Conductivity controller
Committed to process automation solutions
Datasheet
SUP-TDS210-B
1
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
The model SUP-TDS210-B is used for the conductive measurement/control of electrolytic conductivity,
resistivity or the TDS value.Conductivity is a function of ion concentration, ionic charge, and ion mobility.
Ions in water conduct current when an electrical potential is applied across electrodes immersed in the
solution. A controller system consists of a microprocessor-based controller and a conductivity probe.
4 Electrode cells (K=0.01,0.1,1.0 and 10.0)can be connected to the device. Temperature serves as the
second input variable, measured by a NTC10K/ PT1000 probe. Depending on the measured variable, it is
therefore possible to implement specific, automatic temperature compensation.
All adjustments to the current outputs, alarm relays, and calibration of the conductivity and temperature
inputs can be made using the controller's membrane keypad.
Features
Applications
Benefits
■ Affordable
■ Ease of Operation
■ Low Maintenance
■ Ensures Product Quality
2
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
Parameter
EC/TDS/Resistivity: ±1%FS
Accuracy NTC10K: (-10~5℃) ±2℃;(5~60℃) ±0.2℃;(60~130℃) ±2℃
PT1000: (-10~5℃) ±2℃;(5~130℃) ±0.2℃
3
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
Electrode
Suitable for pure water Suitable for drinking water Suitable for rive raw Suitable for sewage waste
ultrapure water testing testing water testing testing
The device offers a far wider dynamic range on the input side, the range must be matched to the operating
range of the cell
Example
A measurement is to be carried out in the 10 µS/cm to 10 µS/cm range. A conductivity cell with
the cell constant K = 0.1 1 is chosen.
Note
4
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
Display
1. Temperature:Compensation temperature
2. Analog output:Analog output
3. Measured value:Real-time measurements value
4. High alarm:High alarm
5. Low alarm:Low alarm
5
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
Wiring
6
Add : 5th floor, Building 4, Singapore Hangzhou
Science Technology Park, Hangzhou Economic
Development Area, 310018, Hangzhou, China
Tel: 86-15158063876 E-mail: [email protected]
www.supmea.com
Ordering Code
Conductivity controller
R1 0-2000 μS/cm
Range
R2 0-20000 μS/cm
K1 K=0.01 0.02 ~ 20.00μS/cm
K2 K=0.1 0.20 ~ 200.0μS/cm
Cell constant
K3 K=1.0 2.00 ~ 2000μS/cm
K4 K=10.0 0.02 ~ 20.00mS/cm
L1 5M
L2 10M
Cable length L3 15M
L4 20M
L5 Other Length
Signal output S1 4-20mA RS485
Relay A1 Two (high and low)
V1 220VAC
Power supply
V2 110VAC