Product Brochure

Network Master™ Series


MT9090A
Mainframe
MU909014A1/B/B1/C/C6
MU909015A6/B/B1/C/C6
µOTDR ModuleTM
1310 nm/1490 nm/1550 nm for Installation,
1625 nm & 1650 nm for Maintenance,
1310 nm/1550 nm plus 1625 nm or 1650 nm
Field Optical Testing Redefined!
MT9090A with MU909014/15 Overview
There are many handheld OTDRs on the market that appear to be a good value until they are put into action and the user quickly finds out
that they lack the performance needed to install and maintain today’s networks.
The new MU909014/15 µOTDR Module series for the MT9090A Network Master™ platform from Anritsu finally addresses this need
by providing all of the features and performance required for installation and maintenance of optical fibers in a compact, modular test
set. The MT9090A represents an unmatched level of value and ease of use, while not compromising performance. Data sampling of 2
centimeters, dead zones of 0.8-meter and dynamic range up to 38 dB ensure accurate and complete fiber evaluation of any network
type – premise to access, metro to core…including PON-based FTTx networks featuring up to a 1 × 64 split.
The MT9090A with MU909014/15 module represents a new era in optical fiber testing!

Key Features
• Tri-wavelength OTDR for both installation and maintenance Mainframe
(1310 nm/1550 nm plus filtered 1650 nm or 1625 nm)
• Built-in PON Power Meter, Loss Test Set and Light Source functions
• High-end OTDR performance in a pocket-size package with unique battery operation
• “Fiber Visualizer” mode simplifies operation, no OTDR knowledge needed
• Complete PON testing through splitters up to 1 × 64
• Bluetooth, Wi-Fi and Ethernet connectivity* µOTDR Module
*: These features use an USB Ethernet converter, USB Wi-Fi dongle, or USB Bluetooth dongle.
The Bluetooth® wordmark and logos are owned by the Bluetooth SIG, Inc. and any use of such marks by Anritsu is under license.

MU909014/15
µOTDR ModuleTM
1310 nm/1490 nm/1550 nm for Installation,
1625 nm & 1650 nm for Maintenance,
1310 nm/1550 nm plus 1625 nm or 1650 nm Network Master™ and µOTDR Module™ are a trademark of
Anritsu Corporation.
2 Product Brochure l MU909014/15
A Truly Revolutionary OTDR!
Introducing the first handheld OTDR that does not compromise performance – the new µOTDR from Anritsu. With performance that rivals
traditional OTDRs that are four times the size and more than double the price, the Network Master MT9090A µOTDR has created a new
class of test instruments. It features 2 cm resolution for accurate mapping of events, dead zones of 0.8-meter (2.6-feet) and a dynamic
range of up to 38 dB – enough to test over 150 km (90+ miles). The MT9090A µOTDR also takes portability to a new level by being the first
handheld OTDR that truly fits in the palm of your hand.

Complete Testing Tool - Premise to Core 0.8-m Dead Zone for Short Fiber Analysis
With a dynamic range of up to 38 dB, the µOTDR evolves far With 0.8-meter dead zones, the MT9090A is perfect for evaluating
beyond the premise/access applications that other handheld OTDRs central office, FTTx and intra building cables.
service. Metro links can be tested with lower pulsewidths which
provides greater detail and better resolution while long haul fibers up Fast Real Time Sweeping
to 175 km (108 miles) can also be completely evaluated. The MT9090A µOTDR features real-time updates as quickly as 0.25
seconds. This is useful for connector and splice optimizations as
FTTx and PON Ready well as verifications of parameter selection.
With splitter-based fiber-to-the-x (FTTx) deployments becoming
more popular, the need for test equipment to thoroughly test and Portable
maintain them has risen. The µOTDR series features the ability to The MT9090A µOTDR takes portability to a whole new level. With
test up to a 1 × 64 split completely from end-to-end and with high dimensions of just 19 cm × 9.6 cm × 4.8 cm (7.5” × 3.8” × 1.9”)
resolution. and a weight of only 700 g (1.54 lbs.), the µOTDR is the smallest
and lightest OTDR on the market. With its lightweight design and
user friendly dimensions, the MT9090A is perfect for the outside
plant environment and can easily be managed with one hand. The
standard soft case with shoulder strap further increases portability
when traveling from the truck to the testing site.

Bluetooth, Wi-Fi and Ethernet Connectivity


The Bluetooth feature enables you to share files between the µOTDR
series and a PC. The Wi-Fi and Ethernet features enable you to
share files as well as use the remote GUI feature. You can connect
the µOTDR and PC, and control the μOTDR series from a browser.

Easy Operation and Analysis Bluetooth:


- Share file folder
“Fiber Visualizer (FV)” is a new fault location function designed to
Wi-Fi and Ethernet:
simplify the entire testing process. Fiber Visualizer automatically - Share file folder
selects the testing parameters to ensure the correct setup and - Remote GUI
provides a simple, graphical summary of the fiber under test within
seconds. A comprehensive PDF report can then be customized and
generated, completing the testing process. Modular Design
The MT9090A features a modular design allowing modules to be
easily changed in the field. Users can interchange different OTDR
modules or perform other optical network testing such as optical
channel analysis with the available CWDM channel analyzer module
or 10/100/1000 MB Ethernet testing on optical or electrical links.
Operation is quite similar between modules so the user is immediately
familiar with operation.

4.3-inch Wide Screen Display for Easy Viewing


The high resolution, full color, 4.3-inch wide screen display is the
perfect format for viewing OTDR results. It also provides excellent
readability both indoors and outdoors.

