S5

F8 Multi-parameter Patient Monitor (12.1 inches)


Technical Data Sheet - Release 2.6

The Unicare F8 patient monitor is designed An optional HDMI output eases


to match the pace and unique needs of adult, observation from a long distance.
pediatric, and neonatal intensive care;
anesthesia and peri-operative care; OR and The unique Accessory Box design can
cardiac care environments. keep the accessories in order at the
backside of the monitor.
An optional USB Mouse/Keyboard
control eases navigation and data
entry.

UNICARE
CE/ISO/RoHS/CFDA
Features

Core Alarm
 Newly advanced A9 main board with Linux OS,
 Three-level acousto-optic alarm
calculating speed is 4 times faster than traditional  Sensor-off alarm
products  Paper out alarm
 Support alarm review
 Support storage of 2160 hours trend table and graph  Support alarm pause
review, 2 hours waveform review, 2000 groups NIBP
review and 2000 alarm events review Linux OS
 Support operation with mouse and
Body computer keyboard (option)
 12.1 inch high-brightness TFT LED
 Support display 9~13 waveforms  Multi-display mode to meet different clinical
 Support 7 channel ECG waveform display requirement, including standard interface, big font,
simultaneously OxyCRG, trend graph, NIBP review and full leads ECG
 Optional HDMI output
interface.

Printer  NIBP self-test mode: including overpressure


 Built-in high-speed 50mm thermal printer test, static pressure test and air leakage test.
(Brand: SEIKO, Japan)  Generate ID automatically when register a new patient
 Support medical history search by patient ID, name
Central System and mobile number
 Optional Build-in wireless network module,  18 types of Arrhythmia analysis and real-time S-T
supporting wired or wireless connection to the central segment analysis and pacemaker detection
monitoring station.
 Drug calculation and titration table
 Optional Support HL7 (Health Level Seven)
 Support multi-language display
 Support online software upgrading by net/USB

ECG Full Lead Big Font Display Standard Display

Oxy CRG Trend Table NIBP Review


Interfac Related Product
Transducer socket
 ECG
 SPO2
 NIBP
 IBP 1 / IBP 2
 TEMP 1 / TEMP 2
 Mainstream EtCO2
 AG (multi-gas)
 Sidestream EtCO2  IBP accessories

LAN / Central monitor system interface


 The LAN port is for connecting the monitor to
a central monitor system network.

Input device interface  Respironics EtCO2 module


 This optional interface provides an USB ports to
enable the monitor to be connected to off-the-
shelf input devices:
 Mouse: any specified trackball or USB mouse may be  Nellcor SpO2 module
used for navigation and data entry.
 Computer keyboard: an USB computer keyboard can
be used for data entry instead of the on-screen pop-
up keyboard.
[Remark: user has to restart the machine after plug and  Masimo Spo2 module
unplug the mouse / computer keyboard.]
 Knob (Standard)

Service Features
 The Support Tool helps technical personnel to  Masimo Phasein AG module
– carry out configuration, upgrades and
troubleshooting on an individual monitor.
– back up the monitor settings.

 The Service Mode is password-protected and ensures


that only trained staff can access service tests and tasks.
It includes Skin Type, Brightness Setting, Clear Data
and NIBP Calibration.  CSM Anaesthesia Depth Monitoring
 The Support Tool uses the USB interface of the
monitor for software upgrading.
Performance Specifications
Dimension and Weight
Trend & Reviewing NIBP
- Dimension: 312mm*212mm*168mm
- Trend: 2160 hours - Method: Oscillometric
- Weight: 4 kg(excluding accessories3
- ARR events: 128 groups of ARR events and - Measure mode: Manual, Auto, STAT
Power Supply
associated waveform - Measure Interval in AUTO Mode 1~480 min
- Voltage: AC100~240V, 50/60HZ, Power!60W - NIBP measurement reviewing: 2000 groups - STAT mode cycle time: Keep 5 minutes, at 5 seconds
Display - Waveform review: 2 hours interval
- 12.1" color TFT LED - Alarm event: 2000 groups of parameter alarms events - Measure and Alarm Range:
- Resolution: 800*600 pixels and associated parameter Adult: SYS: 30 ~ 270 mmHg
Battery (Pluggable) ECG DIA: 10 ~ 220 mmHg

