2 - Imaging With Xray
2 - Imaging With Xray
Properties:
A- Contrast: the ability to distinguish between adjacent areas
B- Spatial resolution: the ability to detect fine details
Tested with (Grid test object), consist of a number of equally spaced bars and the
space between bars is the same width of the bar.
A bar and a space make up a line pair & spatial frequency of a pattern is given as
the number of line pairs per mm
There is several groups of line pairs with progressively reduced spacing & the
smallest visible detail size is about half of the inverse of the given line pairs
example: with resolution 10 lp/mm, smallest detail seen about 0.05 mm ( 50 µm )
C- Noise: (See later)
N.B.:
Whereas the output of the X-ray tube (Intensity) and the skin dose rate are proportional
to kV2, the film-screen dose is nearly proportional to kV4 (exponent is 3-5, dependent on
patient thickness and field size).
MCQ: X-ray exposure to the patient may be reduced by using a higher kVp (T) … Tricky question. It is usually
assumed that when kVp increase, the mAs is reduced and vice versa.
While increasing the FFD necessitates ↑↑ the charge (mAs) to produce the desired
number of photons at the film-screen (to overcome reduction in intensity with
↑distance according to inverse square low), however still the skin dose incurred in
producing an acceptable image is ↓↓, and to a lesser extent the dose to deeper tissues.
MCQ: Changes in focus-to-film distance (FFD) do not alter the entrance surface dose (ESD) because changes
in mA are required to maintain an adequate film dose (F) … Although the mA must be increased as the FFD is
increased, there is an overall reduction in the ESD due to the relationship of X-ray intensity to distance
being an inverse square law
SUBJECT CONTRAST
A structure in the patient is demonstrated by two things:
Resolution, sharpness, or lack of blurring of the image of its boundary
Contrast between it and adjacent tissues caused by differences in the transmission of
X-rays.
We study CONTRAST first, with the aid of a very simple example.
MCQ: Contrast between air and soft tissue is due to differences in their density not atomic number
Contrast media
One of the problems in radiography is the low contrast between soft tissues.
o One way of increasing contrast is to use a lower kV;
o Another is to use a contrast medium.
Contrast-media are chosen to have
1) high atomic number to maximize photoelectric absorption
2) high density.
Ideally, the absorption edge should lie just to the left of the major part of the
spectrum of X-rays leaving the patient. (MCQ)
Last Figure shows that this is the case with iodine (Z = 53, EK = 33 keV).
o Barium (Z = 56, EK = 37 keV) also has a favorably placed absorption edge, and in
addition a high density.
o Air and other gases (negative contrast media) → the use as contrast media relied
on their low density.
2- Air gap:
If, as in Fig., the film-screen is moved some 30 cm away from the patient, much of
the obliquely traveling scatter misses it, and the contrast is improved.
MCQ: An air gap contrast (by reducing the amount of scattered radiation that is able to reach the film)
& dose (As less of the primary radiation reaches the film as a result of the inverse square law, this
requires an increase in the kV/mA to compensate).
3- Flat metal filter:
Such a filter, placed on the cassette, absorbs the softer and obliquely traveling
scatter more than the harder direct rays.
This is not very effective, and necessitates an increase in the mAs.
MCQ: Use of compression reduced contrast seen in the organ being imaged (F) … Not only is the amount of
scatter reduced thus improving contrast, there is less overlying soft tissue above and below the organ or
region of interest. The differing attenuation of the organ therefore becomes more pronounced as it is not
superimposed by other soft tissues.
MCQ: Scatter may be reduced by increasing the kV (F) … Increasing the kV means that any scattered
radiation is more penetrating and may reach the film.
Scatter ≠ Contrast
It absorbs oblique scatter beams, allowing the direct beam only to pass through the
interspaces
Few of the scattered rays S can pass through the channels between the strips of lead and
reach the film (most of them are traveling obliquely and are relatively soft → will be
absorbed by the lead strips).