Integrated Launch Fiber


To further simplify testing, the MU909014/15 series is the only
handheld OTDR that features an integrated launch cable. A ten
meter (30 feet) fiber is built-in so initial fiber connections can be
verified without the need for additional patchcords or launch fibers.

Product Brochure l MU909014/15 3


Reliable. Capable.
When buying products, you tend to choose ones that are innovative and from established companies.
When you need to install and maintain optical networks, this should also apply. With over 50 years of
combined OTDR design, Anritsu, which now includes NetTest, delivers the features that matter.
Having been in the test and measurement business for a long time, we understand the importance of
performance, portability, reliability, easy operation and of course price.

Event Table with User Defined Thresholds This new function inspects the state of the connector end using
PASS/FAIL thresholds for key acceptance criteria such as splice video. It can automatically inspect the end of the optical connector
loss, reflectance and total span loss can be set in the MT9090A for defects and scratches (The automatic pass/fail determination is
allowing technicians to easily assess a fiber’s condition. Failing made in accordance with the IEC61300-3-35 standard.)
values are clearly highlighted in the event table alerting technicians You can also create a PDF report on the μOTDR series.
of potential problems.

Unique Battery Operation


Since AC power is not always available where you need it,
especially at fiber pedestals, the MT9090A typically provides 8 hours
of testing on a single charge. This coupled with an optional car
cigarette lighter cord guarantees the MT9090A is ready when you
are. µOTDR supports widely available NiMH and Alkaline batteries
for truly unique battery operation.

Quick Startup
The MT9090A is ready for measurement in under 15 seconds so
productive work can start immediately. Screen Capture Function
Screen shots are sometimes useful for adding to reports so the
Video Inspection Probe Support
MT9090A features the ability to save screen shots as Bitmap images.
When equipped with the optional connector video inspection probe
(G0306B), the µOTDR becomes a powerful tool for evaluating Functions for FTTx
connector cleanliness and quality. The G0306B can reduce issues One µOTDR module supports FTTx installation and maintenance
by verifying the condition and cleanliness of connector end faces (PON Power Meter, Loss Test Set, Light Source) in addition to
during the installation phase. µOTDR functions. (See page 6 for details.)
The G0306B has added a Pass/Fail analysis function to the
conventional VIP.

Small on price, Not on features!


1 2 3 7 8 9

5
8 8 9
6

1 4 .3-inch high resolution, indoor/


1) 5 Set to select/accept
5)
outdoor color display 6 Menu key for easy access to
6)
2)
2 D edicated function keys for set-ups and mass storage
Top View
performing tasks 7 Visible laser diode∗1
7)
3)
3 S tart key for true one-button 8 OTDR or power meter port∗2
8)
testing ∗1: There are installation restrictions, depending on the model.
9)
9 D ual USB ports for quick and See the Ordering Information for details.
4 A
4)  rrow keys for zooming, cursor easy data transfer ∗2: The number of ports is different, depending on the model.
movement and menu navigation See the Ordering Information for details.

4 Product Brochure l MU909014/15


Installation and Maintenance Simplified
Since the MT9090A is designed for technicians of any level, its hardware and user interface are optimized for simplicity. A customizable
testing sequence and “Full Auto” mode automates testing and guides novice users. Specialized maintenance wavelengths are also available
to eliminate equipment damage and transmission interruptions.

Installation Simplified Maintenance Simplified


The MU909014/15 µOTDR Module series provides easy and accurate Being able to test active fibers is a key requirement for network
verification of fiber installations at 1310 nm, 1490 nm and 1550 nm maintenance since multiple users often share portions of the
to ensure your network is ready for any transmission type. The user network and taking them all out of service is not an option. To
simply connects the fiber, selects “Full Auto” and presses “Start” - all address this need, special modules are available in the MT9090A
settings are automatically selected to ensure accurate and constant µOTDR series. 1650 nm is recommended by the ITU-T L.41 for
results for any skill level. Upon completion, all key fiber characteristics active maintenance since it features 100 nm of isolation from the
are displayed within seconds. Experienced users can also “fine tune” all nearest 1550 nm transmission wavelength. The 1650 nm OTDR also
testing parameters and make manual measurements. features an integrated filter to block transmissions from damaging
the OTDR. 1625 nm is also available and can be used for in-service
Step 1 – Connect fiber and Power on testing or as an “extra” test to verify installation for stresses such as
Step 2 – Select “Full Auto” and Press “Start” macrobends.
All testing parameters are automatically selected.
Added Macro Bending analysis function:
The μOTDR series finds macro bending points by comparing data
from two traces: one 1310 nm and the other 1550 nm.

1310 nm

∗: The screen items depend on the selected module. 1550 nm

Step 3 – Read Results


Test results including all splices and connectors, as well as total
fiber length and loss are shown in an easy to read table.

Network Documentation Simplified


Simple Data Storage
With internal data storage plus support for external USB memory
devices, the MT9090A is more than capable. Add to this auto file
saving and naming for easy, error-free documenting of your network.
Common OTDR Data Format
The MT9090A supports the universal Telcordia SR-4731 format
making it compatible with not only legacy Anritsu and NetTest
products, but with many other vendors data.

Step 4 – View Trace Free and Simple Software Upgrades


View trace if desired to see the complete fiber trace and make any Firmware upgrades are easily performed via USB and available from
manual measurements. the Anritsu website for registered users or through Anritsu customer
support.

Product Brochure l MU909014/15 5


All-in-one FTTx Installation and Maintenance Functions
There are three types of µOTDR module: single wavelength (1625 nm or
1650 nm) for the FTTx maintenance market including Metro networks, dual
wavelength (1310 nm/1550 nm) for the installation market, and triple wavelength
for both these markets.
These all-in-one µOTDR modules support every function required at fiber
installation and maintenance, as well as OTDR functions. The PON Power
Meter and Power Meter are ideal for loss measurements required for quality
measurements and basic fault tests.

