(Rev. 0) MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Revision history
MOVIPLAN iC (Rev. 0)
SYSTEM INTEGRATION GUIDE
Contents
Contents
1 INTRODUCTION 1
2 SAFETY ASPECTS 2
3 SYSTEM CONFIGURATIONS 3
3.1 VDX 3543 PW – Single detector .............................................................. 3
3.2 Power supply connection instructions..................................................... 4
4 SYSTEM CONNECTIONS 7
4.1 General connection scheme .................................................................... 7
4.1.1 G100C RAD – Moviplan iC table .......................................................... 9
4.1.2 G100C RAD – Tele iC Wall Bucky ...................................................... 10
4.1.3 G100C RAD internal connections ...................................................... 11
4.1.4 G100C RAD – Column iC [only with TSC]........................................... 11
4.1.5 Column iC [only with TSC] - DROC acquisition system interface......... 11
4.1.6 G100C RAD – DROC acquisition system interface .............................. 12
4.1.7 G100C RAD – DAP ............................................................................ 12
4.1.8 G100C RAD – Three field solid state AEC device 1
(Claymount or Gilardoni type)............................................................ 13
4.1.9 G100C RAD – Three field solid state AEC device 2
(Claymount or Gilardoni type)............................................................ 14
4.1.10 G100C RAD jumpers and wiring configuration instruction ................. 15
4.1.10.1 H.V. Auxiliary board jumpers ............................................. 15
4.1.11 Accessories connection ...................................................................... 15
4.1.12 G100C RAD – X-ray tube 1 interface (connections to be
done with shielded cable)................................................................... 16
4.2 G100C RAD software SET-UP instruction ............................................. 17
4.2.1 Tube selection ................................................................................... 17
4.2.1.1 X-ray tube present in library .............................................. 17
4.2.1.2 Loading X-ray tube data from GenWare® utility software .... 18
4.2.1.3 X-ray tube not present neither in library nor in
GenWare® utility software.................................................. 18
4.2.2 AEC calibration ................................................................................. 18
4.2.3 G100C RAD image receptor programming worksheet.......................... 19
4.2.4 G100C RAD Receptor 1 I/O configuration worksheet ......................... 20
4.2.5 G100C RAD receptor 2 I/O configuration worksheet .......................... 20
4.2.6 G100C RAD receptor 3 I/O configuration worksheet .......................... 21
4.2.7 G100C RAD receptor 4 I/O configuration worksheet .......................... 21
4.2.8 G100C RAD receptor 5 I/O configuration worksheet .......................... 22
4.3 VDX connection to electric Power Supply.............................................. 23
4.3.1 UPS installation (optional) ................................................................. 24
(Rev. 0) i MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Contents
1 INTRODUCTION
This manual is intended to complete the information present in the
Generator and Moviplan iC documentation.
It explains system components interconnections and system workplaces
configurations.
For X-ray tube and AEC calibration, please make reference to the
generator Service Manual and equipment you have to connect with.
(Rev. 0) 1 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Safety aspects
2 SAFETY ASPECTS
WARNING:
All information contained in the present chapter must be carefully read
and, where necessary, must be applied in order to avoid damage to the
equipment or to people.
WARNING:
• Multiple portable socket-outlets shall not be placed on the floor.
• DO NOT connect any additional multiple portable socket-outlet or
extension cable to the system.
• DO NOT connect items which are not specified as part of the system.
MOVIPLAN iC 2 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System configuration
3 SYSTEM CONFIGURATIONS
Room configuration
• G100C RAD
• Moviplan iC TABLE
• Column iC
• Tele iC (optional)
• VDX 3543 PW
• Room light and door safety switch
Receptor configuration
Receptor 1:
Moviplan iC Table – Tube 1
Receptor 2:
Out of Potter-Bucky (direct X-ray) with Tube 1
Receptor 3:
Tele iC Wall Bucky with Tube 1
Receptor 4:
Tomography - Moviplan iC Table with Tube 1 (available
only on Column iC with touch-Screen Controller)
Receptor 5: Auxiliary X-ray without flat panel detector with Tube 1.
(Rev. 0) 3 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System configuration
Connect the power supply directly to the G100C RAD generator input
terminals under the fuses cover and use an auxiliary contactor in order
to power on all the equipment in the room via generator console.
For every one of these power supply is requested a switch in the room
cabinet to separate the line during technical assistance.
