kvar EXP AVR High total kvar export
HI kVA AVR High total kVA
HI PF LAG AVR Low total PF Lag
HI PF LEAD AVR Low total PF Lead
1-Sec Auxiliary Values
HI In AVR High neutral current
HI FREQ RT High frequency
LO FREQ RT Low frequency
Demands
HI V1/12 DMD 1 High V1/V12 Volt demand
HI V2/23 DMD 1 High V2/V23 Volt demand
HI V3/31 DMD 1 High V3/V31 Volt demand
HI I1 DMD High I1 Ampere demand
HI I2 DMD High I2 Ampere demand
HI I3 DMD High I3 Ampere demand
HI kW IMP BD High block kW import demand
HI kvar IMP BD High block kvar import demand
HI kVA BD High block kVA demand
HI kW IMP SD High sliding window kW import demand
HI kvar IMPSD High sliding window kvar import demand
HI kVA SD High sliding window kVA demand
HI kW IMP ACD High accumulated kW import demand
HI kvar IMP ACD High accumulated kvar import demand
HI kVA ACD High accumulated kVA demand
HI kW IMP PRD High predicted kW import demand
HI kvar IMP PRD High predicted kvar import demand
HI kVA PRD High predicted kVA demand
Time and Date Parameters
DAY OF WEEK Day of week
YEAR Year
MONTH Month
DAY OF MONTH Day of month
HOURS Hours
MINUTES Minutes
SECONDS Seconds
MINUTE INTERVAL Minute interval: 1-5, 10, 15, 20, 30, 60 min
1In
4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-
to-neutral; for any other wiring mode, they will be line-to-line
voltages.
Designation Description
I2 TDD I2 Current TDD
I3 TDD I3 Current TDD
V12 V12 Voltage
V23 V23 Voltage
V31 V31 Voltage
RT TOTAL 1-Cycle Total Values
kW Total kW
kvar Total kvar
kVA Total kVA
PF Total PF
PF LAG Total PF lag
PF LEAD Total PF lead
kW IMP Total kW import
kW EXP Total kW export
kvar IMP Total kvar import
kvar EXP Total kvar export
V AVG 3-phase average L-N/L-L voltage
V LL AVG 3-phase average L-L voltage
I AVG 3-phase average current
RT AUX 1-Cycle Auxiliary Values
In In (neutral) Current
FREQ Frequency
V UNB% Voltage unbalance 2
I UNB% Current unbalance 2
AVR PHASE 1-Second Phase Values
V1 V1/V12 Voltage
V2 V2/V23 Voltage
V3 V3/V31 Voltage
I1 I1 Current
I2 I2 Current
I3 I3 Current
kW L1 kW L1
kW L2 kW L2
kW L3 kW L3
kvar L1 kvar L1
kvar L2 kvar L2
kvar L3 kvar L3
kVA L1 kVA L1
kVA L2 kVA L2
kVA L3 kVA L3
PF L1 Power factor L1
PF L2 Power factor L2
PF L3 Power factor L3
V1 THD V1/V12 Voltage THD 1
V2 THD V2/V23 Voltage THD 1
V3 THD V3/V31 Voltage THD 1
I1 THD I1 Current THD
I2 THD I2 Current THD
I3 THD I3 Current THD
I1 KF I1 K-Factor
Designation Description
I2 KF I2 K-Factor
I3 KF I3 K-Factor
I1 TDD I1 Current TDD
I2 TDD I2 Current TDD
I3 TDD I3 Current TDD
V12 V12 Voltage
V23 V23 Voltage
V31 V31 Voltage
AVR TOTAL 1-Second Total Values
kW Total kW
kvar Total kvar
kVA Total kVA
PF Total PF
PF LAG Total PF lag
PF LEAD Total PF lead
kW IMP Total kW import
kW EXP Total kW export
kvar IMP Total kvar import
kvar EXP Total kvar export
V AVG 3-phase average L-N/L-L voltage 1
V LL AVG 3-phase average L-L voltage
I AVG 3-phase average current
AVR AUX 1-Second Auxiliary Values
In In (neutral) Current
FREQ Frequency
V UNB% Voltage unbalance 2
I UNB% Current unbalance 2
PHASORS Phasors
V1 Mag V1/V12 Voltage magnitude 1
V2 Mag V2/V23 Voltage magnitude 1
V3 Mag V3/V31 Voltage magnitude 1
I1 Mag I1 Current magnitude
