MAMMOMAT Inspiration
SP

Software
System

Software installation (reinstallation)

10140000

© Siemens 2007
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: SPB7-330.816.01.03.02 English


Replaces: SPB7-330.816.01.02.02 Doc. Gen. Date: 10.08
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 2 of 84 Siemens


10.08 CS SD SP/CO
Table of Contents 3
0Table of Contents

1 _______ General Information______________________________________________ 5

Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product-specific remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Conventions and abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Access privileges to the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Required documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Required aids and tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

2 _______ WH AWS software installation ____________________________________ 10

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Installation Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Software installation break . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
System VA10B (Screening) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Software installation VB10B build 114, WH AWS . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Software patch installation, WH AWS VB10B build 115 . . . . . . . . . . . . . . . . . . . . . . 20
System Management Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
System VA10C (Screening) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Software installation VB10B build 114, WH AWS . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Software patch installation, WH AWS VB10B build 115 . . . . . . . . . . . . . . . . . . . . . . 30
Software patch installation, WH AWS VB10C. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Meduser Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Import image processing caf files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
System Management Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
System VA10D (Screening) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Software installation VB10B build 114, WH AWS . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Software patch installation, WH AWS VB10B build 115 . . . . . . . . . . . . . . . . . . . . . . 44
Software Patch Installation, WH AWS VB10D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Meduser Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Import image processing caf files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
System Management Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
System VB10 (Multiapplication) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Software installation WH AWS VB20B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
System Management Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

3 _______ Mammo Unit ___________________________________________________ 62

MU firmware installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Preconditions: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Siemens SPB7-330.816.01.03.02 Page 3 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
4 Table of Contents

Firmware installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63


Final check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Configuring the MammoUnit manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Backup the MammoUnit configuration (from VA10C) . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Restore the MammoUnit configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Preconditions: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Configuring the Mammo Unit Network Parameter (IP Address) . . . . . . . . . . . . . . . . 77

4 _______ Detector Firmware Update _______________________________________ 79

Updating the Detector Firmware System VA10 (Screening) . . . . . . . . . . . . . . . . . . . . . . 79


Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Backup the “contact.cal” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Updating the Detector Firmware System VB10 (Multiapplication) . . . . . . . . . . . . . . . . . 81
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Backup the “contact.cal” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

5 _______ Changes to the previous version __________________________________ 84

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 4 of 84 Siemens


10.08 CS SD SP/CO
General Information 5
1-
Notes
1General Information

Only qualified and system-trained service staff is allowed to perform MAMMOMAT Inspira-
tion system installation, service, maintenance and quality assurance. Ensure that the most
recent version of the service documentation is available.
To avoid any risk of injury to persons and/or damage to the system read and observe the
General Safety Notes (TD00-000.860.01.xx.02).
Please read and observe the “General remarks” which include very important
safety-related information as well as information about the handling of screws and nuts, use
of Loctite, and instructions for torque wrenches.

Safety information 0

Safety labeling

DANGER DANGER indicates an immediate danger that if disregarded


will cause death or serious physical injury.
¹

WARNING WARNING indicates a possible danger that if disregarded


can cause death or serious physical injury.
¹

CAUTION CAUTION used with the safety alert icon indicates a possible
danger that if disregarded will or can lead to minor or moder-
ate physical injury and/or damage to property.
¹

NOTICE NOTICE used without the safety alert icon indicates a possi-
ble danger that if disregarded can or will lead to an undesir-
able outcome or state other than death, physical injury or
property damage.
¹

NOTE NOTE is used to indicate information that explains the proper


way to use devices or to carry out a process, i.e., provides
pointers and tips.

Siemens SPB7-330.816.01.03.02 Page 5 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
6 General Information

Icons
X Warning about ionizing radiation or radioactive substances. Tests and adjustments that
must be performed with the radiation switched on are indicated by this radiation warning
icon.

V Dangerous electrical voltage > 25 VAC or > 60 VDC.

C Caution! General hazard warning.

E ESD: Warning about electrostatically sensitive components.

General remarks 0

NOTICE The screen shots in this document are used for illustration
and orientation purposes.
¹ For this reason, the actual screen may show some dif-
ferences.

NOTICE The installation of software not released or validated for the


system is strictly prohibited.
¹ Otherwise, the syngo software or one or more of its
applications may malfunction.

Product-specific remarks 0

The MAMMOMAT Inspiration is adjusted, programmed and tested in the factory, leaving the
adaptation to the on-site mains voltage, adjustment of the AEC, and the functional tests to
be performed. When the measurements to be made (kV, mAs, etc.) are within the toler-
ances stated in these instructions, this confirms that the settings made in the factory have
not changed and the equipment is fully serviceable.

System condition
Systems are delivered ex factory with complete software.

Installation time
The complete software installation procedure of each component takes no more than 1
hour. The time to configure or restore configuration data, and save patient images, is not
included in the calculation.

Power on the system


When the MAMMOMAT Inspiration is being configured, it must be completely cabled but
not connected to the network.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 6 of 84 Siemens


10.08 CS SD SP/CO
General Information 7

Specific network information


The cables for the network are not supplied in the product delivery volume.
The WH AWS is equipped with two network adapters.
The standard network adapter (first adapter) will be connected and configured to the hos-
pital network (DICOM connectivity). The factory defaults shall be checked and changed in
accordance with the hospital requirements.
The second network adapter (IP address 192.168.0.02) connects the WH AWS to the
MammoUnit (IP address 192.169.0.100).
• Check that the subnet of MammoUnit 192.168.0.0 (second network adapter of the WH
AWS) does not conflict with the hospital network. Contact the local network administra-
tor.
¹ If subnet 192.168.0.0 is reserved by the hospital, change the IP addresses of the
Mammo Unit and the WH AWS (second network adapter).

Test images
Test images are automatically installed during software installation.
The “Service Patient” and corresponding images are shown only when the “Service menu”
is opened and the log-in has been performed.

Acceptance test
The acceptance test of the MAMMOMAT Inspiration is recommended to be performed
whenever a new MAMMOMAT Inspiration system has been installed and whenever
changes that might affect performance have been made to an existing system.
The acceptance test of MAMMOMAT Inspiration involves performing all QA procedures in
this Quality Control Manual, ensuring that a basic minimum image quality criteria is met
before the system is used with patients.
All yearly tests from the Quality Control Manual have to be performed.
Depending on the country where the system is installed, additional tests have to be per-
formed.
• Tests according to §16 of the X-ray Ordinance must be performed in Germany.
• Tests according to DHHS regulations must be performed in the USA.
Maintenance measurements must be made according to the Maintenance Instructions
(SPB7-330.831.01...) and the DHHS Supplements to the instructions for use
(SPB7-330.661.01...)

Conventions and abbreviations 0

Subject Explanation
CBA Control Board Arm
CBS Control Board Stand
DICOM Digital Imaging and Communication in Medicine

Siemens SPB7-330.816.01.03.02 Page 7 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
8 General Information

Subject Explanation
IP address Internet Protocol address
MammoUnit Mammomat/ Stand
TCP/IP Transmission Control Protocol / Internet Protocol
WH AWS Acquisition workstation for MAMMOMAT Inspiration

Access privileges to the system 0

Passwords
Required passwords are not included in this document. You will find them in the CS Knowl-
edge Base on the intranet or you can obtain them from the relevant USC or RSC.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 8 of 84 Siemens


10.08 CS SD SP/CO
General Information 9

Requirements 1.1

Required documents 0

Document Print No.


General Safety information TD00-000.860.01...
User instructions SPB7-330.620.01...
Configuration SPB7-330.843.02.xx.02

Required aids and tools 0

- System Software of the "WH AWS".


- Empty DVD/ CD medium to save backup data.
- If the software is being reinstalled, a valid backup is recommended (refer to section
“Backup” in case a backup has to performed prior).
- BIOS CD
- Detector Firmware on CD
- MammoUnit Firmware on CD

Siemens SPB7-330.816.01.03.02 Page 9 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
10 WH AWS software installation
2-
General
2WH AWS software installation

Installation Prerequisites 0

Monitors
• Make sure that the monitor is connected and switched on.

USB
• Make sure that a USB memory stick or USB hard drive is not connected to the imaging
system PC.

NOTE Be aware that the dongle is not a memory stick.

Network
• The system must be physically disconnected from the network during software installa-
tion.

System serial number


• The system serial number must be available for software installation.

NOTE From WH AWS VB20, the system serial number can not be
modified after software installation!
Ensure that you enter the correct system serial number dur-
ing the software installation!

Patient Images
Patient images that are not saved may be lost during a software installation.
• If necessary, save the patient images.
- Send the patient images to the archive,
- burn them to CD/DVD
- or save them to the USB hard drive.

Password list
• The knowledge base password list must be available.

Entering and saving data

Service key
• Record the syngo service key ......................................................................................
¹ It can be found in the system binder or can be obtained from the responsible
USC.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 10 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 11

BIOS password
• Record the BIOS password...........................................................................................
¹ This is listed in the knowledge base password list.

Computer Name
• Record the computer name..........................................................................................
¹ It can be found in the service software under Configuration/Local host/TCP IP
LAN/Computer name.

Administrator account (Windows OS administrator)


• Record the administrator name ...................................................................................
¹ This is listed in the knowledge base password list.
• Record the administrator password .............................................................................
¹ This is listed in the knowledge base password list.

