Service Report Mednif 100ma Biomedilab

The service report summarizes a field service call to address uneven imaging results from a biomedical lab's medical imaging unit. An engineer inspected the unit and found the issue was caused by an improperly adjusted resistor frequency. The engineer re-adjusted the resistor frequency which resolved the uneven results. The report recommends the unit undergo scheduled maintenance to prevent future issues.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
63 views2 pages

Service Report Mednif 100ma Biomedilab

The service report summarizes a field service call to address uneven imaging results from a biomedical lab's medical imaging unit. An engineer inspected the unit and found the issue was caused by an improperly adjusted resistor frequency. The engineer re-adjusted the resistor frequency which resolved the uneven results. The report recommends the unit undergo scheduled maintenance to prevent future issues.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Service Request Type: Service Request Number:

Service Report Corrective Repair 2022-12-31/103

Customer Details

Customer Name: Bio Medical Lab Location: Contact Cikupa-Indonesia


PO #: Mr.Sapardi Phone: Email -
Contact Nome: Address:
Initial Problem Description:

Equipment Details

System ID: Initial Equipment Status: Results Are Not good


Serial Number: Entitlement: -
Product Description:

Activity Description
Activity #: Activity Close Dote:
Activity Type: Field Support Engineer Name: Sukrisdianto
Tube Slice Reading: Mednif
Problem Description: Uneven Imaging Results

Couse/ Diagnosis: Desember 31-2022 :


- Uneven Imaging Results

Resolution/ Service Performed: Desember -2022 :


- Check and inspect condition Unit OK
- Check and inspect electricity from LAB OK
- Turn On Unit and test
- Test exspose and ON media screen OK
- Check result through CR Uneven Image results
- Turn Off Unit and all check
- Check Condition tube assembly OK
- Check and inspect resistor from cabinet console
- Re Adjustment resistor frequency OK
- Check frequency 50Mhz from resistor Ok
- Turn on unit
- Kv Set and test exsposure on media
- Check result through CR OK
- Test tube in standing condition results Not good
Recommendotion :
“ Maintenance on Schedule”

Page No. 1 of 1
Service Request Type: Corrective Repair
Service Report Request Number: 221004-11

Date Time Hours Labour Type


Des 12, 2022 11:00-17:00 Onsite and
repair

Ports Used Details

Port Number 0tY Description Asset Serial No

Expense Tracker

Expense Dote Type 0tY Comments

Engineer Signoture: Customer Signoture:


31-Des-2022 11:00 Mr.Sapardi
SUKRISDIANTO

Page No. 2 of 2