0% found this document useful (0 votes)
68 views10 pages

New Gentleyag

This document describes cryogen spray alignment procedures for the GentleYAG laser system. It outlines three tests: 1) cryogen alignment to verify proper nozzle alignment, 2) cryogen coverage to verify sufficient spray duration, and 3) cryogen air bubble detection to ensure the system detects bubbles. The alignment test uses distance gauges to check for excess misting. Coverage times are provided for different gauge sizes. The bubble test disconnects the cryogen to induce a fault, verifying the proper error message displays.

Uploaded by

Alonso Garcia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
68 views10 pages

New Gentleyag

This document describes cryogen spray alignment procedures for the GentleYAG laser system. It outlines three tests: 1) cryogen alignment to verify proper nozzle alignment, 2) cryogen coverage to verify sufficient spray duration, and 3) cryogen air bubble detection to ensure the system detects bubbles. The alignment test uses distance gauges to check for excess misting. Coverage times are provided for different gauge sizes. The bubble test disconnects the cryogen to induce a fault, verifying the proper error message displays.

Uploaded by

Alonso Garcia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 10

New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

New GentleYAG
Cryogen Spray Alignment Procedure

8503-01-0845, Revision A Candela Corporation Proprietary Page 1 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

CRYOGEN SPRAY TESTING AND ALIGNMENT PROCEDURE

You will need the following supplies to perform these tests:


• Laser Safety Glasses
• Cryogen Coverage Template 1301-00-8291 available from Candela Corporation.
• VPYAG distance gauge.
• .050 inch Allen Wrench

The following tests are described in this section

1. Cryogen Alignment: Verifies the cryogen spray nozzle is properly aligned with the
distance gauge ring.

2. Cryogen Coverage: Verifies the spray duration required to fill the distance gauge
ring.

3. Cryogen Air Bubble Detector: Verifies that air bubbles in the cryogen line are
detected and the associated “fault” message is displayed on the system.

8503-01-0845, Revision A Candela Corporation Proprietary Page 2 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

8503-01-0845, Revision A Candela Corporation Proprietary Page 3 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation
Test #1 Cryogen Alignment

Purpose:

To verify the cryogen spray nozzle is properly aligned with the distance gauge ring
Procedure:
1. Put the laser in “standby”. Caution- laser should remain in “ Standby” mode
for duration of test
2. Select “DCD –on”.
3. Install GentleYAG 12mm distance gauge PN 7122-00-3109
4. Select 30ms of DCD Spray.
5. Point hand piece away from objects and personnel (toward the floor). View
the contact ring of distance gauge, looking from the hand piece.
6. Press purge button.
-DCD Spray should flow completely through the contact ring. There may be a
minimal spray mist seen hitting the contact ring. No spray should be spraying
beyond outside of the contact ring.

Acceptable-No mist on contact ring

No Cryo clipping on 12
mm distance gauge

Acceptable-Small amount of mist on contact

Small Cryo clipping on 12


mm distance gauge

Unacceptable results-Excess mist on contact ring

Cryo clipping on 12
mm distance gauge

8503-01-0845, Revision A Candela Corporation Proprietary Page 4 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

Results:
Acceptable alignment- no further action needed.
Unacceptable alignment, repeat the test with a different distance gauge

If the test shows acceptable alignment with the new distance gauge discard the
original distance gauge

If the result is still “unacceptable”, Realign the cryogen nozzle.

8503-01-0845, Revision A Candela Corporation Proprietary Page 5 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

Test #2 - Cryogen Coverage:


Purpose:

To verify the proper spray duration required to fill the distance gauge ring. (Note: the
distance gauge ring is larger than the spot size marking).
• If you use multiple size distance gauges, these checks should be made utilizing
the smallest and largest distance gauges in your product configuration.
• The following sizes are available
GentleYAG Distance Gauge P/N (5 Pack)

3mm

8mm 7122-00-3502

10mm 7122-00-3503

12mm 7122-00-3109

15mm 7122-00-3504

18mm 7122-00-3505
• Note: The below tests and values are not intended to represent treatment
parameters, but rather provide a check on proper functionality of the handpiece
and provide a reference for the user to help identify changes in the handpiece
operation.
Procedure:
1. Put on appropriate laser safety glasses
2. Put the laser in “standby Caution- laser should remain in “Standby“mode for
duration of test
3. Select “DCD –on”.
4. Install appropriate distance gauge from chart below.
5. Select “Spray Duration” per chart.
6. Place hand piece distance gauge over desired distance gauge spotson template
(1301-00-8291).
7. Press purge button.
8. Remove hand piece QUICKLY from template .

DCD Spray should completely fill inner spot.


• Spray is acceptable to be found outside of spot as long as inner spot is
completely filled. (This spray may be from reflected spray off the paper).
• As a guide -select the res pective spray duration from the following table:

VPYAG

8503-01-0845, Revision A Candela Corporation Proprietary Page 6 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

“Typical” spray duration


Distance gauge spot size
to minimally fill the inside
ring spot.

3 mm 30 ms

8 mm 30 ms

10 mm 30 ms

12 mm 40 ms

15 mm 50 ms

18 mm 60 ms

Results: If the spot doesn’t fill or a leak is noted, the handpiece assembly should be
replaced or contact Candela Technical Support.

8503-01-0845, Revision A Candela Corporation Proprietary Page 7 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation
Test #3 Cryogen Air Bubble Detection

Purpose:

To ensure that that air bubbles in the cryogen line are detected and the associated
“fault” message is displayed on the system.

Important: This procedure must be done with each GentleYAG handpiece delivery
system in your possession. The subsystem being tested is inside of each delivery
system, not in the laser system.

Procedure

1 Insure that the installed DCD canister is not empty.


Warning: Put on appropriate laser safety glasses
2. Turn on the system and allow the “WARM UP” to
complete.
3. With the Electrical Connector installed, disconnect the Cryogen Connector
(see Figure 3).

Electrical Connector

Cryogen Connector

Figure 3 - Delivery System Connections

4. Set the DCD spray on the front panel to the highest setting.

5. Set the FLUENCE to the lowest setting (this varies by spot size).

6. Enter the “Ready” mode, and calibrate the laser by pressing on the footswitch.

7. Remove the handpiece from the CALPORT and aim in direction away from
personnel (such as the floor).

8. Depress the footswitch to pulse laser continuously until a “PURGE REQUIRED”


window box appears on the user screen and lasing ceases. This should happen
in less than 50 pulses.

8503-01-0845, Revision A Candela Corporation Proprietary Page 8 of 10


New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation

Results Comments Action

If you get This particular You may continue use


“PURGE handpiece of this handpiece
REQUIRED” passed the
test

If “PURGE This particular Discontinue use of


REQUIRED” handpiece this handpiece and
does not failed the test call Candela Tech
occur Support at (508) –358-
7637 ex 336 or (800) –
733-8550 ex 336

9. Reconnect the Cryogen Connector (see Figure 3) and press reset to use the
system, or repeat the test on other GentleYAG handpiece delivery systems.
10. Depress “purge” may be required until the line is refilled.