- Type: Rechargeable lithium battery 14.8V/2200mAh - Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, MEAN: 20 ~ 235 mmHg

- Charge Cycle: "500 times AVL, AVF, V Pediatric: SYS: 40 ~ 235 mmHg

- Working time: 2 hours (optional second-battery for - Protection: Breakdown Voltage 4000VAC 50/60Hz; DIA: 10 ~ 220 mmHg

4-5 hours) Defibrillator proof MEAN: 20 ~ 225 mmHg

Recorder (Option) - Gain: 2.5mm/mV(#0.25), 5.0mm/mV(#0.5), 10mm/mV Neonate: SYS: 30 ~ 135 mmHg

- Method: Thermal printer (#1), 20mm/mV (#2), 40 mm/ mV (#4), Auto DIA: 10 ~ 110 mmHg

- Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s MEAN: 20 ~ 125 mmHg


- Paper width: 50 mm (1.97 in)
- Printing speed: 12.5/ 25/ 50 mm/s - ECG signal range: ±12 mV (Gain # 0.25) - Static pressure accuracy: ±3mmHg

- Trace: Max. 3 tracks - Accuracy: ±1bpm/ ±1%, whichever is greater - Resolution: 1mmHg

- Resolution: 1 bpm - Accuracy: Maximum Mean error ±5mmHg


- Recording way: Real-time Recording, Review
Printing, - Leakage Current < 10 µA Maximum Standard deviation !8mmHg
Periodic Recording, Alarm Recording - Baseline Recovery: - Over pressure Protection: Dual protection via
Alarm ! 3s after defibrillation (Monitor mode) software & hardware

- Level: Low, medium and high ! 1s after defibrillation (Surgery mode) Suntech NIBP
- Indication: Auditory and visual - Bandwidth: Surgery 1 ~ 25 Hz - Measure Method: Oscillometric. Diastolic values
- Alarm volume adjustable Monitor 0.5 ~ 40 Hz correspond to Phase 5 Korotkoff sound
- Alarm pause time: 1min, 2min Diagnostic 0.05 ~ 120 Hz
- Measure Range:
- Parameter alarm type: Latch/ Unlatch - Indication of Electrode Separation: Every electrode
Adult: SYS: 40 ~ 260 mmHg
Input Device (exclusive of RL)
DIA: 20 ~ 220 mmHg
- Knob (Standard) Heart Rate
MEAN: 26 ~ 220 mmHg
- Keypad input (Standard) - Measure range:
Pediatric: SYS: 40 ~ 230 mmHg
- Mouse/ Keyboard input (option) Adult: 0bpm; 10 ~ 300 bpm
DIA: 20~ 160 mmHg
System Output & Extensible Neo/Ped: 0bpm; 10 ~ 350 bpm
MEAN: 26 ~ 183 mmHg
Interface - Resolution: 1 bpm
Neonate: SYS: 40 ~ 130 mmHg
- Accuracy: ± 1% or ± 1bpm, whichever is greater
- Ethernet Network: 1 standard RJ45 socket DIA: 20 ~ 100 mmHg
- USB Port: 1
ST Measurement
MEAN: 26 ~ 110 mmHg
- Video Output: 1 HDMI port (option) - Range: -2.0 ~ +2.0 mV
- PR Range: 30 to 220 BPM
Operating Environment - Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or ±10%,
- PR Accuracy: ±2% or ±3 BPM, whichever is greater
whichever is greater
- Temperature: 5 ~ 40 ? - Pressure Transducer Accuracy: ±3mmHg between 0
- Other range: unspecified
- Humidity: 15% ~ 90% (non-condensing) mmHg and 300 mmHg
- Resolution: 0.01mV
- Atmosphere pressure: 86 KPa ~ 110 KPa for operating conditions between 0� and 50�
Respiration
Transportation and Storage - Deflate Rate: Deflation step size varies with heart
- Method: Impedance between RA-LL, RA-LA
Environment rate, cuff pressure and cuff volume
- Gain: #0.25, #0.50, #1, #2, #4
- Temperature: -20~50 ? - Initial Inflation Pressure:
- Respiration Rate: 0bpm, 6 ~ 150 BrPM
- Humidity: 10%~90% (non-condensing) ADULT: 160 mmHg (default)
- Sweep speed: 6.25 mm/s, 12.5 mm/s, 25mm/s
- Atmosphere pressure: 86 KPa ~ 110 KPa Variable from 120 to 280 mmHg
- Resolution: 1 BrPM
Safety PEDIATRIC: 140 mmHg (default)
- Accuracy: ±2 bpm or ±2%, whichever is greater
- IEC60601-1 Approved, CE marking according to Variable from 80 to 280 mmHg
- Apnea Alarm: 10 ~ 60 s
MDD93/42/EEC NEONATE: 90 mmHg (default)

