Variotherm Plus
Variotherm Plus
Variotherm plus
Operating Instructions
0124
GA1GB.140302.0 2017-03 Index 30
Valid for devices delivered after July 2002.
1.0 Introduction..........................................................3
1.1 Notes on operating instructions.............................3
1.2 Intended use..........................................................4
1.3 Function.................................................................4
1.4 Explanation of symbols..........................................4
4.0 Operation............................................................13
4.1 Adjusting temperatures........................................13
4.2 Selecting temperature levels...............................13
4.3 Adjusting stimulation time....................................13
4.4 Adjusting flow quantity.........................................13
4.5 Description of operating modes...........................14
4.5.1 Thermal process for reducing bacterial count......14
4.5.2 Irrigation mode.....................................................14
4.5.3 Stimulation mode.................................................14
4.5.4 Energy saving mode............................................14
7.0 Troubleshooting.................................................18
10.0 Disposal..............................................................22
ATMOS
• Indicating a list
-- Subdivision of a list/activity
The recommended sequence must be followed in each
MedizinTechnik GmbH & Co. KG case!
Ludwig-Kegel-Straße 16
79853 Lenzkirch )) Indicating particularly important advice!
Germany
ªª Describing the effect of an activity.
Phone: + 49 7653 689-0
Fax:
+ 49 7653 689-190
+ 49 7653 689-393 (Service Centre)
E-mail: [email protected]
Internet: www.atmosmed.de
1.3 Function
• Heating OFF (energy saving mode)
• The Variotherm plus is produced according to IEC 601/ • Only use the warm water jet when a tube tip is installed!
EN 60601 and listed in the following classes: • Take caution to avoid injury to the eardrum when
-- VDE Class of protection 1 introducing the water jet!
-- Class IIa (EEC 93/42).
• For hygienic reasons jet connections must be changed
• The device may only be connected to a properly installed after each patient. This also prevents retrograde
safety socket. germination of the warm water system.
• The unit must be installed by a specialist authorized by • Use only for irrigation of auditory canal!
ATMOS (see chapter 3.3).
• ATMOS is not liable for personal injury and damage to
• The Variotherm plus may only be used under the property if
supervision of skilled staff who have been authorised -- no original ATMOS parts are being used,
by ATMOS and trained in its operation (IEC 601-1 / EN -- the advice for use in these operating instructions is
60601-1). not being observed,
• The mains voltage indicated on the type plate must -- assembly, new settings, alterations, extensions
correspond to the values of the supply network. and repairs have been carried out by personnel not
• Make sure prior to every application of the equipment that authorised by ATMOS.
it is technically safe and in proper condition. Damaged • Please note:
cables must be replaced immediately! A medical insulating transformer with earth leakage
• Correct configuration in assembly of country-specific monitor or any similar safety system acc. to EN 60 601-1
connections: is required, if several devices are connected over one
-- green/yellow: protective conductor (PE) common power supply. The transformer must correspond
-- blue: neutral conductor (N) to the power consumption of all the devices to be
-- black or brown: phase (L) connected.
• It is essential that the country-specific requirements for • The flow rate has to be checked after switch on or at
connection of medical equipment to the public drinking least once a day with a measuring cup. The quantity of
water supply are considered. When in doubt, please 500 ml/min may not be exceeded. The water jet has to be
contact your local ATMOS partner. straight.
• The control panel must be clearly visible and accessible
for the user. Ensure sufficient stability on the installation
surface.
)) In the thermal process for reducing bacterial count hot
water is being ducted through the irrigation handle.
Please do not take the handle from its holder or spray
water!
)) Metal parts can be hot!
)) Prior to spraying, the water temperature must be checked
by the user (display)!
• Switch off main switch after finishing work in practice and
close water valve of supply.
• The Variotherm plus may be operated only in rooms
used for medical purposes, but not in areas subject to
explosion hazards and in oxygen rich environments.
• All additional equipment which is connected to the
analogue and digital interfaces of the device must meet
the requirements of relevant EN specifications (e.g. EN
60950 for data processing equipment and EN 60601 for
electrical medical equipment). In addition, configurations
must satisfy system specification EN 60601-1-1. When
additional equipment is connected to the signal input or
signal output section on the device, the person carrying
out the connection is deemed „a system configuration
operator“ and as such is responsible for meeting the
requirements of system specification EN 60601-1-1. For
answers to additional questions, please contact your local
specialist supplier or the ATMOS Technical Service.
