Calibracion Fluke Certificates Examples
Calibracion Fluke Certificates Examples
more properly termed a Z540.1 calibration certicate. This calibration was conducted using an unbroken chain of standards to the Fluke Volt, which is traceable to the U.S.
representation of the volt, through the internationally accepted value of the Josephson constant Kj=483597.9 GHz/V and a 10
Volt Josephson Array Voltage Standard.
It documents that a specic instrument, identied by Frequency
model and serial number, was tested using Fluke’s This calibration was conducted using an unbroken chain of standards to a GPS disciplined oscillator frequency standard,
traceable to the United States Naval Observatory (USNO), which is traceable to the National Institute of Standards and
international standards, in accordance with ormal AC Voltage (1 MHz and below), Resistance, DC Current, AC Current (10 mA and above), Capacitance, Inductance and
Phase
This calibration was conducted using an unbroken chain of standards which are traceable to NIST.
American Standard named Z540.1. (More specically, AC Voltage ( > 1 MHz) and RF Power
its present version is named ANSI/NCSL Z540.1-1994 This calibration was conducted using an unbroken chain of standards which are traceable to NIST or the National Research
Council Canada (NRC).
(R2002) Part I.) With this certicate, details o the AC Current ( < 10 mA )
individual tests are provided with a variety o sup- This calibration was conducted using an unbroken chain of standards which are traceable to Physikalisch Technische
Bundesanstalt (PTB) (German National Metrology I nstitute).
etc. It doesn’t usually include specic measurement traceability is based on the physical phenomena in which the equilibrium relative humidity values associated with certain
saturated salt solutions are known.
Calibration Certificate
Fluke Corporation 425.446.6378 Fax 425.446.5649 www.fluke.com
Descripton: Multi-Product Calibrator Certificate Number: Sample
Model: 5520A Date of Certificate: 20-Nov-2006 Certificate Number: F3128001 Date of Certificate: 20-Nov-2006 Page 3 of 8
Standards Used
Serial number: Sample Only Temperature: 23 ± 3°C PM02
Customer Name: Fluke Corporation Relative Humidity: Less than 70% RH
DC Voltage
City, State: Everett, WA Output/Range Measurement Deviation TUR 90 Day Spec
0 V, 330 mV 231.169 nV 0.231 uV 1.7 1.000 µV
Procedure Name: 5520A-500 329 mV, 330 mV 329.000 mV 0.000 % 0.002 %
-329 mV, 330 mV -328.998 mV 0.000 % 0.002 %
0 V, 3.30 V 66.681 nV 0.067 uV 3.0 2.000 µV
1 V, 3.30 V 999.998 mV 0.000 % 3.8 0.001 %
Measurement uncertainties at the time of test are calculated in accordance with the method described in NIST TN1297, for a -1 V, 3.30 V -999.995 mV -0.001 % 0.001 %
confidence level of 95% using a coverage factor of approximately k=2. 3.29 V, 3.30 V 3.290 V 0.000 % 3.8 0.001 %
-3.29 V, 3.30 V -3.290 V 0.000 % 3.7 0.001 %
In the attached measurement results deviation may be expressed with units, Measured Value (MV) – Nominal Value (NV) or as 0 V, 33 V -3.027 µV -3.027 µV 20.000 µV
a proportion of the nominal value ((NV-NV)/NV), expressed as percent. 10 V, 33 V 10.000 V 0.000 % 0.001 %
-10 V, 33 V -10.000 V 0.000 % 0.001 %
This calibration certificate applies only to the item identified and shall not be reproduced other than in full, without specific written 32.9 V, 33 V 32.900 V 0.000 % 0.001 %
approval by Fluke Corporation. The user is obliged to have the object recalibrated at appropriate intervals. Calibration -32.9 V, 33 V -32.900 V 0.000 % 3.6 0.001 %
certificates without signatures are not valid. 50 V, 330 V 50.000 V 0.001 % 0.002 %
329 V, 330 V 329.000 V 0.000 % 3.4 0.002 %
-50 V, 330 V -50.000 V 0.000 % 0.002 %
Fluke's calibration system complies with the requirements of ANSI/NCSL Z540A.1-1994 (R2002).
