UVVis PerkinElmer Lambda950 HardwareGuide
UVVis PerkinElmer Lambda950 HardwareGuide
LAMBDA 650/850/950
Hardware Guide
2 . Lambda 650/850/950 Hardware Guide
Release History
Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer Ltd
Chalfont Road
Seer Green
Beaconsfield
BUCKS
HP9 2FX
United Kingdom
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this document, including, but not limited to, the
implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright.
All rights are reserved. No part of this publication may be reproduced in any form whatsoever or
translated into any language without the prior, written permission of PerkinElmer, Inc.
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such,
are protected by law.
Table of Contents
Overview
This hardware guide covers the operating instructions for the Lambda 650, Lambda 850 and
Lambda 950 instruments plus those maintenance routines that do not necessarily require a
PerkinElmer Service Engineer.
NOTE: Responsible Body: Individual or body responsible for the use and maintenance of equipment
and for ensuring that OPERATORS are adequately trained.
There are separate manuals for the Sphere and Universal Reflectance accessories which can
be found on the Lambda 650/850/950 Manuals CD.
Chapter 1 Introduction
This chapter contains a brief introduction on the instrument, the conventions and warnings
used in the manual.
Important safety information is provided in this chapter. It is the same information as given
in the Lambda 650/850/950 Getting Started Guide that shipped with your instrument.
This chapter contains information on the components of the instrument, how it works and
instrument specifications.
Information on installing Lambda 650 instrument. Only the 650 should be customer
installed, the 850 and 950 instruments should be installed a PerkinElmer Service
representative. This is the same information given in the Lambda 650/850/950 Getting
Started Guide that shipped with your instrument.
Startup procedure and general instructions for using the instrument are provided.
Chapter 7 Maintenance
Maintenance and cleaning procedures for the various components of your instrument are
provided.
Lambda 650/850/950 Hardware Guide . 7
UPPERCASE text, for example ENTER or ALT, refers to keys on the PC keyboard. '+' is used
to show that you have to press two keys at the same time, for example, ALT+F.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
NOTE: A note indicates additional, significant information that is provided with some procedures.
8 . Introduction
CAUTION We use the term CAUTION to inform you about situations that could
result in serious damage to the instrument or other equipment. Details
about these circumstances are in a box like this one.
D Caution (Achtung)
Bedeutet, daß die genannte Anleitung genau befolgt werden muß, um
einen Geräteschaden zu vermeiden.
DK Caution (Bemærk)
Dette betyder, at den nævnte vejledning skal overholdes nøje for at
undgå en beskadigelse af apparatet.
E Caution (Advertencia)
Utilizamos el término CAUTION (ADVERTENCIA) para advertir sobre
situaciones que pueden provocar averías graves en este equipo o en
otros. En recuadros éste se proporciona información sobre este tipo de
circunstancias.
F Caution (Attention)
Nous utilisons le terme CAUTION (ATTENTION) pour signaler les
situations susceptibles de provoquer de graves détériorations de
l'instrument ou d'autre matériel. Les détails sur ces circonstances
figurent dans un encadré semblable à celui-ci.
I Caution (Attenzione)
Con il termine CAUTION (ATTENZIONE) vengono segnalate situazioni
che potrebbero arrecare gravi danni allo strumento o ad altra
apparecchiatura. Troverete informazioni su tali circostanze in un
riquadro come questo.
NL Caution (Opgelet)
Betekent dat de genoemde handleiding nauwkeurig moet worden
opgevolgd, om beschadiging van het instrument te voorkomen.
P Caution (Atenção)
Significa que a instrução referida tem de ser respeitada para evitar a
danificação do aparelho.
Lambda 650/850/950 Hardware Guide . 9
We use the term WARNING to inform you about situations that could
result in personal injury to yourself or other persons. Details about
these circumstances are in a box like this one.
WARNING
D Warning (Warnung)
Bedeutet, daß es bei Nichtbeachten der genannten Anweisung zu einer
Verletzung des Benutzers kommen kann.
DK Warning (Advarsel)
Betyder, at brugeren kan blive kvæstet, hvis anvisningen ikke
overholdes.
E Warning (Peligro)
Utilizamos el término WARNING (PELIGRO) para informarle sobre
situaciones que pueden provocar daños personales a usted o a otras
personas. En los recuadros como éste se proporciona información
sobre este tipo de circunstancias.
