MOVIPLAN iC

System Integration Guide

Release 27 July 2015 (Rev. 0)


SYSTEM INTEGRATION GUIDE
Revision history

Revision history Manual code 6998313203

Rev. Date Page/s Modification description

0 27.07.15 - Document approval.

(Rev. 0) MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Revision history

THIS PAGE IS INTENTIONALLY LEFT BLANK

MOVIPLAN iC (Rev. 0)
SYSTEM INTEGRATION GUIDE
Contents

Contents
1 INTRODUCTION 1

2 SAFETY ASPECTS 2

3 SYSTEM CONFIGURATIONS 3
3.1 VDX 3543 PW – Single detector .............................................................. 3
3.2 Power supply connection instructions..................................................... 4

4 SYSTEM CONNECTIONS 7
4.1 General connection scheme .................................................................... 7
4.1.1 G100C RAD – Moviplan iC table .......................................................... 9
4.1.2 G100C RAD – Tele iC Wall Bucky ...................................................... 10
4.1.3 G100C RAD internal connections ...................................................... 11
4.1.4 G100C RAD – Column iC [only with TSC]........................................... 11
4.1.5 Column iC [only with TSC] - DROC acquisition system interface......... 11
4.1.6 G100C RAD – DROC acquisition system interface .............................. 12
4.1.7 G100C RAD – DAP ............................................................................ 12
4.1.8 G100C RAD – Three field solid state AEC device 1
(Claymount or Gilardoni type)............................................................ 13
4.1.9 G100C RAD – Three field solid state AEC device 2
(Claymount or Gilardoni type)............................................................ 14
4.1.10 G100C RAD jumpers and wiring configuration instruction ................. 15
4.1.10.1 H.V. Auxiliary board jumpers ............................................. 15
4.1.11 Accessories connection ...................................................................... 15
4.1.12 G100C RAD – X-ray tube 1 interface (connections to be
done with shielded cable)................................................................... 16
4.2 G100C RAD software SET-UP instruction ............................................. 17
4.2.1 Tube selection ................................................................................... 17
4.2.1.1 X-ray tube present in library .............................................. 17
4.2.1.2 Loading X-ray tube data from GenWare® utility software .... 18
4.2.1.3 X-ray tube not present neither in library nor in
GenWare® utility software.................................................. 18
4.2.2 AEC calibration ................................................................................. 18
4.2.3 G100C RAD image receptor programming worksheet.......................... 19
4.2.4 G100C RAD Receptor 1 I/O configuration worksheet ......................... 20
4.2.5 G100C RAD receptor 2 I/O configuration worksheet .......................... 20
4.2.6 G100C RAD receptor 3 I/O configuration worksheet .......................... 21
4.2.7 G100C RAD receptor 4 I/O configuration worksheet .......................... 21
4.2.8 G100C RAD receptor 5 I/O configuration worksheet .......................... 22
4.3 VDX connection to electric Power Supply.............................................. 23
4.3.1 UPS installation (optional) ................................................................. 24

(Rev. 0) i MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Contents

4.4 PC connections ..................................................................................... 25


4.5.1 RAD BOX jumpers configuration ....................................................... 28
4.6.1 PC Network settings .......................................................................... 29
4.6.2 WAP settings..................................................................................... 30
4.6.3 XRpad 4336 flat panel detector settings............................................. 31

5 DROC SOFTWARE CONFIGURATION 33


5.1 DROC configuration.............................................................................. 33
5.1.1 Protocol ............................................................................................ 34
5.1.1.1 Parameter configuration..................................................... 34
5.1.2 Hardware.......................................................................................... 35
5.1.2.1 Generator configuration ..................................................... 35
5.1.2.2 Detector I configuration ..................................................... 36
5.1.2.3 Detector I: Advance Configuration ...................................... 37
5.1.2.4 Detector I: Calibration data settings ................................... 38
5.1.2.5 Synchronization - SynBox I configuration........................... 40
5.1.2.6 Collimator configuration .................................................... 41
5.1.2.7 Mechanical configuration ................................................... 43
5.1.3 Workflow .......................................................................................... 44
5.1.3.1 Other Settings ................................................................... 44
5.2 Detector Calibration.............................................................................. 45

This publication can only be reproduced, transmitted, transcribed or translated into


any human or computer language with the written consent of VILLA SISTEMI
MEDICALI S.p.A.
This manual in English is the original version.
MOVIPLAN iC ii (Rev. 0)
SYSTEM INTEGRATION GUIDE
Introduction

1 INTRODUCTION
This manual is intended to complete the information present in the
Generator and Moviplan iC documentation.
It explains system components interconnections and system workplaces
configurations.
For X-ray tube and AEC calibration, please make reference to the
generator Service Manual and equipment you have to connect with.