-329 V, 330 V -328.999 V 0.000 % 3.4 0.002 %
334 V, 1 kV 334.002 V 0.001 % 0.002 %
900 V, 1 kV 900.005 V 0.001 % 3.7 0.002 %
1.020 kV, 1 kV 1.020 kV 0.001 % 3.7 0.002 %
-334 V, 1 kV -334.002 V 0.000 % 0.002 %
-900 V, 1 kV -900.004 V 0.000 % 3.7 0.002 %
-1.020 kV, 1 kV -1.02000 kV 0.00045 % 3.7 0.00165 %
Aux Out/Range
0 mV, 300 mV -13.684 µV -0.014 mV 0.350 mV
329 mV, 300 mV 329.008 mV 0.002 % 0.136 %
-329 mV, 300 mV -329.032 mV 0.010 % 0.136 %
330 mV, 3 V 329.990 mV -0.003 % 0.136 %
3.29 V, 3 V 3.290 V 0.001 % 0.041 %
-3.29 V, 3 V -3.290 V 0.001 % 0.041 %
7 V, 7 V 6.999 V -0.008 % 0.035 %
-7 V, 7 V -6.999 V -0.008 % 0.035 %
DC Current
Output/Range Measurement Deviation TUR 90 Day Spec
0 A, 300 µA 206.661 pA 0.207 nA 20.000 nA
190 µA, 300 µA 189.996 µA -0.002 % 0.023 %
Kim Bailey Nicholas Mason -190 µA, 300 µA -189.995 µA -0.002 % 0.023 %
Technician Factory Annex Deputy 329 µA, 300 µA 328.994 µA -0.002 % 0.018 %
-329 µA, 300 µA -328.993 µA -0.002 % 0.018 %
0 A, 3 mA 1.417 nA 1.417 nA 50.000 nA
1.90 mA, 3 mA 1.900 mA 0.000 % 0.011 %
-1.90 mA, 3 mA -1.900 mA -0.001 % 0.011 %
3.29 mA, 3 mA 3.290 mA 0.000 % 0.010 %
-3.29 mA, 3 mA -3.290 mA 0.000 % 0.010 %
0 A, 30 mA 26.477 nA 0.026 uA 2.2 0.250 uA
19 mA, 30 mA 19.000 mA 0.001 % 3.6 0.009 %
-19 mA, 30 mA -19.000 mA 0.000 % 3.6 0.009 %
32.9 mA, 30 mA 32.900 mA 0.001 % 3.6 0.009 %
-32.9 mA, 30 mA -32.900 mA 0.000 % 3.5 0.009 %
Page 1 of 1 0 A, 300 mA 99.708 nA 0.100 uA 2.500 µA
Fluke Corporation 425.446.6378 Fax 425.446.5649 www.fluke.com 190 mA, 300 mA 189.996 mA -0.002 % 2.7 0.009 %
-190 mA, 300 mA -189.994 mA -0.003 % 2.7 0.009 %
329 mA, 300 mA 328.994 mA -0.002 % 2.6 0.009 %
-329 mA, 300 mA -328.992 mA -0.002 % 2.6 0.009 %
Figure 4. Z540 calibration certicate 0 A, 2.90 A -14.165 µA -14.165 µA 2.7 40.000 µA
1.09 A, 2.90 A 1.090 A -0.001 % 0.020 %
-1.09 A, 2.90 A -1.090 A -0.001 % 0.020 %
2.99 A, 2.90 A 2.990 A -0.004 % 0.031 %
-2.99 A, 2.90 A -2.990 A -0.005 % 0.031 %
0 A, 20 A -70.623 µA -0.071 mA 0.500 mA
11 A, 20 A 11.000 A 0.000 % 0.043 %
-11 A, 20 A -11.000 A 0.002 % 0.043 %
20 A, 20 A 19.999 A -0.006 % 0.084 %
-20 A, 20 A -20.001 A 0.007 % 0.084 %
Serial number:
5720A
XXXXXXX
Date of Certificate:
Temperature:
05-Mar-2007
23 ± 3°C
specic instrument was calibrated in a man-
Customer Name: Fluke Corporation Relative Humidity: Less than 70% RH ner traceable to national and international
City, State: Everett, WA standards, in accordance with the practices
Procedure Name: 5720A-150
dened in the international standard ISO/
This calibration certificate may contain data that is not covered by the NVLAP Scope of Accreditation. The unaccredited
material, where applicable, is indicated by an asterisk (*) or confined to clearly marked sections. IEC 17025. (Specically, its present version
Measurement uncertainties at the time of test are calculated in accordance with the method described in NIST TN1297, for a is ISO/IEC 17025:2005). The report identies
confidence level of 95% using a coverage factor of approximately k=2.
