Technical Information

Proline Prosonic Flow 93P


Ultrasonic flow measuring system
Volume flow measurement of liquids in
chemical or process applications

Applications Features and benefits


The sensors are perfectly suited for the non-contact The Prosonic Flow ultrasonic clamp-on system allows
measurement of pure or slightly contaminated liquids, accurate and cost-effective flow measurement from
regardless of the pressure or electrical conductivity. outside the pipe and without the need to interrupt the
process. The flow measurement is bidirectional and
• Suitable for pipe diameters
causes no pressure loss.
from DN 15 to 4000 (½ to 160")
• Suitable for fluid temperatures ranging • Easy, safe and menu-guided sensor mounting
from –40 to +170 °C (–40 to +338 °F) ensures precise measuring results
• Can be used with all metal and plastic pipes lined or • Long-term system integrity thanks to robust sensor
unlined and with composite pipes and industrial mounting kit design
• Ideal solution for applications with chemicals, • Automatic frequency scan for optimized installation
solvents, liquid hydrocarbons, acids and alkalis and maximum measuring performance
• Particularly suitable for corrosive, hazardous or • Remote configuration using Endress+Hauser's
ultrapure fluids thanks to non-contact measurement FieldCare software
concept
• Interface for easy integration into all common
distributed control systems:
– HART
– PROFIBUS DP/PA
– FOUNDATION Fieldbus
• Approval for use in hazardous area Zone 1
(ATEX, FM, CSA) available as standard

TI083D/06/en/07.10
71115159
Proline Prosonic Flow 93P

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Sensor selection and arrangement . . . . . . . . . . . . . . . . . . . . . . . . . 5 Language group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Two-channel operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 25


Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 C-Tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 PROFIBUS DP/PA certification . . . . . . . . . . . . . . . . . . . . . . . . . 25
FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . . 25
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 26
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measuring principle-specific accessories . . . . . . . . . . . . . . . . . . . 27
Communication-specific accessories . . . . . . . . . . . . . . . . . . . . . . 29
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Electrical connection, measuring unit . . . . . . . . . . . . . . . . . . . . . . 9
Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 10 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Connecting the connecting cable . . . . . . . . . . . . . . . . . . . . . . . . 10
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . 31
Connecting cable (sensor/transmitter) . . . . . . . . . . . . . . . . . . . . 12
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Performance characteristics. . . . . . . . . . . . . . . . . . . . 12
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Operating conditions: installation . . . . . . . . . . . . . . . 13


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Operating conditions: environment . . . . . . . . . . . . . . 15


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 15

Operating conditions: process . . . . . . . . . . . . . . . . . . 16


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Medium pressure range (nominal pressure) . . . . . . . . . . . . . . . . . 16
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 17
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

2 Endress+Hauser
Proline Prosonic Flow 93P

Function and system design


Measuring principle The measuring system operates on the principle of transit time difference. In this measurement method,
acoustic (ultrasonic) signals are transmitted between two sensors. The signals are sent in both directions,
i.e. the sensor in question works as both a sound transmitter and a sound receiver.
As the propagation velocity of the waves is less when the waves travel against the direction of flow than along
the direction of flow, a transit time difference occurs. This transit time difference is directly proportional to the
flow velocity.

a
Q b
tb

ta

A0005961

Principle of the transit time difference measurement method


Q=v·A
a Sensor
b Sensor
Q Volume flow
v Flow velocity (v & t )
t Transit time difference (t = ta – tb)
A Pipe cross-sectional area

The measuring system calculates the volume flow of the fluid from the measured transit time difference and
the pipe cross-sectional area. In addition to measuring the transit time difference, the system simultaneously
measures the sound velocity of the fluid. This additional measured variable can be used to distinguish different
fluids or as a measure of product quality.
The measuring device can be configured onsite to suit the specific application using Quick Setup menus.

Measuring system The measuring system consists of one transmitter and two sensors.
The transmitter is used both to control the sensors and to prepare, process and evaluate the measuring signals,
and to convert the signals to a desired output variable.
The transmitter is optionally equipped for two-channel operation  ä 5.
The sensors work as sound transmitters and sound receivers. Depending on the application and version, the
sensors can be arranged for measurement via one or two traverses  ä 5.

Endress+Hauser 3
Proline Prosonic Flow 93P

Transmitter

Prosonic Flow 93 wall-mount housing Prosonic Flow 93 field housing

For mounting in non-hazardous zone, Ex Zone 2 For mounting in Ex Zone 1

A0009629 A0009676

Sensor

Prosonic Flow P Prosonic Flow P

DN 15 to 65 (½ to 2½") DN 50 to 4000 (2 to 160")

A0011484
A0013475

Mounting accessories
The requisite mounting distances must be determined for the sensors. Information on the fluid, the pipe
material used and the exact pipe dimensions is needed to determine these values. The values for the sound
velocity of the following fluids, pipe materials and lining materials are stored in the transmitter:

Fluid Pipe material Lining

• Water • Glycol • Carbon steel • GRP • Mortar


• Sea water • Kerosene • Ductile iron • PVDF • Rubber
• Distilled water • Milk • Stainless steel • PA • Tar Epoxy
• Ammonia • Methanol • Alloy C • PP
• Alcohol • Toluene • PVC • PTFE
• Benzene • Lube oil • PE • Glass pyrex
• Bromide • Diesel • LDPE • Asbestos cement
• Ethanol • Gasoline • HDPE

If you are using fluids, pipe materials or lining materials other than those listed in the table, and you do not
have the corresponding sound velocities for these fluids/materials, you can use the DDU18 and DDU19
sensors to determine the values.

