PLD8000 User Manual-V2.0
PLD8000 User Manual-V2.0
JS-8000-13 V ER2.0
User’s Manual
JS-8000-13 V ER2.0
Electrical equipment security type: This product is a kind of permanently mounted common
equipment with Class I Type B electric shock protection.
The working system is continuous operation with intermittent loading.
The manufacturer will be liable for security, reliability and performance of the equipment
under the following circumstances:
Assembling, adjustment, operation and maintenance are carry out by authorized
personnel;
Electric utilities in the room are complying with related provisions;
The equipment is operated in accordance with user’s manual.
JS-8000-13 V ER2.0
E-mail: [email protected]
Dear customer:
Firstly, we would express our gratitude to you and your company to select our products!
Before operating the equipment, please make sure to carefully read the Product Manual,
especially the following safety precautions, for you to operate this equipment correctly.
● Please carefully read the following precautions for correct operation of the equipments.
● Before using the equipment, please make sure to get fully known about the performance
● Pay attention to the following matters for setup and storage place of the equipment:
ranges;
Pay attention to power supply frequency and voltage as well as allowable current;
60601);
Confirm power supply connection with the equipment and check if the equipment
operates normally;
JS-8000-13 V ER2.0
In case of equipment failure, instead of disassembling, please paste a note written with
“Break down!” on the equipment, and the Company will assign authorized personnel
Before reusing the equipment after a long time out of service, please make sure to
The following advisory symbols will be used throughout this manual. Their application and
The following Security identity symbols will be used the equipment. Their meaning are
described below.
Ionizing radiation
PLD8000User’s Manual
JS-8000-13 V ER2.0
Type-B equipment
Dangerous voltage
GND
disposal equipment.
ii
PLD8000User’s Manual
JS-8000-13 V ER2.0
TABLE OF CONTENTS
CHAPTER1 OVERVIEW....................................................................................................... 1
JS-8000-13 V ER2.0
2.3 Collimator................................................................................................................... 21
3.1.5 Setting AEC backup ms or mAs (applies to non CM-thickness mode only)...33
4.1 Overview.....................................................................................................................42
4.2.4 Flat-panel detector, ionization chamber, the installation of filter wire grid.... 46
JS-8000-13 V ER2.0
6.2 Maintenance that can be carried out and cleaned by the Authorised Technician..... 108
v
PLD8000User’s Manual
JS-8000-13 V ER2.0
CHAPTER1 OVERVIEW
This unit can be used for X-ray routine inspections such as X-ray stomach and intestines
spot film photography, remote control compartment fluoroscopy and other traumatic
photography in medical teaching, scientific research and medical treatment units: It can
conduct digestive tract radiography, biliary tract radiography, urinary tract radiography and
lower limb venae radiography, etc.; it can carry out fluoroscopy inspection and spot film
photography on esophagus, chest, stomach and intestines, abdomen and four limbs; it also
can implement fracture recovery and getting foreign matters under fluoroscopy, etc.
1
PLD8000User’s Manual
JS-8000-13 V ER2.0
Main features:
1. Radiographic table
● Visual display man-machine interface and touch screen control makes man-machine
dialogue more direct-viewing, convenient and easy to understand.
● Programmable control device technology with high control precision, good stability and
excellent anti-interference performance.
● Frequency converter control technology, the table operates with soft startup and soft
shutdown, which is smooth without noise and the positioning is more accurate.
● Reliable angular displacement sensor technology with excellent precision, little
dynamic noise and long mechanical service life.
● High-resolution rotary encoder with automatic slicing accuracy, which is safe and
stable.
● Rotational range: 90° ~ 0° ~ -25°, it can carry out conventional examinations about
alimentary organs, and it also can be used for a variety of special examinations.
● Humanized design of diagnostic table side switch, which is more convenient for
positioning when operating near the console.
2
PLD8000User’s Manual
JS-8000-13 V ER2.0
● Toshiba high-quality image enhancement device, which further improves image clarity
and contrast ratio, and makes it much easier and clearer to identify the nidus.
● The moving range of spot-film device is wide, so we adopt the operating mode with
machine moving but people unmoving, which thus simplifies the implementation of a
series of inspections from the throat and esophagus to the lower abdomen.
● Spot film device load automatically, and the operation process is safe and reliable.
2. Generator
● Generator with X-ray radiography and fluoroscopy functions and its maximum output
power is 80kW.
● Output voltage is 150kV.
● Smaller and lighter block design
● During exposure, you can adjust kV and mA buttons to achieve constant ray output.
● Large screen LCD flat panel display, which is used to display APR conditions and
practical conventional items.
● User-friendly operating system configuration.
● The operator can modify APR technology items.
● APR/ray bulb tube data download function.
● A variety of diagnostic procedures with prompt information for the operator.
● Equipped with serial RS232 communication interface.
● You can carry out operation programming for APR alone or choose programs for APR
and manual technology items.
● Programmable settings, calibration and implementation of APR routine operation (by
connecting it with an external computer).
● It is allowable to select four kV and mA curves for ABS (automatic brightness
stabilization) circuit.
safety
JS-8000-13 V ER2.0
safety
electrical systems
YY/T 0291—2007 Environmental requirements and test methods for medical x-ray
equipment
Medical equipment specification 、Labels and packaging labeling regulations (The state
The manufacturer will be liable for quality of the equipment in aspects of safety, reliability
4
PLD8000User’s Manual
JS-8000-13 V ER2.0
and performance under the following conditions:
1. Installation, adjustment, modification and maintenance are conducted by personnel of our
company or personnel authorized by our company;
2. All electric utilities of the equipment comply with requirements of the Manual;
3. Operators of the equipment are well trained and skilled technical personnel and operate
strictly in accordance with the Manual.
1. Under the precondition that the user complies with above provisions, within one year
after delivery, for any quality problem or failure in normal using of the equipment, the
company will repair or change parts for the user free of charge or even accept return.
equipment exceeding 1 year after delivery, the company may repair or replace parts for
3. The company provides ordinary maintenance service for equipment within 8 years of
This Manual is for the purpose of guiding certified radiographer on safe operation. As
the Manual and cannot be used for other purpose except medical radiography. This
equipment can only be used by trained personnel with needed radiation knowledge. Special
1. The equipment with any electric or mechanical fault will not be used.
2. It is not allowed to remove protective circuit or protective device to make the
equipment work under power.
3. It does not allow using the equipment under moist, inflammable and explosive
5
PLD8000User’s Manual
JS-8000-13 V ER2.0
environment.
4. Consent of our Company or maintenance personnel authorized by the Company shall
be obtained from any modification to the equipment.
5. It shall be ensured that replacing components of the equipment will not endanger any
patient or operator; where necessary, maintenance personnel of our company may be
contacted.
Improper operation may result in damage to the equipment; hence regular inspection and
maintenance are necessary.
The company will be liable for any personal and property losses caused by the equipment,
excluding that due to non-compliance to the Manual or modifying the system without
authorization.
1. Ensure network voltage and current meet requirements of the equipment on power.
2. The equipment shall be put in well-ventilated room without flammable and combustible
gas leakage.
3. Cleaning, disinfection and sterilization shall be conducted when the complete machine
is power off.
4. If the floor where the high voltage generator locates is not made of wood, the generator
shall be put on wood base.
5. Shut off the machine and switch off the power after using.
Every time before exposure, it shall be ensured to adopt measures to prevent X-ray and keep
2. Operate in compartment to the greatest extent to be far away from X-ray source;
3. Low dose exposure to the greatest extent under permitted projection condition.
6
PLD8000User’s Manual
JS-8000-13 V ER2.0
7
PLD8000User’s Manual
JS-8000-13 V ER2.0
When the equipment is retired, the X-ray tube assembly and transformer oil in high voltage
generator shall be reclaimed by specialized industrial waste and waster oil collection station;
while other steel and cooper materials shall be reclaimed by corresponding department.In
printed circuits.
