MYPIA Weight Indicator

1. This document provides instructions for parameter setting on an LM Series Weighing Controller. 2. The parameter setting menu can be accessed by pressing and holding the SET key. This allows modification of alarm settings, output controls, communication settings and more. 3. Alarm settings include ranges from -1999 to 9999 for alarms 1-3, and output control settings for high/low thresholds. 4. Other parameters include weight adjustment calibration, hysteresis values, and communication settings like ID address.

Uploaded by

Grey Mouse
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
304 views2 pages

MYPIA Weight Indicator

1. This document provides instructions for parameter setting on an LM Series Weighing Controller. 2. The parameter setting menu can be accessed by pressing and holding the SET key. This allows modification of alarm settings, output controls, communication settings and more. 3. Alarm settings include ranges from -1999 to 9999 for alarms 1-3, and output control settings for high/low thresholds. 4. Other parameters include weight adjustment calibration, hysteresis values, and communication settings like ID address.

Uploaded by

Grey Mouse
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

MYPIN Parameter setting

Ordering code Operating setting press SET Key until to enter the menu
LM Series Weighing Controller MODEL Functions ↓
Instruction Manual LM LM Series
1 Parameter setting: Alarm 1 setting, range -1999- 9999
Thanks for selecting MYPIN product! 4 48H * 48W *80L
A press and hold SET key until to enter Alarm 1 output controll
Before operating this instrument, please carefully read this manual and SIZE 8 48H * 96W *80L
setting menu. ↓ SET H:High output weight > AL1,AL1 active,
fully understand its contents. If any problems, please contact our sales or 9 96H * 96W *80L
distributors whom you buy from. This manual is subject to change without B Press << /Rst key once, LED flashing weight < AL1-HY1,AL1 inactive
Blank : 220V / 110V AC L: Low output weight < AL1 ,AL1 active
prior notice. Power Press << /Rst to shift LED if need ↓ SET
E 18-30V or 24VDC weighr >AL1+HY1,AL1 inactive
C press Up or Down key to modify data,
Application blank:none
D press SET key to confirm data. Alarm 2 setting, range -1999- 9999
Transmit DC 0-10V
1 Weight measurement and control V E Press SET key to read other parameters, Alarm 2 output controll
output I DC 4-20mA and repeat B,C,D steps to modify data if need
↓ SET H:High output weight > AL2,AL2 active,
*Mixed Batching control R AL3:alarm, no transmit F .press and hold SET key until to quit menu weight < AL2-HY1,AL2 inactive
2 Direct Load :1-4 PCS loadcells, 24 Bits A / D N N: none L: Low output weight < AL2 ,AL2 active
2.Weight Adjustment: ↓SET weight > AL2+HY1,AL2 inactive
3 Digital Input: Start / Stop and Clear Tare AL1 R Relay: AC250V / 3A
Alarm 3 setting(Hidden when no AL3)
4 Output::4-20mA / 0-10V , Relay / SSR / Logic output S SSR : 12 / 24V / 30mA
Alarm A when empty (no load),press DOWN key until
5 Data Peak value holding function for choice. N N: none
↓SET Alarm3 output controll (hidden when no AL3)
‘Oy’ flashing. release DOWN key ; PV LED
6 Communication : RS485 RS232,RJ45(LAN) AL2 R RELAY: AC250V/3A
displays about 0 H:High output weight > AL3,AL3 active,
S SSR,12/24V/30mA weight < AL3-HY1,AL3 inactive
MODBUS-RTU protocol. B Follow step A,placing the weight of this
L: Low output weight < AL3 ,AL3 active
Blank: none calibration, press UP key until SV LED flashing. ↓SET weight > AL3+HY1,AL3 inactive
Panel(for reference) 2 RS232 release UP key.refer parameter setting step
Communication to set SV LED data,SV LED data = the weight
Hysteresis value setting for all alarms.
4 RS485
g Range: ±50.0
AL1 5 LAN RJ45 of this calibration (Equal USP parameter), ↓ SET
PV E.g,the calibrate weight is 50Kg, communication ID address ,Range :000-250,
AL2
T D 5V / 80mA
Setting SV data=50.00, the unit is Kg,
AL3
Kg Excition source B 24V / 3 0mA Baud rate: 9.6K Bit or :19.2Kbit
Setting SV data=0.050, the unit is Ton
E Other:order C Refer Page 2 to lock adjust operation If need ↓ SET
SV SET Reserved,( Tare weight
<<