Angle of acceptance θ:
Def.: The angle within which scattered rays CAN reach the film through the
interspaces (focused grids have smaller angle of acceptance)
Usually small.
With very large fields, especially at a high kV, more scatter is produced, and a
high-ratio grid (12:1 or 16:1) is preferable.
No grid would generally be used with thin parts of the body (extremities), with
children or where there is an air gap.
• Low-frequency grids are used with systems having a moving grid assembly.
• Medium and high-frequency grids are typically used with stationary grid holders (e.g.,
portable radiography and many digital radiography systems)
• High-frequency grid use is particularly important for digital radiography systems to avoid
aliasing artifacts that arise from an insufficient sampling of high-frequency patterns that are
interpreted in the output signal as low frequency (aliased) signals.
CROSSED GRIDS
Unfortunately, scattered rays traveling obliquely to the primary beam but parallel to
the lead strips can pass through the gaps. These rays can be absorbed by the use of
crossed grids.
Structure: two linear grids superimposed with their grid lines at right angle to each
other, having the same focusing distance.
Adv.: More efficient than a single linear grid at removing scattered radiation, as the
pathway for radiation is now a tunnel rather than a channel.
Disadv.:
1. Need greater radiographic exposure.
2. Require very careful centering. If the grids are not at right angles, a coarse
interference pattern (Moire fringes) may be seen on the film.
3. Grid cut-off can also be a problem.
MCQ: In case of crossed grids composed of two linear grids, the resulting grid ratio equals to the sum (not
product) of individual grid ratios
FOCUSED & UNFOCUSED "PARALLEL" GRIDS
Focused grids:
Strips are tilted progressively from the center to the edges of the grid so that they all
point toward the tube focus (N.B. most grids used in radiology are focused)
About 20% of the direct rays are removed & the rest pass.
May be linear or crossed.
Linear focused grids converge at a line in space (convergent line) while crossed
focused grids converge at a point (convergent point).
Focal distance = the perpendicular distance between the grid & convergent line or
point.
Focusing Range is wide for low grid ratio & narrow for high grid ratio.
Dr. Islam Abo Shady 10
MCQ: For a focused grid, the focusing range depends on the grid ratio … The focusing range is wide for
low-ratio grids and narrow for high-ratio grids (the high-ratio grids therefore require more precise
positioning).
Parallel grids:
Lead strips are parallel when viewed in cross-section.
Focused at infinity → no convergent line.
Used effectively with:
1. Small x-ray fields.
2. Long target-grid distances.
It has little use in modern radiology "used in fluoroscopic spot film devices".
If the line density (number of grid lines per millimeter) is sufficiently high → reduce the
definition of fine detail.
MCQ:The line density of a grid (number of lead strips per cm) is usually 30–80 cm–1 (typically 40).
MCQ: Linear grid, reduces contrast in the direction perpendicular to the lead strips (F).
MCQ: In linear grids, it may be possible for X-ray tube to be angled without the effect of grid cut-off.
MCQ: Using grids normally leads to:
a. Reduced scatter dose to the patient (F).
b. Higher mean energy of the beam reaching the film (T).
MCQ: A focused grid may cause radiation cut-off at large field size (F).
Multi-line grid:
Have 7 or more lines per millimeter together with a high grid ratio.
Can be used as a stationary grid without the lines being visible.
It is used when a moving grid cannot be used, and, being thinner, incurs less dose to
the patient.
MCQ:
Concerning antiscatter grid:
-With parallel grid, cut off limits the max field size
-With focused grid, cut off limits the range of FFD
Use of grid may increase patient dose by a factor of 4.
Grid lines in an image occurs with stationary and moving grids (but to less extent).
Use of grids normally leads to:
- scattered dose to the patient (due to mA)
- scattered radiation reaching the film.
- mean energy of the beam reaching the film.
In linear focused grids:
-Wrong FFD central portion of the film will be exposed with progressive increase in cut off towards the
edges of the film.
