INTRA-AORTIC BALLOON
REFERENCE SIZING GUIDE
Approx. ht: < 5'0" Approx. ht: 5'0" to 5'4" Approx. ht: 5'4" to 6'0" Approx. ht: ≥ 5'4"
(< 152cm) (152cm - 162cm) (162cm - 183cm) (≥ 162cm)
6'
5'
4'
Note: This
information is
to be used as a
guidance only.
Clinical information
and patient factors
such as torso length
should be considered
when selecting the
appropriate balloon
size.
MEGA AND LINEAR IAB CATHETER
® ®
1 2 3 4
Firmly attach one-way Apply a 30cc vacuum. Remove syringe while Remove stylet, then manually
valve to male luer fitting keeping one-way valve flush inner lumen with 3-5cc
of IAB catheter. in place. of flush solution.
5
4 6
5 7 8
Insert needle at 45˚ angle Make small incision at Insert vessel dilator over Spread tissue at incision
or less, then insert 0.025" exit of guidewire. guidewire, tapered end first, to facilitate sheathless
(0.06cm) guidewire. then remove. insertion.
9 10 11
Remove IAB catheter from Advance IAB catheter into Secure IAB catheter to
T-handle by pulling STRAIGHT artery using short strokes patient’s leg using STATLOCK®
out to avoid damaging it. until correct placement is Stabilization Device or sutures.
Do not dip, wipe, or handle achieved, then advance
membrane prior to insertion. sheath seal as close to
insertion site as possible.
MEGA AND LINEAR IAB CATHETER
Sheathed Insertion
5 6 7 8
Insert needle at 45˚ angle Make small incision at Insert introducer dilator 1 Advance sheath over
or less, then insert 0.025" exit of guidewire. into sheath hub and twist guidewire into artery
(0.06cm) guidewire. lock in place to secure. using a rotary motion.
2 Withdraw introducer dilator
leaving sheath in place.
9 10 11
Remove IAB catheter from Advance IAB catheter through Secure IAB catheter to
T-handle by pulling STRAIGHT sheath using short strokes until patient’s leg using STATLOCK®
out to avoid damaging it. correct placement is achieved, Stabilization Device or sutures.
Do not dip, wipe, or handle then advance sheath seal into
membrane prior to insertion. hub of sheath.
12 13 14
Connection to IABP
15 16 17
Remove one-way valve Connect IAB catheter’s Connect male luer fitting
from IAB catheter. male luer fitting to female to Safety Disk of IABP.
luer fitting of catheter
extender.
SENSATION PLUS AND SENSATION IAB CATHETER
™ ®
1 2 3 4
Firmly attach one-way Apply a 30cc vacuum. Remove syringe while Remove stylet, then manually
valve to male luer fitting keeping one-way valve flush inner lumen with 3-5cc
of IAB catheter. in place. of flush solution.
5
4 6 7 8
Insert needle at 45˚ angle Make small incision at Insert vessel dilator over Spread tissue at incision
or less, then insert guide- exit of guidewire. guidewire, tapered end first, to facilitate sheathless
wire (8Fr. IAB: 0.025" then remove. insertion.
(0.06cm) / 7Fr. IAB: 0.018"
(0.05cm).
5 6 7 8
Insert needle at 45˚ angle Make small incision at Insert introducer dilator 1 Advance sheath over
or less, then insert guide- exit of guidewire. into sheath hub and twist guidewire into artery
wire (8Fr. IAB: 0.025" lock in place to secure. using a rotary motion.
(0.06cm) / 7Fr. IAB: 0.035" 2 Withdraw introducer dilator
(0.09cm). leaving sheath in place.
9a
9b 10 11 12
7Fr. IAB only: Remove IAB catheter from Advance IAB catheter through Secure IAB catheter to patient’s
9a Remove 0.035" (0.09cm) T-handle by pulling STRAIGHT sheath using short strokes until leg using STATLOCK® Stabilization
guidewire and out to avoid damaging it. correct placement is achieved, Device or sutures (SENSATION
9b Replace with 0.018" Do not dip, wipe, or handle then advance sheath seal into PLUS includes STATLOCK®
(0.05cm) guidewire. membrane prior to insertion. hub of sheath. in IAB box).
Connection to IABP
16 17
15 18
Remove one-way valve Connect IAB catheter’s Connect male luer fitting of
from IAB catheter. male luer fitting to female catheter extender to Safety
luer fitting of catheter Disk of IABP. Insert fiber-optic
extender. sensor connector into IABP’s
sensor input receptacle until
it clicks.