F Warning (Danger)
Nous utilisons la formule WARNING (DANGER) pour avertir des
situations pouvant occasionner des dommages corporels à l'utilisateur
ou à d'autres personnes. Les détails sur ces circonstances sont
données dans un encadré semblable à celui-ci.
I Warning (Pericolo)
Con il termine WARNING (PERICOLO) vengono segnalate situazioni
che potrebbero provocare incidenti alle persone. Troverete
informazioni su tali circostanze in un riquadro come questo.
NL Warning (Waarschuwing)
Betekent dat, wanneer de genoemde aanwijzing niet in acht wordt
genomen, dit kan leiden tot verwondingen van de gebruiker.
P Warning (Aviso)
Significa que a não observância da instrução referida poderá causar
um ferimento ao usuário.
10 . Introduction
Safety Practices
12 . Safety Practices
Overview
This chapter describes the general safety practices and precautions that must be observed
when operating the Lambda 650/850/950.
This advice is intended to supplement, not supersede, the normal safety codes in the user's
country. It is also a supplement to the PerkinElmer standard Safety and Health Policy. The
information provided does not cover every safety procedure that should be practiced.
Ultimately, maintenance of a safe laboratory environment is the responsibility of the analyst
and the analyst's organization.
Please consult all manuals and CDs supplied with the Lambda 650/850/950 and accessories
before you start working with the instrument. Carefully read the safety information in this
chapter and in the other manuals supplied. When setting up the instrument or performing
analyses or maintenance procedures, strictly follow the instructions provided.
Lambda 650/850/950 Hardware Guide . 13
Precautions
WARNING
The following precautions must be observed when using the Lambda 650/850/950:
• Connect the instrument to a correctly installed line power outlet that has a protective
conductor (earth/ground).
• Disconnect the instrument from all voltage sources before opening it for any
adjustment, replacement, maintenance, or repair. If, afterwards, the opened instrument
must be operated for further adjustment, maintenance, or repair, this must only be
done by a skilled person who is aware of the hazard involved.
• Use only fuses with the required current rating and of the specified type for
replacement.
WARNING
14 . Safety Practices
This instrument meets the Canadian Standards Association (CSA) Standard CAN/CSA-C22.2
No. 1010.1-92: Laboratory Equipment.
Only use the Lambda 650/850/950 indoors and under the following conditions:
Temperature 15 oC to 35 oC
If possible, avoid any adjustment, maintenance and repair of the opened, operating
instrument. If any adjustment, maintenance and repair of the opened instrument is
necessary, this must be done by a skilled person who is aware of the hazard involved.
Whenever it is likely that the Lambda 650/850/950 is unsafe, make it inoperative. The
Lambda 650/850/950 may be unsafe if it:
NOTE: The instrument must always be positioned so that it can easily be disconnected from the
power supply.
Lambda 650/850/950 Hardware Guide . 15
Environmental Conditions
The instrument has been designed to be safe under the following conditions:
• Indoor use
• Altitude up to 2000 m
• Ambient temperatures of 5 oC to 40 oC
WARNING
Use, store, and dispose of chemicals that you require for your analyses in accordance with
the manufacturer’s recommendations and local safety regulations.
Hazardous Chemicals
Some chemicals used with this instrument may be hazardous or may
become hazardous after completion of an analysis.
WARNING The responsible body (for example, Laboratory Manager) must take
the necessary precautions to ensure that the surrounding workplace
and instrument operators are not exposed to hazardous levels of toxic
substances (chemical or biological) as defined in the applicable
Material Safety Data Sheets (MSDS) or OSHA, ACGIH, or COSHH
documents.
Venting for fumes and disposal of waste must be in accordance with
all national, state and local health and safety regulations and laws.
Toxic Fumes
If you are working with volatile solvents or toxic substances, you must provide an efficient
laboratory ventilation system to remove vapors that may be produced when you are
performing analyses.
Waste Disposal
Waste containers may contain corrosive or organic solutions and small amounts of the
substances that were analyzed. If these materials are toxic, you may have to treat the
collected effluent as hazardous waste. Refer to your local safety regulations for proper
disposal procedures.
Deuterium lamps and other spectral lamps are maintained under reduced pressure. When
you dispose of lamps that are defective or otherwise unusable, handle them correctly to
minimize the implosion risk.
