Troubleshooting Reference - 7000
Troubleshooting Reference - 7000
Revision: A
Released: 26-August 2011
Author: Adam Milecki
Table of Contents
Troubleshooting Common Failures............................................................................................................... 3
Power-On and Communication ................................................................................................................ 3
Power Switch Not Working – TQ Always On......................................................................................... 3
Pump and Vacuum .................................................................................................................................... 3
Rough Pump Not Turning On ................................................................................................................ 3
Turbo Pump Not Turning On or Not Reaching 100% Speed ................................................................. 3
PCA RF Dipping.......................................................................................................................................... 3
Quad Driver Unable to Dip Within Range ............................................................................................. 3
Collision Cell Driver Unable to Dip Within Range ................................................................................. 4
Autotune ................................................................................................................................................... 4
Autotune Failure – Abundance, Mass Axis, and/or Peak Width........................................................... 4
Autotune Failure – Gain ........................................................................................................................ 5
Front Tailing on Peaks During Scan ....................................................................................................... 5
Dark Current Test ...................................................................................................................................... 6
Air and Water Check ................................................................................................................................. 6
High Air Level ........................................................................................................................................ 6
High Water Level ................................................................................................................................... 7
Chemical Test ............................................................................................................................................ 7
EI & EIEX Signal-to-Noise (SNR)............................................................................................................. 7
EIEX Relative Standard Deviation (RSD) ................................................................................................ 8
Faults ......................................................................................................................................................... 8
1.1 Repeller driver cannot maintain the requested voltage................................................................. 8
1.2 Ion body driver cannot maintain the requested voltage. ............................................................... 8
3.0 L13 (Q1 shroud) lens driver cannot maintain the requested voltage............................................. 8
3.2 Collision cell hexapole DC driver cannot maintain the requested voltage. .................................... 8
3.3 L16 (Iris) lens driver cannot maintain the requested voltage. ........................................................ 9
Appendix ..................................................................................................................................................... 10
Faults ....................................................................................................................................................... 10
Mainboard 1, Fault Register 1............................................................................................................. 10
Mainboard 1, Fault Register 2............................................................................................................. 11
Mainboard 2, Fault Register 3............................................................................................................. 13
Mainboard 2, Fault Register 4............................................................................................................. 14
Fault Masks ......................................................................................................................................... 16
Page | 2
Troubleshooting Common Failures
Power-On and Communication
Power Switch Not Working – TQ Always On
Issue The TQ powers on when the power cord is connected and the power switch cannot
control.
Fix Replace 0950-4826 E-Module power supply.
Issue The turbo pump will not power on or not reach 100% speed.
Fix Try the following fixes starting from the top and working down the list until the issue is
resolved.
1. Verify rough pump powers on and hose is properly connected to turbo pump.
2. When the TQ is powered on, verify a vacuum is pulled on the manifold. If not,
there is a large leak. Common leak locations are: pump hose, analyzer door
(front analyzer may require a light force to help vacuum catch), collision cell
cover, transferline, CI flow module interface, and EI calibrant/vent valve.
3. Verify rough vacuum gauge is properly connected. Gauge may be defective
and require replacement.
PCA RF Dipping
Quad Driver Unable to Dip Within Range
Page | 3
a. Check connector and pins on main cable harness. Rework or replace
G7000-60826 main cable harness.
b. Verify with a voltmeter that 24V is being supplied through the quad
driver power connector yellow pins. If not, replace this cable on the
inside of the E-Module or G7000-61008 bus PCA or the entire E-
Module assembly.
3. Make sure both quad driver coil slugs are even.
4. Verify the bias voltage with a voltmeter. On the quad driver PCA, probe TP11
“Bias Monitor” (next to the +21V LED). It should measure 110mV +/- 5mV. If
not with the instrument power on, reset the bias by adjusting R201 and R202
potentiometers fully counter-clockwise. Then adjust R201 clockwise until TP11
measures 55mV. Finally adjust R202 clockwise until TP11 measures 110mV.
5. Disconnect ribbon cable from E-Module to quad driver PCA. Verify all pins
within connectors are straight and none are missing. Reseat ribbon cable.
6. Replace G7000-61015 quad driver PCA.
7. Verify quad bridges are connected and aligned properly without shorting.
8. Verify quad is defect free and good electrical contact is made when installed.
Fix by reseating or replacing G7000-60155 quadrupole.
Issue Unable to dip collision cell driver PCA within the specification.
Fix Try the following fixes starting from the top and working down the list until the issue is
resolved.
1. Check to see if a fault is raised. Any fault in the system will prevent the quad
driver electronics from turning on.
2. Disconnect ribbon cable from E-Module to collision cell driver PCA. Verify all
pins within connectors are straight and none are missing. Reseat ribbon cable.
3. Replace G7000-61005 collision cell driver PCA.
4. Verify all internal wires to collision cell are properly connected.
5. Verify there are no obstructions inside the collision cell that may be shorting
the rods.
6. Replace G7000-60262 collision cell assembly.
Autotune
Autotune Failure – Abundance, Mass Axis, and/or Peak Width
Issue Autotune failed with one of the following messages: “Check Abundance failed”, “Adjust
Mass Axis failed”, or “Adjust Peak Width failed”.
Fix Try the following fixes starting from the top and working down the list until the issue is
resolved.
