Design Controls For The Medical Device - Industry by MARIE

Download as pdf or txt
Download as pdf or txt
You are on page 1of 253

TLFeBOOK

Design Controls for


the Medical
Device Industry
Marie B. Teixeira
QA/RA Compliance Connection, Inc.
Odessa, Florida, U.S.A.

Richard Bradley
The Global Consulting Network, Inc.
Palm Harbor, Florida, U.S.A.

Marcel Dekker, Inc. New York • Basel


TM

Copyright © 2002 by Marcel Dekker, Inc. All Rights Reserved.

TLFeBOOK
ISBN: 0-8247-0830-X

This book is printed on acid-free paper.

Headquarters
Marcel Dekker, Inc.
270 Madison Avenue, New York, NY 10016
tel: 212-696-9000; fax: 212-685-4540

Eastern Hemisphere Distribution


Marcel Dekker AG
Hutgasse 4, Postfach 812, CH-4001 Basel, Switzerland
tel: 41-61-260-6300; fax: 41-61-260-6333

World Wide Web


http:/ /www.dekker.com

The publisher offers discounts on this book when ordered in bulk quantities.
For more information, write to Special Sales/Professional Marketing at the
headquarters address above.

Copyright  2003 by Marcel Dekker, Inc. All Rights Reserved.

Neither this book nor any part may be reproduced or transmitted in any
form or by any means, electronic or mechanical, including photocopying, mi-
crofilming, and recording, or by any information storage and retrieval sys-
tem, without permission in writing from the publisher.

Current printing (last digit):


10 9 8 7 6 5 4 3 2 1

PRINTED IN THE UNITED STATES OF AMERICA

TLFeBOOK
Preface

Just what is “design control”, and how did it become part of the
vocabulary of the medical device industry? Simply put, design
control may be thought of as a system of checks and balances
that ensure that the product being developed will meet the
important and essential requirements of the company devel-
oping the product. These important and essential require-
ments include ensuring that the product meets the needs of
the end user and is safe and effective for that use. A detailed
design control process was mandated for the medical device
industry in the United States on June 1, 1997, when the U.S.
Food and Drug Administration (FDA) mandated the design

iii

TLFeBOOK
iv Preface

control process as part of the Quality System Regulation


(QSR) for certain classes of medical devices.1
Before the FDA implementation, the words “design con-
trol” had provoked an emotional response from many people
in the industry. People in the medical device industry viewed
the concept as one more obstacle in a long list of unnecessary
burdens being imposed by Washington bureaucrats. This sen-
timent was voiced most often in the small and early-stage
medical product companies. That is not to say that the larger,
and more established, medical device manufacturers were
pleased with the new proposal; they simply thought of the bur-
den in a different way. To many of these larger companies, the
design control regulation was anticipated as higher overhead
and subsequent lower profits. Now, several years after the ini-
tial implementation of the requirement, we need to examine
whether every company affected has implemented an effective
design control system and is comfortable and happy with the
concept mandated by the FDA. Implemented? Probably. Are
they comfortable and happy? Not likely!
In an initial survey taken between June 1997 and 1998,
of 582 device manufacturers, the Center for Devices and Ra-
diological Health (CDRH) discovered that 530 companies had
implemented design control procedures. Good news! The bad
news was that the survey indicated that 396 of these compa-
nies had potential design control deviations.2 That means al-
most 75% of the companies surveyed had potential problems.
What do these numbers really say? Are they just pointing out
“learning curve” problems associated with the implementation
of a new system in the industry? The answer is not clear. The

1
Preproduction design controls were added to the Safe Medical Devices Act
in 1990. This act gave the FDA the authority to add preproduction design
controls to the cGMP regulation. This was felt necessary due to findings
that showed a significant proportion, 44%, of device recalls were attributed
to faulty product design. The proportion was even greater for software-
related recalls at 90%.
2
This QSIT study is available at www.fda.gov/cdrh/gmp/validationrequest.
html.

TLFeBOOK
Preface v

high percentage of problems could also have been due to a mis-


understanding by manufacturers on how to comply with the
new regulations. The poor results could even have been the
result of inspectors who didn’t understand what may or may
not have constituted compliance. However the real issue was
that there was a problem.
But is there still a problem? Recently, the FDA conducted
another survey of its QSIT audits.3 Part of the survey addressed
design control. In 63 audits conducted between October 1999
and May 2000, the FDA found that there were 150 design con-
trol deficiencies. The results are better than was initially seen,
but there is still a noncompliance level approaching 30% in the
medical device industry related to design control. There is still
a problem, and there shouldn’t be.
Throughout this book, we approach design control as a
mandated regulation for the medical device industry in the
United States. Companies that develop and manufacture class
III, class II, and certain class I devices must have a design
control program in place. But there is another, and in some
ways more important definition of design control. Simply
stated, it is a process that allows a company to make a high-
quality product, from inception through production, and be
reasonably sure that the resources spent in the development
will result in a product that works, is useful, and is what the
customer needs. What company wouldn’t want to do that?
Sure, lightning strikes every once in a while, but most of the
very successful companies throughout the world use some
form of design control to increase the odds favoring success
regardless of what industry they happen to be in. This book
takes those successful product development models and super-
imposes the FDA requirements on them to provide a blueprint
for an effective and compliant design control system.

Marie B. Teixeira
Richard Bradley

3
The Gray Sheet. May 10, 1999, p. 4, and June 12, 2000, p. 10.

TLFeBOOK
TLFeBOOK
Contents

Preface iii

1 Overview 1
An Idea Is Born 1
Ask the Customer 2
Design Controls and the FDA 3
Design Controls and Reality 4
Design Controls and the Bottom Line 5
Design Controls and the Customer 7

2 Design and Development Planning 11


Do We Really Need a Plan? 11
What’s a Plan? 13

vii

TLFeBOOK
viii Contents

The FDA, Design Controls, and Plans 14


Planning Techniques 16
Are Projects Really Always Late and
Overbudget? 19

3 Design Input I 25
Input, Who Needs Input? 25
The Foundation—Design Input 27
The FDA and Design Input 27
The Importance of Design Input and FDA
Requirements 28
The Concept Document 29
Product Performance Specification 31

4 Design Input II 35
To Design Control or Not to Design Control 35
Performance Characteristics 36
Product Characteristics 42
Market Requirements 46
Regulatory and Quality Assurance and
Contractual Requirements 47

5 Design Outputs 49
If Design Controls Are Cyclic, Why Didn’t We
Just Cover Outputs? 49
The FDA and Design Outputs 50
There Must Be an End 51
Design Output Requirements 53
The Device Master Record 55

6 Design Review 59
Not Another Meeting! 59
The FDA and Design Reviews 60
Design Review Requirements 61
Design Review Focus 62

TLFeBOOK
Contents ix

Design Review Stages 63


Meeting Dynamics 66

7 Design Verification 77
What Is Design Verification? 77
The FDA and Design Verification 78
Design Verification Requirements 78
Typical Verification Activities 79
Risk Management 80
Human Factors 85
Risk Management Documentation 86

8 Design Validation 89
Why Validate? 89
The FDA and Validation 90
Design Validation Requirements 90
Typical Validation Activities 92
Risk Analysis . . . One More Time 92
Common Risk Analysis Tools 93
Risk Assessment of Medical Device Materials 99
Biocompatibility 102
Regulatory Aspects of Biocompatibility 104
Phases of Biocompatibility Testing 107
Tests to Demonstrate Biocompatibility 111

9 Process Validation 123


It’s No Good If You Can’t Make It 123
The FDA and Process Validation 124
What Do You Call a Group of Processes? 125
Statistical Process Control 130

10 Design Transfer 137


What Is Design Transfer? 137
Check Your Attitude at the Door 138
The FDA and Design Transfer 140
Design Transfer Requirements 141

TLFeBOOK
x Contents

11 Design Changes 143


The Purpose of Design Change Control 143
The FDA and Design Changes 144
Design Change Requirements 144
The Document Change Request 146

12 The Design History File 147


Why Do We Need a Design History File? 147
The FDA and the Design History File 148
Design History File Requirements 148

13 Questions to Expect in an Audit 151


An Audit! Now What? 151
The FDA Design Controls Inspection Objectives 151
Some Questions You May Be Asked 153

Further Reading 157


Appendix A: Implementation Procedure 161
Appendix B: Concept Document 175
Appendix C: Product Specification 179
Appendix D: Product Claims Sheet 191
Appendix E: Risk Analysis: Standard Operating
Procedure 193
Appendix F: Cause-and-Effects Diagram 205
Appendix G: Validation Procedure 209
Appendix H: Material Specification 221
Appendix I: Quality Specification 225
Appendix J: Design Change Procedure 227
Index 239

TLFeBOOK
1
Overview

AN IDEA IS BORN

If you stop to think about how much it costs to research, de-


velop, and then manufacture a new product from the point at
which somebody says it’s needed to the point when the first
product comes off the production line, you might wish you had
a way to ensure that your new widget was the right one and
it worked the first time. Think about that whole process. There
are a lot of steps, and each step uses your company’s most
valuable resource: your people. New-product development has
a voracious appetite. It consumes people, and people use time
and money. Time and money are two of the things you have
to keep an eye on if you want to make a profit and stay in

TLFeBOOK
2 Teixeira and Bradley

business. An easy way to control this is to do it right the first


time—and that is what design controls can help accomplish.
So what typically happens when a new product is devel-
oped, or for that matter when an old product is improved? In
the ideal world, the customers say, “We want this,” or “We
need that and I’m willing to pay more for it.” If they’re not
telling you that directly, then you need to go out and find out
just what it is your customers really want or need. It’s called
market research and it costs money and it takes time. But if
it’s done correctly, you will know what kind of widget you need
to develop to make a profitable sale in the first place and you
won’t waste time and money developing something that you
know how to do but nobody wants.

ASK THE CUSTOMER

At this point, you still don’t know if you can actually do what
your customers want, but at least you know what you should
do. Sometimes, the whole thing starts differently. Occasion-
ally, an inventor has a great idea. It may be for something
entirely new, or it may be a better way to do something that’s
been done before, in some way that’s better than the old way.
This lone inventor then sets off and begins developing the
product. She risks her own money, time, and other things. One
would think that this inventor might want to check to see
whether anybody else thought the product was a good idea
before moving ahead, but that is not often an important ques-
tion for the entrepreneur*, and it’s only a personal risk. Sup-
pose, however, that your company has a department full of
inventors that you call the Research & Development Depart-
ment, and they come up with this really nifty idea for some-
thing new. Do you go ahead and do it because you know, or
at least think, you can? Do you assume you know what’s best
for your customers, or do you ask them? Although the answer
may seem obvious, it should be stated. You need to ask your

* Real entrepreneurs, please forgive us.

TLFeBOOK
Overview 3

customers what they want, and you need to keep asking them
throughout the development process and, in fact, throughout
the commercial life of the product.

DESIGN CONTROLS AND THE FDA

Additionally, as the widget moves forward from idea to pro-


duction, a system of checks—not only with the customer, but
among the various functions of your company such as sales,
QA/RA, manufacturing, engineering, R&D, marketing, and
finance—needs to be implemented to ensure a high-quality
product that will sell at a profit and that can be manufactured
in a reproducible way. But how is this done, and how is it done
in the medical device industry in a manner that accomplishes
all the goals and complies with FDA regulations at the same
time?
Let’s start with the following question: Which medical de-
vices are required to be developed under design controls? The
FDA is clear:
Each manufacturer of any class III or class II device,
and the class I devices listed in paragraph (a)(2) of this
section, shall establish and maintain procedures to con-
trol the design of the device in order to ensure that speci-
fied design requirements are met.
The following class I devices are subject to design
controls:
i. Devices automated with computer software: and
ii. The devices listed in the chart below:

Section Device

868.6810 Catheter, Tracheobronchial, Suction


878.4460 Glove, Surgeon’s
880.6760 Restraint, Protective
892.5650 System, Applicator, Radionuclide,
Manual
892.5740 Source, Radionuclide, Teletherapy

Source: 21 CFR Part 820, Subpart C, Section 820.30(a).

TLFeBOOK
4 Teixeira and Bradley

Things are pretty simple at this point. Medical device


manufacturers must institute a design control program if the
device being designed is any class II or III product or a class
I device identified above. So all of you making the hundreds
of class I devices that can be found in the medical device
industry can now close this book and move on with the
other important things that you need to do to run your busi-
ness.

DESIGN CONTROLS AND REALITY

Hold on a second! You’re a medical device manufacturer and


you would like to make a profit on what you make and the
new products that you plan on making, right? Who doesn’t?
But you make class I devices and you’re not going to do any-
thing that you don’t have to do. So if the FDA says that the
new wound-dressing product line that you’re developing
doesn’t require the process of design control, why bother?
Here’s why:

• Design controls help to identify what your customers


want and what your competition is doing. They can
even help you identify who your competition really is.
• Design controls identify inconsistencies or discrepan-
cies in what you thought you were making when you
started the development process and it identifies these
problems much earlier in the process, thereby reduc-
ing the amount of redesign and rework and improving
the quality of the product. Remember that the fastest
way to so something right is to do it right the first
time, no matter how long it takes. Besides, a well-
planned design control program doesn’t take all that
much more time. For example, one problem notorious
in all industries, including medical devices, is the in-
vention of a product that cannot be manufactured. Us-
ing design controls can help identify, very early in a

TLFeBOOK
Overview 5

project, any process or manufacturing problems that


will occur with a newly designed product. This could
save thousands of dollars and avoid unexpected delays
in the project’s completion.
• Once you’ve identified a problem, it’s usually easier to
fix it and adjust your limited resources early enough
not to waste time and money.
• If done correctly, design controls should make confor-
mance with all other regulatory requirements easier
and simpler.
• Design controls will also make communication be-
tween departments better and ensure that the right
information gets to the people responsible for other as-
pects of the project and therefore eventually becomes
incorporated into the product that is introduced to the
customer.

Our recommendation is that all medical device product


developments should follow design control procedures. Well,
no, not quite. Some really simple products that are classified
as medical devices wouldn’t gain a thing from applying all the
principles of design control. On the other hand, other class I
medical devices that do not require design control procedures
would benefit from the use of design control procedures. Do
you really want to launch a new wound-care line to compete
with the market leaders like Smith & Nephew, Hollister, or
ConvaTec without knowing whether it works or whether there
are interested customers?

DESIGN CONTROLS AND THE BOTTOM LINE

The main objective of any business is to make money. The


question we need to continually ask ourselves as businesspeo-
ple is whether what we are doing is moving us closer or further
away from that objective. The idea that if you “build a better
mousetrap, the world will beat a path to your door” is dead—

TLFeBOOK
6 Teixeira and Bradley

if, in fact it was ever anything more than a cute saying. De-
signing a new medical device requires design engineers and
chemists to formulate the materials that will be used. Unfor-
tunately, many of these otherwise bright people forget the goal
of the business. They have not been hired to make the longest-
lasting, strongest, most cosmetically pleasing medical device.
They have been hired to make a medical device that is safe
and effective for the application for which it is intended to be
used. More importantly, they have been hired to do this and
generate the most profit.
All those things like comfort, safety, effectiveness, ease
of use, and durability are certainly key elements that will con-
tribute to achieving the ultimate goal, but the prime design
criterion is whether the device will make money—at least if
you buy the idea that being in business has profit as its prime
objective. Meeting this objective may be as simple as answer-
ing the question, “Will anybody buy this?” The product devel-
opment process must also address the production, marketing,
financial, and customer expectations required for the product
in addition to all those things that the product must do to be
safe and effective in its application. The only way to ensure
that all these factors are addressed and that they do not con-
flict is through the creation of some sort of master plan that
ensures that all aspects are being looked at and balanced in
relation to each other: in other words, a design control system.
Regardless of whether the design control process is man-
dated by a government agency, such as it has been with medi-
cal devices, it simply makes good business sense to control
what is a very expensive process. No modern company,
whether large or small, can afford the experiment-till-you-
drop-or-find-an-answer approach made famous by Thomas
Edison. Today’s world simply moves too fast and is too expen-
sive. If your company develops and manufactures medical de-
vices, a design control program needs to be implemented not
just because the FDA has mandated it, but also because there
is really no efficient alternative for directing the process of
product development.
Keep in mind that the design control process does not

TLFeBOOK
Overview 7

apply either to basic research, at least not in the context of


this book, or to feasibility studies. However, once someone has
decided that a particular product or design will move forward
toward production, a design control process must be imple-
mented for medical devices.

DESIGN CONTROLS AND THE CUSTOMER

The design control process is a cyclical system of checks and


balances that starts and ends with the customer. Product de-
velopment should start with the identification of what the cus-
tomer or user wants and what he or she needs. Actually, de-
fining the customer can be far more complicated than it seems.
Is the customer the patient, the nurse, the physician, or the
health-care facility? In many cases, the answer is all of these.
Answering that question correctly is one of the major obstacles
associated with developing a new medical device, or improving
an old one. Like many developments in medical devices, the
answer will likely be a combination, and maybe even a com-
promise, among the many requirements each wants and
needs.
So now we know that if we make this device, then the
nurses, doctors, and patients are going to buy it. Great start.
Now what? Rush in and start doing development work? Not
yet. We need a plan. The plan needs to tell us what steps must
be taken in order to produce a finished product. The plan also
needs to tell us who’s responsible for each step, how much it
will cost, how long it will take, and if we have enough of the
right resources to complete the project. There are many ways
to do this including simple flowcharts.

The First Law of Design Controls:


Document Everything
Now all we need to do to get started is to identify what we
would like the product to be or do. We need to identify all the
requirements that would make this product successful:

TLFeBOOK
8 Teixeira and Bradley

• Who will use it?


• What are the risks?
• What kind of environment will it be used in?
• How much can it cost?
• What are the shapes, the colors, and the sizes?
• What tolerances are acceptable?
• What materials can be used?
• How should it be packaged?
• How will it function?
• How do you make a gazillion of them?
• What are the statutory and regulatory requirements?

These things are the design inputs from all the work that was
completed prior to deciding that the development is past the
exploratory or research phase and now requires design con-
trols. These inputs may, in fact, become outputs of additional
testing and design prototypes based on continued work. Re-
member the cyclical nature of the design control process.
These and other questions are identified as the design inputs.
Which of these inputs are critical and essential, which are only
desires (wish lists), which can be modified, which are incom-
plete or vague, and which are contradictory? All this needs to
be written down and dealt with—it needs to be documented.

The Second Law of Design Controls:


Inputs ⫽ Outputs
Sooner or later we do some lab work and perhaps some clinical
use testing on prototypes of the product and we get results
and measurements from additional testing. These results are
the design outputs. How do they compare with the design in-
puts? Did we meet all the goals? If not, these outputs become
new inputs for subsequent tasks. Outputs also establish the
acceptance criteria by which we can verify and/or validate the
design. More importantly, they become the specifications, en-
gineering drawings, quality test methods, standard operating
procedures (SOPs), and manufacturing process controls of the
commercial medical device. So along the way, the design

TLFeBOOK
Overview 9

team—all the people responsible for finishing the product—


needs to get together on a regular basis and go over what’s
been done, what hasn’t been done, and the results. All the in-
formation available at the time of these design review meet-
ings is reviewed. Any inconsistencies and ambiguous or con-
flicting requirements need to be addressed and resolved. Any
risks need to be identified, assessed, and reduced. If objective
and comprehensive, these design reviews keep everything on
track and prevent wasting too much time and money.

The Third Law of Design Controls:


Trust but Verify and Validate Everything
Before the new product can actually get into production, we
need to be sure that everything done up to this point was cor-
rect and can be reproduced time and again. We need to verify
that the design criteria have been met. This verification pro-
cess should confirm whether what we have accomplished (the
design output) is in fact what we said we wanted to accomplish
(the design input). We need to verify that we have met the
acceptance criteria defined by the test methods and specifica-
tions.
Validation follows successful verification. We need to val-
idate everything to be sure we are making what the customer
asked for under actual (or simulated) production conditions.
This is where we get data that says we have accomplished the
developmental goal. Usually this is done through in-vitro per-
formance, functional testing, and in-vivo clinical evaluations
and trials.*

The Fourth Law of Design Controls:


Transfer Is Inevitable
Once this validation is complete, it’s time for one more step:
the step that often causes headaches. The design and all its

* Comparison with predicate devices (form, design, function, use, etc.) and
showing safety and efficacy are also considered forms of design validation.

TLFeBOOK
10 Teixeira and Bradley

specifications, conditions, SOPs, quality data, and drawings


must be transferred to manufacturing. Manufacturing person-
nel need to be trained in the correct way to make the product,
and everything needs to be documented, not simply because
it’s a requirement but because a miscommunication during the
transfer can be costly and time-consuming. Product coming off
the production line must meet specifications and expectations.
Then, and only then, is it ready to launch.
So that’s it in summary. Design controls are a simple pro-
cess that ensures that what you think you are developing is
what you wanted to develop in the first place and that what
finally comes off the production line meets the customer’s
needs and expectations. The rest of this book details each step.

TLFeBOOK
2
Design and Development Planning

DO WE REALLY NEED A PLAN?

Let’s suppose we want to go to Paris for dinner. We decide that


we really can’t stay in France for too long, so we obviously
need to fly. We head off to the nearest international airport
and buy a ticket, get on a plane, and seven or more hours later
we’re in Paris ready for dinner. If that’s all the planning that
we’ve done, the likelihood that we would be in Paris having
dinner is slim to none. We had the goal in mind; we wanted
to have dinner in Paris. We even had a plan: Fly rather than
travel by any other means of transportation. The result should
have been our arrival in Paris.
The problem with this story was that we didn’t have

11

TLFeBOOK
12 Teixeira and Bradley

enough information to begin with and we really didn’t have a


plan at all. You may think that the little story is a bit oversim-
plified, and it may be. But have you ever heard anybody say
to you that they haven’t got the time to do all that fancy plan-
ning that they do in the larger or more profitable companies?
We’ve all heard it. It’s usually followed by, “Besides, I can get
the whole thing down in a few sentences.”
Let’s take a closer look at what really has to happen to
get us to Paris for dinner. First of all, when is it we want to
be there? Today? Tomorrow? Next Thursday? Easy question.
We want to go tomorrow. How long are we going to stay? Now
it begins to get a little more complicated. So we think about
it a little and decide that we really just want to go for dinner
and get back in time to watch our favorite TV show at 9 p.m.
EST. That makes it a lot more difficult. But the time zones
are working with us, at least on the outbound flight, and
there’s always the Concorde. The result now affects several
other things. Does the Concorde still fly? If it doesn’t, then the
desired conclusion is impossible. Even if it does fly, what’s the
schedule? It then also becomes apparent that we just can’t
drive to any airport, but we need to find one of the few that
the Concorde services.
So we call Air France and find out that we can get to Paris
for dinner but there is no flight returning late enough (Paris
time) to get us home for the TV show. We’d have to stay over-
night and the best we could do is be back by noon the next
day. Assuming that we have already decided that the expense
of the Concorde is worth it, we now have to decide whether
the added expense of a Paris hotel and other meals in Paris
are affordable. Paris at night is beautiful. It’s worth it. So we
book the tickets. Now we have to figure out how to get to the
airport from which the Concorde leaves. Not big problems, no
need to discuss it; we just take care of it. Next morning we’re
off and are adjusting the seatbacks and tray tables for the
takeoff. We’re on the Concorde headed for Paris. In a few hours
we’ll be at the Paris airport, then to the hotel, where we’ll
freshen up and then head off to dinner.

TLFeBOOK
Design and Development Planning 13

Did you know that when the plane takes off and heads
toward Paris, it will be off-course most of the time? If some-
body doesn’t do something, there is no way to guarantee that
when we’re ready to land we’ll be anywhere near Paris. The
winds encountered along the way will be enough to make a
difference. But technology is wonderful; with the help of on-
board computers, course adjustments are continually made
and we arrive at our intended destination. Dinner was won-
derful. Paris was charming. The decision to stay overnight
turned out to be the best thing that could have happened.
We could have said that the first thing that needs to be
done in an effective design control program is to plan. Every-
body knows that, but not everybody does it. There is absolutely
no way to complete a complicated project without a plan. It
doesn’t matter how long that planning process takes. It must
be done. Just because someone can articulate the product
doesn’t tell you how to develop that product for manufacture
and sale. If you think about the beginning of the Paris story,
it may not even tell you what the product is! Remember the
goal at first was dinner in Paris. The actual “product” ended
up being dinner and an overnight stay.

WHAT’S A PLAN?

You can’t just walk into the room where you keep your techni-
cal folks and say, “I want a new, fully functional, implantable,
small intestine prosthesis.” You can’t wander on down the hall
to the marketing folks and tell them to get ready for launch
in 24 months, and then stop by the Regulatory office and tell
them to file a 510(k).* We all know better: There are hundreds
of things that need to happen before a product is ready for a
launch. But that’s why we pay all these technical, marketing,
and regulatory people; they know what they have to do. But

* Everybody wants a 510(k); anything else is too much work and takes too
long!

TLFeBOOK
14 Teixeira and Bradley

what they don’t know are the details; they don’t know what is
happening somewhere else. And every time someone changes
something, then everyone else has to adjust to what he or she
is doing. Remember the wind on the way to Paris?
So, what should be in a typical plan? Plan elements vary,
but in general, a plan should contain

• The goals and objectives: Define the product that is


being developed.
• The organizational responsibilities: What are the inter-
faces between departments? How are they interrelated?
• The tasks: What are the major tasks? What tasks de-
pend on others before they can happen?
• The resources: What will it cost? Do we have the re-
sources (money, people, time)? If we hire two more en-
gineers, can we do it faster?
• The time schedule: What tasks should start first? How
long does each task take? What tasks can be done con-
currently? Which tasks, if late, affect the outcome of
the entire project (what are the critical tasks?)?
• The milestones: When do we get together to find out
if there has been any progress? Are there problems?
Do changes need to be made? Do major decisions need
to be made? When will it end?
• The communication: How do we tell everyone who
needs to know what just happened? When should ma-
jor reports be issued? When do change notifications
occur?

THE FDA, DESIGN CONTROLS, AND PLANS

Those are the typical elements of a project plan. There should


be no surprises. So what does the FDA say it wants? The FDA
says

Each manufacturer shall establish and maintain plans


that describe or reference the design and development ac-

TLFeBOOK
Design and Development Planning 15

tivities and define responsibility for implementation. The


plans shall identify and describe the interface with differ-
ent groups or activities that provide, or result in, input to
the design and development process. The plans shall be
reviewed, updated, and approved as design and develop-
ment evolves. [21CFR Part 820, Subpart C, Section
820.30(b)]

The FDA’s wording is deceptively simple. The agency


says you must have a plan that describes the steps of the de-
sign and development being undertaken. The first sentence of
that section also says that the plan must identify the person
or persons responsible for implementing the tasks. Somebody
must be responsible; that’s good business sense. In fact, some-
body must be in charge of each step. The FDA doesn’t preclude
one-person companies, but this does imply a level of expertise
is expected for most of the functions that are parts of a typical
design and development project. Most of us wouldn’t go to a
bagel shop to buy an emerald ring.
The next sentence builds on this idea. The FDA requires
that we identify those persons or groups that will provide
the output. It doesn’t matter whether those resources are
employees, consultants, or other companies. Under design
controls they must be identified. For example, if internal re-
sources cannot tell you if the product you are making is sterile
(assuming it has to be), then we need to find someone or some-
place qualified to provide that testing. But it goes one step
further; it asks that we show how all these resources will in-
teract. Design controls require that we think about and iden-
tify how each group, sometimes working independently in
their own disciplines, will be sure that what they are doing
will be integrated into the tasks of what other resources are
doing. The output from one group is the input for one or more
other resources.
It seems obvious that different groups of people working
on the same project should know what each other is doing.
Everybody knows that and, for the most part, everybody does
it. It’s done most often with the big things, but we sometimes

TLFeBOOK
16 Teixeira and Bradley

miss the smaller, subtle things, and most often we miss the
soft requirements: the expectations and the wishes. In our ex-
perience, as both consultants and employees, the fact that one
group may not know what another department or group ex-
pects is usually not due to anything sinister, but is most often
due to poor communication. The folks in marketing know that
the customer wants this product to be soft to the touch and
have a low profile so it will be unobtrusive when worn. They’ve
talked to users, they’ve run focus groups, and they made sure
it was in the product development goals right from the begin-
ning. The design folks knew about it, too. So why isn’t the
product being developed soft enough or unobtrusive enough
for the marketing group?
Part of the reason may be a poorly written specification.
Does “unobtrusive” mean a profile that’s 3 cm high or 0.3 cm
high? If everyone knew that it meant 0.3 cm, for example, did
the design folks tell marketing (and the rest of the team for
that matter) that some of the other critical objectives couldn’t
be met at a profile that thin? Lots of these small misunder-
standings occur. Remember the greatest problem in communi-
cation is the illusion that it’s been achieved.
So common sense and the FDA says that, “The plans shall
be reviewed, updated, and approved as design and develop-
ment evolves.” Probably the best way to make sure this hap-
pens is to schedule these reviews on a frequent and regular
basis. These meetings do more than just force a review of the
project plan. They force the procrastinators to update and doc-
ument what they’ve done, what they’ve discovered, and what
needs to be changed and modified. It also forces other func-
tional groups to respond to these changes and the person re-
sponsible to approve any changes or initiate a new task to re-
solve the fact that the input no longer quite equals the output.

PLANNING TECHNIQUES

It is beyond the scope of this book to provide an in-depth study


of planning, or even some of the planning methods. But for

TLFeBOOK
Design and Development Planning 17

Figure 1 The action list.

technical projects several techniques have been developed and


work well. They are summarized here.

The Action List


This can be made to work reasonably well for straightforward,
simple product development. It is exactly what it sounds
like—the good old “to-do” list. For small organizations with a
simple project it can be argued that the list is sufficient. It
has several shortcomings; one is that it is time-consuming to
update especially when a task needs to be inserted. Figure 1
shows an action list and some of the things it should contain.
If this simple method is used to track a project, then proj-
ect costs and interactions between people or other resources
need to be documented separately. But perhaps the greatest
problem with this method is that it gives no indication of
which task should happen first or which tasks are dependent
on others, and it has no predictive value. It can’t tell you, once
the work begins, how long it will take for the goals to be
reached.

PERT
PERT is the acronym used for the planning method developed
in the aircraft industry—program evaluation and review tech-
nique. It can be an extremely powerful and complex method.
When it was first introduced, its power was simultaneously
its most onerous characteristic. Each task is placed in a box;

TLFeBOOK
18 Teixeira and Bradley

inside the box is typically the starting date, the duration, and
the completion date, both scheduled and actual. Other infor-
mation is also included: things like the person (or group) re-
sponsible for completing the task and other resources that will
be needed (including costs). The boxes are then arranged so
that those that can be started simultaneously (and do not re-
quire a task to be completed before they can begin) are placed
first. Other dependent tasks (those that require one of these
precedent tasks to be finished) are placed on the table, and
lines are drawn between the different boxes to indicate their
relationships. After this is all laid out, it is relatively simple to
find the longest timeline ending with the overall goal, thereby
giving an estimate of the project length (and cost if you add
up the dollars). This is defined as the critical path. It shows
the tasks that need to be completed on time in order for the
estimated project completion date to remain as that originally
calculated. It is critical from a time point of view only. Some
of the tasks on this critical path may be relatively trivial from
other points of reference. Before the nice people at Microsoft,
Symantec, and several other software companies large and
small programmed software to automate all this, it should be
apparent how difficult it was to record a change, especially if
that change caused a shift in the critical path or the change
required new tasks with different links.
Using one of these programs* makes the most tedious
part of the plan exercise that of doing the initial plan; changes
are relatively simple and automatic. These programs also
allow you to add different and more detailed bits of informa-
tion not envisioned in the original technique such as accruing
costs, visually showing which tasks have slack time,† and lev-
eling resources. Resource leveling (which could be done by old-
fashioned brain power) allows the program to calculate the

* Microsoft Project  or Symantec’s Time Line  are examples.


† The time a task can be delayed from the scheduled start time and still
have no effect on the entire project.

TLFeBOOK
Design and Development Planning 19

amount of additional time it will take in order to complete


tasks and the plan because one or more resources (people)
have been overallocated.* Figure 2 shows the elements of a
typical PERT chart.
One thing that these programs have blurred is the dis-
tinction between PERT diagrams and Gantt charts. You can
switch from one to the other with the press of a button, and
constructing one type automatically builds the other.

