Finals Rad Protect
Finals Rad Protect
Finals Rad Protect
一 NCRP 147 X-ray Room Designs and Shielding Calculations A to G are points used to calculate shielding.
一 Center for Device Regulation Radiation Health and Research
(CDRRHR) under BFAD and DOH
PURPOSE OF SHIELDING
1. To protect:
a. Patients
b. X-ray department staffs
c. Visitors
d. General public
e. Persons working adjacent or near the x-ray
facility RADIATION SHIELDING - CALCULATION
● Based on NCRP report no. 147, structural shielding
RADIATION SHIELDING - DESIGN CONCEPTS design for medical x-ray imaging facilities (2004)
Data required consideration of: ● Assumptions used are conservative, so over-shielding is
1. Type of x-ray equipment common.
2. Usage (workload) ● Software is available giving shielding in thickness of
3. Positioning various materials.
4. Whether multiple tubes or receptors are being used RADIATION SHIELDING PARAMETERS
5. Primary beam access (vs. scatter only) P - design dose per week
6. Operator location ● Usually based on 5 mSv per year for occupationally
7. Occupancy of surrounding areas exposed persons (25% of dose limit), and 1 mSv for the
public.
RADIATION SHIELDING - SHIELDING DESIGN ● Occupational dose must only be used in controlled
Equipment areas, i.e., for radiographers, radiologists, and other
1. What equipment is to be used? radiation workers.
a. General radiography ● Film storage areas (darkrooms) need special
b. Fluoroscopy considerations.
c. Dental (Oral, Cephalometric, or OPG) ● Long periods of exposures will affect the film, but
d. Mammography much shorter periods (i.e., lower doses) will fog film in
e. CT cassettes.
The type of equipment is very important for the following reasons: ● A simple rule is to allow 0.1 mGy for the period the
1. Where the x-ray beam will be directed film is in storage - if this is 1 month the design dose is
2. The number and type of procedures performed 0.025 mGy/week.
3. The location of the radiologic technologist (operator) ● Remember we must shield against three (3) sources of
4. The energy (kVp) of the x-rays radiation. These are:
Usage: ○ Scattered radiation (from the patient)
● Different x-rays have very different usage. ○ Primary radiation (the x-ray beam)
● For example, a dental unit uses low mAs and low kVp ○ Leakage radiation (the x-ray tube)
(~70), and takes relatively few x-rays each week U - use factor
● A CT scanner uses high (~130) kVp, high mAs, and ● Fraction of the time primary beam is in a particular
takes very many scans each week. direction i.e., the chosen calculation point.
● The total mAs used each week is an indication of total ● Must allow for realistic use for all points, sum may
x-ray dose administered. exceed 1.
● The kVp is also related to dose, but also indicates the ● For some x-ray equipment the x-ray beam is always
penetrating ability of the x-rays. stopped by the image receptor thus the use factor is 0 in
● High kVp and mAs means more shielding is required. other directions i.e., CT, fluoroscopy, mammography.
Number of x-ray tubes: ● For general radiographic and fluoroscopic equipment
● Some x-ray equipment may be fitted with more than the primary beam is usually intercepted by the image
one tube. detector. This reduces shielding requirements.
● Sometimes two tubes may be used simultaneously, and ● For radiography, there will be certain directions where
in different directions. the x-ray beam will be pointed:
● This naturally complicates shielding calculation. ○ Towards the floor
RADIATION SHIELDING - POSITIONING ○ Across the patient, usually only in one
The location and orientation of x-ray unit is very important: direction
● Distances are measured from the equipment (inverse ○ Toward the chest Bucky stand
square law will affect dose) ● The type of tube suspension will be important, e.g.,
● The directions the direct (primary) x-ray beam will be ceiling mounted, floor mounted, C-arm, etc.
used will depend on the position and orientation. T - occupancy
RADIATION SHIELDING - DESIGN DETAIL ● T = fraction of time a particular place is occupied by
Must consider: staff, patients, or public.