Service life of the equipment under correct using condition is 8 years; wherein, normal
service life of key component X-ray tube is 10,000 times; damaged X-ray tube may be sent
to our company for replacement. If the equipment is damaged after correct using for 8 years
Symbol description:
voltage radiation
Small Focal Type-B Protective
8
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 1-6
Figure 1-5
9
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 1-6
Figure 1-7
10
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 1-8
Figure 1-9
11
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 1-10
Figure 1-11
12
PLD8000User’s Manual
JS-8000-13 V ER2.0
2. Control room area: 5000×2500mm; the control room shall be opening on to the
machine
room, and the observation window shall be equipped with lead glass.
5. The diagnostic table shall be directly fixed on ground base which shall be horizontal
6. Cable duct shall be set with cover plate parallel to the ground.
7. The machine room shall be equipped with special ground wire with ground resistance
8. Protection capability of the machine room to X-ray shall meet requirements of related
national regulations.
13
PLD8000User’s Manual
JS-8000-13 V ER2.0
1. Power conditions:
2. The power distribution box shall be equipped with three-phase main switch with rated
14
PLD8000User’s Manual
JS-8000-13 V ER2.0
voltage of 500V and rated current of 130A as the switch to break power supply network;
the power distribution box must be firmly fixed on the wall with no less than 1.5m of
height from the indoor floor. Besides, power supply with capacity no less than 85KVA
and independent ground wire (no more than 4 ohm) shall be provided.
OPERATING
15
PLD8000User’s Manual
JS-8000-13 V ER2.0
2. kV range: 40 ~ 150kV
Attention:
Pay attention to that load exceeding the stipulated condition is not allowed,
because overload operation may result in damage to tube sleeve and X-ray tube.
● The assembly will generate X-ray while working under high voltage; therefore
● Using the assembly in the environment with inflammable or poisonous gas is not
allowed.
JS-8000-13 V ER2.0
(50~106)kPa.
● Inside the assembly, there’s glass X-ray tube mounted, hence it shall be handled
● As per the national regulations, installation, adjustment and using of the assembly
● 5. Stator parameters
Item Starting Running
Power frequency (Hz) 50 50
Voltage (V) 240 58
Min acceleration time (S) 0.6 —
Phase-splitting electric 24 24
capacity(μF)
● 6. Insulation resistance of stator between outgoing line and tube case: 2MΩ (Min)
17
PLD8000User’s Manual
JS-8000-13 V ER2.0
18
PLD8000User’s Manual
JS-8000-13 V ER2.0
19
PLD8000User’s Manual
JS-8000-13 V ER2.0
NC: No connection
ET: Grounding terminal
M: Main terminal of stator
A: Auxiliary terminal of stator
20
PLD8000User’s Manual
JS-8000-13 V ER2.0
2.3 Collimator
1. Radiation field:
● Minimum radiation field: when SID is 100 cm, the limit of the Collimator should be not
more than 5 cm x 5 cm
21
PLD8000User’s Manual
JS-8000-13 V ER2.0
3、Tablet parameters:
22
PLD8000User’s Manual
JS-8000-13 V ER2.0
A / D conversion: 14 bit
Input: DC 24V
Weight: 17 kg
5、Environmental conditions
Temperature: 10℃ ~ 35 ℃
Relative humidity: 30 % ~ 85 %
6、Storage conditions
Temperature: -15 ℃ ~ 55 ℃
focus mA kV S
Small focus 10、12.5、16、20、25、32 150 6.3
23
PLD8000User’s Manual
JS-8000-13 V ER2.0
150 5.0
40
145 6.3
150 3.2
50
116 6.3
150 2.5
63
92 6.3
150 1.0
80
72 6.3
150 0.32
100
58 6.3
150 0.125
125
51 5.0
138 0.01
160
48 3.2
110 0.0125
200
42 2.5
88 0.0125
250
50 0.63
10、12.5、16、20、25、32、40、
150 6.3
50、63
150 5.0
80
137 6.3
150 3.2
100
110 6.3
Large focus
150 2.0
125
98 5
150 1.0
160
96 3.2
150 0.5
200
86 2.5
24
PLD8000User’s Manual
JS-8000-13 V ER2.0
150 0.25
250
69 2.5
150 0.125
320
64 1.6
125 0.1
400
56 1.25
100 0.1
500
45 1.25
79 0.1
630
60 0.25
25
PLD8000User’s Manual
JS-8000-13 V ER2.0
LEGEND:
1. Power ON and OFF buttons
2. PREP and X-RAY EXPOSE buttons, and PREP and X-RAY EXPOSURE indicators
3. Radiography controls and display
4. Image receptor buttons
5. Anatomical programming controls and display, BACK / FORWARD buttons, and
6. MENU (reset) button
26
PLD8000User’s Manual
JS-8000-13 V ER2.0
exposure LED, should light briefly. This is an
LED self – test and does NOT indicate the
presence of X-RAYS.The speaker should
briefly beep near the end of theconsole
self-test.
Press the MENU (reset) button to continue if
anyerror messages are presented.
JS-8000-13 V ER2.0
will spin the rotor. Fully depress and hold the hand switch in the EXPOSE position to
make an X-ray exposure.
LEGEND:
1. Technique select button and indicators.
2. Field select buttons and indicators.
3. Film / Screen select button and indicators.
4. Focus select button and indicators.
5. kV up / down buttons and display (non-CM thickness mode).
CM thickness up / down buttons and display (CM thickness mode).
6. mA up / down buttons and display (non-CM thickness mode).
kV up / down buttons and display (CM thickness mode).
7. Density / ms / mAs up / down buttons and display.
8. ms / mAs readout.
9. AEC - ON indicator / post-mAs display.
28
PLD8000User’s Manual
JS-8000-13 V ER2.0
JS-8000-13 V ER2.0
The rate of scrolling for the kV, mA, and ms display increases if the
buttons are pressed continuously. To return to the slowest scroll rate,
release the button(s) and start again.
kV up / down buttons and display (CM up / down buttons and display in CM-thickness mode)
30
PLD8000User’s Manual
JS-8000-13 V ER2.0
To increase kV / CM thickness, press +.
To decrease kV / CM thickness, press -.
Displays demanded kVp of the exposure.
Displays selected CM thickness when
CM-thickness function is active.
mA up / down buttons and display (kV up / down buttons and display in CM-thickness
mode)
To increase mA / kV, press +.
To decrease mA / kV, press -.
Displays demanded mA of the exposure.
This is active for AEC in non CM-thickness
mode or mA/ms techniques only.
Displays kV when the CM-thickness function is
active.
31
PLD8000User’s Manual
JS-8000-13 V ER2.0
setting.
The + / - buttons only adjust the parameter
displayed on top. To swap the density and the
backup ms or mAs displays, press both the -
and + density buttons. Refer toSetting AEC
Backup ms or mAs for further details.
The density range is typically -5 to +5. (Installer
programmable to have a range of up to -8 to +8).
This varies the optical density by changing the
dose.
The % dose change per density step change is
installer programmable.
Displays mAs when mAs is selected.
Displays time (ms) when mA/ms is selected.
ms / mAs readout
The significance of the ms/mAs readout varies, depending on mode selected:
• In AEC mode, with fixed backup time selected, the AEC backup time is displayed.
• In AEC mode, with either variable ms or variable mAs backup time, the backup mAs
or ms value is displayed.
• In mAs mode, the calculated exposure ms will be displayed (based on the mAs
32
PLD8000User’s Manual
JS-8000-13 V ER2.0
selected).
• In mA/ms mode, the calculated mAs of the exposure will be displayed (based on the
mA and ms selected).
AEC - ON indicator / post-mAs display
Displays AEC when AEC is selected.
Displays Post mAs after an AEC exposure.
33
PLD8000User’s Manual
JS-8000-13 V ER2.0
LEGEND:
1. Table, with Bucky, image receptor select.
2. Table, non Bucky, image receptor select.
3. Wall Bucky image receptor select.
4. Auxiliary 1 image receptor select.
5. Auxiliary 2 image receptor select.
JS-8000-13 V ER2.0
35
PLD8000User’s Manual
JS-8000-13 V ER2.0
JS-8000-13 V ER2.0
will be replaced with a rectangular outline around the anatomical view. Simply
overriding the programmed parameters will not change the programmed
parameters. To make permanent changes to the APR, see CHAPTER 4:
PROGRAMMING THE CONSOLE.
