Proline Prosonic Flow 93P
Table of contents
Performance characteristics. . . . . . . . . . . . . . . . . . . . 12
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 17
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2 Endress+Hauser
Proline Prosonic Flow 93P
a
Q b
tb
ta
A0005961
The measuring system calculates the volume flow of the fluid from the measured transit time difference and
the pipe cross-sectional area. In addition to measuring the transit time difference, the system simultaneously
measures the sound velocity of the fluid. This additional measured variable can be used to distinguish different
fluids or as a measure of product quality.
The measuring device can be configured onsite to suit the specific application using Quick Setup menus.
Measuring system The measuring system consists of one transmitter and two sensors.
The transmitter is used both to control the sensors and to prepare, process and evaluate the measuring signals,
and to convert the signals to a desired output variable.
The transmitter is optionally equipped for two-channel operation ä 5.
The sensors work as sound transmitters and sound receivers. Depending on the application and version, the
sensors can be arranged for measurement via one or two traverses ä 5.
Endress+Hauser 3
Proline Prosonic Flow 93P
Transmitter
A0009629 A0009676
Sensor
A0011484
A0013475
Mounting accessories
The requisite mounting distances must be determined for the sensors. Information on the fluid, the pipe
material used and the exact pipe dimensions is needed to determine these values. The values for the sound
velocity of the following fluids, pipe materials and lining materials are stored in the transmitter:
If you are using fluids, pipe materials or lining materials other than those listed in the table, and you do not
have the corresponding sound velocities for these fluids/materials, you can use the DDU18 and DDU19
sensors to determine the values.
DDU18 DDU19
(sound velocity measurement) (wall thickness measurement)
A0009784
A0009673
4 Endress+Hauser
Proline Prosonic Flow 93P
A B
A0001108
The number of traverses required depends on the sensor type, the nominal diameter and the thickness of the
pipe wall. We recommend the following types of mounting:
P-CL-1F*
DN 300 to 600 (12 to 24") 1 MHz (or 2 MHz) 2 traverses
P-CL-2F*
Two-channel operation The transmitter is able to operate two independent measuring channels (measuring channel 1 and measuring
channel 2). A pair of sensors is connected per measuring channel. Both measuring channels operate
independently of one another and are supported by the transmitter to an equal extent.
Two-channel operation can be used for the following measurements:
• Two-channel measurement = flow measurement at two separate measuring points
• Two-path measurement = redundant flow measurement at one measuring point
Endress+Hauser 5
Proline Prosonic Flow 93P
Two-channel measurement
The flow is measured at two separate measuring points in the case of two-channel measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-channel measurement:
• Individual measured values per measuring channel (output independently of one another)
• The difference between the two measured values
• The sum of the two measured values
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
A B
A0001159
Two-channel measurement: example of arranging sensor pairs at two separate measuring points
A Measuring channel 1: mounting the sensor pair for measurement via two traverses
B Measuring channel 2: mounting the sensor pair for measurement via one traverse
Two-path measurement
The flow is measured redundantly at one measuring point in the case of two-path measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-path measurement:
• Individual measured values per measuring channel (output independently of one another)
• The average of the two measured values
The "Averaging" function generally provides you with a more stable measured value. The function is thus
suitable for measurements under conditions that are not ideal (e.g. short inlet runs).
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
It is generally not necessary to individually configure the two measuring channels in the case of two-path
measurement. However, in certain situations individual channel configuration can be used to balance out
application-specific asymmetries.
