Test Report: Model

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Number

Page 6

Test Report

Model
Linear Array 5DL16x2-0.75x5-S49-F5-D3

Code 10030
Serial Number 10030BJH08

Tester PQC003
Date 2020 / 08 / 17
Verifier OQC002
Date 2020 / 08 / 17

Company Address Phone & E-mail


DASHENG PRECISION Room 1403, Tel 020-8565 6213
(GUANGZHOU) Chanzen Center West Tower, E-mail [email protected]
TECHNOLOGY CO., LTD Tianhe District Guangzhou, www.bigprobe.com
Guangdong China
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

Test Specification
Conditions
Signal Generator  5072 P/R

Voltage Excitation 100 V

Test Medium 20mm Rexolite

Oscilloscope Keysight DSO2012A

Gain 10 dB

 PRF MODE: 1kHz

Probe Specification
information Dual Linear
Probe Type
Frequency 5.0MHz

Element Count (n) 32

Pitch(P) 0.75mm

Element width (E) 0.67mm

Notch width (K) 0.08mm

Elevation 5mm

Active Aperture 38.4mm

Medium Rexolite

01
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

Test Result Average Maximum Minimum

Sensitivity STDEV(dB) -47.48 -48.80 -46.35

Center Frequency(MHz) 4.87 4.97 4.77

-6dB Bandwidth(%) 89.64 95.12 73.73

-20dB Pulse Width(μs) 0.74 0.52 0.80

Parameter Measurement Specification Conformance

Center Frequency(-6dB) 4.87 5±0.2 MHz PASS

-6dB Relative Bandwidth 89.64 ≥ 60 % PASS

-20dB Pulse Width(μs) 0.74 ≤ 2ns PASS

Sensitivity Consistency -47.48 -46±3 dB PASS

Order of connection

Coupling response

No leakage

No failure element

02
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

03
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

Case Parameter

Specification

Case Model S49


Case Material SUS304
Case Size 24.5x6.75x20mm
/0.96x0.27x0.79(in)
Distance between
the shell and the first element 3.25mm/0.13in

Distance between
the shell and the last element 3.25mm/0.13in

Connector Specification

Connector D3( 160pin)


Model M2M Gekko/Mantis
OLYMPUS
OmniScan MX/MX2/SX/X3
OLYMPUS Focus LT/PX

Specification
Cable
Material PVC

Length 5m

Outer diameter of cable Φ5mm

External Diameter 60pF/m


Color Black

04
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

Single Element Feature

Element 2

Element 16

Element 31
05
//
Serial Number / 10030BJH08

Model
Linear Array 5L16x2-0.75x5-S49-F5-D3

Recommended Specification
Conditions
Storage Temperature -20°C to 60°C

Operation Temperature 10°C to 40°C

Negative pulse excitation Max 200Vpp

Maximum Frequency Max 10 KHz


Working Voltage 5 KHz 100V

Quality  In general, we offer a one-year warranty for phased array


probes sold. Quality warranty service does not include defects
Guarantee due to normal wear and tear , nor damage due to external
accident caused, as below.

(1) Assemble the probe incorrectly


(2) Incorrect maintenance method
(3) Unreasonable couplant
(4) High pressure or the immersion depth exceeds 1.0 m, and
a depth recommended immersion time as follows:
Standard phased array probes can be used for up to 1 hour
underwater at 0.5 meters (standard design)
The immersion phased array probe can be used continu
ously for up to 4 hours under water for 4 hours (standard
design)
(5) Storage temperature is not in the range of -20° to 60°C
(6) Operating temperature is not in the range of 10° to 40°C
(7) Normal wear and tear of the product
(8) High voltage (exceeding a voltage of up to 200V for probes
with a frequency of 7.5MHz or less; for frequency 10MHz
and above probes, exceeding the maximum voltage of 150V)
(9) The repetitive frequency of the instrument excitation
exceeds 10 KHZ.