UV Radiation
• When the deuterium (UV) lamp is illuminated, do not open the spectrophotometer
covers unless specifically instructed to do so in the manual.
• Always wear UV-absorbing eye protection when the deuterium lamp is exposed.
Compressed Gases
Handle cylinders of compressed gas with care, in accordance with local regulations.
We recommend that gas cylinders be located outside the laboratory and the gases led to the
laboratory through approved gas supply lines.
Use only approved tubing, connectors, and regulators for gas supply lines.
Lambda 650/850/950 Hardware Guide . 17
Electrical Safety
The instrument has been designed to protect the operator from potential electrical hazards.
This section describes some recommended electrical safety practices.
Electrical Hazard
Any interruption of the protective conductor inside or outside the
instrument or disconnection of the protective conductor (earth/ground)
terminal is likely to make the instrument dangerous.
WARNING Intentional interruption is prohibited.
Electrical Protection
Insulation: Class I as defined in IEC 61010-1.
Installation Category: The instruments are able to withstand transient overvoltage according
to Installation Category II as defined in IEC 61010-1 and IEC 664.
Pollution Degree: The equipment will operate safely in environments that contain non-
conductive foreign matter and condensation up to Pollution Degree 2 as defined in IEC
61010-1 and IEC 664.
If any part of the instrument is not installed by a PerkinElmer service representative, make
sure that the line power plug is wired correctly:
International USA
EMC Compliance
EC directive
This product complies with the minimum immunity requirements of IEC 61326 and has been
tested to the relevant parts of the following standards:
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
This product complies with EN 55011 Group 1 Class A and with IEC 61000-3-2 and
61000-3-3.
Warning Labels
WARNING
The following electrical warnings are shown on the rear of the instrument, as identified in
Figure 1:
The following warnings are shown on the inside of the lamp compartment, as identified in
Figure 2:
Lamp compartment
Fuses
Electrical hazard
To prevent potential injury to yourself and damage to the instrument,
switch OFF all instruments in the system and disconnect them from the
line power supply before you alter, or make any new, electrical
WARNING connections.
The fuses are located in a fuse holder at the rear of the instrument, as shown in Figure 3.
Fuse Holder
1. Switch off the instrument and remove the line power cord from the electrical supply.
2. Squeeze in and lever out the two lugs at each side of the fuse holder one at a time
using a flat-headed screwdriver
Fuse Holder
4. Replace the fuses with new ones of the same type and rating (3.15 A time delay).
NOTE: The module has two fuses and you should always replace both at the same time, even if only
one blew, as the other may have been weakened.
5. Replace the fuse holder by aligning the lug at the bottom of the fuse holder with the
slot in the socket and then pushing the holder into place.
A click is heard as each lug snaps into place.
NOTE: If you use the correct fuses but the instrument still does not work correctly, or the fuses
blow repeatedly, contact your PerkinElmer office or representative.
22 . Safety Practices
The instrument weighs about 76 kg. Four people are needed to lift the
instrument.
WARNING
If you must move the instrument any great distance, especially if the
instrument may be subject to vibrations or jolts, contact your local
CAUTION PerkinElmer service representative for guidance or help.
If you are only moving the instrument within the laboratory or to a nearby laboratory then
use four people and lift the instrument using the handholds, as shown in Figure 5.
Placement of hands
under instrument
Overview
The Lambda 650, 850 and 950 are versatile spectrometers operating in the ultraviolet visible
(UV/Vis) spectral ranges. Additionally, the Lambda 950 operates in the near infrared region.
The spectrometer features a double-beam, double monochromator, ratio recording optical
system.
Power switch
Sample
compartment
Optical System
The Lambda 650/850/950 Spectrometer features an all-reflecting, double-monochromator
optical system. The optical components are coated with silica for durability. Holographic
gratings are used in each monochromator for the UV/Vis range and the NIR range.
DL M3
HL
M1 FW
NIR
Monochromator I
M2 G2
UV/Vis
SA Monochromator II
NIR
G1
UV/Vis
M5 M4
M7 M6
CBM
M9
C
R M11
M10 PM (below PbS)
RBA
Dp M12 M13
PbS
M10΄ M14
M11΄
M8
Sample Detector
Compartment Compartment
Two radiation sources, a deuterium lamp (DL) and a halogen lamp (HL), cover the working
wavelength range of the spectrometer.