PON Power Meter (1490 nm/1550 nm) Loss Test Set


Generally, PON communications use three wavelengths: 1310 nm, Combining the µOTDR module light source with the Power Meter
1490 nm, and 1550 nm. Data (1490 nm) and video (1550 nm) signals supports use as a Loss Test Set.
are sent to subscribers through one optical fiber but a general- The loss at both 1310 nm and 1550 nm can be measured with one
purpose optical power meter cannot separate the two wavelengths, µOTDR by looping-back the optical fiber. And both modulation (270 Hz,
making it difficult to locate faults using optical level measurements. 1 kHz, 2 kHz) and CW signals are supported. Just setting the threshold
The PON Power Meter can identify and measure the two 1490 nm and reference value makes PASS/FAIL evaluation easy.
and 1550 nm signals to support PASS/FAIL evaluations based on a
set threshold and reference value. Additionally, power measurements
and µOTDR tests are quick and easy without changing the optical fiber
because the PON Power Meter port is shared with the µOTDR function.

DCFL Function
The Drop Cable Fault Locate (DCFL) mode is a useful function to
investigate faults occurring in a drop cable. It consists of the Power
Meter function and OTDR function, so you are not required to switch
Light Source/Power Meter measuring instruments or applications.
The µOTDR module can be used as a light source to identify an
optical fiber and measure the loss by connecting an optical fiber
identifier and optical power meter at the other end of the fiber.
Since all wavelengths are shared by one µOTDR port, the fiber
identification, loss, and µOTDR measurements can all be performed
as a single task without changing the fiber connection. Both
modulation (270 Hz, 1 kHz, 2 kHz) and CW signals are supported.
The simple power meter function is ideal for checking optical levels
to confirm a fault occurrence using total received power. Setting a
threshold and reference value makes PASS/FAIL evaluation easy
too. In addition, power measurements and µOTDR tests are quick
and easy without changing the optical fiber, because the Power
*: DCFL mode depend on the selected module
Meter port is shared with the µOTDR.

Visible Laser Diode


The optional visible red LD light source makes it easy to spot faults
in splices and connectors and well as to manage fibers.

∗: The PON Power Meter, Light Source, Power Meter, Loss Test Set, and Visible
Laser Diode functions have different menus, depending on the selected
module. See the Ordering Information for details.
∗: The Visible Laser Diode is operated from the µOTDR and Power Meter menus.
∗: The screen items depend on the selected module.

6 Product Brochure l MU909014/15


Specifications
MT9090A Mainframe
190 (W) × 96 (H) × 48 (D) mm (7.5” × 3.8” × 1.9”) (including Mainframe and Module)
Dimensions and Mass
<700 g (1.54 lbs.) (including Mainframe, Module and Standard battery)
Display 4.3-inch TFT Color LCD (480 × 272 pixels, Transmissive)
Interface USB 1.1, Type A × 1 (memory), Type B × 1 (USB mass storage)

µOTDR Module Common (MU909014C/C6, MU909015C/C6, MU909014A1/B/B1 and MU909015B/B1, MU909015A6)


Fiber Type 10 µm/125 µm SMF (ITU-T G.652)
Optical Connector FC/SC/DIN adapter are changeable
Distance Range 0.5, 1, 2.5, 5, 10, 25, 50, 75, 125, 250 km (IOR = 1.500000)
Pulse Width 5, 10, 20, 50, 100, 200, 500 ns, 1, 2, 5, 10, 20 µs
Linearity Which ever is greater ±0.05 dB/dB or ±0.1 dB
Return Loss Measurement Accuracy∗1 ±2 dB
Distance Measurement Accuracy ±1 m ±3 × Measurement distance × 10–5 ±Marker resolution (excluding IOR uncertainty)
Internal memory: 40 MB (<1,000 traces)
Data Storage
External (USB Memory): 1 GB (<30,000 traces)
IOR Setting 1.3000 to 1.7000 (0.0001 steps)
Units km, m, kft, ft, mi
Integrated launch fiber: 10 m (30 ft)
Connection check: Automatic check of OTDR to FUT connection quality
Live fiber detect: Verifies presence of communication light in fiber
Real time sweep: <1 sec (typ.)
Macro bend analysis (without single-wavelength model)
Other Functions
Bluetooth, Wi-Fi and Ethernet connectivity
“Fiber Visualizer (FV)” function
“DCFL” function (differs with selected module)
Password protect function
Video inspection probe (option)
User selectable (English, Simplified Chinese, Traditional Chinese, Korean, Japanese, French, German, Italian, Spanish, Polish,
Language
Portuguese, Finnish, Danish, Swedish, Spanish (Latin America), Russian and Dutch)
Power Supply 9 V(dc), 100 V(ac) to 240 V(ac), Allowable Input voltage range: 90 V(ac) to 264 V(ac), 50 Hz/60 Hz
Fiber Event Analysis Automatic, Displayed in table format based on user defined PASS/FAIL thresholds
Loss Measurement Modes 2-point loss, Splice loss, dB/km Loss LSA, ORL, Event
OTDR Trace Format Telcordia universal (.SOR) issue 2 (SR-4731)
NiMH (Standard battery), NiMH (AA Type), Alkaline Dry Battery (AA Type)∗2
Battery Operating time (Standard battery): 8 hours (typ.)∗3, Telcordia GR-196-CORE Issue2, September 2010
Recharging time: <4 hours (typ.)∗4
EMC EN61326-1, EN61000-3-2
CE LVD EN61010-1
RoHS EN50581