DDU18 DDU19
(sound velocity measurement) (wall thickness measurement)

Nominal diameter range: DN 50 to 3000 (2 to 120") Wall thickness range:


• Steel pipes: 2 to 50 mm (0.08 to 2.0")
• Plastic pipes: 4 to 15 mm (0.16 to 0.60")
(only suitable for PTFE and PE pipes to a certain extent)

A0009784

A0009673

4 Endress+Hauser
Proline Prosonic Flow 93P

Sensor selection and The sensors can be arranged in two ways:


arrangement • Mounting arrangement for measurement via one traverse:
the sensors are located on opposite sides of the pipe.
• Mounting arrangement for measurement via two traverses:
the sensors are located on the same side of the pipe.

A B

A0001108

Sensor mounting arrangement


A Mounting arrangement for measurement via one traverse
B Mounting arrangement for measurement via two traverses

The number of traverses required depends on the sensor type, the nominal diameter and the thickness of the
pipe wall. We recommend the following types of mounting:

Sensor Type Nominal Diameter Sensor Frequency Sensor ID Type of Mounting 1)

DN 15 to 65 (½ to 2½") 6 MHz P-CL-6F* 2 traverses 3)


P-CL-6F*
DN 50 to 65 (2 to 2½") 6 MHz (or 2 MHz) 2 (or 1) traverses 2)
P-CL-2F*

DN 80 (3") 2 MHz P-CL-2F* 2 traverses


Prosonic Flow P P-CL-2F*
DN 100 to 300 (4 to 12") 2 MHz (or 1 MHz) 2 traverses
P-CL-1F*

P-CL-1F*
DN 300 to 600 (12 to 24") 1 MHz (or 2 MHz) 2 traverses
P-CL-2F*

DN 650 to 4000 (26 to 160") 1 MHz P-CL-1F* 1 traverse


1)
The installation of clamp-on sensors is principally recommended in the 2 traverse type installation. This type of
installation allows the easiest and most comfortable type of mounting and means that a system can also be mounted
even if the pipe can only be accessed from one side. However, in certain applications a 1 traverse installation may be
preferred. These include:
• Certain plastic pipes with wall thickness > 4 mm (0.16")
• Pipes made of composite materials such as GRP
• Lined pipes
• Applications with fluids with high acoustic damping
2)
If the pipe nominal diameter is small (DN 65 / 2½" and smaller), the sensor spacing with Prosonic Flow P can be too
small for two traverse installation using sensor P-CL-2F*. In this case, the 1 traverse type of installation must be used.
3) 6 MHz sensors for applications where flow velocity 10 m/s (32.8 ft/s).

Two-channel operation The transmitter is able to operate two independent measuring channels (measuring channel 1 and measuring
channel 2). A pair of sensors is connected per measuring channel. Both measuring channels operate
independently of one another and are supported by the transmitter to an equal extent.
Two-channel operation can be used for the following measurements:
• Two-channel measurement = flow measurement at two separate measuring points
• Two-path measurement = redundant flow measurement at one measuring point

Endress+Hauser 5
Proline Prosonic Flow 93P

Two-channel measurement
The flow is measured at two separate measuring points in the case of two-channel measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-channel measurement:
• Individual measured values per measuring channel (output independently of one another)
• The difference between the two measured values
• The sum of the two measured values
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.

A B

A0001159

Two-channel measurement: example of arranging sensor pairs at two separate measuring points
A Measuring channel 1: mounting the sensor pair for measurement via two traverses
B Measuring channel 2: mounting the sensor pair for measurement via one traverse

Two-path measurement
The flow is measured redundantly at one measuring point in the case of two-path measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-path measurement:
• Individual measured values per measuring channel (output independently of one another)
• The average of the two measured values
The "Averaging" function generally provides you with a more stable measured value. The function is thus
suitable for measurements under conditions that are not ideal (e.g. short inlet runs).
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
It is generally not necessary to individually configure the two measuring channels in the case of two-path
measurement. However, in certain situations individual channel configuration can be used to balance out
application-specific asymmetries.