For operation in the near infrared (NIR) and visible (Vis) ranges, source mirror M1 reflects
the radiation from the halogen lamp onto mirror M2. At the same time it blocks the radiation
from the deuterium lamp.
26 . System Description
For operation in the ultraviolet (UV) range, mirror M1 is raised to permit radiation from the
deuterium lamp to strike source mirror M2. Source change is automatic during
monochromator slewing.
Radiation from the respective source lamp is reflected from mirror M2 via mirror M3 through
an optical filter on the filter wheel assembly (FW) to mirror M4.
From mirror M4 the radiation is reflected through the entrance slit of Monochromator I. All
slits are located on the slit assembly (SA). The radiation is collimated at mirror M5 and
reflected to the grating table G1. Depending on the current wavelength range, the collimated
radiation beam strikes either the UV/Vis grating or the NIR grating (NIR version only).
The radiation is dispersed at the grating to produce a spectrum. The rotational position of
the grating effectively selects a segment of the spectrum, reflecting this segment to mirror
M5 and then through the exit slit. The exit slit restricts the spectrum segment to a near-
monochromatic radiation beam. Grating change is automatic during monochromator slewing.
The exit slit of Monochromator I serves as the entrance slit of Monochromator II. The
radiation is reflected via mirror M6 to the appropriate grating on grating table G2 and then
back via mirror M6 through the exit slit to Mirror M7. The rotational position of grating table
G2 is synchronized to that of G1. The radiation emerging from the exit slit exhibits high
spectral purity with an extremely low stray radiation content.
In the UV/Vis and NIR range a choice is provided between a fixed slit width, a servo slit, and
a slit program. When the servo slit is selected, the slit widths change automatically during
scanning to maintain constant energy at the detector.
From mirror M7 the radiation beam is reflected via toroid mirror M8 to the chopper assembly
(C). As the chopper rotates, a mirror segment, a window segment and two dark segments
are brought alternately into the radiation beam.
When a window segment enters the beam, radiation passes through to mirror M9 and is then
reflected via mirror M10 to create the reference beam (R).
When a mirror segment enters the beam the radiation is reflected via mirror M10΄ to form
the sample beam (S).
When a dark segment is in the beam path, no radiation reaches the detector, permitting the
detector to create the dark signal.
The radiation passing alternately through the sample and reference beams is reflected by
mirrors M11, M12, M13, and M11΄, M12΄, M13΄, respectively of the optics in the detector
assembly onto the appropriate detector. Mirror M14 is rotated to select the required detector.
A photomultiplier (PM) is used in the UV/Vis range while a lead sulfide (PbS) detector is used
in the NIR range. Detector change is automatic during monochromator slewing.
At the cell plane, each radiation beam is approximately 12 mm high. The width of the
radiation beams is dependent on the slit width. At a slit width of 5 nm each radiation beam is
approximately 4.5 mm wide.
Lambda 650/850/950 Hardware Guide . 27
To permit minimum sample volumes to be measured in micro cells, the height of the
radiation beam must be reduced in the active cell area.
A common beam mask (CBM) is mounted between the slit assembly (SA) and mirror M7. This
mask restricts the cross-section of both the sample beam and the reference beam in the
respective cell area. The radiation beam can be reduced from the maximum height of 11.7
mm to 0.0 mm in 50 steps.
During all scanning operations, the monochromators stop slewing while a filter, source, or
detector change is in progress.
NOTE: You can select the source or detector change wavelength within a defined wavelength range.
This facility is useful if a feature of special spectral interest is located at one of the default
change wavelengths.
There is an optional depolarizing filter (DP) accessory, which can be swung into the beam.
There are internal attenuators for the Lambda 850 and 950 which can be swung individually
into the sample beam (SBA) and the reference beam (RBA). The attenuators have the
values: 0%, 1%, 10% and 100%. The Lambda 650 requires manual attenuators to be
placed as required (L6020277 – 10% attenuator, L6020276 – 1% attenuator).
28 . System Description
Principle Double beam, double Double beam, double Double beam, double
monochromator, ratio recording monochromator, ratio monochromator, ratio
UV/Vis/NIR spectrophotometer recording UV/Vis recording UV/Vis
with microcomputer spectrophotometer with spectrophotometer with
electronics, controlled by DELL microcomputer electronics, microcomputer electronics,
PC or compatible personal controlled by DELL PC or controlled by DELL PC or
computer. compatible personal computer. compatible personal
computer.