- With reference to RoHS Directive 2011/65/EU Variable from 60 to 140 mmHg

recasting
Performance Specifications
Temperature
Pulse Rate Multi-gas/O2 (Anesthetic Gas)
- Technique: Thermistor probe (2.25K)
- Measurement range: 25 ~ 240 bpm (Option)
- Channel: Dual-channel, provide T1; T2; !T
- Resolution: 1bpm - Method: Infrared absorption
- Measuring and Alarm Range:
- Accuracy: ±3 bpm (non-motion condition) - Gas sorts: CO2, N2O, Des, Iso, Enf, Sevo, Hal,
0.0 ?~ 50 ? (32fi~ 122fi)
O2 (Optional paramagnetic sensor)
- Unit: Celsius (?), Fahrenheit ( fi)
EtCO2 (Mainstream/ Sidestream) - Calibration: Room air calibration performed
- Resolution: 0.1?or 1fi
(Option) automatically when changing airway
- Accuracy: ±0.1?(exclusive probe)
Airway adapt (< 5 sec)
SpO2 (Digital Technic) - Measure method: Non-dispersive infrared (NDIR)
- Measurement range:
- Measurement Range: 0 ~ 100 % - Measure Range: 0 ~19.7% (0 ~ 150 mmHg)
CO2: 0~25%, N2O : 0~100%
- Resolution: 1 % 0 ~ 20 kPa
O2: 0~100%, Enf, Iso, Hal: 0~25%,
- Response Modes: Low, Medium, High - Resolution: 0.1 mmHg
Sevo, Des: 0~25%
- Accuracy: ±2% (70% ~ 100%) - CO2 Accuracy:
- Data output: Fi and ET values
±3 % (35% ~ 69%) 0 ~ 40 mmHg, ±2 mmHg
- Respiration rate: 0~ 150 BrPM
Unspecified (0 ~ 34%) 41 ~ 70 mmHg, ±5% of reading
- Other: Up to 5 waveforms displayed
- Support Pitch tone and multi-level volume 71 ~ 100 mmHg, ±8% of reading
Agent mixture detection
- User-selectable waveform speed: 101~ 150 mmHg, ±10% of reading
MAC value displayed
6.25, 12.5, 25, 50 mm/s at 760 mmHg, ambient temperature of 25?)
- Respiratory Rate: Range: 3 ~150 BrPM
IBP (Option)
Pulse Rate
Accuracy: ±1 bpm - Max Channel: 2
- Measuring and Alarm Range: 25~250bpm
EtCO2 (Micro-stream) (Option) - Measurement way: Thermal resistance way
- Accuracy: ±1% or ±1 bpm, whichever is greater
- Press Sensor: Sensitivity 5 uV/V/mmHg, ±2%
- Resolution: 1bpm - Measure method: Non-dispersive infrared (NDIR)
Impedance 300 to 3000!
- Measure Range: 0 ~19.7% (0 ~150 mmHg)
- Resolution: 1 mmHg
0 ~ 20 kPa
Nellcor-SpO2 (Option) - Unit: mmKg, kPa, cmH2O
- Sample Rate: 50 mL/min ±10mL/min
- Measurement Range: 0 ~ 100 % - Transducer sites:
- Resolution: 0.1 mmHg (0 ~ 50 mmHg)
- Resolution: 1 % Arterial Pressure (ART)
0.25 mmHg (50 ~ 114 mmHg)
- Accuracy: 70% ~ 100%, ±2 % (adult) Pulmonary Arterial (PA)
- CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg
70% ~ 100%, ±3 % (Neonate) Left Arterial (LAP)
41 ~ 70 mmHg, ±5% of reading
70% ~ 100%, ±2 % (Low Perfusion) Right Arterial (RAP)
71 ~ 100 mmHg, ±8% of reading
0% ~ 69%, unspecified Central Venous Pressure (CVP)
101~ 150 mmHg, ±10% of reading
Pulse Rate Intracranial Pressure (ICP)
at 760 mmHg, ambient temperature of 35?)
- Measurement range: 20 ~ 300 bpm P1/ P2
- Respiratory Rate: Range: 3 ~120 BrPM
- Resolution: 1bpm - Measuring and alarm range:
Accuracy: ±1 bpm
- Accuracy: ART 0 ~ +350mmHg
Depth of Anesthesia (CSI) (Option)
±3 bpm (20 ~ 250 bpm) PA -10 ~ +120
- EEG sensitivity: ± 400µV
unspecified (251~300 bpm) mmHg CVP/ RAP/ LAP/ ICP -10 ~ +40
- Noise: < 2µVp-p, < 0.4µV RMS, 1-250 Hz
Masimo SpO2 (Option) mmHg P1/ P2 -50 ~ +350mmHg
- CMRR: > 140 dB
- Accuracy:
- - Measurement Range: 0 ~ 100 % - Input impedance: > 50 Mohm
Static: ±1mmHg or ±2%, whichever is greater
- - Resolution: 1 % - Sample rate: 2000 samples/sec, (14 bits equivalent)
(exclusive of transducer)
- Accuracy: - BS%: 0-100, filter 1-42 Hz, 1 sec. display update
±4mmHg or ±4%, whichever is greater
70% ~ 100%, ±2 % (adult/ pediatric, non-motion - EMG: 0-100 Logarithmic. Filter 75-85 Hz, 1 sec.
(inclusive of transducer)
conditions) update
Dynamic: ±4mmHg or ±4%, whichever is
70% ~ 100%, ±3 % (neonate, non-motion - Alarms: High / Low with user selectable limit
greater
conditions) - Artifact rejection: Automatic
70% ~ 100%, ±3 % (motion conditions) - Sensor impedance range: 0 - 10 kOhm / measurement
0% ~ 69% unspecified * Specifications subject to change without prior notice
current 0.01 µA
Model Configuration
Standard Config. 12.1 inch LED, 3/5 Lead ECG, NIBP, SpO2, Pulse Rate, Temperature, Respiration