<<

Blank: load sensors


LM MYPIN Clear tare weight : or drop difference value)
Input code 1 0-30mV 3:
2 0-100mV press << / Rst key until ‘oy’ flashs to clear
↓ SET
① PV LED :weight value/ Parameter code display USP=The weight of this calibration
F Other: order tare or short external RST terminal once Range: 0-9999, the unit is determined by this
SV LED: preset /parameter value display Blank: none to clear tare
Application: ALL 350Ω-1KΩ
Peak value display (for Peaking meter) Loadcells or pressure sensors Supply loadcell sensor ↓SET PV LED Decimal point display
T
② Lamps: AL1,AL2,AL3: alarm , g,kg,T: units with the meter 4.The instrument will auto-return to the measuring int :0000 1 -decimal point :000.0
estate without any operation for 25 seconds 2-decimal point :00.00 3-decimal point :0.000,
③ Operation key
E.g LM8-IRRD Weight indicator /controller ↓SET
Note: Zero display (option,refer page 2) Digital filter,range:0-20,factory : sft =001
SET Select OR Confirm key Size: 48H* 96W*80Lmm
when empty or no load Larger the data, more stable is the weight display
<</rst Shift OR Clear Tare key power supply: 220 / 110V AC if need PV window display 0 when no load, then
Up key Down key ↓ SET
Transmit output: DC 4-20mA , set PF2 >= |PV| , Transmit output low value:
>>

>>

Alarm: 2-Relay High or Low output, E.g when no load,PV display range: -0.2--+0.2, E.g weight display range 0.0 -200.0 for output
4-20mA (or 0-10V), set trL=0.0
Specifications input: load cells
set PF2=0.3 > |PV|=0.2, PV will display 0.0,
↓ SET
but if PV > PF2, E.g PV value =0.31> 0.2, Transmit output high value:
Power 220V/110V AC 90-260V AC/DC or 24V AC/DC PV LED will normal display 0.31. E.g weright display range 0.0 -200.0 for output
Consumpsion ≤5VA Mounting and Sizes 4-20mA (or 0-10V), set trH=200.0
Accuracy 0.1%F.S±2digit
↓SET
Sampling speed ≤16times/sec 1 LM4 W48 X H48 X L80mm mounting W45 X H45mm weight unit: : g : Kg : Ton
Relay: Normal Open AC 250V 3A or DC 30V 5A 2 LM8 W48 X H96 X L80mm mounting W45 X H92mm
output
Resistive load ;SSR:12V/24V,MAX30mA
3 LM9 W96 X H96 X L80mm mounting W92 X H92mm ↓ SET Parameter lock, range:0-255
Input 1-4 PCS Loadcells Others:mA / V (special order) 4 LM10 W160X H80 X L70mm
LCK=010,is for read only.
Tramsmit output DC 4-20mA / 0-10V etc. mounting: W152X H76mm
LCK=000 Read and Write.
Communication RS232 / RS485 RJ45 MODBUS RTU protocol
press SET Key until to quit the menu or return AL1
Connections
(If any changed, refer the label on the meter).
load cells
Calibration Lock / Unlock LM MODBUS USER INSTRUCTIONS
RST/TARE
press UP and Down key until to enter menu
AL2 AL1 220V AC
L N Baud Rate 1 The instrument RS485/RS232 MODBUS-RTU data format


com


9.6K Bit/S OR 19.2K Bit/S
+ 10
11

2
12

→ 1
Start bit Data bit Stop bit Parity Baud rate

→3
4
5
6
7
9
8 factory:9.6K Bit/S

110V AC
EXC

↓ Prt=No,Unlock calibration,
SG

SET
LM8-RRD 1 8 1 None 9.6k or 19.2kBitS
Prt=YES,lock calibration,Calibration

load cells
opretion not allowed (refer PAGE 1 2 The format of the data reading and writing is same as standard Modbus protocol.
↓ SET weight adjustment )
RST/TARE AL2 AL1
Definition as follows:
220V AC
PV LED zero value setting. Request:(eg.TX: 01 03 00 62 00 02 65 D5)