I2 Mag I2 Current magnitude
I3 Mag I3 Current magnitude
V1 Ang V1/V12 Voltage angle 1
V2 Ang V2/V23 Voltage angle 1
V3 Ang V3/V31 Voltage angle 1
I1 Ang I1 Current angle
I2 Ang I2 Current angle
I3 Ang I3 Current angle
DEMANDS Present Demands (Power Demands E, EH)
V1 DMD V1/V12 Volt demand 1
V2 DMD V2/V23 Volt demand 1
V3 DMD V3/V31 Volt demand 1
I1 DMD I1 Ampere demand
I2 DMD I2 Ampere demand
I3 DMD I3 Ampere demand
kW IMP BD kW import block demand
kvar IMP BD kvar import block demand
kVA BD kVA block demand
Designation Description
kW IMP SD kW import sliding window demand
kvar IMPSD kvar import sliding window demand
kVA SD kVA sliding window demand
kW IMP ACD kW import accumulated demand
kvar IMP ACD kvar import accumulated demand
kVA ACD kVA accumulated demand
kW IMP PRD kW import predicted sliding window demand
kvar IMP PRD kvar import predicted sliding window demand
kVA PRD kVA predicted sliding window demand
PF IMP@kVA DMD PF (import) at Maximum kVA sliding window demand
kW EXP BD kW export block demand
kvar EXP BD kvar export block demand
kW EXP SD kW export sliding window demand
kvar EXPSD kvar export sliding window demand
kW EXP ACD kW export accumulated demand
kvar EXP ACD kvar export accumulated demand
kW EXP PRD kW export predicted sliding window demand
kvar EXP PRD kvar export predicted sliding window demand
In DMD In (neutral) current demand
SUMM ACC DMD Billing Summary (Total) Accumulated Demands
REG1 ACD Register #1 accumulated demand
REG2 ACD Register #2 accumulated demand
REG3 ACD Register #3 accumulated demand
REG4 ACD Register #4 accumulated demand
SUMM BLK DMD Billing Summary (Total) Block Demands
REG1 BD Register #1 block demand
REG2 BD Register #2 block demand
REG3 BD Register #3 block demand
REG4 BD Register #4 block demand
SUMM SW DMD Billing Summary (Total) Sliding Demands
REG1 SD Register #1 sliding demand
REG2 SD Register #2 sliding demand
REG3 SD Register #3 sliding demand
REG4 SD Register #4 sliding demand
ENERGY Total Energy
kWh IMPORT kWh import
kWh EXPORT kWh export
kvarh IMPORT kvarh import
kvarh EXPORT kvarh export
kVAh TOTAL kVAh total
SUMMARY REGS Billing Summary (Total and single phase) Energy
Registers
SUM REG1 Summary energy register #1
SUM REG2 Summary energy register #2
SUM REG3 Summary energy register #3
SUM REG4 Summary energy register #4
PHASE ENERGY Phase Energy
kWh IMP L1 kWh import L1
kWh IMP L2 kWh import L2
kWh IMP L3 kWh import L3
kvarh IMP L1 kvarh import L1
Designation Description
kvarh IMP L2 kvarh import L2
kvarh IMP L3 kvarh import L3
kVAh L1 kVAh total L1
kVAh L2 kVAh total L2
kVAh L3 kVAh total L3
%HD V1 V1/V12 Harmonic Distortions EH1
V1 %HD01 H01 Harmonic distortion
V1 %HD02 H02 Harmonic distortion
… ...
V1 %HD40 H40 Harmonic distortion
%HD V2 V2/V23 Harmonic Distortions EH1
V2 %HD01 H01 Harmonic distortion
V2 %HD02 H02 Harmonic distortion
… ...
V2 %HD40 H40 Harmonic distortion
%HD V3 V3/V31 Harmonic Distortions EH1
V3 %HD01 H01 Harmonic distortion
V3 %HD02 H02 Harmonic distortion
… ...
V3 %HD40 H40 Harmonic distortion
%HD I1 I1 Harmonic Distortions EH
I1 %HD01 H01 Harmonic distortion
I1 %HD02 H02 Harmonic distortion
… ...