Gantt Charts
A Gantt chart is a graphical representation of all the tasks
and milestones necessary to complete a given plan or project.
The most commonly used graphical representation is a bar
graph. Each task is described in a column along the left side
(the y-axis) and the dates (days, months, years) from the x-
axis of the chart. The bar associated with each task is propor-
tional to the duration of the task and begins on the start date
and ends on the estimated completion date. If something is
late (or early), everything slides appropriately. As with PERT
diagrams, Gantt charts can be constructed to show the critical
path.
Most programs also plot a bar within a bar. The inside
bar grows as the percentage completion of a task moves from
0% to 100%. Figure 3 shows the typical elements of a Gantt
chart.

ARE PROJECTS REALLY ALWAYS LATE


AND OVERBUDGET?

It seems like new-product development projects are almost al-


ways finished later, and cost more, than everyone thought
they would at first. There are dozens of legitimate reasons why

* If your resident genius has been scheduled for three eight-hour tasks on
the same day, she’s been overallocated.

TLFeBOOK
Figure 2 Simple PERT chart.

TLFeBOOK
Figure 3 A simple Gantt chart.

TLFeBOOK
22 Teixeira and Bradley

this could happen. Things change. The product isn’t working


quite the way everybody expected it to work. The competitors
beat us to it and now we have to do something a little different
and a little better. Reimbursement levels have changed. The
product is going to cost us more to manufacture than we origi-
nally thought.
There are probably dozens of reasons that cause a devel-
opment to take longer and cost more than anticipated. There
are several others, however, and those are our fault. We have
formulated these into the following set of rules.

The Teixeira–Bradley Rules for Ensuring a Project


Will Be Completed Later Than Planned

1. If the design team consensus is that the develop-


ment will take 36 months to complete, and you’re
the boss, insist on having it done in 18 months. Time
is money, and you’re the boss; they’ll listen.
2. If you’re a technical type like a chemist or an engi-
neer and the marketing department asks you to de-
velop a product that’s 10 times better than the com-
petition in performance, costs 50% less, and still has
an 85% gross margin, tell them you can do it—espe-
cially if you know it’s impossible, your job is not all
that secure, and you’re supposed to be a genius
anyway.
3. If you’re a marketing type, insist that the testing or
quality types run that 90-day aging study quicker;
show them your technical ability.
4. If you’re a technical type, show them your breadth
of knowledge: Point out that the sales team didn’t
reach the forecast first-year sales for the last prod-
uct until the third year, so why should anybody rush
to market with this one?
5. If you’re a regulatory type, insist on adding another
6 weeks to the schedule to account for the possibility
that the FDA will send the submission document

TLFeBOOK
Design and Development Planning 23

back with a number of obtuse objections. This is


particularly effective if you know it’s the simplest
510(k) the agency will see all year.
6. If you’re a financial type, nitpick every line item in
the proposed budget and hold up the project for
about two weeks—just because you can.
7. If you’re a quality type, insist on a 0.1 AQL for all
final product testing, especially if the tests are for
noncritical cosmetic properties.
8. If you’re the corporate patent attorney, tell everyone
that they can’t do anything in public with the prod-
uct until the patent has been filed and you’ve been
waiting for two months for the engineering depart-
ment to get you some drawings.
9. If you’re the Vice President-in-line-to-be-the-boss,
show them all your concern and agree to head a com-
mittee to decide how long each task will take. Be
sure you mention that the other three members of
the committee will be from a different corporate
business unit.
10. If you’re a human resources type, just smile and tell
them that there is no way you can hire and assimi-
late all those new resources everyone just agreed to
hire to get the project done in the shortest time.

Now, we all know that those tongue-in-cheek rules are


things that never happen in the real world. Sure we do, just
like the fact that the comic strip Dilbert  is funny because the
things it shows never happen either.
One final thought about the planning process and why it
often turns out that the time predictions are always faster
than reality. People always underestimate; somehow they
think it makes them look smarter to say something will hap-
pen faster than they know it probably will.
When assigning the duration times for the tasks, don’t
use the first time that came into your head. Is the time that
you’re thinking of the most probable time, did you account for

TLFeBOOK
24 Teixeira and Bradley

possible unforeseen circumstances or deliveries from suppliers


that might be late? Is anyone going on vacation or taking a
short-term disability leave?
Try getting three different times for each task:

• The worst case, or longest time


• The most probable time, if almost everything happens
correctly
• The best case, or shortest time, if everything goes
right the first time

From these three numbers use a weighted average, and


plug that in as the duration in your project planning—you
might be surprised. Perhaps an average like:

Most probable duration ⫽ 2(worst case time)


⫹ 2(most probable time)
⫹ best case time/5

Oh, and when the president says that the project is too
long—resist cutting the time estimates, at least for a while.*

* You could overestimate to begin with, or perhaps weight the average in


favor of the most pessimistic time, but no one ever does that!

TLFeBOOK
3
Design Input I

INPUT, WHO NEEDS INPUT?

Never mind who needs input. What is input? Well, for one
thing, one person’s input is another’s output. Always keep in
mind that the design control process is cyclical. Also remem-
ber that design controls begin after the initial exploratory
phases of a project are completed.
Design controls, as mandated by the FDA and as covered
in this book, do not apply to what goes on during research and
during other types of feasibility studies. Before design controls
can take effect in the development of a medical device, some-
one has to declare that the prototype, or the material sample,
is likely to become a product. This declaration is usually based

25

TLFeBOOK
26 Teixeira and Bradley

on feasibility testing, both laboratory and clinical use testing,


that shows the product has some promise in meeting the re-
quirements of the end user.
There also should have been a preliminary market survey
that tells the company the product is worth pursuing. Some-
one should have also made a determination, no matter how
crude, that the product could be made at a cost and sold at a
price (don’t forget reimbursement if it applies) that will make
money. All these things become part of the design input.*
Design input is probably the most critical element of the
design control process and not necessarily just from a regula-
tory point of view. It is certainly important to know that the
medical device being developed is safe for human use and is
effective for the intended use. That is, without a doubt, a regu-
latory issue and it is, in fact, the reason why the Pure Food
and Drug Act created the FDA in the first place. But equally
as important to regulatory issues is the simple fact that clearly
defining design inputs is just sound business.
It is also important to want to minimize the risks where
possible. No legitimate company wants to harm its customers
or make a product that doesn’t work. But there are other risks.
A good business doesn’t want to spend more money than it
has, or can get, to develop a product that no one will buy. It
is unimaginable that someone would waste time and money
developing a product whose selling price would be 600%
higher than the market will bear (or that reimbursement
will allow). No one wants to develop a product that is as big
as a microwave that should have been as small as a cell phone
to be acceptable to doctors, nurses, and patients.† These
things can be avoided with clear and accurately defined in-
puts.

* Of course, they are all outputs of the feasibility work. Confused yet?
† Yes we are aware that sometimes this is the only way with really new
technology. Do you remember the first portable PCs from Compaq? They
must have weighed 40 pounds and were bigger than many of today’s desktop
PCs.

TLFeBOOK
Design Input I 27

THE FOUNDATION—DESIGN INPUT

Design input is arguably the most important part of the design


control process. It is the foundation for the entire design and
development activity. If the foundation has basic problems,
then the entire structure will be suspect until those problems
are identified and corrected. The inputs are the physical and
performance characteristics and requirements of a device.
They are the basis for the design. By spending the time and
the resources to get these inputs defined accurately, a com-
pany can save an enormous amount of time and money in the
long term.
Once again it is important to keep in mind the cyclic na-
ture of the design control process. Inputs are themselves the
output of previous work; they are not merely pulled out of the
blue. At least they shouldn’t be in an ideal world. The design
inputs that begin a design control process cannot be all wishes
and desires. Remember, the early exploratory work, the re-
search, the feasibility study (whatever you would like to call
it) are not part of design controls as defined by the FDA. At
the point when design controls become mandatory for a medi-
cal device and this needs to be defined, hopefully you will have
the output from a great deal of preliminary work. As a result,
design controls should typically begin with a prototype that
has been defined as likely to become a commercial medical de-
vice. The output of any testing and studies completed prior to
the implementation of the design control process are the de-
sign inputs for the development process.

THE FDA AND DESIGN INPUT

So what does the FDA have to say about this relatively simple
but extraordinarily important concept?

Each manufacturer shall establish and maintain


procedures to ensure that the design requirements re-

TLFeBOOK
28 Teixeira and Bradley

lating to a device are appropriate and address the in-


tended use of the device, including the needs of the user
and the patient. The procedures shall include a mecha-
nism for addressing incomplete, ambiguous, or conflicting
requirements. The design input requirements shall be
documented and shall be reviewed and approved by desig-
nated individual(s). The approval, including the date and
signature of the individual(s) approving the require-
ments, shall be documented. [21 CFR Part 820, Subpart
C, Section 820.30(c)]

This is the first time that the word “procedures” is used


hand in hand with the concept of design controls. As such, look
at a typical design control procedure. Such a standard op-
erating procedure, an example of which can be found in Appen-
dix A, fulfills the requirement to “ensure that the design re-
quirements relating to a device are appropriate and address the
intended use of the device, including the needs of the user and
the patient. The procedures shall include a mechanism for ad-
dressing incomplete, ambiguous, or conflicting requirements.”

THE IMPORTANCE OF DESIGN INPUT


AND FDA REQUIREMENTS

As we have said, design input is the most important element


of design controls. This importance is emphasized in the proce-
dure. In addition, for complex developments, the design input
phase may account for 30% or more of the entire development
project! Nonetheless, some of the concepts bear repeating for
emphasis in nonregulatory language.

1. Design input is the starting point and the founda-


tion of a successful product design and development.
2. Because they are so important, design inputs need
to be realistic. The critical requirements must be
identified in relation to the customer (the patient
and/or the user) as well as the intended use of the
device.

TLFeBOOK
Design Input I 29

3. The design requirements may be internally or exter-


nally imposed.
4. Design inputs include the product specifications:
• performance characteristics
• product description
• safety and reliability requirements
5. Applicable statutory and regulatory requirements:
• FDA QSR
• ISO 9001
6. Contractual requirements:
• special packaging
• special storage
• special handling and delivery
7. Environmental requirements and limitations:
• energy—electrical, heat
• biological—toxicity, biocompatibility
• electromagnetic interference
• electrostatic discharge
8. Human factors—physical characteristics and con-
straints:
• intended use
• ergonomics and ease of use
• labeling
• packaging
9. Design inputs need to be documented, assigned re-
sources, reviewed and approved.
10. Ambiguous or conflicting requirements need to be
resolved.
11. Finally, a designated individual(s) needs to be given
the authority to sign for “approval” of the design in-
puts.

THE CONCEPT DOCUMENT

The concept document, sometimes referred to as the product


initiation request (PIR), as it is in the example procedure in

TLFeBOOK
30 Teixeira and Bradley

Appendix A, is the first step along the way to an effective de-


sign control process. A model concept document is shown in
Appendix B. The PIR begins to define the requirements, of the
product that is, or is about to be, developed. By its very nature
as a starting point it is often not comprehensive; however, it
should be what its name implies—a written document. It is
not a verbal agreement among a few folks to go off and develop
a new medical device. In fact, even the lone inventor would
benefit from producing a concept document; it would help him
or her to begin to solidify that “light bulb” that went off in his
or her head the day before.
Generally, the concept document is qualitative in nature,
especially when being used to document a new product or ap-
plication for which little is known, and when the product being
developed is “new” for the company undertaking the develop-
ment. It can, however, contain any known quantitative infor-
mation, but that is typically left for the product performance
specification (PPS) document.
In an ideal world, the marketing department of a com-
pany prepares the concept document. In an ideal world, this
same marketing department is in touch with the market it
serves, making this a good place to start. However, it can be
initiated by anyone from any discipline; not all R&D folks and
engineers are clueless.
The concept document’s purpose is to broadly define the
requirements of a new-product idea so that the review of
whether or not that idea should be pursued can be discussed
and approved by other personnel in the company. Several ele-
ments should be included in the concept document:

• A statement of purpose. Why would we want to de-


velop this product . . . is there an opportunity? How
big is the opportunity? What are we going to do?
• A statement of the market position. How is this prod-
uct going to compete? Where? Against whom?
• A statement of essential or desirable characteristic.

TLFeBOOK
Design Input I 31

What does the product do? What does the product need
to do to be successful? What does it look like?
• A statement of the intended claims. What indications
will this product be suitable for? Are there any limita-
tions or exclusions (in demographics, for example)?
• A statement on suitable packaging. Do the intended
users require special packaging for ease of opening?
Does the product design require specific packaging to
ensure stability?
• A statement on the clinical and technical require-
ments. What is the product intended to treat or man-
age? How is the product envisioned to provide the
treatment or management? How does it differ from
other similar products? How is it the same?
• What should the product cost? What does it need to
cost? Will it be reimbursed?

Once this concept document is completed, and approved


to move forward as an “official” development project, then a
design team, composed of personnel representing different
disciplines, should be assembled to prepare the product perfor-
mance specification (PPS).

PRODUCT PERFORMANCE SPECIFICATION

The product performance specification is the last output from


the research or feasibility study that was initiated by the con-
cept document. As such this output becomes the initial input
phase for the final development of the product, transfer to
manufacturing, and finally release for sale.
If there has been a considerable length of time between
the preparation of the concept document and the first PPS,
then changes to the initial concept should be documented in
this PPS. Unlike the concept document, the PPS should be
comprehensive. It should also be quantitative as far as is prac-
ticable. The inputs should also be of a nature that not only

TLFeBOOK
32 Teixeira and Bradley

can be (or have been) measured, but also can be verified. A


model PPS is shown in Appendix C.
The elements that might be found in a product perfor-
mance specification are

1. Performance characteristics
• indications for use of the device
• clinical (or use) procedures
• relevant use settings
• nursing home
• acute-care hospital
• home health care
• hospice care
• physician’s office
• medical specialty of the user
• physician (specialty?)
• registered nurse
• medical technician
• layperson
• patient population—inclusion/exclusion data
• clinical outcome analysis
2. Product characteristics
• physical—color, dimensions, shape
• chemical—materials used
• biomedical/biological—maximum duration of
product use in vivo. Reaction considerations?
• environmental—specific storage, packaging,
transportation and use conditions
• sterilization—type of sterilant or sterilization
process appropriate for the device and its package
• packaging—a description of the specific packag-
ing material and configuration that would be ap-
propriate for processing, including sterilization,
and ease of use by the customer
• ergonomics/user interface—include a description
of any ancillary or adjunct equipment or devices

TLFeBOOK
Design Input I 33

that are necessary for the proper use of the device


being developed
• safety and reliability requirements
3. Market requirements
• intended geographical markets
• contractual requirements?
• intended market segments
• claims
• other labeling requirements
4. Regulatory and quality assurance requirements
• relevant regulatory requirements
• standards and test methods

It should be obvious when these design input elements


are reviewed that it is virtually impossible for a single individ-
ual, or a single department within a corporation, to effectively
document the requirements. Ideally, the project team should
have been consulted during the preparation of the concept doc-
ument. Often that is not the case. But for the creation of the
PPS the interdisciplinary team is essential to success. There
are simply too many specific questions that need to be an-
swered that require an expert in the field. If a company does
not have a specific resource in-house, it needs to find an alter-
native. Think of the time and the money that would be wasted
if a product were developed that was revolutionary for the in-
dication but became trapped in the regulatory approval pro-
cess because the developers were unaware of a certain require-
ment.
There are essentially three categories of design input:

1. Functional
2. Performance—both in usage and in the marketplace
3. Interfacial

Because (1) design inputs are so critical to medical device


development and success, and (2) some inputs are overlooked

TLFeBOOK
34 Teixeira and Bradley

or defined qualitatively when a quantitative measure is both


desired and possible, and (3) folks often confuse necessary
and critical requirements with wishes and desires, the next
chapter reviews in more detail each of the typical elements/
inputs.

TLFeBOOK
4
Design Input II

TO DESIGN CONTROL OR NOT TO DESIGN


CONTROL

Before we look at typical design control elements in some de-


tail, it is worth noting that some medical device manufactur-
ers have difficulty determining when the feasibility phase or
the R&D phase of a project ends and the developmental stage
begins.* So perhaps we should ask, “Why does one do ‘re-

* Others seem to think it saves time to hold off the development phase as
long as possible to avoid all the design control rules. This, of course, not only
ignores the regulatory requirements but also makes little sense from the
viewpoint of the effective use of resources.

35

TLFeBOOK
36 Teixeira and Bradley

search’?” Well, the first reason may be to determine the basic


characteristics of a new material or a new concept. Even in
industrial research, there are unknowns and certain funda-
mental facts that need to be studied, quantified, and explained
before anyone should even think about developing a new prod-
uct. Another reason for a research or feasibility stage would
be to decide whether or not a business opportunity even exists.
The research and development of a new product do not include
just the technical stuff. If there is no business, there is no real
product.*
So, we now enter the development stage and need to insti-
tute the design controls process as we make our first proto-
types. Correct? Not necessarily. It may be quite reasonable to
make several prototypes before the actual development begins,
and sometimes before the input requirements are even par-
tially understood. Do not equate prototype design with finished
product design. The early prototypes lack many of the features
that the final product will have. These early prototypes may
not indicate a process by which they can be made; they are
feasibility models. But, when there is enough information to
think that there may be a new product or a new business op-
portunity, the process of design controls should be initiated
by the concept document followed by the product performance
specification.

PERFORMANCE CHARACTERISTICS

Since everything has to start somewhere, let’s assume that


the starting point for the design inputs is defining the perfor-
mance characteristics of the new product being developed.
This does not necessarily have to be the starting point. We
could begin, for example, with the definition of the market and

* Of course, there are exceptions. Everyone knows that the market feasibil-
ity for the first photocopy machines showed a limited market—and now the
world can’t seem to live without photocopies. Stuff happens!

TLFeBOOK
Design Input II 37

the needs of the patient and the end user and be just as effec-
tive.

Indications for Use


In the last chapter we list the kinds of things that help define
the performance characteristics of the product. The first issue
on the list is the “indications for use.”
Let’s use an example of a wound dressing. At this point in
the development cycle, defining the indications must be more
specific than saying something like, “We want to develop a
wound dressing.” It is simply not enough to say what the prod-
uct is supposed to be or what it is supposed to do in such
general terms. That is why the PPS lists this particular per-
formance characteristic as “indications.” Although the de-
velopment of the product from this point might be primarily
a technical and engineering function, others play an impor-
tant role. Defining the indications clearly, concisely, and accu-
rately is a necessary first step in ensuring that the needs of
the market, the patient, the end user, and the regulatory envi-
ronment for which the product is to be introduced are consid-
ered. Different indications may result in different functional
as well as product requirements.
Suppose we say that our wound dressing should be indi-
cated for use on chronic wounds. That’s a beginning, but
should there be more? Will we be marketing a product that is
indicated for all chronic wounds? Will the product be indicated
for leg ulcers or for bedsores, or for both? What about chronic
wounds that are infected? Certainly while we’re defining the
indications for use, we might as well add the contraindications
that are known. Does our research indicate there is any other
type of wound that might be appropriate for the product?
Don’t forget that these are inputs. They should be based
on data that is relatively concrete. It doesn’t mean we still
can’t have a wish list, but if all the fundamental work has
shown that this new wound dressing does not heal wounds
and that nothing that could be added to it would change that,

TLFeBOOK
38 Teixeira and Bradley

then don’t write the indication as “heals chronic wounds.” If


you did, it could set in motion a series of events that will en-
sure a product development that will fall short of its goals and
perhaps even fail completely.
A word of caution, although it should be apparent to ev-
eryone: This section of the PPS defines the indications for use,
not the claims for the product. The definition of the claims
comes in a later section. Also remember that these are the
inputs from the work that has preceded the development or
are the result of the continuing work on products that are al-
ready commercialized and may have received premarket ap-
proval from the FDA.*
For example, if the inputs suggest that a wound dressing
you are already marketing that is approved for chronic
wounds may also be safe and effective in the management of
nonhealing surgical wounds, a new filing to the FDA is neces-
sary for these new indications even though the product is al-
ready “approved.” It tells the regulatory people, early in the
development process, that they will have to prepare the sub-
mission documentation, but only if this new indication is writ-
ten into the PPS. This new indication may also signal someone
in marketing of the need to come up with a new strategy.†

Clinical or Use Procedures


In the procedures section, once again based on the results of
the preliminary or research work, we need to define how the
product is to be used. We should have enough information to
be fairly specific. Things may still change as the development
continues, but at this point we should know how the product

* New indications for an “old” marketed product require a new filing.


† Think about how clever Bristol-Meyers Squibb was when they added the
migraine indication to Anacin. They created a whole “new” product just by
changing the box.

TLFeBOOK
Design Input II 39

is to be used. In fact, it should be possible to actually write a


version of the “instructions for use.”*
Writing a version of the instructions for use at this point
may be the best way to handle this section of the PPS. The
instructions have to be written sooner or later anyway, and
formatting it in this manner allows that all-important docu-
ment to be reviewed, revised, and approved from the begin-
ning.
Whether you decide to actually document this by at-
taching “instructions for use” or by simply listing the neces-
sary conditions of how the product is to be used, remember
that this section has several audiences. It not only tells the
engineers what design parameters and requirements they
need to include in the ongoing design, but it also should be
written with the patient and/or end user in mind. Thus the
instructions should be written in a style and with vocabulary
suited for people who may not be trained in either medicine
or engineering. The instructions should be kept as simple as
possible even if the device is “intended for use” only by physi-
cians. Starting early in the development cycle and constantly
revising with “simple language” as a goal ensure that the in-
structions for use will be clear and concise by the time they
are included in the commercial product.

Relevant Use Settings


Incorrectly defining the “relevant use setting” characteristic
can destroy a product introduction. This particular input will
have some technological and clinical facets to it, but it should
be thought of as a marketing input. The definition may be
worded by the answer to the simple question: “Where will this
product most likely be used?” The answer may be deceptively
simple.

* The instructions are often called the “package insert” by those who began
in the pharmaceutical business.

TLFeBOOK
40 Teixeira and Bradley

We have the inputs of the indications for use. Where is


this indication usually treated or managed? Several answers
may come to mind such as hospitals, home health care, and
chronic-care facilities. Depending on the device being devel-
oped, the answer may be all three. But those are just general
answers and they need to be clarified and classified even fur-
ther. The answer determines the likely total market for this
particular product.
Where the product is to be used also helps to determine
other resources. These are resources that the company may
or may not have. For example, if the company has a sales force
whose traditional focus is the acute-care hospital, and even
more specifically only the radiography department within that
acute-care setting, then it is likely that those existing sales-
people will not be able to effectively sell a new product that
has nursing homes as its primary sales point. This market
mismatch should have been caught during the feasibility
phase of the project. It is a farfetched example used only to
make a point, but similar things do happen.
The dilemma may be subtler. The new product may have
gone through its feasibility phase aimed all the while at the
current market niche of the company. But somewhere along
the way this new device was found to allow “management and
treatment” in an entirely different segment of the medical in-
dustry. This could be viewed as a marketing opportunity. It
could signal the start of a period of fast growth for the com-
pany as it enters a new fertile area. The one thing this new
input should do is signal the fact that the corporation’s re-
sources need to be checked to be sure that they are at least
adequate to perform new tasks in a new area.

Medical Specialty of the User


This should be a relatively easy input to define. Does the prod-
uct require the intervention of a health-care worker, or can it
be used directly by the patient? If the product, because of other

TLFeBOOK
Design Input II 41

characteristics, requires the intervention of a professional


health-care worker, then the input should define this further.
Does it require a physician, and if so should it be a physician
with a particular specialty? Some devices require that the end
user be a registered nurse, or perhaps a trained medical tech-
nician.

Patient Population—Inclusion/Exclusion Data


This is another characteristic that is usually relatively easy
for a design team to define. Most medical devices are designed
to treat or manage a specific indication, and that simulta-
neously defines the patient population. But its simplicity can
allow the input to be misleading. For example, let’s say that
we are developing a device for the management of urinary in-
continence. Is the patient population then all those people that
suffer from any form of urinary incontinence? The answer is
“probably not.” First, is the device for males or females? If we
continue our example by saying it’s a female urinary inconti-
nence device, then we have just cut the overall patient popula-
tion by more than half. If it’s an external female urinary de-
vice, it has probably cut the remaining 40% of the original
total population in half again. Is the device appropriate for
females with stress incontinence? The answer may lead to an
even smaller fraction of the total population.
The other part of defining the patient population charac-
teristic is defining the contraindications. Is there a group of
patients in the suitable population on which the device should
not be used? Is there a component or an ingredient that may
cause an allergic reaction in some people? Are there situations
or conditions or even other devices that interfere with the
proper and safe function of the new device? Does the new de-
vice interfere with other situations in the surrounding envi-
ronment? The answers to these and similar questions will help
define the contraindications and therefore the ultimate pa-
tient population.

TLFeBOOK
42 Teixeira and Bradley

Clinical Outcome Analysis


Clinical evaluations are not required for all medical devices,
but knowing the result of what will happen as the result of the
use of your product is. It seems unbelievable that any company
would allow the continued development and sometimes even
the commercialization of a medical device without any clinical
use testing at all. Because Leonardo DaVinci could envision
a helicopter hundreds of years ago didn’t mean that it would
fly. Dozens of seemingly well-designed products have been re-
designed and improved after simple, but objective, use tests.
Even for straightforward products, a clinical use test is impor-
tant. How else do you know that your product will work under
the conditions it will see in use? Because a competitor’s prod-
uct that looks the same and has the same components has
been working for years? Maybe it will work, but sometimes it
doesn’t.
Depending on whether any use testing was completed in
the feasibility process, this characteristic may define what the
product is envisioned to do and what clinical use testing needs
to be done in order to verify those expectations.
This is the age of cost consciousness in the medical field.
It is no longer enough to be able to manage or treat a given
condition. The manufacturer must also answer “at what cost?”
The answer is especially true when the selling price of the new
product is higher than the currently available technology. Will
it lower overall costs? What will be the outcome of using this
product on the patient, on the problem being managed or
treated, and on the end user’s wallet?

PRODUCT CHARACTERISTICS
Physical Properties
The product’s physical characteristics such as its exact dimen-
sions, shape, and even color should be clearly and accurately
defined. Everything about the physical characteristics should
be clearly defined in the PPS. This includes not only the di-

TLFeBOOK
Design Input II 43

mensions but also the allowable and acceptable tolerances. If


you believe that these dimensions and tolerances are an issue
only for engineering and manufacturing, think again. Ask
someone with a colostomy if a pouch that is dimensioned awk-
wardly to the point where it is obtrusive is a good product.
Don’t forget that this is the section to define the different sizes
the product may need to have.

Chemical Properties
The chemical properties that should be defined in the PPS in-
clude the chemical composition of all the components that
make up the device. Many of today’s medical devices are com-
posed of polymers or blends of polymers, so it is important to
know the nature of any additives that these polymers contain.
Plastics often contain plasticizers such as oils and other mate-
rials that are used to make the materials more flexible and
soft. In some polymers, notably polyvinyl chloride, or PVC,
these plasticizers are often fugitive; that is, they have a ten-
dency to migrate to the surface of the material, where they
can be exposed to the patient directly.
Part of the identification of the materials that compose
the product should also include identification of the interac-
tion of these materials with other materials with which the
device is likely to come into contact during normal usage. For
example, manufacturing surgical gloves from a polymeric ma-
terial that dissolves in alcohol or other solvents found in an
operating-room setting would not be a wise idea. If the proper
functioning of the device requires a material that is not in-
cluded with the device—for example, “wall oxygen”—then
this material should also be listed in the PPS.

Biomedical/Biological
The biomedical and biological properties of the product in-
clude several things. You must clearly define how long the
product is expected to be used. This duration of use may have
a two-level answer. The product may, as an example, be used

TLFeBOOK
44 Teixeira and Bradley

to treat an indication for a 24-hour duration. It may also be


used every 24 hours for 14 days. This 24-hour duration-of-use
product may in fact be used to manage a chronic condition so
that a single device may be used for 24 hours but changed
every day for the rest of the user’s life.
In this section we also need to define which body fluids
and tissues the device (and each of its components) comes into
contact with during normal usage. Is the device intended for
use on intact skin? It is exposed to the dermal layer of the
skin? Is it exposed to bone or internal tissues and organs? The
answers to these questions determine the nature of the toxic-
ity and biocompatibility testing required on the device and its
components.

Environmental
The environmental property definitions include whether or
not all the components of the device are suitable for operation
in the working environment the device will normally encoun-
ter. In addition to that somewhat obvious definition, it is im-
portant to define the conditions for the storage, packaging, and
transportation of the device, and for that matter the raw mate-
rials used to manufacture the device. This will not only aid in
the determination of the shelf stability of the finished product,
but also ensure that raw materials used in the manufacturing
process are fresh and effective.

Sterilization
The type of sterilant or the sterilization process should be
clearly defined along with a definition of what constitutes
“sterile” for the product upon completion of the sterilization
process. It is important that the sterilization method appro-
priate for the device is chosen. Several polymers, both syn-
thetic and natural, may degrade after being exposed to ioniz-
ing radiation. In fact, although for all practical purposes there
is little difference between sterilization by gamma radiation
and electron-beam processing, some materials behave differ-

TLFeBOOK
Design Input II 45

ently due to a dose rate effect* and therefore may limit the
choice of ionizing radiation to one or the other. It would be
impractical and perhaps even dangerous to use an ethylene
oxide sterilization process on a material or a device that is a
gas barrier. Removal of the EtO gas would be a significant
problem. If the device is intended to be resterilized, the rester-
ilization conditions and the number of cycles the device or ma-
terial can withstand must be clearly understood and defined.

Packaging
A description of the specific packaging material and its con-
figuration needs to be defined from several points of view. The
package should appropriately protect the product from the en-
vironment for the duration of its useful life, including the pro-
tection of the product’s sterility. It should also be designed in
such a manner that would be appropriate for cost-effective
manufacturing.
The packaging should also be clearly defined in terms of
its ease of use for the end user or patient. A device intended
for use in a sterile operating room will find few happy users
if the device is difficult to remove from its package when the
user wears surgical gloves. If the device’s user is a paraplegic
or a quadriplegic, then it is important that the package can
be easily opened without necessarily requiring the assistance
of another person.

Ergonomics/User Interface
This section should include a description of any ancillary or
adjunct equipment or devices necessary for the proper use of
the device being developed. This is especially true if they are
not to be packaged with the device.†

* Sometimes there is a difference between delivering 25 kgy at 1 kgy per


hour than delivering all 25 at one time.
† Remember to add this list to the “instructions for use.”

TLFeBOOK
46 Teixeira and Bradley

Safety and Reliability Requirements


This section should specifically describe any conditions that
will affect the safe and reliable use of the product. This may
include electrostatic discharge hazards, specific voltage and
grounding requirements, or protective clothing and equip-
ment that would be recommended for use when the device is
being applied, used, or removed. In addition to requirements
that are related directly to the device itself, it is important to
include other end user actions that may affect their safety.
For example, any device that uses oxygen during its opera-
tions should take into account the standard safe use of the
gas.

MARKET REQUIREMENTS
Intended Geographical Markets
This section includes the definition of the product’s market in
terms of geography. Is the product to be marketed and sold in
the United States, European Community, Pacific Rim, and/or
Asia? The answer will define not only the market size but also
the regulatory requirements that must be met before the prod-
uct can be introduced into a given area. It is important to in-
volve the engineering function in this definition in order to
define cultural differences that may affect the product design
in different regions and countries.

Intended Market Segments


The market segment into which the device is being marketed
should be clearly defined and quantified. This definition
should include unit product forecasts, currency conversions,
and the anticipated growth rate of the specific target mar-
ket(s). For example, a wound dressing is likely to be developed
and targeted at a specific wound type (e.g., surgical incisions),
and perhaps even at only a fraction of that particular segment
(e.g., nonhealing surgical incisions).

TLFeBOOK
Design Input II 47

Claims
This section includes the definition of the specific claims that
will be made about the device; for example, is it a therapeutic
device, a patient management product? The wording of the
claims should be done carefully and accurately represent any
scientific and clinical findings that are known and verifiable
about the device and its use in a specific indication. This sec-
tion should include a review of all the other labeling related
to the product such as the advertisement slicks, web site, post-
ers, videos, and tradeshow-booth signage.
Appendix D provides a useful form for recording the
claims for use with the PPS and for the design review meetings.

Other Labeling Requirements


Define any additional labeling requirements not mentioned in
the other sections such as precautions and warnings. This is
an ideal section for the review of photographs or line art used
to show the application and function of the device being devel-
oped.

REGULATORY AND QUALITY ASSURANCE


AND CONTRACTUAL REQUIREMENTS
Relevant Regulatory Requirements
This section should include all the relevant regulatory re-
quirements that will be necessary before the device can be
launched in the geographical areas noted earlier in the PPS,
such as FDA QSR, EN 46001, ISO 9001, and 93/42/EEC. The
section should also include the relevant standards and test
methods from ASTM, ANSI, and ASQC. The AQL, sampling
plans, and actual product specifications must also be included.