CS300 INITIAL SET-UP
IAB
SENSOR
INPUT
4
Connect fiber-optic IAB:
for MAQUET/Datascope
fiber-optic IAB.
OR
2
Open helium tank 4
IABP OFF ON (if not already opened).
Connect pressure cable:
1 Turn IABP on. for conventional IAB.
3
Connect ECG
5
cable.
Connect IAB.
ALARM TRIGGER
MESSAGES
ADV ISORIES
HEART RATE
bpm
ECG
U LEAD
SYSTOLIC
No mmHg
U GAIN
Zer o
DIASTOLIC
PRESSURE mmHg
SOURCE
SLOW GAS
AUG.
ALARM mmHg
Verify Press
AUTO START
79
U LEAD
SYSTOLIC
No mmHg
HEART RATE U GAIN
Z ero
AUTO
bpm
Semi Manual Zero START STANDBY
Auto Pressure
PRE SSURE
DIASTOLIC
126
mmHg
SOURCE
SYSTOLIC
No mmHg Pediatric Only Hold 2 Sec.
Zero IA B FIL L MODE MEAN
mmHg
TRIGGER SOURCE
85 DIASTOLIC
mmHg
ALARM
IAB FREQUENCY
MESSAG ES
Intelligent Counterpulsation
A U G M E N TAT I O N
SL OW GA S
A L A RM
IAB INFLATION
TRIGGER
IAB DEFLATION
AUG.
mmHg
.2
96
OPE RATION AUG. ALARM
Defl. MODE mmHg
MEAN ADV ISORIES
mmHg
ECG MAX Earlierbpm
HEART RATE
Later Earlier Later
1:1
140
ECG IA B STA TUS
Pressure U LEAD
AUG. No SYSTOLIC
mmHg
mmHg U GAIN
Zer o
Pacer V/AV 1:2
DIASTOLIC
Pacer A 120
PRESSURE
AUG. ALARM mmHg
mmHg
SOURCE
1:3
OFF
Internal IAB FILL MODE MEAN
mmHg
SLOW G AS
AUG.
ALARM
m/Advisory Helium Battery mmHg
Ref.
mmHg
Verify Press
AUTO START
IAB Aug. Open Pump Freeze
Fill Pressing the START key Alarm Menu Options Display
Automatically purges and fills IAB
Hold 2 Sec.
Automatically selects most appropriate
lead and trigger
User Print
Automatically sets inflation and deflation Preferences Strip
timing
Hold For Continuous
| 13 | IAB Insertion / CS300 Operation | Cardiac Assist | Back CS300 SET-UP
| 14 | IAB Insertion / CS300 Operation | Cardiac Assist |
VAULT COPY CS300 SET-UP
TRIGGER SOURCE
ECG
Pressure
Pacer V/AV
Pacer A
Internal
CS300 TRIGGERS
Triggering CS300 Triggers
A Trigger is the signal the CS300 uses to ECG
identify the beginning of the next cardiac cycle Trigger event is the R-Wave
When the CS300 recognizes the trigger event, Trigger of choice when an adequate R-Wave is present
it will deflate the balloon if not already deflated Pacer spikes are automatically rejected
Trigger Source keys are only active while in Gain can be adjusted while in Semi-Auto or Manual
Semi-Auto operation mode operation mode
Pressure
Trigger event is the systolic upstroke
TRIGGER SOURCE Trigger of choice (with a regular rhythm) when an
adequate R-Wave is not present
ECG
A fixed pressure threshold can be manually set while
Pressure
in Semi-Auto or Manual operation mode
Pacer V/AV
Pacer A Pacer V/AV
Internal Trigger event is the Ventricular pacer spike
Typically used when ECG triggering is unsuccessful
and a V or AV pacer is being used
Must be 100% paced
Only available in Semi-Auto or Manual operation mode
Pacer A
Trigger event is the R-Wave
Recommended only if atrial pacer tails are interfering
with R-Wave detection while in ECG trigger
Only available in Semi-Auto or Manual operation mode
Internal
Trigger event is asynchronous at a fixed rate of 80 BPM
Only used when there is no mechanical cardiac cycle
(i.e.: cardiopulmonary bypass or asystole)
Rate can be adjusted from 40 to 120 BPM
Only available in Semi-Auto or Manual operation mode
THEORY OF COUNTERPULSATION THERAPY
Benefits Benefits
Increases coronary artery perfusion Decreases afterload
Increases mean arterial pressure Decreases cardiac workload
Increases cardiac output
TIMING
Timing refers to the positioning of inflate
and deflate points on the arterial pressure Diastolic
Augmentation
waveform.