I1 %HD40 H40 Harmonic distortion
%HD I2 I2 Harmonic Distortions EH
I2 %HD01 H01 Harmonic distortion
I2 %HD02 H02 Harmonic distortion
… ...
I2 %HD40 H40 Harmonic distortion
%HD I3 I3 Harmonic Distortions EH
I3 %HD01 H01 Harmonic distortion
I3 %HD02 H02 Harmonic distortion
… ...
I3 %HD40 H40 Harmonic distortion
ANG V1 V1/V12 Harmonic Angles EH1
V1 H01 ANG H01 Harmonic angle
V1 H02 ANG H02 Harmonic angle
… ...
V1 H40 ANG H40 Harmonic angle
ANG V2 V2/V23 Harmonic Angles EH1
V2 H01 ANG H01 Harmonic angle
V2 H02 ANG H02 Harmonic angle
… ...
V2 H40 ANG H40 Harmonic angle
ANG V3 V3/V31 Harmonic Angles EH1
V3 H01 ANG H01 Harmonic angle
V3 H02 ANG H02 Harmonic angle
… ...
V3 H40 ANG H40 Harmonic angle
Designation Description
ANG I1 I1 Harmonic Angles EH
I1 H01 ANG H01 Harmonic angle
I1 H02 ANG H02 Harmonic angle
… ...
I1 H40 ANG H40 Harmonic angle
ANG I2 I2 Harmonic Angles EH
I2 H01 ANG H01 Harmonic angle
I2 H02 ANG H02 Harmonic angle
… ...
I2 H40 ANG H40 Harmonic angle
ANG I3 I3 Harmonic Angles EH
I3 H01 ANG H01 Harmonic angle
I3 H02 ANG H02 Harmonic angle
… ...
I3 H40 ANG H40 Harmonic angle
H1 PHASE Fundamental (H01) Phase Values
V1 H01 V1/V12 Voltage 1
V2 H01 V2/V23 Voltage 1
V3 H01 V3/V31 Voltage 1
I1 H01 I1 Current
I2 H01 I2 Current
I3 H01 I3 Current
kW L1 H01 kW L1
kW L2 H01 kW L2
kW L3 H01 kW L3
kvar L1 H01 kvar L1
kvar L2 H01 kvar L2
kvar L3 H01 kvar L3
kVA L1 H01 kVA L1
kVA L2 H01 kVA L2
kVA L3 H01 kVA L3
PF L1 H01 Power factor L1
PF L2 H01 Power factor L2
PF L3 H01 Power factor L3
HRM TOT POW Fundamental and Harmonic Total Power Values
kW H01 Total fundamental kW
kvar H01 Total fundamental kvar
kVA H01 Total fundamental kVA
PF H01 Total fundamental PF
MIN PHASE Minimum 1-Cycle Phase Values
V1 MIN V1/V12 Voltage 1
V2 MIN V2/V23 Voltage 1
V3 MIN V3/V31 Voltage 1
I1 MIN I1 Current
I2 MIN I2 Current
I3 MIN I3 Current
MIN TOTAL Minimum 1-Cycle Total Values
kW MIN Total kW
kvar MIN Total kvar
kVA MIN Total kVA
Designation Description
PF MIN Total PF
MIN AUX Minimum 1-Cycle Auxiliary Values
In MIN In Current
FREQ MIN Frequency
MAX PHASE Maximum 1-Cycle Phase Values
V1 MAX V1/V12 Voltage 1
V2 MAX V2/V23 Voltage 1
V3 MAX V3/V31 Voltage 1
I1 MAX I1 Current
I2 MAX I2 Current
I3 MAX I3 Current
MAX TOTAL Maximum 1-Cycle Total Values
kW MAX Total kW
kvar MAX Total kvar
kVA MAX Total kVA
PF MAX Total PF
MAX AUX Maximum 1-Cycle Auxiliary Values
In MAX In Current
FREQ MAX Frequency
MAX DMD Maximum Demands (Power Demands E, EH)
V1 DMD MAX V1/V12 Maximum volt demand 1
V2 DMD MAX V2/V23 Maximum volt demand 1
V3 DMD MAX V3/V31 Maximum volt demand 1
I1 DMD MAX I1 Maximum ampere demand
I2 DMD MAX I2 Maximum ampere demand
I3 DMD MAX I3 Maximum ampere demand
kW IMP SD MAX Maximum kW import sliding window demand
kW EXP SD MAX Maximum kvar import sliding window demand
kvar IMP SD MAX Maximum kW export sliding window demand
kvar EXP SD MAX Maximum kvar export sliding window demand
kVA SD MAX Maximum kVA sliding window demand
In DMD MAX In (neutral) current maximum demand
MAX SUMMARY DMD Billing Summary (Total) Maximum DemandsE, EH
REG1 MD Summary register #1 maximum demand