Contractual Requirements
These may include ISO 9000 agreements or even supply and
pricing agreements with distributors, venture partners, and
contract manufacturers.

TLFeBOOK
TLFeBOOK
5
Design Outputs

IF DESIGN CONTROLS ARE CYCLIC, WHY


DIDN’T WE JUST COVER OUTPUTS?

It is true that all outputs follow inputs, which may then lead
to new inputs and so on, but we have not yet really covered
the concept of design outputs. Besides the overall cyclic nature
of design controls, another important concept that should al-
ways be remembered is

Inputs ⫽ Outputs

What exactly does that equality mean? Look at it this


way: Sooner or later the design and development activity re-
lated to a product must come to an end. If that end signaled

49

TLFeBOOK
50 Teixeira and Bradley

success, then all the information and specifications that we


had about the product, that is, everything the product had to
do to be safe and effective and meet customer needs, should
have been designed into the product. Furthermore, if success-
ful, it was then manufactured to meet these specifications
(outputs). More than that, these outputs need to be defined in
such a manner that we can objectively say that they are in
conformance with the inputs, that is, they must be able to be
verified and validated. If these things are not true, then a risk
is associated with the product and the development may not
be completed—and the cycle continues.
Outputs are the work product or the “deliverable” that
defines the essential and proper functioning of the device or
product.

THE FDA AND DESIGN OUTPUTS

Let’s see what the people at the FDA have to say about design
outputs.

Each manufacturer shall establish and maintain


procedures for defining and documenting design output in
terms that allow an adequate evaluation of conformance
to design input requirements. Design output procedures
shall contain or make reference to acceptance criteria and
shall ensure that those design outputs that are essential
to the proper functioning of the device are identified. De-
sign output shall be documented, reviewed, and approved
for release. The approval, including the date and signa-
ture of the individual(s) approving the output, shall be
documented. [21 CFR Part 820, Subpart C, Section
820.30(d)]

There are several important requirements in that rela-


tively small paragraph from the FDA. Among the more impor-
tant are these:

TLFeBOOK
Design Outputs 51

1. The output procedures must define and document the


methods or terms for evaluating outputs and de-
termining their conformance in meeting the input re-
quirements. This should be reviewed as part of the
design review meetings and should be documented
and maintained as part of the design history file
(DHF).
2. The critical and/or essential outputs (specifications,
properties) as well as the acceptance criteria must be
clearly defined. These are actually the requirements
for the device itself and its packaging and labeling.
3. The outputs must be approved prior to release.

The design output is the result of the effort at each design


phase and at the end of the total design effort. The design out-
puts include the tests and procedures that may have been de-
veloped, adapted, or simply used to show conformance with
the defined design inputs.

THERE MUST BE AN END

One particular characteristic of design output implied in its


name is that a design and development program must come
to an end sooner or later. We all assume an eventual successful
conclusion to the design activities. The successful conclusion
expected is that the development project will meet its goals
and that the result is a safe and effective new device. Although
not strictly a topic in a discussion about design controls, end-
ing a project is an important accomplishment. In words, ex-
pressing that successful end is easy. In real life it is not often
clear when a project is at an end, particularly if it has not
reached the anticipated conclusion of a safe and effective prod-
uct that meets all the design specifications. This problem is
not specific to the medical device industry; it happens in any
industry that develops products using an R&D and engi-

TLFeBOOK
52 Teixeira and Bradley

neering department. The fault lies with several (groups of)


people.
In the first instance, it is almost impossible to get a re-
search person or an engineer to admit to either of the fol-
lowing:

1. That something cannot be done with just a little more


work, a little more money, or a little more something.
2. That a product, especially the one being developed,
could not be made a little (or a lot) better with just
more time.

At the risk of irate communications from some readers,


all technical people have made similar statements at least
once in their careers, if not at least once during the course of
every project they have worked on. Why do they do that? The
answer is disarmingly simple. They do it because they really
believe that virtually nothing is impossible and that things
can always be made better. It would be difficult to do the job
of developing in the absence of such a philosophy. They may
know that in some particular case the desired outcome will
not happen, but somewhere in their makeup they believe it
can be done.*
The problem in a finite development situation is that re-
sources are limited. Remember, the goal of any company is
to make money. If a company is wasting money pursuing a
development project that will not be achieved in a finite period
of time or within a defined expenditure, it is not pursuing its
goal. Countless startup companies have failed not just because
they didn’t know what they were doing, but because they pur-

* It is possible to believe that something you want to happen badly enough


will happen, even though at any particular minute in time you know it won’t.
It is the basis of every philosophy in the world. Without a strong belief that
you can accomplish what you have set out to do, or get what you want, then
there is no real motivation to get you through the difficult periods.

TLFeBOOK
Design Outputs 53

sued an elusive dream and exhausted their resources. So what


can be done about the problem?*
First, remember that the people charged with the devel-
opment are focused on the design goals of the project; they
are not focused on the resource allocation problems. Second,
remember that these people are probably pretty bright. So put
them in a smaller box. Remind them about the overall goal.
We are not suggesting that you beat them over the head with
the fact that they’re spending money and time and getting no-
where (although once in a while that couldn’t hurt if done cor-
rectly). What we are suggesting is that you include this “tech-
nical” group in the decision process of resource allocation.
Quantify it. Is there some “rate of return” (ROR) on internal
resources that you use to determine whether too much is being
expended? There better be. If you’re the boss and the ROR is
in your head, share it! You may be pleasantly surprised. If you
put an engineer or a chemist (or any bright person) in a posi-
tion where he or she has to take into account that other things
might be even more successful if resources were adjusted to
accomplish them instead of continuing with the project he or
she is working on, you might get a response that is more in
touch with the reality of the business. This will allow people
to reassess the probability of success of a given developmental
step or, in fact, the whole development project without having
to admit failure.

DESIGN OUTPUT REQUIREMENTS

As we said, design activities must come to an end one way or


another. These design outputs need to be defined in such a
manner that an adequate evaluation of conformance with de-
sign input requirements can be determined. They must be in

* This problem is by no means limited to technical people. Marketers, ad-


ministrators, and professionals of nearly any discipline can be susceptible
to the pursuit of the “white rabbit.”

TLFeBOOK
54 Teixeira and Bradley

a format that can be verified and/or validated. These design


outputs are confirmed as meeting design input requirements
during design verification and validation. They are ensured
during design review.
Design output procedures or specifications need to stipu-
late or refer to acceptance criteria and identify the critical
measures/outputs for the proper functioning of the device.
Critical characteristics may include special handling, storage,
and/or maintenance of the device. Design outputs must also
include specifications for the manufacturing process, assem-
bly drawings, quality assurance procedures and specifications,
device labeling and packaging, and the methods used for con-
trol. Design output documents need to be reviewed and ap-
proved (signature and date of approving individuals) prior to
release. Finished design outputs are documented in the device
master record (DMR).
Outputs typically become inputs for the next design
stage. Although there may not be an output for every input,
there should be an input traceable to each output. Remember
that only approved outputs need to be maintained as part of
the device history file (DHF).

Typical Design Outputs


• Product specifications
• Assembly drawings
• Component and material specifications
• Production and process specifications
• Work instructions
• Quality assurance specifications and procedures
• Packaging and labeling specifications and methods
used
• Software code
• Installation and servicing procedures
• Results of risk analysis
• Biocompatibility results
• Bioburden test results
• Results of verification activities

TLFeBOOK
Design Outputs 55

THE DEVICE MASTER RECORD

The development project output is documented in the device


master record (DMR). This is an extremely important docu-
ment to both the FDA and the company undertaking the devel-
opment. The DMR is a compilation of records that contains
the procedures and specifications for a finished device. Note:
The DMR is not a requirement of a product under development
but is the record of the finished device. People have argued,
based on this one word, “finished,” that the DMR is one of the
last steps in the developmental cycle. The argument has some
validity in that things will constantly change during the devel-
opment process. However, if the DMR material is compiled
during the development process and revised as new informa-
tion becomes available, it is less likely that important informa-
tion will get lost.
In Subpart M on Records (Section 820.181), the FDA de-
fines this important document in more detail.

Each manufacturer shall maintain device master records


(DMRs). Each manufacturer shall ensure that each DMR
is prepared and approved in accordance with Sec. 820.40.
The DMR for each type of device shall include, or refer to
the location of, the following information:

• device specifications including appropriate drawings, com-


position, formulation, component specifications, and soft-
ware specifications
• production process specifications including the appropriate
equipment specifications, production methods, production
procedures, and production environment specifications
• quality assurance procedures and specifications including
acceptance criteria and the quality assurance procedures to
be used
• packaging and labeling specifications, including methods
and processes used
• installation, maintenance, and servicing procedures and
methods

TLFeBOOK
56 Teixeira and Bradley

Think of it this way: If you want to tell someone exactly


(1) what your product is, (2) what it is made of, (3) how to make
your product correctly, (4) what equipment will be needed to
do it, (5) what constitutes acceptable quality and how to test
for that quality level, and (6) even how to maintain and install
the proper equipment, then simply give him or her the DMR
for the product. This makes the DMR one of the most proprie-
tary files in your entire company. It essentially contains every-
thing, even those trade secrets that allow your manufacturing
process to work better than conventional processes. Treat the
DMR with a high degree of confidentiality.
Remember that the DMR is not a single document, like
the PIR and the PPS; it is a compilation of all documents that
relate to the finished product. Take advantage of this and of
the fact that the FDA allows the DMR to refer to the location
of some of the contents of the DMR: Do not keep the entire

Figure 1 A generalized device master record.

TLFeBOOK
Design Outputs 57

record in one discrete file or location. The added work neces-


sary to ensure that revisions to other documents and specifi-
cations are updated in the DMR file is negligible when consid-
ering the damage that can be done by fire, theft, or other
catastrophic loss. Figure 1 depicts a typical device master
record.

TLFeBOOK
TLFeBOOK
6
Design Review

NOT ANOTHER MEETING!

Although a face-to-face meeting is not a requirement, periodic


design reviews are. Face it, this is American business; most of
us wouldn’t know what to do for a living without the occasional
meeting. It is possible to have a design review without having
people in the room. It could all be done, at least theoretically,
by shifting reams of paper back and forth and having every-
body sign off on everything, and you could always do it on the
Internet, but unless at least the majority of these design re-
view meetings are held in person, the real benefit of having
them will be lost.
Remember from the beginning of this book we have men-
tioned two things several times. They are

59

TLFeBOOK
60 Teixeira and Bradley

• In the design controls process what goes around comes


around.
• The development of a product is a team effort of people
from different disciplines.

In order for everyone involved to be doing his or her


“thing” correctly, each person needs to know what everyone
else is doing and has done. They need to hear what is being
said, not just see (or read) it. Design reviews should be con-
ducted with as much personal interaction among team mem-
bers as possible; otherwise things get lost in the “translation.”
By the time I tell another person what happened, add my little
agenda, and she tells the next person, and so on, the original
information is lost or distorted. People need to know as far in
advance as possible what is about to happen so that if it affects
their contribution to the developmental effort they can plan
for the change.

THE FDA AND DESIGN REVIEWS

The FDA is clear and concise in its statement on design re-


views.

Each manufacturer shall establish and maintain


procedures to ensure that formal documented reviews of
the designed results are planned and conducted at appro-
priate stages of the device’s design development. The pro-
cedures shall ensure that participants at each design re-
view include representatives of all functions concerned
with the design phase being reviewed and an individual(s)
who does not have direct responsibility for the design
stage being reviewed, as well as any specialists needed.
The results of a design review, including the identification
of the design, the date, and the individual(s) performing
the review, shall be documented in the design history file
(the DHF). [21 CFR Part 820, Subpart C, Section 820.0(e)]

TLFeBOOK
Design Review 61

Thus, the design review is a documented, comprehensive,


and systematic examination of a product design, or design
phase, in order to evaluate the adequacy of the design require-
ments, to assess the capability of the design to meet those re-
quirements, and to identify problems.
Design reviews should be conducted at strategic points in
the design process, which the FDA implies at the completion of
certain design phases. Each review should primarily provide a
careful assessment of results at that date. It should also pro-
vide feedback and information on existing or emerging prob-
lems related to the product or its development. And, like all
business meetings, the design review should provide and up-
date the team on the project’s progress. It should confirm
readiness to proceed to the next phase or identify the need for
new tasks to be added to an action plan.

DESIGN REVIEW REQUIREMENTS

As the design is developed, it must periodically be reviewed.*


Design reviews should be conducted at major decision points
or milestones in the device’s development cycle. These stages
must be formally defined, and the timing of the design review
should correspond in most cases to completion of the mile-
stones of the development project plan.
The design review meeting should include representa-
tives of all functions concerned with the design stage being
reviewed. This is intended to prevent “fantasy” designs from
entering production or wasting the valuable resources of a
company.
Each design review needs to include an individual who
is independent from the design stage being reviewed. This in-
dependent participant is necessary to bring objectivity to the

* ANSI/ASQC D1160-1995—Formal Design Reviews provides detailed


guidance for conducting formal design reviews.

TLFeBOOK
62 Teixeira and Bradley

process that might otherwise be missing. In addition, other


specialists in the topic being reviewed should also be present.
This design review is not merely a management meeting. The
specialist may have no particular management responsibilit-
ies within the company but could be a leading expert in steril-
ization, for example, and his or her participation could be in-
valuable. Design reviews must be comprehensive for the
design phase being reviewed, and the participation of all these
people should help to achieve this particular goal.
In addition to professional expertise in a given field, de-
sign team members should ideally possess the following quali-
ties:

• Competency
• Objectivity
• Sensitivity

Design reviews need to include a review of verification


data in order to determine whether the design outputs actu-
ally meet the functional and operational requirements. In ad-
dition, the verification data review will determine (1) if the
design is compatible with all the components and any other
accessories, (2) whether the safety requirements are being
achieved, (3) whether the reliability and maintenance require-
ments are being met, and (4) if the manufacturing, installa-
tion, and servicing requirements are compatible with the de-
sign specifications.

DESIGN REVIEW FOCUS

The design review has a dual focus:

• Internal focus—feasibility of design and produceabil-


ity of design
• External focus—user requirements

TLFeBOOK
Design Review 63

It may be easier to decide when a design review should


be conducted if we create several somewhat artificial stages
and look at what the purpose of the review would be at that
point in the developmental project and some of the more im-
portant points that might be included in an agenda. The criti-
cal elements of a design review that must be kept in mind dur-
ing each review are the following:

• Design evaluation
• Resolution of concerns
• Implementation of corrective actions

DESIGN REVIEW STAGES


Stage 1. Reviews Immediately After Initial
Design Inputs Are Approved
The purpose of this initial design review meeting is to formally
define and confirm the design inputs and expected outputs. It
is also used to initiate the development/manufacture phase.
The product performance specification (PPS) is a critical ele-
ment of the initial design review. The initial design review
meeting will also formally define the design project team. All
members of the project team must be present at the initial
design review meeting. Additionally, an individual who does
not have direct responsibility for the design stage being re-
viewed needs to be present. As we have mentioned, the FDA
requires the presence of specialists who are capable of provid-
ing specific expert guidance in critical areas. To review what
we have said before, this initial design review may include a
review of design verification data to determine (1) whether the
design outputs meet the functional and operational require-
ments of the project, (2) whether the design is compatible with
all its components, (3) whether the safety requirements are
being achieved, (4) whether reliability and maintenance re-
quirements are being met, (5) whether the labeling and other
regulatory requirements have and will be met, and finally (6)

TLFeBOOK
64 Teixeira and Bradley

whether the manufacturing process is compatible with design


specifications.
A typical agenda for this initial design review might in-
clude a review of

1. The design inputs


2. The expected outputs and any known outcomes
3. The project plan
4. The risk analysis
5. The PPS
6. Any other pertinent information

Stage 2. Mid-Project Design Reviews


The objective of these mid-project design reviews is to deter-
mine whether prototypes produced by processes that are iden-
tical to (or simulate) actual production methods and proce-
dures have performed adequately in simulated use testing and
clinical evaluations.
A typical agenda for a mid-project design review might
include the following:

1. Verification that the proposed design satisfies prod-


uct or process requirements
2. Examination of the results of analyses, calculations,
and tests
3. Evaluation of the cost-effectiveness of the product
4. Evaluation of the product performance
5. Evaluation of the manufacturability of the product

Stage 3. Final Design Review Meeting


After all verification and validation activities have been com-
pleted, a final design review meeting should be conducted. The
design review should be held immediately prior to the transfer
of the product to manufacturing for production. The last de-
sign review meeting is the final confirmation that the overall
design output has met the overall design input. All project

TLFeBOOK
Design Review 65

team members should be present at the final design review


meeting. Any changes or conflicting, ambiguous, or incomplete
requirements shall be documented.
A typical agenda for a final design review meeting should
include:

1. A final review of the risk analysis to assess any addi-


tional real or potential hazards associated with the
device under normal and fault conditions
2. Any required updates that need to be made and ap-
proved by all participating design team members
3. Any necessary changes that need to be implemented

Figure 1 Design review meeting record.

TLFeBOOK
66 Teixeira and Bradley

Figure 2 Design review meeting minutes comment form.

prior to transferring the design and development


specifications and procedures to production specifi-
cations and procedures

Figure 1 shows a format for recording the minutes of a


design review meeting. This format allows the dissemination
of the main points of the meeting to other members of the de-
sign team and to appropriate members of management. Fig-
ure 2 shows a format for members of the design team to show
their agreement or disagreement with the design review min-
utes.

MEETING DYNAMICS

Because a successful design review depends almost as much


on conducting a successful meeting as on having accurate

TLFeBOOK
Design Review 67

data, it seems appropriate to review some things that we can


do to ensure a successful meeting. The comments and concepts
reviewed in this section are useful to all members of the design
review team, including managers. They are, in fact, useful,
tested, and verified concepts that will work in any business
communication setting.

Communication Skills
The design review runs on information. People need clear, con-
cise, and complete information to plan, organize, and execute
their responsibilities. Whether you’re leading the meeting,
evaluating the outcome, or making a presentation, you will
succeed or fail based on your ability to communicate. Words
are the vehicles people use to communicate their goals, objec-
tives, and performance standards. Unfortunately, many words
are ambiguous and are often interpreted in different ways.
The definition of what’s a “good wage” may depend on whether
you’re paying it or receiving it. The 500 most commonly used
words in the English language have an estimated 10,000 dif-
ferent meanings. When an engineer says, “I will complete this
assignment as soon as possible,” does that mean in the next
two minutes, two hours, two days, or two months? We need
to define our terms to make sure the receiver and sender are
on the same wavelength.
Being sure that the product has “superior quality” is
a great idea, but it may not explicitly communicate the de-
sired behavior or results the team hopes to accomplish. Staff
members may have their own idea of what “superior quality”
means. Each person may leave the meeting ready to imple-
ment his or her own definition of “superior quality.” When
sending and receiving information, make sure the meaning
is clear. Try saying something like, “My definition of supe-
rior quality is . . .” or “As it applies here, superior quality
means . . .”
The vocabulary of product development includes many
abstract ideas and concepts like “superior quality.” Do not only
define the terms, but also provide concrete examples to help

TLFeBOOK
68 Teixeira and Bradley

explain the abstract idea. Examples and illustrations can pro-


vide tangible reference points to drive home the point.
Acronyms and jargon also pose potential hindrances to
effective communication. In technical areas they cannot al-
ways be avoided; but because the design team is composed of
people from different disciplines, be sure that acronyms and
jargon are explained and clear. “The team is writing a revision
to the PPS. We have new information from ASQC and ANSI
that suggests this will be wise when we prepare the PMA for
the CDRH.” Chances are, some people will not know what all
the initials mean. Spell out or define what acronyms mean, or
your project may end up FUBAR.* This extra step can be the
difference between understanding and confusion.

Did They Get It?


Remember, the biggest problem with communication is the il-
lusion it has been achieved. Very often, defining a single word
or concept is all that is needed to successfully communicate a
thought or feeling. At other times verifying that someone
heard what you said is prudent to ensure the message was
interpreted as intended. The content and delivery of a mes-
sage are obviously important, but what really counts is what
the receiver heard or interpreted.
Verifying the message is a simple technique whereby the
message sender asks the receiver to explain his or her inter-
pretation of the message. If the receiver’s interpretation is ac-
curate, then a successful communication has occurred. If the
interpretation is inaccurate, the sender needs to clarify and
correct the misunderstanding. People are much more likely to
pay attention, concentrate, and listen carefully if they know
they may be called upon to give their understanding of the
message. Going one extra step to check out the receiver’s un-
derstanding can often save a lot of grief. If the mirrored re-

* For this publication that means “fouled up beyond all recognition.” For a
more accurate meaning, ask someone who has been in the military.

TLFeBOOK
Design Review 69

sponse is incorrect, the sender knows the message needs to be


restated. Of course, some breakdowns occur simply because
the listener wasn’t paying attention.
The design team sits at the communication center of the
development program and is particularly vulnerable to com-
munication breakdowns. The difficulty of consistently ver-
balizing clear and accurate messages is immense. Never un-
derestimate that problem. Even the most carefully worded
message can be misunderstood. Periodic message verification
can eliminate confusion and misunderstanding and can pre-
vent the small and large blunders that result from communi-
cation breakdowns.

Listen and Validate


Design team members are often on the receiving end of many
bits of information. It’s estimated that team members spend
up to 40% of their time listening. Being able to listen and accu-
rately understand every message received is easier said than
done. Listening is not easy. It requires focus, concentration,
and a motivation to understand the point being made. Suc-
cessful people realize that effective listening is as important
as effective speaking.
How do you become an effective listener? For one thing,
making eye contact with the speaker helps one to focus and
concentrate. Facing the speaker puts the receiver in a good
position to observe body language and other aspects of deliv-
ery. Words tell us the intellectual content of the message. Tone
of voice and body language tell us the sender’s emotional and
energy levels. Actively observing how the message is delivered
is often critical to understanding the total message.
Effective listening is an active process, not a passive one.
The mind must be fully engaged when listening. No other
thought should be permitted to enter your mind while lis-
tening. However, too frequently other thoughts do enter our
mind and we lose focus. Some people are too busy listening to
themselves to listen to someone else. Others begin to think of

TLFeBOOK
70 Teixeira and Bradley

their response to the speaker even before they understand the


speaker’s point.
The best managers do not only concentrate and listen to
the message; they also periodically validate their own inter-
pretation of the message. They feed back to the speaker their
understanding of what is being said. Factors like the complex-
ity and importance of the message determine how often vali-
dation should occur. In the design review setting it should oc-
cur frequently. There are three levels of validation:

Level I: Validate by feeding back the exact words the


speaker used.
Level II: Validate by feeding back a paraphrase of the
message.
Level III: Validate by feeding back your interpretation of
the words and body language.

Team members need to listen, not only to the words of


the message but also to the tone of voice and body language.
At times it’s necessary to feed back what you think the sender
really means but has not said. As shared understanding
builds, the sender is often motivated to share additional
thoughts and feelings.

Accept the Bad News


Bad news may well be the most useful information design
team members receive. One common reaction to bad news is
to take it out on the person delivering the message. Blaming
the customer is another common response to bad news: “The
product failed because they didn’t follow our directions.”
A third response is to deny, deny, deny.* Data indicating
quality problems is criticized for being inaccurate or incom-
plete. Negative feedback from the customer is rationalized. All

* And if you are really adept, you follow that with counteraccusations and
rationalize.

TLFeBOOK
Design Review 71

these reactions are counterproductive. The bad news can be


helpful. It’s a way of telling the design team that something
is wrong and that a change is needed. An upset customer
sends two types of messages. One has to do with the facts. The
other has to do with feelings. If the team reacts defensively,
and doesn’t listen, there is no opportunity to improve the situ-
ation. If, on the other hand, the members listen with an open
mind and acknowledge the problem, they have taken a step
toward improving the situation.

Monitor and Measure


Measurements show how much progress has been made and
what remains to be done. The design team’s ability to mea-
sure, monitor, and control work product is directly related to
the team members’ ability to identify potential problems and
to take corrective action if needed. Decisions based on data are
a lot better than those based on speculation. Objective data
removes emotion and preconceived ideas from an issue. An ef-
fective design team gets the data before making a decision.
Decisions based on good data are a lot better than those based
on emotions in this setting.
The team needs to draw out its members and find out who
did what, where, why, and how. Separate the facts from the
feelings, assumptions, and opinions. This information should
be verified for accuracy with feedback received from others
and from personal observations. The purpose is not to affix
blame but rather to determine the facts. Given all the facts,
almost anyone can make a proper decision.

Don’t Confuse Motion with Progress


The design review is meant to achieve results. The results are
the bottom-line measurement of performance for the design
team. Activity, effort, and hard work are noteworthy factors,
but the correct output is what really counts. What has been
accomplished? What has been implemented? Some members
of the team can project the appearance of productivity through

TLFeBOOK
72 Teixeira and Bradley

the beehive atmosphere of activity. The atmosphere is tense,


people speak rapidly, the phone rings frequently, and other
employees come and go providing various input. Lots of mo-
tion, but is there any progress?
The chaos of a product development can often make it feel
like a lot of work is being accomplished. Activity and effort
can sometimes mislead a team into thinking goals are being
met. By focusing on results, the team can ensure that activi-
ties are not ends in themselves.
The design team needs to learn a lesson from the military.
A long time ago the military learned that personally going,
looking, and listening are the only reliable feedback. A good
military officer who has given an order goes out and sees for
him or herself whether that order has been carried out, and
how well it has been implemented.
Seeing and hearing things not only give the design team
direct, unfiltered feedback; they also convey to everyone in-
volved in the project an interest in them, their ideas, and the
work they perform.

Meeting Minutes
Many design meetings end with confusion as to who has to do
what and when.* During a design review meeting, as new and
different action items are identified, they can be written down
on a flipchart or whiteboard. The due date and person(s) re-
sponsible for each item should also be listed. At the end of the
meeting, the team leader can do a quick review of all action
items on the list. The team members can then leave the meet-
ing with a clear understanding of who is responsible for which
action items, and when they are due.

Making Decisions That Solve Problems


The last thing we need to discuss before returning to more
information specific to design controls is decisions. Effective

* Especially when the meetings run longer than their allotted time.

TLFeBOOK
Design Review 73

design teams make decisions, from simple to complex, so it’s


important to know how the process works. The ability to define
and solve problems leads to progress and improvement. Deal-
ing with symptoms only wastes time, effort, and money. The
process includes the following steps:

• Define the problem.


• Collect and analyze data.
• Generate alternatives, that is, brainstorm.
• Evaluate alternatives.
• Select an alternative.
• Implement the alternative.
• Evaluate the result.

In design review meetings it is imperative to identify


where the project is at any given time so that a correction, if
necessary, can be made.* Many problems may require multi-
ple meetings, and different team members will frequently be
at different points in the process. Little progress is made when
everyone comes at a problem from a different point in the pro-
cess. An effective team leader brings the group together by
defining where they are in the process. Doing that improves
productivity by focusing the group to concentrate on one task.
Multitasking is largely a myth developed in Redmond, Wash-
ington.†
Team leaders usually prepare what they are going to say,
but often they don’t predetermine what questions they will
ask. An important aspect of leadership is the ability to ask the
right questions at the right time and to insist on answers that
make sense. The answers to the right questions provide the
team leader with the information he or she needs to make deci-
sions. These questions often touch on sensitive or unpleasant
topics. The questions should be simple, direct, and focused.

* Remember the plane on its way to Paris?


† You remember Microsoft and Windows .

TLFeBOOK
74 Teixeira and Bradley

They should be framed to elicit a concrete, specific response.


How the questions are asked (choice of words, tone of voice,
body language) is critical. It must be done in a way that doesn’t
cause defensive or adversarial responses. The questions
should be straightforward and asked from a neutral point of
view, meaning not aggressively or with strong emotion, but
not from a weak or passive position either.
Followup questions are often essential to achieve specific
answers to the questions asked. If a person is unresponsive
or vague, be persistent. Keep asking and probing. Rephrase
the question to get to the core of the issue.
Beware of people who know the solution before they un-
derstand the problem. Some people have the tendency to go
off designing systems, forms, and procedures without under-
standing the real problem.
Design teams are confronted with a wide range of prob-
lems ranging from broken machinery, to late deliveries, to un-
happy customers, to conflicts among target properties or peo-
ple. Some problems are defined in such a way that there is
only one obvious solution: “The problem is that the engineers
aren’t working hard enough.” Other problems are really symp-
toms of more basic underlying problems. Still other problems
can be defined in such grand or global terms that it paralyzes
the ability to act. In addition, design teams are often presented
with problems that inspire futility. After all, if the presented
problem didn’t inspire futility, it would have already been
solved. Right?
The hardest part of problem solving is figuring out what
the real problem is. As indicated, sometimes the presented
problem really isn’t the problem that needs to be solved at all.
Assess the accuracy of the data. Come to grips with reality,
as opposed to the images and perceptions. Break through the
generalizations. Complex problems should be broken down
into smaller, simpler ones.
The problem statement should be free of both causes or
solutions. The problem should describe what currently exists
versus what is desired. The more specific and measurable the

TLFeBOOK
Design Review 75

description, the better that description is. A problem state-


ment such as, “How do we reduce the scrap rate from 5% to
3%?” provides specific and measurable criteria in the problem
statement.
Once the problem is defined, the additional steps include

• Collect and analyze the data.


• Generate options or solutions.
• Evaluate those options or solutions.
• Make a decision.
• Implement that decision.
• Evaluate and measure the result to see if the problem
has been solved.

TLFeBOOK
TLFeBOOK
7
Design Verification

WHAT IS DESIGN VERIFICATION?

Design verification provides objective evidence that design re-


quirements have been met; if they have not been met, it shows
to what extent they have or have not been achieved. It shows
that the design outputs meet the design input requirements.
The verification step confirms whether the outputs meet the
device’s functional and operational specifications. Verification
is intended to show that the design is both safe and reliable
and that the labeling and other regulatory requirements are
satisfied. Verification answers the questions, “Did I make the
product correctly, according to specifications, and can I prove
it?”

77

TLFeBOOK
78 Teixeira and Bradley

THE FDA AND DESIGN VERIFICATION

Each manufacturer shall establish and maintain


procedures for verifying the device design. Design verifi-
cation shall confirm the design output meets the design
input requirements. The results of the design verification,
including identification of the design, method(s), the date,
and the individuals performing the verification, shall be
documented in the DHF.* [21 CFR Part 820, Subpart C,
Section 820.30(f )]

Of and by itself, this section really doesn’t say much ex-


cept that somehow we are required to verify the design being
developed and document the verification, including the meth-
od(s) used. By insisting on the documentation of the methods
used for the verification, the FDA has implied that we need
to do something besides saying, “Yup, that design is what we
were aiming for!” The problem is, of course, defining the word
“verification.” Verification means confirmation by examination
and provision of objective evidence that specific requirements
have been fulfilled.† That last sentence is the definition ac-
cording to the FDA. Read it again. Sometimes you have to
wonder what they were thinking when they wrote that sen-
tence. It may have been a little clearer if it was simply stated
as, “Test your design, preferably with standard methods, and
document the results.”

DESIGN VERIFICATION REQUIREMENTS

Procedures need to define verification methods. Any approach


that establishes conformance with a design input requirement
is an acceptable means of verifying the design with respect

* Design history file.


† 21 CFR Part 820, Section 820.3(aa).

TLFeBOOK
Design Verification 79

to that requirement. The results of all verifications must be


recorded and include identification of the design, method(s),
date, and individual performing the verification. This forms
part of the DHF.

TYPICAL VERIFICATION ACTIVITIES

• Design reviews
• Inspection/testing
• Biocompatibility testing
• Risk analysis
• Worst-case analysis
• Thermal analysis
• Fault tree analysis (FTA)
• FMEA
• Package integrity tests
• Bioburden tests
• Comparison to similar product designs [510(k)s]
• Measurements
• Demonstrations
• Alternative calculations
• Documentation review

The FDA likes to see verification activities shown in matrix


form:

Input Output Result

(Requirement) (Test method) (Outcome ⫽ input


requirement?)
Min. flow rate ⫽ 500 ml/hr ASTM XXXX FR ⱖ ⫽ 500 ml/hr

Some typical verification examples may include the fol-


lowing:

TLFeBOOK
80 Teixeira and Bradley

Input Output Review/Result

Noncytotoxic Agar diffusion Cytotoxicity report—Pass


method
Product sizes Product specifica- Inspection—Pass
tion (2 cm ⫻ 2
cm, 4 cm ⫻ 4 cm)
Flow rate (500 ml/hr) Test method Test—Pass
24-hr wear time Clinical protocol Clinical study report
2-year expiration Stability protocol Stability report
CE mark Declaration of con- CE technical file
formity
Prescription Product labeling 510(k)

RISK MANAGEMENT

Risk management is a process of applying policies and proce-


dures to identify, analyze, control and monitor risks.* Several
things need to be done to identify and manage these risks:

• Identify the risk or potential risk in the design (or the


process) that may result in a hazard.
• Assign a degree or level of risk to the hazard, including
the probability of its occurrence and the severity of the
hazard if it occurs.
• Eliminate or reduce the problem or risk.
• Evaluate the controls and the solutions and determine
whether the solution caused new problems or risks;
repeat the steps if it has.
• Document the process.