• The irrigation jet must not get into contact with
contaminated material.
2.1 Important instructions for the maintenance of the hygiene status of warm water
irrigation units
For loosening cerumen in the auditory meatus and for the stimulation of the labyrinthine system, ATMOS offers the warm water
units Hygrotherm plus (37°C) and Variotherm plus (20° - 47°C).
These units heat the drinking water which comes from the household connection to the preselected temperature. The operating
instructions must be followed, or the increased number of pathogens in the rinse water can affect the health of sensitive
patients.
Requirement for taking into use
The water which is provided by the household connection must at least meet the WHO guidelines or the country-specific
guidelines for drinking water.
Connection
Fresh water must flow through the wall connection for a period of approx. 1 minute prior to the unit being connected.*
Ensure that the installation is hygienically carried out! Disinfect the threaded connections with 70 % alcohol before screwing
them into place.*
Wear disposable gloves when carrying out maintenance work on parts which come into contact with the water.*
Carry out thermal cleaning at least 3 times before patient use.*
* This work is carried out by the ATMOS service technician.
Running operation
Before beginning the surgical operation, switch on the unit and await the thermal cleaning process.
After the device had been idle for a long period of time, (weekend, during holidays, etc.) the thermal cleaning process must
be initiated and repeated.
We recommend that the device be switched off and on every 2 hours and not longer than 4 hours to start the thermal
cleaning mode.
Exchange and disinfection required for contact parts such as the jet connection after each patient (prevention of retrograde
germification).
Before every use, spray water and check the temperature.
Service
Ensure that the suspended particle filter is hygienically changed (see commissioning) (dispose of filter and carefully clean the
filter glass). Ensure that no contact is made with the contaminated parts.
Carry out thermal cleaning after each service.
We recommend that the number of germs in the water be regularly determined at the input and output of the unit. In the case
of a complaint, data measurements must be prove whether the contaminants are "water" or "skin” bacteria.
3.2 Illustrations
11
30 37
[3 0 [3 0 [4 4
Variotherm plus
F 14 AH
Fig. 2. Rear view
1 Type plate
2 Control output for controlling a nystagmograph
3 Device plug with fuse compartment
4 Equipotential bonding connection
p = +2000...+6000hPa
t= +7...+19,5°C
*
10
3.3 Connections
Fig. 5.
11
12
• Standard settings:
-- Level for irrigation mode: 37°C fixed
-- Level for cold stimulation: 30°C
-- Level for warm stimulation: 44°C
13
Purpose:
4.5.1 Thermal process for reducing bacterial
count Stimulation of the vestibular organ.
Properties:
Purpose: • Temperature: corresponds to the preselected cold or
Reduction of bacteria count in the warm water circuit to warm stimulation level
prevent or clear a contaminated system. • Position of the adjusting lever: Low (see fig. 6, page 13).
Activation: • Duration: as pre-set by the timer.
Occurs each time the unit is switched on. Activation:
• Condition for starting the correct thermal process for • Initially select the type of stimulation by actuating either
reducing bacterial count: warm-stimulation or cold-stimulation key (, , fig. 1)
-- The irrigation handle must be inserted in the holder (see section 4.1 for pre-setting of temperature)
on the side (jet pointing to reverse side of the unit).
ªª Water with pre-set temperature value flows to handle.
ªª If not, a warning signal is given (warning also sounds
when the handle is removed during the thermal • Wait for 20 seconds.
cleaning procedure). • Operate the "timer-start-key".
ªª Additionally, the error code "F0" is shown (disappears • Preparation for stimulation:
when handle is returned). -- Operate unit 15 seconds in holding position, so that
)) In the thermal process for reducing bacterial count hot the jet can be positioned correctly in the auditory
water is being ducted through the irrigation handle. canal.
Please do not take the handle from its holder or spray )) During this period, the water is directed, via the handle,
water! into the drain in order to guarantee an optimum
)) Metal parts of the irrigation handle and of the double- temperature stability; with an optical indication by the
barrelled hose will be hot! Please do not touch metal flashing figure in temperature display.
parts during the thermal cleaning procedure! Do not start stimulation by operating the release lever at
the handle until you hear the audible signal!