For CM thickness mode:
Anatomical section and anatomical view selection buttons.
37
PLD8000User’s Manual
JS-8000-13 V ER2.0
38
PLD8000User’s Manual
JS-8000-13 V ER2.0
• Pressing this button will clear error messages.
39
PLD8000User’s Manual
JS-8000-13 V ER2.0
40
PLD8000User’s Manual
JS-8000-13 V ER2.0
41
PLD8000User’s Manual
JS-8000-13 V ER2.0
4.1 Overview
The equipment must be installed and adjusted by professional personnel of our Company or
Before opening the packing case, it shall be stored in the rain-proof and dust-free
environment; and boxboard shall be opened reasonably while opening the case. While
handling the parts, attention shall be paid to avoid scratching parts surface; check according
to the packing list after opening the case and installation shall only be conducted after
confirming the machine is defect and omission; any problem shall be informed to the
42
PLD8000User’s Manual
JS-8000-13 V ER2.0
According to the figure 4-1 measurement and installation, put the bed well ,radiographic
table directly fixed on the ground,the ground level and can carry 1500 kg load,to build the
basis of using the cement is not less than 400 "label,the ground of the casting thickness not
1—Column components 2—Hex cylinder head screw M12×30 3—pin 4—Hex cylinder head screw M12
×30 5—Scale identification plate 6—Swing arm components 7—Calibration plate bracket
43
PLD8000User’s Manual
JS-8000-13 V ER2.0
1, Post installation. Let the column component (1) pin and swing arm components (6) of
pin hole on a straight line, after inserting pin positioning, Then use M10 x 25 hex
cylinder head screw(5, number of 2), flat washer and spring washer screw in tight, the
2, Scale identification plate installation. Scale will identify plate (5) and scale identification
plate bracket (7, number 2)through M5×10 hex head screw (4, number of 8) connection,
1—Column components 2—Hex cylinder head screw M12×30 3—Hex cylinder head screw M10×20 4—
Hex cylinder head screw M6×20 5—Tube connection seat 6—Hex cylinder head screw M4×12 7— Tube
outer garment 8—Collimator plate 9—Tube cover mounting plate B 10—Tube lid 11—Collimator
1,Preparation: first loosen the two M10 x 20 hex cylinder head screw (3), remove the fixed
bulb connected seat within the M12 x 30 hexagonal cylindrical screw (2), and Remove the
2, The installation of the tube connection seat: Connecting lamp holder (5) and bulb
44
PLD8000User’s Manual
JS-8000-13 V ER2.0
components with four M6 x 20 cylinder head screw (4) fixed connection, screw down the
screws, Note the tube connection block chamfering direction with the ball tube installed in
3, The installation of plate and the cover cap: according to figure the order of the limit of
the collimator plate (8), tube cover mounting plate (9) and B ball tube (10), the limit of the
beam splitter connecting plate (11) and a light bulb components have four M6 * 30 philips
4, Light bulbs and column component connection: The first three steps have assembly bulb
insert column component mounting holes , with M12 x 30 cylinder head screw (2), two
M10 x 20 hex cylinder head screw (3) fixed connection, screw not preloaded, convenient
5, The installation of the Collimator: Will be limited within the Collimator on four
hexagonal cylindrical screw M5 x 40 screw out but can't fall off, turn the bulb components
to make the ball tube output window level up, lock bulb turn fastening screws, careful drop
collimator bayonet center line connected to the collimator plate and inserted into the slot,
M5 Hex cylinder head screw, after confirm the connection is firm, loosen the bulb turn
fastening screw (3), turn the bulb components make the bulb output window down.
6, Bolt pre-tightening : after adjust the bulb and tablet of collimation, tighten the pieces of
7, the installation of tube outer garment: all the installation is completed, the ball tube cover
(7) T slot insert ball tube mounting plate B notches, with two cross trough pan head screws
45
PLD8000User’s Manual
JS-8000-13 V ER2.0
wire grid
1—Disc 2—The downside cover 3—Philips countersunk head screwM3×10 4—Philips countersunk
head screwM3×6 5—Philips countersunk head screwM3×6 6—Tablet power installation plate 7
— Tablet power 8—Hex cylinder head screw 9—Probe mounting plate 10—Flat-panel detector
11—Ionization chamber 12—Ionization chamber pressure plate 13—Box front cover components
14—Philips countersunk head screw M4 x 10 15—Line filter tray components 16—Hex cylinder head
screw M4 x 10
Figure 4-4: Flat-panel detector, ionization chamber, filter wire grid installation
diagram
1, Plate, the combination of the ionization chamber: as shown in the above, First the
flat-panel detector detector installed in the (9), The mounting holes with ten M4 x 10 philips
countersunk head screw (14) add flat washer and spring washer is fixed, Put ionization
chamber detector, determine the correct position and direction, with ionization chamber
pressure plate (12) and 8 M4 x 16 hex cylinder head screw (8) fixed flat washer and spring
2, Box front cover components, the gear cover, flat power installation plate disassembly:
46
PLD8000User’s Manual
JS-8000-13 V ER2.0
Loosen the four M3x10 philips countersunk head screw (3), a box cover components will be
removed, pay attention to protect box cover component of wire. Then four M3x6 philips
countersunk head screw (5), decomposition gear cover (2), and finally loosen the four M3x6
philips countersunk head screw (4), took the tablet power installation plate (6).
3, The installation of tablet power supply:The tablet PC power supply box on the tablet PC
power installation plate (6), the installation hole, with 4 M3 x 6 cross trough pan head
screws (5) combined with fixed flat washer and spring washer fixed connection, tighten the
4, The installation of flat component: The first step has been completed flat component with
six M4 x 10 cylinder head screw (16) and flat washer and spring washer is fixed on the box
(1), confirmed after its aligned with the center line of the ball tube screw down the screws.
5, The installation of tablet power components: Will complete the assembly step 3 tablets
power components with 4 M3 x 6 philips countersunk head screw (2) fixed on the box (1),
6, Cover the downside of Ann: Will be descending cover (2) using 4 cubic meters x 6 cross
trough pan head screws (5) combined with flat washer is fixed on the box (1), screw down
the screws.
7, The installation of box front cover components: The box cover components (13) with 4
M3 x 10 philips countersunk head screw (3) fixed to the boxes (1), screw down the screws.
8, The installation of line filter tray components: Filter tray component (15) from the box
cover component opened-toe inserted into the box about the line filter chute, the assembly is
complete.
47
PLD8000User’s Manual
JS-8000-13 V ER2.0
1— Adornment casing 2 2—Adornment casing 3 3—Swing arm components 4—Trough pan head
screw M4 x 10 5—Piece of box cover
1, The shell is installed: With six M4 x 10 cross trough pan head screw washer peace will
decoration housing 2 (1) fixed ,with four M4 x 10 cross trough pan head screw washer for
peace will decorate the shell 3 (2) fixed to the column component, screw down the screws。
2, Box cover installation: After the machine debugging installation box cover,remove the
bed surface to the outside , With four M4 x 10 cross trough pan head screw washer for
peace will box cover (5) and pills box fixed components..
A. more than soft wire, materials for BVR10mm2, light blue color and length is 8 m, are
connected by A generator power terminal N to zero wall gate line, as shown in figure 4-6
"N".
