Users Manual EPVLF-110V60Hz
Users Manual EPVLF-110V60Hz
EPVLF
USER’S GUIDE
Cautions
accident and any electrical shock, do not perform any service beyond
off the tester, and cut off the power supply of whole testing units.
⚫ Please connecting the wire according the wiring diagram strictly, and
wire to the ground reliably. And please cordon off the test area to make
sure the safety and prevent unqualified person entrance, and avoid
accident occurred.
environment. Keep the surface clean and dry, ventilative area free from
corrosive gas.
⚫ Connect cables correctly. The order is to connect the device with ground
⚫ Test operator should keep safe distance from high voltage electrified
apparatus. Don't touch exposed contacts and apparatus when test item
electrified.
⚫ Don’t touch the metal of the test leads to avoid electric shock.
⚫ Please turn off the power after the testing is finished or the user left the
test area.
EPVLF 2
⚫ Please make sure the equipment is upright before opening. Don't drop
⚫ Please confirm the replaced fuse is same type and rated as the old one.
Warranty
The warranty period for this series is one year from the date of shipment,
warranty dates. EP Hipot Electric warrants to the original purchaser that this
product will be free from defects in material and workmanship under normal
use. Throughout the warranty period, provide that such defects are not
Packing List
1 Controller 1
EPVLF 3
2 HV Booster 1
3 HV Cable 1
4 LV Cable 1
5 Power Cable 1
6 Discharge Rod 1
7 Grounding Wire 1
8 Capacitor 1
9 Fuse 4
10 Print Paper 2
11 User’s Manual 1
12 Test Reports 1
13 Warranty Cert 1
I. Overview
The withstand voltage test can be divided into AC withstand voltage test and
DC withstand voltage test. The AC withstand voltage test can be divided into
power frequency, frequency conversion and 0.1hz ultra-low frequency test
shortcomings of similar products (see table 1), and its cost performance is
generator and motor, etc.), which conforms to the national standard of electric
power industry, general technical conditions for ultra-low frequency and high
the power frequency withstand voltage test. In the power frequency withstand
voltage test of large and medium-sized generators, motors, power cables and
other test objects, due to their large capacitance in the insulation layer, a
of these huge equipment is not only heavy and expensive, but also
inconvenient to use. The method of reducing the test frequency and the
capacity of the test power supply is widely used in the world. Replacing power
voltage test can not only have the same equivalence, but also greatly reduce
the volume and weight of the equipment, theoretically, the capacity is about
small size, light weight and low cost, the DC voltage withstand test is also the
most destructive to the insulation of the tested object. Therefore, the latest
stipulated that DC high voltage is no longer used for voltage withstand test of
2. Technical parameters
3) Load capacity:
4) Measurement accuracy: 3%
- 10 ℃ ~ + 40 ℃; humidity: ≤ 85% RH
8) Power fuse: see Table 2
Note: if the portable generator is used for power supply, the output
power is more than 3KW, the frequency is 60Hz, and the voltage is 110V)
EPVLF 7
10) The electrical capacity of the test object shall not exceed the maximum
value of the rated electrical capacity of the instrument. See Table 2 for
the value.
voltage generator.
Rated
Model Load capacity Fuse Weight
voltage
0.1Hz ≤1.1µF
0.02Hz ≤5.5µF
0.1Hz ≤1.1µF
EPVLF- 40kV controller:4kg
10A
0.05Hz ≤2.2µF
40/1.1 (peak) Booster:40kg
0.02Hz ≤5.5µF
0.1Hz ≤1.1µF
EPVLF- 50kV Controller:4kg
0.05Hz ≤2.2µF 20A
50/1.1 (peak) Booster:45kg
0.02Hz ≤5.5µF
0.02Hz ≤5.5µF
0.1Hz ≤1.1µF
EPVLF- 80kV Controller: 4 kg
0.05Hz ≤2.2µF 20A
80/1.1 (Peak) Booster:50 kg
0.02Hz ≤5.5µF
subjects
When in use, the capacitance of the test object shall not exceed the rated
capacitance of the instrument. If the capacitance of the test object is too small,
the output waveform will be affected. If it is less than 0.05 μ F, the instrument
will not output normally. At this time, 0.05 μ f capacitor (provided by EP Hipot
Electric) can be connected in parallel for auxiliary output. See Table 4 and
(when acting)
Room ≤ 10ms), the instrument is safe and reliable.