This manual includes the interconnections list, the hardware and


software configuration for a typical radiological rooms using a generator
G100C RAD with Digital Panel.

(Rev. 0) 1 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
Safety aspects

2 SAFETY ASPECTS

WARNING:
All information contained in the present chapter must be carefully read
and, where necessary, must be applied in order to avoid damage to the
equipment or to people.

WARNING:
• Multiple portable socket-outlets shall not be placed on the floor.
• DO NOT connect any additional multiple portable socket-outlet or
extension cable to the system.
• DO NOT connect items which are not specified as part of the system.

MOVIPLAN iC 2 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System configuration

3 SYSTEM CONFIGURATIONS

3.1 VDX 3543 PW – Single detector

 Room configuration
• G100C RAD
• Moviplan iC TABLE
• Column iC
• Tele iC (optional)
• VDX 3543 PW
• Room light and door safety switch

 Receptor configuration
Receptor 1:
Moviplan iC Table – Tube 1
Receptor 2:
Out of Potter-Bucky (direct X-ray) with Tube 1
Receptor 3:
Tele iC Wall Bucky with Tube 1
Receptor 4:
Tomography - Moviplan iC Table with Tube 1 (available
only on Column iC with touch-Screen Controller)
Receptor 5: Auxiliary X-ray without flat panel detector with Tube 1.

(Rev. 0) 3 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System configuration

3.2 Power supply connection instructions

Connect the power supply directly to the G100C RAD generator input
terminals under the fuses cover and use an auxiliary contactor in order
to power on all the equipment in the room via generator console.

As showed in the schematic, the room cabinet requirements are:


• N°1 Power supply 400Vac to connect generator and Moviplan IC
system
• N° 1 Power supply 230Vac 16A to connect the detector battery
charger
• N° 1 Power supply 230Vac 16A to connect the Acquisition system
(DROC).

For every one of these power supply is requested a switch in the room
cabinet to separate the line during technical assistance.

NOTE:
 If the indicated auxiliary contactor is not available, use a remote
controlled switch with the following characteristics:
– Coil: 220 – 230 Vac 50 Hz
230 – 240 Vac 60 Hz
– Contact: 4 N.O. contacts, 16 A.

MOVIPLAN iC 4 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System configuration

Power Line G100C RAD


400 Vac Main contactor TERMINALS

R R
S S
T T
N N

J2.3 ( 24Vdc )

220Vac AUXILIARY CONTACTOR


* CODE 4291515200
G100C RAD
Generator
Control Board

J2.4 ( 0Vdc )

Column iC

L X0 Moviplan iC
N Table

Tele iC
Dedicated Main power line 230 Vac 16°
for detector battery charger

L2 Detector battery
charger
N2

synchronizer
Dedicated Main Power Line
230Vac 16A for DROC cabinet
UPS
L3 L3 PC
N3 N3
monitor

(Rev. 0) 5 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System configuration

The main contactor plate, supplied with the installation kit, has to be
installed inside G100C RAD cabinet, rear side, as for Figure 1,
assembling on it contactor and relays.

Fixing screws

TB1 screw connector

Figure 1

Connect the components as for the following diagram:

Figure 2: Connection diagram

MOVIPLAN iC 6 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4 SYSTEM CONNECTIONS

4.1 General connection scheme

Here below you can find the system connection scheme, which indicates how to connect the system components each other.
The first row lists all the system components, and below, you can find connectors names (bold), signals names and directions (Italic Gray); for example X-Ray request on table goes
from Auxiliary Board on the G100C RAD to the Interface PCB A26 inside the Moviplan iC Table.
In the following paragraphs, each connection shown in the "General connection scheme" will be detailed wire by wire. It is also shown to which connector each wire must be
connected (a checkbox will help you to keep trace of the connections done).