1. Check calibrant level, make sure calibrant bulb is free of cracks, and verify
calibrant valve cable is connected.
2. For CI, make sure reagent gas flow is on and flow readback is correct.
3. Tighten transferline nut.
4. Check quad driver dip level at 1000 AMU is less than 73.3%.
5. Verify all internal detector wires are connected (HED, EMV, Red Signal).
Page | 4
6. For CI, check filament wire straightness. Replace G1999-80070 CI filament if
warped. Also, make sure the source lens stack is tight.
7. Replace G7000-61015 quad driver PCA.
8. Replace G7000-60155 quadrupole.
Issue Autotune failed with the following message: “Adjusting gain failed, insufficient valid
data points”.
Fix Run Dark Current Check against the following specifications.
• Baseline average <= 60.0
• Baseline threshold <= 50.0
• Pulse count with HED on <= 50.0
• Pulse count with HED off <= 25.0
Try the following fixes starting from the top and working down the list until the issue is
resolved.
1. Reseat all detector external cables and internal wires (HED, EMV, Signal).
Check MS2 heater wire and make sure there is a minimum of 2mm clearance
between all metal parts.
2. Replace G2571-60806 signal cable.
3. From where tune left off, rerun gain curve using a different mass. For too few
pulses, decrement mass by 1. For too many pulses, increment mass by 1.
4. Clean face of HED, horn of EM, and crevices of structure.
5. Replace G7000-80102 detector assembly.
Page | 5
Dark Current Test
Issue Air and Water Check failed with ratio N2 to m/z 69 > 10%
Fix Generally, the air level decreases quickly over time so if this is the case, wait longer for
the high temperatures and vacuum to remove remaining nitrogen.
In atmospheric air, the ratio of O2 to N2 is about 27%. If close to this ratio exists within
the system, there is likely an air leak. Try the following fixes starting from the top and
working down the list until the issue is resolved.
1. Verify collision cell N2 was off and allotted time to pump out before recording
Page | 6
abundances.
2. Verify proper tune abundances for m/z 69 (>200,000 for EI, >400,000 for EIEX).
If too low, there may be an issue with the EI calibrant valve.
3. Use a can of compressed air and start MS1 profile on the target mass (ie Aero-
duster is m/z 83). Spray air slowly at different areas of the system including
transferline, CI flow module if applicable, front window, EI calibrant valve,
edges of analyzer doors, feed-through on analyzer doors and manifold,
collision cell cover and tubing, rear cover, and turbo pump.
4. If a leak is not found on the TQ, perform the same leak checking on the GC and
gas tubing.
5. Replace collision cell EPC module.
Issue Air and Water Check failed with ratio H2O to m/z 69 > 16%
Fix Generally, the water level decreases over time so if this is the case, wait longer for the
high temperatures and vacuum to remove water.
If the water level is not decreasing over time or is taking longer than expected, try the
following fixes starting from the top and working down the list until the issue is
resolved.
1. Verify GC heater zones are turned on.
2. Bake and purge the system. Set TQ temperatures to 300/180/180. Set GC
temperatures to 250/150/250, GC column flow to 2mL/min, and GC collision
cell gases to 2mL/min N2 and 3mL/min He. Allow the system to sit for at least
1 hour then return to normal operating setpoints.
3. Inject a large volume of solvent to flush the system.
Chemical Test
Issue Failing EI SNR < 100:1 or EIEX SNR < 500:1 using the checkout method with 3 MRM
transitions and gain of 100.
Fix Most new systems require some conditioning prior to achieving the required results.
This may include high GC temperatures, high column flow, and/or large volume
injections.
If you have run 4 or more injections and still fail, check to see if the results show an
improving trend. If so, continue to inject until passing or until leveling off.
Page | 7
2. Follow Dark Current Test troubleshooting instructions for detector noise
issues.
3. Follow High Air Level troubleshooting instructions for leak issues.
Issue Failing EIEX RSD > 8% of 8 sequential injections using the checkout method with 3
MRM transitions and gain of 100.
Fix Look through the individual areas of each injection. Determine if one or two injections
are outliers and if so continue to run sequential injections until these are purged out of
the 8 that are used for RSD.
For a downward trend throughout the sequence, bake and purge the GC column as
well as the TQ source for a couple hours.
For erratic results, check the peak shape. If the top of the peaks are being missed, it
may be due to an issue with the method. For tailing, perform a column cut on the inlet
end.
Faults
Page | 8
3.3 L16 (Iris) lens driver cannot maintain the requested voltage.
• End of clip that holds EM is shorting to iris lens.
• An insulator is missing from the iris lens.
• Metal shavings inside iris lens gold screw hole.
Page | 9
Appendix
Faults
Mainboard 1, Fault Register 1
Page | 10
Mainboard 1, Fault Register 2
Page | 11
is needed to clear condition.
2.26 Zone 3 Runaway MS1 heater has exceeded its maximum temperature. Power cycle is
needed to clear condition.
2.27 n/a masked
2.28 n/a masked
2.29 n/a masked
2.30 INTERNAL masked
SHUTDOWN
2.31 SHUTDOWN masked
Page | 12
Mainboard 2, Fault Register 3
Page | 13
Mainboard 2, Fault Register 4
Page | 14
SHUTDOWN
4.31 SHUTDOWN masked
Page | 15
Fault Masks
Page | 16