48
PLD8000User’s Manual
JS-8000-13 V ER2.0
B. more than soft wire, materials for BVR13mm2, color is black, length is 8 m, the
generator power terminal "L1" "L2" "l" to connect to wall brake line, as shown in figure 4-6
Ground: more than soft wire, materials for BVR4mm2, yellow-green color and length is 8
2, power cable photography bed PLD80.00.01:4 core cable, materials for RVV1.5 was x 4,
length is 8 m.Photography bed after the power cable from the generator and the terminal
row of TB - 1504 - l terminal connected to photography bed base line of the X1 "L1" "L2"
"l" and "N".As shown in figure 4-7 "1 - ground" 2 to the power cord ""
Ground: more than soft wire, materials for BVR4mm2, yellow-green color and length is 12
m.Protective earthing terminal connected to ground by photography bed base wall brake
Figure 4-7 :Photography bed power cable and protection of ground wire connection
49
PLD8000User’s Manual
JS-8000-13 V ER2.0
diagram:
3, probe power cable PLD80.00.03:2 core cable, materials for RVV0.5 was x 2, length is 8
m.Probe power cable terminals from the photography bed base and row X2-2 X2-2
connection and power supply box of EPS plate - F001A.As shown in figure 4-15 "1".
4, other protective grounding line: console protection grounding line, the length is 5 m, the
protective earthing terminal connected to the console base wall brake independent
photography base protective earthing terminal connected to the site of the protective
composed of high voltage generator connected to the protective earthing terminal of the ball
the user must provide independent meet the national standards for
be no greater than 4 Ω.
Wall brake wiring, must ensure that the switch is in the open
does not
allow closing!
1, anode startup cable PLD80.00.07: from the high voltage generator AUX J7 PWB
figure 4-8.
50
PLD8000User’s Manual
JS-8000-13 V ER2.0
Shielding E J7-3
2, exposure control cable DRA20.04: from the high voltage generator AUX J2 PWB
8 "3".
RLS_2 8 J2-5
RLS_1 7 J2-6
XPB1- 2 J2-7
XPB1+ 1 J2-8
3, synchronous box POWER cable DRA20.02: from the high voltage generator AUX J11
PWB BOARD interface connected to synchronous POWER socket box interface, as shown
in figure 4 to 8 "4".
4, generator CONTROL cable PLD80.00.05: from the high voltage generator CONTROL
BOARD interface J19 to connect to the console interface XZ1 CONTROL box, as shown in
figure 4 to 8 "5".
51
PLD8000User’s Manual
JS-8000-13 V ER2.0
ON XZ1-9 J19-9
5, generator serial communication cable PLD80.00.06: from the high voltage generator
figure 4 to 8 "6".
RXD 3 J22-2
TXD 2 J22-3
GND 5 J22-5
6, high voltage cable: connect to photography from the high voltage generator fuel tank bed
52
PLD8000User’s Manual
JS-8000-13 V ER2.0
bulb components, as shown in figure 4-9 and 1 – anode 2 - the cathode. Two cables
non-polarity restrictions, but before the installation must be prepared to tag, prevent the
anode and the cathode cable meet the cross. Check whether the plugs and high pressure
outlet is clean, or with anhydrous alcohol or ethyl ether scrub clean. Evenly at the top of the
high voltage plug insulation silicon grease, alignment hole reciprocating slowly inserted,
rule out the air of the socket. Finally, tighten gland and locked tightly screw fixation.
High voltage cable bending is not less than 10 times its diameter, as
safety.
8, ionization chamber cable PLD80.00.08: from the high voltage generator AEC BOARD
J11 connecting to the ionization chamber extension cable plug amplifier, as shown in figure
4 to 10.
53
PLD8000User’s Manual
JS-8000-13 V ER2.0
There is no connection 1 1
Reset/exposure 4 4
-12VDC 7 7
+12VDC 8 8
GND (Shielding) 9 9
54
PLD8000User’s Manual
JS-8000-13 V ER2.0
1 ball tube pillar, photography bed and lathe bed cable connection
1) the bulb out of limit switch cable: swing arm unit left terminal row 9-1, 9-2 and 9-3.
2) the ball tube retracted limit switch cable: swing arm unit left terminal row 10-1, 10-3-2
and 10.
3) ball tube telescopic motor cable: swing arm unit right terminal line of 419 and 420.
1) the console communication power supply cable PLD80.00.04: from the console
55
PLD8000User’s Manual
JS-8000-13 V ER2.0
operation box DB9 serial port/F connecting to photography TABLE bed PCB terminal
terminal row X6 J32 and photography bed base, as shown in figure 4-11 "1".
A XZ2-6 J32-5
B XZ2-7 J32-7
2) audio power cable PLD80.00.11 from control box air outlet XZ3 connected to terminal
row X5 photography bed base, as shown in figure 4-12 "2";Speaker cable PLD80.00.12
from control box air outlet XZ3 connected to the speaker, as shown in figure 4-12 "3".
0V XZ3-3 X5-11
Figure 4-12 :Photography bed and control box cable connection diagram
56
PLD8000User’s Manual
JS-8000-13 V ER2.0
1) vision light cable PLD80.00.10: from terminal diagnosis bed row X5 connection to the
limit of 2 needle beam splitter XZ2 air outlet (2 needle aviation plug), as shown in figure
4-13 "1".
2) limit beam splitter PLD80.00.09: communication power supply cable from photography
TABLE bed PCB terminal J31, power supply and driver PCB terminal J4, diagnosis bed
terminal row X5 of prefabricated cable connector (17 stitches aviation plug) connected to
the limit of beam splitter 17 stitches XZ1 air outlet, as shown in figure 4-13 "2".
Figure 4-13 : Photography bed and the limit of beam splitter cable connection diagram
2, flat-panel detector control cable DRA20.05: from flat panel detector EXT_1NF interface
3, a serial port communication cable DRA20.03: from synchronous box serial COM_PC
4, tablet power cable: from the tablet power supply box connected to the flat panel detector,
57
PLD8000User’s Manual
JS-8000-13 V ER2.0
Photography base installation fixed bed, the ball tube - the limit of beam splitter of the
installation, detector and AEC ionization chamber install accurate, confirms the accuracy of
Check equipment ac power supply circuit and the dc power supply circuit for short circuit,
especially the power line can't bare or touch metal case.External power supply voltage is
58
PLD8000User’s Manual
JS-8000-13 V ER2.0
59
PLD8000User’s Manual
JS-8000-13 V ER2.0
On the workstation at the same time press the SID of 1.0 m and 1.8 m button, and then press
According to 1.0 m and 1.8 m button switch can according to positive and negative
direction indicator state changes (1:light),operate the movement control key,measure the
mast swing Angle or bed body up the pour point of view , to adjust to each state in the
column automatically return button to save the current state of the AD value, successfully
saved nearly control plate column automatically return the light 500 ms.
Table 5-1 each AD and LED status indicates the corresponding table
LED7(Columns
LED1(1.0m) LED2(1.8m) Instructions
automatically return)
JS-8000-13 V ER2.0
the AD
of
Debug state, the Collimator control are described in the following table 5-2 functions.
Collimator transverse
KEY_XSQ_H_C Debug mode, manually bulb SID
close button
Point moving bed body up down buttons (guarantee the bed body is not limit location), the
same direction is marked down direction and key of the bed body the direction is right, if on
the contrary, to exchange the bed body up any two input line of the motor, can make the
61
PLD8000User’s Manual
JS-8000-13 V ER2.0
motor reverse.
Bed body eventually using photoelectric protection restrictions on both ends of the coupling
switch as, potentiometer on testing, on the basis of the restrictions on the rotating
1/ 2-- Body down restrictions on both ends of the bed 3—Pour the potentiometer
casualties.
Debugging:
Confirm lathe bed in level state, the demolition of the limit of protection pour photoelectric
62
PLD8000User’s Manual
JS-8000-13 V ER2.0
coupling switch mechanism, loosen the screws of the potentiometer far away from the big
gear, trimmer potentiometer and measure the potentiometer output voltage of 1.2 V at the
same time, fixed screw lock potentiometer, save the bed body in accordance with table 5-1
0 ° bit AD value;
C, slow dumping of lathe bed, check whether confirm lathe bed level automatically stop; Or
a potentiometer to adjust the bed down, save 0 ° and 90 ° AD value, and the standard of the
lathe bed between dumping and the vertical position can be automatically stop.