1. Appropriate calculation points, covering all critical ● Has to be conservative
locations. ● Ranges from 1 for adjacent offices and work areas, to
2. Design parameters such as workload, occupancy, use 1/20 for public toilets, and 1/40 for outdoor areas with
factor, leakage, target dose. transient traffic.
3. These must be either assumed or taken from actual data.
4. Use a reasonable, worst case scenario (conservatively
high estimates), since under-shielding is worse than
over-shielding.
● Gypsum or high atomic number (Z) plasterboard
● Concrete block
● Leaded glass or acrylic
ROOM SHIELDING - CONSTRUCTION PROBLEMS
● Brick walls - mortar joints
● Use of lead sheets nailed to timber frame
● Lead inadequately bonded to backing
W - workload ● Joints between sheets with no overlap
● A measure of radiation output in one week ● Use of hollow core brick or block
● Measured in mA - minutes ● Use of plate glass where lead glass specified
● Varies greatly with assumed maximum kVp of x-ray PROBLEMS IN SHIELDING - BRICK WALLS AND
unit MORTAR JOINTS
● Usually a gross overestimation ● Bricks should be solid and not hollow
● Actual dose / mAs can be estimated ● Bricks have very variable x-ray attenuation
● For example: a general radiography room ● Mortar is less attenuating than bricks
● The kVp used will be in the range 60-120 kVp ● Mortar is often not applied across the full thickness of
● The exposure for each film will be between 5 mAs and the brick
100 mAs. PROBLEMS IN SHIELDING - LEAD INADEQUATELY
● There may be 50 patients per day, and the room may be BONDED TO BACKING
used 7 days a week. ● Lead must be fully glued (bonded) to a backing such as
● Each patient may have between 1 and 5 films. wood or wallboard
Estimating W: ● If the lead is not properly bonded, it may peel off after a
1. Assume an average of 50 mAs per film, 3 films per few years
patient ● Not all glues are suitable for lead (oxidation of the lead
2. Thus, W = 50 mAs x 3 films x 50 patients x 7 days surface)
= 52,500 mAs per week PROBLEMS IN SHIELDING - JOINTS BETWEEN SHEETS
= 875 mA - min per week WITH NO OVERLAP
We could also assume that all this work is performed at 100 kVp. ● There must be 10-15 mm overlap between adjoining
EXAMPLES OF WORKLOAD sheets of lead
● Without an overlap, there may be relatively large gaps
for the radiation to pass through
● Corners are a particular problem
● Penetrations for electrical boxes and ducts are of
concern
PROBLEMS IN SHIELDING - USE OF PLATE GLASS
● Leaded glass or acrylic should be used for windows
● Laminated layers of plate glass can be used where
radiation levels are low, e.g., for the wall at the foot of a
WORKLOAD - CT CT scanner
● CT workloads are best calculated from NCRP 147 ROOM SHIELDING - CONSTRUCTION
● Remember that new CT spiral units, or multi-slice CT, ● Continuity and integrity of shielding is very important
could have higher workloads. ● Problem areas:
● A typical CT workload is about 28,000 mA - min per ○ Joints
week ○ Penetration in walls and floor
TUBE LEAKAGE ○ Window frames
● All x-ray tubes have some radiation leakage - there is ○ Door and frames
only 2-3 mm lead in the housing. PENETRATIONS
● Leakage is limited in most countries to 1 mGy hr-1 at 1 ● Means any hole cut into the lead for cables, electrical
meter, so this can be used as the actual leakage value connectors, pipes, etc.
for shielding calculations. ● Unless the penetration is small (~2-3 mm), there must
● Leakage is specified at the maximum rated continuous be additional lead over the hole, nails and screws used
tube current, which is about 3-5 mA at 150 kVp for to fix bonded lead sheets to a wall do not require
most radiographic x-ray tubes. covering.
WINDOW FRAMES
● The lead sheet fixed to a wall must overlap any lead
glass window fitted
● It is common to find a gap of up to 5 cm, which is
unacceptable
SHIELDING DOORS AND FRAMES