Risk analysis is an important part of the overall design


project and an integral part of the design controls process. A

* Design Control Guidance Document for Medical Device Manufacturers,


FDA, March 11, 1997.

TLFeBOOK
Design Verification 81

sample SOP for risk analysis is shown in Appendix E. In sim-


ple terms, only two outcomes are possible in the development
of a medical device regardless of its specific intended use. The
finished device is “safe and effective,” or it is not. The FDA
calls the identification and correction of use-related hazards
associated with the development of a medical device human
factors engineering (HFE).* The FDA has a stated goal for us-
ing HFE: “To minimize use-related hazards, assure that the
intended users are able to use the devices safely and effec-
tively throughout the product life cycle, and to facilitate re-
view of new device submissions and design control documenta-
tion.Ӡ
So how do you do this? First, the design team needs to
review all the device characteristics and determine whether
they are required for the intended use and whether they could
pose a real or potential hazard under normal and/or fault con-
ditions. The team then needs to determine the degree of risk
associated with the hazard. Most people have no trouble iden-
tifying the really big problems. Actually, most design engi-
neers “design out” major hazards from the beginning. Some of
the problems and risks that are often overlooked are the minor
ones or those associated with fault conditions.‡ As medical de-
vice manufacturers we can’t always account for every dumb
or unexpected use that someone tries with a product, but we
must try. Although these minor problems don’t necessarily
render the device unsafe or ineffective, they are often the dif-
ference between a good product and a great product. How
many product prototypes have you seen that were intended to
be placed on the human body for a couple of days and had an
outline that contained right angles? Is it difficult to “round”

* Kay, R. and J. Crowley, Medical device use-safety: Incorporating human


factors engineering into risk management, CDRH Guidance Document, July
18, 2000.
† Ibid., p. 5.
‡ Operating conditions other than “normal.”

TLFeBOOK
82 Teixeira and Bradley

those edges and make the device a bit more comfortable for
the person who will wear it?
Risk analysis is not difficult if the team remembers to
think of the customers. The new-device design has accom-
plished all the hard stuff (like managing a fistula), but what
happens to the patient when he or she puts it on? It may help
manage or treat whatever problem it was intended to address,
but shouldn’t it be as comfortable as possible? It should also
be as reliable as possible; does the design account for that?
What about the needs of the physician, the nurse, or the care-
giver in the home health environment? Does the device be-
come ineffective and perhaps create a hazard for any of them?
What happens to the device if it is used exactly as you in-
tended and instructed? It should work perfectly, and most de-
sign teams think about that. What happens if someone uses
the device in a manner that isn’t typical? What happens if the
device is used in an environment the team did not take into
account? What happens when you put all these things to-
gether—the user, the environment, and the device—and they
interact in a manner you didn’t think of ?
The reason for risk management is to answer questions
such as those above. If the “hazard” created by an atypical use
of a device can be reduced or eliminated in the design stage,
then do it. Sometimes an identified hazard cannot be handled
by changing the design. In that case it can be managed by
the “warnings” and “cautions” in the labeling. A simple way
to collect and document an overall risk analysis is shown in
Table 1. The document is called the risk analysis master rec-
ord. Although for space considerations it is shown here as a
single-page table, it can be, and should be, an extensive and
exhaustive document. Its purpose, besides the obvious docu-
mentation use, is to focus in one document all the hazards that
are or could be associated with the device being developed. It
implies testing and change and helps verify that the design is
both safe and effective for the intended use.
Keep in mind that completing the last column “Action
Taken to Reduce Risk” may entail a great deal of work and

TLFeBOOK
Design Verification 83

Table 1 Risk Analysis Master Record

possible redesign of the product itself. This record may go on


for several pages. It seems almost condescending to mention
it but there are risks associated with most medical treatments
and there may be risks and hazards associated with the use
of a particular device. However, these major risks are almost
always known to the people doing the design development and
are either eliminated or minimized by the design itself or han-
dled in the labeling if the associated risk is outweighed by the

TLFeBOOK
84 Teixeira and Bradley

benefits delivered to the patient when using the device. Most


engineers have completed a reliability analysis related to the
use of a particular device. They know what will happen to the
patient, the user (if other than the patient), and the sur-
rounding environment should there be a device failure. What
usually needs more attention is what will happen if someone
tries an unexpected use of the product. These problems may
include

• Mechanical problems—pinch points, break points, etc.


• Chemical problems—toxicity
• Biological—allergy, infection, biocompatibility issues
• Heat—burns from high temperatures in or around the
device
• Electrical—shock, electromagnetic interference (EMI)
• Radiation

Remember, the team has been designing the device for


use in a very particular way and most likely with a specific
environment in mind, but someone could use the device in an
entirely different way or in an unexpected environment. Sup-
pose that you manufacture a skin cleanser that has been for-
mulated into a cream. What will happen if the end user con-
fuses the cleanser with his or her toothpaste tube and uses it
to clean his or her teeth? You may not be able to prevent the
occurrence, but have you done anything to minimize it?* The
development team and the manufacturer both have an obliga-
tion to minimize risks associated with the use of the device
even if, the end user uses the device inappropriately.
It should be noted at this point that this is not often easy
to do, since it is sometimes considered impossible to anticipate
every way that a device may be used. Even when a set of cir-
cumstances can be envisioned, it is often difficult to then esti-

* An oversimplified answer may be to design the package so that it doesn’t


even come close to resembling toothpaste (or denture cleaner) packaging.

TLFeBOOK
Design Verification 85

mate how severe that particular hazard may be. But it must
be done. The design team should:

• Take any existing information and use this informa-


tion to identify use related hazards. This information
can come from the anticipated use of a device and from
any existing information that may be available on
similar devices. In this way, hazards that may be too
dangerous to actually cause to happen in a user evalu-
ation can be addressed. There are some hazards that
can actually be identified through actual or simulated
use testing. If actual use testing is possible, it is pref-
erable to use subjects that are representative of the
people who will actually be using the device.
• After the information is gathered, the design team
should estimate the severity of the problem; that is,
the risk should be assessed.
• The team must then develop controls that will mini-
mize or eliminate the identified hazard in the best pos-
sible manner.
• The team must now make sure that the controls im-
plemented to control or eliminate a particular hazard
have not introduced new hazards and are truly effec-
tive.
• The team should now verify that the functional and
operational requirements of the device are still being
met.
• Safety and efficacy should be validated.
• Finally, everything needs to be clearly and accurately
documented.

HUMAN FACTORS

In its guidance document on human factors and risk manage-


ment the FDA outlined the following human factors consider-
ations for use in risk management:

TLFeBOOK
86 Teixeira and Bradley

• User preferences do not necessarily indicate safety


and effectiveness.
• Rare and unusual scenarios resulting in serious conse-
quences often pose the greatest threat to safe and ef-
fective device use.
• Direct inspection or paper-based analyses might not
identify all hazards.
• The amount of thought that a user exerts when using
the device may cause hazards during periods of high
stress.
• Indicator lights, alarms, and other signals may not be
appropriate in all environments. This may be espe-
cially true when there are multiple alarms on the
same device.
• A device that is safe for use with one intended user
may not be safe for all users. Be sure you take into
account the physical and mental conditions of all pos-
sible users.
• Design the device interface to tolerate errors, and
allow for backup mechanisms and systems where pos-
sible.

RISK MANAGEMENT DOCUMENTATION

As a rule, the level of documentation should be in proportion


to the level of device-related hazards. In general, it should in-
clude

• The purpose and operation of the device; the patient


population including those who should be excluded; a
description of the device and a comparison with other
similar devices if possible
• The characteristics of the user interface and any la-
beling
• How the user is expected to interact with the inter-

TLFeBOOK
Design Verification 87

face; device maintenance and any other steps the user


or patient is expected to take
• Training information for the user or patient
• Use-related hazards and their solutions
• Testing and evaluations (verification) and validation
of these results

TLFeBOOK
TLFeBOOK
8
Design Validation

WHY VALIDATE?

Validation is necessary to demonstrate that the design is a


product the marketplace needs. As a result, its purpose is to
provide objective evidence that design specifications (outputs)
confirm predetermined user needs and intended uses. Design
validation is essentially showing by objective evidence that de-
vice specifications conform to defined user needs and the in-
tended use(s) of the device. The design team needs to ask and
answer questions such as, “Did I make the right product for
the user, and can I prove it?” In terms of validation of the pro-
cess used to make the device, “Does the process consistently
produce a product meeting predetermined specifications, and

89

TLFeBOOK
90 Teixeira and Bradley

can I prove it?” The key phrase in both of these questions is,
“Can I prove it?” If you can’t, your design has not been vali-
dated.

THE FDA AND VALIDATION

What does the FDA say about validation?

Each manufacturer shall establish and maintain


procedures for validating the device design. Design vali-
dation shall be performed under defined operating condi-
tions on initial production units, lots, or batches, or their
equivalents. Validation shall ensure that devices conform
to defined user needs and intended uses and shall include
testing of production units under actual or simulated use
conditions. Design validation shall include software vali-
dation and risk analysis, where appropriate. The results
of the design validation including identification of the de-
sign, method(s), the date, and the individual(s) per-
forming the validation shall be documented in the DHF.
[21 CFR Part 820, Subpart C, Section 820.30(g)]

Now don’t be misled by the statement “where appro-


priate.” As a manufacturer of a medical device regulated by
the FDA, for you it is always appropriate unless you can docu-
ment a justification for not doing it. For example, it would
seem almost impossible (and foolish) to try and justify deci-
sions made about safety without a formal risk analysis.

DESIGN VALIDATION REQUIREMENTS

Remembering that design validation can only follow success-


ful design verification, the requirements for validation are
listed below.

• Design validation must be performed under defined


operating conditions. This means that the conditions

TLFeBOOK
Design Validation 91

under which the product is made for validation pur-


poses, which may include clinical use tests, should be
the same conditions anticipated for actual full-scale
production. In general, three production runs
(batches, lots, etc.) is the accepted norm.
• Validation activities must not be thought of as simply
a checkmark. The chosen validation activities must
provide discrete data (values or measurements).
• Validation activities must address the needs of all rel-
evant parties (i.e., patient, health-care worker, etc.)
and be performed for each intended use. If you have
developed an ostomy pouch that may be used by pa-
tients with a colostomy and patients with a urinary
diversion, then the clinical validation of the design
must be performed on both types of subjects. Of
course, reliable in-vitro testing could be used in place
of an actual clinical trial if there is sufficient scientific
and statistical evidence that the test is predictive of
patient use.
• Validation activities should address the design out-
puts of labeling and packaging.
• Risk analysis must be completed in the design valida-
tion stage. If it was started during verification, then
this is the phase when it should be rechecked to deter-
mine whether selected solutions to hazards actually
work for the patient and the caregiver and whether
the solutions themselves haven’t caused additional
problems.
• Software used to control manufacturing processes
needs to be validated. Validation needs to start in
the design phase but may finish in the production
phase.
• Design validation is completed when clinical evalua-
tion is complete. Clinical evaluation is not just a clini-
cal trial. It may include testing for safety and effec-
tiveness, literature searches, and 510(k)/PMA
reviews.

TLFeBOOK
92 Teixeira and Bradley

• The results of validations must be documented and in-


clude the design method(s), date, and individual(s)
performing the validation. This forms part of the de-
sign history file.
• Design validation may expose deficiencies in original
assumptions (e.g., device specs/outputs) regarding
user needs and intended uses. If this is true, then all
discrepancies must be addressed and resolved (e.g.,
change in design output required or change in user
need or intended use).

TYPICAL VALIDATION ACTIVITIES

Validation activities may include any or all of the following:


• Clinical studies/trials conducted through institutional
review boards (IRBs) and with an investigational de-
vice exemption (IDEs). For nonsignificant risk devices
an IRB approval process is usually sufficient.
• 510(k)/PMA historical database search for predicate
devices.
• Stability studies on both the product and the pack-
aging.
• Clinical evaluations in clinical or nonclinical settings
(market evaluation).
• Literature search (published journal articles).
• Review of labels, labeling, packaging, and product his-
tory.
• Simulated use testing.
• Transit trials.
• Performance tests.
• Functional tests.
• Biocompatibility tests.

RISK ANALYSIS . . . ONE MORE TIME

Risk analysis, including the final identification of possible haz-


ards associated with the design in both normal and fault con-

TLFeBOOK
Design Validation 93

ditions, must be completed in this phase of the design controls


process. Any risks associated with those hazards, including
those resulting from user error, should then be calculated in
both normal and fault conditions. If a risk is deemed unaccept-
able, it must be reduced to acceptable levels by the appropriate
means, which might include redesign or the addition of warn-
ings and contraindications to the product labeling. Remember,
as we have said before, it is important to ensure that changes
made to eliminate or minimize hazards have not introduced
new hazards.

COMMON RISK ANALYSIS TOOLS

In addition to the methodology mentioned in the last chapter,


other risk assessment tools can be used and should be ex-
plored. These are

• FTA—fault tree analysis


• FMA—failure-mode analysis
• FMEA—failure-mode effects analysis
• FMECA—failure-mode effects criticality analysis
• HACCP—hazard analysis and critical control points*

FTA: Fault Tree Analysis


A fault tree analysis is a graphical representation of the major
and critical faults associated with a design, the cause for those
faults, and the potential solutions to eliminate or minimize
the hazards. This tool can help identify major hazards and
areas of concern in a new product design, as well as corrective
actions and strategies to minimize or eliminate the problems.
The nature of FTA makes it a good tool for the risk manage-
ment of those devices that are less likely to cause major prob-

* HAACP is currently under study by the CDRH/FDA as a system to mini-


mize the impact of the manufacturing process on product safety and perfor-
mance.

TLFeBOOK
94 Teixeira and Bradley

lems for patients and health-care workers, that is, most class
I devices.
It is relatively easy to construct an FTA:
1. Construct a box at the top of the FTA diagram. List
the component to be analyzed inside the box.
2. Identify the critical hazards and faults associated
with the design component. This can be done by having the
group brainstorm or by creating a cause-and-effect, or “fish-
bone,” diagram.* (A cause-and-effect diagram is explained in
Appendix F.) These hazards should be placed in separate
boxes below the component box being analyzed.
3. Identify the causes for each hazard or fault. These
causes should be placed in ovals (or some other shape) below
the fault and connected to the particular fault box with lines.
4. Work toward a controllable cause. The design team
should continue working toward identifying causes until a
cause that can be controlled is reached. This has sometimes
been referred to as a root cause.
5. Identify a correction for each root cause. Upon identi-
fying the underlying cause of a hazard, the design group can
then identify corrective actions that will eliminate or mini-
mize the problem. Boxes for each of these solutions should be
drawn below the controllable cause and linked by a line.
A typical fault tree analysis document would look like
Figure 1.

FMA: Failure-Mode Analysis


Failure-mode analysis is a procedure to determine which mal-
function symptoms appear immediately before or after a fail-
ure of a critical parameter in the product design. After all the
possible causes for each symptom are listed, the product may
need to be redesigned to eliminate the problem, or other strat-
egies may be applied to minimize the fault.

* The FTA can be used as a summary diagram of an FMEA. See the next
section.

TLFeBOOK
Design Validation 95

Figure 1 Fault tree diagram.

FMEA: Failure-Mode Effects Analysis


Failure-mode effects analysis is a procedure in which each po-
tential failure mode in every component of a design is analyzed
to determine its effect on other components and on the re-
quired function of the design. It can be useful in identifying
critical product or process factors and designing the solutions
to potential problems. It can also establish controls that may
be implemented to prevent process problems and to help en-
sure reliability.
One method for using the FMEA analysis is to create a
spreadsheet. Frequently, a multicolumn spreadsheet as
shown in the Figure 2 is ideal for creating the analysis. The
design team should identify all the possible modes of failure
and list each of them on a separate line in column 1 of the
spreadsheet (Mode of failure). The team can then create a list
of the possible causes for each of the failures and list each

TLFeBOOK
96 Teixeira and Bradley

Figure 2 FMEA chart.

cause in column 2 (Cause of failure). Column 3 could be used


to list the impact of the failures on the customer, the product,
or the process (Failure effect). In columns 4 to 6 a table of
relative values* is assigned for each failure with regard to fre-
quency of occurrence, impact on the customer, and likelihood
of detection. These values, which the consensus of the team

* On a scale of 1 to 10, a 1 would indicate a frequency of less than 2 and a


10 would indicate a frequency of greater than 25. Using the same scale for
“severity” a score of 1 indicates a trivial problem, a 10 indicates loss of the
customer, and a 5 might indicate a formal complaint. For “detection” a 1
would indicate certain detection while a 10 would indicate there was no pos-
sible way to predict the problem would occur.

TLFeBOOK
Design Validation 97

(or actual data) determines, are placed in the appropriate


spreadsheet cell. A risk factor is then calculated by multi-
plying the values in each row of columns 4, 5, and 6. This risk
factor is recorded in the next column of the spreadsheet. The
remedy is then recorded in column 7 (Design action), and the
method for verifying that the action has been accomplished is
shown in column 8.

FMECA: Failure-Mode Effects and Criticality


Analysis
Failure-mode effects and criticality analysis is a procedure
performed with an FMEA analysis to classify each potential
failure relative to a list of critical factors and helps determine
what function (or resource) is responsible for the critical fac-
tors.
A spreadsheet method is also useful for preparing the
criticality analysis. Once again, a multicolumn spreadsheet is
created. Column 1 of the spreadsheet is used to identify the
features and/or components, and all the remaining columns
are used to identify the critical factors. The features or compo-
nents of the device are listed in the first column. The critical
factors for each of these features are then listed in the appro-
priate row. The number of columns of the spreadsheet depends
upon the critical factors identified. A value of high (H), me-
dium (M), or low (L) representing the risk associated with each
feature is placed in the appropriate cell of the spreadsheet.
The lower portion of the spreadsheet is used to indicate which
internal function or other resource controls is responsible for
a particular factor. Figure 3 shows a criticality analysis
spreadsheet.

HACCP: Hazard Analysis and Critical


Control Points
The FDA has begun a feasibility study to determine if HACCP
can help the CDRH in

TLFeBOOK
98 Teixeira and Bradley

Figure 3 Criticality analysis, manufacturing process (example).

1. Premarket submissions in both the product and pro-


cess review
2. Identifying and correcting problems for devices cur-
rently on the market that have performance issues
3. Streamlining premarket inspections

According to the FDA, the key elements of the system in-


clude

• Identification of all potential hazards to safety and


performance for both the product and the process
• Identification of the corresponding corrective actions
for the identified hazards
• Identification and control of the hazards at critical
steps during the manufacturing process, the so-called
critical control points
• Documentation that this control is maintained on an
ongoing basis

TLFeBOOK
Design Validation 99

It is believed that this system will identify hazards more


quickly and specifically and allow preventative measures to
be taken sooner, thereby reducing reliance on government in-
spection. Because it is a well-known fact that you cannot in-
spect quality into a product, that is a belief with which it is
difficult to argue.*

RISK ASSESSMENT OF MEDICAL


DEVICE MATERIALS

The principles of risk management mentioned in the previous


sections also apply to the materials used as or in medical de-
vices. A number of other things should be explored when eval-
uating materials.† Questions need to be answered when se-
lecting a material for use as or in a medical device. These
questions include
How will the material be exposed to the patient, and for
how long? For example, the material (or the device) may be
in contact with a patient for only a few minutes or an hour.
Some materials may be in contact with a patient for 24 hours
and may be used for only a matter of days or for the rest of
the patient’s life.
Will the material be in contact with intact skin? Blood?
Internal organs? It seems fairly obvious that applications that
will contain materials that will always be in contact with in-
tact skin need a different level of evaluation than a material
that may be in contact with the blood supply or an internal
organ.

* As the FDA has not yet adopted HACCP, the methodology is not included.
Information about the program can be obtained from the FDA by contacting
Adrianne Galdi, the team leader for the study at the CDRH.
† These are preferably evaluated at the beginning of the material selection
process.

TLFeBOOK
100 Teixeira and Bradley

Is the material stable in the environment intended for


use? Will other chemicals or materials used in the treatment
setting cause the material to change? Even for simple materi-
als this can be a difficult question to answer, as it is often dif-
ficult to predict the environment in which the end user will
ultimately use the material. This is the reason for the risk
assessment. A lot is known about materials and their stability
under a wide variety of conditions. Even if it seems impossible
that your material or device could not possibly encounter ad-
verse conditions, they should be evaluated and assessed. Un-
der some circumstances, perhaps the best you can do is to
warn potential users of unsafe environmental conditions and
put that warning in the product labeling.
Will other materials or conditions cause the material to
leach some of its components? Strictly speaking, this question
may be part of the environmental considerations of a material.
However, some polymeric materials are themselves a physical
blend of a variety of other materials used to impart specific
properties to the finished entity. These properties might in-
clude flexibility, specific adherence, absorption, color, and any
number of other things. Many of the ingredients used to make
these property modifications are not polymers themselves but
are relatively low-molecular-weight materials. Low-molecu-
lar-weight materials often migrate to the surface of the poly-
mer that forms the matrix. Temperature, pressure, and con-
tact with other materials can cause this migration to occur
faster. Are these “minor” materials a source of irritation and
biocompatibility problems?
Does the material change its characteristics (physical
and chemical) when sterilized? Do different types of steriliza-
tion have different effects? Are there long-term changes to the
material because of the sterilization process? The choice of the
sterilization method can have a profound effect on the safety
and performance of a material or a device. If you use EtO ster-
ilization on a material that is a significant barrier to gas, you
may have a sterile product in the package, but if the ethylene
oxide is trapped within the material you may have a potential

TLFeBOOK
Design Validation 101

toxicity problem.* Is there a difference† between electron-


beam sterilization and gamma ray irradiation? Well, except
for a few quantum mechanics, not a whole lot of scientific dif-
ferences exist for the practicality of sterilization. However,
some materials respond differently to the dose rate at which
they are irradiated. Could this cause a potential problem? For
most commonly used materials in medical devices the answer
is probably not. But it’s not true for all materials and should
be considered. Polymers cross-link, degrade, or do nothing in
the presence of radiation.‡ People often consider cross-linking
as “good” and degradation as “bad.” In fact, neither is either.
If cross-linking results in a material that is now more brittle
than is appropriate for a medical device, is that good? Still
degradation is a problem and it doesn’t always occur immedi-
ately. It wasn’t that long ago when polypropylene syringes
were sterilized by irradiation and were found to crack some
months later.
Are there interactions of the materials with drugs that
the patient may take? Have you ever seen a medical-grade
adhesive that was working really well on a particular person
for years and then one day stopped sticking? After an exhaus-
tive study the only thing that seemed to have changed was
that the person was receiving a new prescription drug. Was
this a coincidence or a drug interaction? It was hard to tell,
because there just wasn’t sufficient information. An enormous
amount of information is available today on drug interactions.
Drugs are materials like anything else. Could they interact
with your device or material?
Does anything in the literature suggest that the material
or a closely related material has biocompatibility problems or

* This may be in addition to better-known packaging material problems


with EtO.
† This is, of course, neglecting such mundane things as cost and product
shape, etc.
‡ Actually, they do a little of everything.

TLFeBOOK
102 Teixeira and Bradley

is toxic? There are databases that contain a wealth of informa-


tion on the toxicity and biocompatibility issues associated with
different materials. The suppliers of the commercial products
that you use can get you useful information from the very be-
ginning. Some times it’s even as simple as reading the mate-
rial safety data sheet (MSDS) that came with the sample you
requested. Sometimes you need a chemist to tell you the risk
associated with your material based on its similarity to some-
thing else.
Many more questions need to be answered before a mate-
rial could be considered safe and effective for use as a medical
device itself or as a component of such a device. The funda-
mental question revolves around whether the device or mate-
rial is sufficiently biocompatible for its intended application.

BIOCOMPATIBILITY

Biocompatibility is generally determined by tests using toxico-


logical principles that provide information on the potential
toxicity of materials in a clinical application. Many classical
toxicological tests, however, were developed for a pure chemi-
cal agent and are not applicable to biocompatibility testing of
materials and devices.
A biomaterial is usually a complex entity, and the mate-
rial toxicity is affected by both physical and chemical proper-
ties. Toxicity from a biomaterial or polymer formulation often
comes from components that migrate to the surface and are
extracted from the material. The chemical composition of the
material is often not known. Sometimes toxicological informa-
tion on the material and its chemical composition is not avail-
able, and the possible interactions among the components in
any given biological test system are often not known.
Biocompatibility cannot be defined by a single test. It is
necessary to test as many biocompatibility parameters as pos-
sible. It is also important to test as many samples of the mate-
rial as possible. Suitable positive and negative controls should

TLFeBOOK
Design Validation 103

be used and produce a standard response in repeated tests.


The use of an exaggerated challenge, such as using higher
dose ranges and longer contact durations or multiple insults
that are many times more severe than the actual use condi-
tion, is important. Adopting an acceptable exposure level that
is multiple factors below the lowest toxic level has also been
a general practice.
Most of the biocompatibility tests to establish acute toxic-
ity are short-term tests. Data from these short-term tests
should not be overextended to cover the areas where no test
results are available. Biocompatibility testing should be de-
signed to assess the potential adverse effects under actual use
conditions or specific conditions close to the actual use condi-
tions. The physical and biological data obtained from biocom-
patibility tests should be correlated to the device and its use.
Accuracy and reproducibility of these tests depend on the
method and equipment used and often on the investigator’s
skill and experience.
You must consider several toxicological principles before
planning biocompatibility testing. Biocompatibility depends
on the tissue that contacts the device. For example, the re-
quirements for a blood-contacting device would be different
from those applicable to an external urinary catheter. The de-
gree of biocompatibility assurance also depends on the
involvement and the duration of contact with the human body.
Some materials, such as those used in orthopedic im-
plants, are meant to last for a long period in the patient. In
this case a biocompatibility test needs to show that the im-
plant does not adversely affect the body during the long period
of use. The possibility of biodegradation of the material or de-
vice should not be ignored. Biodegradation by the body can
change an implant’s safety and effectiveness. The materials
leached from plastic used during a hemodialysis procedure
may be very low, but the patient who is dialyzed three times
a week may be exposed to a total of several grams during his
or her lifetime. Therefore, cumulative effects should also be
assessed when appropriate.

TLFeBOOK
104 Teixeira and Bradley

Two materials having the same chemical composition but


different physical characteristics—for example, particle
size—may not induce the same biological response. Also, past
biological experience with seemingly identical materials is not
necessarily indicative of biocompatibility in a new application.
Toxicity may come from leachable components of the material
due to differences in formulation and manufacturing proce-
dures.
The empirical correlation between biocompatibility test-
ing results and actual toxic findings in humans and the ex-
trapolation of the quantitative results from short-term in-vitro
tests to quantitative toxicity at the time of use are controver-
sial. These need careful and scientifically sound interpretation
and adjustment.
The challenge of biocompatibility testing is to use the
knowledge to reduce the degree of unknowns and to help make
the logical decisions. The hazard presented by a substance,
with its inherent toxic potential, can only be truly known when
the material is actually exposed to a patient. Therefore, risk
is a function of toxic hazard and exposure. The safety of any
materials that may migrate from a device or are contained in
the device or on its surface can be evaluated by determining
the total amount of potentially harmful substance, estimating
the amount reaching the patient tissues, assessing the risk of
exposure, and performing a risk-versus-benefit analysis.
When the potential harm from the use of the biomaterial is
identified from the biocompatibility tests, this potential must
be compared against the availability of an alternate material.

REGULATORY ASPECTS
OF BIOCOMPATIBILITY

Basic biocompatibility guidance is found in the document Tri-


partite Biocompatibility for Medical Devices, which provides
a suitable framework to test for biocompatibility. This docu-
ment is from the Toxicology Subgroup of the Tripartite Sub-

TLFeBOOK
Design Validation 105

committee on Medical Devices of Canada, the United King-


dom, and the United States. The Tripartite guidance
recommends the application of fundamental principles to de-
termine the appropriateness of biocompatibility testing.
To evaluate the safety of medical devices, the guidance
categorizes medical devices based on the type of body contact.
The biological tests required depend on the body contact and
contact duration. See Table 1, which lists the biological tests
that might be applied for evaluating the safety of medical de-
vices made of polymers. It does not imply that all tests listed
under each category are necessary. The tests can be modified
to accommodate devices made of metals, ceramics, or biologi-
cal materials.
Medical device manufacturers need to be careful about
using generally recognized as safe (GRAS) substances. GRAS

Table 1 Biological Tests—Biomaterials


Level I
Acute Screening tests Cytotoxicity
USP biological tests
Hemolysis
Other tests Irritation
Sensitization
Implantation
Hemocompatibility
Mutagenicity
Reproductive
Pyrogenicity
Subchronic Irritation
& chronic Sensitization
Implantation
Hemocompatibility
Reproductive & developmental
Level II Chronic Implantation
Reproductive & developmental
Carcinogenesis
Additional tests based on Level I,
e.g., pharmacokinetics
Level III Clinical studies

TLFeBOOK
106 Teixeira and Bradley

substances can be found in 21 CFR Part 182; they are applica-


ble to food but are not directly applicable to medical devices.
Any material or substance in the GRAS list cannot automati-
cally be assumed as safe and effective for medical devices.

Biocompatibility Testing Programs


Manufacturers must devise a biocompatibility test program
that involves some or all of the following activities:

• Gather information on the material and the finished


device.
• Complete a physio-chemical characterization of the
material.
• Identify rapid, sensitive, cost-effective screening tests.
• Monitor incoming raw materials, the final product,
and the manufacturing process.
• Define the product release tests and the pass/fail cri-
teria.

In addition, the manufacturer should select reliable,


state-of-the-art bioassays to demonstrate safety for the in-
tended use of the device.
Regulatory issues are equally as complex as the scientific
considerations. Some regulatory issues are

• Anticipated human exposure to the device


• Biological resistance to chemical insult
• Testing variables
• Species differences
• Relevance of the test to the device and its use
• Substantiating the accuracy and predictive values of
the test
• Proper interpretation
• The use of no observed biological responses (negative
results) to chemical insult(s) to predict biocompati-
bility

TLFeBOOK
Design Validation 107

Undesirable extremes should be avoided during the de-


sign of biocompatibility testing programs. It is important not
to attempt to demonstrate biocompatibility by a single test,
and a biocompatibility program should be based on the in-
tended use of the device. The large number of tests and test
samples is as important as the accuracy, specificity, signifi-
cance, and economy of the testing. Medical devices vary widely
in their types, uses, functions, exposures, and contact ions.
Therefore, one test system cannot accommodate all applica-
tions. Manufacturers do not have to repeat extensive biocom-
patibility testing programs simply to fill the files of evidence
of safety if the device is constructed of well-known, previously
well-tested materials, or only uses materials with a long, safe
history for the same intended use.
Some tests may be inappropriate or unnecessary for the
intended use of the device. For example, pyrogenicity tests are
appropriate for intravenous catheters but not for topical de-
vices that contact only intact external surfaces. Figure 4 lists
suggested biocompatibility tests.

PHASES OF BIOCOMPATIBILITY TESTING

Good biocompatibility testing programs for medical devices


should follow three levels of testing. See Table 1.

• Level I tests include information on the physical and


chemical characterization of the materials and screen-
ing toxicological tests. Level I tests are generally not
difficult to perform, require readily available equip-
ment, and are considered a minimal characterization
of biomaterials. These tests have broad application
and low resolution and are recommended for screen-
ing during the early stages of development and contin-
ued monitoring of new lots of materials.
• Level II tests involve acute toxicity tests and some
subchronic and chronic tests if needed. Level II testing

TLFeBOOK
108
Teixeira and Bradley

Figure 4 Suggested biocompatibility testing.

TLFeBOOK
Design Validation 109

is basically an extension of Level I and involves a vari-


ety of in-vitro and in-vivo testing of devices that re-
quire additional testing based on the Level I screening
tests. This includes extensive preclinical tests such as
pharmacokinetic studies and lifetime bioassays or
special testing due to complexity and/or intended use
of the device.
• Level III testing involves clinical studies. Manufactur-
ers should determine whether or not to proceed to
Level II or Level III testing depending on the results
of Level I tests.