Optional Config. 1/2 IBP, 2-Temperature, Nellcor SpO2, Masimo SpO2, Suntech NIBP,
EtCO2 (Mainstream/Sidestream/Microstream),
Anesthesia Depth Monitoring (CSM module),
Multi-gas (AG) Monitoring

Optional Accessories: Printer, USB Mouse/Computer Keyboard Input Function, Touch Screen (Resistive type)
Central Monitoring Station,
Neo/Ped Accessories,
HDMI Output, Wall-mounting, Trolley,

Remark: 1) User can NOT choose USB mouse function and WIFI at the same time.

Gallery for Optional Accessories:

N#$$c&r )*+#nsi&n /a1$# 2#3sa1$# N#$$c&r N#& 4r&1# 2#3sa1$# 4#d6 7892 4r&1# ;is8&sa1$# 4#d 7892
;is8&sa1$# N#& 7892 4r&1#
4r&1#

<=4 ;is8&sa1$# >i+s H;@< +& ABC +ransD#r+#r /#n+ra$ @&ni+&rinE 7+a+i&n Fr&$$#G Hi+I =asJ#+ Ka$$LM&3n+ Hi+I 1asJ#+

4rin+#r CB @ains+r#aM 4r&1# CB <2@C CirHaG Cda8+#r


CB 7id#s+r#aM Cna$Gs#r ;is8&sa1$# )$#c+r&d#s &P
NCd3$+O<nPan+Q
Cna#s+I#sia ;#8+I

@&ni+&rinE
Standard Packing List:

