Sample Laboratory Report - Fluorescence Activity
Sample Laboratory Report - Fluorescence Activity
FLUORESCENCE SPECTROSCOPY
1 ABSTRACT
2 INTRODUCTION
energy states and are characteristic of the molecular structure. Fluorescence occurs when
a molecule absorbs photons from the UV - Visible light spectrum (200-900 nm), causing
transition to a high-energy electronic state and then emits photons as it returns to its
initial state, in less than 10-9 sec. (Fink, 1982)
The emission spectrum provides information for both qualitative and quantitative
analysis. As shown in Figure 1, when light of an appropriate wavelength is absorbed by a
molecule (i.e., excitation), the electronic state of the molecule changes from the ground
state to one of many vibrational levels in one of the excited electronic states. The excited
electronic state is usually the first excited singlet state, S1 (Figure 1). Once the molecule
is in excited state, relaxation can occur via several processes. Fluorescence is one of these
processes and results in the emission of light. (Skoog, 2004)
where Efluor is the energy of the emitted light, E abs is the energy of the light absorbed by
the molecule during excitation, and E vib is the energy lost by the molecule from
vibrational relaxation. The Esolv.relax term arises from the need for the solvent cage of the
molecule to reorient itself in the excited state and then again when the molecule relaxes
to the ground state. As can be seen from Equation (1), fluorescence energy is always less
than the absorption energy for a given molecule. Thus the emitted light is observed at
longer wavelengths than the excitation (Lakowicz, 1983).
Fluorescence quantitative analysis can be made for both excitation and emission
spectrum. In this experiment, fluorescence quantitative analysis was performed using
Rhodamine B. Rhodamine B belongs to a family of fluorone dyes that all share the same
basic fluorone skeleton (Figure 2) that is commonly used as an active medium in tunable
lasers due to its high fluorescence quantum yield. It also finds a wide range of
applications in materials science, chemistry and biology as a sensitizer in solar cells, as a
molecular probe, as an electrochemical luminescence sensitizer, as a water-tracing agent,
as a biological stain, etc. (Setiawan, 2010)
Pipet, 25 mL and 10 mL
Beaker, 250 mL
Volumetric flasks, 50 mL and 100 mL
Stirring rod
Wash Bottle
From the above secondary stock solution, a 100 mL of 0.8 ppm Rhodamine B
solution was prepared by appropriate dilution in a 100 mL volumetric flask. Then from
the 0.8 ppm, the remaining standards (0.6, 0.4, 0.2 ppm) were prepared by serial
dilutions, all in 100 mL volumetric flasks. Each of these standard solutions was carefully
transferred into a properly labelled, clean amber bottle.
All fluorescence intensity readings were determined using the instrument JASCO
FD-777 Spectrofluorometer. Initially, the instrument was pre-heat for about fifteen (15)
minutes prior to use. A peak search on Rhodamine B was performed using 0.6 ppm
6
standard Rhodamine B solution. The excitation spectrum of the 0.6 ppm standard
Rhodamine B solution was taken. From the excitation spectrum, the λ ex was determined.
The emission spectrum was also taken and λem was determined from the obtained
emission spectrum. Then, the fluorescence intensity of each of the standard solutions of
Rhodamine B (0.2, 0.4, 0.6, 0.8 ppm) was measured. Three cycles for each trial was
performed and three trials were done for intensity measurements of each standard
solution. The fluorescence of the blank (distilled water) was also determined. The
obtained blank’s intensity corrected the measured intensity of fluorescence of the
standard solutions. A calibration curve (concentration vs. fluorescence intensity) was then
constructed. The intensity of the unknown Rhodamine B solution was also taken and the
concentration of the unknown was calculated based on the constructed calibration curve.
Pipet, 25 mL and 10 mL
Buret, 25 mL
Beaker, 250 mL
Volumetric flasks, 100 mL
Test tubes, 10 mL
Stirring rod
Vortex
Wash Bottle
several times with distilled water. Stock solutions were kept in a clean 250mL amber
bottle.
From previously prepared secondary stock solution (section 3.1.2.2), the 0.8 ppm
Rhodamine B solution was prepared by quantitatively transferring 8 mL of secondary
stock (10 ppm) into 100-mL volumetric flask and diluting it to mark with distilled water.
The resulting solution was shaken and stored into a properly labelled, clean amber bottle.
Initially, all reagents as stated above were prepared. A set of 16 solutions (Table
1) with different ratios of Rhodamine B and Hg(II) was then prepared.
8
The above mentioned 16 solutions were prepared through a burette and were
contained in labelled test tubes. Fluorescence intensity was determined for each solution.