A B
A0001160
6 Endress+Hauser
Proline Prosonic Flow 93P
Input
Measured variable Flow velocity (transit time difference proportional to flow velocity)
Output
Output signal Current output
• Galvanically isolated
• Active/passive selectable
– Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 )
– Passive: 4 to 20 mA, max. 30 V DC, Ri 150
• Time constant selectable (0.01 to 100 s)
• Full scale value adjustable
• Temperature coefficient: typ. 0.005 % o.r./°C (o.r. = of reading)
• Resolution: 0.5 μA
Pulse/frequency output
• Galvanically isolated
• Active/passive selectable
– Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100
– Passive: open collector, 30 V DC, 250 mA
• Time constant selectable (0.05 to 100 s)
• Frequency output
– End frequency: 2 to 10000 Hz (fmax = 12500 Hz)
– End frequency for EEx ia 2 to 5000 Hz
– On/off ratio 1:1, pulse width max. 10 s
• Pulse output
– Pulse value and pulse polarity selectable
– Max. pulse width adjustable (0.05 to 2000 ms)
– As of a frequency of 1 / (2 × pulse width), the on/off ratio is 1:1
PROFIBUS DP interface
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding = NRZ Code
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), zero point adjustment, measuring mode, totalizer control
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
• Available output combination ä 9
Endress+Hauser 7
Proline Prosonic Flow 93P
PROFIBUS PA interface
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP)
• Galvanically isolated
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Current consumption = 11 mA
• Error current FDE (fault disconnection electronic) = 0 mA
• Signal encoding = Manchester II
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), operation control, totalizer control, zero point adjustment
control, display value
• Bus address can be set via DIP switch on device
Low flow cutoff Switch points for low flow are selectable
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
8 Endress+Hauser
Proline Prosonic Flow 93P
Power supply
Electrical connection,
measuring unit A B
d
g
b
a
a
b d/(g) (d)
– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20
N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b
A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g
N (L-) 2 c N (L-) 2 c
L1 (L+) 1 L1 (L+) 1
b b
A0013851
Connecting the transmitter, cable cross-section max. 2.5 mm2 (14 AWG)
A Aluminum field housing
B Wall-mount housing
*) Permanent assignment communication boards
**) Flexible assignment communication boards
a Cover of the connection compartment
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 1: N for AC, L- for DC
c Ground terminal for protective ground
d Signal cable: see terminal assignment
Fieldbus:
- Terminal 26: DP (B) / PA (+), FF (+) with reverse polarity protection
- Terminal 27: DP (A) / PA (–), FF (–) with reverse polarity protection
e Ground terminal for signal cable / fieldbus cable
f Service connector for connecting service interface FXA 193 (Fieldcheck, FieldCare)
g Signal cable: see terminal assignment
Cable for external termination (only for PROFIBUS DP with permanent assignment communication board)
- Terminal No. 24: +5V
- Terminal No. 25: DGND
Endress+Hauser 9
Proline Prosonic Flow 93P
Electrical connection, The inputs and outputs on the communication board can be either permanently assigned or variable, depending
terminal assignment on the version ordered (see table). Replacements for modules which are defective or which have to be replaced
can be ordered as accessories.
1 2 3 4
A
A
A0011390
10 Endress+Hauser
Proline Prosonic Flow 93P
4
3
2
1
A
A
A0008314
Sensor
Powered by the transmitter
A0008152
Endress+Hauser 11
Proline Prosonic Flow 93P
Connecting cable (sensor/ Only use the connecting cables supplied by Endress+Hauser.
transmitter)
Different versions of the connecting cables are available ä 27.
• Cable material:
– Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"): PVC (standard) or PTFE (for higher temperatures)
– Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"): TPE-V
• Cable length:
– For use in a non-hazardous zone: 5 to 60 m (16.4 to 196.8 ft)
– For use in a hazardous zone: 5 to 30 m (16.4 to 98.4 ft)
! Note!
To ensure correct measuring results, route the connecting cable well clear of electrical machines and switching
elements.
Performance characteristics
Reference operating • Fluid temperature: +20 to +30 °C
conditions • Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
• Sensors and transmitter are grounded
• The measuring sensors are correctly installed
Maximum measured error The measured error depends on a number of factors. A distinction is made between the measured error of the
device (Prosonic Flow 93 = 0.5 % of the measured value) and an additional installation-specific measured error
(typically 1.5 % of the measured value) that is independent of the device.