Standard Packing List


No. Parts No. Description QTY
1 8189007070 5-lead snap connector ECG cable (AHA standard) 1
2 8189007002 Adult disposable adhesive electrodes (snap connector, 25pcs/pack) 1
3 8310400014 Unicare reusable adult SpO2 finger sensor 1
4 8189007010 NIBP PVC hose (3M), with connector 1
5 8189007074 Reusable adult NIBP cuff with connector (25cm-35cm), Sac-type 1
6 8189007072 Skin temperature probe with plug 2.25K 1
7 8112900002 Power cord 1
8 8310100008 Ground cable 1
9 8510100001 User's manual 1
10 8510120002 Checking list 1

Options:
No. Parts No. Description
ECG
1 8189007070 5-lead snap connector ECG cable (AHA standard)
2 8189007064 5-lead ECG cable (crocodile type) (AHA standard)
3 N/A 5-lead snap connector ECG cable (IEC standard)
4 N/A 5-lead ECG cable (crocodile type) (IEC standard)
5 8189007002 Adult Disposable Adhesive Electrodes (Snap Connector, 25pcs/pack)
6 8189007003 Pediatric/Neonatal Disposable Adhesive Electrodes (Snap Connector, 25pcs/pack)
Nellcor SPO2
7 8310300058 Nellcor SpO2 Module, extension cable 1pc and adult sensor 1pc
8 8189007075 Nellcor reusable adult SpO2 finger sensor
9 8189007059 Nellcor reusable neonatal SpO2 wrapping sensor
10 8189007060 Nellcor disposable neonatal SpO2 sensor
11 8310400031 SpO2 extension cable with Nellcor connector
Masimo SPO2
12 N/A Masimo SpO2 Module, extension cable 1pc and adult sensor 1pc
13 N/A Masimo reusable adult SpO2 finger sensor
14 N/A Masimo reusable neonatal SpO2 wrapping sensor
15 N/A Masimo disposable neonatal SpO2 sensor
16 N/A Masimo extension cable with Masimo connector
Unicare SPO2
17 8310400014 Unicare reusable adult SpO2 finger sensor
18 8189007024 Unicare reusable pediatric SpO2 finger sensor
19 8189007029 Unicare reusable neonatal SpO2 wrapping sensor
20 8189007060 Unicare disposable neonatal SpO2 sensor
21 8189008003 Unicare reusable SpO2 adult soft sensor
22 8189008051 Unicare reusable SpO2 pediatric soft sensor
23 N/A Unicare reusable SpO2 ear clip sensor
24 N/A Unicare reusable Y-Type SpO2 sensor
NIBP
25 8189007063 Reusable neonatal cuff with connector (10cm~15cm), Sac-type
26 8189007047 Reusable pediatric cuff with connector (18cm~26cm), Sac-type
27 8189007074 Reusable adult NIBP cuff (4) with connector (25cm~35cm), Sac-type
28 8189007069 Reusable adult NIBP cuff (5) with connector (33cm~47cm), Sac-type
29 8189007073 Reusable adult NIBP leg cuff (6) with connector (46cm~66cm), Sac-type
30 8189007028 Disposable neonatal NIBP cuff
31 8189007010 NIBP PVC hose (3M), with connector
Temperature
32 8189007005 Rectal probe
33 8189007071 Esophageal probe
34 8189007004 Reusable skin probe, 10k
35 8189007072 Reusable skin probe, 2.25k
IBP
36 81800100078 2-IBP module with accessories kit (Pressure transducer interface cable with connector 2pcs;Disposable
pressure transducer kit 2sets)

37 8189007033 UTAH IBP transducer interface cable with connector (2 sets!


38 8189007032 UTAH IBP disposable transducer (PT-01) (Made in china)
Main-stream ETCO2
39 8180100059 USA Respironics Main Stream EtCO2 probe with Disposable airway adapter - Adult / Pediatric 1pc
40 8189008007 USA Respironics Main Stream EtCO2 sensor
41 8189008008 Disposable airway adapter - Neonatal/Infant
42 8189008009 Disposable airway adapter - Adult / Pediatric
Side-stream ETCO2
43 8180100064 USA Respironics Side Stream EtCO2 Module with accessories kit (bracket 1pc, water trap 1pc, Disposable
CO2 tube 1pc, three way tap 1pc)

44 8189008010 Bracket
45 8189008011 Water trap
46 8189008012 Disposable CO2 sample tube
47 8189008013 Three way tap
Side-stream ETCO2 (China module)
48 8180100067 China Module Side Stream EtCO2 Module with accessories kit (bracket 1pc, water trap 1pc, Disposable CO2
tube 1pc, three way tap 1pc)