Screening Tests
Screening tests allow rapid and relatively inexpensive rejec-
tion of incompatible materials at an early stage and can be
used as a monitoring device of the manufacturing process. The
cytotoxicity tests, intracutaneous and/or skin irritation tests,
and hemolysis tests are good candidates for screening. In addi-
tion, many cell or tissue culture systems can be custom-tai-
lored to biomaterials. Unless clearly contraindicated, both di-
rect-contact tests and tests with extract with polar and
nonpolar extraction media should be considered.
Some manufacturers overestimate the importance of
screening tests alone as a proof of biocompatibility. These
screening tests are not intended to demonstrate that the mate-
rial is biocompatible, but to reject grossly incompatible mate-
rials.

Acute Toxicity Tests


Biomaterial testing strategy has been broadly established as
acute, subchronic, and chronic.

• Acute toxicity is generally a toxic effect resulting from


short-term exposure, usually less than 30 days, to a
chemical substance. Biological tests for acute toxicity
include estimates of irritation, sensitization potential,

TLFeBOOK
110 Teixeira and Bradley

cytotoxicity, hematologic profile, and evaluation of the


local toxic effect on living tissue by surgical implanta-
tion.
• Subchronic toxicity is generally a toxic effect resulting
either from prolonged exposure up to 90 days or from
multiple exposures.
• Chronic toxicity is generally a toxic effect from contin-
uous and prolonged exposure.

The most common cause of acute toxicity of a biomaterial


is the presence of biologically active leachable substances, and
the acute toxic responses are closely related to the kind and
quantity of leachable substances in or on the materials or de-
vices. Although the absence of leachable substances is desired,
extraction procedures will yield detectable quantities of chem-
ical substances. It is important to assess the potential toxic
effect of the material and to keep in mind the potential toxic
effect of apparently minor or undetected substances.
Toxicity testing should be used in conjunction with chem-
ical and physical analysis to prevent costly development of un-
satisfactory materials. The acute toxic response of materials
comes from the presence and leachability of toxic substances.
Therefore, the detection of leachable substances should be the
principal focus of the test systems. Tests can be performed di-
rectly on the material, on the device, or on the extracts.
If a leachable substance is contained and cannot migrate
to the surface at the time of the testing, the results will appear
to be satisfactory for a short test. The substance will be ex-
posed and becomes leachable when the polymer degrades after
a longer period of time. Devices intended for longer duration
must be tested for longer periods according to subchronic or
chronic test methods.
Acute toxicity does not necessarily predict potential
harmful effects from longer exposures, multiple exposures, bi-
oaccumulation, biodegradation, delayed responses, carcinoge-
nicity, and reproductive toxicity. The duration of subchronic

TLFeBOOK
Design Validation 111

or chronic studies should be appropriate to the expected dura-


tion of use.
If the FDA requires Level II subchronic or chronic tests,
manufacturers should consult the CDRH’s Office of Device
Evaluation before running the test programs.

TESTS TO DEMONSTRATE
BIOCOMPATIBILITY
Cytotoxicity and Cell Cultures
Cell culture tests including cytotoxicity are a good predictor
of biocompatibility when used together with other appropriate
tests. Several highly specialized cell culture methods are
available to monitor the biocompatibility of the raw materials
used in the manufacturing of the device or auditing of the
manufacturing process. Cell or tissue culture testing offers
several advantages:

• It is simple, rather inexpensive, and easy to perform.


• It allows testing of a biomaterial on human tissue.
• It is sensitive to toxic material.
• It is easy to manipulate and allows more than one end-
point investigation.
• It can be used to construct a dose-response curve.
• It can give quick and quantitative results and allows
direct access or direct observation or measurement.

Although cell or tissue cultures offer many advantages,


their use is limited to screening the biomaterials. Therefore,
cytotoxicity results should be used for biocompatibility only in
conjunction with other tests.
Many tissue culture methods are available for testing of
biomaterials. These are divided into two major groups: One
tests the toxicity of a soluble extract of the material, and the
other tests the toxicity by the direct contact of cells with the
material or components of the device.

TLFeBOOK
112 Teixeira and Bradley

Examples of cytotoxicity test methods using extracts in-


clude

• Fluid medium tissue culture that evaluates the cellu-


lar damage caused by the test extract on a confluent
monolayer culture. The test extract is incorporated
into the culture medium. The toxic effect on the mono-
layer, such as cell lysis and microscopic observation of
cell morphology changes, is usually checked after 24
and 48 hours. Cell lysis can be scored by direct micro-
scopic observation or with the use of radiolabels or
blue-dye uptake.
• The inhibition of cell growth is a more informative test
requiring more time and skill. Distilled water extract
is incorporated into the tissue culture medium and in-
oculated with the cells in the tissue culture tubes.
After 72 hours the extent of cell growth is determined
by total protein assay on the cells removed from the
individual tubes.
• The cloning efficiency assay is even more informative,
sensitive, and quantitative and requires even more
skill. The cloning efficiency assay’s procedure and end-
point is similar but is more accurate, sensitive, and
direct than the cell growth inhibition or fluid medium
method. The cloning efficiency assay normally uses a
Chinese hamster ovary cell line and a single-cell clon-
ing technique to estimate the toxic insult-induced re-
duction in cloning efficiency. The cytotoxic effect of the
extract is determined by measuring the ability of the
treated cells to form colonies during seven subsequent
days of incubation. The cloning efficiency of the
treated cultures is compared to that of the control. The
agar overlay method can be used to evaluate the toxic-
ity of the extracts, but it is primarily used for the di-
rect-contact cytotoxicity tests of the solid test sample.

TLFeBOOK
Design Validation 113

Several tests are available to test cytotoxicity by direct


contact. These include

1. ASTM F 813 Practice for Direct Contact Cell Culture


Evaluation of Materials for Medical Devices
2. ASTM F 1027 Standard Practice for Assessment of
Tissue and Cell Compatibility on Prosthetic Materi-
als and Devices
3. NIH Publication No. 85-2185 Guidelines for Blood–
Material Interactions
4. HIMA Report Guidelines for the Preclinical Safety
Evaluation of Materials Used in Medical Devices
5. Others including many device specific toxicity guid-
ance documents on toxicity testing
In addition, the agar overlay tissue culture method and
fluid medium tissue culture method can be used for direct-con-
tact cytotoxicity testing. In the fluid medium method the test
material or device is placed directly on the growing monolayer
cell surface. In the agar overlay method the solid test sample
is placed on or in the agar layer containing vital stain such as
neutral red over the growing monolayer of cells. The response
is evaluated grossly and microscopically and graded according
to the zone index, the size of the cytopathic area, the lysis in-
dex, and percent of cell. (See Table 2) Proper cytotoxicity test-
ing must include at least one test with extract and one direct-
contact test.
In addition, differentiated cells are used to evaluate the
effects materials may have on specific tissues. Differentiated
cells are generally nonfibroblastic cells that are different from
transformed and fibroblastic cell lines such as L-cells used in
ASTM cytotoxicity test methods. Differentiated cells have or-
gan-specific or tissue-specific functions and have specific bio-
logical endpoints or measurable characteristics. Liver cells,
which are differentiated cells, have all or some liver func-
tions.

TLFeBOOK
114 Teixeira and Bradley

Table 2 Cell Lysis Zone Index


Lysis index Extent of cell lysis
0 None observable
1 Less than 20%
2 Less than 40%
3 Less than 60%
4 Greater than 60%
Zone Index Description of zone
0 No detectable zone
1 Limited to area under sample
2 Less than 0.5 cm from sample
3 Less than 1.0 cm from sample
4 Greater than 1.0 cm

Note: A score other than 0 should be studied further.


A score greater than 2 should be evaluated for poten-
tial harm.

Most cells in culture are fibroblasts. Primary cells that


are taken directly from an animal are difficult to establish in
culture and become fibroblasts, losing the normal functions of
growing differentiated cells. Numerous conditions have to be
optimized for obtaining good growth in differentiated cells.
Most cultured cells have a fibroblastic appearance, although
they may not be fibroblasts. The fibroblasts in culture can take
over cultures because they readily grow on plastic surfaces.
The recent success in growing differentiated cells was par-
tially due to technique that have been developed to remove
and limit the growth of fibroblasts to allow other cells to grow.
The properties of the cell cultures usually depend on the culti-
vation conditions, and normal cells grow in culture only for a
limited number of generations.
The test in differentiated cells is important for at least
two reasons. First, the tissue-type specific features of differen-
tiated cells may modulate the effects of chemicals on the fun-
damental properties of cells. Second, it is important to deter-
mine the effects of chemicals on specific cell function
responses.

TLFeBOOK
Design Validation 115

Many companies consider that the in-vitro cytotoxicity


test correlates well with other screening acute toxicity tests
such as USP VI, but is more sensitive. The FDA considers that
cytotoxicity tests are good replacements for the systemic test
in mice, which the FDA no longer supports. All materials, in
general, should be subjected to a cytotoxicity test, which is the
in-vitro test intended to predict possible toxicity. The failure
of a device or material to pass an in-vitro tissue culture test
needs thorough review.

USP Biological Tests


Several USP biological reactivity tests are designed to test the
suitability of materials intended for use in fabricating contain-
ers or accessories for parenteral preparations test the suitabil-
ity of polymers for medical use in implants.
• The systemic injection test and intracutaneous test
are designed to determine the biological response of
animals to plastics by single-dose injection of specific
extracts prepared from a sample.
• The implantation test is designed to evaluate the reac-
tion of living tissue to the plastic by the implantation
of the sample into animal tissue. The use of the sys-
temic injection test in mice is no longer recommended
by the FDA’s CDRH.
• The intracutaneous test is designed to score the in-
flammatory reaction that may occur after the test ex-
tract has been injected into the rabbit’s skin. The in-
tramuscular implantation in rabbits macroscopically
examines the area of the tissue surrounding the center
portion of each implant strip.
Many manufacturers use the USP Class VI test for dem-
onstrating biocompatibility to the FDA. Class VI requires ex-
tracts using sodium chloride, alcohol in sodium chloride, poly-
ethylene glycol, and vegetable oil. Class IV requires one less
extraction, which is polyethylene glycol. Even though the

TLFeBOOK
116 Teixeira and Bradley

Class VI is a well-recognized predictor of biocompatibility, it


may have little relevance to the biocompatibility evaluation of
the intended use of the device. Other tests for biocompatibility
can be used to resolve specific questions regarding intended
use.

Irritation Tests
Intracutaneous Tests
The intracutaneous irritation Test is a sensitive acute toxicity-
screening test and is generally accepted for detecting potential
local irritation by extracts from a biomaterial. Extracts of ma-
terial obtained with nonirritating polar and nonpolar extrac-
tion media are suitable, and sterile extracts are desirable.
Rabbits are the most commonly used animals. A more detailed
description can be found in Biological Reactivity Tests, In
Vivo, Intracutaneous Test, and ASTM F749 Standard Practice
for Evaluating Material Extracts by Intracutaneous Injection
in the Rabbit. The irritation results are evaluated on a scale
of 0 to 4, from “no erythema” (0) to “severe” (4). A similar scale
is used for edema. The result is the average value obtained
from five animals. Any score greater than 2 should be evalu-
ated for potential harm. The methods of testing primary (skin)
irritant substances can be found in the Code of Federal Regu-
lations: 16 CFR 1500.40, 1500.41, and 1500.42.
Skin irritation testing is performed to demonstrate the
potential toxicity of the device, that is, initiating or aggravat-
ing damage through its contact with the skin. Primary skin
irritation is usually done according to the regulations of the
Consumer Product Safety Commission, Title 16, Chapter II,
Part 1500. The purpose of the test is to determine the dermal
irritation potential of the article to the intact and abraded skin
of the rabbit.

Dermal Sensitization Study (Animal Study)


Dermal sensitization is performed to demonstrate the poten-
tial of the device for eliciting an immunological response

TLFeBOOK
Design Validation 117

through its contact with the skin. This reaction is due primar-
ily to substances that could leach out of a material. Guinea
pigs are used because they have been shown to be the best
animal model for human allergic contact dermatitis.*

Modified Draize Test (Human Study)


John Draize of the FDA developed the Draize test. The test
uses albino rabbits to test cosmetics, toiletries, pesticides,
drugs, and other products. The test was designed to assess the
eye irritancy of chemicals and mixtures of chemicals. The test
results are scored according to Title 16, Chapter II, Parts
1500.41 and 1500.3(c)(4).

Ocular Irritation tests


The requirements for the USP Eye Irritation Test and Three
Week Ocular Irritation Test in Rabbits can be found in the
FDA’S Guidance Document for Class III Contact Lenses. This
document prepared by the Office of Device Evaluation can be
obtained through the Division of Small Manufacturers Assis-
tance (DSMA).

Sensitization Tests
Sensitization tests are used to determine adverse effects medi-
ated by immunological mechanisms from material after re-
peated or prolonged exposure. Sensitization is a delayed hy-
persensitivity reaction and is manifested in a variety of
clinical complications. Hypersensitivity is the condition of a
primed individual who tends to give an exaggerated immuno-
logical response on further exposure to the relevant antigen.
When an individual has been immunologically primed or sen-
sitized, further contact with the antigen can not only lead to
secondary boosting of the immune response but can also cause

* ASTM Standard F-720 Standard Practice for Testing Guinea Pig Contact
Allergens, Guinea Pig Maximization Test.

TLFeBOOK
118 Teixeira and Bradley

tissue-damaging reactions. There is no test to reliably and ac-


curately predict the sensitization potential of biomaterials.
However, the ASTM Standard Practice for Testing Guinea
Pigs for Contact Allergens; Guinea Pig Maximization Test
(ASTM F720) is widely used to determine the potential for a
substance or material extract to elicit contact dermal allergen-
icity. Immediate allergy skin testing (IAST) is a very useful
and sensitive hypersensitivity test. Two such in-vitro tests,
the lymphocyte transformation test and inhibition of macro-
phage migration, have been used as an alternative to the skin
patch testing.

Hemocompatibility Tests
The term “hemocompatibility” should be used only in a defined
context. Few materials have consistently shown good hemo-
compatibility in both arterial and venous blood-flow environ-
ments. Results obtained from laboratory animals may not
apply to man, and results from one test system cannot be cor-
related to those obtained from a different test system. Any
hemocompatibility statement should be linked to the intended
use and conditions for which the statement is valid.
The blood–material interaction can range from minimal
protein adsorption to activation of coagulation, complement,
and destruction of cells. Complicated mechanisms exist in the
cardiovascular system that interacts with medical devices.
Devices vary enormously in type, function, and duration of
blood contact. Therefore, a multidisciplinary approach to hem-
ocompatibility testing is important. This includes in-vitro
static and dynamic tests, acute extracorporeal tests, tests of
cardiovascular devices in animals, and clinical studies. This
section will be limited to static in-vitro tests. For other, more
specialized tests, manufacturers should consult with hemo-
compatibility experts or the Office of Device Evaluation (ODE).

Hemolysis Tests
Hemolysis tests evaluate the acute in-vitro hemolytic proper-
ties of materials, especially those led for use in contact with

TLFeBOOK
Design Validation 119

blood. The concentration of substances that produce hemolysis


is generally higher than that needed to produce a cytotoxic
effect. The results of hemolysis testing can be correlated with
acute in-vivo toxicity tests. A hemolysis test is rapid, requires
simple equipment and easily interpretable quantitative re-
sults, and can be performed in the presence of the material
of the extract. The results are compared to the controls and
expressed as percent hemolysis. The ASTM F756 Standard
Practice for Assessment of Hemolytic Properties of Materials
can be used to measure hemolytic potential.

Erythrocyte Stability
The erythrocyte stability test provides a sensitive measure of
the interaction of leachable substances with the plasma mem-
brane of erythrocytes and is reflected as changes in the os-
motic fragility of the erythrocytes. This test can detect leach-
ables at concentrations slightly below the levels of many
cytotoxicity systems.

Protein Adsorption
The adsorption of plasma protein is generally the first event
that occurs when blood contacts a foreign surface. This protein
layer has a great influence on the thrombogenicity of a mate-
rial. One of the more commonly used techniques is the radio
labeling of protein with 125 I. This technique provides a direct
measurement of the amount of protein adsorbed on a surface.

Implantation Tests
Two examples of implantation tests are the Standard Practice
for Short-Term Screening Implant Material (ASTM F763) and
Standard Practice for Assessment of Compatibility Biomateri-
als (Nonporous) for Surgical Implants with Respect to Effect
of Materials on Muscle and Bone (ASTM F981).

Mutagenicity Tests
Genetic and cell culture techniques are available to test muta-
genic properties to demonstrate the presence or lack of ability

TLFeBOOK
120 Teixeira and Bradley

of the test material to cause mutation or chromosomal damage


or cancer. The material intended for intimate contact and long
exposure should not have any mutagenic properties. The pres-
ence of nonpolymerized materials and traces of monomers,
oligomers, additives, or biodegradation products can cause
mutations.
Mutation can be either a point mutation or a chromo-
somal rearrangement caused by DNA damage. Therefore, the
material’s ability to cause point mutation, chromosomal
change, or evidence of DNA damage should be tested. Correla-
tions exist between mutagenic and carcinogenic properties.
Most, if not all, carcinogens are mutagens, but not all muta-
gens are human carcinogens.

• The Ames salmonella/microsome test is a principal


sensitive mutagen-screening test. Compounds are
tested on the mutants of Salmonella typhimuriun for
reversion from a histidine requirement back to proto-
trophy. A positive result is seen by the growth of re-
vertant bacteria. A microsomal activation system
should be included in this assay. The use of all five
bacterial test strains is generally required. Two non-
bacterial mutagenicity tests are generally required to
support the lack of mutagenic or carcinogenic poten-
tial.

Pyrogenicity Tests
Pyrogenicity testing is generally required if

• The label claims that the device is pyrogen-free.


• The device comes in contact with blood or spinal fluid.
• The device is an intraocular lens.

Reproductive and Developmental Toxicity


Irreversible toxic effects are associated with adverse reactions
of the reproductive system and in the offspring. Reproductive

TLFeBOOK
Design Validation 121

anomalies range from sterility to problems in the production


of the mother’s milk. The adverse influence may occur at any
time within the reproductive cycle of the organism. Toxic ef-
fects to the offspring range from mortality to morbidity as sub-
tle as decreased body weight at birth. Pronounced reproduc-
tive anomalies are teratic or morphological defects observed
in offspring at birth. Teratology has expanded to include not
only morphological alterations but also biochemical, immuno-
logical, and behavioral deficits.

Carcinogenesis Bioassays
The carcinogenesis bioassays observe the test animal over a
major portion of its life span for the development of neoplastic
lesions. Because of the long latent period required for induc-
tion and manifestation of tumors, it is necessary to expose the
test animals when young and continue for duration of the ex-
periment. The organs and tissues of both treated and control
animals should be examined for morphological changes. The
absence of a carcinogenic effect cannot be assured unless all
organ systems have been examined with high accuracy in all
animals, and all grossly visible suspect lesions examined mi-
croscopically. More comprehensive carcinogenesis protocols
can be found in the FDA’s Toxicological Principles for the
Safety Assessment of Direct Food Additives and Color Addi-
tives Used in Food.

TLFeBOOK
TLFeBOOK
9
Process Validation

IT’S NO GOOD IF YOU CAN’T MAKE IT

Strictly speaking, process validation is not part of design con-


trols. But the best product design in the world is useless if
it can’t be made in a consistent and efficient way and at an
appropriate cost. At least it’s not part of Subpart C on design
controls although design validation must occur on initial pro-
duction units or their equivalent. So it would make little sense
to get that far along in the developmental process without hav-
ing developed and validated the process by which the product
is going to be made. In fact, the very next step after design
validation in formal design control is design transfer, which
involves getting everything out of development and moving it

123

TLFeBOOK
124 Teixeira and Bradley

into manufacturing. The time to start worrying about how the


product will ultimately be made is Day 2 of the development
program; the day after the product is designated as a probable
commercial product, not the day you start the design valida-
tion run.
Back at the beginning of the book, we mention in passing
that very few people would go to a bagel store to buy an emer-
ald ring. Well, there are a few jewelers that you probably
wouldn’t go to either. In order to make a perfect emerald ring,
you need more than a good natural emerald and some gold.
You need a manufacturing process. Natural emeralds are hard
to find, and once you’ve found one that is valuable, you don’t
want someone who’s cutting emeralds for the first time work-
ing on your particular stone. You want to have a system that
will cut the emerald correctly the first time it is struck. You
also want it polished and set correctly. Therefore, you wouldn’t
turn the gem over to someone who has never cut and polished
a natural emerald before. Neither should you attempt to man-
ufacture a medical device without first knowing how you in-
tend to make it and having the assurance that the process will
produce the product according to the specifications set during
the design control process.

THE FDA AND PROCESS VALIDATION

The FDA does talk about process validation in 21 CFR 820,


not in Subpart C for Design Controls, but in Subpart G, Pro-
duction and Process Controls. In Section 820.75 it says

(a) Where the results of a process cannot fully be verified


by subsequent inspection and test, the process shall
be validated with a high degree of assurance and ap-
proved according to established procedures. The vali-
dation activities and results, including the date and
signature of the individual(s) approving the valida-
tion and where appropriate the major equipment val-
idated, shall be documented.

TLFeBOOK
Process Validation 125

(b) Each manufacturer shall establish and maintain pro-


cedures for monitoring and control of process param-
eters for validated processes to ensure that specifi-
cation requirements continue to be met.
(1) Each manufacturer shall ensure that validated
processes are performed by qualified individuals.
(2) For validated processes, the monitoring and the
control methods and data, the date performed, and,
where appropriate, the individual(s) performing the
process or the major equipment used shall be docu-
mented.
(c) When changes or process deviations occur, the manu-
facturer shall review and evaluate the process and
perform revalidation where appropriate. These activ-
ities shall be documented.

A model operating procedure for process control is shown in


Appendix G.

WHAT DO YOU CALL A GROUP


OF PROCESSES?

If you can have a litter of puppies or a pride of lions, what do


you call a group of processes strung together to make a medi-
cal device? Manufacturing, of course! In the manufacturing of
many medical devices, several different operations are tied to-
gether to make the device. Let’s look in Figure 1 at the manu-
facturing process for making a colostomy pouch.
Although oversimplified, Figure 1 shows the need for vali-
dation of this manufacturing process, or at least some of the
five subprocesses that are part of this manufacturing setup.
We have identified five separate processes necessary for the
manufacture of a relatively simple medical device. This device
is actually a class I product, so it is not necessary to establish
a design control program for it in the first place. But as we
said before there are many reasons, including quality and the
company’s bottom line, for doing so.

TLFeBOOK
126 Teixeira and Bradley

Figure 1 Typical manufacturing process for a one-piece colostomy


pouch.

The first question we need to ask is which, if any, of these


processes should be validated? The FDA says, “Where the re-
sults of a process cannot fully be verified by subsequent in-
spection and test, the process shall be validated.”
The first step of the overall manufacturing process is the
mixing of the medical-grade adhesive that serves as the means
for holding the pouch on the patient and protecting the pa-
tient’s skin. Typically such an adhesive formulation might
contain up to a dozen separate ingredients.* These ingredients
are mixed together in separate batches according to a proce-
dure that specifies the order of addition, the temperature at
critical points in the mixing, and the length of time the mixing
is allowed to run for each addition of raw materials or group
of raw materials. Change any of those things beyond a certain
range and the resultant adhesive may have different proper-
ties.
Assume that the development work has led to an adhe-
sive with an ideal set of properties for the application and that
this was verified through standard adhesive testing such as

* If the adhesive is not an acrylic or a urethane but is the typical “hydrocol-


loid” adhesive used in many ostomy pouches.

TLFeBOOK
Process Validation 127

tack, shear, and peel adhesion.* Assume further that this was
validated by skin adhesion tests on volunteer subjects and
with an actual clinical use test of the device. How do you know
whether each batch of material made in production is the
same as the adhesive made and tested during the development
of the product? Can you test (verify) that each batch is the
same as the one before and after and that each of these is what
was developed and verified previously? The answer is “yes,”
maybe. An appropriate sample of each batch can be taken
when the mix is completed and subject to a battery of adhesion
tests in order to verify that they meet the specifications that
were developed. If you have a good correlation between these
in-vitro tests and in-vivo results, you have verification, thus
validating that this subprocess is not necessary. There are,
however, several potential problems. First, the correlation be-
tween adhesion in the lab and adhesion to people probably
isn’t as good as you think. The same person using the same
adhesive batch may have different results from day to day.
Diet, drugs, exercise, ambient temperature, and humidity
change things. Stainless steel, the most often used test sub-
strate, does not undergo these changes. Having human sub-
jects hanging around to act as test substrates doesn’t seem
quite practical either. Second, the batch mixing process is slow
compared to the other processes that come after it. It may not
be practical to have everyone wait for the slowest Boy Scout
marching† along in the line before proceeding with the rest of
the process. So maybe a validated mixing process is a benefit.
Of course, there is at least one way to verify that the batch
is consistent with the product specifications. During the devel-
opment a correlation between both the in-vivo and in-vitro ad-

* Let’s assume the testing was done at 40°C to simulate body temperature
and we weren’t misled by results observed at room temperature.
† If you want to know where the Boy Scout came from and about optimizing
production, read The Goal by Goldratt and Cox, North River Press, 1984.

TLFeBOOK
128 Teixeira and Bradley

hesion results could have been made with some property in-
trinsic to the material. If the right instrumentation is
available, this method could be used to verify that each batch
of medical-grade adhesive made according to the established
procedure and containing the right amounts of all the required
materials meets the product specifications.*
All in all, it may be a good idea to validate this batch
subprocess. It would allow a smoother flow into the rest of the
manufacturing steps for a subprocess that is already time-con-
suming. Here’s what can be done:
1. For each of the ingredients one needs to determine
the added range beyond which the formula no longer achieves
the specified properties using a constant mix procedure. When
this allowable variation for each ingredient is known, the mix
process itself can be validated.
2. Using a designed experiment,† the process parame-
ters (time, order of ingredient addition, temperature) should
be evaluated at several points above, below, and at the opti-
mum conditions set during the development. This set of condi-
tions encompassing upper and lower processing limits and cir-
cumstances, including those within standard operating
procedures, pose the greatest chance of process or product fail-
ure when compared to “ideal” conditions. Remember that such
conditions do not necessarily induce product or process failure.
3. Each batch should then be carefully tested according
to standard test methods and the results statistically evalu-
ated to determine the parameters of the process. A multiple
regression analysis of the results may even illuminate which
parameter is more critical than the others.
Now what about the extrusion? Validate or rely on verifi-
cation. Here the answer depends primarily on what your ex-

* If you don’t know how to do this, we’re not telling you. We’re consultants,
remember? All you have to do is hire us to find out.
† A good example of the value of DOE in process validation can be seen in
M.J. Anderson and P.J. Anderson, Design of experiments for process valida-
tion, Medical Device & Diagnostic Industry, January 1999, pp. 193–199.

TLFeBOOK
Process Validation 129

truder has in terms of inline inspection instrumentation or


how much you would be willing to spend to equip the unit to
make the necessary measurements while the extruder is run-
ning. Many of the physical properties of the exudate such as
width and thickness can be quickly and easily made offline
using appropriate sampling techniques during a run. At the
same time inline measurement gauges are readily available
and continuously monitor thickness and keep necessary docu-
mentation. Remember that the use of such equipment will re-
quire validation of that component. But what about the effects
of the extrusion temperature profile, the residence time in the
barrel (screw speed), and the backpressure? Assuming the de-
velopment work was done correctly, these parameters were
set and are all interrelated. So the product of the extrusion
could be easily handled with simple testing. The exception is
the extrusion of a heat-sensitive material that significantly
changes properties depending on its heat history. In this case
the extrusion subprocess should also be validated in a manner
similar to that as outlined in the batch mixing subprocess.
The laminating process seems to be more easily handled
through verification that the lamination has developed suffi-
cient bond strength between the layers to prevent separation
during anticipated use conditions. For large-scale, multishift
operations a case could be made for validation of the process
and control of the specified result by evaluating the nip roll
pressures and temperatures as well as the line speed through
the unit. One could then arrive at a set of conditions that
would allow reasonable assurance that the specified lamina-
tion was accomplished as long as the parameters remained
within the validated values.
The remaining die cutting and bonding* subprocesses are
easily verifiable against product specifications using standard
test methods, a proper sampling technique, and appropriate
maintenance of the cutting dies. A strong case could be made

* These techniques include RF welding, heat sealing, or ultrasonic welding.

TLFeBOOK
130 Teixeira and Bradley

for the validation of any bonding machine such as an ultra-


sonic welder especially if this bonding occurs offline. If the
bonding of the pouch outline weld was made through an RF
weld, verification would be absolutely necessary whether or
not the RF generator was validated, as RF welding in such
an application is known to cause pinholes in the film due to
arcing.
In general it should be remembered that

• All processes vary.


• Some processes vary more than others.
• Each variation has a cause, and many of these causes
can be identified.
• To the extent that the causes can be identified and
understood, they can be controlled.
• Only a few of the causes are significant; the greater
the extent of control on these critical causes, the better
control there is of the manufacturing process.
• The product design must be flexible enough to take up
the remaining process variations.
• If it can’t be manufactured efficiently with a high level
of reproducible quality at a cost that will allow a profit,
then you haven’t got a product.

Once a process is validated, the need for inspection and testing


is decreased dramatically and manufacturing can take advan-
tage of statistical process control (SPC) methods to increase
efficiency and throughput.

STATISTICAL PROCESS CONTROL


Processes and Process Variability
The concept of process variability, as outlined in the previous
section, is the basis of statistical process control. As an exam-
ple, suppose a marksman shoots 100 rounds at a target every
day. He would not get the same number of shots at the center

TLFeBOOK
Process Validation 131

of the target during each daily session. Some days he would


get 89 of 100, some days 73 of 100, some days 84 of 100, etc.
All manufacturing processes have this kind of variation.
This process variation can be divided into two compo-
nents. Natural process variation, or system variation, is the
naturally occurring variation inherent in all processes. In the
case of the shooter, this variation would vary around his long-
term percentage of bulls-eye’s hit. Special cause variation is
caused by some problem or unusual occurrence in the system.
In the case of our shooter, an eye infection would cause him
to miss a larger than “normal” number of shots on that day.
Statistical process control is a method for graphically rep-
resenting the process and determining when a process is “out
of control.” A process is out of control by this definition not
because of its normal or inherent variation but because a spe-
cial cause variation is present. Something unusual is oc-
curring in the process. The process can then investigated to
determine the base cause of this out-of-control condition.
When the cause of the problem is determined, an action can
be identified to correct it. The investigation and corrective ac-
tion are a team process.
It is management’s responsibility to reduce system varia-
tion as well as special cause variation. This is often accom-
plished through

• Process improvement techniques


• Investment in new technology
• Reengineering the process to have fewer steps and
therefore less variation

Reduced variation makes the process more predictable,


with process output closer to the specified value. This goal of
minimal variation mandates working toward the goal of re-
duced process variation.
The process in Figure 2 is in apparent statistical control.
All points lie within the upper control limit (UCL) and the
lower control limit (LCL). This process is showing only normal

TLFeBOOK
132 Teixeira and Bradley

Figure 2 A process in statistical control.

system variation. The process charted in Figure 3 is out of


statistical control. A single point can be found outside the con-
trol limits. This means that a special occurrence likely caused
the variation above the upper control limit. There is small
probability that this variation happened by chance, so when
a point is found outside the control limits, it is very probable
that a source of special cause variation occurred. This needs
to be isolated and corrected. A single point outside the control
limits is easily detectable as an out-of-control condition.

Figure 3 A process out of statistical control.

TLFeBOOK
Process Validation 133

Figure 4 A process improvement cycle.

Figure 4 illustrates an SPC improvement cycle. The pro-


cess begins out of specification (out of control), then, as causes
of variation are found and corrective action is implemented,
the process comes into statistical control. Through continuous
process improvement, variation is reduced and a new specifi-
cation representing a higher-quality product can be intro-
duced. Eliminating special cause variation keeps the process
in control; process improvement reduces the process variation
and moves the control limits toward the center of the product
specification.

Control Chart Zones


Control charts can be broken into three zones—A, B, and C—
on each side of the process center specification. This is illus-
trated in Figure 5. Rules exist that are used to detect condi-
tions in which the process is behaving in an out-of-control con-
dition.

TLFeBOOK
134 Teixeira and Bradley

Figure 5 SPC control chart zones.