D, lathe bed fall protection AD negative Angle value associated with the protection of
column vertically down, due to bed up negative Angle protection AD value and lathe bed
backward down the negative Angle AD values, are saved in the following method: column
vertical downward movement to limit down protection, dumping of lathe bed, slow
measurement bed down potentiometer output to 1 v, in accordance with the table 5-1 save
bed fall protection a negative Angle AD value; Column vertical upward movement to limit
from down protection, dumping of lathe bed to slow - 14 °, measurement bed down
potentiometer output is about 0.3 V, in accordance with the table 5-1 save bed down a
coupling switch block slice, observe 90 ° limit and 15 ° protect limit state of the indicator
lights on the corresponding PCB, covering photoelectric coupling limit switch, the lights
went out after normal adjustment 90 ° protection limit photoelectric switch position
coupling and lathe bed up automatically stop and block slice after a difference of 1 mm do
not limit, adjust to 15 ° protection limit photoelectric switch position coupling and lathe bed
has fallen to negative Angle after automatic stop and catch a difference of 1 mm is not limit.
Inching pillar longitudinal motion buttons (ensure that column is not limit location),
63
PLD8000User’s Manual
JS-8000-13 V ER2.0
direction of motion with the same direction is marked key pillar longitudinal motion
direction is correct, if on the contrary, to exchange any two of the pillar longitudinal motion
Ball tube pillar longitudinal motion using photoelectric coupling switch as the limit of the
A, make moves down, the post in limit barrier pillar down limit switch photoelectric
B, post up mobile, in limit barrier pillar up limit switch photoelectric coupling column
C, adjust the post up limit and pillar down limit, make the pillar of the moving distance of
970 mm.
D, check in bed fall under protection of a negative Angle, column vertical down only to
Point moving bed lateral movement keys (guarantee the bed surface is not limit state),
moving direction and buttons the same direction is marked on the surface of the bed
movement direction is correct, if on the contrary, to exchange bed lateral motion input
64
PLD8000User’s Manual
JS-8000-13 V ER2.0
electric draw stem drive line, can make the reverse movement.
A, make the bed for mobile, in limit barrier bed face left when the limit switch photoelectric
B, make the bed for mobile, right at the limit piece of shade the bed surface limit switch
photoelectric coupling bed surface should be able to automatically stop the movement.
C, adjust the bed left limit moves to the right of the bed and the limit, make the bed surface
Mast swing inching button (guarantee not limit state) mast swing, swing direction with the
same direction is marked key pillar swing direction is correct, if on the contrary, to
exchange any two input of the mast swing motor drive line, can make the motor reverse.
Pay attention to the potentiometer AD value is in the state allows motor movements.
65
PLD8000User’s Manual
JS-8000-13 V ER2.0
1-- Mast swing left limit 2-- Mast swing right limit 3-- Mast swing potentiometer
A. to bed dumped on the horizontal pillar longitudinal motion to the middle position,
confirm the ball tube column in a vertical state, adjust the mast swing limit switch
photoelectric coupling position outside the range of swing, loosen the potentiometer mast
swing set screw make them away from the big gear and fine-tuning potentiometer and
measure the potentiometer output voltage of 2.50 V at the same time, fixed screw lock
B. swing to the left column to + 45 °, measuring pillar potentiometer output is about 4.5 V,
the mast swing save in accordance with table 5-1 + 45 ° AD value; To the right column to -
30 °, save in accordance with table 5-1-30 ° a mast swing AD value; Continue to the right
column to - 45 °, measuring pillar potentiometer output is about 0.5 V, the mast swing save
C. swinging the ball tube pillar, check in the plus or minus 0 ° and 45 ° position can be
automatically stop, or to perform a and b.To observe the mast swing control limit switch
corresponding to the indicator lights on the PCB board, cover the photoelectric coupling
limit switch, the lights went out after normal adjust the mast swing left limit switch position
the ball tube mast swing to + 45 ° automatically stopped and does not limit block slice a
difference of 1 mm; Adjust the mast swing position limit switch tube mast swing the ball to
the right - 45 ° after automatic stop and catch a difference of 1 mm is not limit.
66
PLD8000User’s Manual
JS-8000-13 V ER2.0
In debug state, inching limit beam splitter longitudinal open button (ball tube elevator is not
in the limit state), the tube back (SID) movement is correct, if on the contrary, the exchange
Enter debug state press "limit beam transverse off" the ball tube in a set to manual operation,
detection limit, photoelectric coupling switch corresponding to the indicator lights on the
A, make the bulb into smaller (SID), in the limit block slice keep out ball tube into limit
switch photoelectric coupling tube should be able to automatically stop motion;Make the
bulb back movement (SID), in the limit block slice keep out ball tube withdrawal limit
photoelectric coupling when the switch pipe should be able to automatically stop motion.
B, adjust the pipe into the limit and the ball ball back limit light coupling switch, make the
C, debugging is completed, can be set up according to demand the ball tube in a movement
67
PLD8000User’s Manual
JS-8000-13 V ER2.0
that the electric draw stem movement is not beyond its limit
position,
When the film holder car is in the ”PARK” position, push the radioscopy footswitch under
console, the system will display images in the real-time. When loosen the footswitch, it will
Load the film magazine, the allowable spot film division forms of this kind of film
magazine will be shown on the corresponding position of the interface; if you do not want
to spot film on this film magazine, then you can click "In/Out film" button to take out the
film magazine; it you want to spot film on this film magazine, then click the spot film
division form, then the interface enters the corresponding spot film division form of this
kind of film magazine, after that, you can click "re-selection" to return to the former
If you select the full spot film mode, then you can implement fluoroscopy or directly press
down the shooting button to implement exposure; after the exposure, click "In/Out film"
If you select 2PCS or larger spot film mode, then there would be two kinds of spot film
functions for you to choose from: 1."spot film by area"; 2."automatic spot film".
68
PLD8000User’s Manual
JS-8000-13 V ER2.0
1. "Spot film by area": this function allows you to choose any shading area i.e. to implement
exposure only to the selected area; this method requires the operator to remember on
which shading area that he has implemented exposure; otherwise he would implement
Click "spot film by area" to enter the interface of spot film by area, it prompts " click to
choose the exposure area directly", click area "1" or area "2", or even area "3" or "4"
among the shading icons, and the selected area is corresponding to the actual film
magazine, after that, you can implement fluoroscopy or press down the shooting button
to implement exposure; click "Back" button, then you can exit from the function of spot
film by area, or you can click "In/Out film" button to take out the film magazine directly;
after selecting the exposure area, click "Cancel", then you can exit from this area; after
exposure, you can return to select another area or take out the cartridge according to
prompts.
2."Automatic spot film": this function is to implement spot film according to area sequence
i.e. according to the area sequence marked with icons, i.e. you can implement exposure
on the next area only when it has been done on the former area, by parity of reasoning,
until it has implemented exposure on all areas, and the operator does not have to worry
about repeated exposure of the same area. The operating method is as follows:
Click "automatic spot film" to enter the interface of automatic spot film, it prompts
"exposure area 1 has been set", and you can implement fluoroscopy or press down the
automatic spot film mode, and exposure has been done, then it goes to the next area, and
prompts "exposure area X has been done" and "exposure area X has been set", until it has
69
PLD8000User’s Manual
JS-8000-13 V ER2.0
Fast shooting function: each of the spot film modes has fast shooting function; if you need
fast shooting function, then click "preset cartridges" button, the interface enters fast
shooting page, the trolley moves to the corresponding spot film position, and the collimator
automatically traces to the right position, and now you can press down the shooting button
to implement exposure; fluoroscopy is not allowed during fast shooting; when you want to
cancel fast shooting function, click "Back", then it would return to the previous state
interface;
This operation interface consists of characters combined with icons, thus it makes the
Radiographic table is in a supine position when it is in photography, the body lie on the
table, put into the cassette, light Collimator field lamp, move the table surface and
longitudinal movement tube column left/right, select the part which need to be photography,
and then select the photography condition. Press the exposure handbrake I file, the tube
anode is start-up, the filament temperature gets high, and the film car moves to the exposure
position. Until the light of “ READY ” is on, press the exposure handbrake II, generator
implement exposure, exposure indicator light is on and "beep" sound is sent out. End of
JS-8000-13 V ER2.0
1、Before power up the host system, please powers up the X-ray generator ,
detector and all the other x-ray system components needed.