HIPAA administrator account (only if HIPAA is configured)

NOTE If a login window appears after the system startup, then


HIPAA is configured.

• Record the HIPAA administrator account .....................................................................


• Record the HIPAA administrator password...................................................................
1. If the administrator password is not known for certain, switch the system on and log into
the service application.
2. Click <Configuration>-<Next>-<Users>.
3. Enter the password (e. g. default password, see SP password list) in the "Password"
and “"Confirm" fields below "Account for Administrator".
¹ With HIPAA configured, the administrator password is necessary for the first
HIPAA login after "SW_Settings 02" is restored.

Backup
If reinstalling the software, back up the site-specific data:

NOTE All network settings, including the static route, are a part of
"SW-Settings02".

1. Open local service.


2. Insert an empty CD-R into the DVD drive.
3. Click <Backup & Restore>
4. Select "Command" and "Backup".
5. As the backup drive, select "(R) CD-R" in the "Drives" list field.

Siemens SPB7-330.816.01.03.02 Page 11 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
12 WH AWS software installation

6. Back up the following packages:


¹ SW-Settings02
¹ DXMG-Mammomat
¹ WHAWS
¹ Security-Settings (Option HIPAA)
¹ SW Distribution (optional)
¹ MammoUnit (only with WH AWS VB20)
7. Remove the CD-R from the DVD drive, and write the system serial number, software
version, date, and "Backup" on the CD-R.
8. Click <Home>.

Software installation break 0

NOTE Under no circumstances may the system be switched off dur-


ing installation. If the system is switched off, follow the
instructions below.

Procedure when the installation has been aborted


• Disconnect the PC power cable, wait about 5 seconds, and reconnect the power cable.
• Switch on the imaging system PC using the power button on the front side of the PC.
• Start again with "Software Installation VB10x"

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 12 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 13

System VA10B (Screening) 2.1

This section describes the software installation for MAMMOMAT Inspiration, system state
VA10B, WH AWS software version VB10B.
These systems shall be updated according to the Update Instructions SP009/08/P.

Software installation VB10B build 114, WH AWS 0

Installation Prerequisites
• Ensure that any network connection is disconnected from the hospital network.
• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The BIOS CD is on hand.
• Switch the system off.
• Within the WH AWS VB10B, the system serial number, e.g. #1022, has to be entered
with 5 digits e.g. #11022.
¹ That problem is solved with WH VB10C.

Loading the software

Loading the WH AWS BIOS CD

1. Turn ON the PC and press F2 to enter the BIOS menu.


¹ The BIOS login window is displayed.

Fig. 1: Bios_PW
2. Enter the BIOS password, and confirm it by pressing the "Enter" key.
3. Press F9 in the BIOS menu to load the default BIOS settings and confirm with YES.
4. Set the current date and time. For SW installation on the PC, type M440, select the
installed hard drive(s) on the BIOS main menu window, and disable the "Noise Man-
agement (Silent Mode)" feature for each installed hard drive.
5. Insert the BIOS CD into the CD-ROM drive.
Press the ESC button and leave the BIOS via Save Changes & Exit. Confirm with
YES. The PC reboots.

Siemens SPB7-330.816.01.03.02 Page 13 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
14 WH AWS software installation

6. The PC boots from the CD-ROM and loads the BIOS version, if required. The follow-
ing messages are displayed.
"BIOS Installation Done - System Ready for Software Installation"
"Please remove BIOS CD!"
"Please turn off or reset system!"
7. Remove the BIOS CD.

Loading the WH AWS VB10B Build 114 software

1. Insert the "System Software VB10B Build 114" DVD into the DVD drive.
2. Reboot the PC.
MS-DOS (as part of Windows 98) is started from the CD-ROM.
Starting Win98
After a short time, a start screen is displayed in text mode. In the top right corner, a mes-
sage indicates that the EIDE, SCSI and CD-ROM drivers are loaded.
If any drivers cannot be found with this hardware configuration, an error message is
generated and can be ignored.
3. After a short time, the screen changes and the PC starts to beep.
¹ The following screen is shown:

SELECTION MENU INFORMATION WINDOW


Turn off computer to interrupt Creating a new primary partition on the first
or press any key to confirm hard drive. This partition will be converted to
NTFS during Windows XP installation. Con-
operation and to firm to DESTROY DATA on the first partition
of the first hard drive.
CONTINUE INSTALLATION

HARDWARE DETECTION IN PROGRESS


Checking first hard drive...

4. Press Enter to continue with the installation. A new primary partition is created and the
PC is restarted.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 14 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 15

5. Leave the installation CD in the CD-ROM drive. After a short time, the following screen
is displayed:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: * Operating system fixes
Create partitions on hard drive 2 Additional device drivers
Setup type: AWS Additional files
Med Software
Installation in progress
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

6. The following settings need to be made:


- Machine name
Enter the appropriate machine name.

NOTE The machine name must be changed and has to be unique


within the hospital network environment.
If you do not change the machine name '* or MedPC’, the
syngo installation cannot be properly completed.

- Partitioning
Move the cursor to “Keep data on other partitions” and toggle to “Create partitions
on hard drive 2”.

NOTE If reinstalling the software, “Keep data on other partitions” is


optional to save patient images on disk.
Be sure that all patients are archived.

- Continue with the installation


Move the cursor to CONTINUE INSTALLATION and press Enter.

Siemens SPB7-330.816.01.03.02 Page 15 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
16 WH AWS software installation

7. The final text on the screen changes. Press Enter again to continue with the installa-
tion. It is normal for the PC to start beeping again.

SELECTION MENU INFORMATION WINDOW


No pause in installation Now destroying all data and creating further
UDB-file: on CD-ROM partitions on hard drive 1-2.

Machine Name: YBHH007154


Create partitions on hard drive 2 Please confirm to continue.

Setup type: AWS

Press any key to continue.


HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

8. The installation files are copied from the CD-ROM to the hard drive. The screen
changes as follows:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: YBHH007154 Operating system fixes
Setup type: AWS Additional device drives.

Create further partitions


on hard drives 1-2
Installation in progress...
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

¹ After the installation, files are prepared and the CD-ROM is unloaded. The PC is
then rebooted.

NOTE The CD-ROM must be removed immediately, otherwise the


PC automatically reloads the CD after rebooting and installa-
tion begins again. If this happens, remove the CD-ROM with-
out striking any key and restart the PC.

¹ Windows XP continues preparing the files. The PC is restarted several times for:

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 16 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 17

- Collecting information
- Dynamic update
- Preparing installation
- Installing Windows
- Finalizing installation
¹ After the WH AWS software installation is complete, the system reboots again.
“Please enter license” is displayed when ready.

Fig. 2: Starting the Local Service Configuration

NOTE If no backup is on hand, configure the system according to


the configuration guide.

9. Perform the “Configuration” or the “Restore” if a backup is available.

Configuration

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.

Siemens SPB7-330.816.01.03.02 Page 17 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
18 WH AWS software installation

5. Perform the configuration as described in the document “Configuration,


SPB7-330.843.02..., section WH AWS software configuration”.
6. Restart the WH AWS.

Country-specific keyboard layout and user interface

1. In the "Configuration" panel of the user interface, double-click “Regional and Lan-
guage”.
2. Select “Regional Options”.
3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result,


“Apply” is not available.

6. Select the “Languages” tab and click <Details>.


7. Under “Installed services”, click “Add”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “Input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 18 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 19

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the “Text Services and Input Languages” window.
12. Click <Apply> and <OK> to close the “Regional and Language Options” window.
¹ If the language has been changed, the following message appears:
The following is displayed if the values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change, click <OK>. This will apply the settings and the sys-
tem restarts.
13. Restart the WH AWS.

Restore

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.
5. Insert the backup CD into the DVD drive.
6. In the Service menu, click <Backup & Restore>.
7. Under “Command”, select “Restore” and select “(-R-) CD-R” under “Drives”.

SW-Settings 02

1. Under “Archive”, select the correct (newest) backup file of “SW-Settings 02”
2. Highlight all the files in “Groups” (press the “Shift” key and select the first and last entry
of groups).
3. Click <Go> .

AWS Settings

1. Under “Archive”, select the correct backup file of the “WHAWS Settings”
2. Highlight all the files in "Groups”.
3. Click <Go> .

Siemens SPB7-330.816.01.03.02 Page 19 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
20 WH AWS software installation

Security-Settings (only if HIPAA is configured)

NOTE If HIPAA is configured, an HIPAA administrator account is


required. Refer to section “Entering and Saving Data” if
needed.

1. Under “Archive”, select the correct backup file of “Security Settings”.


2. Click <Go> .
¹ Now the user names and passwords stored in the security settings are valid
again.
¹ Next time, log in with the user name/password obtained from the HIPAA adminis-
trator.

SW-Distribution (optional)

1. Under “Archive”, select the correct backup file of “SW-Distribution”.


2. Click <Go> .

Remove backup CD

1. Remove the backup CD from the DVD drive.

Activating the restored values

1. Restart the system via the Windows menu bar, click <Start>-<Shut Down>, select
<Restart> and click <OK>.
¹ The patient database creation starts if “Create partitions on hard drive 2” was pre-
viously selected, and takes about 15 minutes.
¹ Wait until the patient registration dialog or the patient browser is displayed.
2. Close the patient registration dialog or patient browser.
3. Click the window “Warning: Database...” and press the “Enter” key.
¹ The window “Warning: Database...” disappears.