MU909014C/C6 and MU909015C/C6 µOTDR Module


MU909015C/C6-057 MU909015C/C6-058 MU909015C/C6-059 MU909014C/C6-057 MU909014C/C6-058
Model Name
MU909015C/C6-067 MU909015C/C6-068 MU909015C/C6-069 MU909014C/C6-067 MU909014C/C6-068
1310/1550 ±20 nm∗6 1310/1550 ±20 nm∗6 1310/1550 ±20 nm∗6 1310/1550 ±20 nm∗6
Center Wavelength∗5 1310/1490/1550 ±20 nm∗6
1625 ±15 nm 1650 ±15 nm 1625 ±15 nm 1650 ±15 nm
PW = 20 µs 38 dB/37 dB/35 dB∗9, ∗10 38 dB/37 dB/35 dB∗9, ∗10 36 dB/35 dB/35 dB 32.5 dB/31 dB/32.5 dB∗9, ∗11 32.5 dB/31 dB/32.5 dB∗9, ∗11
Dynamic Range∗7, ∗8
PW = 500 ns 27 dB/26 dB/25 dB∗9, ∗10 27 dB/26 dB/24 dB∗9, ∗10 25 dB/24 dB/24 dB 24.5 dB/23 dB/24 dB∗9, ∗11 24.5 dB/23 dB/24 dB∗9, ∗11
Dead Zone∗12 Fresnel: ≤0.8 m (Typical)
(IOR = 1.500000) Backscatter: ≤4.0 m (1310 nm, Typical), ≤4.5 m (1490/1550/1625/1650 nm, Typical)
Number of Sampling Points∗13 <250,001 pts (Course: <7,501 pts, Medium: <20,001 pts, Fine: <250,001 pts)
Sampling Resolution 2 cm (min.)
OTDR (Full automatic, Manual, Real time), Power Meter, [Video Inspection Probe (Option)]
Testing Modes
[PON Power Meter, Loss Test Set, Light Source (MU909015C6, MU909014C6)]
Power Meter Please refer to the spec “Power Meter”
PON Power Meter
Please refer to the spec “PON Power Meter”
(only for MU909015C6/14C6)
Light Source
Please refer to the spec “Light Source”
(only for MU909015C6/14C6)
Loss Test Set
Please refer to the spec “Loss Test Set”
(only for MU909015C6/14C6)
Operating temperature and humidity: –10° to +50°C, <95% (no condensation)
Environment Storage temperature and humidity: –30° to +70°C, <95% (no condensation)
Vibration: MIL-T-28800E Class 3, Dust and Drip proof: IP51
Laser Safety∗14 IEC Pub 60825-1: 2007 Class 1M, 21CFR1040.10

Product Brochure l MU909014/15 7


MU909014A1/B/B1 and MU909015B/B1 µOTDR Module
MU909015B/B1-056 MU909014B/B1-056 MU909014A1-053 MU909014A1-054
Model name
MU909015B/B1-066 MU909014B/B1-066 MU909014A1-063 MU909014A1-064
Center Wavelength∗5 1310/1550 ±20 nm∗6 1625 ±15 nm 1650 ±15 nm
PW = 20 µs 37 dB/36 dB 32.5 dB/31 dB 32.5 dB∗9, ∗11
Dynamic Range∗7, ∗8
PW = 500 ns 28 dB/26 dB 24.5 dB/23 dB 24.5 dB∗9, ∗11 24 dB∗9, ∗11
Dead Zone∗12 Fresnel: ≤1 m
(IOR = 1.500000) Backscatter: ≤5 m
Number of Sampling Points∗13 <125,001 pts (Course: <6,251 pts, Medium: <25,001 pts, Fine: <125,001 pts)
Sampling Resolution 5 cm (min.)
Testing Modes OTDR (Full automatic, Manual, Real time), Power Meter, [Visible Fault Locator (Option)], [Video Inspection Prove (Option)]
Power Meter
Please refer to the spec “Power Meter” Not applicable
(only for MU909014B/B1/15B/15B1)
Connector: 2.5 mm universal
Visible Fault Locator Wavelength: 650 ±15 nm (CW, +25°C)
(only for MU909014A1/B1/15B1) Output power: 0 ±3 dBm (CW, +25°C)
Modulation: CW, 1 Hz
Operating temperature and humidity: –5° to +40°C, <80% (no condensation)
Environment Storage temperature and humidity: –20° to +60°C, <80% (no condensation)
Vibration: MIL-T-28800E Class 3, Dust and Drip proof: IP51
IEC Pub 60825-1: 2007 Class 1, IEC Pub 60825-1: 2007 Class 1M, IEC Pub 60825-1: 2007 Class 3R (VLD Option),
Laser Safety∗14
21CFR1040.10

MU909015A6 µOTDR Module


MU909015A6-053 MU909015A6-054
Model Name
MU909015A6-063 MU909015A6-064
Center Wavelength∗5 1625 ±15 nm 1650 ±15 nm
PW = 20 µs 35 dB∗9, ∗10
Dynamic Range∗7, ∗8
PW = 500 ns 25 dB∗9, ∗10 24 dB∗9, ∗10
Dead Zone∗12 Fresnel: ≤0.8 m (Typical)
(IOR = 1.500000) Backscatter: ≤4.5 m (Typical)
Number of Sampling Points∗13 <250,001 pts (Course: <7,501 pts, Medium: <20,001 pts, Fine: <250,001 pts)
Sampling Resolution 2 cm (min.)
OTDR (Full automatic, Manual, Real time), Power Meter, [Video Inspection Probe (Option)]
Testing Modes
[PON Power Meter, Light Source]
Power Meter Please refer to the spec “Power Meter”
PON Power Meter Please refer to the spec “PON Power Meter”
Light Source Please refer to the spec “Light Source”
Operating temperature and humidity: –10° to +50°C, <95% (no condensation)
Environment Storage temperature and humidity: –30° to +70°C, <95% (no condensation)
Vibration: MIL-T-28800E Class 3, Dust and Drip proof: IP51
Laser Safety∗14 IEC Pub 60825-1: 2007 Class 1, 21CFR1040.10

