Saturation of GC Detectors
Saturation of GC Detectors
Saturation of GC Detectors
This technical article relates to the saturation of GC detectors. It addresses the causes of saturation, what effect saturation might have on analytical
results, how to know when it is happening, and how to avoid it.
Each style of GC detector will detect compounds based on some physical or A detector becomes “saturated” if its increase in signal as a function of
chemical characteristic of the sample. In flame ionization detectors (FID), increasing amount of analyte is less than expected. Figures 2 and 3 illustrate
organic compounds are burned in an H2/air flame and form ions that are the roll off in response at the high end of a calibration curve as the detector
collected on a polarized electrode (the “collector”). Other detectors respond saturates. Each detector will have its own root cause(s) for saturation
in different ways, but the response always ends up as an amplified electrical depending on the underlying chemical and physical phenomena relating to
signal (which then is usually digitized and stored). its response. In the case of an FID, saturation usually results from the flame
Every detector is designed for a specific intended use paradigm that becoming “starved”; it is no longer able to completely combust the large
involves all setpoint variables and physical design characteristics. Take the amount of analyte passing through it. This lack of complete combustion can
FID, for example. The primary source of signal is the efficient combustion be influenced by many variables such as limited fuel (H2) or oxidant (air),
of column effluent in a hydrogen flame, as illustrated in Figure 1. Column not enough pre-heat, inefficient mixing, etc. An extreme indication of poor
effluent is directed to the flame and mixed with H2 as it passes through mixing and insufficient combustion is when the flame self-extinguishes
a small orifice (the “jet”). In order to yield the theoretically 6-7 orders of when the solvent peak elutes. In many cases, increasing air and/or fuel
magnitude response that are possible with an FID, all the physical aspects flows and increasing detector temperature can yield another factor of
of the detector must work together. So, the dimensions of the detector, 5-10 of proportional detector response at the high end of the calibration
the orifice size, dimensions and position of the jet, the temperature profile, curve; however these changes would at the same time make LOD worse
the flow-rates of gases, electrical and electronics design, etc. — all (increasing noise).
these parameters are designed and specified for a “typical” use paradigm. Poor mixing of column effluent with the H2 fuel prior to entering the
Saturation occurs when some of the actual parameters of use fall outside flame can happen if the jet dimensions are not well matched to the column
the normal operating range. flow (a bigger orifice is needed for higher column flows, especially if using
packed columns). In this case, a change in jet orifice size can help. Using too
big of a jet orifice when unnecessary, on the other hand, degrades LOD if
used with lower flow columns.
Figure 1
Figure 2
www.sepscience.com
Figure 3
This technical article is produced in collaboration with Dr Matthew S. Klee,
internationally recognized for contributions to the theory and practice
of gas chromatography. His experience in chemical, pharmaceutical and
instrument companies spans over 30 years. During this time, Dr Klee’s
work has focused on elucidation and practical demonstration of the many
processes involved with GC analysis, with the ultimate goal of improving
the ease of use of GC systems, ruggedness of methods and overall quality of
results.
www.sepscience.com