-Decentering tends to produce overall lighter film.
Mobile units often use stationary grids with low grid ratios.
In single phase X-ray units, the timing of grid movement must not coincide with that of the pulses, otherwise
it will appear as if the grid is stationary (Stroboscopic effect) not required in medium or high phase
systems.
Grids not used in pediatrics & extremity radiography.
N.B.:
1ry transmission indicates only the amount of 1ry radiation absorbed by a grid.
Bucky factor indicates the absorption of both 1ry & 2ry radiation.’
Grid-front cassette
Special X-ray cassettes with a grid build into the front of the cassette.
Used for portable radiography.
Most are focused with a low grid ratio (4:1 or 8:1) & long focal distance "focusing
range"
MCQ: Using a lead backed cassette Compton scatter and photoelectric effect
Focal spot size is one factor that must be considered when selecting an x-ray tube for a
specific application. Tubes with small focal spots are used when high image visibility of
detail is essential and the amount of radiation needed is relatively low because of small
and thin body regions as in mammography.
The line focus principle → the surface of target is inclined so that it forms an angle with the
plane to incident beam.
Angle θ → the angle between the central ray and the target face
The size of the projected focal spot is directly related to the sine of the angle of the
anode.
The smaller the angle of the anode, the smaller the apparent focal spot
🟉 There is a limit to which the anode angle can be decreased as dictated by the heel effect
(the point of anode cutoff).
MCQ: For general diagnostic radiography done at a 40-inches focus-film distance (1 m), the anode angle
is usually no smaller than 15°.
MCQ: The focal spot size increases with as increase in the tube current … due to Blooming.
MCQ: The effective focal spot is governed by:
a. the target size (F).
b. the applied kV (T).
Focal spot size is expressed in terms of the apparent / projected / effective focal spot;
sizes of 0.3, 0.6, 1.0, and 1.2 mm are commonly employed.
Usually, an X-ray tube has two filaments and two focal spots of different sizes which
are selected from the control panel.
The smaller focal spot is selected where small fields are needed & for better resolution
"in mammography and in cineradiography with a small field image intensifier" and the
larger one for thicker parts of the body where a greater intensity of X-rays is needed "in
general radiography using large films"
The focal spot size can limit the spatial resolution "geometric unsharpness", depends on the
location of the object in the source-to-detector direction.
The resolution impact of the focal spot increases with geometric magnification, i.e.increasing
distance between the object and the film or detector if FFD is fixed.
Thus, a small focal spot is desired in order to optimize spatial resolution.
The focal spot size also sets the upper limit on X-ray tube current or output rate (heatloading).
If an X-ray tube is operating at its instantaneous power limit, decreasing the size of thefocal spot will
require a decrease in the tube current (radiation output).
There is a trade-off between spatial resolution due to the size of the focal spot & imagenoise in a
fixed exposure time due toTypical
the decreasedeffective focal
X-ray intensity sizestime.
and imaging (mm)
Macromammography 0.1
Mammography 0.3
Macroradiography 0.3
Radiography 0.6-1.2
Fluoroscopy 0.6
MCQ: X-ray output DOES NOT depend on focal size, only sharpness and effective field of view do.
The focal spot is shorter when measured at the anode side than cathode side.
A small focal spot means a sharp image
As focal spot increase local heating of X-ray tube / geometrical unsharpness.
Measured by Pinhole camera : for actual focal spot size
Star test object: measure resolving capacity of focal spot.
The effective focal spot is governed by:
- Target angle (not target size)
- Line focus principle
- Filament size
- Applied KV
Dr. Islam Abo Shady 18
Target angle angle between the target surface and the central beam.
the anode angle actual focal spot size
Typical size of focal spot in general radiography is 1 mm.
Focal spot size has no effect on contrast.
BLOOMING
Def unwanted increase in focal spot size which occurs when the tube is operated
at high mA. Occurs particularly at low kV values and with small focal spots.