Optical system All reflecting optical system All reflecting optical system All reflecting optical system
(SiO2 coated) with holographic (SiO2 coated) with holographic (SiO2 coated) with
grating monochromator with grating monochromator with holographic grating
1440 Lines/mm UV/Vis blazed 1440 Lines/mm UV/Vis blazed monochromator with
at 240 nm and 360 Lines/mm at 240 nm, Littrow mounting, 1440 Lines/mm UV/Vis
NIR blazed at 1100 nm, Littrow sample thickness compensated blazed at 240 nm, Littrow
mounting, sample thickness detector optics. mounting, sample thickness
compensated detector. compensated detector
optics.
Beam Splitting Chopper (46+ Hz, Cycle: Chopper (46+ Hz, Cycle: Chopper (46+ Hz, Cycle:
System Dark/Sample/Dark/ Reference, Dark/Sample/Dark/ Reference, Dark/Sample/Dark/
Chopper Segment Signal Chopper Segment Signal Reference, Chopper
Correction CSSC). Correction CSSC). Segment Signal Correction
CSSC).
Detector Photomultiplier R6872 for high Photomultiplier R6872 for high R955 Photomultiplier, giving
energy in the whole UV/Vis energy in the whole UV/Vis high energy throughout the
wavelength range. wavelength range. whole UV/Vis range.
Stray Light
Wavelength
Reproducibility
Photometric Accuracy
NIST 930D Filters 0.5 A +/- 0.002 A +/- 0.002 A +/- 0.002 A
Photometric Linearity
(Addition of filters
UV/Vis at 546.1 nm, Slit
2 nm, 1-sec. integration
time)
Photometric
Reproducibility
Photometric Range 8A 8A 6A
Bandpass 0.05 nm–5.00 nm in 0.01 nm 0.05 nm–5.00 nm in 0.01 nm 0.17 nm–5.00 nm in 0.01
increments UV/Vis range increments UV/Vis range nm increments UV/Vis range
Photometric Noise
RMS
(2 nm Slit, 1-sec.
integration time, Gain 1
NIR)
Sample Compartment 200 mm x 300 mm x 220 mm 200 mm x 300 mm x 220 mm 200 mm x 300 mm x 220 mm
Instrument (W x D x H)
Purging
Instrument Dimensions 1020 mm x 630 mm x 300 mm 1020 mm x 630 mm x 300 mm 1020 mm x 630 mm x 300 mm
(W x D x H)
Instrument Weight 76 kg 76 kg 76 kg
Light Beam 90 mm above the base plate 90 mm above the base plate 90 mm above the base plate
120 mm beam separation 120 mm beam separation 120 mm beam separation
3 mm–12 mm beam height 3 mm–12 mm beam height 3 mm–12 mm beam height
Instrument
Requirements
Power 90 VAC–250 VAC, 50/60 Hz; 90 VAC–250 VAC, 50/60 Hz; 90 VAC–250 VAC, 50/60 Hz;
250 VA 250 VA 250 VA
Recommended 10–70% relative humidity, 10–70% relative humidity, 10–70% relative humidity,
Humidity non-condensing non-condensing non-condensing
32 . System Description
Lambda 650 Installation
34 . Lambda 650 Installation
Overview
CAUTION For the Lambda 850 or 950 please arrange for installation with your
local PerkinElmer Service Representative.
Requirements
For maximum stability and minimum maintenance, observe the following requirements when
choosing where to site the instrument:
• Enough space around and underneath the instrument for efficient air circulation, as
shown in Figure 8.
• A suitable source of electrical power should be located in the vicinity of the instrument,
in this case a proper earth-grounded 3-wire electrical outlet.
• The standard sample compartment baseplates have drain holes in them to run off
spilled liquids to the benchtop underneath the instrument. If required, place a sheet of
thick filter paper under the instrument.
Unpacking
Take great care when installing your Lambda 650, and follow the
procedures described in this manual. If you require assistance, contact
CAUTION your local PerkinElmer Service Representative.
The instrument weighs about 76 kg. Four people are needed to lift the
instrument.
WARNING
1. Unpack the components carefully, using four people to lift the instrument itself.
Keep the packing materials for possible future storage or reshipment.