REG2 MD Summary register #2 maximum demand
REG3 MD Summary register #3 maximum demand
REG4 MD Summary register #4 maximum demand
AO RAW Raw Analog Outputs (A/D Units)
AO1 Analog output AO1
AO2 Analog output AO2
AO3 Analog output AO3
AO4 Analog output AO4
TOU PRMS TOU Parameters E, EH
ACTIVE TARIFF Active TOU tariff
ACTIVE PROFILE Active TOU profile
TOU REG1 Billing TOU Energy Register #1 E, EH
REG1 TRF1 Tariff #1 register
REG1 TRF2 Tariff #2 register
… …
REG1 TRF8 Tariff #8 register
Designation Description
TOU REG2 Billing TOU Energy Register #2 E, EH
REG2 TRF1 Tariff #1 register
REG2 TRF2 Tariff #2 register
… …
REG2 TRF8 Tariff #8 register
TOU REG3 Billing TOU Energy Register #3 E, EH
REG3 TRF1 Tariff #1 register
REG3 TRF2 Tariff #2 register
… …
REG3 TRF8 Tariff #8 register
TOU REG4 Billing TOU Energy Register #4 E, EH
REG4 TRF1 Tariff #1 register
REG4 TRF2 Tariff #2 register
… …
REG4 TRF8 Tariff #8 register
TOU MAX DMD REG1 Billing TOU Maximum Demand Register #1 E, EH
REG1 TRF1 MD Tariff #1 maximum demand
REG1 TRF2 MD Tariff #2 maximum demand
… …
REG1 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG2 Billing TOU Maximum Demand Register #2 E, EH
REG2 TRF1 MD Tariff #1 maximum demand
REG2 TRF2 MD Tariff #2 maximum demand
… …
REG2 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG3 Billing TOU Maximum Demand Register #3 E, EH
REG3 TRF1 MD Tariff #1 maximum demand
REG3 TRF2 MD Tariff #2 maximum demand
… …
REG3 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG4 Billing TOU Maximum Demand Register #4 E, EH
REG4 TRF1 MD Tariff #1 maximum demand
REG4 TRF2 MD Tariff #2 maximum demand
… …
REG4 TRF8 MD Tariff #8 maximum demand
1In 4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-
to neutral; for any other wiring mode, they will be line-to-line voltages.
NOTE
PAS does not allow to store data in files using different data names. If
you have a file uploaded with a previous version of PAS using long data
names, either continue using long data names, or store data in a new
file.
See table below for a list of parameters with short and long names.
Short Data Long Data Name Description
Name
kW IMP ACD kW IMP ACC DMD Accumulated demand
kW IMP PRD kW IMP PRD DMD Predicted sliding window demand
PF IMP@kVA MD PF IMP@kVA MXDMD PF (import) at maximum kVA demand
REG1 ACD SUM REG1 ACC DMD Billing summary (total) register accumulated demand
REG1 BD SUM REG1 BLK DMD Billing summary (total) register block demand
REG1 SD SUM REG1 SW DMD Billing summary (total) register sliding demand
REG1 SUM REG1 Billing summary (total) energy register
REG1 MD SUM REG1 DMD MAX Billing summary (total) register maximum demand
REG1 TRF1 TOU REG1 TRF1 Billing tariff energy register
REG1 TRF1 MD DMD1 TRF1 MAX Billing tariff register maximum demand
TRF1 SEASON TRF1 Generic billing tariff energy register
TRF1 MD SEASON TRF1 Generic billing tariff register maximum demand
The second column shows data abbreviations used in the PAS data log
reports. Data log file sections are highlighted in bold.
Table 28: Daily Billing/TOU Profile Data Log (Data Log #16)