∗1: Design assurance. Distance range: 25 km, Pulse width: 2 µs, 20 km open the ∗9: 1490 nm/1550 nm cut filter included (1625 nm or 1650 nm port)
fiber-end. BSC: –78.5 (1310 nm), –80.1(1490 nm), –81.5 (1550 nm), ∗10: Specified without background light (1625 nm, 1650 nm)
–82.5 (1625 nm/1650 nm) ∗11: In service Signal is –20 dBm (CW) at 1310 nm/1550 nm
∗2: All specifications are guaranteed by standard battery. ∗12: Return Loss 45 dB, +25°C
∗3: Back light low, Sweeping halted, +25°C Fresnel: PW = 5 ns, 1.5 dB down from the peak of Fresnel
∗4: +10 to +30°C, Power off Backscatter: PW = 5 ns, Deviation ±0.5 dB
∗5: At +25°C, 1 µs, except charging battery
∗6: Typical value, ±25 nm is Guaranteed
∗7: Typical value, Distance range: 125 km, Averaging: 180 sec, SNR = 1, +25°C,
Except while charging battery, Subtract 1 dB for guarantee
∗8: Dynamic range (one-way back-scattered light)
SNR = 1: The level difference between the RMS nose level and the level
where near end back-scattering occurs.

∗13: Either medium and fine density value is selected depends on distance range
∗14: Safety measures for laser products
This option complies with optical safety standards, in Class1, 1M, 3R of
IEC 60825-1; the following descriptive labels are affixed to the product.

8 Product Brochure l MU909014/15


Other Functions
Light Source
Models MU909015C6/14C6, MU909015A6
1310/1550 ±25 nm (MU909015C6/14C6)
1490 ±25 nm (MU909015C6-059, MU909015C6-069)
Wavelength∗15
1625 ±25 nm (MU909015C6/14C6-057, MU909015A6-053, MU909015C6/14C6-067, MU909015A6-063)
1650 ±25 nm (MU909015C6/14C6-058, MU909015A6-054, MU909015C6/14C6-068, MU909015A6-064)
Fiber Type 10 µm/125 µm SMF (ITU-T G.652)
Output port Shared with OTDR port
Output power∗15, ∗16 –5 ±1.5 dBm
Output stability∗17 ≤0.2 dB
Modes of Operation CW, 270 Hz, 1 kHz, 2 kHz
Laser Safety Same as OTDR

Power Meter
Models MU909015C6/14C6, MU909015A6 MU909015C/14C MU909015B/B1, MU909014B/B1
Setting Wavelength 1310/1490/1550/1625/1650 nm 1310/1490/1550 nm 1310/1490/1550/1625/1650 nm
Fiber Type 10 µm/125 µm SMF (ITU-T G.652)
–50 to +26 dBm (CW) –50 to –5 dBm (CW)
Measurement range∗18
–40 to +13 dBm (270 Hz, 1 kHz, 2 kHz)
Shared with OTDR port Shared with OTDR port
Measurement port 1625 nm or 1650 nm OTDR port 1310 nm/1550 nm OTDR port (Except options 059 and 069)
Dedicated port (Options 059 and 069) 1310 nm/1490 nm/1550 nm OTDR port (Options 059 and 069)
Measurement Accuracy∗19 ±0.5 dB
Modes of Operation CW, 270 Hz, 1 kHz, 2 kHz CW

PON Power Meter (1490 nm/1550 nm)


Models MU909015C6/14C6, MU909015A6
Wavelength 1490 nm/1550 nm
Fiber Type 10 µm/125 µm SMF (ITU-T G.652)
–50 to +13 dBm (1490 nm, CW)
Measurement range
–50 to +26 dBm (1550 nm, CW)
Shared with OTDR port (1625 nm or 1650 nm)
Measurement port
Dedicated port (Options 059 and 069)
Measurement Accuracy∗20 ±0.5 dB
Isolation∗21 1490 nm: >35 dB, 1550 nm: >50 dB

Loss Test Set


Models MU909015C6/14C6
Fiber Type 10 µm/125 µm SMF (ITU-T G.652)
Light Source: Shared with OTDR port
1310 nm/1550 nm OTDR port (Except options 059 and 069)
1310 nm/1490 nm/1550 nm OTDR port (Options 059 and 069)
Measurement port
Power Meter: Shared with OTDR port
1625 nm or 1650 nm OTDR port (Except options 059 and 069)
Dedicated port (Options 059 and 069)
Light Source
1310 ±25 nm, 1550 ±25 nm (Except options 059 and 069)
Wavelength
1310 ±25 nm, 1490 ±25 nm, 1550 ±25 nm (Options 059 and 069)
Output Power∗15, ∗16 –5 ±1.5 dBm (CW, 25°C)
Output stability∗17 ≤0.2 dB
Modes of Operation CW, 270 Hz, 1 kHz, 2 kHz
Laser Safety Same as OTDR
Power Meter
Setting Wavelength 1310/1490/1550/1625/1650 nm
–50 to +26 dBm (CW)
Measurement range∗18
–40 to +13 dBm (270 Hz, 1 kHz, 2 kHz)
Measurement Accuracy∗19 ±0.5 dB
Modes of Operation CW, 270 Hz, 1 kHz, 2 kHz
∗15: At +25°C, CW
∗16: Fiber length 2 m, after the warm-up.
∗17: Wavelength 1310 nm/1550 nm, CW, ±1°C at one point within –10° to +50°C, deference between the largest value and shortest value for one minute, single mode fiber
2 m, when the optical power meter with return loss of 40 dB or more is used. After the warm-up time (10 minutes) passed.
∗18: At 1550 nm
∗19: 1310 nm/1490 nm/1550 nm, CW, –20 dBm, 25°C, on master connector fiber (FC) use, after zero offset execution.
∗20: 1490 nm/1550 nm, CW, –20 dBm, +25°C, on master connector fiber (FC) use, after zero offset execution.
∗21: Design assurance.