Loading the BIOS CD


Loading the BIOS CD
1. Load the BIOS CD. Reboot the PC.
2. The following message is displayed: <Installation Done - HIPAA Installed>
¹ The boot procedure from CD-ROM is disabled again.
3. After the restart, the syngo login screen (when HIPAA is activated) or the examination
task card is displayed.

Software patch installation, WH AWS VB10B build 115 0

This section describes the software patch installation for “Single Touch” functionality.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 20 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 21

Software installation break

NOTE Under no circumstances may the system be switched off dur-


ing installation.

Installation prerequisites

NOTE Ignore subsequent software installation user messages


about detector temperature warnings, because the detector
firmware is updated at a later time!

• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• With an optional monitor, disconnect the DVI cable for the “High Resolution Moni-
tor” during the complete update of the system!
¹ Either on the monitor side or WH AWS side!
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The prerequisite is a successfully installed WH AWS VB10B build 114 version.

Patch installation procedure


The patch CD WH AWS VB10B Build 115 is installed as follows:
1. Switch on or reboot the WH AWS.
2. Perform “Restart Application” and hold down the Shift key until the Logon window
appears.
3. Log on as administrator:
User name: administrator Password:*********
4. Insert the patch CD “WH AWS VB10 build 115 – Patch CD” into the CD/DVD drive.
5. Open the file explorer (My Computer) and select the CD/DVD drive.
6. Execute “update.cmd” by double-clicking it.
¹ A DOS window shows the progress of the installation.
¹ After installation of the bug fix has finished (approx. 1 min.), the DOS window
remains open.
¹ No error messages should be present.
7. Close the DOS window by pressing the ENTER key.
8. Remove the CD.
9. Restart the WH AWS. During boot up, new software version VB10B build 115 will be
shown on the screen.

Siemens SPB7-330.816.01.03.02 Page 21 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
22 WH AWS software installation

System Management Registration 0

After the complete WH AWS software installation, a registration of the System Manage-
ment services (Siemens Remote Service is activated) has to be performed.
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01....).
• Perform System Management (when Siemens Remote Service is activated).

Final work steps 0

Patient Images
• If necessary, restore the patient images.

Second monitor (optional)


• If necessary, reconnect and configure the second monitor as described within the
appropriated monitor document.

Detector calibration
• The Detector Calibration procedure has to be performed for “Contact” and “Magnifica-
tion”. Please see the MAMMOMAT Inspiration user manual for more details on how to
calibrate the system.
¹ Wait at least 30 minutes to warm up the detector.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 22 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 23

System VA10C (Screening) 2.2

This section describes the software installation for MAMMOMAT Inspiration, system state
VA10C, WH AWS software version VB10C.
These systems shall be updated according to the Update Instruction SP025/08/P.

Software installation VB10B build 114, WH AWS 0

Installation Prerequisites
• Ensure that any network connection is disconnected from the hospital network.
• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The BIOS CD is on hand.
• Switch the system off.
• Within the WH AWS VB10B, the system serial number, e.g. #1022, has to be entered
with 5 digits e.g. #11022.
¹ That problem is solved with WH VB10C.

System Management deregistration


To avoid inconsistencies of the System Management database, perform the deregistration
of the System Management services prior to the software re-installation (if applicable).
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01...
• Perform the System Management deregistration within the local service/ System Man-
agement/ Agent Control, deselect the checkboxes, and click “Go” (if applicable).

Loading the software

Loading the WH AWS BIOS CD

1. Turn ON the PC and press F2 to enter the BIOS menu.


¹ The BIOS login window is displayed.

Fig. 3: Bios_PW
2. Enter the BIOS password and confirm it by pressing the “Enter” key.

Siemens SPB7-330.816.01.03.02 Page 23 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
24 WH AWS software installation

3. Press F9 in the BIOS menu to load the default BIOS settings and confirm with YES.
4. Set the current date and time. For SW installation on the PC, type M440, select the
installed hard drive(s) on the BIOS main menu window, and disable the "Noise Man-
agement (Silent Mode)" feature for each installed hard drive.
5. Insert the BIOS CD into the CD-ROM drive.
Press the ESC button and leave the BIOS via Save Changes & Exit. Confirm with
YES. The PC reboots.
6. The PC boots from the CD-ROM and loads the BIOS version, if required. The follow-
ing messages are displayed.
“BIOS Installation Done - System Ready for Software Installation”
“Please remove BIOS CD!”
“Please turn off or reset system!”
7. Remove the BIOS CD.

Loading the WH AWS VB10B Build 114 software

1. Insert the “System Software VB10B Build 114” DVD into the DVD drive.
2. Reboot the PC.
MS-DOS (as part of Windows 98) is started from the CD-ROM.
Starting Win98
After a short time, a start screen is displayed in text mode. In the top right corner, a mes-
sage indicates that the EIDE, SCSI and CD-ROM drivers are loaded.
If any drivers cannot be found with this hardware configuration, an error message is
generated and can be ignored.
3. After a short time, the screen changes and the PC starts to beep.
¹ The following screen is shown:

SELECTION MENU INFORMATION WINDOW


Turn off computer to interrupt Creating a new primary partition on the first
or press any key to confirm hard drive. This partition will be converted to
NTFS during Windows XP installation. Con-
operation and to firm to DESTROY DATA on the first partition
of the first hard drive.
CONTINUE INSTALLATION

HARDWARE DETECTION IN PROGRESS


Checking first hard drive...

4. Press Enter to continue with the installation. A new primary partition is created and the
PC is restarted.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 24 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 25

5. Leave the installation CD in the CD-ROM drive. After a short time, the following screen
is displayed:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: * Operating system fixes
Create partitions on hard drive 2 Additional device drivers
Setup type: AWS Additional files
Med Software
Installation in progress
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

6. The following settings need to be made:


- Machine name
Enter the appropriate machine name.

NOTE The machine name must be changed and has to be unique


within the hospital network environment.
If you do not change the machine name '* or MedPC’, the
syngo installation cannot be properly completed.

- Partitioning
Move the cursor to “Keep data on other partitions” and toggle to “Create partitions
on hard drive 2”.

NOTE If reinstalling the software, “Keep data on other partitions” is


optional to save patient images on disk.
Be sure that all patients are archived.

- Continue with the installation


Move the cursor to CONTINUE INSTALLATION and press Enter.

Siemens SPB7-330.816.01.03.02 Page 25 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
26 WH AWS software installation

7. The final text on the screen changes. Press Enter again to continue with the installa-
tion. It is normal for the PC to start beeping again.

SELECTION MENU INFORMATION WINDOW


No pause in installation Now destroying all data and creating further
UDB-file: on CD-ROM partitions on hard drive 1-2.

Machine Name: YBHH007154


Create partitions on hard drive 2 Please confirm to continue.

Setup type: AWS

Press any key to continue.


HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

8. The installation files are copied from the CD-ROM to the hard drive. The screen
changes as follows:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: YBHH007154 Operating system fixes
Setup type: AWS Additional device drives.

Create further partitions


on hard drives 1-2
Installation in progress...
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

¹ After the installation, files are prepared and the CD-ROM is unloaded. The PC is
then rebooted.

NOTE The CD-ROM must be removed immediately, otherwise the


PC automatically reloads the CD after rebooting and installa-
tion begins again. If this happens, remove the CD-ROM with-
out striking any key and restart the PC.

¹ Windows XP continues preparing the files. The PC is restarted several times for:

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 26 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 27

- Collecting information
- Dynamic update
- Preparing installation
- Installing Windows
- Finalizing installation
¹ After the WH AWS software installation is complete, the system reboots again.
“Please enter license” is displayed when ready.

Fig. 4: Starting the Local Service Configuration

NOTE If no backup is on hand, configure the system according to


the configuration guide.

9. Perform the “Configuration” or the “Restore” if a backup is available.

Configuration

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.

Siemens SPB7-330.816.01.03.02 Page 27 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
28 WH AWS software installation

5. Perform the configuration as described in the document “Configuration,


SPB7-330.843.02..., section WH AWS software configuration”.
6. Restart the WH AWS.

Country-specific keyboard layout and user interface

1. In the "Configuration" panel of the user interface double-click “Regional and Lan-
guage”.
2. Select “Regional Options”.
3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result,


“Apply” is not available.

6. Select the “Languages” tab and click <Details>.


7. Under “Installed services”, click “Add”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “Input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 28 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 29

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the “Text Services and Input Languages” window.
12. Click <Apply> and <OK> to close the “Regional and Language Options” window.
¹ If the language has been changed, the following message appears:
The following is displayed if the values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change, click <OK>. This will apply the settings and the sys-
tem restarts.
13. Restart the WH AWS.

Restore

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.
5. Insert the backup CD into the DVD drive.
6. In the Service menu, click <Backup & Restore>.
7. Under “Command”, select “Restore” and select “(-R-) CD-R” under “Drives”.

SW-Settings 02

1. Under “Archive”, select the correct (newest) backup file of “SW-Settings 02”
2. Highlight all the files in “Groups” (press the “Shift” key and select the first and last entry
of groups).
3. Click <Go> .

AWS Settings

1. Under “Archive”, select the correct backup file of the “WHAWS Settings”
2. Highlight all the files in "Groups”.
3. Click <Go> .