A B

A0001160

Two-path measurement: examples of arranging sensor pairs at one measuring point


A Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via two traverses
B Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via one traverse

6 Endress+Hauser
Proline Prosonic Flow 93P

Input
Measured variable Flow velocity (transit time difference proportional to flow velocity)

Measuring range Typically v = 0 to 15 m/s (0 to 50 ft/s)

Operable flow range Over 150 : 1

Input signal Status input (auxiliary input)


U = 3 to 30 V DC, Ri = 5 k, galvanically isolated
Configurable for:
totalizer(s) reset, measured-value suppression, error-message reset

Output
Output signal Current output
• Galvanically isolated
• Active/passive selectable
– Active: 0/4 to 20 mA, RL < 700  (for HART: RL 250 )
– Passive: 4 to 20 mA, max. 30 V DC, Ri 150 
• Time constant selectable (0.01 to 100 s)
• Full scale value adjustable
• Temperature coefficient: typ. 0.005 % o.r./°C (o.r. = of reading)
• Resolution: 0.5 μA

Pulse/frequency output
• Galvanically isolated
• Active/passive selectable
– Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 
– Passive: open collector, 30 V DC, 250 mA
• Time constant selectable (0.05 to 100 s)
• Frequency output
– End frequency: 2 to 10000 Hz (fmax = 12500 Hz)
– End frequency for EEx ia 2 to 5000 Hz
– On/off ratio 1:1, pulse width max. 10 s
• Pulse output
– Pulse value and pulse polarity selectable
– Max. pulse width adjustable (0.05 to 2000 ms)
– As of a frequency of 1 / (2 × pulse width), the on/off ratio is 1:1

PROFIBUS DP interface
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding = NRZ Code
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), zero point adjustment, measuring mode, totalizer control
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
• Available output combination  ä 9

Endress+Hauser 7
Proline Prosonic Flow 93P

PROFIBUS PA interface
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP)
• Galvanically isolated
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Current consumption = 11 mA
• Error current FDE (fault disconnection electronic) = 0 mA
• Signal encoding = Manchester II
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), operation control, totalizer control, zero point adjustment
control, display value
• Bus address can be set via DIP switch on device

FOUNDATION Fieldbus interface


• FOUNDATION Fieldbus H1, IEC 61158-2
• Galvanically isolated
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Current consumption = 12 mA
• Error current FDE (fault disconnection electronic) = 0 mA
• Signal encoding = Manchester II
• Function blocks: 8 × Analog Input (AI), 1 × Discrete Output, 1 × PID
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, signal strength channel 1 or 2, average volume flow,
average sound velocity, average flow velocity, volume flow sum, difference, volume flow,
totalizer 1 to 3
• Input data: positive zero return (ON/OFF), reset totalizer, zero point adjustment control
• Link master function (LAS) is supported

Signal on alarm • Current output  failsafe mode selectable


• Pulse/frequency output  failsafe mode selectable
• Relay output  "deenergized" in the event of a fault or if the power supply fails

Load See "Output signal"

Switching output Relay output


• NC or NO contact available
Factory setting: relay 1 = NO contact, relay 2 = NC contact
• Max. 30 V / 0.5 A AC; 60 V / 0.1 A DC
• Galvanically isolated
• Configurable for: error messages, flow direction, limit values

Low flow cutoff Switch points for low flow are selectable

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.

8 Endress+Hauser
Proline Prosonic Flow 93P

Power supply
Electrical connection,
measuring unit A B

d
g
b

a
a
b d/(g) (d)

HART* PROFIBUS PA*


FOUNDATION Fieldbus*

– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20

N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b

PROFIBUS DP* PROFIBUS DP**

A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g

N (L-) 2 c N (L-) 2 c
L1 (L+) 1 L1 (L+) 1
b b

A0013851

Connecting the transmitter, cable cross-section max. 2.5 mm2 (14 AWG)
A Aluminum field housing
B Wall-mount housing
*) Permanent assignment communication boards
**) Flexible assignment communication boards
a Cover of the connection compartment
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 1: N for AC, L- for DC
c Ground terminal for protective ground
d Signal cable: see terminal assignment
Fieldbus:
- Terminal 26: DP (B) / PA (+), FF (+) with reverse polarity protection
- Terminal 27: DP (A) / PA (–), FF (–) with reverse polarity protection
e Ground terminal for signal cable / fieldbus cable
f Service connector for connecting service interface FXA 193 (Fieldcheck, FieldCare)
g Signal cable: see terminal assignment
Cable for external termination (only for PROFIBUS DP with permanent assignment communication board)
- Terminal No. 24: +5V
- Terminal No. 25: DGND

Endress+Hauser 9
Proline Prosonic Flow 93P

Electrical connection, The inputs and outputs on the communication board can be either permanently assigned or variable, depending
terminal assignment on the version ordered (see table). Replacements for modules which are defective or which have to be replaced
can be ordered as accessories.

Order variant Terminal No. (inputs/outputs)


20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
Fixed communication boards (fixed assignment)
93***-***********A – – Frequency output Current output HART
93***-***********B Relay output 2 Relay output 1 Frequency output Current output HART
93***-***********F – – – PROFIBUS PA, Ex i
93***-***********G – – – FOUNDATION Fieldbus, Ex i
93***-***********H – – – PROFIBUS PA
93***-***********J – – – PROFIBUS DP
93***-***********K – – – FOUNDATION Fieldbus
93***-***********S – – Frequency output, Ex i Current output, Ex i,
passive, HART
93***-***********T – – Frequency output, Ex i Current output, Ex i,
passive, HART
Flexible communication boards
93***-***********C Relay output 2 Relay output 1 Frequency output Current output HART
93***-***********D Status input Relay output Frequency output Current output HART
93***-***********L Status input Relay output 2 Relay output 1 Current output HART
93***-***********M Status input Frequency output Frequency output Current output HART
93***-***********P Current output Frequency output Status input PROFIBUS DP
93***-***********V Relay output 2 Relay output 1 Status input PROFIBUS DP
93***-***********W Relay output Current output Current output Current output HART
93***-***********2 Relay output Current output Frequency output Current output HART
93***-***********4 Current input Relay output Frequency output Current output HART
93***-***********6 Relay output Relay output Current output Current output HART