A calibration curve of ppm Hg2+ (x-axis) against fluorescence intensity (y-axis). The
fluorescence intensity of the unknown solution containing Hg2+ was also determined. The
unknown solution was treated as solution no.16 where its composition ratio is 1:4
(volume of Rhodamine B:volume of unknown solution). From the plot, the
concentration of the unknown Hg2+ solution was determined.
Micropipet, 1000 µL
Pipet, 25 mL and 10 mL
Beaker, 250 mL
Buret, 25 mL
9
4.1 Data
Table 2. Obtained Wavelength for the Peak Search on Excitation and Emission
Spectrum of Rhodamine B.
Spectrum Wavelength with Maximum Peak Height, nm
Excitation 452
Emission 575
11
Using the wavelengths obtained during peak search, the fluorescence intensities
were determined for the working calibration curve as well as for the blank and unknown
solution and were shown in the Table 3.
Again, emission and excitation spectra were initially scanned. Peak search results
were tabulated in Table 4.
Table 4. Obtained Wavelength for the Peak Search on Excitation and Emission
Spectrum of Rhodamine B.
Spectrum Wavelength with Maximum Peak Height, nm
Excitation 450
Emission 575
Using the obtained wavelengths, the fluorescence intensities were measureed for
the working calibration curve as well as for the blank and unknown solution and were
indicated in the Table 5.
12
At λex and λem indicated in the procedure which is 420 nm and 530 nm
respectively, fluorescence intensities were obtained and indicated in Table 6.
13
Table 6. Measured fluorescence intensity at λex, 420 nm and λem, 530 nm.
Fluorescence Intensity
Solutions Trial Average
cycle 1 cycle 2 cycle 3
Blank 1 0.592 0.592 0.467 0.550
1 6.802 6.770 6.739 6.770
0.5 ppm Mg2+ 2 5.365 5.303 5.21 5.293
3 4.336 4.274 4.275 4.295
1 5.460 5.397 5.367 5.408
1.0 ppm Mg2+ 2 5.241 4.774 4.742 4.919
3 5.147 4.742 4.772 4.887
1 5.147 5.147 5.147 5.147
1.5 ppm Mg2+ 2 4.866 4.866 4.928 4.887
3 4.772 4.773 4.679 4.741
1 5.022 5.085 5.147 5.085
2.0 ppm Mg2+ 2 4.866 5.179 5.116 5.054
3 4.835 4.928 4.836 4.866
1 4.992 5.024 4.992 5.003
3.0 ppm Mg2+ 2 4.835 4.929 4.897 4.887
3 4.991 4.898 4.929 4.939
Unknown 1 4.024 4.055 4.117 4.065
2 4.151 4.306 4.181 4.213
3 3.900 3.774 3.868 3.847
4.2 Calculation
For the preparation of standard solutions, serial dilution was used and to
determine the exact volume needed simple dilution formula was employed:
M1V1 = M2V2
B in the solution. As such to prepare 0.8 ppm Rhodamine B solution from secondary
stock (10 ppm), the calculation of needed volume was shown below:
M1V1 = M2V2
V1 = 8 mL
Same calculation has been done for the rest of the standard solutions. The calculated
volumes required for the preparation of the standard solutions were indicated in Table 7.
All fluorescence intensity observed was at first corrected by the intensity of blank
solution.
The same calculation has been made for the rest of the standard solutions and corrected
fluorescence intensities were tabulated below (Table 8). Then, a calibration curve (Figure
4) was constructed by plotting the concentration of Rhodamine B against the corrected
fluorescence intensity.
From the equation of the working calibration curve, the concentration of unknown
Rhodamine B solution was determined. The obtained fluorescence intensity for the
unknown Rhodamine B solution was 53.96. Substituting this value (y-variable) to the
equation will finally yield the concentration (x-variable) of the unknown as shown below:
y = 59.513x - 0.444
x = 0.9142
At first, a primary stock solution of Hg2+ which concentration is 100 ppm. The
calculation below showed the exact amount of HgCl 2 quantitatively weighed for the
preparation of primary stock solution.