G100C RAD DROC Column iC DAP X-ray Tube


TB1 Screw connector Tele iC Moviplan iC Table AEC
(inside generator)
| | | | | | | | |
[ Auxiliary board ] X-Ray request on table ► [ Interface PCB A26 ] | | |
| | | | | | | | |
[ Generator Control Board ] ◄ Emergency signals ► [ Main terminal strip X0 ] | | |
| | | | | | | | |
| |:| ◄ Bucky ready signal [ Interface PCB A26 ] | | |
| | | | | | | | |
[ Auxiliary board ] X-Ray request on Wall Bucky ► [ Wall Bucky PCB A23 or A04 ] | | | | |
| | | | | | | | |
| |:| ◄ Bucky ready signal [ Wall Bucky PCB A23 or A04]| | | | |
| | | | | | | | |
[ Auxiliary board ] ◄ X-Ray release |:| | | | | | | |
| | | | | | | | |
[ Auxiliary board ] ◄ Tomo Signals [ Column PCB A06 TSC ] | | | |
| | | | | | | | |
| | [ RAD BOX ] CAN Bus [ Column PCB A06 TSC ] | | | |
| | | | | | | | |
| |:| ◄ Bucky ready / ► [ RAD BOX ] | | | | | |
| | X-ray release | | | | | | |
[ Generator Control Board ] ◄ Serial communications ►[ PC ] | | | | | |
| | | | | | | | |
[ Generator Control Board ] ◄ DAP Communications ► [ DAP ] | |
| | | | | | | | |
[ AEC board ] ◄ AEC Communications ► [ AEC device 1 ] |
| | | | | | | |
[ AEC board ] ◄ AEC Communications ►[ AEC device 2 ] |
| | | | | | | |
[ Auxiliary board ] ◄ TUBE Signals ► [ X-RAY TUBE ]

(Rev. 0) 7 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

MOVIPLAN iC
System 1
interconnection
Page 1 of 1

MOVIPLAN iC 8 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.1.1 G100C RAD – Moviplan iC table

Generator Control board CABLE: GENERIC Main terminal board X0


Connector Connected Wire n° Signal name Wire n° Connected Connector
J2.1 __ Emergency signal __ X0.1
J2.2 __ Emergency signal __ X0.2

Auxiliary board CABLE: GENERIC Interface board A26


Connector Connected Wire n° Signal name Wire n° Connected Connector
J2.10 __ Bucky1 select (+0V) __ X47.2
J2.7 __ Bucky1 select (+24V) __ X47.1

NOTE:
 The voltage selected on J2.7 – J2.10 is 24Vdc.

TB1 screw connector CABLE: GENERIC Interface PCB A26


(Figure 1)
Connector Connected Wire n° Signal name Wire n° Connected Connector
TB1.1 __ Bucky 1 (table) ready __ X47.3
TB1.2 __ Bucky 1 (table) ready __ X47.4

(Rev. 0) 9 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.1.2 G100C RAD – Tele iC Wall Bucky

Auxiliary board CABLE: GENERATOR Wall Bucky board A23


Connector Connected Wire Signal name Wire Connected Connector
J4.1 Yellow Bucky 2 select (+24V) Yellow ✔ X10.1
J4.4 Green Bucky 2 select (+0V) Green ✔ X10.2

OR

Auxiliary board CABLE: GENERATOR Wall Bucky board A04


Connector Connected Wire Signal name Wire Connected Connector
J4.1 Yellow Bucky 2 select (+24V) Yellow ✔ X11.1
J4.4 Green Bucky 2 select (+0V) Green ✔ X11.2

TB1 screw connector CABLE: GENERIC Wall Bucky PCB A23


(Figure 1)
Connector Connected Wire n° Signal name Wire n° Connected Connector
TB1.1 Brown Wall Bucky ready Brown ✔ X10.3
TB1.2 Blue Wall Bucky ready Blue ✔ X10.4

OR

TB1 screw connector CABLE: GENERIC Wall Bucky PCB A04


(Figure 1)
Connector Connected Wire n° Signal name Wire n° Connected Connector
TB1.1 Brown Wall Bucky ready Brown ✔ X11.3
TB1.2 Blue Wall Bucky ready Blue ✔ X11.4

MOVIPLAN iC 10 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.1.3 G100C RAD internal connections

