Daniel Schoop 539B Calibtation
Daniel Schoop 539B Calibtation
Daniel Schoop 539B Calibtation
June 1995
Use this procedure to test and calibrate the Hickok Model 539B or 539C
transconductance tube tester. All of the readings are taken with a 1000
ohms per volt meter. If an accurate 1000 ohms per volt meter is not
available a modern high impedance analog or digital voltmeter can be used
with appropriate shunt resistors in parallel with the input to simulate
proper loading. The following resistor values should be used: 10 volt scale
use 10K, 50 volt scale use 51K, 200 volt scale use 200K. All resistors are
1/2 watt 5% carbon composition. Calibration will be easier if you supply
AC power through a constant voltage regulation type transformer to do the
tests, but this is not essential.
Remove the screw from the bottom of the case. Remove all of the screws
around the outside of the panel that hold the tester to the case. Remove the
tester from the case and set it up on spacers so that the front panel is facing
up in the normal operating position. Before applying power adjust the
mechanical zero on all of the meters to set the pointers exactly at the zero
line on the scale.
POWER ADJUST control until the AC panel meter needle is exactly over
the red line. Remove the external AC voltmeter from the AC supply.
SHUNT POSITION
2.
Rotate the SHUNT control fully counter clockwise and verify that the
pointer lines up with the 0 mark on the scale.
SHORTS TEST
3.
Set the SHORTS switch to the TUBE TEST position and the SHORT
TEST switch to the OTHER TUBES position. Set both the CATHODE and
the SUPPRESSOR switches to the number 1 position. Rotate the SHORTS
switch counter clockwise. The SHORTS lamp should light in all positions
except number 1. Repeat the procedure setting both the CATHODE and
SUPPRESSOR switches simultaneously in positions 2 through Z. Return
the SHORTS switch to the TUBE TEST position.
METER= NORMAL
SELECTORS:
FILAMENT= H SCREEN= 4
FILAMENT= S CATHODE= 8
GRID= 5 SUPPRESSOR= 1
PLATE= 3
Figure 1
Set the FUNCTION switch to each of the positions shown in the chart
below. Push P4 and adjust the voltage source as indicated in the chart.
Observe the reading on the main meter and verify that the reading matches
the value given in the chart. After each reading set the voltage back to
zero, release P4 and go to the next setting.
ADJUSTMENT CHART
1.
LINE ADJUST METER: No adjustment. High readings can be brought
into calibration by the addition of a selected resistor placed in series with a
meter lead. Low readings cannot be raised.
2.
SHUNT POSITION: Loosen the set screw on the knob and reposition the
knob to the correct location. Retighten the set screw.
3.
No adjustment.
4.
SHORTS LIGHT SENSITIVITY: Perform the test and adjust R45 until the
SHORTS light just comes on.
5.
No adjustment.
6
No adjustment.
7.
No adjustment
8.
For the model 539B, the adjustment for the lower screen voltage is made
by loosening the screw on the left sliding tap of R18 and sliding it to the
position which gives the correct reading. For the 539C this setting is not
adjustable. This adjustment will also effect the maximum bias voltage
which is tested in the next step.
9.
If necessary, adjust the maximum bias voltage by loosening the screw on
the right sliding tap of R18 and sliding it to the position which gives the
correct reading. This adjustment will also effect the screen voltage which
was tested in the previous step and the screen voltage should be retested if
an adjustment is made.
10.
No adjustment.
11.
Calibration of this function is controlled by the 600 ohm spool resistor
R42.
12.
Calibration of this function is controlled by the 2 ohm spool resistor R44.
13.
No adjustment.
14.
Adjust R15.
15.
Adjust R8.