Connecting the connecting Connecting the wall-mount housing


cable

1 2 3 4

A
A

A0011390

Connecting the connecting cable, wall-mount housing


1 Channel 1 upstream
2 Channel 1 downstream
3 Channel 2 upstream (only available when ordered)
4 Channel 2 downstream (only available when ordered)

10 Endress+Hauser
Proline Prosonic Flow 93P

Connecting the field housing

4
3

2
1
A
A
A0008314

Connecting the connecting cable, field housing


1 Channel 1 upstream
2 Channel 1 downstream
3 Channel 2 upstream (only available when ordered)
4 Channel 2 downstream (only available when ordered)

Supply voltage Transmitter


HART
• 85 to 260 V AC, 45 to 65 Hz
• 20 to 55 V AC, 45 to 65 Hz
• 16 to 62 V DC
PROFIBUS DP/PA
• 9 to 32 V DC
FOUNDATION Fieldbus
• 9 to 32 V DC

Sensor
Powered by the transmitter

Cable entry Power supply and signal cables (inputs/outputs)


• Cable gland M20 × 1.5
– Cable gland for cables 8 to 12 mm (0.31 to 0.47")
– Cable gland for cables 6 to 12 mm (0.24 to 0.47")
• Thread for cable entry ½" NPT, G ½"

Connecting cable (sensor/transmitter)


Cable gland for one connecting cable per cable entry, 1 × 8 mm (1 × 0.31")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"
Cable gland for two connecting cables per cable entry, 2 × 4 mm (2 × 0.16")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"

A0008152

Cable gland for two connecting cables per cable entry

Endress+Hauser 11
Proline Prosonic Flow 93P

Connecting cable (sensor/ Only use the connecting cables supplied by Endress+Hauser.
transmitter)
Different versions of the connecting cables are available  ä 27.
• Cable material:
– Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"): PVC (standard) or PTFE (for higher temperatures)
– Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"): TPE-V
• Cable length:
– For use in a non-hazardous zone: 5 to 60 m (16.4 to 196.8 ft)
– For use in a hazardous zone: 5 to 30 m (16.4 to 98.4 ft)

! Note!
To ensure correct measuring results, route the connecting cable well clear of electrical machines and switching
elements.

Power consumption AC: < 18 VA (incl. sensor)


DC: < 10 W (incl. sensor)
Switch-on current
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC

Power supply failure Lasting min. 1 power cycle


HistoROM/T-DAT save measuring system data if the power supply fails

Potential equalization For potential equalization, no special measures are necessary.

Performance characteristics
Reference operating • Fluid temperature: +20 to +30 °C
conditions • Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
• Sensors and transmitter are grounded
• The measuring sensors are correctly installed

Maximum measured error The measured error depends on a number of factors. A distinction is made between the measured error of the
device (Prosonic Flow 93 = 0.5 % of the measured value) and an additional installation-specific measured error
(typically 1.5 % of the measured value) that is independent of the device.
The installation-specific measured error depends on the installation conditions on site, such as the nominal
diameter, wall thickness, real pipe geometry, fluid, etc.
The sum of the two measured errors is the measured error at the measuring point.

%
3.5

c
2.0

0.5
a
0 2 4 6 8 10 12 m/s

A0011347

Example of the measured error in a pipe with a nominal diameter DN > 200 (8")
a Measured error of the device (0.5 % o.r. ± 3 mm/s)
b Measured error due to installation conditions (typically 1.5 % o.r.)
c Measured error at the measuring point: 0.5 % o.r. ± 3 mm/s + 1.5 % o.r. = 2 % o.r. ± 3 mm/s

12 Endress+Hauser
Proline Prosonic Flow 93P

Measured error at the measuring point


The measured error at the measuring point is made up of the measured error of the device (0.5 % o.r.) and the
measured error resulting from the installation conditions on site. Given a flow velocity > 0.3 m/s (1 ft/s) and
a Reynolds number > 10000, the following are typical error limits:

Installation-specific Error limits at the measuring


Nominal diameter Device error limits + 
error limits (typical) point (typical)

DN 15 (½") ±0.5 % o.r. ± 5 mm/s + ±2.5 % o.r.  ±3 % o.r. ± 5 mm/s

DN 25 to 200 (1 to 8") ±0.5 % o.r. ± 7.5 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 7.5 mm/s

> DN 200 (8") ±0.5 % o.r. ± 3 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 3 mm/s
o.r. = of reading

Measurement Report
If required, the device can be supplied with a factory measurement report. To certify the performance of the
device, a measurement is performed under reference conditions. Here, the sensors are mounted on a pipe with
a nominal diameter of DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") or DN 100 (4") respectively.

The measurement report guarantees the following error limits of the device
[at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]:

Nominal diameter Guaranteed error limits of the device

DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") ±0.5 % o.r. ± 5 mm/s

DN 100 (4") ±0.5 % o.r. ± 7.5 mm/s


o.r. = of reading

Repeatability ±0.3 % for flow velocities > 0.3 m/s (1 ft/s)

Operating conditions: installation


Installation instructions Mounting location
Correct flow measurement is possible only if the pipe is full. It is preferable to install the sensors in a riser.