Occurs because the negative charge of the focusing cup is less effective, so electrons
emitted from the filament are not well focused in a regular beam → hit a larger area
> actual focal spot.
Rotating Anode:
Rotating anode is used to produce x-ray tubes capable of withstanding the heat
generated by large exposures.
The anode of a rotating anode tube consists of a large disc of tungsten or an alloy of
tungsten "tungsten-rhenium alloy" → better thermal characteristics than pure
tungsten and does not roughen with use as quickly.
Typical disc diameters measure 75, 100, or 125 mm.
The diameter of the tungsten disc determines the total length of the target track→
affects the maximum permissible loading of the anode.
Any area of the tungsten disc is found opposite the electron stream only once every 1/60
sec & during the remainder of the time heat generated during the exposure can be
dissipated.
The tungsten disc has a beveled edge. The angle of the bevel may vary from 6 to 20°.
The bevel is used to take advantage of the line focus principle.
The purpose of the rotating anode is to spread the heat produced during an exposure
over a large area of the anode while the apparent or effective focal spot size has
remained the same.
Heat loading:
measure of amount of energy deposited during an exposure
The heat loading of an X-ray tube (calculated in joules or heat unit HU)
= kV x mAs for a constant potential (three phase).
= 0.7 x kV x mAs for a pulsating single-phase generator.
Heat capacity:
A higher heat capacity means the temperature of the material rises only a small amount with
a large increase in heat input.
Temperature rise = energy applied / heat capacity
Heat (Tube) rating: Maximum allowable power (kV and mA) that can be used for a given
exposure time (0.1 sec) before the system overloads.
Each machine has a different capacity for dissipating heat before damage is caused. The
capacity for each focal spot on a machine is given in tube rating graphs provided by the
manufacturer.
Remember, heat rating is different from tube to tube, heat capacity is fixed
Heat loading (opposite of) Tube rating / Heat rating
The rapidity with which a series of such exposures can be made depends on:
a) The maximum amount of heat that can be temporarily stored in the anode in
particular & the tube housing as a whole; and
b) The rate at which they lose heat by cooling.
Dr. Islam Abo Shady 21
The heat storage capacity of the anode may be increased by soldering to the back of
the tungsten plate a disk of molybdenum and/or solid graphite (both have a higher
heat capacity per unit mass than tungsten).
Microprocessor in control circuit calculates the max. total number of exposures
allowed.
If the anode heat capacity (typically 0.2 MJ) has been reached → need at least 15 min
to cool down completely;
& the entire assembly (typical heat capacity 1.0 MJ) may need an hour.
MCQ:
Heal effect
Only occurs along the anode cathode line (axis) occurs in direction parallel to anode cathode axis
Variation in beam attenuation beam change in hardness (quality) and intensity (quantity)
It is more noticeable at low KVp, as photons are less penetrating & the differences in attenuation on the
cathode anode side are more pronounced.
It is useful in spine radiographs.
Used in mammography higher beam intensity on cathode side is directed towards chest wall where
greater penetration is required.
Dosimetry:
- Tube KV tested by: digital KV meter
-Tube output tested by: ionization chamber or solid state detectors
-Tube filtration tested by: HVL
-Light beam diaphragm alignment should be checked every 1-2 m.
MCQ: Subtraction of the high kV image from the low kV image will improve detection of calcified structures
(calcified lung nodule) … not the opposite.
MCQ: Dual-energy radiography always requires two separate exposures (F) … This is true for dual-exposure
systems that obtain two radiographs sequentially at different kVp. There are also single-exposure systems
where two image plates divided by a copper filter (‘sandwiched detectors’) are simultaneously exposed, with
the back detector receiving filtered, higher energy photons.
MCQ: In dual-exposure systems for chest radiography, tube potentials of 60 and 120 kV are usually used
MCQ: Dual-exposure systems produce selective tissue images with a better signal-to-noise ratio than single-
exposure systems