Safety and reliability: the controller is connected with the low voltage of
the high voltage generator, with photoelectric control, safe and reliable
use.
The high and low voltage closed-loop negative feedback control circuit
Large test range: 0.1Hz, 0.05Hz and 0.02hz multi frequency selection,
Small volume and light weight: it is very convenient for outdoor operation.
The instrument consists of two parts: controller and booster. The structure
1. The layout of each part of the controller panel is shown in Figure 1, and
"Switch" - power switch: built-in indicator light, on when on, off when
"Capacitive touch screen" - display test data and output waveform, and can
Wiring instructions: connect the control output special line, high voltage
output special line and grounding wire equipped with the product according
to the method shown in Figure 3. The power cord of the power socket is
switch can be turned on. After the microcomputer is powered on or reset, the
turn off the power supply. The power socket is equipped with a safety tube. If
EPVLF 13
there is no display on the start-up screen, check whether the fuse is broken
first, and the size of the fuse shall be replaced according to the data provided
in Table 3.
First, click the "setting" button on the screen of Figure 4 to display the setting
★ There are three options for frequency: 0.1, 0.05 and 0.02 in Hz.
test, in minutes.
★ Set voltage: range from 0 to rated value, in kV. It sets the test voltage we
want to raise. When the instrument rises to this set voltage limit, it will no
longer boost voltage, and maintain constant amplitude sine wave output at
★ Set voltage limit: the setting range of voltage protection value is from 0 to
rated value, and in kV. It specifies the upper limit value of the voltage
passing the test object. When the voltage exceeds this setting, the
★ Setting current limit: the setting range of current protection value is from
0 to rated value, and in mA. It specifies the upper limit value of current
passing through the test object. When the current exceeds this setting, the
(Note: the above voltage, current and measurement data displayed by the
After pressing the "start" key in Figure 4, the instrument will carry out the
the prompt message "load not connected" in the status bar of Figure 6
2) Boosting process
After the self-test is successful, the instrument will automatically enter the
boosting state, as shown in Figure 7, and the prompt message in the status
The controller will raise the voltage to the set value in several cycles, and
the instrument will output the equal amplitude, as shown in Figure 8. The
4) Downtime
When the timing reaches the set time, the instrument will stop automatically,
as shown in Figure 9, and the prompt message in the status bar is "stop the
test".
EPVLF 17
the test object, as shown in Figure 10. The prompt message in the status bar
is "discharging".
After the shutdown, as shown in Figure 11, the prompt message in the
★ timed shutdown: when the timing reaches the set time, the instrument will
stop automatically.
In the process of test, when the output high pressure exceeds the set limit,
the instrument will automatically cut off the output after starting and shutting
down the command, and then execute the data history saving. After shutting
down, as shown in Figure 12, the prompt message in the status bar is "over
In the process of test, when the output current exceeds the set limit, the
instrument will automatically cut off the output after starting and stopping the
command, and then execute the data history saving. After the shutdown, as
shown in Figure 13, the prompt message in the status bar: "overcurrent
Click the "print" key in Figure 3 to print the data on the display into a test
report.
Under the status of viewing historical data, click the "print" key to print the
All the data instruments that have been shut down by timing, clicking the
"stop" key, over-voltage protection and over-current protection will
measured data, and the previous 64 times will be deleted automatically. Click
the "view" key in Figure 3 to open the interface in Figure 13 to view the
Click the "clock" button in Figure 3 to open the setting interface in Figure
15, which is used to set the date and time of the system.
2. Use a megger to test the insulation parameters of each phase of the cable,
and carry out the ultra-low frequency withstand voltage test only after the test
is qualified.
3. Set the test voltage value: Umax = 3uo, where uo is the rated phase
Example 1: cable parameter: the rated line voltage is 10kV, and the rated
Note: UN is the rated voltage of the cable, and uo is the phase voltage of
the cable.