Siemens SPB7-330.816.01.03.02 Page 29 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
30 WH AWS software installation

Security-Settings (only if HIPAA is configured)

NOTE If HIPAA is configured, an HIPAA administrator account is


required. Refer to section “Entering and Saving Data” if
needed.

1. Under “Archive”, select the correct backup file of “Security Settings”.


2. Click <Go> .
¹ Now the user names and passwords stored in the security settings are valid
again.
¹ Next time, log in with the user name/password obtained from the HIPAA adminis-
trator.

SW-Distribution (optional)

1. Under “Archive”, select the correct backup file of “SW-Distribution”.


2. Click <Go> .

Remove backup CD

1. Remove the backup CD from the DVD drive.

Activating the restored values

1. Restart the system via the Windows menu bar, click <Start>-<Shut Down>, select
<Restart> and click <OK>.
¹ The patient database creation starts if “Create partitions on hard drive 2” was pre-
viously selected, and takes about 15 minutes.
¹ Wait until the patient registration dialog or the patient browser is displayed.
2. Close the patient registration dialog or patient browser.
3. Click the window “Warning: Database...” and press the “Enter” key.
¹ The window “Warning: Database...” disappears.

Loading the BIOS CD


Loading the BIOS CD
1. Load the BIOS CD. Reboot the PC.
2. The following message is displayed: <Installation Done - HIPAA Installed>
¹ The boot procedure from CD-ROM is disabled again.
3. After the restart, the syngo login screen (when HIPAA is activated) or the examination
task card is displayed.

Software patch installation, WH AWS VB10B build 115 0

This section describes the software patch installation for “Single Touch” functionality.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 30 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 31

Software installation break

NOTE Under no circumstances may the system be switched off dur-


ing installation.

Installation prerequisites

NOTE Ignore subsequent software installation user messages


about detector temperature warnings, because the detector
firmware is updated at a later time!

• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• With an optional monitor, disconnect the DVI cable for the “High Resolution Moni-
tor” during the complete update of the system!
¹ Either on the monitor side or WH AWS side!
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The prerequisite is a successfully installed WH AWS VB10B build 114 version.

Patch installation procedure


The patch CD WH AWS VB10B Build 115 is installed as follows:
1. Perform “Restart Application” and hold down the Shift key until the Logon window
appears.
2. Log on as administrator:
User name: administrator Password:*********
3. Insert the patch CD “WH AWS VB10 build 115 – Patch CD” into the CD/DVD drive.
4. Open the file explorer (My Computer) and select the CD/DVD drive.
5. Execute “update.cmd” by double-clicking it.
¹ A DOS window shows the progress of the installation.
¹ After installation of the bug fix has finished (approx. 1 min.), the DOS window
remains open.
¹ No error messages should be present.
6. Close the DOS window by pressing the ENTER key.
7. Remove the CD.
8. Restart the WH AWS. During boot up, new software version VB10B build 115 will be
shown on the screen.

Siemens SPB7-330.816.01.03.02 Page 31 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
32 WH AWS software installation

Software patch installation, WH AWS VB10C 0

Software installation break

NOTE Under no circumstances may the system be switched off dur-


ing installation.

Installation prerequisites

NOTE Ignore subsequent software installation user messages


about detector temperature warnings, because the detector
firmware is updated at a later time!

• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• With an optional monitor, disconnect the DVI cable for the “High Resolution Moni-
tor” during the complete update of the system!
¹ Either on the monitor side or WH AWS side!
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The prerequisite is a successfully installed WH AWS VB10B build 115 version.

Patch installation procedure


The patch CD WH AWS VB10C is installed as follows:
1. Perform “Restart Application” and hold down the Shift key until the Logon window
appears.
2. Log on as administrator:
User name: administrator Password:*********
3. Insert the patch CD “WH AWS VB10C – Patch CD” into the CD/DVD drive.
4. Open the file explorer (My Computer) and select the CD/DVD drive.
5. Execute “update.cmd” by double-clicking it.
¹ A DOS window shows the progress of the installation.
¹ After installation of the bug fix has finished (approx. 1 min.), the DOS window
remains open.
¹ No error messages should be present.
6. Close the DOS window by pressing the ENTER key.
7. Remove the CD.
8. Restart the WH AWS. During the boot-up, the new software version VB10C is shown
on the screen.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 32 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 33

Configuring the System Serial Number (4 digits) in the WH AWS Software


Since installation of the WH AWS VB10B software, configuration of the system serial num-
ber (4 digits) in the WH AWS software fails and for configuration purposes, the number has
to be increased by one digit to 5.
For WH VB10C, configuration of the valid 4-digit system serial number is required, e.g.
#11022 is corrected to #1022.
Proceed as follows to correct the system serial number in the WH AWS software:
1. Log in to local service.
2. Select “Configuration” and then “Site Info” in the menu.
3. Correct the system serial number to 4 digits as found in the “System binder” or on the
back of the MammoUnit (label).
4. Click “Home”, and confirm the message box indicating that the computer will be reboo-
ted.

Meduser Password 0

Ensure that the default meduser password is activated.


1. Log into Local Service.
2. Select “Configuration/ Users”.
3. Select the check box “Meduser”, “Default Password” and Auto Login.
¹ If required, ask the customer whether a meduser password exists and make a
note of it!

Fig. 5: Meduser password


4. Click “Save”.

Siemens SPB7-330.816.01.03.02 Page 33 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
34 WH AWS software installation

5. Continue with the next section “Importing image processing caf files” before reboot-
ing the system.

Import image processing caf files 0

1. Log in to local service.


2. Select “Acquisition System/ Image Processing” and click “View” to activate the
“Advanced Settings”.

Fig. 6: Show advanced settings


3. Insert the delivered CD “WH AWS VB10C” into the CD drive.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 34 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 35

4. Click “Browse” for the Engineering Parameters and go to the folder


“Patch_CD_VB10C124_P16” of the installation CD WH AWS VB10C, e.g.
“R:\.Patch_CD_VB10C124_P16”.

Fig. 7: Engineering parameters


5. Select the “preprocessing.caf” file.
6. Click Import. The file preprocessing.caf is copied to “C:\AWS\config\SiemensIP”.
7. Remove the CD from the DVD/ CD drive.
8. Restart the WH AWS.

System Management Registration 0

After the complete WH AWS software installation, a registration of the System Manage-
ment services (Siemens Remote Service is activated) has to be performed.
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01....).
• Perform System Management (when Siemens Remote Service is activated).

Final work steps 0

Patient Images
• If necessary, restore the patient images.

Siemens SPB7-330.816.01.03.02 Page 35 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
36 WH AWS software installation

Second monitor (optional)


• If necessary, reconnect and configure the second monitor as described within the
appropriated monitor document.

Detector calibration
• The Detector Calibration procedure has to be performed for “Contact” and “Magnifica-
tion”. Please see the MAMMOMAT Inspiration user manual for more details on how to
calibrate the system.
¹ Wait at least 30 minutes to warm up the detector.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 36 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 37

System VA10D (Screening) 2.3

This section describes the software installation for MAMMOMAT Inspiration, system state
VA10D, WH AWS software version VB10D.

Software installation VB10B build 114, WH AWS 0

Installation Prerequisites
• Ensure that any network connection is disconnected from the hospital network.
• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The BIOS CD is on hand.
• Switch the system off.
• Within the WH AWS VB10B, the system serial number, e.g. #1022, has to be entered
with 5 digits e.g. #11022.
¹ That problem is solved with WH VB10C.

System Management deregistration


To avoid inconsistencies of the System Management database, perform the deregistration
of the System Management services prior to the software re-installation (if applicable).
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01...
• Perform the System Management deregistration within the local service/ System Man-
agement/ Agent Control, deselect the checkboxes, and click “Go” (if applicable).

Loading the software

Loading the WH AWS BIOS CD

1. Turn ON the PC and press F2 to enter the BIOS menu.


¹ The BIOS login window is displayed.

Fig. 8: Bios_PW
2. Enter the BIOS password and confirm it by pressing the “Enter” key.
3. Press F9 in the BIOS menu to load the default BIOS settings and confirm with YES.

Siemens SPB7-330.816.01.03.02 Page 37 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
38 WH AWS software installation

4. Set the current date and time. For SW installation on the PC, type M440, select the
installed hard drive(s) on the BIOS main menu window, and disable the "Noise Man-
agement (Silent Mode)" feature for each installed hard drive.
5. Insert the BIOS CD into the CD-ROM drive.
Press the ESC button and leave the BIOS via Save Changes & Exit. Confirm with
YES. The PC reboots.
6. The PC boots from the CD-ROM and loads the BIOS version, if required. The follow-
ing messages are displayed.
“BIOS Installation Done - System Ready for Software Installation”
“Please remove BIOS CD!”
“Please turn off or reset system!”
7. Remove the BIOS CD.

Loading the WH AWS VB10B Build 114 software

1. Insert the “System Software VB10B Build 114” DVD into the DVD drive.
2. Reboot the PC.
MS-DOS (as part of Windows 98) is started from the CD-ROM.
Starting Win98
After a short time, a start screen is displayed in text mode. In the top right corner, a mes-
sage indicates that the EIDE, SCSI and CD-ROM drivers are loaded.
If any drivers cannot be found with this hardware configuration, an error message is
generated and can be ignored.
3. After a short time, the screen changes and the PC starts to beep.
¹ The following screen is shown:

SELECTION MENU INFORMATION WINDOW


Turn off computer to interrupt Creating a new primary partition on the first
or press any key to confirm hard drive. This partition will be converted to
NTFS during Windows XP installation. Con-
operation and to firm to DESTROY DATA on the first partition
of the first hard drive.
CONTINUE INSTALLATION

HARDWARE DETECTION IN PROGRESS


Checking first hard drive...