In the attached measurement results deviation may be expressed with units, Measured Value (MV) – Nominal Value (NV) or as
the instrument tested by model and serial
a proportion of the nominal value ((NV-NV)/NV), expressed as percent. number. The testing used Fluke’s applicable
This calibration certificate applies only to the item identified and shall not be reproduced other than in full, without specific written
approval by Fluke Corporation. The user is obliged to have the object recalibrated at appropriate intervals. This certificate shall procedures. Details o the individual tests
not be used to claim product endorsement by NVLAP or any agency of the U.S. Government. Calibration certificates without
signatures are not valid. are provided with a variety o supporting
Fluke's calibration system complies with the requirements of ANSI/NCSL Z540A.1-1994 (R2002) and ISO/IEC 17025:2005. inormation. These details may include some
or all o, but are not limited to, the specic
test points, measured values, the appropriate
specication limits, measurement uncertainty,
etc. Also included are specic testing dates,
Technician Factory Annex Deputy
specic testing instrumentation details, trace-
ability inormation, and reers to the 17025
accreditation body with accreditation details.
An example o this document is included in
Figure 5.
Page 1 of 5
Fluke Corporation 425.446.6378 Fax 425.446.5649 www.fluke.com
Certificate Number: Sample Date of Certificate: 05-Mar-2007 Page 2 of 5 Certificate Number: Sample Date of Certificate: 05-Mar-2007 Page 3 of 5
Traceability Information Standards Used
DN46
DC Voltage DN47
This calibration was conducted using an unbroken chain of standards to the Fluke Volt, which is traceable to the U.S.
representation of the volt, through the internationally accepted value of the Josephson constant Kj=483597.9 GHz/V and a
10 Volt Josephson Array Voltage Standard. DC Voltage
Output/Range Measurement Deviation Uncertainty 24 Hour Spec
Frequency 0 V, 220 mV -0.30 uV -0.30 uV 0.21 uV 0.50 uV
This calibration was conducted using an unbroken chain of standards to a GPS disciplined oscillator frequency standard, 0 V, 2.2 V -0.11 uV -0.11 uV 0.16 uV 0.80 uV
traceable to the United States Naval Observatory (USNO), which is traceable to the National Institute of Standards and 0 V, 11 V -0.9 uV -0.9 uV 0.6 uV 3.0 µV
Technology (NIST). 0 V, 22 V -0.6 uV -0.6 uV 1.0 µV 5.0 µV
0 V, 220 V -21 µV -21 µV 16 µV 50 µV
AC Voltage (1 MHz and below), Resistance, DC Current, AC Current (10 mA and above), Capacitance, Inductance and 100 mV, 220 mV 99.9996 mV -0.0004 % 0.0005 % 0.0010 %
Phase -100 mV, 220 mV -99.9998 mV -0.0002 % 0.0007 % 0.0010 %
This calibration was conducted using an unbroken chain of standards which are traceable to NIST. 1 V, 2.2 V 0.9999994 V -0.0000636 % 0.0001207 % 0.0004300 %