TB1 screw connector CABLE: GENERIC Auxiliary board


(Figure 1)
Signal name
Connector Connected Wire n° Wire n° Connected Connector
DROC side CPI side
Bucky 1
TB1.3 __ RLS1_1 __ J2.5
ready
Bucky 1
TB1.4 __ RLS1_2 __ J2.6
ready

4.1.4 G100C RAD – Column iC [only with TSC]

Auxiliary board CABLE: GENERATOR Column board A6


Connector Connected Wire Signal name Wire Connected Connector
Tomo
J2.1 Yellow
Exposure/interlock2
Yellow ✔ X29.8

Tomo
J2.2 Green
Exposure/interlock2
Green ✔ X29.5

4.1.5 Column iC [only with TSC] - DROC acquisition system


interface

Column board A6 CAN BUS CABLE RAD BOX


Connector Connected Signal name Connected Connector
X34 CAN Bus CAN_CLM

(Rev. 0) 11 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.1.6 G100C RAD – DROC acquisition system interface

TB1 screw connector CABLE: GENERIC RAD BOX


(Figure 1)
Connector Connected Wire Signal name Pin Connected Connector
TB1.1 XPB1+ XPB1+ 1
TB1.2 XPB1- XPB1- 2
I/O GEN
TB1.3 RLS1_1 RLS1_1 7
TB1.4 RLS1_2 RLS1_2 8

Generator Control Board RCABLE: RS232 PC


Connector Connected Signal name Connected Connector
J22 Serial Line COM3

4.1.7 G100C RAD – DAP

Generator Control Board CABLE: DAP CABLE DAP


Connector Connected Signal name Connected Connector
J4 DAP RJ45 plug

MOVIPLAN iC 12 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.1.8 G100C RAD – Three field solid state AEC device 1


(Claymount or Gilardoni type)

NOTE:
 If the AEC interface PWB it is supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

AEC Board (type 737992) AEC CABLE AEC Device 1


Connector Connected Wire Signal name Wire Connected Connector
J11.1 Pink Anode Left Pink ✔
J11.2 Grey Cathode Left Grey ✔
J11.3 White Anode Middle White ✔
Device
J11.4 Brown Cathode Middle Brown ✔ plug
J11.5 Green Anode Right Green ✔
J11.6 Yellow Cathode Right Yellow ✔
J11.7 Shield Ground Shield ✔

(Rev. 0) 13 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.1.9 G100C RAD – Three field solid state AEC device 2


(Claymount or Gilardoni type)

NOTE:
 If the AEC interface PWB it is supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

AEC Board (type 737992) AEC CABLE AEC Device 2


Connector Connected Wire Signal name Wire Connected Connector
J12.1 Pink Anode Left Pink ✔
J12.2 Grey Cathode Left Grey ✔
J12.3 White Anode Middle White ✔
Device
J12.4 Brown Cathode Middle Brown ✔ plug
J12.5 Green Anode Right Green ✔
J12.6 Yellow Cathode Right Yellow ✔
J12.7 Shield Ground Shield ✔

MOVIPLAN iC 14 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.1.10 G100C RAD jumpers and wiring configuration instruction

4.1.10.1 H.V. Auxiliary board jumpers

The BUCKY DRIVE output MUST be configured to +24 VDC connecting E11 to
E5 and E10 to E8.

4.1.11 Accessories connection

NOTE:
 Connect the external Emergency button to the connector J2.1 and J2.2
in order to switch off all the equipments of the radiological room.

Auxiliary board J4 DEDICATED CABLES Room accessories


Connector Connected Wire n° Signal name Wire n° Connected Accessory
J4.9 __ Room Door Interlock __ Door contact
J4.10 __ Room Door Interlock __ Door contact
J11.3 __ Room light contact __ Room Light
J11.4 __ Room light contact __ Room Light

NOTE:
 The generator supplies a dry contact closure when the room light is to be
activated. Room lights must be connected to an external 220Vac.

(Rev. 0) 15 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.1.12 G100C RAD – X-ray tube 1 interface (connections to be done


with shielded cable)

Auxiliary board J7 TUBE CABLE X – ray tube


Connector Connected Signal name Connected Connector
J7.4 Common Common
J7.5 Principal/inductive Principal
J7.6 Auxiliary/capacitive Auxiliary
J7.1 Thermal switch
Thermal switch
J7.2 Thermal switch

NOTE:
 Connect the HV cables following the instructions on the G100C RAD
Service Manual paragraph 2.6.6.