! Note!
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
For this reason, avoid the following mounting locations:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipe. Risk of partial pipe filling.

A0001103

Endress+Hauser 13
Proline Prosonic Flow 93P

Orientation

Vertical
Recommended orientation with upward direction of flow (View A). With this orientation, entrained solids will
sink and gases will rise away from the sensor when the fluid is stagnant. The piping can be completely drained
and protected against solids buildup.

Horizontal
In the recommended installation range in a horizontal installation position (View B), gas and air collections at
the pipe cover and problematic deposits at the bottom of the pipe have a smaller influence on measurement.

A C C

B
A0001105

A Recommended orientation with upward direction of flow


B Recommended installation range with horizontal orientation
C Recommended installation range max. 120°

Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following inlet and outlet runs is required in order to ensure measuring accuracy.

³ 15 x DN ³ 3 x DN

³ 20 x DN
1

³ 20 x DN
2

3 ³ 15 x DN

A0013459

Inlet and outlet runs (top view)


1 Valve (2/3 open)
2 Pump
3 Two pipe bends in different directions

14 Endress+Hauser
Proline Prosonic Flow 93P

Operating conditions: environment

Ambient temperature range Transmitter


• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)

Prosonic Flow P sensor


Prosonic Flow P (DN 15 to 65 / ½ to 2½")
• Standard: –40 to +100 °C (–40 to +212 °F)
• Optional: –40 to +150 °C (–40 to +302 °F)
Prosonic Flow P (DN 50 to 4000 / 2 to 160")
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)

DDU18 sensor (accessories: sound velocity measurement)


• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)

DDU19 sensor (accessories: wall thickness measurement)


–20 to +60 °C (–4 to +140 °F)

Connecting cable (sensor/transmitter)


Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"):
• Standard (TPE-V): –40 to +80 °C (–40 to +176 °F)
Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"):
• Standard (PVC): –20 to +70 °C (–4 to +158 °F)
• Optional (PTFE): –40 to +170 °C (–40 to +338 °F)

! Note!
• It is permitted to insulate the sensors mounted on the pipe.
• Mount the transmitter in a shady location and avoid direct sunlight, particularly in warm climatic regions.

Storage temperature The storage temperature corresponds to the ambient temperature range.

Degree of protection Transmitter


IP 67 (NEMA 4X)

Sensor
IP 68 (NEMA 6P)

DDU18 sensor (accessories: sound velocity measurement)


IP 68 (NEMA 6P)

DDU19 sensor (accessories: wall thickness measurement)


IP 67 (NEMA 4X)

Shock and vibration resistance According to IEC 68-2-6

Electromagnetic compatibility Electromagnetic compatibility (EMC requirements) according to IEC/EN 61326 "Emission to class A
(EMC) requirements" and NAMUR Recommendation NE 21 and NE 43.

Endress+Hauser 15
Proline Prosonic Flow 93P

Operating conditions: process


Medium temperature range Prosonic Flow P sensor
Prosonic Flow P (DN 15 to 65)
• Standard: –40 to +100 °C (–40 to +212 °F)
• Optional: –40 to +150 °C (–40 to +302 °F)
Prosonic Flow P (DN 50 to 4000)
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)

DDU18 sensor (accessories: sound velocity measurement)


• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)

DDU19 sensor (accessories: wall thickness measurement)


–20 to +60 °C (–4 to +140 °F)

Medium pressure range No pressure limitation, however perfect measurement requires that the static fluid pressure is higher than vapor
(nominal pressure) pressure.

Pressure loss There is no pressure loss.

16 Endress+Hauser
Proline Prosonic Flow 93P

Mechanical construction
Design, dimensions Transmitter wall-mount housing

D
Esc
- + E

B
C
F G
A E

H J K J

N
S S

M
R

L
O
P Q P

A0001150

Dimensions (SI units)

A B C D E F G H J K

215 250 90.5 159.5 135 90 45 > 50 81 53


1)
L M N O P Q R S T

95 53 102 81.5 11.5 192 8 × M5 20 2 ×  6.5


1)Securing screw for wall mounting: M6 (screw head max. 10.5 mm)
All dimensions in [mm]

Dimensions (US units)

A B C D E F G H J K

8.46 9.84 3.56 6.27 5.31 3.54 1.77 > 1.97 3.18 2.08
1)
L M N O P Q R S T

3.74 2.08 4.01 3.20 0.45 7.55 8 × M5 0.79 2 ×  0.26


1)
Securing screw for wall mounting: M6 (screw head max. 0.41")
All dimensions in [inch]

Endress+Hauser 17
Proline Prosonic Flow 93P

There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory ( ä 27). The following installation variants are possible:
• Panel-mounted installation
• Pipe mounting

Panel-mounted installation

+0.5 (+0.019)
–0.5 (–0.019)
210 (8.27)

+0.5 (+0.019)
–0.5 (–0.019)
245 (9.65)

mm (inch) ~110 (~4.33)

A0001131

Pipe mounting

Ø 20…70
(Ø 0.79…2.75)