• During the thermal process for reducing bacterial count,
the temperature display shows a 'd' on the first segment • Carrying out the thermo-stimulation by actuating release
and the remaining time (in minutes) on the second lever at handle (, fig. 4).
segment. • At the end of the stimulation period a control signal for a
• After cooling down to 37°C the unit will automatically recording unit is issued at the nystagmograph output.
switch into irrigation mode. • After completion of the stimulation the unit remains in a
holding position for 15 seconds so that the handle may
4.5.2 Irrigation mode be removed from the auditory canal.
ªª During this time, water flows into the drain; again an
Purpose: optical indication is made by flashing temperature
Rinsing of the auditory canal. value.
• Repeated actuation of the "timer-start-key" leads to
Properties:
stopping timer operation.
• Temperature: 37°C, permanently set.
• Second actuation of the currently active key effects
• Position of the adjusting lever: High (see fig. 6, page 13). deactivation of the pertaining stage.
Activation: ªª Heating is switched off completely.
• Automatically after completion of thermal cleaning ªª Stimulation with cold water (feed temperature).
procedure and taking out the handle or by key (, fig. 1).
• When the irrigation mode is not used (handle remains 4.5.4 Energy saving mode
in holder) the Variotherm plus changes into the energy
saving mode after 5 min. Purpose:
• Reactivation of the irrigation mode by taking the handle Reduction of energy consumption.
out of the holder or activating a key on the temperature Activation:
control unit.
• Actuation of key "heating on/off” (, fig. 1).
• If the 37°C push button is repeatedly pressed, the heating
is switched off. ªª Heating is switched off.
ªª Water with feed temperature is available. • After fi ve minutes of non-use of the unit.
14
5.1 General information on cleaning and You may also use disinfectant sprays or disinfectant tissues
for cleaning and disinfection.
disinfection
)) Set main switch of the device to OFF prior to cleaning
The jet connection which comes into contact with the and disinfection!
patient must be disinfected after each use. Wipe the unit surface with a cloth moistened with a cleaning
The hose tips must be exchanged after each use. or disinfecting solution. Take care that no liquid penetrates
The surfaces of the Variotherm plus resist most of the the device. The cleaning agents and disinfectants listed in
common surface disinfectants. section 5.2 are all suitable.
However, do not use )) Spilled liquid must immediately be dry wiped.
• disinfectants which contain concentrated organic or )) Always observe the instructions for use by the
inorganic acids as they could cause corrosion damage. manufacturer of the disinfectants, including all
concentration specifications.
• Disinfectants containing chloramides, phenol derivatives
or anionic tensides, as these may cause stress cracks in )) The described action relating to cleaning and disinfection
the material used for the housing of the unit. resp. sterilisation do not substitute the relevant
instructions which must be adhered to prior to operation.
When using disinfectants containing aldehyde and amine at the same object colour changes may occur.
15
5.3 Cleaning method for handle with jet connection for water irrigation tip
The ATMOS appliances Variotherm and ATMOS Hygrotherm are used, according to the rules, with disposable hose tips
(disposables). These tips must be disposed off after each application to the patient.
When using the hose tips ATMOS recommends a reprocessing in line with the below cleaning and disinfection plan.
application
Details
After each
C D S with reprocessing.