49 8189008027 Bracket
50 8189008028 Water trap
51 8189008029 Disposable CO2 sample tube
52 8189008030 Three way tap
Micro stream ETCO2
53 8180100065 Micro Stream EtCO2 Module with accessories kit (Disposable CO2 Nasal Cannula -adult 1 pc, Disposable
CO2 Airway Adapter Cannula -adult 1 pc )
54 8189008014 Micro Stream EtCO2 sensor
55 8189008015 Disposable CO2 Airway Adapter - Adult
56 8189008016 Disposable CO2 Nasal Cannula - Infant
57 8189008017 Disposable CO2 Nasal Cannula - Pediatric
58 8189008018 Disposable CO2 Nasal Cannula - Adult
Multi-gas (main Stream)
59 8189008019 Multi-gas Probe with accessories kit (Disposable airway adapter Adult/pediatric 1pc)
60 8189008020 IRMA AX+ Multi-Gas sensor (Main Stream)
61 8189008021 Disposable airway adapter - Neonatal/Infant (10pcs/pack)
62 8189008022 Disposable airway adapter - Adult / Pediatric (25pcs/pack)
63 8189008023 Extension cable
Multi-gas (side stream)-ISA OR+
64 8189008019 Multi-gas Probe with accessories kit (Disposable Nomoline Sampling line 1pc)
65 8180100066 ISA OR+ Multi-Gas Module (Side Stream, Paramagnetic O2)
66 Servomex Paramagnetic Oxygen Sensor (note: needs calibration when paried with ISA)
67 8189008024 Nomoline Sampling line with Luer Lock connector, 2 m.
68 8189008025 ISA module Modura holder
69 8189008026 ISA module Clamp Adapter
Multi-gas (side stream)-ISA AX+
70 Multi-gas Probe with accessories kit (Disposable Nomoline Sampling line 1pc)
71 ISA AX+ Multi-Gas Module (Side Stream, CO2,N20, 5 x AA, Auto Agent ID)
72 8189008024 Nomoline Sampling line with Luer Lock connector, 2 m.
73 8189008025 ISA module Modura holder
74 8189008026 ISA module Clamp Adapter
CSM (Depth of Anesthesia Awareness Index)
75 N/A Cerebral state monitor module with accessories kit (disposable electrode 1 set)
76 N/A Disposable electrode 1set
Others
77 8600000007/2 Central Monitoring System (software + dongle)
78 8510110001 Service Manual
79 8112900002 Power cord
80 8212010010 Rechargeable Lithium-ion Battery
81 8700010103 Net cable for Wire Central Monitoring (5M)
82 9100010003 Switch Board
83 8490000031 Wireless AP (transmitter & receiver) (Four patient monitors need one set of transmitter & receiver)
84 8490000030 Wireless AP (transmitter) (Each monitor needs one transmitter)
85 8310500003 Built-in thermal printer
86 8111810002 Thermal printing Paper 50mm
87 8212900053 Wall mounting rack
88 8490000001 Trolley with basket and wheels
Remark: Multi-gas has included CO2 monitoring, if you choose Multi-gas module, EtCO2 module is needless.

Shenzhen Unicare Electronic Co., Ltd.


Add: 3D, Building4, Tian-an Industrial Zone, Nan You,
Nanshan District, Shenzhen, 518054, P.R. China
Tel: +86-(0) 755-3336 9088 / 3336 2088 / 3330 3188
Fax: +86-(0) 755-3336 6818
Email: [email protected]
Web: www.unicare-med.com/en

© 2017 Shenzhen Unicare Electronic Co., Ltd. All

rights are reserved.

Unicare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or

obligation and will not be liable for any consequences resulting from the use of this publication.

Printed in The People’s Republic of China.

January, 2017

PT. Fortuna sembada makmur medika


Kompleks Agung Sedayu Blok K No 28
JL Auto Ringroad Lingkar Luar Barat
Cengkareng,Jakarta Barat,11730 DKI Jakarta ,indonesia.

www.fortunasembada.com E-mail : [email protected] Telp. (021) 22956888