MOVIPLAN iC 16 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.2 G100C RAD software SET-UP instruction

After the hardware connections indicated in the previous paragraphs, it's


necessary to proceed with a software configuration of the generator that
includes the information about the installed X-ray tube (that has to be
calibrated) and the data about the room configuration that must
correspond to the hardware connections (receptor set-up).

4.2.1 Tube selection

Prior to beginning the X-ray tube calibration, the tube used in


installation must be properly selected and the generator limits has to be
programmed.

4.2.1.1 X-ray tube present in library

If the X-ray tube used in the installation is present in the library of the
generator refer to the G100C RAD Service Manual, chapter 3C
"Interfacing and Programming" in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G100C RAD Service Manual, chapter 2.7.0 "Low speed starter tube
compatibility" to record the necessary information to complete the
configuration.

NOTE:
 The tube library, concerning the I.A.E. X-ray tubes, contains only high
speed tube types. When a low speed I.A.E. X-ray tube has to be installed,
an high speed tube must be selected in the library, but it’s necessary to
force the generator to use only the low speed tube (see section 3C in the
Service Manual G100C RAD). The generator and the tube limits must be
set according with the tube type installed.

(Rev. 0) 17 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.2.1.2 Loading X-ray tube data from GenWare® utility software

If the tube is not in the generator library, it can be found in a wider


selection available in the GenWare® Software. In order to use GenWare®
and upload the tube library, a serial communication connection must be
established between the generator and a laptop PC. Refer to G100C RAD
service manual, chapter 3C.12.0 "Data link" for details.

4.2.1.3 X-ray tube not present neither in library nor in GenWare® utility
software

If the X-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the
X-ray tube that is going to be installed. An X-ray tube present in the
library or in the GenWare® utility software, as similar as possible to the
one is going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the X-ray tube. Refer to
the G100C RAD Service Manual, chapter 3C.7.1 "Tube selection” and
3C.7.2 “Generator limits" in order to record all the necessary information
to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G100C RAD Service Manual, chapter 2.7.0 "Low speed starter tube
compatibility" to record the necessary information to complete the
configuration.

4.2.2 AEC calibration

The installation of the AEC chambers and their calibration are covered in
the VDX system Service Manual, chapter 12 "AEC calibration (Automatic
exposure control)".

The installation of the AEC chambers and their calibration are covered in
the G100C RAD Service Manual, chapter 3 "Automatic exposure control
(AEC) calibration".

MOVIPLAN iC 18 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.2.3 G100C RAD image receptor programming worksheet

IMAGE RECEPTOR PROGRAMMING WORKSHEET


DATE:
SERIAL#:
FUNCTION RECEPTOR 1 RECEPTOR 2 RECEPTOR 3 RECEPTOR 4 RECEPTOR 5

RECEPTOR NAME

TUBE NUMBER 1 1 1 1 1
TOMO ENABLE NO NO NO YES NO
FLUORO ENABLE N/A N/A N/A N/A N/A
INTERFACE OPTS 0 0 0 0 0
FLUORO
N/A N/A N/A N/A N/A
HANGOVER
RAD HANGOVER 0 0 0 0 0
LAST IMAGE HOLD N/A N/A N/A N/A N/A
MEMORY OFF OFF OFF OFF OFF
REM TOMO BUT 2500 ms 2500 ms 2500 ms 4000 ms 4000 ms
AUTO FOCUS ON ON ON ON ON
AEC BACKUP MS FIXED MS FIXED FIXED
AEC BACKUP mAs 500 500 500 500 500
AEC BACKUP ms 500 3200 500 4000 3200
AEC LOCK EXP: N/A N/A N/A N/A N/A
DEFAULT VOLTAGE 75 KV 75 KV 75 KV 75 KV 75 KV
DEFAULT CURRENT 20 mA 20 mA 20 mA 20 mA 20 mA
DEFAULT TIME (ms) 200 ms 200 ms 200 ms 200 ms 200 ms
TECHNIQUE mAs mAs mAs ms mAs
DEFAULT FOCUS large large large large large
DEFAULT DENSITY 0 0 0 0 0
DEFAULT FILM
I I I I I
SCREEN
DEFAULT FIELDS RCL RCL RCL RCL RCL
FUNCTIONAL OPTS 0 0 0 0 0

NOTE (*):
 It is necessary to calibrate the AEC chambers following the instructions
in the G100C RAD Service Manual, section 3D.