~155 (~ 6.1)

mm (inch)

A0001132

18 Endress+Hauser
Proline Prosonic Flow 93P

Transmitter Field housing


There is a separate mounting kit for the field housing. It can be ordered from Endress+Hauser as an accessory
( ä 27). The following installation variants are possible:
• Wall mounting
• Pipe mounting

Wall mounting

A
A*
B C
B* D

e r Sp annu
ht unt ng
Nic öffnen

Nicht-eigensichere

Ke ircuit s a re aliv ile

Kee ircuits are ali hile


Stromkreise durch

ove r tight wh
e

c
IP40-Abdeckung geschützt

p cover tight w
Non-intrinsically safe
circuits Ip40 protected

E
ep c
Esc
Boucles de courant

c
– +

ve
E
sans sécurité intrinsèque
protégées par Ip40

sou s nsion l Ne
pas o t e pparei
uvrir l’a

M
L
J

K
F
G

A0006999

Dimensions in SI units

A A* B B* C D E F G H J K L M

8.6
265 242 240 217 206 186 178 100 130 100 144 170 355
(M8)
* Blind version (without local display)
All dimensions in [mm]

Dimensions in US units

A A* B B* C D E F G H J K L M

8.6
10.4 9.53 9.45 8.54 8.11 7.32 7.01 3.94 5.12 3.94 5.67 6.69 13.9
(M8)

* Blind version (without local display)


All dimensions in [inch]

Endress+Hauser 19
Proline Prosonic Flow 93P

Pipe mounting

240 (9.45)

ht unter Spannun
Nic öffnen g

Nicht-eigensichere

Ke ircuits are alive le

Kee ircuits are ali hile


i
Stromkreise durch

over tight wh

c
IP40-Abdeckung geschützt

p cover tight w
Non-intrinsically safe
circuits Ip40 protected
Esc

ep c
Boucles de courant – + E

ve
c
sans sécurité intrinsèque
protégées par Ip40

Ne sous tension reil


pas ou r l’appa
vri

a
mm (inch)

A0005157

a = Pipe mounting (separate mounting set, see accessories  ä 27)

20 Endress+Hauser
Proline Prosonic Flow 93P

Prosonic Flow P sensor (DN 15 to 65 / ½ to 2½")

A C
D

E
F
A0011502

Dimensions in SI units

A B C D E F

72 331 39 28 233 450

All dimensions in [mm]

Dimensions in US units

A B C D E F

2.83 13.03 1.54 1.10 9.17 17.72

All dimensions in [inch]

Endress+Hauser 21
Proline Prosonic Flow 93P

Prosonic Flow P sensor (DN 50 to 4000 / 2 to 160")

F
I

E
C
A

B J D

A0001154

Mounting arrangement for measurement via two traverses

Dimensions in SI units

A B C D E F G H

56 62 145 111  58 max. 872 min. 0.5 439 to 790

I J

Depends on the measuring point conditions (pipe, fluid, etc.).


Dimension "I" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)

All dimensions in [mm]

Dimensions in US units

A B C D E F G H

2.20 2.44 5.71 4.37  2.28 max. 34.3 min. 0.2 16.3 to 31.1
I J

Depends on the measuring point conditions (pipe, fluid, etc.).


Dimension "I" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)

All dimensions in [inch]

22 Endress+Hauser
Proline Prosonic Flow 93P

E
C
A
B H D

A0001155

Mounting arrangement for measurement via one traverse

Dimensions in SI units
A B C D E F
56 62 145 111  58 max. 872
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
All dimensions in [mm]

Dimensions in US units
A B C D E F
2.20 2.44 5.71 4.37  2.28 max. 34.3
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
Pipe outer diameter
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
All dimensions in [inch]

Endress+Hauser 23
Proline Prosonic Flow 93P

Weight Transmitter
• Wall-mount housing: 6.0 kg (13.2 lbs)
• Field housing: 6.7 kg (14.8 lbs)

Sensor Prosonic Flow P


• DN 15 to 65 (½ to 2½") (incl. mounting material): 1.2 kg (2.65 lbs)
• DN 50 to 4000 (2 to 160") (incl. mounting material): 2.8 kg (6.2 lbs)

Sensor (accessories)
• Prosonic Flow DDU18 (incl. mounting material): 2.4 kg (5.3 lbs)
• Prosonic Flow DDU19 (incl. mounting material): 1.5 kg (3.3 lbs)

! Note!
Weight information without packaging material.