• The probability of having two out of three consecutive


points either in or beyond zone A is an extremely un-
likely occurrence when the process mean follows a nor-
mal distribution.
• The probability of having four out of five consecutive
points either in or beyond zone B is also an extremely
unlikely occurrence when the process mean follows
the normal distribution.
• The probability of having eight or more consecutive
points either above or below the centerline is also an
extremely unlikely occurrence when the process fol-
lows the normal distribution.
• The probability of six or more consecutive points show-
ing a continuous increase or decrease is also an ex-
tremely unlikely occurrence when the process follows
the normal distribution.
• The probability of having 14 or more consecutive
points oscillating back and forth is also an extremely

TLFeBOOK
Process Validation 135

unlikely occurrence when the process follows the nor-


mal distribution.
• The probability of having eight or more consecutive
points occurring on either side of the centerline and
not entering zone C is also an extremely unlikely oc-
currence when the process follows the normal distri-
bution and signals an out-of-control condition.*
• The probability of having 15 or more consecutive
points occurring outside zone C is also an extremely
unlikely occurrence when the process follows the nor-
mal distribution and tells you that you have an out-
of-control condition.

* This occurs when more than one process is being followed on the same
chart, when using improper sampling techniques.

TLFeBOOK
TLFeBOOK
10
Design Transfer

WHAT IS DESIGN TRANSFER?

When design controls first made their appearance, many peo-


ple looked at the section concerning design transfer and
pointed to it as proof of the bureaucratic nature of the new
regulation. What else would you do with a newly developed
product but transfer it to production? The problem is that it’s
not what you do at this point in the development but how—
and sometimes even when—you do it. The purpose of the
design transfer phase is to make sure design outputs are
adequately and correctly translated into production specifica-
tions.

137

TLFeBOOK
138 Teixeira and Bradley

CHECK YOUR ATTITUDE AT THE DOOR

We are sure that nothing we are about to say in this section


applies to your company, but you know it happens at Company
“AB,” and of course everyone has heard of the internal compe-
tition at Corporation “BA”! Design transfer is probably the
most obvious phase of the design control cycle, and that is the
root of the problem. Senior management causes the other fun-
damental problem of internal competition whether it does it
consciously or unconsciously.
If management does not have a sincere commitment to
the design control process and the teamwork and high quality
the process represents, then this transfer, while obvious, be-
comes the source of fundamental problems. If management be-
lieves that some level of internal competition between func-
tional groups brings out the best in everybody, then no one
should be surprised when the product that was developed and
looked so good in clinical evaluations isn’t quite the one that
actually got launched and is in inventory. It may look the same
and have the same components, but somehow it isn’t quite
what it was.
Don’t blame this on the fact that the product inevitably
loses a little when it gets into production. If the design control
process was followed and management created a team that
took possession of the device from beginning through produc-
tion, then it will not have changed.
Poor transfer happens more often than you might think.
One of the scenarios for this poor transfer often starts with
manufacturing personnel and is tacitly agreed to by general
management. If there is a process engineering presence in the
company, it almost always happens. Management decides
that the participation of manufacturing personnel on the de-
sign team needs to be either limited or, worse, not at all. The
rationalization for this is usually something like, “We are
working three shifts a day and overtime. Manufacturing peo-
ple simply don’t have the time, and the company can’t afford

TLFeBOOK
Design Transfer 139

to take them away from their jobs for a meeting. They’re too
critical!”*
During the design process the manufacturing people hear
all sorts of things about the new product, and most of them
are out of context. Sometimes design engineers ask them ques-
tions, by design engineers and a mental picture begins to form
of a new product that will be difficult to integrate into manu-
facturing at best and a nightmare at worst. As time goes on
the manufacturing people begin to form a defense against this
new product that often takes the form of, “When it gets to man-
ufacturing, we’ll fix it.”
Never happens, right? It happens. Here are two instances
that really occurred. Both were at well-known and well-re-
spected companies. In the first case, a device was developed
in-house, and R & D entirely handled the technical side of the
development. The manufacturing process was designed, de-
bugged, and validated by process engineering. The product
was virtually demanded by the customers, and so its success
was guaranteed. The first manufacturing line was built in-
house and tested, and the validation was without significant
problems. The machine was transferred to the manufacturing
plant. For a period of three weeks on-specification product was
made while the manufacturing personnel were trained. Pro-
cess engineering stayed with manufacturing and the product
for the entire three-week period. After everyone signed off his
or her approval, the process engineers left. Within one month
of the transfer, all hell broke loose. A critical design dimen-
sional specification could not be held and manufacturing
sought approval to loosen the specification. When the original
process design engineers went to the facility, the reason for
the sudden inability to hold the specification was apparent.

* The other extreme is, “We have problems with costs and quality. Manufac-
turing people simply don’t have the time, and the company can’t afford to
take them away from their jobs for a meeting. They’re too critical!”

TLFeBOOK
140 Teixeira and Bradley

The manufacturing engineers had redesigned the production


equipment to increase efficiency and save floor space.* In do-
ing so they cut the framework of the machine to rearrange its
components. From that point on, the warped frame made it
impossible to hold the dimensional specification. In another
instance an outside source designed and built the manufactur-
ing process. The line was installed and validated. Everyone
was in heaven except, as it turned out, some of the manufac-
turing people didn’t like the design of the machine and
thought they had a better idea. So the machine began to act
erratically. The cause for this sudden erratic behavior in the
process was that someone in the plant disconnected a circuit.†
The people responsible for the sabotage said they knew the
machine was no good and they hoped that when it failed they
would get the chance to build their own.‡
How could this happen? The reason is simple really. The
manufacturing people were never involved; there was no own-
ership of the product from their point of view. Whose fault was
it? That answer is also simple. It was the fault of the senior
managers. It was their responsibility to ensure quality. It was
senior management’s responsibility to build the team. And one
more important point, although the two examples happened
to put manufacturing folks in a poor light: It could and does
happen with other disciplines also. Manufacturing was used
as the example because they are the one’s most often left out.

THE FDA AND DESIGN TRANSFER

The FDA says, “Each manufacturer shall establish and main-


tain procedures to ensure that the device design is correctly

* They had a legitimate floor space problem. Business was growing so fast
there was no room.
† By the way, the cause was uncovered by SPC due to the peculiar nature
of the out-of-specification data points. It told the engineers that someone
either had changed the machine or was faking the data.
‡ Yes, they were terminated.

TLFeBOOK
Design Transfer 141

translated into production specifications” [21 CFR Part 820,


Subpart C, Section 820.30(h)]. There is no point in beating this
one to death. It’s simple and concise.

DESIGN TRANSFER REQUIREMENTS

Before we move on, it seems appropriate to list the require-


ments for an effective design transfer. A typical material spec-
ification and its corresponding quality specification are shown
in Appendices H and I.

• When the design is complete, the design specifications


need to be transferred to manufacturing in the form
of production specifications. As always, the system for
the transfer needs to be defined and documented.
• Production specifications typically consist of written
documents. These documents include
• product and assembly drawings
• material specifications
• inspection (quality) and test specifications
• manufacturing instructions
• training materials
• tooling drawings

In addition, all verification and validation activities


should be completed and all appropriate personnel training
conducted before initial production begins. It is also an ideal
time for a design review to ensure that all aspects of the design
process were reviewed and transferred correctly. Now the
product is ready to launch.

TLFeBOOK
TLFeBOOK
11
Design Changes

THE PURPOSE OF DESIGN CHANGE


CONTROL

Design change control is just one more effort to ensure the


highest quality in the product development process. Anyone
that has ever been intimately involved with the development
of a medical device is aware of how many changes happen be-
tween the original concept and the product that finally reaches
production. If we remember that design control is a cyclic pro-
cess, then it is important to know where we’ve been. For one
thing, knowing where we’ve been tells us where we are, which
in turn tells us where we can go.

143

TLFeBOOK
144 Teixeira and Bradley

Just like the pitfalls that can occur during design trans-
fer, similar events can happen if the design team does not have
control of the changes that occur along the way. If an overzea-
lous design engineer decides to change a dimension without
telling anyone, and worse without noting the change, a disas-
ter can occur. It happens and you know it. Remember the first
law of design controls: Document everything.
The section on design changes in the quality system regu-
lation is simply about documentation. It is the foundation of a
good product development cycle and the cornerstone of design
controls.

THE FDA AND DESIGN CHANGES

In another concise and clear paragraph, the FDA has this to


say about the subject:

Each manufacturer shall establish and maintain


procedures for the identification, documentation, valida-
tion or where appropriate verification, review and ap-
proval of design changes before their implementation. [21
CFR Part 820, Subpart C, Section 820.30(i)]

DESIGN CHANGE REQUIREMENTS

Once you approve your initial design inputs for the design
project, changes must be documented and controlled for the
life of the product, not just during the design controls cycle.
The system for making design changes needs to be defined and
documented in a procedure. This system should address how
changes are identified and documented and how it is deter-
mined whether verification and/or validation is necessary.
The degree of design change control depends on the signifi-
cance of the change and the risk presented by the device.
Changes need to be reviewed and approved by the same
individuals or functions associated with the original design to

TLFeBOOK
Design Changes 145

ensure the device will continue to perform as intended. Design


changes include

• Changes made to the device itself


• Labeling changes

Figure 1 Documentation change request.

TLFeBOOK
146 Teixeira and Bradley

• Performance changes
• Packaging changes
• Changes resulting from complaints, etc.

When a change is made to a specification, method, or pro-


cedure, the manufacturer should evaluate whether the change
requires the submission of a premarket notification or a sup-
plement to a PMA. An evaluation of many small/minor
changes should also be considered in the same light. Records
of this evaluation and the results need to be maintained.
Companies are required to maintain a defined and docu-
mented design change procedure even if they have not com-
pleted any design projects or have no ongoing or planned de-
sign projects or changes. A design change procedure is shown
in Appendix J.

THE DOCUMENT CHANGE REQUEST

The heart of controlling design changes is the document


change request (DCR), shown in Figure 1. This simple form
provides the necessary control by requiring the originator of
any change to think about what he or she is doing and what
other documents, specifications, and tests the change may af-
fect. In addition, the people responsible for the various corpo-
rate functions must review the changes, add any resultant
changes of which they may be aware, and affix their signature
in approval.
By keeping a log of all the DCRs in a central location,
duplication will be improbable and tracing the origins of the
changes to any product can be accomplished.

TLFeBOOK
12
The Design History File

WHY DO WE NEED A DESIGN


HISTORY FILE?

The prime reason for the design history file is to provide a


record or evidence of compliance with design control require-
ments. The design history file is the final section of the design
control topics in the quality system regulation and is by far
the most simple. As manufacturers of medical devices, we are
required to keep a complete record of the development of every
medical device developed under the design control system.
The design history file may be defined as a compilation of the
records (drawings, formulations, test methods, etc.) that de-
scribes the design history of the finished device.

147

TLFeBOOK
148 Teixeira and Bradley

Of course, there are other sound reasons to compile this


historical record of the development beyond the regulatory re-
quirement. Some of these are as follows:

• The historical record will allow subsequent develop-


ment teams to take advantage of things that were dis-
covered by studying objective records.
• Postdevelopment review of the records will provide a
unique vantage point for the continuous improvement
of the product development cycle in general.
• The design history file provides an excellent eviden-
tiary source in the event of patent disputes.
• In the event of unanticipated future product problems
or complaints, the device history file may provide a
valuable resource for solving the problem.

THE FDA AND THE DESIGN HISTORY FILE

In its last words on design control, the FDA addresses the con-
cept of the device history file:

Each manufacturer shall establish and maintain a


DHF for each type of device. The DHF shall contain or
reference the records necessary to demonstrate that the
design was developed in accordance with the approved de-
sign plan and the requirements of this part [21 CFR Part
820, Subpart C, Section 820.30(j)]

DESIGN HISTORY FILE REQUIREMENTS

A design history file needs to be maintained for each device a


manufacturer develops. It contains the records necessary to
demonstrate that the design was developed in accordance with
the approved design plan and the established design controls.
The elements of the design history file include

TLFeBOOK
The Design History File 149

• The plan
• Design inputs
• Design outputs
• Design review records
• Verification records and methods
• Validation protocols and results

TLFeBOOK
TLFeBOOK
13
Questions to Expect in an Audit

AN AUDIT! NOW WHAT?

So you’re about to be audited. What can you expect? First of


all, if you’ve been doing everything correctly, there is not much
to worry about. It will happen sooner or later. If things are
running smoothly, you should have conducted several internal
audits before the FDA shows up for its inspection.

THE FDA DESIGN CONTROLS INSPECTION OBJECTIVES

In the quality system inspection technique (QSIT) Guide to


Inspections of Quality Systems,* the FDA lists the objectives
regarding design controls. There are 15:

* August 1999.
151

TLFeBOOK
152 Teixeira and Bradley

1. Select a single design project. Note: If the project se-


lected involves a device that contains software, con-
sider reviewing the software’s validation while pro-
ceeding through the assessment of the firm’s design
control system.
2. For the design project selected, verify that design
control procedures that address the requirements of
Section 820.30 of the regulation have been designed
and documented.
3. Review the design plan for the selected project to
understand the layout of the design and develop-
ment activities including assigned responsibilities
and interfaces. Note: Evaluate the firm’s conduct of
risk analysis while proceeding through the assess-
ment of the firm’s design control system.
4. Confirm that the design inputs were established.
5. Verify that the design outputs essential for the
proper functioning of the device were identified.
6. Confirm that the acceptance criteria were estab-
lished prior to the performance of verification and
validation activities.
7. Determine if the design verification confirmed that
the design outputs met the design input require-
ments.
8. Confirm that the design validation data shows that
the approved design met the predetermined user
needs and intended uses.
9. Confirm that the completed design validation did
not leave any unresolved discrepancies.
10. If the device contains software, confirm that the
software was validated.
11. Confirm that risk analysis was performed.
12. Determine if design validation was accomplished
using initial production devices or their equivalents.
13. Confirm that the changes were controlled including
validation or, where appropriate, verification.
14. Determine if design reviews were conducted.
15. Determine if the design was correctly transferred.

TLFeBOOK
Questions to Expect in an Audit 153

SOME QUESTIONS YOU MAY BE ASKED

In addition to the material mentioned above, there are some


questions that may arise during an audit.

General Design Controls


• What initiates a design project?
• When does the actual design and development begin
(e.g., design controls)?

Design and Development Planning


• How is each design and development activity identi-
fied and documented?
• How are responsibilities defined?
• Are design and development activities assigned to
qualified personnel?
• Are plans updated as the design evolves?
• Are the organizational and technical interfaces be-
tween different groups that input into the design pro-
cess identified?
• Do procedures exist for the documentation, transmit-
tal, and review of interdepartmental data exchanges?

Design Input
• Do design inputs include customer requirements?
• Do design inputs include applicable statutory and reg-
ulatory requirements for those countries in which you
are intending to market?
• Do design inputs address intended uses, including the
needs of the user and the patient?
• Are design inputs reviewed and approved?
• Does the company have a procedure/method to ad-
dress incomplete, ambiguous, or conflicting require-
ments with those responsible for imposing these re-
quirements?

TLFeBOOK
154 Teixeira and Bradley

Design Output
• Is design output documented and expressed in terms
that can be verified and validated against design input
requirements?
• Are design outputs approved?
• Does design output documentation
• provide evidence that the final design meets input
requirements?
• identify or make reference to acceptance criteria?
• identify characteristics of the design that are cru-
cial to the safe and proper functioning of the prod-
uct (for example, operating, handling, mainte-
nance, storage, and disposal requirements)?

Design Review
• At what stages are formal documented reviews per-
formed?
• Do design reviews include representatives of all func-
tions concerned with the design stage being reviewed
as well as an individual independent of the design
stage being reviewed?
• Are records of design reviews maintained? If so, for
how long?
• Do records include individuals in attendance, date, de-
sign reviewed?
• Do design reviews address, as applicable,
• comparison of customer needs with technical speci-
fications for materials, products, and processes?
• validation of the design through prototype tests?
• considerations of unintended use and misuse?
• safety and environmental compatibility?
• compliance with regulatory requirements, national
and international standards, and corporate prac-
tices?
• comparison with similar designs, especially analy-

TLFeBOOK
Questions to Expect in an Audit 155

sis of internal and external problem history to


avoid repeating problems?
• permissible tolerances and comparison with pro-
cess capabilities?
• product acceptance/rejection criteria?
• manufacturability of the design, including special
process needs, mechanization, automation, assem-
bly, and installation of components?
• capability to inspect and test the design, including
special inspections and test requirements?
• specification of materials, components, and subas-
semblies, including approved suppliers?
• packaging, handling, storage, and shelf-life re-
quirements?
• safety factors relating to incoming and outgoing
items?
• How are problems or action items identified during a
review handled?

Design Verification
• Do verification activities identify the method, date,
and individual performing verification?
• What types of design verification activities were per-
formed?

Design Validation
• What methods were used to validate the design?
• Were the first three production lots tested under ac-
tual or simulated use conditions?
• If design validation is done on nonproduction devices,
how were the devices shown to be equal to production
devices?
• How were unresolved discrepancies handled?
• If the device has software, how was the software vali-
dated?

TLFeBOOK
156 Teixeira and Bradley

Design Reviews
• Are there alternative calculations to verify the correct-
ness of original calculations and analysis (i.e., risk as-
sessment, AMPE, etc.)?
• Are there experimental runs?
• Are there tests and demonstrations (model or proto-
type test)?
• Is there a comparison to similar proven designs?
• Are design verification records maintained as part of
the design history file?
• If output ≠ input, how was discrepancy resolved?

Design Transfer
• How are design specifications translated into produc-
tion specifications?

Design Changes
• When do changes to product design begin to fall under
design control?
• How are design changes controlled?
• Are all design changes identified, documented, re-
viewed, and approved by authorized personnel prior
to implementation?
• Are design changes under document control?
• How do design changes trace back to the initial design
project?

Design History File


• What documents make up your design history file?

TLFeBOOK
Further Reading

Anderson, M.J. and P.J. Andrews, Design of experiments for process


validation, Medical Device & Diagnostic Industry, Jan. 1999,
193–199.

ANSI/ASQC D1160-1995, Formal design review.

Augustine, R., A structural approach to rapid process development


and control, Medical Device & Diagnostic Industry, Jan. 2000,
112–127.

Design control guidance for medical device manufacturers, FDA,


1997.

Dzog, H., Risk management in medical device regulations, Medical


Device & Diagnostic Industry, Oct. 1997, 112–117.

157

TLFeBOOK
158 Teixeira and Bradley

Edwards, F., Before design, thoroughly evaluate your concept, Medi-


cal Device & Diagnostic Industry, Mar. 1997, 46–50.
FDA design input guidance, The Silver Sheet, June 1997, 11–14.
Federal Register, 21 CFR Part 820, 1996.
Final Q.C. inspectional strategy, CDRH, February 1998, 1–10.
Fleischer, M. and J.K. Liker, Concurrent Engineering Effectiveness,
Hanser-Gardner, Cincinnati, 1997.
Giantini, R.E., Developing safe reliable medical devices, Medical De-
vice & Diagnostic Industry, Oct. 2000, 60–67.
Goldratt, E.M. and J. Cox, The Goal: Excellence in Manufacturing,
North River Press, New York, 1984.
Gosbee, J., The discovery phase of medical device design, Medical
Device & Diagnostic Industry, Nov. 1997, 79–82.
Guide to inspections of quality systems, FDA, Aug. 1999.
Knepel, P., Integrating risk management with design controls, Med-
ical Device & Diagnostic Industry, Oct. 1998, 83–88.
Kuhn, M.A., Implementing design controls, Medical Device & Diag-
nostic Industry, Feb. 2000, 40–49.
McCay, B., Design control measures that can boost return on invest-
ment, Medical Device & Diagnostic Industry, May 1998, 158–
164.
Meyers, S.L., Data Analysis for Scientists and Engineers, Wiley &
Sons, New York, 1975.
Munro, B.H., Statistical Methods for Health Care Research, Lippin-
cott, New York, 1997.
Oliveri, D., Developing design control strategies to meet technology
advances, Medical Device & Diagnostic Industry, Sept. 2000,
77–85.
Reich, R., How design controls affect sterilization process develop-
ment & validation, Medical Device & Diagnostic Industry, Oct.
1997, 107–111.

TLFeBOOK
Further Reading 159

Rubin, I.I., Product design, Plastics Engineering, April 2000, 101–


106.
Sheratt, D., Taking a risk based approach to medical device develop-
ment, Medical Device & Diagnostic Industry, Sept. 1999, 84–
95.
Teixeira, M.B., Design Controls Training Module, QARA Compli-
ance Connection, Odessa, NY, 2000.

TLFeBOOK
TLFeBOOK
Appendix A
Implementation Procedure

Effective
Rev. DCR* Date Date Originator Description
A 07-20-98 08-14-98 M. Teixeira DCR 98-0131
Approval signature: Date:

PURPOSE

The requirement to initiate a design control project may arise


for a variety of reasons such as the identification of a new prod-
uct, a marketing need to satisfy a customer’s request/problem,
a cost savings to the customer or company, the potential for a

* Document change request.

161

TLFeBOOK
162 Teixeira and Bradley

process improvement, or a change that is imposed by external


circumstances. In all cases, the design control process is to be
carried out under controlled conditions.
The purpose of this procedure is to define the system for
controlling the design control process at Any Company, Inc.,
in order to ensure that specified design requirements are un-
derstood and met.

SCOPE

This procedure applies to all products developed by or for Any


Company, Inc., that are required by the FDA Quality System
Regulation to meet design control requirements.

DEFINITIONS

Design control project: a self-contained program of work initi-


ated to either introduce a new product or process or to make
a change to an existing product or process that may require
redevelopment of that product or process.
Design history file: a compilation of records (drawings,
formulations, test methods, etc.) that describes the design his-
tory of a finished device.
Design input: the physical and performance require-
ments of a device that are used as a basis for device design.
Design output: the results of a design effort at each design
phase and at the end of the total design effort. The finished
design output is the basis for the device master record. The
total finished design output consists of the device, its packag-
ing and labeling, and the device master record. Design outputs
include the tests/procedures, etc. that are developed/utilized
to show/meet the design input requirements.
Design review: a documented, comprehensive, systematic
examination of a design to evaluate the adequacy of the design
requirements, to evaluate the capability of the design to meet
these requirements, and to identify problems.

TLFeBOOK
Appendix A: Implementation Procedure 163

Design validation: establishing by objective evidence that


device specifications conform to user needs and intended
use(s). (This may include laboratory testing.)
Design verification: a confirmation by examination and
provision of objective evidence that specified requirements
have been fulfilled and to what extent they have been
achieved.
Human volunteer study: a study in which a test article
is used on human volunteers to assess efficacy or user accep-
tance of the test article. The study may also yield limited infor-
mation regarding test article safety; however, it is not per-
formed to obtain such data.
Healthy human volunteer study: a product evaluation
performed on volunteers that are usually in-house (laboratory
based) and generally not suffering from the condition for
which the device is intended to treat.
Clinical study: a product evaluation performed on patient
volunteers.
Marketing evaluation (consumer preference test or field
test): a study in which a marketed or nonmarketed test article
is not used on humans but instead gathers input from hu-
mans, to assess physical and chemical characteristics of the
test article; or a study in which a marketed test article is used
on humans, but no individual medical data is collected (e.g.,
questionnaire).
Preliminary evaluation: work done to establish the basic
merit of an idea.
Product evaluation: study designed to assess effective-
ness and/or user acceptance of a product or intermediate (test
article). These studies are usually in-vivo based, and per-
formed on humans.

RESPONSIBILITIES

The R&D Department is responsible for planning, organizing,


and managing design control projects. Project team members

TLFeBOOK
164 Teixeira and Bradley

are responsible for acting as department liaisons in support


of project requirements. Management is responsible for adher-
ing to the requirements of this procedure and ensuring that
employees comply. The project team leader is responsible for
coordinating all aspects of the design control project.

REQUIRED EQUIPMENT

N/A

MATERIALS

N/A

RECORDS

Product initiation request (PIR)


Product performance specification (PPS)
Project design change form
Risk analysis master record (RAMR)
Design review meeting record
Design project plan
Meeting minutes comment form
Approval for sale

REFERENCES

Record retention
Project planning software or equivalent manual calcula-
tions
Risk analysis
Engineering change notice procedure
Document control procedure

TLFeBOOK
Appendix A: Implementation Procedure 165

PROCEDURE

There are typically four phases of any design control project:

Preliminary evaluation phase


Feasibility phase
Development/manufacture phase
Market introduction phase

Preliminary Evaluation Phase


As previously discussed, the requirement to initiate a design
control project may arise for a variety of reasons and from a
number of sources. Any offered products/technology or ideas
may have some preliminary work performed to establish the
basic merit of the idea, that is, a preliminary evaluation.

Feasibility Phase
If an idea appears to have merit, a product initiation request
(PIR) shall be initiated to begin the feasibility phase. The Mar-
keting Department is responsible for initiating the product
initiation request. The PIR defines the basic requirements for
a product and serves as the input for the feasibility phase. Any
results of a preliminary evaluation should be incorporated into
the PIR. Marketing should meet with the Technology Depart-
ment to determine the feasibility of the requirements and the
proper nomenclature.
The PIR form shall document the preliminary require-
ments. The following requirements shall be established (Note:
A distinction should be made between desirable attributes and
essential requirements):

Purpose/indication: Why develop product? What are the


estimated market size and the potential exploita-
tion/impact?

TLFeBOOK
166 Teixeira and Bradley

Market position: How is product to compete, against


whom, and where?
Product: Essential and/or desirable requirements and pa-
rameters; for example:
delivery system
compatibility
sterilization requirements
preferred physical attributes (size, shape, color)
Claims: Those required to successfully and competitively
market the product.
Packaging.
Clinical/technical.
Product costs (target).

Project team members shall be involved as necessary dur-


ing the feasibility stage to support or offer other inputs and
requirements; for example, regulatory, environmental, perfor-
mance, labeling, etc.
Once the feasibility work is completed—that is, it is felt
there is a viable product capable of meeting basic require-
ments—a report of the feasibility results (i.e., PPS, inclusive
of financial information) will be generated and the project
leader will call a formal design review meeting.

Development/Manufacture/First Clinical Use


(Start of Formal Design Control) Phase
At the completion of the feasibility phase, the project leader is
responsible for completing the PPS with assistance from other
project team members as needed. The PPS serves as an output
from the feasibility phase and as a primary input for the
development/manufacture phase. The PPS shall document
the product requirements. Any necessary changes regarding
initial design inputs/outputs identified on the PIR shall be in-
corporated into the PPS. All inputs defined by the PPS shall
be verifiable/quantifiable where possible, and any essential
outputs for a device to ensure its proper performance, func-

TLFeBOOK
Appendix A: Implementation Procedure 167

tioning, or use shall be defined. (Note: A distinction should be


made between desirable attributes and essential require-
ments. Layman terms should be used. Customer expectations,
safety, and satisfaction must be taken into consideration. If
certain requirements are unique for a market, such as the Eu-
ropean Community, then it should be indicated as such within
the appropriate section. Additional description areas may be
added as required.)
The PPS shall define the following:

Performance characteristics:
• indications for use of product
• clinical procedure for use
• relevant setting/use environment (e.g., nursing
home, home care, hospital)
• medical specialty of user (e.g., doctor, nurse, end
user/layperson)
• patient population inclusion/exclusion criteria
• clinical outcomes analysis (trial data)
Product characteristics
• Physical characteristics (e.g., dimensional, color,
etc.)
• chemical characteristics (describe the direct
chemical interactions of the product in prepara-
tion for and during a procedure, e.g., drugs, wipe
downs, if applicable)
• biological characteristics (indicate the maximum
duration of product use in-vivo (wear time) and
body fluids and tissue to which the item will be
exposed and toxicity and biocompatibility re-
quirements)
• environmental characteristics (describe antici-
pated conditions in transportation, storage, and
use, e.g., temperature, humidity, and/or any limi-
tations, etc.)
• sterilization characteristics [describe the type of
sterilization to which the product will be exposed

TLFeBOOK
168 Teixeira and Bradley

and the number of resterilization (dose) the prod-


uct must be able to withstand]
• packaging (describe the packaging material and
configuration in which the product will be steril-
ized and/or shipped)
• equipment interface (describe any accessories
necessary for use of the product/device)
• safety and reliability requirements
Marketing requirements
• intended marketplace (United States, Europe,
etc.)
• labeling (precautions, warnings, or contraindica-
tions)
Claims
Regulatory/quality assurance requirements
• relevant regulatory or statutory requirements
[e.g., standards/test methods: FDA QSR, ISO
9001, EN 46001, 93/42/EEC (CE Mark), ASTM,
etc.]
Financial requirements
• The market in which the product/device is to per-
form and market’s size
• cost projection
• competitive environment (e.g., competitors,
strengths, and weaknesses)
• proposed forecast/profit
• capital projection
• percent share of market (estimated shares)
• total opportunity
• resource assessment (engineering, packaging,
regulatory, etc.)

The PPS shall be reviewed and approved by all project team


members as part of the initial design review. Changes from
this point forward shall be documented and controlled using
the project design change form.
The project leader, or designee, shall create a project plan

TLFeBOOK
Appendix A: Implementation Procedure 169

using standard off-the-shelf project planning software to de-


fine project tasks and activities, associated timeframes, and
project team member/department responsibilities and inter-
relationships. The detail in which planning is carried out and
documented may vary depending on the complexity and times-
cale of the project. All project team members shall approve the
project plan as part of the initial design review. Project plans
shall be reviewed, updated, and approved as the project
evolves.
A risk analysis shall be conducted at the initial design
review meeting in order to assess the risks and hazards associ-
ated with the use of a device/product. The risk analysis master
record (RAMR) shall be approved by all project team members
as part of the initial design review. The RAMR shall be re-
viewed at each subsequent design review.
Formal design review meetings shall be conducted at sig-
nificant milestones to verify and/or validate that design out-
puts meet design input requirements. At a minimum, design
reviews shall occur at the end of feasibility, the end of develop-
ment, and prior to market introduction/product launch.
The purposes of the initial design review meeting are to
formally define and confirm the design inputs and expected
outputs and to initiate the development/manufacture phase.
The PPS shall be a critical element of the initial design review.
The initial design review meeting shall also formally define
the design project team. All members must be present at the
initial design review meeting, and each design review shall (1)
consist of an individual who does not have direct responsibility
for the design stage being reviewed and (2) include any spe-
cialists needed. Design reviews may include the review of de-
sign verification data to determine whether (1) the design out-
puts meet functional and operational requirements, (2) the
design is compatible with components, (3) the safety require-
ments are achieved, (4) reliability and maintenance require-
ments are met, (5) labeling and other regulatory requirements
are met, and (6) the manufacturing process is compatible with
design specifications.

TLFeBOOK
170 Teixeira and Bradley

The initial design review meeting will review the design


inputs, the expected outputs, and any known outcomes. The
project plan, risk analysis, PPS, and any other pertinent infor-
mation shall be included in the design review. The following
format will be used to document design review meetings (de-
sign review meeting record). (Note: All inputs should be ex-
pressed in verifiable/quantifiable terms. Outputs should allow
for an adequate evaluation/verification of conformance to de-
sign input requirements. As such, outputs should contain or
make reference to acceptance criteria and identify characteris-
tics of the design that are crucial to the safe and proper func-
tioning of the product.)

Inputs Outputs Review/Results

(Inputs ⫽ outputs) (Verification/validation)


Requirement Specification/test Outcome

Noncytotoxic Agar diffusion method Cytotoxicity report—Pass


Product sizes Product specification Inspection—Pass
Flow rate Test method Test—Pass
Wear time Clinical protocol Clinical study report
Expiration date Stability protocol Stability report
Package integrity ASTM method Transit trial report
CE mark Declaration of conformity CE technical file
Prescription Product labeling 510(k)

Any incomplete, ambiguous, or conflicting requirements


shall be resolved and documented as part of the design review
meeting record.
A design review meeting record shall be generated to doc-
ument the results of design review meetings.
If it is determined at the design review meeting that the
product is not viable, not cost-effective, etc., the design project
will be terminated. All associated data shall be retained in
accordance with the record retention procedure. If the team
decides to move forward with the design project, the design

TLFeBOOK
Appendix A: Implementation Procedure 171

review meeting record shall be approved by all project team


members. This will formally signify the initiation of the
development/manufacture phase and the implementation of
compliance with design controls.
The project team leader shall call project team meetings
as necessary as inputs evolve and the project moves forward.
Prior to each meeting an agenda will be issued that identifies
topics of discussion and project team members required in at-
tendance. The project leader or designee will record all design
project meeting proceedings in minutes. The minutes will in-
clude copies of all tests or reports presented at that meeting.
Project team members shall sign off on the minutes as to their
agreement or disagreement by way of the meeting minutes
comment form.
Any changes or conflicting, ambiguous, or incomplete re-
quirements shall be addressed formally at the subsequent de-
sign review meeting.
As the design evolves in the development cycle, various
methods of verification shall be used to determine whether the
design outputs meet functional and operational requirements
or design inputs. Design verification shall be performed on
prototypes. Verification may be accomplished via a variety of
methods, including design reviews, inspection/testing under
simulated use conditions (i.e., in-vivo testing), biocompatibil-
ity testing, package integrity tests, risk analysis, comparison
to similar designs [510(k)s], tests, and demonstrations, etc.
Verification activities shall identify the method of verification,
the date, and the individual performing the verification.
The project team leader shall call a design review meet-
ing after development and verification of prototypes. Any
changes that need to be implemented shall be documented and
approved by all project team members on the design review
meeting record. Manufacture of product under simulated use
conditions (production runs) shall then commence.
Validation shall be performed under defined operating
conditions on initial production lots. Design validation is per-
formed to ensure devices conform to defined user needs and

TLFeBOOK
172 Teixeira and Bradley

includes testing of production units under actual or simulated


use conditions. The results of all validation activities shall in-
clude the identification of the design, the method(s), the date,
and the individual performing the validation. Validation ac-
tivities may include stability studies, validation (process/
product), clinical evaluation, clinical studies, literature stud-
ies—(i.e., published journal articles), 510(k) comparison, tran-
sit trial, product/market evaluation, review of labels/labeling,
etc. [Note: Process and/or product validation is a critical ele-
ment of design control. Process/product validations are per-
formed to validate production processes, for example, that
manufacturing standard operating procedures (output) meet
input requirements.]