(Rev. 0) 19 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.2.4 G100C RAD Receptor 1 I/O configuration worksheet

FUNCTIONS STANDBY PREP GEN RDY RAD EXP

***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR ITLK OFF ON ON ON
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON

4.2.5 G100C RAD receptor 2 I/O configuration worksheet

FUNCTIONS STANDBY PREP GEN RDY RAD EXP

***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON

MOVIPLAN iC 20 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.2.6 G100C RAD receptor 3 I/O configuration worksheet

FUNCTIONS STANDBY PREP GEN RDY RAD EXP

***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF ON
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF ON
ROOM LIGHT OFF ON ON ON

4.2.7 G100C RAD receptor 4 I/O configuration worksheet

FUNCTIONS STANDBY PREP GEN RDY RAD EXP

***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF ON
BUCKY 1 READY OFF OFF OFF OFF
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF ON
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON

(Rev. 0) 21 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.2.8 G100C RAD receptor 5 I/O configuration worksheet

FUNCTIONS STANDBY PREP GEN RDY RAD EXP

***INPUTS***
THERMAL SW ON ON ON ON
ROOM DOOR
OFF ON ON ON
ITLK
INTERLOCK 1 OFF OFF OFF OFF
INTERLOCK 2 OFF OFF OFF OFF
BUCKY 1 READY OFF OFF OFF OFF
BUCKY 2 READY OFF OFF OFF OFF
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON

MOVIPLAN iC 22 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.3 VDX connection to electric Power Supply

VDX 4343TS / VDX 3543TC / VDX 3543 PW systems are set up for a
nominal power supply voltage of 230 V - 50/60 Hz.

The VDX 4343TS/VDX 3543TC / VDX 3543 PW system power supply is


not taken from the radiological table, nor from the X-ray generator, but it
is taken directly from an uninterruptible power supply source.

The VDX 4343TS/VDX 3543TC / VDX 3543 PW can be equipped with an


UPS with Output Power Capacity 540 Watts / 900 VA providing
uninterruptible power supply to the following components:
• Personal computer
• Monitor
• Synchronization and Control Box

VDX 4343TS/VDX 3543TC require a second permanent 230V mains


dedicated to the FPD power directly supplied by the room electrical
cabinet.

The suggested configuration assures a constant operating quality; the


detector reaches the optimum operating conditions at thermal
equilibrium, it is therefore advisable to keep it powered on during times
out of work (even night, if possible).
In case of power supply interruption, the UPS can avoid loss of data
during the image transfer and matching with exam data.

The VDX 4343TS / VDX 3543TC / VDX 3543PW systems cables are
usually 15 m long, in order to connect the operative console in protected
area to the image receptor.

(Rev. 0) 23 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.3.1 UPS installation (optional)

The UPS is already installed inside the DROC cabinet. Connect the cables
to the relevant sockets as listed below.

Figure 3: UPS rear panel

Legend:
5 Power input 230V
6 Output for RAD box v. 2.2 (Battery backup outlets with surge
protection)
7 Output for monitor
8 Output for PC

NOTE:
 The meanings of the not listed sockets is described in the UPS relevant
Manual.

MOVIPLAN iC 24 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.4 PC connections

In addition to the typical connections for keyboard, mouse and monitor,


the PC is connected to the FPD by GigaEthernet network adapter; it is
also connected to the RAD box v.2.2 by COM port serial communication
port, it is connected to the X-Ray generator by serial ports, it is
connected to the RIS/HIS network by Ethernet port.

LAN (RIS/HIS)
USB

COM 1
COM 4

COM 3
FPD

Figure 4: Example of PC rear panel

(Rev. 0) 25 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.5 SYNCHRONIZER RAD box version 2.2


The main function of ESYN-2000T Synchronization box is to connect the
high-voltage generator, FPD, collimator and PC via USB\RS232\
CAN bus.
The box analyzes the commands from the DROC software and
synchronizes the communication with the X-ray generator and FPD, so
that the digital X-ray imaging system can work at required timing
sequence.
On the front there is a series of green LEDs which light-up when the
relative signal is ON.