4. Press Enter to continue with the installation. A new primary partition is created and the
PC is restarted.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 38 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 39

5. Leave the installation CD in the CD-ROM drive. After a short time, the following screen
is displayed:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: * Operating system fixes
Create partitions on hard drive 2 Additional device drivers
Setup type: AWS Additional files
Med Software
Installation in progress
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

6. The following settings need to be made:


- Machine name
Enter the appropriate machine name.

NOTE The machine name must be changed and has to be unique


within the hospital network environment.
If you do not change the machine name '* or MedPC’, the
syngo installation cannot be properly completed.

- Partitioning
Move the cursor to “Keep data on other partitions” and toggle to “Create partitions
on hard drive 2”.

NOTE If reinstalling the software, “Keep data on other partitions” is


optional to save patient images on disk.
Be sure that all patients are archived.

- Continue with the installation


Move the cursor to CONTINUE INSTALLATION and press Enter.

Siemens SPB7-330.816.01.03.02 Page 39 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
40 WH AWS software installation

7. The final text on the screen changes. Press Enter again to continue with the installa-
tion. It is normal for the PC to start beeping again.

SELECTION MENU INFORMATION WINDOW


No pause in installation Now destroying all data and creating further
UDB-file: on CD-ROM partitions on hard drive 1-2.

Machine Name: YBHH007154


Create partitions on hard drive 2 Please confirm to continue.

Setup type: AWS

Press any key to continue.


HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

8. The installation files are copied from the CD-ROM to the hard drive. The screen
changes as follows:

SELECTION MENU INFORMATION WINDOW


No pause in installation Copying OEM setup files
UDB-file: on CD-ROM Operating system setup files
Machine Name: YBHH007154 Operating system fixes
Setup type: AWS Additional device drives.

Create further partitions


on hard drives 1-2
Installation in progress...
HARDWARE DETECTION RESULTS
NVidia chipset-based graphic adapter (Quadro NVS-285 PCI-E (NV34))
found

¹ After the installation, files are prepared and the CD-ROM is unloaded. The PC is
then rebooted.

NOTE The CD-ROM must be removed immediately, otherwise the


PC automatically reloads the CD after rebooting and installa-
tion begins again. If this happens, remove the CD-ROM with-
out striking any key and restart the PC.

¹ Windows XP continues preparing the files. The PC is restarted several times for:

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 40 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 41

- Collecting information
- Dynamic update
- Preparing installation
- Installing Windows
- Finalizing installation
¹ After the WH AWS software installation is complete, the system reboots again.
“Please enter license” is displayed when ready.

Fig. 9: Starting the Local Service Configuration

NOTE If no backup is on hand, configure the system according to


the configuration guide.

9. Perform the “Configuration” or the “Restore” if a backup is available.

Configuration

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.

Siemens SPB7-330.816.01.03.02 Page 41 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
42 WH AWS software installation

5. Perform the configuration as described in the document “Configuration,


SPB7-330.843.02..., section WH AWS software configuration”.
6. Restart the WH AWS.

Country-specific keyboard layout and user interface

1. In the "Configuration" panel of the user interface double-click “Regional and Lan-
guage”.
2. Select “Regional Options”.
3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result,


“Apply” is not available.

6. Select the “Languages” tab and click <Details>.


7. Under “Installed services”, click “Add”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “Input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 42 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 43

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the “Text Services and Input Languages” window.
12. Click <Apply> and <OK> to close the “Regional and Language Options” window.
¹ If the language has been changed, the following message appears:
The following is displayed if the values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change, click <OK>. This will apply the settings and the sys-
tem restarts.
13. Restart the WH AWS.

Restore

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.
5. Insert the backup CD into the DVD drive.
6. In the Service menu, click <Backup & Restore>.
7. Under “Command”, select “Restore” and select “(-R-) CD-R” under “Drives”.

SW-Settings 02

1. Under “Archive”, select the correct (newest) backup file of “SW-Settings 02”
2. Highlight all the files in “Groups” (press the “Shift” key and select the first and last entry
of groups).
3. Click <Go> .

AWS Settings

1. Under “Archive”, select the correct backup file of the “WHAWS Settings”
2. Highlight all the files in "Groups”.
3. Click <Go> .

Siemens SPB7-330.816.01.03.02 Page 43 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
44 WH AWS software installation

Security-Settings (only if HIPAA is configured)

NOTE If HIPAA is configured, an HIPAA administrator account is


required. Refer to section “Entering and Saving Data” if
needed.

1. Under “Archive”, select the correct backup file of “Security Settings”.


2. Click <Go> .
¹ Now the user names and passwords stored in the security settings are valid
again.
¹ Next time, log in with the user name/password obtained from the HIPAA adminis-
trator.

SW-Distribution (optional)

1. Under “Archive”, select the correct backup file of “SW-Distribution”.


2. Click <Go> .

Remove backup CD

1. Remove the backup CD from the DVD drive.

Activating the restored values

1. Restart the system via the Windows menu bar, click <Start>-<Shut Down>, select
<Restart> and click <OK>.
¹ The patient database creation starts if “Create partitions on hard drive 2” was pre-
viously selected, and takes about 15 minutes.
¹ Wait until the patient registration dialog or the patient browser is displayed.
2. Close the patient registration dialog or patient browser.
3. Click the window “Warning: Database...” and press the “Enter” key.
¹ The window “Warning: Database...” disappears.

Loading the BIOS CD


Loading the BIOS CD
1. Load the BIOS CD. Reboot the PC.
2. The following message is displayed: <Installation Done - HIPAA Installed>
¹ The boot procedure from CD-ROM is disabled again.
3. After the restart, the syngo login screen (when HIPAA is activated) or the examination
task card is displayed.

Software patch installation, WH AWS VB10B build 115 0

This section describes the software patch installation for “Single Touch” functionality.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 44 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 45

Software installation break

NOTE Under no circumstances may the system be switched off dur-


ing installation.

Installation prerequisites

NOTE Ignore subsequent software installation user messages


about detector temperature warnings, because the detector
firmware is updated at a later time!

• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• With an optional monitor, disconnect the DVI cable for the “High Resolution Moni-
tor” during the complete update of the system!
¹ Either on the monitor side or WH AWS side!
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The prerequisite is a successfully installed WH AWS VB10B build 114 version.

Patch installation procedure


The patch CD WH AWS VB10B Build 115 is installed as follows:
1. Perform “Restart Application” and hold down the Shift key until the Logon window
appears.
2. Log on as administrator:
User name: administrator Password:*********
3. Insert the patch CD “WH AWS VB10 build 115 – Patch CD” into the CD/DVD drive.
4. Open the file explorer (My Computer) and select the CD/DVD drive.
5. Execute “update.cmd” by double-clicking it.
¹ A DOS window shows the progress of the installation.
¹ After installation of the bug fix has finished (approx. 1 min.), the DOS window
remains open.
¹ No error messages should be present.
6. Close the DOS window by pressing the ENTER key.
7. Remove the CD.
8. Restart the WH AWS. During boot up, new software version VB10B build 115 will be
shown on the screen.

Siemens SPB7-330.816.01.03.02 Page 45 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
46 WH AWS software installation

Software Patch Installation, WH AWS VB10D 0

Software installation break

NOTE Under no circumstances may the system be switched off dur-


ing installation.

Installation prerequisites

NOTE Ignore any software installation user messages about detec-


tor temperature warnings, because the detector firmware will
be updated at a later point.

• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• If an optional monitor is installed, Disconnect the DVI cable of the “High Resolution
Monitor” before performing a complete update of the system.
¹ Either on the monitor side or WH AWS side!
• Ensure that a backup of the current system configuration is on hand in case reinstalla-
tion is necessary.
• Ensure that all patients are archived.
• Prerequisite is a successfully installed WH AWS VB10B build 115 or later version.

Patch installation procedure


The WH AWS VB10D patch CD is installed as follows:
1. Perform the “Application Restart” and hold down the Shift key until the Logon window
appears.
2. Log on as administrator:
User name: administrator Password:*********
3. Insert the “WH AWS VB10D – Patch CD” into the CD/DVD drive.
4. Open the file explorer ("My computer") and select the CD/DVD drive.
5. Execute “aws-update.vbs” by double-clicking it.
¹ A window appears.

Fig. 10: Start VB10D update

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 46 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 47

6. Click “OK” to start the installation.


7. After the installation (approximately 10 min.), confirm the message box (indicating that
the installation was successful) with “OK”.
¹ No error messages should appear.

Fig. 11: Installation successfully completed


¹ In case of error, click “OK”, and start the installation procedure again.

Fig. 12: Update failed


8. Remove the CD.
9. Restart the WH AWS. During the bootup, the new software version VB10D will be
shown on the screen.

Meduser Password 0

Ensure that the default meduser password is activated.


1. Log into Local Service.
2. Select “Configuration/ Users”.

Siemens SPB7-330.816.01.03.02 Page 47 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
48 WH AWS software installation

3. Select the check box “Meduser”, “Default Password” and Auto Login.
¹ If required, ask the customer whether a meduser password exists and make a
note of it!