Materials Transmitter
• Wall-mounted housing: powder-coated die-cast aluminum
• Field housing: powder-coated die-cast aluminum

Sensor Prosonic Flow P


DN 15 to 65 (½ to 2½"); DN 50 to 4000 (2 to 160")
• Sensor holder: stainless steel 1.4308/CF-8
• Sensor housing: stainless steel 1.4301/304
• Strapping bands/bracket: stainless steel 1.4301/304
• Sensor contact surfaces: chemically stable plastic

Sensor (accessories)
Prosonic Flow DDU18; Prosonic Flow DDU19
• Sensor holder: stainless steel 1.4308/CF-8
• Sensor housing: stainless steel 1.4301/304
• Strapping bands/bracket: stainless steel 1.4301/304
• Sensor contact surfaces: chemically stable plastic

Connecting cable (sensor/transmitter)


Prosonic Flow 93P (DN 15 to 65 / ½ to 2½")
• TPE-V connecting cable
– Cable sheath: TPE-V
– Cable connector: stainless steel 1.40301
Prosonic Flow 93P (DN 50 to 4000 / 2 to 160")
• PVC connecting cable
– Cable sheath: PVC
– Cable connector: nickeled brass 2.0401
• PTFE connecting cable
– Cable sheath: PTFE
– Cable connector: stainless steel 1.4301

24 Endress+Hauser
Proline Prosonic Flow 93P

Human interface
Display elements • Liquid crystal display: illuminated, four lines each with 16 characters
• Custom configuration for presenting different measured values and status variables
• 3 totalizers

Operating elements • Local operation with three optical keys


• Application specific Quick Setup menus for straightforward commissioning

Language group Language groups available for operation in different countries:


• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and Eastern Asia (SEA):
English, Japanese, Indonesian
• China (CN):
English, Chinese
You can change the language group via the FieldCare operating program.

Remote operation Operation via HART, PROFIBUS DP/PA, FOUNDATION Fieldbus

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

C-Tick mark The measuring system is in conformity with the EMC requirements of the "Australian Communications and
Media Authority" (ACMA).

Ex approval Information on the currently available Ex-rated versions (ATEX, IECEx, FM, CSA, NEPSI, etc.) can be supplied
by your Endress+Hauser Sales Center on request. All information relevant to explosion protection is available
in separate documents that you can order as necessary.

PROFIBUS DP/PA The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS/DP User Organization). The device thus meets all the requirements of the following
specifications:
• Certified to PROFIBUS DP/PA, profile version 3.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers (interoperability)

FOUNDATION Fieldbus The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 5.01 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus Foundation

Other standards and • EN 60529


guidelines Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use
• IEC/EN 61326
"Emission in accordance with Class A requirements".
Electromagnetic compatibility (EMC requirements).

Endress+Hauser 25
Proline Prosonic Flow 93P

• ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and Related Equipment -
General Requirements. Pollution Degree 2, Installation Category II.
• CAN/CSA-C22.2 No. 1010.1-92
Safety Requirements for Electrical Equipment for Measurement and Control and Laboratory Use.
Pollution degree 2, Installation Category II
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with analog output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics.

Ordering information
The Endress+Hauser sales and service organization can provide detailed ordering information and information
on the order codes on request.

26 Endress+Hauser
Proline Prosonic Flow 93P

Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organization can provide detailed information on the order codes
on request.

Device-specific accessories Accessory Description Order code

Wall-mount housing, Transmitter for replacement or for stock. Use the order code to Single-channel version:
transmitter define the following specifications: 93XXX - XX1XX********
Prosonic Flow 93 • Approvals
• Degree of protection/version Two-channel version:
• Cable entry 93XXX - XX2XX********
• Display / power supply / operation
• Software
• Outputs / inputs

Conversion kit, Conversion kit with appropriate plug-in point modules for DK9UI - **
inputs/outputs converting the current input/output configuration to a new
version.

Sensor P DN 15 to 65 (½ to 2½")
(DN 15 to 65 / ½ to 2½") • –40 to +100 °C (–40 to 212 °F) DK9PS - 1*
Clamp-on version • –40 to +150 °C (–40 to 302 °F) DK9PS - 2*

Sensor P DN 50 to 300 (2 to 12")


(DN 50 to 4000 / 2 to 160") • –40 to +80 °C (–40 to 176 °F) DK9PS - B*
Clamp-on version • 0 to +170 °C (+32 to 338 °F) DK9PS - F*

DN 100 to 4000 (4 to 160")


• –40 to +80 °C (–40 to 176 °F) DK9PS - A*
• 0 to +170 °C (+32 to 338 °F) DK9PS - E*

Sensor DDU18 Sensor for sound velocity measurement


• –40 to +80 °C (–40 to 176 °F) 50091703
• 0 to +170 °C (+32 to 338 °F) 50091704

Sensor DDU19 Sensor for wall thickness measurement. 50091713

Measuring principle-specific Accessory Description Order code


accessories
Mounting kit for Mounting kit for wall-mount housing. DK9WM - A
aluminum field housing Suitable for:
• Wall mounting
• Pipe mounting
• Panel mounting

Mounting kit for field housing Mounting kit for aluminum field housing: DK9WM - B
Suitable for pipe mounting (¾ to 3")

Sensor holder set • Prosonic Flow P (DN 15 to 65 / ½ to 2½"): DK9SH - 1


Sensor holder, clamp-on version

• Prosonic Flow P (DN 50 to 4000 / 2 to 160")


– Sensor holder, fixed retaining nut, clamp-on version DK9SH - A
– Sensor holder, removable retaining nut, clamp-on version DK9SH - B

Endress+Hauser 27
Proline Prosonic Flow 93P

Accessory Description Order code

Clamp-on installation set Sensor fastening for Prosonic Flow P


(DN 15 to 65 / ½ to 2½")
• U-shaped screw DN 15 to 32 (½ to 1¼") DK9IC - 1*
• Strapping bands DN 40 to 65 (1½ to 2½") DK9IC - 2*