2、Make sure that the peripheral X-ray system components are on and working
properly.
3、 Power up the Host PC and login the PC’s operating system
4、Log in to the DROC application.
The following subsections provide specific instructions for each of those steps:
Step 1: Powering up the X-ray generator, detector , etc
Please refer to the related instructions from corresponding manufacturers.
Step2: Checking the System
Once the X-ray generator is powered on, perform the following checks:
● Remove any patient supports or other objects so that they do not interfere
with the movement of the X-ray tube's suspension arm.
● Switch on the collimator’s light to ensure that the area to be exposed to the
X-ray beam is illuminated.
● Check the releases, if applicable, on the X-ray tube's suspension arm, the
Bucky , and the table to ensure that the controls are operating properly.
You can now power up DROC.
Step3: Powering Up the Host PC
Power up the HOST PC and Log in to the PC's operating system;; if the monitor is
not on, press the power button on the monitor.
71
PLD8000User’s Manual
JS-8000-13 V ER2.0
To sign into the application, choose your level from Level Icon (Technician,
Application, Manager) and your name from the User Name drop-down list,
In case you CAN NOT see this window , and a message box
No Dongle available or Dongle is invalid! DROC System will
exit!” shows up instead. Check your USB license key to make
sure it is well installed or contact our engineers for further
support.
72
PLD8000User’s Manual
JS-8000-13 V ER2.0
The initialization window will show the initialization status of the DROC, shown as
below:
There are some initialization tasks of DROC system; the result of each task
(successful or failed) will be displayed one after another.
From System menu (located in the Menu Bar), click User Management menu to
open the User Management window, shown below. When you login with the
manager privilege account, all users are listed. Otherwise, only your own account is
listed.
73
PLD8000User’s Manual
JS-8000-13 V ER2.0
In this window, you can create a new account, edit the existing user or delete an
existinguser.
74
PLD8000User’s Manual
JS-8000-13 V ER2.0
2. Input the new user’s information and choose a user level you want to create
3. Click OK button to accept the new account, and the New User window closes.
Then the new account information displays in the User list on the User
Management window.
Click Clear button to empty all of new user information, and reenter the information
in New User window.
Click Cancel button to discard the entry of the new user.
To delete an existing user
In the User Management window, choose the user you want to delete and click
Delete button.
To edit user information
1. Choose the specific user you want to edit.
2. Click Edit button located at the bottom of User Management window. The Edit
User window is showed as below.
75
PLD8000User’s Manual
JS-8000-13 V ER2.0
The general workflow to acquire and send images to output devices involves the
following steps.
Steps:
1、Identifying the patient and Study;
2、Taking an exposure;
3、 Viewing and accepting the image; Optionally, you can reject the image and
re-capture till an acceptable image is acquired;
4、After all images captured, the data will be transmitted to the default node.
Alternatively, users can send the images to other node or print the images;
5、Closing the patient/Study;
The following subsections provide specific instructions for each of those steps.
Step 1: Identifying the Patient and Study
76
PLD8000User’s Manual
JS-8000-13 V ER2.0
Before capturing images, you must identify the patient with whom they are to be
associated. Proper identification ensures that the image data is associated with the
correct patient information.
Depending on the system configuration at your site, there are three ways to identify
the patient from whom you want to acquire images, as described below.
JS-8000-13 V ER2.0
If your system supports bar code scanning, you can use the bar code scanner to
read the patient ID or some other unique identifying number. If the ID is found in
worklist (if this feature is available) or in the local DROC patient database, the
corresponding patient’s study will be opened. If the ID is not found, you must enter
that patient manually.
The patient/study list in the Worklist window is shown as below:
You can capture images for the patient/study of SCHEDULED or IN PROGRESS status;
even can super-add images to a COMPLETED status patient/study as long as it still
existing in the Local database.
NOTE: Make sure you enter patient names and IDs correctly.
Incorrect entries can cause duplicate patient records for the
same patient. As a result it will be more difficult to find the
patient in future search operations.
To add a new patient without a barcode, fill the information in New Patient page
78
PLD8000User’s Manual
JS-8000-13 V ER2.0
on Worklist window.
79
PLD8000User’s Manual
JS-8000-13 V ER2.0
The New Patient page is displayed below. Please note that, for your site, the fields in this
window may be different.
In this page, specify the patient information and determine the examination that
patient is going to take.
Fields labeled with an asterisk (*) are required to be filled in. You can switch to the
next field by pressing Tab key. After birthday entered, age is calculated and showed
by system automatically. The name will be displayed in the format
LAST^MIDDLE^FIRST. If you don’t want “^“ character to appear in name, you can
input all content of name in Last Name field.
● Click Exam button. The patient information will be loaded on the Exam
80
PLD8000User’s Manual
JS-8000-13 V ER2.0
window for technicians to capture images.
● Click Save To Local button. The patient information save to local worklist
database and system is still in Worklist window.
When specifying the protocol, click Add Protocol button , the Add Protocol
window shows like below
● Select a Body Part from the left list, the corresponding protocol will be listed in the
upper right area. If select “Often Used” item in the left list, only protocols marked
“often used” display in the upper right area (The often used protocol is defined by
DROC configuration).
● Click the protocol to load the corresponding views into the lower right area.
To delete the view from the lower right area, click Delete Selected View
81
PLD8000User’s Manual
JS-8000-13 V ER2.0
82
PLD8000User’s Manual
JS-8000-13 V ER2.0
Only the Manager can edit the information. (More details about user account, please
refer to relevant topic of Section E)
1)From the Local page in the Worklist window, select the patient/Study whose
fundamental information you want to correct.
83
PLD8000User’s Manual
JS-8000-13 V ER2.0
84
PLD8000User’s Manual
JS-8000-13 V ER2.0
4)Click Accept button to save the correction, the information will be renewed and
shown in the Local page as
85
PLD8000User’s Manual
JS-8000-13 V ER2.0
From the Query Panel in the Local/Worklist page, you can query a patient/Study by
various searching criteria such as Patient ID, Patient Name, Accession Number
and date.
Input the key information associated to the desired patient/Study, and then click
Query button. DROC application will search the local database, and then the result
will be displayed in the Local/Worklist page.
86
PLD8000User’s Manual
JS-8000-13 V ER2.0
If multiple patients/studies match the criteria, they will be listed in the Local/Worklist
page.You can locate the exact Study or patient to perform the next operation.
To manually refresh the Worklist from Radiology Information System (RIS)
If your site is connecting to RIS, DROC can retrieve the worklist in background and
automatically refresh the Worklist page. You can override this operation manually
by clicking the Query button.
Select the patient/Study that you want to delete from the Local/Worklist page; then
87
PLD8000User’s Manual
JS-8000-13 V ER2.0
88
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 40: Study loaded
To load MULTIPLE studies:
From the Local/Worklist page, select several Studies belongs to same patient by
pressing Ctrl + left click mouse; Click the Exam button in Next Step area; Studie
relating to the desired patient are loaded in the Exam Window:
The UI in your site may not show the some buttons after its
installation due to the customization. Please refer to your
support engineer for more details.
You can select a view by highlighting the view icon, and the Study Description will
change as well when you shift between views.
At the time the system is installed, default exposure settings and capturing
89
PLD8000User’s Manual
JS-8000-13 V ER2.0
sequences are configured according to every specific view. You can use manual
mode or photo timingmode to adjust the default exposure settings.
Technically , you could simply use the default parameters. They should only be
changed for special circumstances, such as a very small person or pulmonary
emphysema.
If you need to adjust the exposure parameters or capturing sequence, refer to Section
C Adjusting exposure parameters for detailed instructions.
Adjusting kV
The range of kV is confined by the generator.