NOTE:
If the indicated auxiliary contactor is not available, use a remote
controlled switch with the following characteristics:
– Coil: 220 – 230 Vac 50 Hz
230 – 240 Vac 60 Hz
– Contact: 4 N.O. contacts, 16 A.
MOVIPLAN iC 4 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System configuration
R R
S S
T T
N N
J2.3 ( 24Vdc )
J2.4 ( 0Vdc )
Column iC
L X0 Moviplan iC
N Table
Tele iC
Dedicated Main power line 230 Vac 16°
for detector battery charger
L2 Detector battery
charger
N2
synchronizer
Dedicated Main Power Line
230Vac 16A for DROC cabinet
UPS
L3 L3 PC
N3 N3
monitor
(Rev. 0) 5 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System configuration
The main contactor plate, supplied with the installation kit, has to be
installed inside G100C RAD cabinet, rear side, as for Figure 1,
assembling on it contactor and relays.
Fixing screws
Figure 1
MOVIPLAN iC 6 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
4 SYSTEM CONNECTIONS
Here below you can find the system connection scheme, which indicates how to connect the system components each other.
The first row lists all the system components, and below, you can find connectors names (bold), signals names and directions (Italic Gray); for example X-Ray request on table goes
from Auxiliary Board on the G100C RAD to the Interface PCB A26 inside the Moviplan iC Table.
In the following paragraphs, each connection shown in the "General connection scheme" will be detailed wire by wire. It is also shown to which connector each wire must be
connected (a checkbox will help you to keep trace of the connections done).
(Rev. 0) 7 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC
System 1
interconnection
Page 1 of 1
MOVIPLAN iC 8 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
NOTE:
The voltage selected on J2.7 – J2.10 is 24Vdc.
(Rev. 0) 9 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
OR
OR
MOVIPLAN iC 10 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
Tomo
J2.2 Green
Exposure/interlock2
Green ✔ X29.5
(Rev. 0) 11 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC 12 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
NOTE:
If the AEC interface PWB it is supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
(Rev. 0) 13 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
NOTE:
If the AEC interface PWB it is supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
MOVIPLAN iC 14 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
The BUCKY DRIVE output MUST be configured to +24 VDC connecting E11 to
E5 and E10 to E8.
NOTE:
Connect the external Emergency button to the connector J2.1 and J2.2
in order to switch off all the equipments of the radiological room.
NOTE:
The generator supplies a dry contact closure when the room light is to be
activated. Room lights must be connected to an external 220Vac.
(Rev. 0) 15 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
NOTE:
Connect the HV cables following the instructions on the G100C RAD
Service Manual paragraph 2.6.6.
MOVIPLAN iC 16 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
If the X-ray tube used in the installation is present in the library of the
generator refer to the G100C RAD Service Manual, chapter 3C
"Interfacing and Programming" in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G100C RAD Service Manual, chapter 2.7.0 "Low speed starter tube
compatibility" to record the necessary information to complete the
configuration.
NOTE:
The tube library, concerning the I.A.E. X-ray tubes, contains only high
speed tube types. When a low speed I.A.E. X-ray tube has to be installed,
an high speed tube must be selected in the library, but it’s necessary to
force the generator to use only the low speed tube (see section 3C in the
Service Manual G100C RAD). The generator and the tube limits must be
set according with the tube type installed.
(Rev. 0) 17 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
4.2.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the X-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the
X-ray tube that is going to be installed. An X-ray tube present in the
library or in the GenWare® utility software, as similar as possible to the
one is going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the X-ray tube. Refer to
the G100C RAD Service Manual, chapter 3C.7.1 "Tube selection” and
3C.7.2 “Generator limits" in order to record all the necessary information
to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G100C RAD Service Manual, chapter 2.7.0 "Low speed starter tube
compatibility" to record the necessary information to complete the
configuration.
The installation of the AEC chambers and their calibration are covered in
the VDX system Service Manual, chapter 12 "AEC calibration (Automatic
exposure control)".
The installation of the AEC chambers and their calibration are covered in
the G100C RAD Service Manual, chapter 3 "Automatic exposure control
(AEC) calibration".