Market Introduction
After all verification and validation activities have been com-
pleted, the project team leader shall call a final design review
meeting. This meeting is the final confirmation that the over-
all design output has met the overall design input. All project
team members are required at the final design review meet-
ing. Any changes or conflicting, ambiguous, or incomplete re-
quirements shall be documented on the design review meeting
record, which shall be reviewed and approved by all project
team members.
The final design review meeting shall include a final re-
view of the risk analysis to assess any additional potential
hazards associated with the device under normal and fault
conditions. The risk analysis master record shall be updated
and approved by all participating project team members. Any
necessary changes shall be implemented prior to transferring
design and development specifications/procedures to produc-
tion specifications/procedures and sign-off of the approval for
sale form.

Design Transfer
Transfer of product development specifications/procedures to
production specifications/procedures shall be done using a

TLFeBOOK
Appendix A: Implementation Procedure 173

document change request form in accordance with the docu-


ment control procedure. Any required training shall be con-
ducted and documented.

Approval for Sale


An approval for sale form is completed prior to product launch.
The purpose of the approval for sale form is to document the
confirmation from all project team members that all docu-
ments and data pertinent to their area(s) of responsibility and
necessary to ensure that the product is ready for distribution
are in place. The approval for sale form shall indicate in what
countries the product has been approved for sale and at what
price.

Design History File


A design history file (DHF) shall be established and main-
tained for each type of device. The DHF shall contain or refer-
ence the location of all documents/records needed to demon-
strate compliance with the design plan and design control
requirements. The elements of the DHF shall include, as ap-
plicable, the PIR, PPS, feasibility report, project team meeting
minute and comment forms, design review meeting records,
project design change forms, risk analysis master record,
drawings/formulations, product development specifications/
procedures, clinical testing, stability studies, project plans,
safety testing, market evaluations, validations, transit trials,
and approval for sale. The DHF shall be retained in accor-
dance with the record retention procedure.

Design Changes
Any changes after the design has been transferred to Manu-
facturing shall be in accordance with the engineering change
notice procedure.

TLFeBOOK
TLFeBOOK
Appendix B
Concept Document

PRODUCT INITIATION REQUEST

Product name: New Product X

Purpose/intended use: New Product X is a family of


products intended to hermetically cover a site and apply heal-
ing entities at the site. New Product X also acts as a device
that protects these sites from external contaminants. In-
tended launch is Quarter 1 2001.

Marketing position: (Check One)


Is product intended to be sold into the European Union
as a medical device?

175

TLFeBOOK
176 Teixeira and Bradley

YES X No
New Product X is intended for sale initially in the United
States and Canada. It is recommended that all issues be ad-
dressed for sale of the product line in the European Union. By
providing regulatory compliance for entry into the European
Union from the outset, product components, design, packag-
ing, and clinical research will be in place to take advantage
of worldwide licensing opportunities. The device will be posi-
tioned as an adjunct therapy in whatever management and
treatment and for the protection of various sites. In some mar-
ket segments as, for example, prepackaged kits, the product
will be offered to manufacturers of such products as an OEM
product. The device will be positioned as a “positive manage-
ment entity.”

Product: Various sizes are planned, each with its own


special adaptations for a particular market segment. This is
a new indication for a class III device already approved.

INDICATIONS

U.S. marketed product: New Product X: A protective device


for certain sites that provides a physical barrier to external
contaminants.

European Union marketed product, if applicable:


The same claims as in the United States.

Packaging: Sterile, single-unit, blister pack.

Clinical/technical: Initial clinical trial for safety and ef-


ficacy has been completed. Demonstration of positive patient
management compared with competitor A and wear time have
been completed.

Product Cost: Depending on product size and complex-

TLFeBOOK
Appendix B: Concept Document 177

ity of catheter connection, a direct manufacturing cost of be-


tween $X.XX and $X.XX for should be targeted.

SIGNATURE OF AUTHOR(S):

DATE:

TLFeBOOK
TLFeBOOK
Appendix C
Product Specification

PRODUCT PERFORMANCE SPECIFICATION

Product name: New Product X


General description of product or system: New Product X is
a family of products intended to hermetically cover a site and
apply a therapeutic entity at a flow rate of about 1 liter per minute.
The New Product X device is a derivative design of commercial
product A, approved for marketing as a class III device indicated
for the treatment of indication F, indication G, and indication H.
New Product X is additionally designed as a device that pro-
tects these indications from external or environmental contami-
nants.

179

TLFeBOOK
180 Teixeira and Bradley

REVISION NUMBER: 0700-B

AUTHOR(S): DATE:
APPROVALS:

VP R&D:

VP Regulatory Affairs:

VP Sales and Marketing:

VP Manufacturing:

President/CEO:

Product Performance Specification (cont.)


I. Performance definitions: Distinction should be made
between attributes and essential requirements. Layman
terms should be used. Customer expectations, safety, and sat-
isfaction must be taken into consideration. If certain require-

TLFeBOOK
Appendix C: Product Specification 181

ments are unique for a market (e.g. Europe), then it should be


indicated as such within the appropriate section. Additional
description areas may be added as required.

Clinical Terms
Indications for Use of Product
New Product X [general]: indications F, G, and H; plus use as
a protective device for these sites.

New Product X: indicated for use as a protective device


over specific location. Provides all the benefits listed
in the filing for Product A and provides a protective
entity around the indication that protects the site
from environmental contaminants.

Clinical Procedure for Use


Refer to instructions for use for Product A. The instructions
(current revision) for use for New Product X are also included
in this PPS by reference.

Relevant Setting/Use Environment (e.g., nursing home,


home care, hospital)
Acute-care hospital, trauma centers, skilled nursing facility,
chronic care settings, hospice, and home health care.

Medical Specialty of User (e.g., Doctor, Nurse, End


User/Layperson)
Nurse, doctor, and home health care provider.

Patient Population Inclusion/Exclusion Criteria


Excludes all patients allergic to the ingredients or components
of the product and patients excluded from using Product A. It
is expected that New Product X would continue to be “for sale

TLFeBOOK
182 Teixeira and Bradley

by or on the order of a physician.” It would remain a prescrip-


tion product.

Clinical Outcomes Analysis (Trial Data)


Initial safety and efficacy. Initial safety and efficacy testing
was included in the filing for Product A. The components of
New Product X are similar, if not identical. A Phase 1 clinical
study on the wear time, ease of use, and patient comfort levels
is underway and the results will be used for any necessary
design changes. The results of the trial are expected to be com-
pleted by (date), and will be kept on file.

Product Characteristics
Physical Characteristics (Dimensional, Color, etc.)
New Product X comes in various sizes and model types de-
pending on the specific application/use. It is a sterile, single-
use, disposable device. In general, it will be rectangular with
rounded corners. A specific medical pressure-sensitive adhe-
sive is used to anchor the device to the skin. Six medical-grade
polyurethane tubes are attached. Four are inlet and outlet
tubes for the introduction of the protective entity into and out
of the device. The remaining two tubes are for the connection
through the device into the indication. All tubes have a plastic
clip to control flow through the tubes. The New Product X de-
vice has a clear, see-through polyurethane bubble that allows
the clinician an unobstructed view of the site and areas adja-
cent to the indication.
The initial product will have overall rectangular dimen-
sions of 25.50″ length by 22.25″ width. The four corners of the
rectangle will be rounded to a radius of 2 in. There will be an
adhesive border with a 5.5″ width on the outside perimeter of
the product. The adhesive side mounts toward the body and
anchors the device in place on the patient. The center of the
rectangle that becomes the working area has dimensions of

TLFeBOOK
Appendix C: Product Specification 183

15.0″ ⫻ 12.75″. Six polyurethane tubes are attached to the


chamber. Four are the inlet and outlet tubes for the introduc-
tion of the protective entity into and out of the device. These
two tubes are approximately 16″ to 17″ long and have a 0.175″
outside diameter and a 0.125″ inside diameter. Each has a clip
for closing that tube. The two remaining tubes are for the con-
nection of the device into the indication. This tube is 10″ to
12″ long (with 4″ to 5″ remaining inside the device and 6″ to
7″ of its length outside the device). This tube has an inside
diameter of 0.125″ and an outside diameter of 0.175″. It also
has a pinch clip to shut off flow through the tube. The distal
end of this tube is terminated in a male Luer lock, while the
proximal end is terminated in a female Luer lock connector
with a male lock cap. A product drawing LTS 091300-00 is
attached.

Chemical Characteristics (Describe the Direct Chemical


Interactions of the Product in Preparation for and
During a Procedure, e.g., Drugs, Wipe Downs, if
applicable)
Formulation of the adhesive barrier must be able to withstand
the presence of IPA and other liquids and materials found in
hospitals. Operates with material XX. Suitable operating
pressure will be determined by in-vitro and in-vivo testing.
Clinical testing to date indicates that a flow rate of 1 l/min is
sufficient to keep the device operational.

Biological Characteristics [Indicate the Maximum


Duration of Product Use in-Vivo (Wear Time), Body
Fluids and Tissue to Which the Item Will Be Exposed,
and Toxicity and Biocompatibility Requirements]
Product will be replaced every 3 to 5 days or when it is neces-
sary for the health-care professional to have access to the indi-
cation. It will generally be placed on intact, dry skin. Several
devices may be used for several weeks and could possibly be

TLFeBOOK
184 Teixeira and Bradley

used up to 30 days. The device should not be in direct contact


with circulating blood. All components contacting the patient
should be nonirritating, noncytotoxic, and hypoallergenic.
Data included with the original filing indicates that the com-
ponents meet these criteria. Data on alternative materials, if
any, will be received from the manufacturer prior to use.
The components that would normally contact the skin
would be the medical-grade adhesive, the connectors on the
centerline attachment, and the tubing of the centerline attach-
ment. It is possible that in the event of a failure to maintain
sufficient operating conditions that the device film could also
contact the patient.
One suitable adhesive is manufactured by XX and con-
verted by YY. Their tape is essentially polyethylene film
coated on both sides with a hypoallergenic pressure-sensitive
acrylate adhesive. It is supplied on a paper liner, which is
bleached and silicone-coated on both sides. XX has certified
the safe and effective use of its No. 575783 Medical Tape for
its intended use. Biocompatibility tests completed include in-
vitro cytotoxicity (agar overlay), in-vitro hemolysis, acute pri-
mary skin irritation in albino rabbits, repeated insult patch
(Draize) in humans, 21-day cumulative irritation in humans,
intracutaneous irritation in albino rabbits, and acute systemic
toxicity in mice.
Suitable polymer, for both tubing and molded connectors,
is supplied by TT Compounds, a subsidiary of ZZZ. Testing
from NAMSA on plaques included acute systemic toxicity, in-
tracutaneous toxicity, and implantation test. This data is on
file at Any Company, WWW Medical (contract manufacturer
of New Product X prototypes), and at the manufacturer of the
components. Material is class VI. This safety testing is also
applicable to the polymer-molded connectors.
Input and output tubing is medical-grade polyurethane
with a Shore A hardness of 80, which is the same tubing used
to manufacture Product A. This tubing may also be used for
the other tubing. The device film is medical-grade polyure-

TLFeBOOK
Appendix C: Product Specification 185

thane—Trademark 329480 or equivalent. Specifications are


currently maintained at WWW Medical in City, State along
with appropriate safety testing on urethane components. (tub-
ing, film, and connectors) WW documentation 24185-6503391
Rev B. Transfer of documentation to Any Company is un-
derway.
All fittings and connectors will be plastic and molded
from acrylic, polycarbonate, nylon, or high-density polyethyl-
ene as received from the suppliers of these medical connectors
and clamps. Safety data and certificates of compliance are on
file. This data is in addition to the data cited above on the
materials themselves.

Environmental Characteristics (Describe Anticipated


Conditions in Transportation, Storage, and Use, e.g.
Temperature, Humidity, and/or Any Limitations, etc.)
It is not anticipated that the product will be subjected to any
unusual conditions in storage and/or transportation. During
use the product must withstand a positive pressure of 10 cm
(H 2 O). Normal precautions must be taken with regard to op-
erating. See Product A instructions for use. The adhesive has
a 2-year shelf life and should be stored between 50° to 80° F
at 40% to 60% humidity.

Sterilization Characteristics (Describe the Type of


Sterilization to Which the Product Will Be Exposed
and the Number of Resterilization (Dose) the Product
Must Be Able to Withstand]
These are intended to be single-use products. New Product X
will require gamma or electron-beam sterilization due to the
nature of the materials. EtO will be excluded, as it will be im-
possible to remove this gas from the device.
Gamma sterilization should be anticipated at a dose of
2.5 MRad, with a sterility assurance level of 10 ⫺6. Dosimetry
is required at 3 MRad ⫾ 10%. Sterility at a lower dose is possi-

TLFeBOOK
186 Teixeira and Bradley

ble, as is dosimetric sterility release, depending on the biobur-


den levels of the manufacturing facility.

Packaging (Describe the Packaging Material and


Configuration in Which the Product Will Be Sterilized
and/or Shipped)
Sterile, single-use, PETG blister packs with Tyvek  for each
unit.

Equipment Interface (Describe Any Accessories


Necessary for Use of the Product/Device)
N.A.

Safety and Reliability Requirements


Color-coded fittings will be used to differentiate the inlet and
outlet tubes from the site line. Transparent tubes, film, and
fittings are used to ensure visibility. Failure analysis will be
conducted and documented on completion of first production
run.

Marketing Requirements
Intended Marketplace
U.S. Market (Domestic): New Product X is expected to grow
from $1.5 million in sales during the first full year of sales to
$83.5 million in year 5.
International Market (List Countries): United King-
dom, Germany, Scandinavia, Italy, Belgium, Netherlands,
France, and Spain.
Labeling (Precautions, Warnings, Contraindica-
tions): Currently as they appear in the instructions for use
and/or product labeling.
Claims: New Product X: all claims of Product A plus use
as a protective device for F, G, and H. The insert is attached.
The device excludes external contaminants.

TLFeBOOK
Appendix C: Product Specification 187

Regulatory/Quality Assurance Requirements


Relevant Regulatory or Statutory Requirements [e.g.,
Standards/Test Methods—FDA QSR, ISO 9001, EN
46001, 93/42/EEC (CE Mark), ASTM, etc.]

FDA 21 CFR Part 807: Device listing and establishment


registration
FDA 21 CFR Part 820: FDA quality system regulation
EN 46001: Europe’s QS requirements for medical devices
FDA 21 CFR Part 801: Labeling
FDA 21 CFR Part 803: Medical device reporting
MDD: Medical device vigilance requirements
FDA 21 CFR Parts 7: Recall/remedial action
FDA 21 CFR 806: Reports of corrections and removals
ISO 10993: Biocompatibility testing
93/42/EEC: Medical device directive
ANSI/AAMI/ISO 11037 or 11035: Sterilization (gamma
or EtO)
EN 550: EtO
EN 552: Irradiation (gamma)
EN 554: Moist heat
EN 980 and EN 1041: Labeling
EN 556: Requirements for terminally sterilized devices
to be labeled “sterile”
ISO 11067: Packaging for terminally sterilized medical
devices
EN 868-1: General requirements and test methods—
packaging materials and systems for medical de-
vices, which are to be sterilized (this is a series of
standards)
EN 1441: Risk analysis
PrEN 12442-1,2,3: Animal tissue materials
FDA Docket No. 98D-0924: Animal tissue materials
FDA validation guidance document
Validation of sterilization: AAMI/ISO 11137 Method 1

TLFeBOOK
188 Teixeira and Bradley

Package integrity: ASTM F88 seal strength; ASTM


D1140 burst strength
ASTM Test Methods—Film: ASTM D638: Tensile proper-
ties; ASTM D1003: Test method for haze and lumi-
nous transmission of transparent plastic; ASTM
D1004: Initial tear resistance; ASTM D1044: Resis-
tance of transparent plastics to surface abrasion;
ASTM D 1239: Resistance of plastic film to extrac-
tion by chemicals; ASTM D1709: Test method for im-
pact resistance of film (falling dart); ASTM D2582:
Puncture-propagation tear resistance of plastic film.
Adhesive: ASTM D896: Resistance of adhesive bonds to
chemical reagents; ASTM D903: Peel strength;
ASTM D1002: Shear strength; ASTM D1151: Effect
of moisture and temperature on bond strength;
ASTM D3121: Tack (rolling ball).

It may not be necessary to complete all these particular test


methods; substitution by tests performed by the manufacturer
of the components purchased by LifeTech may be sufficient.
Quality test development should use these properties as a
guide in determining testing of components and final product.

Financial Requirements
Market the Product/Device Is to Perform in and Size
The total market in the United States is $64 billion and grow-
ing at 16.4% per year.

Cost Projection
Total operating costs are approximately $3 million.

Competitive Environment (e.g., Competitors, Strengths,


and Weaknesses)
A predicate device table is included with the FDA submission
and is incorporated here by reference.

TLFeBOOK
Appendix C: Product Specification 189

Proposed Forecast/Profit
See business plan.

Capital Projection
See business plan. Product will be produced at the Maui fa-
cility.

Percent Share of Market (Estimated Share)


Less than 5%.

Total Opportunity
Current estimate of market size is $14 billion.

Resource Assessment (Engineering, Packaging,


Regulatory, etc.)
Will require the following:

• Design (product and process)


• Regulatory
• Clinical
• Sales and marketing
• Quality assurance/testing
• Reimbursement consultant
• Patent attorney
• Outside source

TLFeBOOK
TLFeBOOK
Appendix D
Product Claims Sheet

191

TLFeBOOK
192 Teixeira and Bradley

Product Name:

Intended Use(s)

Product Claims

Claims Supporting Data

Precautions and Warnings:

Contraindications:

Compiled By:

Director Quality Assurance/


Regulatory Affairs

Approval/Date:

Director Sales and Marketing

Approval/Date:

Director Clinical Studies

Approval/Date:

Revision and Revision Date:

TLFeBOOK
Appendix E
Risk Analysis: Standard Operating
Procedure

RISK ANALYSIS

Effective
Rev. DCR Date Date Originator Description
A 04-02-00 04-27-00 M. Teixeira DCR 00-0115

Approval Signature: Date:

1.0 PURPOSE

1.1 Judgments relating to the safety, including the ac-


ceptability of risks, are necessary in order to de-
193

TLFeBOOK
194 Teixeira and Bradley

termine the suitability for use of a product. Such


judgments take into account the manufacturer’s in-
tended use; the performance, risks, and benefits of
the product; and the risks and benefits associated
with the clinical performance.
1.2 The criteria that must be evaluated/met regarding
the safety, function, and quality of the product are
as follows:

The product must be designed and manufactured in


such a way that, when used under the conditions and for
the purpose intended, it will not compromise the clinical
condition or the safety of the patients, or the safety and
health of users or, where applicable, other persons, pro-
vided that any risks that may be associated with their use
constitute acceptable risks when weighed against the ben-
efits to the patient and are compatible with a high level
of protection of health and safety.

This procedure outlines the requirements for per-


forming a risk analysis assessment to investigate
the safety of a product by identifying hazards and
estimating the risks associated with the product.

2.0 SCOPE

2.1 This procedure applies to those individuals respon-


sible for performing a risk analysis assessment in
support of the CE mark or any other design control
project.

3.0 DEFINITIONS

3.1 Harm: physical injury and/or damage to health or


property
3.2 Hazard: a potential source of harm

TLFeBOOK
Appendix E: Risk Analysis: Standard Operating Procedure 195

3.3 Risk: the probable rate of occurrence of a hazard


causing harm and the degree of severity of the
harm
3.4 Risk analysis: the investigation of available infor-
mation to identify hazards and to estimate risks
3.5 Safety: freedom from unacceptable risk of harm

4.0 RESPONSIBILITIES

4.1 Director of Quality Assurance and Regulatory Af-


fairs: is responsible for coordinating the completion
of the risk analysis, ensuring that all the necessary
documentation is available, and arranging final re-
view and approval of the risk analysis report.

5.0 REQUIRED EQUIPMENT

5.1 N/A

6.0 ASSOCIATED MATERIALS

6.1 N/A

7.0 RECORDS

7.1 Risk analysis master record

8.0 REFERENCES

8.1 Medical device directive (93/42/EEC)


8.2 EN 1441 Medical devices—risk analysis

TLFeBOOK
196 Teixeira and Bradley

8.3 Guide to identification of characteristics which


could affect safety, Appendix I
8.4 Examples of possible hazards and contributing fac-
tors, Appendix II
8.5 SOP 100-003 Record retention
8.6 SOP 100-004 Complaint handling procedure
8.7 SOP 100-007 Document control
8.8 SOP 700-001 CE technical files
8.9 SOP 700-002 Medical device vigilance reporting
8.10 SOP 700-008 CE document control
8.11 Design control procedures (400 series)
8.12 Risk analysis master record
8.13 Risk analysis matrix

9.0 PROCEDURE

9.1 Once it has been determined that an existing device


is to be CE marked or that a new product to be devel-
oped is to be CE marked, the hazards and risks asso-
ciated with the use of the device/product need to be
analyzed. A risk analysis shall be the method used
to analyze and assess the device’s risks and hazards.
The risk analysis shall be performed by individuals
from various departments (e.g., Technology, Mfg,
Mktg, etc.).
9.2 For new products the risk analysis should be done
after the feasibility stage. It should be updated as
necessary, as the development process evolves.
9.3 Flow diagram of risk analysis procedure
9.4 Refer to the RAMR form
9.4.1 Heading: The heading of the RAMR con-
tains the product name, product descrip-
tion, intended use, and the signatures of the
individuals involved in the risk analysis
process.
9.4.2 Characteristics that could affect safety: List

TLFeBOOK
Appendix E: Risk Analysis: Standard Operating Procedure 197

Figure 1 Carrying out the above procedure results in a risk analy-


sis master record (RAMR).

all the characteristics that could affect the


device’s safety. A nonencompassing list,
which can be used as a guide to identify
characteristics, is given in Appendix I.
9.4.3 Possible hazard associated with character-
istics: The hazard is a potential source of
harm resulting from a characteristic that
could affect safety. Compile a list of poten-
tial hazards associated with the device in
both normal and fault conditions. A nonen-

TLFeBOOK
198 Teixeira and Bradley

compassing list of potential hazards is given


in Appendix II.
9.4.4 Estimation of risks: For each possible haz-
ard, estimate the associated risk(s) under
both normal and fault conditions. Consider
the implications (severity) of the hazard ac-
tually happening and the probability of the
hazard occurring. To estimate the risks, re-
fer to the risk assessment matrix.

Severity level:

• Catastrophic 1. Loss of life. Severe, per-


manent damage to health.
• Critical 2. Serious injury, possibly life
threatening, but reversible. Requires
medical intervention (e.g., severe burns).
• Marginal 3. Minor injury, but cause sig-
nificant discomfort (e.g., minor burns).
• Negligible 4. Minor injury with minor
discomfort, readily reversible.

In the column “Risk assessment,” the resul-


tant analysis should be in the format High/
Medium/Low and indicate the probability/
severity (e.g., Low (C4), Medium (E2), etc.
derived from risk assessment matrix).

9.4.5 Acceptability of risk/risk reduction: If the


risk is high (as determined on the risk as-
sessment matrix), this is unacceptable and
needs to be addressed. This unacceptable
risk should be reduced to acceptable levels
by appropriate means. Examples are given
below.
• Direct safety means (design).
• Indirect safety means (safeguarding).

TLFeBOOK
Appendix E: Risk Analysis: Standard Operating Procedure 199

Examples of safeguarding are re-


stricting accessibility (e.g., for radiation
hazards) and shielding from the hazard
(e.g., by means of a protective cover).
• Descriptive safety means (e.g., re-
stricting period or frequency of use of the
device, restricting application, lifetime,
or environment).
• Redefining intended use.
• Record how the risk is addressed. If the
risk is medium, it will commonly be ad-
dressed by warnings/instructions, etc. on
the device packaging or the instruction
leaflet. If the risk is low, it is optional
whether warnings/instructions are pro-
vided. When possible, risks should be re-
duced or eliminated. When the risk has
been identified as high, the risk should be
reduced to acceptable levels.
9.4.6 Evaluations for new products/existing prod-
ucts: For existing products, perform a com-
plaint review and a nonconformance review.
The complaint review should consider how
long the product has been on the market, the
number of complaints on record, and the na-
ture of these complaints. The nonconform-
ance review should consider the nonconform-
ances related to relevant product and any
risk(s) associated with these nonconform-
ances.
9.4.7 Comments: Indicate any additional com-
ments here (e.g., product status with re-
gards to the FDA, etc.).
9.4.8 Sign-off: Upon completion of the RAMR, the
signature of the Director of Quality Assur-
ance and Regulatory Affairs and the date
are required.

TLFeBOOK
200 Teixeira and Bradley

9.4.9
Supporting documentation: All documenta-
tion supporting the risk analysis process
will be attached to the RAMR.
9.4.10 Storage of risk analysis master records:
RAMRs will be maintained in Regulatory
Affairs. RAMRs will be retained according
to appropriate procedure.
9.5 Amendments/changes to the RAMR

The RAMR may require amendment/change when:

• An additional indication/intended use is made for


the product.
• Additional product codes are added to the product
line.
• Changes are made to the product.
• Complaints are received on the product.
• New risks are identified (e.g., as a result of tech-
nology).
• A change in the device classification occurs.

The revision status and control of the RAMR is per appli-


cable control procedure. The reason for change will be noted
in the change history.

RISK ASSESSMENT MATRIX


SEVERITY
Catastrophic 1
Critical 2 High Medium
Marginal 3
Negligible 4 Low
A B C D E
Probability
Frequent Probable Occasional Remote Improbable

TLFeBOOK
Appendix E: Risk Analysis: Standard Operating Procedure 201

SUPPLEMENT 1 RISK ASSESSMENT SOP


Guide to Identification of Characteristics That
Could Affect Safety
1. What is the intended use, and how is the device to
be used?
2. Is the device intended to contact the patient or other
persons?
3. What materials and/or components are incorpo-
rated in the device or are used?
4. Are substances delivered to and/or extracted from
the patient?
5. Is the device supplied sterile or intended to be steril-
ized by the user, or are other microbiological con-
trols applicable?
6. Is the device intended to modify the patient’s envi-
ronment?
7. Is the device intended to control or to interact with
other devices or drugs?
8. Is the device susceptible to environmental influ-
ences?
9. Are there essential consumables or accessories asso-
ciated with the device?
10. Does the device have a restricted shelf life?
11. Are there possible delayed and/or long-term use
effects?
12. To what mechanical forces will the device be sub-
jected?
13. What determines the lifetime of the device?
14. Is the device intended for single use or reuse?
15. Can the device be used by a high-risk patient (e.g.,
infant, elderly, immunocompromised, etc.)?
16. How is the device sold (e.g., sterile, nonsterile, pre-
served, etc.)?

TLFeBOOK
202 Teixeira and Bradley

SUPPLEMENT 2 RISK ANALYSIS SOP


Estimation of Toxicological Risk
The toxicological risk analysis should take account of

• The chemical nature of the materials


• The identity, concentration, availability, and toxicity
of all constituents (e.g., additives, processing aids, mo-
nomers, catalysts, reaction products, etc.)
• The influence of biodegradation and corrosion on the
material
• Prior use of the materials
• Biological safety test data

SUPPLEMENT 3 RISK ANALYSIS SOP


Examples of Possible Hazards and Contributing
Factors
Environmental Hazards
• Likelihood of operation outside prescribed environ-
mental conditions
• Incompatibility with other devices
• Accidental mechanical damage
• Contamination due to waste products and/or device
disposal

Biological Hazards
• Bioburden
• Biocontamination
• Biocompatibility
• Incorrect formulation (chemical composition)
• Toxicity
• (Cross) infection
• Pyrogenicity
• Inability to maintain hygienic safety
• Degradation

TLFeBOOK
Appendix E: Risk Analysis: Standard Operating Procedure 203

Hazards Related to the Use of the Device


• Inadequate labeling
• Inadequate operating instructions
• Inadequate specification of accessories
• Inadequate specification of preuse checks
• Overcomplicated operating instructions
• Unavailable or separated operating instructions
• Use by unskilled/untrained personnel
• Reasonably foreseeable misuse
• Insufficient warning of side effects
• Inadequate warning of hazards likely with reuse of
single-use devices
• Incompatibility with consumables/accessories/other
devices

Hazards Arising from Functional Failure, Maintenance,


and Aging
• Inadequacy of performance characteristics for the in-
tended use
• Lack of adequate determination of end of device life
• Loss of mechanical integrity
• Inadequate packaging (contamination and/or deterio-
ration of the device)

TLFeBOOK
TLFeBOOK
Appendix F
Cause-and-Effects Diagram

A cause-and-effects diagram illustrates the relationship be-


tween an outcome and all the factors that influence that out-
come. Because of their shape, these diagrams have sometimes
been called a “fishbone diagram.”* The diagram is intended
to show the relationship of the parts to the whole by

• Determining the factors that cause an outcome or ef-


fect, whether these factors are positive or negative

* The cause-and-effect diagram is also referred to as an “Ishikawa diagram.”

205

TLFeBOOK
206 Teixeira and Bradley

Figure 1 Typical cause-and-effect diagram.

• Focusing on a specific issue without resorting to irrele-


vant discussion*
• Determining the controllable cause of an effect
• Identifying areas in which there is inadequate or no
objective data

* In this case irrelevant discussion is a polite way of saying complaining,


blame shifting, and CYA activities.

TLFeBOOK
Appendix F: Cause-and-Effects Diagram 207

The first step in preparing a cause-and-effect diagram is


to specify what problem or hazard needs to be analyzed. This
is placed in a box on the right side of the diagram. The second
step is to list the major factors that influence the hazard be-
ing analyzed. These factors may include things like person-
nel, materials, components, process, machinery, maintenance,
specifications, design flaws, etc. The next step is to identify
subfactors within each of the major categories already identi-
fied. The final step is to prioritize the list of causes, that is, to
identify which of the causes is the basis for the hazard. Keep
in mind that the placement of a given cause on a diagram is
not an indicator of its importance. The diagram should also
help identify factors for which you may need to collect addi-
tional information, especially if one of the factors was not pre-
viously considering during the development.
Figure 1 is an example of a typical cause-and-effect dia-
gram.

TLFeBOOK
TLFeBOOK
Appendix G
Validation Procedure

Effective
Rev. DCR Date Date Originator Description
A 08-19-93 R. Bradley Quality assur-
ance policy
B 02-03-97 06-30-97 R. Bradley Original release
of policy
C 06-08-98 06-15-98 M. Teixeira DCR 98-0094
D 08-18-98 08-28-98 M. Teixeira DCR 98-0147

Approval Signature: Date:

209

TLFeBOOK
210 Teixeira and Bradley

PURPOSE

To establish a documented procedure and process for ob-


taining, recording, and interpreting results required to show
that a process will consistently produce a product that meets
predetermined specifications.

SCOPE

This procedure applies to all products. The details and extent


of process validation will vary depending on the nature of the
device and the nature and complexity of the process being vali-
dated. Process validation covers three activities: qualification
of process installation and setup; qualification of process capa-
bility; and process long-term stability (i.e., IQ, OQ, and PQ).