-1 V, 2.2 V -1.0000005 V 0.0000473 % 0.0001200 % 0.0004300 %
AC Voltage ( > 1 MHz) and RF Power 10 V, 11 V 9.999998 V -0.000023 % 0.000070 % 0.000280 %
This calibration was conducted using an unbroken chain of standards which are traceable to NIST or the National Research -10 V, 11 V -9.999997 V -0.000030 % 0.000070 % 0.000280 %
Council Canada (NRC). 10 V, 22 V 10.000001 V 0.000013 % 0.000070 % 0.000300 %
-10 V, 22 V -10.000000 V 0.0000 05 % 0.000070 % 0.000300 %
AC Current ( < 10 mA ) 100 V, 220 V 99.99999 V -0.00001 % 0.00010 % 0.00040 %
This calibration was conducted using an unbroken chain of standards which are traceable to Physikalisch Technische -100 V, 220 V -100.00001 V 0.00001 % 0.00010 % 0.00040 %
Bundesanstalt (PTB) (German National Metrology Institute). 1 kV, 1.1 kV 1.0000010 kV 0.0000997 % 0.0001400 % 0.0005500 %
-1 kV, 1.1 kV -1.0000008 kV 0.0000842 % 0.0001400 % 0.0005500 %
Temperature
This calibration was conducted using an unbroken chain of standards to the Hart Scientific Metrology Laboratory, which is Resistance
traceable to NIST and/or to Hart maintained intrinsic standards. Output Measurement Deviation Uncertainty 24 Hour Spec
0 Ω -8 µΩ -8 µΩ 10 µΩ 50 µΩ
Humidity 999.890 m Ω 999.874 m Ω -0.002 % 0.002 % 0.009 %
This calibration was conducted using an unbroken chain of standards traceable to the Fluke Everett Service Center, whose 1.89981 Ω 1.89983 Ω 0.00109 % 0.00202 % 0.00850 %
traceability is based on the physical phenomena in which the equilibrium relative humidity values associated with certain 9.99963 Ω 9.99957 Ω -0.00059 % 0.00100 % 0.00230 %
18.99974 Ω 18.99963 Ω -0.00058 % 0.00085 % 0.00230 %
saturated salt solutions are known.
100.0003 Ω 100.0005 Ω 0.0001 % 0.0003 % 0.0010 %
190.0010 Ω 190.0009 Ω -0.0001 % 0.0003 % 0.0010 %
1.0000400 k Ω 1.0000404 k Ω 0.0000427 % 0.0003132 % 0.000 8000 %
1.900061 k Ω 1.900065 k Ω 0.000231 % 0.000418 % 0.000800 %
9.999987 k Ω 9.999985 k Ω -0.000020 % 0.000208 % 0.000800 %
19.00080 k Ω 19.00080 k Ω 0.00003 % 0.00025 % 0.00090 %
99.99946 k Ω 99.99955 k Ω 0.00009 % 0.00035 % 0.00090 %
189.9923 k Ω 189.9925 k Ω 0.0001 % 0.0003 % 0.0009 %
999.985 k Ω 999.987 k Ω 0.000 % 0.000 % 0.002 %
1.899927 M Ω 1.899927 M Ω 0.000013 % 0.000477 % 0.001700 %
9.99904 M Ω 9.99904 M Ω 0.00005 % 0.00102 % 0.00330 %
18.99941 M Ω 18.99925 M Ω -0.00082 % 0.00105 % 0.00430 %
100.010 M Ω 100.009 M Ω -0.001 % 0.006 % 0.010 %
AC Voltage
Output/Frequency Measurement Deviation Uncertainty 24 Hour Spec
1.9 mV, 1 kHz 1.8993 mV -0.0367 % 0.0964 % 0.2716 %
1.9 mV, 20 kHz 1.8998 mV -0.0091 % 0.1230 % 0.2716 %
19 mV, 40 Hz 18.9972 mV -0.0149 % 0.0090 % 0.0348 %
19 mV, 1 kHz 18.9974 mV -0.0139 % 0.0090 % 0.0348 %
19 mV, 20 kHz 18.9990 mV -0.0051 % 0.0090 % 0.0348 %
19 mV, 100 kHz 18.999 mV -0.005 % 0.028 % 0.082 %
19 mV, 300 kHz 18.996 mV -0.020 % 0.042 % 0.163 %