MOVIPLAN iC 18 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
RECEPTOR NAME
TUBE NUMBER 1 1 1 1 1
TOMO ENABLE NO NO NO YES NO
FLUORO ENABLE N/A N/A N/A N/A N/A
INTERFACE OPTS 0 0 0 0 0
FLUORO
N/A N/A N/A N/A N/A
HANGOVER
RAD HANGOVER 0 0 0 0 0
LAST IMAGE HOLD N/A N/A N/A N/A N/A
MEMORY OFF OFF OFF OFF OFF
REM TOMO BUT 2500 ms 2500 ms 2500 ms 4000 ms 4000 ms
AUTO FOCUS ON ON ON ON ON
AEC BACKUP MS FIXED MS FIXED FIXED
AEC BACKUP mAs 500 500 500 500 500
AEC BACKUP ms 500 3200 500 4000 3200
AEC LOCK EXP: N/A N/A N/A N/A N/A
DEFAULT VOLTAGE 75 KV 75 KV 75 KV 75 KV 75 KV
DEFAULT CURRENT 20 mA 20 mA 20 mA 20 mA 20 mA
DEFAULT TIME (ms) 200 ms 200 ms 200 ms 200 ms 200 ms
TECHNIQUE mAs mAs mAs ms mAs
DEFAULT FOCUS large large large large large
DEFAULT DENSITY 0 0 0 0 0
DEFAULT FILM
I I I I I
SCREEN
DEFAULT FIELDS RCL RCL RCL RCL RCL
FUNCTIONAL OPTS 0 0 0 0 0
NOTE (*):
It is necessary to calibrate the AEC chambers following the instructions
in the G100C RAD Service Manual, section 3D.
(Rev. 0) 19 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR ITLK OFF ON ON ON
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON
***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON
MOVIPLAN iC 20 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF ON
ROOM LIGHT OFF ON ON ON
***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF ON
BUCKY 1 READY OFF OFF OFF OFF
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON
(Rev. 0) 21 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF OFF
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON
MOVIPLAN iC 22 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
VDX 4343TS / VDX 3543TC / VDX 3543 PW systems are set up for a
nominal power supply voltage of 230 V - 50/60 Hz.
The VDX 4343TS / VDX 3543TC / VDX 3543PW systems cables are
usually 15 m long, in order to connect the operative console in protected
area to the image receptor.
(Rev. 0) 23 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
The UPS is already installed inside the DROC cabinet. Connect the cables
to the relevant sockets as listed below.
Legend:
5 Power input 230V
6 Output for RAD box v. 2.2 (Battery backup outlets with surge
protection)
7 Output for monitor
8 Output for PC
NOTE:
The meanings of the not listed sockets is described in the UPS relevant
Manual.
MOVIPLAN iC 24 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
4.4 PC connections
LAN (RIS/HIS)
USB
COM 1
COM 4
COM 3
FPD
(Rev. 0) 25 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC 26 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
(Rev. 0) 27 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC 28 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
(Rev. 0) 29 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC 30 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections
Open a browser and enter the IP of the detector into the URL field, for
example 192.168.95.4
The XRpad Detector’s Administration Interface should open.
Click Network -> Network Settings and a prompt will appear requiring
login credentials, enter:
• User: admin
• Password: PerkinElmer.
NOTE:
Password is case sensitive.
(Rev. 0) 31 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections
MOVIPLAN iC 32 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
You can enter DROC configuration tool through the following way:
• Run the "DROCConfig.exe" file which is located in the same directory
of DROC
• Select the "DROC Configuration" item in "System" menu on DROC
interface.
Select as "User Name: mgr" from the User Name drop down list and input
the as "Password: mgr".
Click "OK" button to log in the configuration tool.
(Rev. 0) 33 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
5.1.1 Protocol
NOTE:
In case of single panel detector, please make sure the detector selected
on "Workstation: Wall" is "detector1" and NOT "detector2".
Then select "Technical Parameter File" and "Image Process Parameter
File" as on "Workstation: Table".
MOVIPLAN iC 34 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
5.1.2 Hardware
(Rev. 0) 35 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
MOVIPLAN iC 36 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
(Rev. 0) 37 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
MOVIPLAN iC 38 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
(Rev. 0) 39 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
MOVIPLAN iC 40 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
(Rev. 0) 41 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
MOVIPLAN iC 42 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
(Rev. 0) 43 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
5.1.3 Workflow
MOVIPLAN iC 44 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration
NOTE:
It is possible to perform detector calibration also without aluminium filter
but in this case kV values will be low and ms values will be short if
compared to normal use.
(Rev. 0) 45 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration
The calibration procedure begins. It will lead you to acquire flat field
images at one dose level. The operator must select correct exposure
parameters in order to reach the target grey level (e.g. 24000 in the
picture).
At the end of image acquisitions click on "Save" and restart DROC
application.
MOVIPLAN iC 46 (Rev. 0)