01 GNP
01 GNP
TANGO-R Matrix-F
TANGO-T
MPA II
-Bruker Confidential- 2
Optimize all Settings to get the best
Spectra
Good Results
Good Spectra
Signal/Noise Ratio
Preamp Gain
32 16 8 128 64 32
cm-1 cm-1 cm-1 Scans Scans Scans
-Bruker Confidential- 3
Good NIR Methods are based on good
Spectra and Reference Values
• In the moment you realize that the spectra are not good, the samples are
probably gone.
-Bruker Confidential- 4
Good NIR Methods are based on good
Spectra and Reference Values
-Bruker Confidential- 5
What is the Goal for the Spectrum?
• The spectrum should represent the sample. More scans will be needed if the
sample is heterogeneous.
• For good Chemometric evaluation methods, the signal to noise (S/N) ratio
will directly affect the repeatability and accuracy of model, i.e.
• Thresholds in Ident methods
• RMSECV/RMSEP of Quant methods
(noisy spectra -> noise in PLS factors -> noise in regression coefficients)
NOTE: Using derivatives as data preprocessing can improve results but noise
will increase
-Bruker Confidential- 6
Background Measurements
-Bruker Confidential- 7
What is the Goal for the Spectrum?
Optimize the
Signal/Noise Ratio
Sample Measuring
heterogeneity speed
Sample Spectral
absorptivity characteristics
(peak width)
-Bruker Confidential- 8
Amplitude and Signal/Noise Ratio
• In the case of low peak amplitude, change the setting to lower resolution
(for example, from 16cm-1 to 32cm-1) could help improve Signal/Noise
ratio
-Bruker Confidential- 9
High Amplitude, good Signal/Noise
• Preferable not to
exceed: 25000
-Bruker Confidential- 10
How to check Signal/Noise Ratio
visually
• Check
smoothness of
single channel
spectrum on flat
area, e.g. around
10,000 to
8,000cm-1.
• NOTE: in check
signal mode
always 13cm-1
resolution is
used!
Improvement of
S/N by lowering
resolution is not
visible here, only
in measured
spectra!
-Bruker Confidential- 11
How to check Signal/Noise Ratio
visually
• Checking the
0,0002
noise level by
taking 1st
derivative of a
sample
spectrum.
Absorbance Units
-0,0004
-0,0008
-Bruker Confidential- 12
How to check Signal/Noise Ratio
visually
0,0002
is done
depending on
-0,0004
the used
resolution.
• Common
settings are
-0,0008
8cm-1 = 17pt
16cm-1 = 13pt
32cm-1 = 9pt
64cm-1 = 5pt
10500 10000 9500 9000 8500 8000
-Bruker Confidential- 13
Important Measurement Parameters
• The following parameters are most important for the quality of spectra
(i.e. Signal/Noise Ratio) and are fundamental for the robustness of
calibrations and repeatability of NIR predictions.
• Resolution
• Number of scans (measurement time)
• Preamp gain settings for sample and background
• Scanner velocity (only in process)
-Bruker Confidential- 14
Resolution
-Bruker Confidential- 15
Resolution
• The FT technology with mirror allows resolution changing which has direct
effects and consequences on model performance.
• Lower resolution requires shorter acquisition time for the same number of
scans; therefore, for the same measurement time (for example 30 sec),
double number of scans can be used for low resolution set up.
• 8cm-1, 32 scans requires 22 seconds for MPA II/Matrix-F
• 16cm-1, 64 scans also requires 22 seconds for MPA II/Matrix-F
-Bruker Confidential- 17
Spectra measured at different resolution
Log (1/R)
8cm-1, 64 scans
16cm-1, 64scans
-Bruker Confidential- 18
Spectra measured at different resolution
First derivative
8cm-1, 64 scans
16cm-1, 64scans
-Bruker Confidential- 19
Benefits of high resolution
(White bright samples!)
-Bruker Confidential- 20
Information loss in resolution
-Bruker Confidential- 21
Talc OH-Bands with different
Resolutions
0.40
-Bruker Confidential-
Recommended Resolution/Scan Settings
-Bruker Confidential- 23
Recommended Resolution/Scan Settings
-Bruker Confidential- 24
Sample Gain Settings
• Use always
sample gain x1
instead of
‘automatic’.
-Bruker Confidential- 25
IMPORTANT: Preamp Gain Settings
• Separate settings
for preamp gain for
Sample and BGR
measurements.
Different preamp
gains can be used
for Sample and
BGR.