Fig. 13: Meduser password


4. Click “Save”.
5. Continue with the next section “Importing image processing caf files” before reboot-
ing the system.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 48 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 49

Import image processing caf files 0

1. Log in to local service.


2. Select “Acquisition System/ Image Processing” and click “View” to activate the
“Advanced Settings”.

Fig. 14: Show advanced settings


3. Insert the delivered CD “WH AWS VB10C” into the CD drive.

Siemens SPB7-330.816.01.03.02 Page 49 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
50 WH AWS software installation

4. Click “Browse” for the Engineering Parameters and go to the folder


“Patch_CD_VB10C124_P16” of the installation CD WH AWS VB10C, e.g.
“R:\.Patch_CD_VB10C124_P16”.

Fig. 15: Engineering parameters


5. Select the “preprocessing.caf” file.
6. Click Import. The file preprocessing.caf is copied to “C:\AWS\config\SiemensIP”.
7. Remove the CD from the DVD/ CD drive.
8. Restart the WH AWS.

System Management Registration 0

After the complete WH AWS software installation is finished, the System Management ser-
vices (Siemens Remote Service is activated) has to be registered.
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01...
• Perform the System Management registration within the local service/ System Man-
agement/ Agent Control, select the checkboxes, and click “Go” (when Siemens
Remote Service is activated).

Final work steps 0

Patient Images
• If necessary, restore the patient images.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 50 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 51

Second monitor (optional)


• If necessary, reconnect and configure the second monitor as described within the
appropriated monitor document.

Detector calibration
• The Detector Calibration procedure has to be performed for “Contact” and “Magnifica-
tion”. Please see the MAMMOMAT Inspiration user manual for more details on how to
calibrate the system.
¹ Wait at least 30 minutes to warm up the detector.

Siemens SPB7-330.816.01.03.02 Page 51 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
52 WH AWS software installation

System VB10 (Multiapplication) 2.4

This section describes the software installation for MAMMOMAT Inspiration, system state
VB10B, WH AWS software version VB20B.

Software installation WH AWS VB20B 0

Installation Prerequisites
• Ensure that any network connection is disconnected from the hospital network.
• Ensure that a USB memory stick or USB hard drive is not connected to the imaging sys-
tem PC (the USB dongle is not a USB stick).
• Ensure that the monitor is connected and switched on.
• Ensure that a backup of the current system configuration is on hand in case there is a
reinstallation.
• Ensure that all patients are archived.
• The BIOS CD is on hand.
• Switch the system off.
• Within the WH AWS VB20B the system, the serial number 1022 has to be entered dur-
ing the installation and cannot be modified later!.

Loading the software

Loading the WH AWS BIOS CD

1. Turn ON the PC and press F2 to enter the BIOS menu.


¹ The BIOS login window is displayed.

Fig. 16: Bios_PW


2. Enter the BIOS password and confirm it by pressing the “Enter” key.
3. Press F9 in the BIOS menu to load the default BIOS settings and confirm with YES.
4. Set the current date and time. For SW installation on the PC, type M460, select the
installed hard drive(s) on the BIOS main menu window, and disable the "Noise Man-
agement (Silent Mode)" feature for each installed hard drive.
5. Insert the BIOS CD into the CD-ROM drive.
Press the ESC button and leave the BIOS via Save Changes & Exit. Confirm with
YES. The PC reboots.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 52 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 53

6. The PC boots from the CD-ROM and loads the BIOS version, if required. The follow-
ing messages are displayed.
“BIOS Installation Done - System Ready for Software Installation”
“Please remove BIOS CD!”
“Please turn off or reset system!”
7. Remove the BIOS CD.

Loading the WH AWS VB10B Build 114 software

1. Insert the “System Software VB20B” DVD into the DVD drive.
2. Reboot the PC.
After a short time, a start screen is displayed in text mode. In the top right corner, a mes-
sage indicates that the EIDE, SCSI and CD-ROM drivers are loaded.
If any drivers cannot be found with this hardware configuration, an error message is
generated and can be ignored.
3. Confirm the message “Press any key to boot from CD" and press any key to proceed.
4. After a short time, a message box informs you: “„ Confirm to destroy data on parti-
tion C”. Confirm with “OK” to format drive “C:”.

Fig. 17: Format partition drive C

Siemens SPB7-330.816.01.03.02 Page 53 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
54 WH AWS software installation

5. After a short time, the dialog "COEM Main Setup Mask” will be displayed.

Fig. 18: COEM settings


6. Enter/ select mandatory data as follows:
- Machine Name: e.g.: ZWAWS184
- Partition Settings (radio button): “Keep data on other partition(s)” to “NO”
- Distribution Type: Customer
- UDB- File: CD- ROM
- MUI Installation: No
- Keyboard Layout: US- International
7. Click “OK” to install.
¹ Installation data are copied!
¹ The DVD will be rejected!
¹ The PC will be rebooted!
¹ During the Windows XP Setup, the installation process will be indicated, and the
PC will be rebooted automatically several times!.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 54 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 55

8. When the Internet Explorer opens with “Please enter License”, remove the DVD from
drive and insert the “Bios CD” into the drive again and reboot the PC.
¹ “BIOS Installation Done – HIPAA installed” will be displayed during reboot.
¹ Boot from CD/ DVD is deactivated!

Fig. 19: Please enter license


9. Remove “Bios-CD” from drive.

Siemens SPB7-330.816.01.03.02 Page 55 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
56 WH AWS software installation

10. The Internet Explorer opens with “Please enter License” again.
¹ The installation is successfully finished.
¹ Perform the “initial setup”

Fig. 20: Please enter license


11. Perform the “Configuration” or the “Restore” if a backup is available.

Configuration

NOTE During the initial setup, the CORRECT system serial number
has to entered. This number cannot be modified anymore!

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.
5. Perform the configuration as described in the document “Configuration,
SPB7-330.843.02..., section WH AWS software configuration”.
6. Restart the WH AWS.

Country-specific keyboard layout and user interface

1. In the "Configuration" panel of the user interface double-click “Regional and Lan-
guage”.
2. Select “Regional Options”.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 56 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 57

3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result,


“Apply” is not available.

6. Select the “Languages” tab and click <Details>.


7. Under “Installed services”, click “Add”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “Input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the “Text Services and Input Languages” window.

Siemens SPB7-330.816.01.03.02 Page 57 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
58 WH AWS software installation

12. Click <Apply> and <OK> to close the “Regional and Language Options” window.
¹ If the language has been changed, the following message appears:
The following is displayed if the values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change, click <OK>. This will apply the settings and the sys-
tem restarts.
13. Restart the WH AWS.

Restore

1. Select “Click here to start the Local Service Configuration”.


2. Enter the service key in the service window that appears (under password).
3. Select <Set as default>.
4. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.
5. Insert the backup CD into the DVD drive.
6. In the Service menu, click <Backup & Restore>.
7. Under “Command”, select “Restore” and select “(-R-) CD-R” under “Drives”.

SW-Settings 02

1. Under “Archive”, select the correct (newest) backup file of “SW-Settings 02”
2. Highlight all the files in “Groups” (press the “Shift” key and select the first and last entry
of groups).
3. Click <Go> .

AWS Settings

1. Under “Archive”, select the correct backup file of the “WHAWS Settings”
2. Highlight all the files in "Groups”.
3. Click <Go> .

Security-Settings (only if HIPAA is configured)

NOTE If HIPAA is configured, an HIPAA administrator account is


required. Refer to section “Entering and Saving Data” if
needed.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 58 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 59

1. Under “Archive”, select the correct backup file of “Security Settings”.


2. Click <Go> .
¹ Now the user names and passwords stored in the security settings are valid
again.
¹ Next time, log in with the user name/password obtained from the HIPAA adminis-
trator.

SW-Distribution (optional)

1. Under “Archive”, select the correct backup file of “SW-Distribution”.


2. Click <Go> .

Remove backup CD

1. Remove the backup CD from the DVD drive.

Activating the restored values

1. Restart the system via the Windows menu bar, click <Start>-<Shut Down>, select
<Restart> and click <OK>.
¹ The patient database creation starts if “Create partitions on hard drive 2” was pre-
viously selected, and takes about 15 minutes.
¹ Wait until the patient registration dialog or the patient browser is displayed.
2. Close the patient registration dialog or patient browser.
3. Click the window “Warning: Database...” and press the “Enter” key.
¹ The window “Warning: Database...” disappears.

System Management Registration 0

After the complete WH AWS software installation is finished, the System Management ser-
vices (Siemens Remote Service is activated) has to be registered.
For more details, refer to the document, “Configuration of Siemens Remote Service,
SPB7-330.843.01...
• Perform the System Management registration within the local service/ System Man-
agement/ Agent Control, select the checkboxes, and click “Go” (when Siemens
Remote Service is activated).

Final work steps 0

Patient Images
• If necessary, restore the patient images.

Siemens SPB7-330.816.01.03.02 Page 59 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
60 WH AWS software installation

Second monitor (optional)


• If necessary, reconnect and configure the second monitor as described within the
appropriated monitor document.