The installation-specific measured error depends on the installation conditions on site, such as the nominal
diameter, wall thickness, real pipe geometry, fluid, etc.
The sum of the two measured errors is the measured error at the measuring point.
%
3.5
c
2.0
0.5
a
0 2 4 6 8 10 12 m/s
A0011347
Example of the measured error in a pipe with a nominal diameter DN > 200 (8")
a Measured error of the device (0.5 % o.r. ± 3 mm/s)
b Measured error due to installation conditions (typically 1.5 % o.r.)
c Measured error at the measuring point: 0.5 % o.r. ± 3 mm/s + 1.5 % o.r. = 2 % o.r. ± 3 mm/s
12 Endress+Hauser
Proline Prosonic Flow 93P
DN 25 to 200 (1 to 8") ±0.5 % o.r. ± 7.5 mm/s + ±1.5 % o.r. ±2 % o.r. ± 7.5 mm/s
> DN 200 (8") ±0.5 % o.r. ± 3 mm/s + ±1.5 % o.r. ±2 % o.r. ± 3 mm/s
o.r. = of reading
Measurement Report
If required, the device can be supplied with a factory measurement report. To certify the performance of the
device, a measurement is performed under reference conditions. Here, the sensors are mounted on a pipe with
a nominal diameter of DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") or DN 100 (4") respectively.
The measurement report guarantees the following error limits of the device
[at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]:
! Note!
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
For this reason, avoid the following mounting locations:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipe. Risk of partial pipe filling.
A0001103
Endress+Hauser 13
Proline Prosonic Flow 93P
Orientation
Vertical
Recommended orientation with upward direction of flow (View A). With this orientation, entrained solids will
sink and gases will rise away from the sensor when the fluid is stagnant. The piping can be completely drained
and protected against solids buildup.
Horizontal
In the recommended installation range in a horizontal installation position (View B), gas and air collections at
the pipe cover and problematic deposits at the bottom of the pipe have a smaller influence on measurement.
A C C
B
A0001105
Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following inlet and outlet runs is required in order to ensure measuring accuracy.
³ 15 x DN ³ 3 x DN
³ 20 x DN
1
³ 20 x DN
2
3 ³ 15 x DN
A0013459
14 Endress+Hauser
Proline Prosonic Flow 93P
! Note!
• It is permitted to insulate the sensors mounted on the pipe.
• Mount the transmitter in a shady location and avoid direct sunlight, particularly in warm climatic regions.
Storage temperature The storage temperature corresponds to the ambient temperature range.
Sensor
IP 68 (NEMA 6P)
Electromagnetic compatibility Electromagnetic compatibility (EMC requirements) according to IEC/EN 61326 "Emission to class A
(EMC) requirements" and NAMUR Recommendation NE 21 and NE 43.
Endress+Hauser 15
Proline Prosonic Flow 93P
Medium pressure range No pressure limitation, however perfect measurement requires that the static fluid pressure is higher than vapor
(nominal pressure) pressure.