Monthly
(Please fill in the
Weekly
Parts to be reprocessed Cleaning Disinfection Sterilization
responsible person
Daily
-> use a water-based
overhead marker)
Recommended disinfectants
Surface disinfection for
3)
Other surfaces: 4)
Manual disinfection of 5)
Automatic disinfection of instruments:
coated surfaces: instruments:
• Green & Clean SK (ATMOS) • Dismozon® plus (Bode Chemie) • Korsolex® AF (Bode Chemie) • Dismoclean® 21 clean (Bode Chemie)
• Dismozon® plus (Bode Chemie) • Kohrsolin® FF (Bode Chemie) • Korsolex® basic (Bode Chemie) • Dismoclean® 24 Vario (Bode Chemie)
• Kohrsolin® FF (Bode Chemie) • Bacillocid® rasant (Bode Chemie) • Korsolex® plus (Bode Chemie) • Dismoclean® 28 alka one (Bode Chemie)
• Perform® (Schülke & Mayr) • Mikrobac® forte (Bode Chemie) • Korsolex® extra (Bode Chemie) • Dismoclean® twin basic/twin zyme (Bode Chemie)
• Terralin® Protect (Schülke & Mayr) • Perform® (Schülke & Mayr) • neodisher® Septo MED (Dr. Weigert) • neodisher® FA (Dr. Weigert)
• Terralin® Protect (Schülke & Mayr) • neodisher® Septo 3000 (Dr. Weigert) • neodisher® FA forte (Dr. Weigert)
• Surface disinfectant FD 312 (Dürr • Sekusept® PLUS (Ecolab) • neodisher® MediClean forte (Dr. Weigert)
Dental) • Sekusept® aktiv (Ecolab) • Thermosept® alka clean forte (Schülke & Mayr)
• Quick disinfection B 30 (Orochemie) • Gigasept® Instru AF (Schülke & Mayr) • Thermosept® RKN-zym (Schülke & Mayr)
• Gigazyme® (Schülke & Mayr)
• Gigasept FF neu (Schülke & Mayr)
For concentrations, contact time, temperature, material compatibility, please see the relevant information from the manufacturer.
Important information
Wipe cleaning and disinfection: All surfaces have to be wiped with a clean (disposable) wipe which is damped with disinfectant solution; the
entire surface has to be wiped thoroughly and may not be dried afterwards.
1)
Please observe the manufacturer‘s operating instructions.
2)
Preferred: machine cleaning and disinfection in the washer disinfector
6)
Material dimensionally stable at 134°C
16
• Decalcification and changing of filter (see sections 6.1, 6.2) should be carried out immediately by the user when required.
Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge and
are familiar with the product. To carry out these measures the person must have the necessary test devices and original spare
parts.
ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This ensures that repairs and
testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
• Carry out an inspection according to the manufacturer‘s specifications every 12 months.
6.1 Decalcification
If the tap water in the area, where the unit is used, is known for mineral precipitation, preventive decalcification should be carried
out:
• Close water tap.
• If applicable, dismantle filter when mounted vertically at water tap.
• Unscrew screwed glass of filter.
• Fill screwed glass with approx. 100 ml of decalcifier on acetic / citric, or lactic acid base (for inst. Citrosteril, by Fresenius).
• Screw glass on again.
• Install filter vertically again, if applicable.
• Open water tap again.
• Switch on the device.
ªª Decalcification takes place automatically in the thermal process for reducing bacterial count. After the unit has been in this
mode for 5 minutes the water passages have been decalcified and cleaned, and also sufficiently flushed. The Variotherm
plus is again fully ready for operation. When water passages are heavily calcified, a modified procedure may become
necessary, with a prolonged reaction period for the decalcifier.
Recommendation: Calcification safety device 502.0995.0.
17
"F0" Handle is not in its support • Insert handle in its support to allow thermal process for reducing
bacterial count to start (jet connection must point to the back of
the unit.
• Replace handle (Art.No. 502.0963.0)
"F1" No water (water pressure < 0,5 bar) • Check, whether the water supply delivers a pressure of at least
2 bar (did you open the water-tap?).
• Filter clogged?
"F2" -5 V is missing (supply voltage on the control • Inform the service staff.
board)
"F3" Break of the safety NTC • Inform the service staff.
"F4" Safety switching does not react (49°C • Have the safety switching (temperature controller) checked by the
comparator) service staff.
"F5" Break of the regulating NTC • Inform the service staff.
"F6" Different switch settings in the handle holder • Handle must be correctly inserted in its support.
• Otherwise inform the service staff.
"F7" Temperature too high (>48°C), display only in • Check, whether temperature setting is too high. If necessary,
the stimulation and irrigation mode but not in adjust desired temperature to a value of <48°C by means of the
the thermal cleaning procedure coding switches.
• Inform the service staff.
"F8" Short-circuit of the regulating NTC • Have temperature feeler of the regulating NTC checked by the
service staff.
"F9" No heating performance (excess temperature • Inform the service staff (internal excess temperature switch might
switch), def. Semi-conductor relay or regulator) have released).
)) If errors cannot be corrected with the assistance of the trouble-shooting list, please inform the service staff or send in the
device for repair. Do not start any attempts to repair the unit yourself!