Ref: 1x (1x)
A: 3x (2.1x)
B: 30x(8.2x)
C: 300x (15x)
The number in () is
for MPA II Int sphere
For integrating
sphere of MPA
1/Matrix-I, changing
of preamp gain is not
possible
-Bruker Confidential- 26
Preamp Gain setting for Background
(not for Int sphere of MPA 1/Matrix-I)
• Different settings might be used for other probes, long fiber, and High
Intensity MPA.
• By the preamp gain the signal is amplified directly in the detector electronic,
without amplifying the noise (noise is manly introduced by the electronic
after the detector).
• The optimal preamp gain settings for background and sample can only be
found by trial and error. For inhomogeneous and/or moving samples this
must be done during movement of the sample.
-Bruker Confidential- 27
Dark Sample, Same Resolution and
Measurement time, different Gains
Gain Ref
4
Gain A
Gain B
Gain C
3
Absorbance Units
2 1
0
-Bruker Confidential- 28
Preamp Gain
Moving Samples with proper Gain
-Bruker Confidential- 29
Preamp Gain
Moving Samples with too high Gain
-Bruker Confidential- 30
Preamp Gain
Moving Samples Spectra changes
-Bruker Confidential- 31
Saturation of Detectors
• In real live the detector can be overloaded even when the amplitude is
below 32,000. If the signal is exceeding the digital limit it can be converted
to a negative contribution leading to a lower amplitude.
• E.g. an amplitude of 28,000 can be that the real value is 36,000, but the
4,000 counts above 32,000 are flashed back and reduce the shown
amplitude.
That’s why you should always check the Single Channel Spectrum in the
Check Signal mode!
-Bruker Confidential- 32
Check signal (align mode)
• Display limits
set down to
0 cm-1!
-Bruker Confidential- 33
Check signal (align mode)
• Normal shape
of detector
signal with
sample in
place
-Bruker Confidential- 34
Saturation of detector (overloaded)
Signal overload
can be detected
by
• Strange shape
around 10,000
cm-1
• Signal goes up
from zero
beyond detector
limit (here
4,000 cm-1)
-Bruker Confidential- 35
Saturation of detector (overloaded)
Signal goes up
again below the
detector limit
after reaching
zero line (here
4,000 cm-1)
-Bruker Confidential- 36
Optical Slit & Total Absorbance
-Bruker Confidential- 37
Which Optical Path Length should I use?
-Bruker Confidential- 38
Total Absorbance at Transmission
Measurements
-Bruker Confidential- 39
Pentaerythritol and Dipentaerythritol in
Water
Signal Saturation
4.0
OBVIOUS
3.5
3.0
Can we use
this region?
Absorbance Units
2.0 2.5
1.5
1.0
0.5
-Bruker Confidential- 40
Move to SSC
0.10
0.08
Single channel
0.06
0.04
0.02
0.00
-Bruker Confidential- 41
Comparison SSC and AB Spectrum
Signal Saturation!!
4
3
Absorbance Units
2
SSC
1
AB
0
-Bruker Confidential- 42
Comparison SSC and AB Spectrum
Detail
• Peak
3.0
around
Signal Saturation!!
7.000
cm-1
2.5
cannot
be used!
2.0
Absorbance Units
1.5 1.0
SSC
AB
0.5
-Bruker Confidential- 43
Fiber Optical Cables
Cut-Off
-Bruker Confidential- 44
Long Wavelength Cutoff due to Fiber
Length
0.5
5m
10m
15m
0.4
20m
30m
50m
0.3
Single channel
100m
0.2
0.1
0.0
-Bruker Confidential- 45
Fiber Optical Cable
Cut-Off Wavenumbers
0.16
0.14
0.12
0.10
500m 150m 2m
Single channel
0.08 0.06
0.04
0.02
0.00
-Bruker Confidential- 46
Fiber Optical Cable
Cut-Off Wavenumbers
-Bruker Confidential- 47
Influence of Fiber Optical Cable
Length/Coupling
-Bruker Confidential-
40 kHz Measurements
-Bruker Confidential- 49
In-line Fermentation Spectra with
Scanner Velocities 10 and 40 kHz
10 kHz
1min
4
scan time
40 kHz
3
Absorbance Units
1min
scan time
2 1
0
-Bruker Confidential- 50
In-line Spectra of Fermentation with
different Scanner Velocities: O-H-Band
2.6
10 kHz
1min
2.5
scan time
40 kHz
2.4
Absorbance Units
1min
scan time
2.3 2.2
2.1
2.0
-Bruker Confidential- 51
Aspects of NIR Analysis of Liquids
-Bruker Confidential- 52
Aspects of NIR Analysis of Liquids
• Pathlength
• 1, 2, 5 or 10mm?