Hg 2+¿ 1
weight of HgCl2=100 mg x Total volume of solution x ¿
mL 2+¿ 1 mmol HgCl 2
MW , Hg x ¿
1 mmol Hg2+¿ x MW , HgCl 2 ¿
2+¿
Hg mmol
weight of HgCl2=100 mg x 0.100 L x ¿
L 2+ ¿ 1mmol HgCl2
200. 59 mg Hg x ¿
2+¿ 271.496 mg
1 mmol Hg x ¿
mmol HgCl 2
weight of HgCl2=13.53 mg
Then, the calculation below illustrated how 5 ppm of Hg 2+ was prepared from 100 ppm
primary stock solution via simple dilution method:
M1V1 = M2V2
M1V1 = M2V2
V1 = 5 mL
Initially, the concentration of Hg2+ on each test tube was determined using the
simple dilution formula:
M1V1 = M2V2
16
M1V1 = M2V2
M2 = 0.5 ppm
Same calculation has been done for the rest of the solutions and calculated Hg 2+
concentrations were tabulated below. Next, all fluorescence intensity were corrected by
the intensity of the blank solution which is distilled water. All corrected intensity was
tabulated below (Table 10).
y = -0.4217x + 10.732
x = 0.5502 ppm
This obtained value will be further corrected by the dilution factor in such a way that the
final concentration of Hg2+ in the unknown solution will be:
¿¿
¿¿
mg 8−hydroxyquinoline
weight of 8−hydroxyquinoline=100 x Total volume of solution
L
mg 8−hydroxyquinoline
weight of 8−hydroxyquinoline=100 x 0.250 L
L
mg MgCl 2 • 6 H 2 O
weight of MgCl 2 •6 H 2 O=210 x Total volume of solution
L
mg MgCl 2 • 6 H 2 O
weight of MgCl 2 •6 H 2 O=210 x 0.100 L=21.0 g MgCl 2 •6 H 2 O
L
¿¿¿
¿¿¿
¿¿¿
Next, the volume of MgCl2•6H2O needed for the preparations of the solutions with
known concentrations of Mg2+ was then calculated using simple dilution formula as
shown below:
M1V1 = M2V2
M1V1 = M2V2
V1 = 122 µL
Next, all measured fluorescence intensities were corrected by the intensity of the blank
which is the distilled water. That is for the solution containing 0.5 ppm of Mg 2+ (trial 1),
its corrected fluorescence intensity is
The same calculation has been made for the rest of the calibration solutions and for the
unknown solution as well. All corrected fluorescence intensities were tabulated below
(Table 13). Then, a calibration curve was constructed by plotting concentration of Mg2+
(x-axis) against the fluorescence intensity (y-axis) (Figure 6)
From the equation of the working calibration curve, the concentration of unknown
Mg2+ solution was determined. The obtained fluorescence intensity for the unknown Hg2+
solution was 3.491. Substituting this value (y-variable) to the equation will finally yield
the concentration (x-variable) of the unknown as shown below:
y = -0.3477x + 5.2508
x = 5.060 ppm
This obtained value will be further corrected by the dilution factor in such a way that the
final concentration of Mg2+ in the unknown solution will be:
¿¿
¿¿
The figure below revealed the excitation and emission spectra of Rhodamine B.
Then, calculated volumes for the preparation of standard solutions via serial
dilutions were listed in the table below.
Table 7. Calculated volumes for the preparation of standard solutions via serial dilutions.
Standard Solutions, ppm Calculated Volume
0.8 8 mL of secondary stock solution (10 ppm)
0.6 75 mL of 0.8 ppm solution
0.4 66.7 mL of 0.6 ppm solution
0.2 50 mL of 0.4 ppm solution
20
50
45 f(x) = 59.5133333333334 x − 0.444000000000003
40 R² = 0.997562187816178
35
30
Corrected Flu- 25
orescence In-
tensity 20
15
10
5
0
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
Concentration of Rhodamine B, ppm
Table 10. Calculated Hg2+ Concentration and its Corrected Fluorescence Intensity
Corrected Averaged
Concentration
Solution Fluorescence Fluorescence
of Hg2+, ppm
Intensity Intensity
1 10.58
2 0.0 11.07 10.92
3 11.09
4 10.38
10.36
5 0.5 9.888
6 10.80
7 10.67
10.07
8 1.0 9.682
9 9.848
11 9.557 9.826
2.0
12 10.05
13 9.305
14 3.0 9.557 9.511
15 9.670
16 4.0 8.962 8.962
12
Fluorescence 6
Intensity
4
0
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
Hg2+ concentration, ppm
Using the above working calibration curve finally determined the concentration of
Hg2+ in the unknown solution. Result was tabulated in the table below.
Table 12. Calculated Weight for the Prepartion of 8-hydroxyquinoline and MgCl 2•6H2O
Solutions.