18
8.1 Accessories
Description......................................................................................................................................................REF
Calcification safety device.................................................................................................................................502.0995.0
Jet connection extra long, straight (110 mm)....................................................................................................508.0429.0
Rinsing connection (Teflon) for stimulating the vestibular organ....................................................................... 501.0355.0
Extension of supply hose G3/4a-G3/4i, L =1,5 m..............................................................................................501.0315.6
Waste water hose, special length, G1/4a-G1/4a (please indicate desired length)............................................ 502.0881.1
Drainage fittings as to DIN 1988.......................................................................................................................502.0880.0
Connecting cable for a nystagmograph............................................................................................................502.0850.0
Connecting cable for the equipotential compensation, L = 5m......................................................................... 008.0596.0
19
20
21
• Packaging material, cardboard and/or PE foam, can be fully recycled or returned to your supplier.
• The Variotherm plus does not contain any hazardous materials.
• The housing is recyclable.
• The component parts of the Variotherm plus must be disposed of correctly and the materials are to be separated carefully.
• The electronics circuit boards must be fed into the appropriate recycling process.
• Used hose tips, which no longer can be disinfected, must be discarded into domestic waste immediately.
22
i • Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following
EMC notes.
• Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other converters and cables than stated may lead to an increased emission or a reduced
interference immunity of the equipment or system.
The ATMOS® Variotherm plus is designed for operation in the environment specified below. The customer or user of the ATMOS
Variotherm plus should ensure that it is used in such an environment.
i The device may not be used directly next to other devices or piled up with other devices. If operation next to or piled with other
devices is necessary, please watch the device to check its intended operation in this arrangement.
Electromagnetic Environment -
Immunity Test IEC 60601- Test Level Compliance Level
Guidance
Electrostatic discharge ± 6 kV Contact ± 6 kV Contact Floors should be made of wood or
(ESD) according to IEC concrete or tiled with ceramic tiles. If
61000-4-2 floors are synthetic, the relative humidity
± 8 kV Air ± 8 kV Air
should be at least 30 %.
EFT IEC 61000-4-4 ± 2 kV Mains ± 2 kV Mains Mains power quality should be that
of a typical commercial or hospital
environment.
± 1 kV I/Os Inapplicable
Surges IEC 61000-4-5 1 kV 2 kV Mains power quality should be that
Differential Differential of a typical commercial or hospital
environment.
1 kV 1 kV
Common Common
Magnetic field at power 3 A/m Inapplicable Power frequency magnetic fields should
frequency 50/60 Hz acc. be that of a typical commercial or
to IEC 61000-4-8 hospital environment.
23
Immunity Test IEC 60601- Test Level Compliance Level Electromagnetic Environment -
Guidance
Voltage Dips / Dropout < 5 % UT < 5 % UT Mains power quality should be that of a
IEC 61000-4-11 (> 95 % Dip of the UT) (> 95 % Dip of the UT) typical commercial or hospital environment.
If the user of the ATMOS® Variotherm plus
for 0.5 Cycle for 0.5 Cycle
requires continued operation upon the
occurrence of disruptions in the energy
40 % UT 40 % UT supply, the ATMOS® Variotherm plus should
(60% Dip of the UT) (60% Dip of the UT) make use of an uninterruptible power
supply or a battery.
For 5 cycles For 5 cycles
70% UT 70% UT
(30 % Dip of the UT) (30 % Dip of the UT)
For 25 cycles For 25 cycles
< 5 % UT < 5 % UT
(>95 % Dip of the UT) (>95 % Dip of the UT)
for 5 s for 5 s
NOTE UT is the mains alternating current prior to application of the test levels.
24
NOTE 1 By 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by absorption and
reflection of buildings, objects and people.
a
The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radio equipment,
amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly. To determine the electromagnetic
environment in regard to stationary transmitters, a study of the location is to be considered. If the field strength measured at
the site where the ATMOS® Variotherm plus is used exceeds the compliance level above, the ATMOS® Variotherm plus must
be observed to demonstrate proper function. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the ATMOS® Variotherm plus.
b
Within the frequency range of 150 kHz to 80 MHz the field strength should be below 3 V/m.
NOTE 1
With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by absorption and
reflection of buildings, objects and people.
25
26
27