Product Brochure l MU909014/15 9


Ordering Information
Please specify the model/order number, name and quantity when ordering.
The names listed in the chart below are Order Names. The actual name of the item may differ from the Order Name.

1) Select Mainframe
Includes battery pack, AC charger/adapter, standard soft case, strap and protector.
Model/Order No. Description
MT9090A Mainframe
Standard Accessories
G0202A Replacement NiMH Battery Pack
G0203A Replacement AC Charger/Adapter
B0601B For MT9090A with/without Protector. This soft case is standard accessory for MT9090A main frame.
Z1023A Replacement Strap
B0663A Protector

2) Select Base Module


Model/Order No. Description
MU909014A1∗1 µOTDR (Single wavelength, 30 dB class OTDR with VLD)
MU909015A6∗2 µOTDR (Single wavelength, 35 dB class OTDR with PM, PON-PM and LS)
MU909014B∗1 µOTDR (2-wavelength, 30 dB class OTDR)
MU909014B1∗1 µOTDR (2-wavelength, 30 dB class OTDR with VLD)
MU909015B∗1 µOTDR (2-wavelength, 35 dB class OTDR)
MU909015B1∗1 µOTDR (2-wavelength, 35 dB class OTDR with VLD)
MU909014C∗4 µOTDR (3-wavelength, 30 dB class OTDR)
MU909014C6∗4 µOTDR (3-wavelength, 30 dB class OTDR with PM, PON-PM, LTS and LS)
MU909015C∗3, ∗4 µOTDR (3-wavelength, 35 dB class OTDR)
MU909015C6∗4, ∗5 µOTDR (3-wavelength, 35 dB class OTDR with PM, PON-PM, LTS and LS)
∗1: One OTDR port (any of 1310 nm/1550 nm, 1625 nm, 1650 nm) and visible light source (option) (Fig. 1)
∗2: One OTDR port (1625 nm or 1650 nm) (Fig. 2)
∗3: One OTDR port (1310 nm/1490 nm/1550 nm; Options 059 and 069) (Fig. 2)
∗4: Two OTDR ports (1310 nm/1550 nm, and 1625 nm or 1650 nm; Except options 059 and 069) (Fig. 3)
∗5: One OTDR port and dedicated power meter port (1310 nm/1490 nm/1550 nm, and power meter; Options 059 and 069) (Fig. 3)


Fig. 1 Fig. 2 Fig. 3

3) Select Module, Connector Interface and Testing Options


Includes operation manual and quick reference guide.
Model/Order No. Description PM PON-PM LTS LS∗7 VLD∗8 FV DCFL
UPC type APC type Installation and Maintenance Models
MU909014C-057 MU909014C-067 µOTDR (1310/1550/1625 nm, 32.5/31/32.5 dB) ∗1 
MU909014C-058 MU909014C-068 µOTDR (1310/1550/1650 nm, 32.5/31/32.5 dB) ∗1 
MU909015C-057 MU909015C-067 µOTDR (1310/1550/1625 nm, 38/37/35 dB) ∗1 
MU909015C-058 MU909015C-068 µOTDR (1310/1550/1650 nm, 38/37/35 dB) ∗1 
MU909015C-059 MU909015C-069 µOTDR (1310/1490/1550 nm, 36/35/35 dB) ∗1 
MU909014C6-057 MU909014C6-067 µOTDR (1310/1550/1625 nm, 32.5/31/32.5 dB) ∗1 ∗3 ∗5   
MU909014C6-058 MU909014C6-068 µOTDR (1310/1550/1650 nm, 32.5/31/32.5 dB) ∗1 ∗3 ∗5   
MU909015C6-057 MU909015C6-067 µOTDR (1310/1550/1625 nm, 38/37/35 dB) ∗1 ∗3 ∗5   
MU909015C6-058 MU909015C6-068 µOTDR (1310/1550/1650 nm, 38/37/35 dB) ∗1 ∗3 ∗5   
MU909015C6-059 MU909015C6-069 µOTDR (1310/1490/1550 nm, 36/35/35 dB) ∗2 ∗4 ∗6  
UPC type APC type General Purpose Models
MU909014B-056 MU909014B-066 µOTDR (1310/1550 nm, 32.5/31 dB) ∗1 
MU909014B1-056 MU909014B1-066 µOTDR (1310/1550 nm, 32.5/31 dB) ∗1  
MU909015B-056 MU909015B-066 µOTDR (1310/1550 nm, 37/36 dB) ∗1 
MU909015B1-056 MU909015B1-066 µOTDR (1310/1550 nm, 37/36 dB) ∗1  
UPC type APC type Maintenance Models
MU909014A1-053 MU909014A1-063 µOTDR (1625 nm, 32.5 dB)  
MU909014A1-054 MU909014A1-064 µOTDR (1650 nm, 32.5 dB)  
MU909015A6-053 MU909015A6-063 µOTDR (1625 nm, 35 dB) ∗1 ∗3   
MU909015A6-054 MU909015A6-064 µOTDR (1650 nm, 35 dB) ∗1 ∗3   
∗1: PM (Power Meter) function shared with OTDR port. ∗8: VLD (Visible Laser Diode) function with visible light source port operated from
∗2: Dedicated PM port. OTDR or Power Meter.
∗3: PON-PM (PON Power Meter) shared with 1625 nm or 1650 nm OTDR port.
Identifies and measures 1490 nm and 1550 nm wavelengths.
∗4: Dedicated PON-PM port. Identifies and measures 1490 nm and 1550 nm
wavelengths.
∗5: LTS (Loss Test Set) function for measuring 1310/1550 nm wavelengths. Light
source shared with 1310/1550 nm OTDR port. Power meter shared with 1625 nm
or 1650 nm OTDR port.
∗6: LTS function for measuring 1310/1490/1550 nm wavelengths. Light source
shared with 1310/1490/1550 nm OTDR port. Power meter is dedicated port.
∗7: LS (Stabilized Light Source) shared with OTDR port for each wavelength.
Top Menu differs with selected module