Sensor fastening for Prosonic Flow P


(DN 50 to 4000 / 2 to 160")
• Without sensor fastening DK9IC - A*
• Strapping bands DN 50 to 200 (2 to 8") DK9IC - B*
• Strapping bands DN 200 to 600 (8 to 24") DK9IC - C*
• Strapping bands DN 600 to 2000 (24 to 80") DK9IC - D*
• Strapping bands DN 2000 to 4000 (80 to 160") DK9IC - E*

• Without mounting tools DK9IC - *1


• Spacing ruler DN 50 to 200 (2 to 8") DK9IC - *2
• Spacing ruler DN 200 to 600 (8 to 24") DK9IC - *3
• Mounting rail DN 50 to 200 (2 to 8") DK9IC - *4
• Mounting rail DN 200 to 600 (8 to 24") DK9IC - *5
• Fastening, 1 traverse DN 50 to 4000 (2 to 160") DK9IC - *6

Conduit adapter for Prosonic Flow P (DN 15 to 65 / ½ to 2½")


connecting cable • Conduit adapter incl. cable entry M20 × 1.5 DK9CB - BA1
• Conduit adapter incl. cable entry ½" NPT DK9CB - BA2
• Conduit adapter incl. cable entry G ½" DK9CB - BA3

Prosonic Flow P (DN 50 to 4000 / 2 to 160")


• Conduit adapter incl. cable entry M20 × 1.5 DK9CB - BB1
• Conduit adapter incl. cable entry ½" NPT DK9CB - BB2
• Conduit adapter incl. cable entry G ½" DK9CB - BB3

Connecting cable Prosonic Flow P (DN 15 to 65 / ½ to 2½")


5 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) DK9SS - BAA
10 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) DK9SS - BAB
15 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) DK9SS - BAC
30 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) DK9SS - BAD

Prosonic Flow P (DN 50 to 4000 / 2 to 160")


5 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) DK9SS - BBA
10 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) DK9SS - BBB
15 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) DK9SS - BBC
30 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) DK9SS - BBD

5 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBE
10 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBF
15 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBG
30 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBH

Acoustic coupling fluid • Coupling fluid 0 to 170 °C (+32 to 338 °F), DK9CM - 2
high temperature, standard
• Adhesive coupling fluid –40 to +80 °C (–40 to +176 °F) DK9CM - 3
• Water-soluble coupling fluid –20 to +80 °C (–4 to +176 °F) DK9CM - 4
• Coupling fluid DDU 19, –20 to +60 °C (–4 to +140 °F) DK9CM - 6
• Coupling fluid –40 to +80 °C (–40 to +176 °F), standard, DK9CM - 7
type MBG2000

28 Endress+Hauser
Proline Prosonic Flow 93P

Communication-specific Accessory Description Order code


accessories
HART handheld terminal Handheld terminal for remote configuration and for obtaining DXR375 - *******
DXR375 measured values via the HART current output (4 to 20 mA)
and FOUNDATION Fieldbus.

Contact your Endress+Hauser representative for more


information.

Fieldgate FXA320 Gateway for remote interrogation of HART sensors and FXA320 - *****
actuators via Web browser:
• 2-channel analog input (4 to 20 mA)
• 4 binary inputs with event counter function and frequency
measurement
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values

Fieldgate FXA520 Gateway for remote interrogation of HART sensors and FXA520 - ****
actuators via Web browser:
• Web server for remote monitoring of up to 30 measuring
points
• Intrinsically safe version [EEx ia]IIC for applications in
hazardous areas
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values
• Remote diagnosis and remote configuration of
connected HART devices

Endress+Hauser 29
Proline Prosonic Flow 93P

Service-specific accessories Accessory Description Order code

Applicator Software for selecting and planning flowmeters. DXA80 - *


The Applicator can be downloaded from the Internet or
ordered on CD-ROM for installation on a local PC.

Contact your Endress+Hauser representative for more


information.

Fieldcheck Tester/simulator for testing flowmeters in the field. 50098801


When used in conjunction with the "FieldCare" software
package, test results can be imported into a database, printed
out and used for official certification.

Contact your Endress+Hauser representative for more


information.

FieldCare FieldCare is Endress+Hauser's FDT-based plant asset See the product page on
management tool. It can configure all intelligent field units in the Endress+Hauser
your system and helps you manage them. website:
By using the status information, it is also a simple but effective www.endress.com
way of checking their status and condition.

FXA193 Service interface from the measuring device to the PC for FXA193 – *
operation via FieldCare.

Communication cable Communication cable for connecting the Prosonic Flow 93P DK9ZT – A
transmitter to the FXA193 service interface.

30 Endress+Hauser
Proline Prosonic Flow 93P

Documentation
• Flow measurement (FA005D)
• Operating Instructions for Prosonic Flow 93 (BA070D and BA071D)
• Operating Instructions for Prosonic Flow 93 PROFIBUS DP/PA (BA076D and BA077D)
• Operating Instructions for Prosonic Flow 93 FOUNDATION Fieldbus (BA078D and BA079D)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI

Registered trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
HistoROM™, T-DAT™, F-CHIP®, FieldCare®, Fieldcheck®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Endress+Hauser 31