16 Endress+Hauser
Proline Prosonic Flow 93P
Mechanical construction
Design, dimensions Transmitter wall-mount housing
D
Esc
- + E
B
C
F G
A E
H J K J
N
S S
M
R
L
O
P Q P
A0001150
A B C D E F G H J K
A B C D E F G H J K
8.46 9.84 3.56 6.27 5.31 3.54 1.77 > 1.97 3.18 2.08
1)
L M N O P Q R S T
Endress+Hauser 17
Proline Prosonic Flow 93P
There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory ( ä 27). The following installation variants are possible:
• Panel-mounted installation
• Pipe mounting
Panel-mounted installation
+0.5 (+0.019)
–0.5 (–0.019)
210 (8.27)
+0.5 (+0.019)
–0.5 (–0.019)
245 (9.65)
A0001131
Pipe mounting
Ø 20…70
(Ø 0.79…2.75)
~155 (~ 6.1)
mm (inch)
A0001132
18 Endress+Hauser
Proline Prosonic Flow 93P
Wall mounting
A
A*
B C
B* D
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
E
ep c
Esc
Boucles de courant
c
– +
ve
E
sans sécurité intrinsèque
protégées par Ip40
sou s nsion l Ne
pas o t e pparei
uvrir l’a
M
L
J
K
F
G
A0006999
Dimensions in SI units
A A* B B* C D E F G H J K L M
8.6
265 242 240 217 206 186 178 100 130 100 144 170 355
(M8)
* Blind version (without local display)
All dimensions in [mm]
Dimensions in US units
A A* B B* C D E F G H J K L M
8.6
10.4 9.53 9.45 8.54 8.11 7.32 7.01 3.94 5.12 3.94 5.67 6.69 13.9
(M8)
Endress+Hauser 19
Proline Prosonic Flow 93P
Pipe mounting
240 (9.45)
ht unter Spannun
Nic öffnen g
Nicht-eigensichere
over tight wh
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
Esc
ep c
Boucles de courant – + E
ve
c
sans sécurité intrinsèque
protégées par Ip40
a
mm (inch)
A0005157
20 Endress+Hauser
Proline Prosonic Flow 93P
A C
D
E
F
A0011502
Dimensions in SI units
A B C D E F
Dimensions in US units
A B C D E F
Endress+Hauser 21
Proline Prosonic Flow 93P
F
I
E
C
A
B J D
A0001154
Dimensions in SI units
A B C D E F G H
I J
Dimensions in US units
A B C D E F G H
2.20 2.44 5.71 4.37 2.28 max. 34.3 min. 0.2 16.3 to 31.1
I J
22 Endress+Hauser
Proline Prosonic Flow 93P
E
C
A
B H D
A0001155
Dimensions in SI units
A B C D E F
56 62 145 111 58 max. 872
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
All dimensions in [mm]
Dimensions in US units
A B C D E F
2.20 2.44 5.71 4.37 2.28 max. 34.3
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
All dimensions in [inch]
Endress+Hauser 23
Proline Prosonic Flow 93P
Weight Transmitter
• Wall-mount housing: 6.0 kg (13.2 lbs)
• Field housing: 6.7 kg (14.8 lbs)
Sensor (accessories)
• Prosonic Flow DDU18 (incl. mounting material): 2.4 kg (5.3 lbs)
• Prosonic Flow DDU19 (incl. mounting material): 1.5 kg (3.3 lbs)
! Note!
Weight information without packaging material.
Materials Transmitter
• Wall-mounted housing: powder-coated die-cast aluminum
• Field housing: powder-coated die-cast aluminum
Sensor (accessories)
Prosonic Flow DDU18; Prosonic Flow DDU19
• Sensor holder: stainless steel 1.4308/CF-8
• Sensor housing: stainless steel 1.4301/304
• Strapping bands/bracket: stainless steel 1.4301/304
• Sensor contact surfaces: chemically stable plastic
24 Endress+Hauser
Proline Prosonic Flow 93P
Human interface
Display elements • Liquid crystal display: illuminated, four lines each with 16 characters
• Custom configuration for presenting different measured values and status variables
• 3 totalizers
C-Tick mark The measuring system is in conformity with the EMC requirements of the "Australian Communications and
Media Authority" (ACMA).
Ex approval Information on the currently available Ex-rated versions (ATEX, IECEx, FM, CSA, NEPSI, etc.) can be supplied
by your Endress+Hauser Sales Center on request. All information relevant to explosion protection is available
in separate documents that you can order as necessary.
PROFIBUS DP/PA The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS/DP User Organization). The device thus meets all the requirements of the following
specifications:
• Certified to PROFIBUS DP/PA, profile version 3.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers (interoperability)
FOUNDATION Fieldbus The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 5.01 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus Foundation
Endress+Hauser 25
Proline Prosonic Flow 93P
• ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and Related Equipment -
General Requirements. Pollution Degree 2, Installation Category II.
