GR611M
GR611M
Technical Data
FEATURES :
The Gamma Ray Spectrometer (Microcontroller based) Type GR 611M consists of a MINIM based modular counting unit and a
Scintillation Detector Type SD 151 / 152F /152W / 153F / 153W. The Spectrometer can also be used for Gamma Counting
applications. This MINIM based system has added advantage of savings in cost and also conserves bench space because of its
optimal design.
This system is configured around MINI BIN and Power Supply Type: MB 403 with the following Modules:
High Voltage Unit Type: HV 502
Linear Amplifier Type: LA 520
Single Channel Analyser Type: SC 530
& Counter Timer Type: CT 541A
(Microcontroller based), having unique built-in user programmable features for data acquisition & data outputting.
II. HIGH VOLTAGE UNIT (HV 502) : f. Pulse Height Discrimination :100mV - 10V by a preset
a. Output voltage variable continuously from 0V to 2000 volts provided on front panel
b. Output current (max) 1mA g. Counts Indication : 16x2 dotmatrix LCD display
c. Load & Line regulations : Better than 0.05% of full scale h. Modes of Data Acquisition : a. Counts for a preset time
d. Indefinite over load & short circuit protections and self b. CPS
recovery c. CPM
e. Output ripple less than 10mV (rms).
i. Storage facility : upto 1000 readings
f. Dimensions : Two bit module
j. TIMER
Preset Time Setting : Programmable through tuctile
III. LINEAR AMPLIFIER (LA 520) : switch control buttons
a. Input Polarity : Positive or Negative k. Preset Time / Elapsed Time :On 16x2 dot matrix LCD
b. Total gain (Typical) : 800 (approx) display
l. Preset Time Range : 1 to 9999 Sec
c. Output (Unipolar) : 0 to 8V
m. Built-in printer port facilitating direct data printing
d. Max. output (Unipolar) : 12V
n. Additional Options :
e. Dimensions : Two bit module
i. Data communication software for down loading of data
can be given at extra cost.
IV. SINGLE CHANNEL ANALYSER (SC 530) :
a. Input : Unipolar or bipolar with a +ve VI. APPLICATIONS :
leading edge 0 to 10V
The system finds wide range of applications in nuclear
b. Pulse Pair resolution (approx): 0.6 micro sec. research and academic fields which include:
c. Output pulse polarity : +ve Analysis of gamma radiation Identification of unknown
Pulse amplitude : +5V isotopes and their relative abundance.
Pulse width : 0.5 micro sec Measures the strength of radio activity of sample.
d. LLD output pulse amplitude : +5V Useful in radiotracer techniques.
Output pulse width : 0.5 micro sec Can be used for protein bound iodine studies in medicine
Base line variable by : 10 turn helipot dial using well type scintillation head etc.
Swipe sample counting in Health Physics Labs.
Base line variation : 0 to 10V by helical
potentiometer
e. Window width : Continuously variable by helical
potentiometer
(0 - 1) V in window mode
(0 - 10) V in normal mode
f. Dimensions : Single bit module
Page 2 of 4
NUCLEONIX
ACCESSORIES FOR
GAMMA RAY SPECTROMETER SYSTEM
SCINTILLATION DETECTORS
Nucleonix Systems offers wide range of NaI Scintillation Detectors
of different sizes both with flat & well type crystals, to meet the
requirements of wide range of users for Gamma ray spectrometry
measurements.
Important Note: All NaI Scintillation Detectors manufactured & supplied by Nucleonix systems, use
integral assemblies of Saint Gobain who are world leaders.
Page 3 of 4
NUCLEONIX
Gamma Reference Standard Set Type: GS290 consists of a set of FIVE Gamma
sources evaporated and sealed on 25mm dia x 5mm plastic discs ranging from
100KeV upto 1.33MeV energies with activity in the range of 2 to 5 micro curie. A
reference chart for this is given below. The accuracy of these sources is in the range
of +/-10%.All these disc sources are enclosed in a neatly polished wooden box.
This consists of 40mm thick lead shielding cylindrical rings assembled according to the detector. There is a provision in the bottom
ring through which system connections are given to the detector, which is placed inside the lead shield.
For 1”X1” detector (9 lead assembling parts) For 2”X2 detector (10 lead assembly parts) For 3”X3” detector(11 lead assembly parts)
The top ring has a holding knob through which sample can be loaded on to the scintillation detector, and this closes the lead castle
from top side.
Page 4 of 4