10 Product Brochure l MU909014/15


4) Select Connector Adapter
Adapter included at no charge – must be added as a separate line item.
Model/Order No. Description
MU909014A/B/C-025 FC-APC Connector key width 2.0 mm
MU909015A/B/C-025 (APC: Models -063, 064, 066, 067, 068, and 069)
MU909014A/B/C-026 SC-APC Connector
MU909015A/B/C-026 (APC: Models -063, 064, 066, 067, 068, and 069)
MU909014A/B/C-037 FC Connector
MU909015A/B/C-037 (UPC: Models -053, 054, 056, 057, 058, and 059)
MU909014A/B/C-039 DIN 47256 Connector
B0601B Standard Soft Case B0602B Deluxe Soft Case
MU909015A/B/C-039 (UPC: Models -053, 054, 056, 057, 058, and 059) This standard accessory Full Network Master operation without
accommodates the mainframe with removal from the case.
MU909014A/B/C-040 SC Connector
fitted protector. Provides excellent protection for use in
MU909015A/B/C-040 (UPC: Models -053, 054, 056, 057, 058, and 059) hash conditions.
This does not accommodate the
mainframe if the protector is fitted.
5) Select Accessories
Must be added as separate line items.
Model/Order No. Description
Z1580A∗1 Protector & Soft Case
B0663A∗2 Protector
G0203A AC Adapter (for Replacement)
G0202A NiMH battery pack (for Replacement)
B0602B Deluxe Soft Case (for MT9090A)
B0601B Standard Soft Case
B0600B Hard Case (for MT9090A)
Z1023A Strap B0600B Hard Case
J1402A Car Plug Cord This accommodates two mainframes (with or without fitted protector), accessories
J1530A SC Plug-in Converter (UPC(P)-APC(J)) (light source or power meter, backup battery, fiber cleaner, etc.).
J1531A SC Plug-in Converter (APC(P)-UPC(J))
J1532A FC Plug-in Converter (UPC(P)-APC(J))
J1533A FC Plug-in Converter (APC(P)-UPC(J))
J1534A LC-SC Plug-in Converter (for SM, SC(P)-LC(J))
J1535A LC-SC Plug-in Converter (for MM, SC(P)-LC(J))
W3585AE Quick Reference Guide (English, Printed)
W3586AE Operation Manual (English, Printed)
Operation Manual
Z1579A
(English and Japanese, Electronic (CD-R))
G0306B Video Inspection Probe (× 400) Mainframe with Protector
NETWORKS PC Emulation Software for Data Analysis and Reporting
B0663A Protector
∗1: The protector (B0663A) and standard soft case (B0601B) from a set. The mainframe with fitted protector.
The protector includes a shoulder strap.
∗2: The shoulder strap can be used to hang the instrument around the neck while
working.

6) Replacement Adaptors
Must be added as separate line items.
Model/Order No. Description
J0617B FC (UPC: Models -053, -054, -056, -057, -058)
J0618E DIN (UPC: Models -053, -054, -056, -057, -058)
Replaceable Optical Connector SC
J0619B (UPC: Models -053, -054, -056, -057, -058)
(APC: Models -063, -064, -066, -067, -068)
J0739A FC (APC: Models -063, -064, -066, -067, -068) J1530A to J1535A G0306B
Replaceable optical connector (SC) Plug-in Converter Video Inspection Probe (× 400)
J1602A
Phosphor bronze
(The photo shows the J1534A)
Replaceable optical connector (FC)
J1603A
Phosphor bronze

Product Brochure l MU909014/15 11


MT9090 Series
MU909020A OCA Module
Compact CWDM channel analyzer to verify power levels, drift and channel presence of CWDM networks.

MU909060A GigE Module


Dedicated field test solution for installation and troubleshooting Ethernet links in the access network.

CMA5 Series MT9083 Series


For Optical Fiber Installation and ACCESS Master Mini-OTDR
Maintenance. All-in-one test tool for fiber construction and
maintenance.