DEFINITIONS

Installation qualification: establishing by objective evi-


dence that all key aspects of the process, process
equipment, and ancillary system installation adhere
to the approved design criteria and that the recom-
mendations of the manufacturer of the equipment
are suitably considered.
Operation qualification: establishing by objective evi-
dence parameters that result in a product that meets
all predetermined requirements.
Operating parameters: If the equipment being validated
has variable control settings or parameters, the
range over which the process can operate and pro-
duce within specified AQLs should be clearly stated.
If the upper and lower process limits have been chal-
lenged by the protocol, the resulting parameters
should be identified.
Performance qualification: establishing by objective evi-
dence that the process, under anticipated conditions,

TLFeBOOK
Appendix G: Validation Procedure 211

including worst-case conditions, consistently pro-


duces a product that meets all predetermined re-
quirements.
Process validation: establishing by objective evidence
that the process consistently produces a result or
product meeting its predetermined specifications.
Process validation protocol: a document stating how vali-
dation will be conducted, including test parameters,
product characteristics, manufacturing equipment,
and decision points on what constitutes acceptable
test results.
Retrospective process validation: validation of a process
for a product already in distribution based on accu-
mulated production, testing, and control data.
Validation: confirmation by examination and provision of
objective evidence that the particular requirements
for a specific intended use are fulfilled.
Verification: confirmation by examination and provision
of objective evidence that specified requirements
have been fulfilled.
Worst case: a set of conditions encompassing upper and
lower processing limits and circumstances, including
those within standard operating procedures, that
pose the greatest chance of process or product failure
when compared to ideal conditions. Such conditions
do no necessarily induce product or process failure.

RESPONSIBILITIES

The validation team has the responsibility for planning the


validation approach and defining the requirements. The vali-
dation team may include representatives from or personnel
with experience in Quality Assurance, Engineering, Manufac-
turing, Research and Development, Clinical Affairs, etc.
The Quality Assurance Department has the responsibil-
ity to prepare installation/operation and performance proto-

TLFeBOOK
212 Teixeira and Bradley

cols and reports in accordance with this procedure and to pro-


vide technical support for process validations.
The Technology Department has the responsibility for as-
sisting in defining and fulfilling the requirements of the
installation/operation and performance qualification protocols
(e.g., identifying operating parameters, determining/devel-
oping procedures, developing cleaning and maintenance and
calibration requirements, etc.).

EQUIPMENT

N/A

MATERIALS

N/A

RECORDS

Validation risk assessment


Process validation decision tree
Installation qualification
Operation qualification
Performance qualification

REFERENCE

SOP 100-003 Record retention


SOP 700-005 Risk analysis
Global harmonization task force study group #3—“Draft
process validation guidance”

PROCEDURE
General
Validation requires documented evidence that a process con-
sistently conforms to requirements. It requires that you first

TLFeBOOK
Appendix G: Validation Procedure 213

establish a process that can consistently conform to require-


ments and that you then run studies demonstrating that this
is the case. A number of statistical tools are available to aid
in performing both of these tasks (e.g., acceptance sampling,
FMEA, challenge tests, capability studies, etc.).
Those processes for which the product cannot be fully ver-
ified typically require that process validation be performed. As
a guide, the process validation decision tree shall be used in
determining whether or not a process should be validated.
Note: Each process should have a specification describing
both the process parameters and the output desired. Consider-
ation should then be given to whether the output can be veri-
fied by inspection and/or test (A). If the answer is positive,
then the consideration should be made as to whether or not
verification alone is a sufficient and cost-effective solution (B).
If yes, the output should be verified and the process should be
appropriately controlled (C). If the output of the process is not
verifiable, then consideration with regard to the risk to the
patient of the process or the final product (D) should be given.
If the risk is high, then the decision should be to validate the
process (G). If the risk is low, then justification for not validat-
ing the process should be considered (E). Consequently, man-
agement may decide to validate a process even though the out-
put of the process is verifiable (F).
Processes that should normally be validated include

1. Sterilization processes
2. Clean-room ambient conditions
3. Sterile packaging sealing processes
4. Plastic injection molding processes

Processes that may be satisfactorily covered by verifica-


tion include

1. Manual cutting procedures


2. Testing for color, turbidity, total pH
3. Visual inspection/testing of assemblies

TLFeBOOK
214 Teixeira and Bradley

Processes for which the decision tree may be useful in de-


termining the need for validation include

1. Cleaning processes
2. Certain human assembly processes
3. Numerical control cutting processes
Note: Software utilized in manufacturing or testing pro-
cesses should be validated for its intended use.

Determining the Need for Validation


Prior to the initiation or consideration of any process valida-
tion efforts of a new process or a product/process change, a
validation risk assessment form shall be completed by the
party/group proposing a new piece of equipment or a change.
As a minimum, the validation risk assessment should
consider the following:

1. Impact on the product/process


2. Risk to the patient
3. Review of final product specifications
4. Review of machine design capability (i.e., is equip-
ment properly designed to accomplish end goal?)
5. Equipment instrumentation and required calibration
6. Environmental considerations
7. Applicable procedures (e.g., process specifications,
cleaning, maintenance, etc.)
8. Desired performance and quality parameters

As a minimum, the Vice President of Technology and the


Director of Quality Assurance/Regulatory Affairs shall ap-
prove all validation risk assessment forms.
To determine the impact or risk to the patient, the risk
analysis procedure should be referenced with regards to use
of the risk analysis master record and the risk assessment ma-
trix.

TLFeBOOK
Appendix G: Validation Procedure 215

Once all applicable information has been delineated on


the validation risk assessment form, the process validation de-
cision tree (Appendix A) should be used to assist in determin-
ing if validation is required. If it is determined that no valida-
tion is required, the reason/justification for not doing so shall
be documented on the validation risk assessment form and the
form forwarded to Quality Assurance for filing. If a decision
is made to validate, the validation risk assessment form
should be used as a tool in developing the validation proto-
col(s).

Validation Protocol Generation


Detailed protocols for performing validations are essential to
ensure that the process is adequately validated. A validation
package shall contain, although not limited to, an installation
qualification protocol and a performance qualification pro-
tocol.
For all phases of a process validation (IQ, OQ, PQ), the
following should be determined:

1. What to verify/measure
2. How to verify/measure
3. How many to verify/measure (i.e., statistical signifi-
cance)
4. When to verify/measure
5. Acceptance/rejection criteria
6. Required documentation

Validation Protocol Format


1. Title page.
2. Identify the process, product, or equipment undergo-
ing validation/qualification.
3. Identify the author of the protocol.
4. Abstract—briefly describe the intent of the protocol
[i.e., identification of device(s) to be manufactured us-
ing this process].

TLFeBOOK
216 Teixeira and Bradley

5. Protocol approval. Sign-off by Technology, Quality


Assurance, and Manufacturing is required prior to
performing the installation protocol.

Installation Qualification Protocol


1. Introduction
2. Objective
3. Scope/intended use
4. Applicable part numbers
5. Background
6. Description of the process
7. Installation qualification protocol (main text)
8. Objective
9. Equipment verification
10. Equipment calibration requirements
11. Equipment maintenance requirements
12. Equipment cleaning requirements
13. Process specifications/operating parameters
14. Performance parameters
15. Quality parameters
16. Training

Installation Qualification Report


1. Title page
2. Identify the process, product, or equipment under-
going validation/qualification.
3. Identify the author of the protocol.
4. Abstract—briefly describe the intent of the protocol.
5. Protocol approval. Sign-off by Technology, Quality
Assurance, and Manufacturing is required prior to
performing the installation protocol.
6. Summary of results.
7. Objective.
8. Equipment verification.
9. Equipment calibration requirements.
10. Equipment maintenance requirements.

TLFeBOOK
Appendix G: Validation Procedure 217

11. Equipment cleaning requirements.


12. Process specifications/operating parameters.
13. Performance parameters.
14. Training.

Performance Protocol
1. Title page.
2. Identify the process, product, or equipment under-
going validation/qualification.
3. Identify the author of the protocol.
4. Abstract—briefly describe the intent of the protocol.
5. Protocol approval. Technology, Quality Assurance,
and Manufacturing approval is required prior to
performing the performance protocol.
6. Introduction.
7. Objective.
8. Scope.
9. Background.
10. Performance qualification.
11. Procedure.
12. Performance parameters.
13. Inspection and test acceptance criteria.
14. Quality parameters (sampling).
15. Documentation.
16. List required SOPs (calibration, QA monitoring,
production, maintenance requirements).
17. List required manuals, schematics.
18. List “as-built” drawing modification requirements,
if applicable.
19. List spare parts requirements, if applicable.

Performance Protocol Report (Main Text)


1. Title page.
2. Identify the process, product, or equipment under-
going validation/qualification.
3. Identify the author of the protocol.

TLFeBOOK
218 Teixeira and Bradley

4. Abstract—briefly describe the intent of the protocol.


5. Validation package approval.
6. Sign-off by Technology, Quality Assurance, and Pro-
duction.
7. Summary of results.
8. Objective.
9. Scope.
10. Documentation verification.
11. Calibration verification.
12. Performance verification/testing.
13. Conclusion.

Validation Package Approvals (General)


Installation qualification and performance protocols will be
routed for approval. Area managers requesting revisions will
document their requests. A protocol design review meeting
will be initiated in cases of complex validations requiring in-
put from all departments. Final approval of the validation
package will be made after all documentation and appropriate
personnel have completed training. Completed validations
will be filed in Quality Assurance.

Equipment/Area Maintenance
All equipment must be fabricated in conformance with GMPs.
Equipment should be free of sharp edges that could damage
the product, should not generate excessive particulates while
operating, and should not contaminate product with lubri-
cants or by-products during operation. The equipment/area
should be properly cleaned prior to the protocol run.

Revalidation
Revalidation may be necessary when

• There is a change in the actual process that may affect


quality.

TLFeBOOK
Appendix G: Validation Procedure 219

• There is an investigation of a negative trend in quality


indicators.
• A change in the product design which affects the pro-
cess.
• A process is transferred from one facility to another.
• The scope of application of the process has changed.

Protocol Execution
Prior to protocol execution, the following must occur to prevent
compromising the study:

1. Equipment must be debugged and producing what


appears to be acceptable product.
2. All equipment must be calibrated (as applicable).
3. Maintenance requirements during the run should be
defined.

Retrospective Process Validation


In some cases a product may have been on the market without
sufficient premarket process validation. In these cases, it may
be possible to validate, in some measure, the adequacy of the
process by examination of accumulated test data on the prod-
uct and records of the manufacturing procedures used.
Retrospective validation can also be useful to augment
initial premarket prospective validation for new products or
changed processes. In such cases, preliminary prospective val-
idation should have been sufficient to warrant product mar-
keting. As additional data is gathered on production lots, such
data can be used to build confidence in the adequacy of the
process. Conversely, such data may indicate a declining con-
fidence in the process and a commensurate need for corrective
changes.
Test data may be useful only if the methods and results
are adequately specific. As with prospective validation, it may
be insufficient to assess the process solely on the basis of lot-
by-lot conformance to specifications if test results are merely

TLFeBOOK
220 Teixeira and Bradley

expressed in terms of pass/fail. Specific results, on the other


hand, can be statistically analyzed and a determination can
be made of what variance in data can be expected. It is impor-
tant to maintain records that describe the operating charac-
teristics of the process (e.g., time, temperature, humidity, and
equipment settings). Whenever test data is used to demon-
strate conformance to specifications, it is important that the
test methodology be qualified to ensure that test results are
objective and accurate.

TLFeBOOK
Appendix H
Material Specification

MATERIAL SPECIFICATION 100-02


Vistanex LM-MH LC (Low Color) Polyisobutlylene Polymer

Effective
Rev. DCR Date Date Originator Description

A 02-08-98 02-09-98 R. Bradley Material specification


B 02-10-98 02-11-98 A. Sotnick As per DCR 98-0002

Approved by: Date:

DESCRIPTION DETAILS

Vistanex LM-MH LC is a water-white to pale yellow, tacky,


semi-solid polymer.
221

TLFeBOOK
222 Teixeira and Bradley

SPECIFICATIONS

Appearance: Water-white to pale yellow; Exxon Test


Method #AMS 83-006
Penetration, MM @ 25(C: 15.4 -11.5; Exxon Test Method
#AMS 210-10
Nonvolatile matter, WJ %: 97.0 min.; Exxon Test Method
#AMS 82-003

APPLICABLE REFERENCE DOCUMENTS

Exxon “Vistanex LM Polyisobutylene, Low Color Grades”; Ex-


xon Publication #102-1096-0101C dated October 1996; Exxon
MSDS #84320000 dated April 13, 1996

RAW MATERIAL CODE

Vistanex LM-MH LC; Test RMC 90002

SAFETY AND HANDLING


Personnel Safety
Avoid eye and skin contact. Wear safety glasses with side
shields, long sleeves, and chemical-resistant gloves. Avoid any
contact with heat, open flames, or oxidizing materials. For eye
contact flush with large amounts of clean water until irritation
subsides. Then seek medical attention if irritation persists.

STORAGE REQUIREMENTS

Storage should be at ambient temperature. Electrostatic accu-


mulation hazard should be minimized by proper grounding
procedure during storage.

TLFeBOOK
Appendix H: Material Specification 223

MATERIAL SAFETY DATA SHEETS

Exxon Chemical Company Polymers Group. MSDS No.


84320000

SAMPLING AND TESTING


Sampling
Per Quality Assurance Specification No. QS 700-90002

PACKAGING
Primary Package
A cylindrical paper fiber tube measuring 141/2″ diameter ⫻
24” height, with a close-fitting cap or the same material. The
inside of the tube has a release agent to allow transfer of the
contents out of the tube. The net content is 100# of product.

IDENTIFICATION MARKING

Exxon material name and lot number when received. BioDerm


part number, lot number, and QC receiving inspections status
when QC released.

VENDOR CERTIFICATION

A certificate of analysis will accompany each manufacturer’s


lot of product.

STATEMENTS AND NOTES

Safety and handling data (MSDS)

TLFeBOOK
224 Teixeira and Bradley

QUALITY CHARACTERISTICS

Chemical and physical properties and acceptable ranges:


Exxon Publication 102-1096 0101C dated Oct. 96
Characteristics, dimensions, color, mechanical require-
ments, general properties, and test requirement
quality specification 700-90002.*

* See Appendix I.

TLFeBOOK
Appendix I
Quality Specification

225

TLFeBOOK
TLFeBOOK
Appendix J
Design Change Procedure

DOCUMENT CONTROL PROCEDURE

Effective
Rev. DCR Date Date Originator Description

A 08-19-93 M.B. Robbins Draft QA Policy


B 02-03-97 06-30-97 R. Bradley Original release
C 06-25-97 06-30-97 J. Passero DCR 97-0012
D 04-29-98 05-29-98 M. Teixeira DCR 98-0064
E 08-12-98 08-14-98 R. Bradley DCR 98-0148
F 09-14-99 10-08-99 M. Teixeira DCR 99-0057

Approval Signature: Date:

227

TLFeBOOK
228 Teixeira and Bradley

1.0 PURPOSE

1.1 This procedure describes the system to be used for


controlling quality system documents and data.

2.0 SCOPE

2.1 This procedure applies to all quality system docu-


ments identified within the quality system. It in-
cludes the initiation, review, approval, and distribu-
tion of new documents as well as changes to existing
documents and obsolescence of documents (e.g.,
quality assurance policies, interim specifications,
standard operating procedures, label text specifica-
tions, quality assurance specifications, material
specifications, engineering drawings, forms, product
specifications, technical files, device master records,
quality plans, etc.).

3.0 DEFINITIONS

3.1 Approved document: An approved document is a


document with approval from all required approv-
ing authorities.
3.2 Denied document: A denied document is a docu-
ment not acceptable to an approving authority(s).
3.3 Withdrawn document: A withdrawn document is a
document removed from the handling and ap-
proval cycle system (void).
3.4 Vendor or supplier verification: This is a response
from a vendor or supplier defining its agreement to
conform to the written requirements of a proposed
specification/drawing. (This includes the accep-
tance of a purchase order.)
3.5 Specification/procedure books: These are con-
trolled books assigned by Quality Assurance to

TLFeBOOK
Appendix J: Design Change Procedure 229

departments/areas, which contain the current of-


ficial approved copies of all assigned documents.
3.6 Master indexes: These lists, where practical, are
maintained by Quality Assurance to identify the
current revision of documents (e.g., QS index, SOP
index, MS index, forms index, PD index, etc.). In
most cases, the document number, title, revision
level, and/or effectivity date are identified. The revi-
sion for each quality assurance policy (QAP) section
is indicated in the table of contents. The effective
date identified serves as both the revision and ef-
fectivity dates. These lists are updated as necessary
in conjunction with the Notice of New Issue/Revi-
sion memo. (Note: Device master record revision ⫽
the date signed. Quality plan revision ⫽ revision
letter. Technical file elements are per procedure.)
3.7 Distribution matrix: This is a matrix detailing the
documents assigned to individual specification/
procedure books or departments/areas.
3.8 Change history: This is the historical information
for a document(s). The change history includes, at
a minimum, a description of the change(s), identi-
fication of any affected documents, the signature
and date of the approving individual, and the date
the change becomes effective. The change history
information for BioDerm documents is reflected on
the documents and the completed DCR form.
3.10 Documentation change request (DCR) form: This
is the document used to initiate changes to quality
system documentation. Changes to quality system
documentation include changes/revisions, new
issues/additions, deviations, and obsolescence/de-
letions.
3.11 Approving authorities: Those individuals responsi-
ble for reviewing, approving, or denying a change
to quality system documentation are the approving
authorities.

TLFeBOOK
230 Teixeira and Bradley

4.0 RESPONSIBILITIES

Quality Assurance is responsible for controlling and


maintaining quality system documents in accor-
dance with this procedure, excluding engineering
drawings.
Engineering is responsible for controlling and main-
taining engineering drawings in accordance with
this procedure.
4.2 All employees are responsible for adhering to the re-
quirements of this procedure when submitting new
or revised documents into the quality system.

5.0 REQUIRED EQUIPMENT

N/A

6.0 MATERIALS

N/A

7.0 RECORDS

Document change request form (DCR)

8.0 REFERENCE DOCUMENTS

Document indexes
SOP 100-003—Record retention procedure
SOP 100-008—Standard operating procedure
SOP 100-009—Quality assurance specifications
SOP 100-012—Label/labeling text specifications
SOP 100-013—Material specifications
SOP 100-017—Product specification procedure
SOP 100-024—Device master records
SOP 700-001—Technical file

TLFeBOOK
Appendix J: Design Change Procedure 231

Quality plans
Engineering drawings
Document distribution matrix

9.0 PROCEDURE
Making Changes to Quality System
Documentation
The originator/requestor of a change to quality system docu-
mentation shall obtain a copy of a document change request
(DCR) form from Quality Assurance.
The originator/requestor shall allocate the next available
DCR number from the DCR log. (The DCR log is located in
Quality Assurance.)
The originator/requestor shall complete the log for the
DCR number (number, originator, date DCR initiated, and
brief description).
As applicable, the originator/requestor will obtain an un-
controlled current copy of the quality system document that
requires a change or revision from a controlled book.
The originator will make the necessary changes to the
existing document by redlining the document to be changed.
In extenuating circumstances, work to a red-line document
will be permitted if all DCR form approvals have been given,
while processing/editing the document occurs.
Documents should reference the DCR number in the
header section of the document, if applicable, under “DE-
SCRIPTION.”

9.7 New documents or extensive changes/revisions to


documents should be done on disk in the appropriate
format.
9.8 When a change or revision affects more than one
document, the originator/requestor shall identify
responsibility for and/or submit changes for all af-
fected documents.

TLFeBOOK
232 Teixeira and Bradley

9.9 Documentation change request DCR form:

The originator/requestor shall complete the DCR form as


follows:
9.9.1 Record the request number, request date,
and originator/requestor.
9.9.2 Identify documentation types affected (e.g.,
SOP, MS, QS, LT, etc.).
9.9.3 Identify request type (e.g., change/revision,
new, deviation, obsolescence, etc.).
9.9.4 Identify effect code (if known).
9.9.5 Identify material disposition (if known).
9.9.6 Provide a complete and accurate reason for
the requested change. (As applicable, in-
clude any reference to a nonconformance
number, complaint number, audit observa-
tion number, etc.) Sufficient information is
needed for approval authorities to enable
them to have a complete understanding of
the nature and impact of the request.
9.9.6.1 If introducing a new document,
state the new documents applica-
tion.
Example: This specification pro-
vides a process for assembling. . . .
.
9.9.7 The originator/requestor must provide a
complete and accurate description of the re-
quested change.
Example:
Change from: paragraph 3.3, maximum
thickness 1.000 ⫹/- .005″
Change to: paragraph 3.3, maximum thick-
ness 1.000 ⫹/- .002″
Note: Define and indicate the change com-
pletely and precisely.

TLFeBOOK
Appendix J: Design Change Procedure 233

9.9.7.1 A description of the required


changes for complex/extensive doc-
umentation changes should be
neatly printed or typed on a sepa-
rate sheet of paper or on the back
side of the DCR form. Attach the
addendum sheet to the DCR Form.
9.9.8 Identify all known documents affected by
indicating title and revision, and assigned
to appropriate personnel.
9.9.9 The originator/requestor should acquire the
signature of his or her department manager
on the DCR form to indicate approval to pro-
ceed with the request and submit all re-
quired documentation to Quality Assur-
ance.
9.9.10 DCR processing, review, approval, and dis-
tribution.
9.9.11 Quality Assurance will log in and assign a
sequential request number for the DCR if
not already done.
9.9.12 Quality Assurance will review all DCRs and
associated documents for completeness and
to ensure that no request for a change com-
ing from other individuals or departments
exists and/or will vary the intent of the orig-
inal request.
9.9.13 Quality Assurance or designee will tran-
scribe the document/document changes us-
ing a word processor.
9.9.14 Engineering is responsible for revising and
controlling drawings. Engineering will re-
vise specified drawings as needed and for-
ward with the completed DCR form to Qual-
ity Assurance.
9.9.15 Quality Assurance or originator/requestor
will route the DCR form along with the new

TLFeBOOK
234 Teixeira and Bradley

or changed documents and all required at-


tachments to the applicable approving de-
partments as defined in Figure 1. The de-
partment head is considered the approving
authority unless otherwise designated/del-
egated. (Note: Changes to quality system
documents receive the same review and ap-
proval as the original document unless oth-
erwise indicated. Approval designation is at
the discretion of the Director of QA/RA.)
9.9.16 The approving authority is required to add
additional documents, not identified, that
may need to be revised as a result of the re-
quested change. As an example, an engi-
neering drawing change of a critical dimen-
sion may require changes in manufacturing
SOPs, material specifications, QA specifi-
cations, label text specs, etc. The approving
authority should add affected documents to
the DCR form and assign the person respon-
sible for making the appropriate changes.
All additional changes should be made and
submitted with the DCR form.
9.9.17 The approving authority approves docu-
ments by signing and dating the DCR form.
Any minor or conditional changes to be
made should be indicated by the approving
authority. The approving authority denies
documents by attaching a written explana-
tion of the reasons for disapproval. The ap-
proving authority must provide written in-
formation to indicate any changes that
would make the document acceptable (if ap-
propriate).
9.9.18 Quality Assurance issues, withdraws, or re-
turns the documents to the originator,
based on DCR approval status.

TLFeBOOK
Appendix J: Design Change Procedure 235

(QAP) (SOP) (MS) (LT) (PD) (QS) (IS) (DEV)

Sales x * - * - - * *

Business x * - x - - * *
Development

Technology x x x x x x x x

Quality x x x x x x x x
Assurance/Regulatory Affairs

Manufacturing x x x * x - x x

Finance x * x x x x * *

Clinical Affairs x * - x - - * *

Engineering x * * - x x * *

Facilities x * - - - - * *

President/CEO x * - x - - * *

Note: Approval can always be delegated upward without written notification.


QAP ⫽ Quality Assurance policy; SOP ⫽ standard operating procedure; MS ⫽ mate-
rial specification; LT ⫽ label text specification; PD ⫽ product specification; QS ⫽
Quality Assurance specification; IS ⫽ interim specification; EDS ⫽ Engineering
drawing specification; DEV ⫽ deviation; x ⫽ approval required; * ⫽ approval if de-
partment is affected (QA determines)

Figure 1 Necessary approving departments for certain forms.

TLFeBOOK
236 Teixeira and Bradley

9.9.19 The originator may review the denied DCRs


with the appropriate approval authorities
and may resubmit a revised DCR for consid-
eration by documenting the necessary
change(s) and recirculating the documents
in accordance with this procedure.

SIGNIFICANCE OF APPROVAL

Approval of a DCR signifies that the designated approving au-


thority has considered the best interests of the company and
has determined that the approval of the DCR, including any
deviations, will not adversely affect the safety and efficacy of
any item or product.

DELEGATING APPROVAL AUTHORITY

To delegate approval authority, permanently or temporarily,


forward a memo to Quality Assurance specifying the person(s)
authorized to approve documents, the dates for which the au-
thorization is valid, and any limitations (e.g., document types).
Quality Assurance will assign document numbers for new
documents. Quality Assurance will print the document in the
correct documentation form, route for approval, and distribute
copies of the approved document as per the distribution ma-
trix. Obsolete copies of all documents will be destroyed. The
“obsolete” master document will be stamped obsolete and filed
in Quality Assurance in accordance with the record retention
procedure.
When desirable, quality system documents shall be color-
coded to help rapidly identify them; the following colors will
be used as follows:

Quality Assurance policy Yellow/Melon


Standard operating procedures White

TLFeBOOK
Appendix J: Design Change Procedure 237

Material specifications Green


Product specifications Blue
Interim specifications Pink
Label text specs White
Quality Assurance specs Orange

Note: Only documents with a red or blue “Copy” stamp


are considered controlled documents.
Communication of the issuance of a new document or a
change to an existing document will be done by issuance of a
“Notice of New Issue/Revision” memo to department heads/
specification/procedure book holders. This memo lists the doc-
uments affected and effective date and serves as notification
of a change. Department heads are responsible for communi-
cating the changes to their employees and providing training
as necessary. The effective date of the document is also indi-
cated on the document.
Note: Label/labeling text specifications may become ef-
fective prior to issuance of the Notice of New Issue/Revision
Memo. (Only one master book is maintained in Q.A.) These
documents will be reflected on the Notice of New Issue/Revi-
sion Memo at the time of next document update.
Finance is responsible for forwarding any copies of mate-
rials specifications, drawings, label text specifications, etc.
that a supplier may need to meet purchase order or contract
requirements. Finance will request vendor or supplier verifi-
cation of the change. A vendor’s or supplier’s reply is expected
within 4 to 6 weeks; nonreply is construed as agreement to
the change, “as is.”
At the time of distribution of any documents, Quality As-
surance will update applicable document index(es).

10.0 SECURITY CONTROL


CLASSIFICATION

10.1 All documents that make up the quality system are

TLFeBOOK
238 Teixeira and Bradley

considered “restricted” documents. Controlled cop-


ies of documents are issued by Quality Assurance
and identified by a red or blue “Copy” stamp im-
print. Controlled copies are to be made from the
master document only.
10.2 Copies of documents for informational use or re-
dlining may be made from controlled books. These
copies are considered uncontrolled and may be
stamped in red “Uncontrolled Copy.”
Copies of documents for external distribution shall be
stamped in red “Confidential Copy.” External distribution of
documents should be limited and authorized by department
managers.

11.0 ELECTRONIC DOCUMENT CONTROL

11.1 All documents controlled by this procedure are


stored via computer.
11.2 Entering new documents, deleting obsolete docu-
ments, and making revisions to current documents
may only be accomplished by authorized personnel
(primarily Quality Assurance; Engineering for
drawings).
11.3 Electronic documents will be periodically backed
up.

TLFeBOOK
Index

Acceptance criteria, 8, 50, 51, 21 CFR Part 820, Subpart C,


54, 55, 152, 154, 170 Section 820.30(a),3
Acronyms, 68 21 CFR Part 820, Subpart C,
Approval for release, 51 Section 820.30(b),15
Audit, 151 21 CFR Part 820, Subpart C,
Section 820.30(c), 28
Basic research, 7
21 CFR Part 820, Subpart C,
Biocompatibility, 29, 44, 54,
Section 820.30(d), 50
79, 84, 92, 100, 102–109, 183
21 CFR Part 820, Subpart C,
Biological properties, 43
Section 820.30(e), 60
Carcinogenesis, 105 21 CFR Part 820, Subpart C,
21 CFR Part 820, Subpart Section 820.30(f ), 78
C,Section 820.30, 3, 15, 28, 21 CFR Part 820, Subpart C,
50, 60, 78 Section 820.30(g), 90

239

TLFeBOOK
240 Index

21 CFR Part 820, Subpart C, Design transfer, 123, 137–141


Section 820.30(h), 141 Design validation, 89–121,
21 CFR Part 820, Subpart C, 123, 152, 155, 163, 171
Section 820.30(i), 144 Device history file, 54,148
21 CFR Part 820, Subpart C, Device master record, 54, 55–
Section 820.30( j), 148 57, 162, 228–230
Claims, 31, 38, 47, 168, 188 Document change request
Class I medical devices control (DCR), 145–146
exemption, 3 Documentation, 38, 78, 81, 86,
Clinical use test, 26, 42, 91, 129, 144–145
127 Draize test, 117
Communication, 5, 14, 67
Concept document, 29, 30 Effective listening, 69
Contractual requirements, 47 Erythrocyte stability, 119
Controllable cause, 94, 206
Critical path, 18, 19 Flowcharts, 7
Customer, 2, 28, 32, 50, 70, Failure-mode effects analysis
96, 161 (FMA), 93–94
Cytotoxicity, 80, 108, 111–115 Failure-mode effects analysis
(FMEA), 93, 95–97
Failure-mode effects criticality
Dermal sensitization, 116 analysis (FMECA), 93, 97
Design change, 143 Fault tree analysis (FTA), 93–
Design control definition, 10 94
Design controls, first law, 7
Design controls, fourth law, 9 Gantt chart, 19
Design controls, second law, 8 Generally recognized as safe
Design controls, third law, 9 (GRAS), 103
Design history file, 51, 60, 92,
147–149, 158, 162, 173 Hazard analysis and critical
Design inputs, 8, 25–47, 63, control points (HACCP), 93,
144, 149, 152–153 97–99
Design outputs, 8, 49–57, 91, Hazards, 46, 65, 81, 84, 87,
137, 149, 154 93, 93, 94, 95, 96, 169, 194,
Design requirements, 3, 27, 196, 202–203
61, 77, 162 Hemocompatibility, 108, 118
Design review, 8, 47, 51, 54, Hemolysis, 108–109, 118–119
59–75, 79, 141, 154–156, Human factors engineering,
162, 164, 169–172 81

TLFeBOOK
Index 241

Implantation tests, 115–119 Product initiation request, 29,


Indications for use. See Prod- 164, 165, 175–177. See also
uct specification Concept document
Instructions for use, 39, 181 Product specifications, 29, 47,
Interactions, 17, 101, 113, 167 128, 214
Pyrogenicity, 107, 108, 120–
121
Labeling, 29, 33, 47, 51, 54,
55, 63, 77, 80, 82, 84, 87,
91, 92, 96, 100, 144, 145, Quality system inspection tech-
162, 166, 168, 186, 203 nique (QSIT), 151
Leachable substances, 110,
119 Relevant use setting, 32, 39–
40
Reliability analysis, 84
Market survey, 26
Reproductive toxicity, 110
Marketing, 6, 13, 16, 22, 30,
Risk analysis, 54, 65, 79, 81,
38, 40, 161, 163, 165
82, 83, 91, 99, 139, 149, 152,
Material safety data sheet,
153, 154, 155, 157, 160, 161,
102
167, 169
Material specifications, 54,
Risk analysis master record,
141, 221–224
83, 169
Meeting dynamics, 66
Meeting minutes, 65, 72
Mutagenicity, 108, 119 Screening tests, 105, 109
Sensitization, 108, 110, 116,
117
Output. See Design outputs Skin irritation, 110, 116
Special cause variation, 151
Packaging, 29, 31, 32, 44, 45, Standard methods, 79
54, 91 Statistical process control,
Patient population, 32, 41 151–153
Program evaluation and re- Sterilization process, 32, 48,
view technique (PERT), 17– 100
19, 20 System variation, 151–153
Planning techniques, 16–24
Problem solving, 74 Tasks, 8, 14, 15, 17, 18, 19,
Process validation, 127–135 22, 60, 168, 211
Product development, 1, 5, 6, Toxicity, 17, 48, 84, 101, 103,
15, 67, 72, 143, 172 107, 110–118

TLFeBOOK
242 Index

Tripartite biocompatibility, USP biological reactivity tests,


103 115

Unresolved discrepancies, 152, Validation, 9, 54, 65, 70, 88,


155 89–121, 123
Use conditions, 32, 89, 102, Verification, 8, 9, 54, 55, 77–
148, 155, 170 88

TLFeBOOK

You might also like