Reagent Calculated weight required to prepare the solution, grams
8-hydroxyquinoline 25.0 x 10-3
MgCl2•6H2O 21.0
23
5
f(x) = − 0.347720720720722 x + 5.25081981981982
R² = 0.559623098517589
4
Fluorescence 3
intensity
2
0
0 0.5 1 1.5 2 2.5 3 3.5
concentration of Mg2+, ppm
As shown above in Figure 3, two spectra were obtained namely Excitation and
Emission Spectra. These spectra made the fluorescent compounds or fluorophors such as
Rhodamine B to be identified and quantified on the basis of their excitation and emission
properties. It is very prominent in Figure 3 that Rhodamine B has detectable emission
intensity for a broad excitation range (440- 460 nm). Maximum emission occurs at a
unique excitation wavelength of 452 nm. Emitted light is detected for a broad
wavelength range (500–600 nm), however, when excited at 452 nm, maximum emission
occurs at 550 nm. These obtained excitation and emission properties of Rhodamine B are
fixed and can be used for identification and quantification. It is also necessary to conduct
peak search for both spectra since at the absorption maximum the maximum number of
light quanta is absorbed and therefore the maximum number of molecules is excited per
unit of time. It must be noted that the number of molecules fluorescing is proportional to
the number of excited molecules, maximumfluorescence intensity is achieved when the
exciting wavelength is at the absorption maximum.
concentration
6.2 Quenching Method of Analysis
When a molecule arrives at the lowest vibrational level of the excited electronic
state, the possibility is so high that the energy quantum for the return to the ground state
will be accepted by a molecule present in the solution. In this case, some of the energy
quantum for the return to the ground state was transferred to Hg 2+ as manifested by the
decreasing emission intensity after subsequent addition of Hg2+ into the system. The
divalent only confirmed its good capability as a quencher of fluorescence of Rhodamine
B. There are three major mechanisms for bimolecular quenching being collisional
deactivation, energy transfer and electron transfer.
As indicated in the Table 13, only two out of three trials obtained in emission
intensity were used for the construction of calibration curve. The instrument has been
consistently provided an outlier of the three trials conducted. Sudden drift of emission
intensity (usually in the third trial) has been attributed to the fluctuation current sources.
Thus, spectrofluorometer provided a variation of emission intensity.
The biggest source of error in the measurement is probably read out of the
fluorescence spectrometer since the errors in weighting a sample of a substance are
usually quite small and since there is completely confidence those, who prepared the
standard solutions the errors coming from these solutions probably can be ignored. The
micropipettes used to conduct the experiment are so precise, that there as well the errors
not coming from human failure can be ignored. So it is virtually impossible to read the
values any exacter than 1 % intensity. The instrument Spectrofluorometer also initially
provides erratic reading which might be attributed to fluctuation electric current sources.
7 CONCLUSION
Fluorescence spectroscopy can yield low detection limits, high sensitivity and
high specificity. The high specificity is largely due to the fact that fluorophores exhibit
specific excitation (absorption) and emission (fluorescence) wavelengths. These
wavelengths can be determined via the collection of two spectra, an excitation spectrum
and an emission spectrum. In this experiment, the excitation and emission spectra for the
Rhodamine B and 8-hydroxyquinoline was measured. Fluorescence capability of
Rhodamine B was determined using a standard calibration curve. Quantitative
fluorescence analysis revealed that unknown solution contains 0.9142 ± 0.0277 ppm of
Rhodamine B. Linear least square and other statistical method confirmed that standard
addition method was repeatable and fulfilled precision criteria as indicated in a very low
standard deviation in concentration which corresponds to 0.0277 ppm. In addition, the
effect of divalent mercury ion (Hg2+) to quench the fluorescence of Rhodamine B was
used determine the concentration of Hg2+ in an unknown solution. The quenching
capability of Hg2+ was very evident as the emission intensity decreases after subsequent
addition of Hg2+. Calibration curve from the plot of concentration of Hg 2+ against
emission intensity revealed that the Hg2+ concentration in the unknown solution is 0.6877
ppm. Aside from Rhodamine B, fluorescence activity of 8-hydroxyquinoline was studied.
Magnesium metal, which is non-fluorescent, able to induce a fluorescent by forming
chelate with 8-hydroxyquinoline. Thus, the concentration of Mg2+ in an unknown
solutionwas determined employing the fluorescent complex of Mg2+- 8-
hydroxyquinoline. Calibration curve from the plot of concentration of Mg2+ against
emission intensity revealed that the Mg2+ concentration in the unknown solution is 80.96
ppm.
28
8 REFERENCES
6. J. Slavik, Fluorescent Proves in Cellular and Molecular Biology (Boca Raton, FL:
CRC Press, 1994).
7. M. M. Krahn and others, Environ. Sci. Technol., 27 (1993), 699.