Figure 5: Front side


Legend:
1. POWR: Power state
2. Rest: Box Rest status
3. USB: USB connection status
4. XPB1: Generator Ready signal 1
5. XPB2: Generator Ready signal 2
6. PREP: Generator prepare signal
7. RLS1: Generator synchronizing signal 1
8. RLS2: Generator synchronizing signal 2
9. P_PREP: Generator prepare signal
10. PW_GD180: Grid 1 status (Wall Mode)
11. PW_GD120: Grid 2 status (Wall Mode)
12. PW_EXP: Detector Synchronizing Signal (Wall Mode)
13. PW_OK: Detector Ready Signal (Wall Mode)
14. FW_TXD: NC
15. FW_RXD: NC
16. PT_GD180: Grid 1 status (Table Mode)
17. PT_GD120: Grid 2 status (Table Mode)
18. PT_EXP: Detector Synchronizing Signal(Table Mode)
19. PT_OK: Detector Ready Signal (Table Mode)
20. FT_TXD: NC
21. FT_RXD: NC

MOVIPLAN iC 26 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

Figure 6: Connector side


Legend:
• USB: Connect to USB of PC
• Power: synchronization box power supply interface
• COM_PC: Connect to the computer serial port
• CAN_CLM: Connect to CAN interface of collimator (only for systems
with automatic collimator)
• COM_PC_1: Connect to the computer serial port NOT USED
• COM_GEN_1: Serial interface Connecting Generator NOT USED
• COM_PC_2: Connect to the computer serial port NOT USED
• COM_GEN_2: Serial interface Connecting Generator NOT USED
• IO_GEN: Generator cable interface
• CAN_GEN: CAN BUS interface for Generator NOT USED
• DET_TABLE: Detector 1 cable interface
• DET_WALL: Detector 2 cable interface (only for dual detector
configuration)

The Synchronizer board is equipped with jumpers to set communication


mode and signal type.
The jumpers configuration are reported on paragraph 4.5.1.

The synchronizer RAD box v.2.2 is connected to the PC by serial RS-232


short cable.
Connect COM-PC to COM 1 port on the computer rear side (Figure 4).

(Rev. 0) 27 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.5.1 RAD BOX jumpers configuration

MOVIPLAN iC 28 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.6 VDX Network settings

4.6.1 PC Network settings

(Rev. 0) 29 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

4.6.2 WAP settings

Open a browser and enter the IP of the WAP: 192.168.95.3.


The browser will request login credentials; enter:
• Username: Cisco
• Password: Cisco

The main router web admin page will now load.

• Under Wireless -> Wireless Settings -> Configuration View, choose


Manual radio button
• Configure the 5GHz Wireless Settings
• Choose Network Mode: Wireless-N Only
• Network Name (SSID): VDX3543PW
• Security Mode: WPA2 Personal
• Passphrase: XRpad4336
• Channel Width: 20 MHz Only
• Channel: 40 – 5.200 GHz
• SSID Broadcast: Enabled
• 2.4 GHz can be disabled
• DHCP server can be disabled, if enabled set starting IP address to
192.168.95.100 end IP address to 192.168.95.200
• Click "Save Settings" at the bottom
• Click "Continue".

MOVIPLAN iC 30 (Rev. 0)
SYSTEM INTEGRATION GUIDE
System connections

4.6.3 XRpad 4336 flat panel detector settings

Open a browser and enter the IP of the detector into the URL field, for
example 192.168.95.4
The XRpad Detector’s Administration Interface should open.
Click Network -> Network Settings and a prompt will appear requiring
login credentials, enter:
• User: admin
• Password: PerkinElmer.

NOTE:
 Password is case sensitive.

The pull-down menu at the top will display "*Configuration 1" or


"*Configuration 2". Click the down-arrow and choose "*Configuration 1".
All the fields below will now be editable.

(Rev. 0) 31 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
System connections

Check the following settings:


• "Type": Station
• "IP address" for the XRpad will have the same network ID as the
router and PC, but a unique host ID.
For example, choose an IP of 192.168.95.4.
• Leave the "Netmask" as the default 255.255.255.0
• Leave "Gateway" and "Nameserver" empty.
• Enter a description of the external wireless router which will be used.
For example VDX3543PW
• "SSID" of the router: VDX3543PW
• "WPA2 passphrase" of the router: XRpad4336
• "Image Transfer" to use WLAN. This will enable the detector to
transfer images in wireless mode
• Click "Apply" at the bottom
• Click "Save as default" and "Restart Network" at the top of the
interface to save the changes as the default settings.