• CAN/CSA-C22.2 No. 1010.1-92
Safety Requirements for Electrical Equipment for Measurement and Control and Laboratory Use.
Pollution degree 2, Installation Category II
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with analog output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics.
Ordering information
The Endress+Hauser sales and service organization can provide detailed ordering information and information
on the order codes on request.
26 Endress+Hauser
Proline Prosonic Flow 93P
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organization can provide detailed information on the order codes
on request.
Wall-mount housing, Transmitter for replacement or for stock. Use the order code to Single-channel version:
transmitter define the following specifications: 93XXX - XX1XX********
Prosonic Flow 93 • Approvals
• Degree of protection/version Two-channel version:
• Cable entry 93XXX - XX2XX********
• Display / power supply / operation
• Software
• Outputs / inputs
Conversion kit, Conversion kit with appropriate plug-in point modules for DK9UI - **
inputs/outputs converting the current input/output configuration to a new
version.
Sensor P DN 15 to 65 (½ to 2½")
(DN 15 to 65 / ½ to 2½") • –40 to +100 °C (–40 to 212 °F) DK9PS - 1*
Clamp-on version • –40 to +150 °C (–40 to 302 °F) DK9PS - 2*
Mounting kit for field housing Mounting kit for aluminum field housing: DK9WM - B
Suitable for pipe mounting (¾ to 3")
Endress+Hauser 27
Proline Prosonic Flow 93P
5 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBE
10 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBF
15 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBG
30 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBH
Acoustic coupling fluid • Coupling fluid 0 to 170 °C (+32 to 338 °F), DK9CM - 2
high temperature, standard
• Adhesive coupling fluid –40 to +80 °C (–40 to +176 °F) DK9CM - 3
• Water-soluble coupling fluid –20 to +80 °C (–4 to +176 °F) DK9CM - 4
• Coupling fluid DDU 19, –20 to +60 °C (–4 to +140 °F) DK9CM - 6
• Coupling fluid –40 to +80 °C (–40 to +176 °F), standard, DK9CM - 7
type MBG2000
28 Endress+Hauser
Proline Prosonic Flow 93P
Fieldgate FXA320 Gateway for remote interrogation of HART sensors and FXA320 - *****
actuators via Web browser:
• 2-channel analog input (4 to 20 mA)
• 4 binary inputs with event counter function and frequency
measurement
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values
Fieldgate FXA520 Gateway for remote interrogation of HART sensors and FXA520 - ****
actuators via Web browser:
• Web server for remote monitoring of up to 30 measuring
points
• Intrinsically safe version [EEx ia]IIC for applications in
hazardous areas
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values
• Remote diagnosis and remote configuration of
connected HART devices
Endress+Hauser 29
Proline Prosonic Flow 93P
FieldCare FieldCare is Endress+Hauser's FDT-based plant asset See the product page on
management tool. It can configure all intelligent field units in the Endress+Hauser
your system and helps you manage them. website:
By using the status information, it is also a simple but effective www.endress.com
way of checking their status and condition.
FXA193 Service interface from the measuring device to the PC for FXA193 – *
operation via FieldCare.
Communication cable Communication cable for connecting the Prosonic Flow 93P DK9ZT – A
transmitter to the FXA193 service interface.
30 Endress+Hauser
Proline Prosonic Flow 93P
Documentation
• Flow measurement (FA005D)
• Operating Instructions for Prosonic Flow 93 (BA070D and BA071D)
• Operating Instructions for Prosonic Flow 93 PROFIBUS DP/PA (BA076D and BA077D)
• Operating Instructions for Prosonic Flow 93 FOUNDATION Fieldbus (BA078D and BA079D)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI
Registered trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
HistoROM™, T-DAT™, F-CHIP®, FieldCare®, Fieldcheck®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Endress+Hauser 31
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI083D/06/en/07.10
71115159
FM+SGML6.0 ProMoDo