Placement of hands
under instrument
NOTE: The two manual attenuators are only supplied with a Lambda 650.
• Check the entire outer cabinet of the spectrometer for damage, and make sure
that terminals, fuse holders, etc. are not damaged.
• Open and close the sample compartment cover, checking that it moves freely
without binding.
• Make sure that the compartment is free of dust or other foreign matter.
NOTE: In the event of damage or missing parts, file an immediate claim with the authorized carrier,
and inform your PerkinElmer office or representative.
Lambda 650/850/950 Hardware Guide . 37
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument,
first make any electrical connections in the system before connecting to
WARNING the line power supply.
The spectrometer automatically adjusts to the correct operating voltage.
Before starting the instrument for the first time, make sure that the
correct fuse is fitted to your line power supply.
• If you have a different fuse, change it for the correct one.
• Do not connect the spectrometer to the line power supply if the
wrong fuse is fitted.
4. Make sure that the correct fuses are fitted in the holder at the rear of the spectrometer
(see Fuses on page 21).
5. Make sure that the plug fitted to the line power cord provided with the spectrometer is
suitable for your local electrical outlets.
If it is not, remove it and fit a plug conforming to the local regulations.
6. After all connections have been made between the various components of the system,
make certain that all Power switches are set to off, then connect the line cords to the
electrical power supply.
The Power switch is located at the top right-hand rear of the spectrometer.
NOTE: To prevent interferences caused by earth loops when operating with ancillary instruments
(for example, printers), connect all components of the system to the same phase of the
electrical supply via a multisocket distributor.
38 . Lambda 650 Installation
Electrical Hazard
• The PC and the printer may be placed either side of the spectrometer, but must not be
placed on top of the spectrometer.
When placing instruments side by side, always leave a small gap between them.
• Make sure that air can circulate freely over and under the system as well as behind it.
• Do not place anything on or under any of the components that could hinder free air
circulation.
Connecting to the PC
¾ The PC is connected using the connection marked COMPUTER on the right side of the
instrument, as shown in Figure 10.
PC connection
Figure 10 PC Connection
NOTE: We suggest that you place the ferrite choke supplied with the instrument around the lead.
NOTE: You can use an external printer for hard-copy printouts of the analytical results. The printer
is connected via the PC.
To install the software, refer to the instructions that shipped with your software package.
Lambda 650/850/950 Hardware Guide . 39
NOTE: If your instrument is Service installed you should not need to do this.
1. With the instrument switched on, in the Main pane of the UV WinLab Explorer select
Instruments.
The Instruments page is displayed.
4. Click Next.
40 . Lambda 650 Installation
5. Select the instrument type to add from the drop-down list, and if required, select
Make this the default instrument.
6. Click Next.
7. From the drop-down list, select the Port the instrument is connected to.
8. Click Next.
The software will automatically perform a check to see if the correct instrument is
attached to the selected port and switched on, and will display a Warning message if
not.
You may then be asked to confirm the serial number of the instrument.
9. Check the serial number on the right-hand side of the instrument, where the PC
connects to the instrument, correct it if necessary, and click Next.
Lambda 650/850/950 Hardware Guide . 41
NOTE: Manual Attenuators are only available for the Lambda 650, the Lambda 850 and 950 have
automatic attenuators. The Common Beam Depolarizer requires a firmware command to be
enabled.
13. If all the settings are correct, click Finish to close the wizard and add the instrument.
The Wizard closes, a confirmation message that the instrument has been successfully
installed is displayed, and the instrument is displayed in the Explorer.
OR
Click Back to return to the previous page(s) and amend the settings as required.
42 . Lambda 650 Installation
NOTE: Make sure that the instrument has been switched on for at least an hour before calibrating it.
14. From the Instruments page of the Explorer, right-click on the new instrument and select
Calibrate.
The Calibration Utility dialog is displayed.
The dialog displays the names of the routines, a description, the last calibration date of
the routine, and whether the test is enabled.
15. To view (and change) the settings for a particular routine, highlight the routine in the
table and then click Settings.
The settings dialog for the selected routine is displayed.
18. When the calibration has finished, click Close to close the dialog.
The date and time of the calibration is recorded in the instrument event log.
¾ Select Manual set, Manual set 1 peak, Manual set 2 peaks, Auto search 1 peak,
or Auto search 2 peaks.