Detector calibration
• The Detector Calibration procedure has to be performed for “Contact” and “Magnifica-
tion”. Please see the MAMMOMAT Inspiration user manual for more details on how to
calibrate the system.
¹ Wait at least 30 minutes to warm up the detector.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 60 of 84 Siemens


10.08 CS SD SP/CO
WH AWS software installation 61

Final work steps 2.5

Backup 0

• Perform a backup as described in the document, “Configuration, SPB7-330.843.02...”.

Siemens SPB7-330.816.01.03.02 Page 61 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
62 Mammo Unit
3-
MU firmware installation
3Mammo Unit

This section describes how to install the MammoUnit firmware:

Preconditions: 0

1. WH AWS is connected to the Mammo Unit.


2. The MammoUnit “Flash Master program (FlashMaster_Vxxxx.exe)”, which will install
the stand firmware version.
¹ Backup of the MammoUnit since firmware version VA10C available!

Application System version MU firmware version Backup


Screening VA10B VA00D no backup available
Screening VA10C VA10C Yes
Screening VA10C VA10D Yes
Screening VA10D VB00B Yes, but not compati-
ble to VA00D data
format!
Multiapplication VB10B VB01F Yes

3. Backup file of the MammoUnit.

General Information 0

Example when an error occurs:


¹ Example of error! If an error occurs, click the "Ok" button and continue the update
from work step 1.

Fig. 21: Example of error

NOTE If an error occurs during the update, an error dialog is dis-


played. Follow the instructions in the dialog, restart the
Mammo Unit and start the update from work step 1!
An example of an error dialog is shown in (Fig. 21 / p. 62).

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 62 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 63

Firmware installation procedure 0

1. Switch on the Mammo Unit using the On button on the control box.
2. Perform “Restart application” of the WH AWS and hold down the Shift key until the
Login window appears.
¹ A Login window appears.
3. Log on as administrator.
¹ adm$pwd$4$med
¹ The MammoUnit is started and is now in the "Initial stop" state. This can be veri-
fied by looking at the stand display where the text “MAMMOMAT Inspiration” and
the state “Initial stop”, “Stop”, or “Halt” are displayed.
4. Insert the CD which contains the valid system “FlashMaster_Vxxx.exe” file.
5. Open the Windows Explorer, and double click the R:\FlashMaster_Vxxxx.exe to run
the “Flash master program” (Fig. 22 / p. 63), where “R” is the CD drive.
¹ The program will start and a dialog will be displayed on the computer monitor.

Fig. 22: Flash master dialog


6. Verify that the flash master program states the correct firmware version
(Fig. 23 / p. 63).

Fig. 23: Verify correct firmware version

Siemens SPB7-330.816.01.03.02 Page 63 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
64 Mammo Unit

7. Verify that the IP address is correct (default address is “192.168.0.100”)


(Fig. 24 / p. 64).
¹ The IP address will be also displayed on the MammoUnit display, here shown
with firmware VA00D, (Fig. 25 / p. 64).

Fig. 24: Verify correct IP address


¹ The stand display shows the configured IP address, Mammo Unit firmware ver-
sion and the Mammo Unit status (Fig. 25 / p. 64).

Fig. 25: MU - IP address

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 64 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 65

8. Click on the “Check” button (Fig. 26 / p. 65).


¹ Wait about one minute until the text "Please press start" is displayed in the dialog
(Fig. 29 / p. 66).
¹ The Mammo Unit will be in the standby mode (Fig. 28 / p. 65).

Fig. 26: "Check" button

Fig. 27: Status preparing


¹ The information displayed in the MammoUnit stand display (Fig. 28 / p. 65)
appears there before the "Check" button is pressed. During the "Check" opera-
tion, the information appears as shown in (Fig. 31 / p. 67).

Fig. 28: MU - Standby

Siemens SPB7-330.816.01.03.02 Page 65 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
66 Mammo Unit

9. Verify that the “Check” is successful, the current version of the boards CBS and CBA
are displayed (Fig. 29 / p. 66). If not, restart system and continue with work step 1.

Fig. 29: Start button and version

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 66 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 67

10. Click on the “Start” button to start the firmware update (Fig. 29 / p. 66).
¹ The update will be started, a progress bar is displayed informing that the update is
“Preparing...” (Fig. 30 / p. 67).
¹ Wait while the update is in progress, as shown in the dialog (Fig. 32 / p. 67),
(Fig. 31 / p. 67) and (Fig. 33 / p. 68).
¹ The complete update time is 5 to 10 minutes.

Fig. 30: Status Preparing

Fig. 31: MU - Firmware update

Fig. 32: Status update

Siemens SPB7-330.816.01.03.02 Page 67 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
68 Mammo Unit

Fig. 33: Status update


11. When the update is finished, a dialog is displayed confirming that the update was suc-
cessful (Fig. 34 / p. 68). Click “OK” to close the program.
¹ The update is now successfully finished. The Mammo Unit will restart itself when
the update is completed.

Fig. 34: Update successfully finished


12. Restart the complete system.

Final check 0

1. The system will start without an error message.


¹ IF Error 42 is displayed after rebooting the MammoUnit, the configuration of the
MammoUnit is invalid or incomplete.
In that case, perform the “Restore the MammoUnit configuration” as described in
the next section, or configure the MammoUnit manually.
2. Firmware version Vxxxx will be displayed within the “Local Service/ Acquisition Sys-
tem/ Service Web Interface”.
3. Perform a partial function test of the Mammo Unit movements (Compression, Lift, Rota-
tion).
4. Register a patient and release at least one exposure in -90 degrees and one exposure
in +90 degrees.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 68 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 69

Configuring the MammoUnit manually 3.1

If the MammoUnit configuration is lost, perform the configuration manually.


The table indicates which configuration to perform and where to find the procedure:

Configuration document Information


swivel arm position Adjustments, Perform that adjustment in
SPB7-330.842.01... E_state_STOP, if possible at
0 degree position!
Generator adjustment Adjustments, Perform the main inverter
SPB7-330.842.01... and all filament adjustments!
Collimator adjustment Adjustments, If needed, set the collimator
SPB7-330.842.01... to factory default before start-
ing the adjustment!
Select “WebServiceInter-
face/ Collima-
tor/Details/Reset to Start
Value”!
Single touch Configuration, Default disabled!
SPB7-330.843.02...
AEC Threshhold Configuration, Default 12%!
SPB7-330.843.02...

Siemens SPB7-330.816.01.03.02 Page 69 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
70 Mammo Unit

Backup the MammoUnit configuration (from VA10C) 3.2

Since the MammoUnit Firmware VA10C, the Backup/ Restore functionality of the EEPROM
data is available.
MammoUnit configuration data of MammoUnit firmware VA00D have to configured manu-
ally or adjusted again.
The data of the MammoUnit network “Config IP ” page are not included within this backup
procedure. Note down the data manually.
• Log into “Local service” and select “Acquisition System/ Web Service Interface”.
¹ The “Web Service Interface” opens.

Fig. 35: Web Service main menu


• For MammoUnit Firmware VA10C/ VA10D/ VB00B, select “System” and then
“EEPROM”.
For MammoUnit Firmware Version VB01F, select “Configuration Data”.

Fig. 36: System EEPROM data

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 70 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 71

• Scroll down and click “Backup” in the window.

Fig. 37: Backup


¹ A window pops up to select the destination path of the backup.

Fig. 38: File download


• Click “Save”.
¹ The “Choose File” box will be opened.

Siemens SPB7-330.816.01.03.02 Page 71 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
72 Mammo Unit

• Select the destination path “C:\Temp” or “C:\AWS\MammoUnit” or USB, enter the file
name e.g. “MU_Config_YYYY_MM_DD.mcf” (e.g. MU_Config_2008_06_09.mcf) and
click “Open” to start the backup procedure.

Fig. 39: Configuration file name


• Confirm the message box with “Close” when the file download is completed.

Fig. 40: Download complete

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 72 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 73

• Select “System” and then “Config IP”.

Fig. 41: Web IP Config


• Make a note of the data for the network configuration:

Address Names Default Current


IP Config (IP address of the 192.168.0.100
Mammo Unit)
Network Mask config (net mask of 255.255.255.0
the Mammo Unit)
MAC Config (Mac address of the n.a.
Mammo Unit)
AWS IP Config (IP address of the 192.168.0.2
WH AWS)
AWS MAC Config (Mac address of n.a.
the WH AWS)

Siemens SPB7-330.816.01.03.02 Page 73 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
74 Mammo Unit

Restore the MammoUnit configuration 3.3

This section describes how to restore the MammoUnit configuration data from MammoUnit
firmware version VA10C.
Switch on the system:
- the Mammo Unit will start with default IP address 192.168.0.100
¹ If the default IP address is not used, refer to the section “Appendix/ Additional
Information on the Mammo Unit IP Address”.
- The system may start with error messages Error 42 and lost detector communication!
- The firmware version, e.g. VA10D, will be displayed within the “Local Service/ Acqui-
sition System/ Web Service Interface”.

Preconditions: 0

1. WH AWS is connected to the Mammo Unit.


2. The default IP address 192.168.0.2 is configured within the WH AWS service software,
second network adapter
(refer to section “Additional Information of the the MammoUnit IP address”).
3. The previously made backup of the Mammo Unit exists on C:\temp of the WH AWS file
system or USB drive.

Procedure 0

1. Switch on the Mammo Unit using the On button on the control box.
C ¹ The power supply of the detector must still be unplugged!
2. Perform the “Restart Application” of the WH AWS and hold the “Shift button” of the
keyboard until the login dialog appears.