EPVLF 23
4. Test time: 60 minutes for handover test and 15 minutes for preventive test.
Io=2πfCU=2×3.14×0.1CU(mA)………………… (formula 1)
C is the electric capacity of the cable to the ground, unit: UF; u is the effective
Example 2: a 10KV (UN = 10kV, uo = 8.7kv) cable is 4km long, with a single
If the low frequency test voltage is 26kv (peak value), the leakage current is
approximately:
Io=2πfCU=2×3.14×0.1CU=0.628×0.21×4×26/ 2 =9.69(mA)
I=kIo………………… (formula 2)
6. Test wiring: connect the test equipment and test cable with the special
connection attached with the machine according to the method shown in
Figure 13. Close the power supply after carefully checking the correct wiring,
set the test frequency, time and voltage again, as well as the over-current
protection value and over-voltage protection value on the high-voltage side,
monitored to see if there is any abnormal sound in the test cable. When the
test voltage is raised, the instrument will automatically record the test time
7. When the test time is up, the instrument will stop automatically. If no
EPVLF 25
destructive discharge occurs in the test, it is considered to pass the withstand
voltage test.
the voltage is unstable, and the test cable has peculiar smell, smoke,
abnormal sound or flashing, etc., stop boosting immediately, and find out the
cause after shutdown. If these phenomena are caused by the weak insulation
determined that the test cable is polluted due to air humidity or surface dirt,
9. During the test, if the insulation defect of non-test cable causes over-
1. This test can be carried out during handover, overhaul, local replacement
EPVLF 26
of winding and routine test. Using 0.1Hz ultra-low frequency to carry out the
voltage withstand test on the motor, the defects of the end insulation of the
generator are more effective than the power frequency voltage withstand test.
The reason is that under the power frequency voltage, the capacitance
current flowing from the wire rod causes a large voltage drop when it flows
through the semiconductor corona proof layer outside the insulation, so that
the voltage on the wire rod insulation at the end is reduced; while under the
the voltage drop on the semiconductor corona proof layer is also greatly
Defects found.
2. Wiring method: conduct the test in phases, pressurize the tested phase,
and short-circuit the non-tested phase. The test wiring is shown in Figure 14.
EPVLF 27
3. According to the regulations, the peak value of test voltage can be
For example: for the synchronous motor with rated voltage of 10.5kv, the
follows:
Umax= 2 ×1.2×1.5×10.5≈26(kV)
4. Test time setting: the same as the test time of power frequency withstand
smoke and unstable data display, it can be considered that the insulation has
withstood the test. In order to better understand the insulation situation, it is
necessary to monitor the surface state of the insulation, especially the air-
cooled unit. Experience shows that the appearance monitoring can find the
abnormal insulation phenomena of synchronous motor which cannot be
IX. Precautions
1. During the test, please strictly abide by the safety specifications of the high-
pressure test, and operate by professional personnel.
3. After shutdown, discharge the test object fully with discharge rod. After the
(1) Transportation
The product must be packed during transportation. Cartons can be packed
in cartons or wooden boxes, and foam proof layer should be placed in the
The warehouse shall be protected from rain, dust and mechanical damage.
(2) Storage
than 85%, ventilation and no corrosive gas. Storage should not be close to
In humid areas or wet seasons, if the instrument is not used for a long time,
it is required to turn on and power on once a month (about two hours) to make
the moisture emission and protect the components.
possible.
The operation of replacing the paper roll printer is very simple. It does not
need to take out the whole printer. Just press the open button to open the
front cover. Open the front cover as shown in Figure A and take out the
remaining paper core. Then install the new paper roll as shown in Figure B
Note: when closing the front cover, make the printing paper stretch out from
the paper outlet, put it in the middle position, and let the rubber shaft fully
press the printing paper, otherwise it can't be printed. When the thermal
printer is loading paper, it must be confirmed that the thermal coating of the
thermal paper is on it, and then put the thermal paper into the printer paper
bin. If the thermal coating is not on the correct side, no writing will be printed.
If the printing paper deviates, you can open the front cover again and adjust
the position of the printing paper.
EPVLF 31