90
PLD8000User’s Manual
JS-8000-13 V ER2.0
The Automatic Exposure Control (AEC) system monitors X-radiation dosage which
is passing through ion chambers and automatically terminates the exposure after
theX-radiation dosage reaches a predetermined amount of radiation.
91
PLD8000User’s Manual
JS-8000-13 V ER2.0
You can select AEC detector fields in each ion chamber individually or in any
combination.You can vary the density setting between +4 and -4 (0 is neutral).
To change the density setting, click on the Increase Density button or Decrease
Densitybutton located to the right of Density value field, as shown above. For the
increases or decreases amplitude of each time, please refer to the Service Manual of
Generator.
Adjusting Exposure timing using MAS setting
You can manually specify exposure timing by setting the mAs value (shown below).
With mAs button selected, the exposure is terminated when the specified mAs value
is reached.
To specify the exposure by mAs, ensure the mAs button in the Exposure Control
panel is selected, then click on the Increase mAs button or Decrease mAs button
92
PLD8000User’s Manual
JS-8000-13 V ER2.0
located to the right of MAS value field to set the desired value. The value changes by
one unit each time. Pressing and holding the button causes the value to increase or
decrease automatically until the button is released. If the value reaches maximum or
minimum, the value will remain unchanged.
With the mAs button selected, the TIME controls are dimmed. Since the TIME
settings and MAS setting are dependent to each other.
You can manually specify exposure timing by setting the mA and the ms value. When
using the TIME settings (shown below), the exposure is terminated when the specified
time value is reached. This mode is used when a specific time must be achieved,
such as in a breathing study or for stop motion. Regarding the reciprocal grid’s
requirement on exposure window, the ms value should be guaranteed above 50 when
the system employs reciprocal Grid.
To specify the exposure by Time setting, make sure the Time button in the
Exposure Control panel is selected (shown below), then click on the Increase
mA/ms button or Decrease mA/ms button to set the desired values. The value
increases and decreases in steps of one unit. Pressing and holding the button causes
the value to increase or decrease automatically until the button is released. If the
value reaches maximum or minimum, the value will remain unchanged.
With the Time button selected, the mAs controls are dimmed. Since the TIME
settings and mAs setting are dependent to each other. Changing mA/ms values
causes the mAs value changes.
93
PLD8000User’s Manual
JS-8000-13 V ER2.0
If the selected Focal Spot Size is not compatible with the tube
and focal spot, the system inhibits exposure and displays an
overload condition. An overload condition may exist if the
technique selected exceeds the maximum or is below the
minimum allowable X-ray tube focal spot rating. The tube
parameters are consistent with the X-ray tube manufacturer's
rating charts.
Hardware Status
The status of hardware is displayed in the Common Message area, as shown below.
JS-8000-13 V ER2.0
In Process window, you can see the patient/study information on the top left of the
image, and the real exposure dose information on the top right of the image.
To add a marker to the image, click on the marker button. The added marker
displays on the top left corner of the image. The added marker could be dragged to
wherever you want on the image within the clipping area.
Click Define Mark button, Labe list displays as the following figure. You can add
information into image by clicking an item from the list.
95
PLD8000User’s Manual
JS-8000-13 V ER2.0
96
PLD8000User’s Manual
JS-8000-13 V ER2.0
Figure 54: The image with added label
97
PLD8000User’s Manual
JS-8000-13 V ER2.0
Clip an Image
When an image has been captured and is displayed in the Process window, you can
clipthe image for output to a DICOM print or storage device.
When you accept the image, the full image along with any clipping information, is
stored in the local disk and can be recalled for clipping again within a period of time.
Meanwhile, the clipped image is sent to the DICOM output device.
To remove the clip line and not clip the image: click the Remove Clip button.
To fix clip line: click Save Clip button, the current clip line set to be the default clip
line.
To change clip size: choose one item from the clip size drop-down list which can be
triggered by clicking Define Clip Size button,the items in the drop down list represent
standard film size with inch being unit.
98
PLD8000User’s Manual
JS-8000-13 V ER2.0
To Rotate the image by Rotate Any Angle button, make sure the Rotate Any Angle
button is clicked (shown below), then click the Increase button or Decrease button to
set the desired values. The value increases and decreases in steps of one unit.
Pressing and holding the button causes the value to increase or decrease
automatically until the button is released. If the value reaches maximum or minimum,
the value will remain unchanged.
99
PLD8000User’s Manual
JS-8000-13 V ER2.0
100
PLD8000User’s Manual
JS-8000-13 V ER2.0
Click the LUT, Symphony button or Custom Symphony, system would use the default
parameters to process the image.
When previewing an image, you can modify the LUT settings of the image. The LUT
controls the minimum and maximum densities used in the printed or viewed image.
101
PLD8000User’s Manual
JS-8000-13 V ER2.0
When the system is installed, customized default LUT settings are configured for
each procedure type. The image’s original LUT is created by applying the procedure’s
specified LUT settings to the image data.
Typically, you need to change the LUT only for patients with health conditions that
may affect the image quality, such as emphysema for chest views.
To modify an image’s LUT: drag one of the three square handles of the graph line
until the image shows the way you want. If the top or the bottom one is moved out of
display are and is invisible, one the middle one can be dragged. You can also press
and hold the right mouse button, and move the mouse on image preview area to
change the LUT. Vertical movement changes window width of LUT, horizontal
movement changes window level of LUT.
102
PLD8000User’s Manual
JS-8000-13 V ER2.0
5.2.4.5 Outputting Images manually (optional)
After captured, the images will be sent to default destinations or printers automatically.
But you can still re-send the images to different destinations or resend images from
Output window.
There are two types of output destinations: archive server for storage and printer
for printing. These two destinations (may contain zero or multiple nodes) can be
configured for all images belong to the same procedure. Furthermore, the film
orientation, size and format can be configured for each different procedure. When an
exam is opened, these configurations are loaded automatically. You can also change
them manually for current Study .
For more information, please refer to Section D: Working with Worklist Studies.
You can quit the Exam window to perform other operations, no matter whether all
the images are captured or not.
To quit the Exam window, click the button in Next Step area.
● If all of the images in the current study has being acquired, the status of this
Study will become Completed in Worklist window;
● Otherwise, the status will become In Progress in Worklist window.
You can pick up the incomplete Study from the Worklist window to proceed with the
exposures whenever you like.
The fact that a Study is incomplete does not prevent the patient
record and associated images from being deleted which is part
of the system's storage space reclamation process. For more
information about the reclamation, and how to protect a
patient record and associated images from being deleted,
refer to Section E.When Study is closed and no new patient is
selected, the detector will perform offset calibration; if no Study
103
PLD8000User’s Manual
JS-8000-13 V ER2.0
To log off the DROC application, click the Exit button from the Menu Bar. The DROC
Close dialog displays.
Typically, you should leave the system components on. The DROC HOST PC, monitor,
and Detector are able to enter an energy-saving "sleep" mode1 when they are not
used for a period of time.
JS-8000-13 V ER2.0
To power off the system, it is recommended that you perform the power off tasks in
the following sequence,
Steps:
105
PLD8000User’s Manual
JS-8000-13 V ER2.0
CHAPTER6 MAINTENANCE
For the disinfection work, unit power must be shut down! And
106
PLD8000User’s Manual
JS-8000-13 V ER2.0
disinfection liquid should be paid attention that it can’t flow into the unit interior, lest the
phenomenon of leakage and short circuit is occured, which will lead to all kinds of malfunction,
even cause injuries.
107
PLD8000User’s Manual
JS-8000-13 V ER2.0
(1)、All surfaces of photography table should be kept clean, dry , and variety of
residues can be scrubbed with a soft cloth dipped with few absolute alcohol
(2)、Regular check if bearing or movement orbit is short of oil .After cleaning , it
should coat with lubricant (grease) in time.
(3)、No flammable liquids, gases near the equipment to prevent explosion.
(4)、Protect grounding facilities, and inspections should be done every half year.
(5)、Diagnostic table’s sterilization method are same as the unit’s.
JS-8000-13 V ER2.0
check out the problem, it can only be used after problem solved.