Unassisted Assisted Systole
Proper IABP Timing Systole
INFLATION
Occurs at the dicrotic notch
Appears as a sharp “V”
Ideally diastolic augmentation rises
above systole
Unassisted End
DEFLATION Diastolic Pressure
One Cardiac Cycle Assisted End
Occurs just prior to the next systolic event Diastolic Pressure
Results in a reduction in assisted end
diastolic pressure
Results in a reduction in assisted systolic
pressure
TIMING ERRORS
Aortic Dicrotic
Valve Notch
Closure
TIMING ERRORS
1 2 3 4
Fully close helium tank Slowly loosen yoke Remove helium tank. Replace washer,
valve clockwise. T-handle counter- if available.
clockwise.
Helium
5 6 7 8
Install fresh helium tank. Fully tighten yoke T-handle Slowly open helium Verify full helium level
clockwise. tank valve counter- via indicator on monitor
clockwise. display.
126
ECG MAX E a rl i e r Lat er Ear l ier Lat er
SYSTOLIC 1:1
P r e ssu r e N o
mmHg
Z e ro
P ac e r V /AV 1:2
ALARMS
85
P ac e r A 1:3
OFF
In te r n al DIASTOLIC
mmHg
2.2
96
Augmentation BelowLine
Limit Set Autofill Failure
Defl.
Alarm M E A Help
N Ref. ECG/AP Inflation
Mute mmHg Sources Inter val
116
IAB
Fill
Hold 2 Sec.
AUG.
mmHg
Aug.
Alarm
Open
Menu
Pump
Options
Freeze
Display
ALARMS
User Print
Preferences Strip
22 min in
Standby START START
IAB STATUS check tubing
for blood
ALARMS
Unable to Calibrate IAB Optical Sensor IAB Optical Sensor Calibration Expired
Patient’s pulse pressure is When patient’s pulse A calibration update has Assess patient to
inadequate for calibration. pressure improves, press been intentionally post- determine if a brief pause
ZERO PRESSURE key poned because either in assist would be toler-
for 2 seconds while the IABP patient’s mean arterial ated, and if so, press
is assisting. pressure may be too low ZERO PRESSURE key for
to pause assist or less 2 seconds while IABP is
Provide alternate A.P. source than 15 minutes have assisting.
(i.e.: radial). elapsed since last
Provide alternate A.P.
calibration.
source (i.e.: radial).
Extender tubing or balloon Relieve restriction. Pump is either in Verify that IAB FILL mode
catheter may be restricted. STANDBY or the IAB FILL is set to AUTO.
Attempt calibration by mode is set to MANUAL.
pressing ZERO PRESSURE Resume pumping, then
key for 2 seconds while IABP press ZERO PRESSURE
is assisting. key for 2 seconds to initiate
a calibration.
IAB FILL mode is set to If appropriate, set IAB FILL
MANUAL. mode to AUTO via PUMP
OPTIONS menu.
ALARMS
PATIENT ASSESSMENT
Release button
PULL
1 2 3 4
Unlock console by press- Lift pump console Detach monitor from cart Attach monitor to top of pump
ing tab to right of console straight up and off cart. handle by pressing button console. Extend console handle
release handle and pull on rear of monitor. until it locks, tilt pump and pull to
handle straight out. transport.
PORTABLE OPERATION
Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician. Refer to
STATLOCK is a Registered Trademark of C. R. Bard, Inc · Copyright 2011 Datascope Corp. or its affiliates. All rights
Instructions for Use for current indications, warnings, contraindications, and precautions.
reserved. 12.11
Datascope Corp.
1300 MacArthur Blvd.
Mahwah, NJ 07430, USA GETINGE GROUP is a leading global provider of products and
Tel: +1 800 777 4222 systems that contribute to quality enhancement and cost efficiency
Fax: +1 800 258 8762 within healthcare and life sciences. We operate under the three brands
Tel: +1 973 244 6100 of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on
patient mobility and wound management solutions. GETINGE provi-
Fax: +1 973 244 6279
des solutions for infection control and prevention within healthcare
and life sciences. MAQUET specializes in solutions, therapies and
http://ca.maquet.com products for surgical interventions and intensive care.