NOTE The syngo application must be stopped before restoring the


MammoUnit configuration procedure!

3. Login as administrator.
¹ Find the standard password in the knowledge base.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 74 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 75

4. Open Internet Explorer and enter “192.168.0.100” in the URL line.


¹ The “Web Service Interface” opens.

Fig. 42: Web Service main menu


5. For firmware version VA10C/ VA10D/ VB00B, select “System” and then “EEPROM”,
For firmware version VB01F, select “Configuration data”.

Fig. 43: System EEPROM data

Siemens SPB7-330.816.01.03.02 Page 75 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
76 Mammo Unit

6. Within the window scroll down.

Fig. 44: Backup


7. In the Menu, click “Browse...”, select the valid MammoUnit backup file
“MU_Config_2008_MM_DD.mcf” stored in “C:\temp, or C:\AWS\MammoUnit, or the
USB drive of the WH AWS file system and click “Open”.

Fig. 45: Configuration file name

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 76 of 84 Siemens


10.08 CS SD SP/CO
Mammo Unit 77

8. Click “Send/ Restore” to transfer the backup data to the MammoUnit.

Fig. 46: Upload Web configuration


9. Confirm the message box and wait 1 minute until the MammoUnit is rebooted.
10. Refresh the Web Browser with the “F5” function key and select “Back to the Main
Menu”.
11. Within the main menu of the Web Service Interface select “Configuration data” and
ensure that all states are set to “valid”.

Fig. 47: Configuration data


12. Switch the Mammo Unit off and then on and again ensure that all parameters are set to
valid in the “Configuration Data” menu.
13. Restart the complete system.

Configuring the Mammo Unit Network Parameter (IP Address) 0

This section describes configuration of the Mammo Unit network adapter:


1. In the “Web Service Interface”, click “Back to Main Menu” and then “System”.
2. Select “Config IP”.

Siemens SPB7-330.816.01.03.02 Page 77 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
78 Mammo Unit

3. Reconfigure the data previously noted in the section “Backup the MammoUnit configu-
ration”.
- IP address, net mask (if needed) of the Mammo Unit
- MAC address of the Mammo Unit
- IP address and MAC address of the WH AWS

Fig. 48: Web IP Config


4. If required, reconfigure the IP address (second network adapter) in the WH AWS ser-
vice software.
- Log into “Local Service” of the WH AWS, and select “Configuration/ TCP/IP”.
- Select the above second network adapter in the menu (connection to the Mammo
Unit).
- Change the IP address to the one noted in section “Backup the MammoUnit configu-
ration”, “WH AWS IP config”.
- Click “Home”, and confirm the message that the computer will be rebooted.
5. Switch the MammoUnit off and then on.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 78 of 84 Siemens


10.08 CS SD SP/CO
Detector Firmware Update 79
4-
Updating the Detector Firmware System VA10 (Screening)
4Detector Firmware Update

This section describes how to update the detector firmware and the medic using the WH
AWS VB10C or VB10D service software.

Prerequisites 0

• With system restart, system communication to the detector is available, no error mes-
sages should appear.
• The “Test-Image result” CD (Anrad) of the detector is available.
¹ The “Contact.cal” file will be reinstalled after detector firmware update.

Backup the “contact.cal” 0

Prior to updating the detector firmware, ensure that the detector “Test-Image result” CD
(Anrad) is available!
• Check that the “Contact.cal is on CD “Test-Image result”.
¹ If yes, go to section “Procedure”
¹ If not, perform the following:
• Log in to local service.
• Select “Utilities/ Escape to OS”.
• Enter “start explorer” into the command line.
¹ The Windows Explorer will be opened.
• Copy (backup) the “Contact.cal”, and the “mag.cal” to c:\temp.

Procedure 0

1. Switch the system on and wait until the system is completely booted up.
2. Log in to local service.
3. Select “Acquisition System”.
4. Select “Upgrade Firmware”.

NOTE Do not interrupt the update process before the update of the
firmware is completely finished.
Otherwise communication to the detector may be lost and an
initial firmware installation is required.

Siemens SPB7-330.816.01.03.02 Page 79 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
80 Detector Firmware Update

5. Double-click to select the “SystemConfig.upd” from the “C:\AWS\MOSAIC\config” or


from an additional CD (valid detector firmware delivered with the system).

Fig. 49: Detector SW update


6. Click “Upgrade Firmware” to start the update procedure.
¹ Confirm the message box which informs you that the update procedure has suc-
cessfully started.
¹ Ignore the Info line in the service window that the update has been successfully
performed!
7. Wait about 30 minutes until the update process has successfully finished.
8. Confirm the messages box, which informs you that the update is successfully finished.
9. Close the Service menu via “Home”.
10. Using Windows Explorer, copy the “Contact.cal” from the “Test - Image result” of the
detector to “C:\AWS\mosaic\calfiles”.
¹ If the CD is not available ,copy the backed up files of Contact.cal from C:\temp to
C:\AWS\mosaic\calfiles.
11. Using Windows Explorer, copy the “contact.cal” of the “Test - Image result” CD to
“C:\AWS\mosaic\calfiles\mag.cal”.
¹ If the CD is not available, copy the backed up file mag.cal from C:\temp to
C:\AWS\mosaic\calfiles.
12. Restart the complete system.
13. Perform the detector calibration for “Contact” and “Magnification”, if applicable, as
described in the user's manual.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 80 of 84 Siemens


10.08 CS SD SP/CO
Detector Firmware Update 81

Updating the Detector Firmware System VB10 (Multiapplication) 4.1


This section describes how to update the detector firmware and the medic using the WH
AWS VB20B service software.

Prerequisites 0

• With system restart, system communication to the detector is available; no error mes-
sages should appear.
• The “Test-Image result” CD (Anrad) of the detector is available.
¹ The “Contact.cal” file will be reinstalled after detector firmware update.

Backup the “contact.cal” 0

Prior to updating the detector firmware, ensure that the detector “Test-Image result” CD
(Anrad) is available!
• Check that the “Contact.cal is on CD “Test-Image result”.
¹ If yes, go to section “Procedure”
¹ If not, perform the following:
• Log in to local service.
• Select “Utilities/ Escape to OS”.
• Enter “start explorer” into the command line.
¹ The Windows Explorer will be opened.
• Copy (backup) the “Contact.cal”, and the “mag.cal” to c:\temp.

Procedure 0

1. Switch the system on and wait until the system is completely booted up.
2. Log in to local service.
3. Select “Acquisition System”.
4. Select “Upgrade Firmware”.

NOTE Do not interrupt the update process before the update of the
firmware is completely finished.
Otherwise communication to the detector may be lost, and
an initial firmware installation is required.

Siemens SPB7-330.816.01.03.02 Page 81 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
82 Detector Firmware Update

5. Double-click to select the “Bootloader.upd” from the “C:\AWS\MOSAIC\config” or


from an additional CD (valid detector firmware delivered with the system).

Fig. 50: Detector SW update


6. Click “Upgrade Firmware” to start the update procedure.
¹ Confirm the message box which informs you that the update procedure has suc-
cessfully started.
¹ Ignore the Info line in the service window that the update has been successfully
performed!

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 82 of 84 Siemens


10.08 CS SD SP/CO
Detector Firmware Update 83

7. Double-click to select the “SystemConfig.upd” from the “C:\AWS\MOSAIC\config” or


from an additional CD (valid detector firmware delivered with the system).

Fig. 51: Detector SW update


8. Click “Upgrade Firmware” to start the update procedure.
¹ Confirm the message box which informs you that the update procedure has suc-
cessfully started.
¹ Wait several minutes until the installation is successfully completed.
9. Wait about 30 minutes until the update process has successfully completed.
10. Confirm the messages box, which informs you that the update is successfully finished.
11. Close the Service menu via “Home”.
12. Using Windows Explorer, copy the “Contact.cal” from the “Test - Image result” of the
detector to “C:\AWS\mosaic\calfiles”.
¹ If the CD is not available ,copy the backed up files of Contact.cal from C:\temp to
C:\AWS\mosaic\calfiles.
13. Using Windows Explorer, copy the “contact.cal” of the “Test - Image result” CD to
“C:\AWS\mosaic\calfiles\mag.cal”.
¹ If the CD is not available, copy the backed up file mag.cal from C:\temp to
C:\AWS\mosaic\calfiles.
14. Restart the complete system.
15. Perform the detector calibration for “Contact” and “Magnification”, if applicable, as
described in the user's manual.

Siemens SPB7-330.816.01.03.02 Page 83 of 84 MAMMOMAT Inspiration


10.08 CS SD SP/CO
84 Changes to the previous version
5Changes to the previous version 5-

Chapter Changes
General Information Update the recommended document list.
WH AWS software The installation description of the different software
installation installation procedures of VA10 systems have been
updated, and were divided into several sections.
Update the instructions for installing the WH AWS
VB20 software.
MammoUnit Update the firmware installation procedure with all
available firmware versions in the field.
Update the Backup/ Restore functionality for all ver-
sions.
Detector Update the firmware installation procedure with all
available firmware versions in the field.
Changes to previous ver- This chapter has been updated with respect to doc-
sion ument changes.

MAMMOMAT Inspiration SPB7-330.816.01.03.02 Page 84 of 84 Siemens