5、 Do not press heavily on all buttons of controller to avoid dislocation, injury or poor
contact, even damaged.
6 、 Check frequently protective earth of each components is in good condition
(grounding resistance is very small).
7、Check power supply on a regular basis, and check power supply resistance value
(or the power supply voltage value is changed), to make sure it meet machine’s
requirement.
8、After using the controller, it should be cut off the power in time.
9、Avoid striking of x-ray tube components during operation.
10、 When X-ray tube components are used for the first time or idled for a longer time,
it must be progressive loading before X-ray tube used ,and do the sophisticated
test to avoid damaging to the X-ray tube by instantaneous load.
11 、 When machine works continuously, please pay attention to the X-ray tube
assembly to allow Thermal capacity, or check surface temperature of X-ray tube
assembly , it can’t exceed 70 ℃.
109
PLD8000User’s Manual
JS-8000-13 V ER2.0
1 .Diagnostic Table
Fault
Adjustment method
Code
EROR0 Touch screen displays cue
2.HV Generator
Fault
information problem Adjustment method
Code
Consult your service
E001 GEN EPROM ERR representative.
E004 GEN RTC ERROR Generator CPU real Reset time and date.
110
PLD8000User’s Manual
JS-8000-13 V ER2.0
time clock doesn't Consult your service
work. representative.
Consult your service
E005 PS CONTACT ERR
representative.
Power unit off and retry rotor
start.
E006 ROTOR FAULT
Consult your service
representative.
Consult your service
E007 FILAMENT FAULT
representative.
Consult your service
E008 KV/MA FAULT
representative.
Power supply is not Retry exposure.
E009 PS NOT READY ready to make an Consult your service
exposuer representative.
Consult your service
E010 LOW KV FAULT
representative.
Consult your service
E011 HIGH MA FAULT
representative.
Consult your service
E012 LOW MA FAULT
representative.
Re-take exposure if
Operator released
necessary .
E013 MANUAL TERMIN exposure switch
Consult your service
during exposure .
representative.
Check exposure
AEC exposure technique settings.
E014 AEC BUT ERROR exceeded Consult your service
allowed back up time. representative.
Check exposure
AEC exposure
technique settings.
exceeded
E015 AEC BUT ERROR Consult your service
allowed back up mAs.
representative.
Tomo exposure
Check exposure
exceeded back-up
E016 TOMO BUT ERROR technique settings.
time.
JS-8000-13 V ER2.0
JS-8000-13 V ER2.0
representative.
Switch OFF
generator. Prevent
further use of
E043 KV ERROR
generator, then.
consult your service
representative.
Received
communication
E044 COMM ERROR Reset error.
message not valid and
ignored.
Received message is
valid,
E045 NOT SUPPORTED but not supported by Reset error.
this
system.
Attempting to change
CM
thickness in AEC
mode if
the receptor is not Consult your service
E046 MODE INHIBITED programmed for mAs representative.
backup. Reset error.
Received message is
valid,
but not allowed during
present state.
Consult your service
E048 FOCUS MISMATCH
representative.
Requested function
has not
Consult your service
E049 NOT ENABLED been programmed to
representative.
be
enabled.
Consult your service
E050 GEN DATA ERROR
representative.
Generator has Check that X-Ray
detected no tube is pointing at
E051 AEC DEVICE ERR feedback signal from correct AEC device.
AEC Consult your service
device. representative.
Consult your service
E052 HIGH SF CURRENT
representative.
E053 HIGH LF CURRENT Consult your service
113
PLD8000User’s Manual
JS-8000-13 V ER2.0
representative.
AEC OUT OF Consult your service
E054
RANGE representative.
AEC enabled but no
E055 NO FIELDS ACTIVE fields Select AEC field(s).
are selected.
NO TUBE Consult your service
E056
SELECTED representative.
Consult your service
E057 AEC STOP ERROR
representative.
CONSOLE BUT Consult your service
E058
ERR. representative.
X-Ray tube housing
HOUSE HEAT
E059 has Wait for tube to cool.
LIMIT
exceeded limit.
Consult your service
E060 EXP. KV HIGH
representative.
Consult your service
E061 EXP. KV LOW
representative.
Consult your service
E062 EXP_ SW ERROR
representative.
FACTORY Consult your service
E063
DEFAULTS representative.
Generator duty limit Consult your service
E067 PS DUTY LIMIT
reached. representative.
MAS OVERTIME Consult your service
E069
ERR representative.
SOFTWARE KEY Consult your service
E070
ERR representative.
The accumulated DAP
Reset the DAP.
DAP DOSE value exceeds the
E071 Consult your service
OVERFLW display
representative.
limit.
The DAP device is not Consult your service
E072 DAP DEVICE ERR.
functional. representative.
Consult your service
E073 DAP DATA ERROR
representative.
INVERTER 1 Consult your service
E074
ERROR representative.
INVERTER 2 Consult your service
E075
ERROR representative.
INVERTER 3 Consult your service
E076
ERROR representative.
114
PLD8000User’s Manual
JS-8000-13 V ER2.0
Consult your service
E077 RES. CIRCUIT ERR
representative.
BUCKY1 Consult your service
E078
INTERLOCK representative.
BUCKY2 Consult your service
E079
INTERLOCK representative.
INTERLOCK 1 Consult your service
E080
OPEN representative.
INTERLOCK 2 Consult your service
E081
OPEN representative.
KV OVER Consult your service
E082
VOLTAGE representative.
Consult your service
E083 ANODE MA FAULT
representative.
CATHODE MA Consult your service
E084
FAULT representative.
ROTOR MAIN Consult your service
E085
FAULT representative.
ROTOR SHFT Consult your service
E086
FAULT representative.
115
PLD8000User’s Manual
JS-8000-13 V ER2.0
CHAPTER7 ATTACHMENT
Quantit
NO Name Unit
y
1 Fuse(380V/16A): pcs 3
2 Fuse(220V/2A): pcs 2
3 Fuse(220V/5A): pcs 1
116
PLD8000User’s Manual
JS-8000-13 V ER2.0
7.2 Safety Performance Inspection Report
1. Electrolyte strength under normal working temperature
Test condition: AC Voltage 1500V Time 1 min
Conclusion:
2 .Grounding resistance
Protective earthing and protective grounding of all possible resistance between the metal parts
touch, maximum: Ω<0.1Ω
Conclusion:
Conclusion:
JS-8000-13 V ER2.0
Packing list(1)
NO Name Unit Quantity
1 Radiographic table pcs 1
2 Grid (South Korea) Pcs 1
3 Collimator (auto/ manual) Pcs 1
4 Console Pcs 1
5 Connection wires: Pcs
XJZ00-01 generator control wire Pcs 1
XJZ00-02 PLC-EVIEW communication cable Pcs 1
XJZ00-03 boot, brake, CCD, EVIEW power wire pcs 1
XJZ00-04 console control cable pcs 1
XJZ00-05 brake, image intensifier, foot switch, motor pcs 1
cable
XJZ00-06 collimator, load indicator cable pcs 1
XJZ00-07 cassette motor cable pcs 1
XJZ00-08 table control signal cable pcs 1
XJZ00-10 ground wire Pcs 7
XJZ00-11 Radiographic table power cable Pcs 1
XJZ00-12 heat protection and rotary anode wire Pcs 1
XJZ00-15 speaker wire 2 Pcs 1
XJZ00-17 microphone wire pcs 1
XJZ00-19 speaker wire 1 pcs 1
XJZ00-20 cassette AD wire pcs 1
XJZ00-05A three-phase AC motor line pcs 1
XJZ00-23 video cable Pcs 2
XJZ00-35 In the control of power cable Pcs 1
ZDC20-02 the Collimator lamp wire Pcs 1
ZDC20-04 Radiographic table internal magazine signal pcs 1
cable
X-ray tube components pcs 1
Image Intensifier pcs 1
High voltage generator(CPI) pcs 1
6
7
8
JS-8000-13 V ER2.0
Company: Nanjing Perlove Medical Equipment Co., Ltd. Tel: 025-87187780
packing list(2)