• Manual set enables you to enter the offset and factor (without the need to
determine the values in advance).
• Manual set 1 peak enables you to change the offset. A well known peak can be
shifted to an exact wavelength by entering values of Old Peak 1 and New Peak
1.
Lambda 650/850/950 Hardware Guide . 43
• Manual set 2 peaks enables you to change the offset and factor by specifying
the old and new values of 2 well known peaks by entering values of Old Peak 1,
New Peak 1, Old Peak 2 and New Peak 2.
• Auto search 1 peak enables you to change the offset. The software performs an
automatic search for the D2 peak at 656.1 nm. The measured peak is then shifted
to the exact wavelength.
• Auto search 2 peaks enables you to change the offset and factor. The software
performs an automatic search for the D2 peak at 656.1 nm and the peak at 0.0 nm.
The measured peaks are then shifted to the exact wavelengths.
0%T
OR
Startup procedure
NOTE: If you have purchased accessories you may want to install these first before commencing
with the Startup procedure. An overview of accessories is available on page 70, see also the
specific documentation for the accessory.
Do not turn power to the instrument on and off quickly as this may
damage the power supply.
CAUTION
Wait at least 30 seconds before re-starting.
2. Make sure that the beam paths are free from obstruction.
This means that no objects (for example, cables) project into the beam paths, no
samples are in the sample compartment, and any accessories are properly installed.
NOTE: If the sample compartment is obstructed during the startup procedure, the spectrometer will
not initialize correctly.
4. Switch on the power switch located at the top right hand corner of the instrument as
shown in Figure 11.
The switch will light up when there is power to the instrument.
Power switch
6. Wait until all instrument initialization is complete before starting the UV WinLab
software.
This will take approximately three minutes from switching the power on.
NOTE: We recommend that for optimal performance of the instrument you should leave the
spectrometer switched on for approximately one hour to allow the lamps to warm up and
stabilize before starting analysis.
Lambda 650/850/950 Hardware Guide . 47
Description
There are two single-cell holders provided with the instrument, one for the sample beam and
one for the reference beam. The single-cell holders are mounted on a plinth to bring them in
line with the radiation beam.
Locking screw
for horizontal
alignment
Vertical adjustment
screw
Milled post
Lifter
Locking screw
for horizontal
alignment
1. Lower the plinth so that the two locating holes slip onto the two locating pins on the
baseplate in the sample compartment, as shown in Figure 13.
The plinth only fits one way round.
Thumbscrew Thumbscrew
3. Orientate the holder so that the word LAMBDA is toward the front of the sample
compartment (see Figure 14 below).
4. Lower the holder so that the two locating holes slip onto the two studs on the two
locating pins on the plinth in the sample compartment.
48 . Using the Instrument
Retaining Screw
Cover Plate
Sample Holder
and Plinth
Tube Ports
5. Move the milled posts a little to locate the threaded holes in the baseplate, and then
tighten the milled posts.
You can lead tubes and connecting cables for accessories into the sample compartment
through either the tube ports, located at the front of the sample compartment, or the
opening located at the left rear of the sample compartment under the cover plate.
When not in use, you should always insert the caps into the tube ports and keep the
cover plate fixed in place.
The cover plate is fixed in place by a retaining screw.
3. Insert one cell into the sample cell holder and one into the reference cell holder.
Make certain that the cell is pushed down fully.
NOTE: The alignment procedure is for a given cell in a given holder. After alignment, the cell should
always be used in the same holder.
4. From The UV WinLab Explorer, select the instrument you are using and select Manual
Control from the Tools menu.
5. From the Manual Control window, select Data Collection in the treelist and then select
Alignment Mode on the page.
6. Block the sample and reference beam windows on the right side of the sample
compartment with a card to prevent white light from saturating the detector.
Lambda 650/850/950 Hardware Guide . 49
7. By holding a piece of matt white paper behind each cell holder, visually examine the
light spot to see that the radiation beam is passing through the cell sample area.
Diffraction patterns become apparent if the radiation beam impinges on the cell wall.
8. If the radiation beam is not centered exactly, loosen the two locking screws and the two
milled posts on the relevant cell holder and shift the cell holder plate to center the
radiation beam.
Then retighten the two milled posts and the two locking screws.
9. Visually check the vertical alignment of the radiation beam in the cell sample area.
Alignment is correct when the radiation beam is just above the floor of the cell sample
area (minimum 2 mm) or covers the cell window.
