iU22iE33 Software Reload A-C Carts
iU22iE33 Software Reload A-C Carts
Tools:
Software CD with proper software level
Licensed Options disk of the system
System Back-up
Software reload:
Boot System
Press DEL to go into BIOS on keyboard
Log into BIOS
o Enter BIOS Password
Select Advanced BIOS Features
Set CD drive as first boot
o Select device USB CD-drive
Install software disk into drive
Press F10 (Text/Cursor Button)
o Save and exit BIOS
System will boot to a DOS prompt box on screen
o Select Y to retain the hostname
o Select H when prompted
o Select Y to continue
Install message
o Remove/eject software CD
When Screen goes black and there is no CD/DVD activity, Remove Disk 1
Note:
If installing on an iE33, insert disk 2
If loading Vision 2010 S/W 5.2.0.289 or higher, then you will have to insert the Drivers and
Printers disks when prompted.
If unable to configure Bluetooth – hit cancel. The system does not have a Bluetooth dongle
installed.
Once configured, system will shut down
To access diagnostics:
Boot System
o System should boot to an error 11.xxx.xxx.
Select setups
Select Tech admin tab, if not already selected
o User name “us_techuser” temporary user name for 24 hours only
o Password “Jc2Dw7KbrTx7” temporary password for 24hours only
Security tab – to create a permanent account
o Add techadmin
Create user
Set user as “ultrasound”
Create password
Set password as “ultrasound”
1
iU22/iE33 Software Reload A-C Carts
Select Install/Upgrade Tab
o Select Manage System Button
Select Check/Update System Configuration Button
Select Check All
Select Configure
Select OK, on pop up window
o Select Quit , in the upper right hand side
o System will reboot
On the reboot:
Go into BIOS
o Set HDD-1 back to first boot device
Log into Tech Admin with newly created Tech Admin
o Name: ultrasound
o Password: ultrasound
Select Test/Utilities
o Calibrate Touch Panel
Select arrow next to Platform
Select arrow next to Touch Panel
Select Calibration
Select Execute
o Run Diagnostics
Select System
Select Execute