19 mV, 1 MHz 18.954 mV -0.244 % 0.110 % 0.422 %
190 mV, 40 Hz 190.003 mV 0.002 % 0.003 % 0.013 %
190 mV, 1 kHz 189.999 mV -0.001 % 0.006 % 0.013 %
190 mV, 20 kHz 190.000 mV 0.000 % 0.008 % 0.013 %
190 mV, 100 kHz 190.00 mV 0.00 % 0.01 % 0.06 %
190 mV, 300 kHz 190.00 mV 0.00 % 0.02 % 0.10 %
190 mV, 1 MHz 189.97 mV -0.02 % 0.07 % 0.30 %
600 mV, 40 Hz 599.995 mV -0.001 % 0.003 % 0.006 %
600 mV, 1 kHz 599.994 mV -0.001 % 0.002 % 0.006 %
600 mV, 20 kHz 600.002 mV 0.000 % 0.002 % 0.006 %
600 mV, 100 kHz 600.03 mV 0.01 % 0.01 % 0.02 %
600 mV, 300 kHz 599.95 mV -0.01 % 0.01 % 0.05 %
600 mV, 1 MHz 599.4 mV -0.1 % 0.1 % 0.2 %
1 V, 40 Hz 1.00001 V 0.00057 % 0.00200 % 0.00550 %
Fluke Corporation 425.446.6378 Fax 425.446.5649 www.fluke.com Fluke Corporation 425.446.6378 Fax 425.446.5649 www.fluke.com
What are “as ound” data and “as let” data? What calibration certifcate does Fluke rec-
Several dierent measurement test data types can ommend?
be ound on certicates. They can be interpreted as: As a general practice, Fluke does not recommend
one certicate over another. It is up to the us er o
• Asfound.Calibrationdatacollectedbeforethe
the instrument to determine what is appropriate
unit is adjusted and/or repaired.
or the quality and metrology processes which
• Asleft.Calibrationdatacollectedaftertheunitis support their product or service. It is important to
adjusted and/or repaired. be consistent. For example, a laboratory which
• Asfound/asleft.Calibrationdatacollectedwith- supplies 17025 accredited calibrations to their
out any adjustment and/or repair perormed. customers should obtain similar 17025 accredited
calibrations or their instruments.
It is important or instruments that are being rou- I there is some doubt on what type o calibra-
tinely recertied, or certied ollowing a repair, to tion report is needed, it is good practice to obtain
have the as ound data documented as well the as the more ormal 17025 calibration certicate over
let data documented. In this way, the complete less stringent calibration certicate alternatives.
perormance prole o the calibrated instrument is This ensures the highest quality and best situation
known, and any uture corrective actions can be to minimize an organization’s risk against providing
more easily taken. Both as ound/as let data are unreliable test results, services and/or products or
provided with 17025 certicates, but it is not neces- their end customer’s expectations.
sarily included in Z540 certicates unless requested.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more inormation call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Arica +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fuke.com
©2007 Fluke Corporation. All rights reserved.
Printed in U.S.A. 6/2007 3029357 D-EN-N Rev A
8 Fluke Corporation Fluke Calibration Certicates Pub_ID: 11249-eng Rev 01