Atomic Absorption Spectrometry: How It Works
Atomic Absorption Spectrometry: How It Works
M° M* M*
+ + + +
- Ne° Ne+ - - M° Ne+ - M° Light
Ne+
Figure 2
1
e.
Sample Cell
Figure 3
A typical atomic absorption instrument holds several constant monitoring between the reference beam and the light
lamps each for a different element. The lamps are housed source. To ensure that the spectrum does not suffer from loss
in a rotating turret so that the correct lamp can be quickly of sensitivity, the beam splitter is designed so that as high a
selected. proportion as possible of the energy of the lamp beam passes
through the sample.
The optical system and detector
A monochromator is used to select the specific wavelength
of light – ie spectral line – which is absorbed by the Atomisation of the sample
sample, and to exclude other wavelengths. The selection of Two systems are commonly used to produce atoms from the
the specific light allows the determination of the selected sample. Aspiration involves sucking a solution of the sample
element in the presence of others. The light selected by the into a flame; and electrothermal atomisation is where a drop
of sample is placed into a graphite tube that is then heated
monochromator is directed onto a detector that is typically electrically.
a photomultiplier tube. This produces an electrical signal
proportional to the light intensity.(Fig. 3) Some instruments have both atomisation systems but share
one set of lamps. Once the appropriate lamp has been selected,
Double beam spectrometers it is pointed towards one or other atomisation system.
Modern spectrometers incorporate a beam splitter so that
one part of the beam passes through the sample cell and the
other is the reference (Fig. 4). The intensity of the light Flame aspiration
source may not stay constant during an analysis. If only a Figure 5 shows a typical burner and spray chamber. Ethyne/
single beam is used to pass through the atom cell, a blank air (giving a flame with a temperature of 2200–2400 °C) or
reading containing no analyte (substance to be analysed) ethyne/dinitrogen oxide (2600–2800 °C) are often used. A
would have to be taken first, setting the absorbance at zero. flexible capillary tube connects the solution to the nebuliser.
If the intensity of the source changes by the time the sample At the tip of the capillary, the solution is ‘nebulised’ – ie
is put in place, the measurement will be inaccurate. In the broken into small drops. The larger drops fall out and drain
double beam instrument there is a off while smaller ones vaporise in the flame. Only ca 1% of
the sample is nebulised.
Reference beam
Sample beam
Figure 4
2
e.
Sample hole
Flow spoiler
Mixing chamber
with burner head
Nebuliser
Impact bead Light
End cap
Figure 5 Figure 6
Electrothermal atomisation
of the original sample. Figure 7 shows a flame
Figure 6 shows a hollow graphite tube with a
atomic absorption spectrometer with an autosampler
platform. 25 µl of sample (ca 1/100th of a raindrop) is
and flow injection accessory.
placed through the sample hole and onto the platform
When making reference solutions of the element
from an automated micropipette and sample changer.
under analysis, for calibration, the chemical
The tube is heated electrically by passing a current
environment of the sample should be matched as
through it in a pre-programmed series of steps. The
closely as possible – ie the analyte should be in the
details will vary with the sample but typically they
same compound and the same solvent. Teflon
might be 30–40 seconds at 150 °C to evaporate the
containers may be used when analysing very dilute
solvent, 30 seconds at 600 °C to drive off any volatile
solutions because elements such as lead are sometimes
organic material and char the sample to ash, and with a
leached out of glass vessels and can affect the results.
very fast heating rate (ca 1500 °C s-1) to 2000–2500 °C
for 5–10 seconds to vaporise and atomise elements
(including the element being analysed). Finally Background absorption
heating the tube to a still higher temperature It is possible that other atoms or molecules apart from
those of the element being determined will absorb or
– ca 2700 °C – cleans it ready for the next sample.
scatter some radiation from the light source. These
During this heating cycle the graphite tube is flushed
species could include unvaporised solvent droplets, or
with argon gas to prevent the tube burning away. In
electrothermal atomisation almost 100% of the compounds of the matrix (chemical species, such as
sample is atomised. This makes the technique much anions, that tend to accompany the metals being
more sensitive than flame AAS. analysed) that are not removed completely. This
means that there is a background absorption as well as
Sample preparation that of the sample.
Sample preparation is often simple, and the chemical One way of measuring and correcting this
form of the element is usually unimportant. This is background absorption is to use two light sources, one
because atomisation converts the sample into free of which is the hollow cathode lamp appropriate to the
atoms irrespective of its initial state. The sample is element being measured. The second light source is a
weighed and made into a solution by suitable dilution. deuterium lamp.
Elements in biological fluids such as urine and blood The deuterium lamp produces broad band
are often measured simply after a dilution radiation, not specific spectral lines as with a hollow
Figure 7
3
e.
cathode lamp. By alternating the measurements of the two light sources generally at 50 –100 Hz the total
absorption (absorption due to analyte atoms plus background) is measured with the specific light from the hollow
cathode lamp and the background absorption is measured with the light from the deuterium lamp. Subtracting the
background from the total absorption gives the absorption arising from only analyte atoms.
Calibration
A calibration curve is used to determine the unknown concentration of an element eg lead in a solution.
The instrument is calibrated using several solutions of known concentrations. A calibration curve is produced
which is continually rescaled as more concentrated solutions are used the more concentrated solutions absorb
more radiation up to a certain absorbance. The calibration curve shows the concentration against the amount of
radiation absorbed (Fig. 8(a)).
The sample solution is fed into the instrument and the unknown concentration of the element eg lead is then
displayed on the calibration curve
(Fig. 8(b)).
Absorbance
Absorbance
Concentration Concentration