MT1000A Network Master Pro


MU100020A OTDR Module 1310/1550 nm SMF
MU100021A OTDR Module 1310/1550/850/1300 nm SMF/MMF
MU100022A OTDR Module 1310/1550/1625 nm SMF
MU100010A 10G Multirate Module
Installing an OTDR Module MU100020A/MU100021A provides the OTDR functions required for optical fiber I&M. Work efficiency is
increased by all-in-one support for optical fiber tests and data communications network commissioning.
I&M tests of 1.5-Mbps to 10-Gbps communications networks can be executed by simultaneously installing the MU100010A. In addition
to supporting Ethernet, OTN, etc., networks, Mobile base station CPRI and OBSAI, as well as SyncE protocols are also supported.
MU100040B CPRI RF Module
Installing the CPRI RF Module MU100040B in the MT1000A supports analysis of IQ signal frequency characteristics included in CPRI
signals between the LTE base station RRH and BBU. This can be used to check operation of the RRH after installation.
MU100040B supported BBU emulation for RRH.

6SHFLILFDWLRQVDUHVXEMHFWWRFKDQJHZLWKRXWQRWLFH

x8QLWHG6WDWHV x,WDO\ x,QGLD


$QULWVX&RPSDQ\ $QULWVX6UO $QULWVX,QGLD3ULYDWH/LPLWHG
(DVW&ROOLQV%OYG6XLWH5LFKDUGVRQ 9LD(OLR9LWWRULQL5RPD,WDO\ QG UG)ORRU%LQQDPDQJODVW6WDJH
7;86$ 3KRQH ,QGLUDQDJDUIW5RDG%DQJDORUH,QGLD
7ROO)UHH )D[ 3KRQH
3KRQH
x6ZHGHQ )D[
)D[
$QULWVX$% x6LQJDSRUH
x&DQDGD .LVWDJnQJHQ%.,67$6ZHGHQ $QULWVX3WH/WG
$QULWVX(OHFWURQLFV/WG 3KRQH &KDQJ&KDUQ5RDG6KULUR+RXVH
6LOYHU6HYHQ5RDG6XLWH.DQDWD )D[ 6LQJDSRUH
3KRQH
2QWDULR.9&&DQDGD
3KRQH x)LQODQG )D[
)D[ $QULWVX$% x35&KLQD 6KDQJKDL
x%UD]LO 7HNQREXOHYDUGL),9$17$$)LQODQG
3KRQH $QULWVX &KLQD &R/WG
$QULWVX(OHWURQLFD/WGD )D[ 5RRP7RZHU$
1HZ&DRKHMLQJ,QWHUQDWLRQDO%XVLQHVV&HQWHU
3UDoD$PDGHX$PDUDO$QGDU
%HOD9LVWD6DR3DXOR63 x'HQPDUN 1R*XL3LQJ5RDG6KDQJKDL35&KLQD
%UD]LO $QULWVX$6 3KRQH
3KRQH 7RUYHSRUWHQ9DOE\'HQPDUN )D[
)D[ 3KRQH x35&KLQD +RQJ.RQJ
x0H[LFR )D[
$QULWVX&RPSDQ\/WG
$QULWVX&RPSDQ\6$GH&9 x5XVVLD 8QLW)*UHHQILHOG7RZHU&RQFRUGLD3OD]D
1R6FLHQFH0XVHXP5RDG7VLP6KD7VXL(DVW
$Y(MpUFLWR1DFLRQDO1R3LVR&RO*UDQDGD $QULWVX(0($/WG
.RZORRQ+RQJ.RQJ35&KLQD
0p[LFR')0p[LFR 5HSUHVHQWDWLRQ2IILFHLQ5XVVLD 3KRQH
3KRQH 7YHUVND\DVWUEOGWKIORRU
)D[ )D[
0RVFRZ5XVVLD
x8QLWHG.LQJGRP 3KRQH x-DSDQ
)D[ $QULWVX&RUSRUDWLRQ
$QULWVX(0($/WG
&DSDELOLW\*UHHQ/XWRQ%HGIRUGVKLUH/8/88. x6SDLQ 7DPXUDFKR$WVXJLVKL.DQDJDZD-DSDQ
3KRQH
3KRQH $QULWVX(0($/WG )D[
)D[
5HSUHVHQWDWLRQ2IILFHLQ6SDLQ x.RUHD
x)UDQFH (GLILFLR&X]FR,93RGHOD&DVWHOODQD3WD
0DGULG6SDLQ $QULWVX&RUSRUDWLRQ/WG
$QULWVX6$ )/3DQJ\R\HRNUR%XQGDQJJX6HRQJQDPVL
DYHQXHGX4XpEHF%kWLPHQW,ULV6LOLF 3KRQH
)D[ *\HRQJJLGR.RUHD
9,//(%21685<9(77()UDQFH 3KRQH
3KRQH
)D[
x8QLWHG$UDE(PLUDWHV )D[
$QULWVX(0($/WG x$XVWUDOLD
x*HUPDQ\ 'XEDL/LDLVRQ2IILFH $QULWVX3W\/WG
$QULWVX*PE+ $XURUD7RZHU 8QLW5LFNHWWV5RDG
1HPHWVFKHN+DXV.RQUDG=XVH3ODW] 32%R['XEDL,QWHUQHW&LW\ 0RXQW:DYHUOH\9LFWRULD$XVWUDOLD
0QFKHQ*HUPDQ\ 'XEDL8QLWHG$UDE(PLUDWHV 3KRQH
3KRQH 3KRQH )D[
)D[
)D[
x7DLZDQ
$QULWVX&RPSDQ\,QF
)1R6HF1HL+X5G7DLSHL7DLZDQ
3KRQH
)D[ 
3ULQWHGRQ5HF\FOHG3DSHU Catalog No. MU909014x/15x-E-A-1-(13.00) Printed in Japan 27/FEB/2018 ddcm/CDT

You might also like