L N


com
If |PV| < PF2,the PV disiplay is 0
+ 10
11

2
12

→ 1
→3
4

↓SET
5
6
7
9
8

01 03 0098(0062H) 0002 26069 (65D5)


110V AC
EXC
SG

RS232 RS485
Reserved weight=Ct*USP-PVF
ADD COM PV Counts CRC
19
→1 8
17

17
→1 8


LM8-RR4D/RR2D SET
GND RXD TXD B- A+ Response: (eg.RX: 01 03 04 6D 96 49 F3 71 66)
Calibration password for factory
load cells 01 03 04 6D96 49F3 7166
RST/TARE AL2 AL1 220V AC press UP and Down key until to quit menu ADD COM Count PV CRC
L N


com
+ 10
11

2
12

→ 1
→3
4
5
6
7
9
8

Return data : 2 word,PV= 6D96 49F3=6D96.49F3H = INT 6D96H+ POINT 49F3H=28054+18931/65536=28054.2888


110V AC
EXC
SG

(49F3H=18931 6D96H=28054)When Max bit is “1”,means negative, E.g PV=ED96 49F3 = - 6D96.49F3H= -28054.2888
→1 8
17

LM8-IRRD/VRRD - + 3, Can read OR write 1 parameter only every time,when writing ,Please convert the demical data to hexadecimal data,
OUTPUT 4-20mA OR 0-10V
.For example,100.5,INT 100=0064H,0.5=0.8000H,therefore 100.5=0064.8000H,100.5 data=00 64 80 00
Complete products contains 4 Clear Tare: Parameter PVF(Address 10H), TX: 01 06 00 10 00 00 88 0F, 01:ID number,,06:COM ,ingnore writing data
★1 copy of user manual, 5, Commands: 03H: read holding registers parameters 06H: write single holding register parameter value
  ★1 inspection QC label,
  ★2 installing brachets,
10H: write multi holding registers parameters value
We are responsible for the overal repairment for the failure of
manufactorying quality within 12 months since the date of purchase. 5, Communication parameters:LM Weighing meter reading and writing parameter
More information:PLS Down load from : www.mypinchina.com Parameters data address count data address count
(HEX) (w o r d s )
functions Parameters (w o r d s ) functions
IP/wechat: 86-18689341985, Email:[email protected] (HEX)
PV 62H 2 weight value,read only Add 0FH 1 ID address,0-255,R/W
Main Products AL1 00H 2 AL1Value, R/W USP 14H 2 the weight of calibration, R/W
Counter & length meter AM1 03H 1 AL1 control, 0:L,1:H, R/W dp 17H 1 PV decimal point, R/W
Time relay
Temperature controller AL2 04H 2 AL2Value, R/W SFt 18H 1 software filtering Value,R/W
Panel meter
Frequency/Tacho/Line-speed meter AM2 07H 1 AL2 control, 0:L,1:H, R/W TRL 1CH 2 Transmit low value, R/W
Proximity sensor Photo-electrical sensor
Ampere & Voltage Power meter AL3 08H 2 AL3Value, R/W TRH 20H 2 Transmit high value, R/W
Transmitter
Universal Sensor indicator
AM3 0BH 1 AL3 control, 0:L,1:H, R/W Unt 23H 1 weight unit, 0: g, 1: Kg, 2: T, R/W
Power adjustor
HY1 0CH 2 AL1 AL2 AL3 control hysteresis, R/W LCK 24H 1 Parameter lock, R/W