• Sample presentation and handling
• Vial, flow cell or probe?
• Sample temperature
• Temperature control required?
• Water content of sample
• Viscosity
• Influences the selection of
pathlength and sample
presentation/handling
• Opaque or cloudy liquids
• Transmission measurements
limited; Transflection as
alternative
-Bruker Confidential- 53
Aspects of NIR Analysis of Liquids
-Bruker Confidential- 54
NIR Spectrum of CHCl3
-Bruker Confidential- 55
Overtone of combination band and 2nd
Overtone of C-H vibration in CHCl3
-Bruker Confidential- 56
Ethanol NIR Spectra
AB Spectra
optical slit
1 mm
2 mm
4
5 mm
10 mm
3
Absorbance Units
2
1
0
-Bruker Confidential- 57
Ethanol NIR Spectra
Single Channel Sample Spectra
optical slit
0.14
1 mm
2 mm
5 mm
0.12
10 mm
0.10
Single channel
0.06 0.08
0.04
0.02
0.00
-Bruker Confidential- 58
Cyclohexane NIR Spectra
AB Spectra
optical slit
5
1 mm
2 mm
5 mm
10 mm
4
Absorbance Units
2
1
0 3
-Bruker Confidential- 59
Cyclohexane NIR Spectra
Single Channel Sample Spectra
optical slit
1 mm
0.25
2 mm
5 mm
10 mm
0.20
Single channel
0.15
0.10
0.05
0.00
-Bruker Confidential- 60
NIR AB Spectra of Water
measured with Transmission Probes
5
optical slit
1 mm
2 mm
5 mm
4
10 mm
3
Absorbance Units
2
1
0
-Bruker Confidential- 61
NIR Spectra of Water
Single Channel Sample Spectra
optical slit
1 mm
2 mm
0.20
5 mm
10 mm
0.15
Single channel
0.10
0.05
0.00
-Bruker Confidential- 62
NIR Spectra of Water
Single Channel Sample Spectra
optical slit
1 mm
2 mm
0.05
5 mm
10 mm
0.04
Single channel
0.03
0.02
0.01
0.00
-Bruker Confidential- 63
Small Amount of C-H in O-H
3.5
Optical Slit
2 mm
5 mm
3.0
2.5
Absorbance Units
2.0
O-H
1.5
1.0
C-H
C-H
0.5
O-H
10000 9000 8000 7000 6000 5000 4000
Wavenumber cm-1
-Bruker Confidential- 64
NIR Spectra of Water in Range of
0 to 95oC
1.4
1.2
Absorbance Units
1.0
0.8
0.6
0.4
0.2
0.0
-Bruker Confidential- 65
Toluene NIR Spectra in Range of
0 to 100oC
0.06
0.05
0.04
Absorbance Units
0.03
0.02
0.01
-0.01 0.0
-Bruker Confidential- 66
Changes of Water Spectra with
Temperature
0 °C
5 °C
10 °C
1.45
15 °C
20 °C
25 °C
30 °C
Absorbance Units
1.40
35 °C
40 °C
45 °C
50 °C
1.35
55 °C
60 °C
65 °C
70 °C
1.30
75 °C
80 °C
85 °C
90 °C
1.25
95 °C
7100 7000 6900 6800 6700
100 °C
Wavenumber cm-1
-Bruker Confidential- 67
Changes of Toluene Spectra with
Temperature1.165
25 °C
30 °C
1.160
35 °C
40 °C
Absorbance Units
45 °C
1.155
50 °C
55 °C
1.150
60 °C
65 °C
70 °C
1.145
75 °C
80 °C
85 °C
1.140
90 °C
95 °C
1.135
-Bruker Confidential- 68
OPUS Practical Session
• Datablock
• Color assignment
• Peak picking
• Zoom in/out
• Show parameter including checking gain, amp, time for BG
• Compare parameter
• Change background
• Check signal and gain set up
• Velocity
• Correlation
• OVP Set up, PQ Ref, LWN
• User setting and GLP mode
• Stray corr
-Bruker Confidential- 69
Innovation with Integrity
©Copyright
Copyright Bruker
© 2011 Bruker Corporation.
Corporation. All rights
All rights reserved. reserved.-Bruker Confidential-
www.bruker.com