MOVIPLAN iC 32 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5 DROC SOFTWARE CONFIGURATION

5.1 DROC configuration

You can enter DROC configuration tool through the following way:
• Run the "DROCConfig.exe" file which is located in the same directory
of DROC
• Select the "DROC Configuration" item in "System" menu on DROC
interface.

The DROC Configuration Login window will be displayed:

Select as "User Name: mgr" from the User Name drop down list and input
the as "Password: mgr".
Click "OK" button to log in the configuration tool.

In the following pages it is described the configuration for VDX 3543 PW


single panel detector.

(Rev. 0) 33 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.1 Protocol

5.1.1.1 Parameter configuration

NOTE:
 In case of single panel detector, please make sure the detector selected
on "Workstation: Wall" is "detector1" and NOT "detector2".
Then select "Technical Parameter File" and "Image Process Parameter
File" as on "Workstation: Table".

MOVIPLAN iC 34 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2 Hardware

5.1.2.1 Generator configuration

Please set workstation as follows:

(Rev. 0) 35 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.2 Detector I configuration

Please set Single Detector configuration as follows:

MOVIPLAN iC 36 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.3 Detector I: Advance Configuration

Click on "OK" button and then click on "Apply" in Detector I page.

(Rev. 0) 37 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.4 Detector I: Calibration data settings

Close "DROCConfig" application and go to "C:\DROC\conf\" folder, open


"PanelOprConf.xml" with Notepad and check the 5 lines highlighted in
the picture below.
<HighPowerTimeOut> parameter set to "180" means that XRpad 4336
detector will go to "Sleep Mode" after 3 minutes in DROC "Exam" page
without image acquisition.
This setting is necessary to keep detector temperature under the limits
set in "Advanced Configuration".
Do not input values higher than "300".

MOVIPLAN iC 38 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

(Rev. 0) 39 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.5 Synchronization - SynBox I configuration

MOVIPLAN iC 40 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.6 Collimator configuration

In order to set the automatic collimation, set the following parameters in


"C:\DROC\conf\" files:
• TabOprConf: <Data key="Reserved1" value="1" />

(Rev. 0) 41 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

• OTCOprConf: <Data key="Reserved1" value="1" />

• GenInfoParameter: <Data key="Exam_ARP_SetCollimator"


value="YES" />

MOVIPLAN iC 42 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.2.7 Mechanical configuration

Select "OTC Type" as "ACCELERAYUARM" and DO NOT connect it.

(Rev. 0) 43 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.1.3 Workflow

5.1.3.1 Other Settings

Enable "Detector I Battery", Detector I Wifi" and "Detector I Temperature"


icon.

MOVIPLAN iC 44 (Rev. 0)
SYSTEM INTEGRATION GUIDE
DROC software configuration

5.2 Detector Calibration

At the end of DROC installation and set-up it is possible to keep old


detector calibration.
Detector calibration files are saved in "C:\DROC\references" folder.

If new detector calibration is necessary, follow the instruction below:


1. Take XRpad 4336 out of Potter-Bucky and put it on the tabletop.
2. Set SID to 120 cm or higher.
3. Center the X-ray beam on flat panel detector.
4. Open wide the collimator, so that the detector is fully illuminated.
5. Put 10 mm Al filter at collimator exit.

NOTE:
 It is possible to perform detector calibration also without aluminium filter
but in this case kV values will be low and ms values will be short if
compared to normal use.

6. Select the workplace/detector you want to calibrate; "Table" in case


of single detector configuration.
7. Click on "Begin" calibration to start image acquisition.

(Rev. 0) 45 MOVIPLAN iC
SYSTEM INTEGRATION GUIDE
DROC software configuration

The calibration procedure begins. It will lead you to acquire flat field
images at one dose level. The operator must select correct exposure
parameters in order to reach the target grey level (e.g. 24000 in the
picture).
At the end of image acquisitions click on "Save" and restart DROC
application.

MOVIPLAN iC 46 (Rev. 0)
Cod. 6998313203_Rev.0