NOTE: The center of the window for micro flowcells should be ideally approximately 15 mm above
the base of the cell.
Min. 2 mm
Figure 15 Correct alignment of the radiation beam in the cell sample area
10. If alignment is required, turn the vertical adjustment screw on the lifter either clockwise
to raise the cell, or counterclockwise to lower the cell.
11. Recheck the horizontal alignment of the radiation beam through the cell and correct if
necessary.
13. Remove the card blocking the sample beam window and close the sample compartment
cover.
This completes the coarse alignment of the cell holders. If necessary, proceed with the
fine alignment as described below.
Fine Alignment
If fine alignment is necessary, proceed as follows:
5. Make horizontal fine alignment to the sample cell holder (locking screws and milled
posts loosened) to obtain the highest possible transmittance reading on the display.
Close sample compartment cover while measuring transmittance.
6. Make fine alignment using the vertical adjustment screw again to obtain the highest
possible reading.
Close sample compartment cover while measuring transmittance.
7. When you are satisfied with the alignment, tighten the milled posts and the locking
screws on the cell holder.
10. Repeat steps 4 to 6 with the reference cell holder, but this time obtain the lowest
possible transmittance reading on the display.
This completes the fine alignment procedure.
NOTE: When the cell holder has been aligned once, you can take it out and reinstall it without
aligning it again.
The minimum sample volume required is a function of the cell internal width or volume and
is specified below.
Lambda 650/850/950 Hardware Guide . 51
5 µL 1.0 cm 10 µL B0505823
30 µL 1.0 cm 50 µL B0190608
NOTE: You should align microcells very carefully in the radiation beam by following the procedures in
Aligning the Single-cell Holder on page 48. When aligning microcells, fill each cell with the
minimum volume of liquid specified in the above table to make sure that the liquid meniscus
is not in the radiation beam.
52 . Using the Instrument
Always purge the instrument with an inert gas such as Nitrogen. Never
use a flammable gas such as Hydrogen.
WARNING
Oxygen absorbs radiation in the UV range below 190 nm, while water vapor absorbs
radiation in the NIR range between 1350 nm and 1450 nm, 1850 nm and 1950 nm, and also
between 2520 nm and 3000 nm. Best accuracy for measurements in these spectral regions
can only be obtained when the spectrometer is purged with nitrogen.
The entire optical compartment of the spectrometer is sealed. Radiation from the source
lamps enters the optical compartments through a window. The sample compartment is also
sealed from the optical compartment by windows. Thus the entire optical radiation path from
virtually the source lamp to the detector can be purged largely free of water vapor and
oxygen.
Fittings are provided on the left side of the spectrometer for connection of a supply of
nitrogen, as shown in Figure 16. The inlet nearest the rear of the instrument is for the
sample compartment and the one nearest the front of the instrument is for the main optical
compartment. After purging these compartments the nitrogen escapes to atmosphere.
Purge inlets
You should keep the sample compartment closed during routine operation to prevent the
ingress of water vapor and oxygen. Do not leave the sample compartment cover open for
longer than is necessary to perform operations within the sample compartment. After closing
the sample compartment purge for several minutes before starting the measurement.
When the spectrometer has been standing for some time unpurged, an initial flow of 20
L/min is recommended. After the spectrometer has been thoroughly purged, the flow can be
reduced to between 6 L/min to 7 L/min.
When you have a good deal of work in a spectral region where water vapor or oxygen
absorb the radiation, we recommend that you purge the spectrometer more or less
continuously. This has the advantage that startup times are markedly shorter. This
advantage is somewhat offset by the increased consumption of nitrogen.
Lambda 650/850/950 Hardware Guide . 53
NOTE: A Nitrogen Purge Accessory Kit, comprising of a filter and a flow controller with flowmeter, is
offered as an option. With this kit you can control the gas flow more accurately.
NOTE: The Lambda 650 does not have this purge facility.
54 . Using the Instrument
Maintenance
56 . Maintenance
Overview
This chapter describes the routine maintenance procedures required to keep your instrument
in proper working condition and to ensure the highest possible level of performance. It
includes maintenance checks that should be done on a daily basis and maintenance
procedures that should be done periodically depending on instrument use. This chapter is
divided into several sections, each section covering maintenance procedures for a particular
component of the system.
You should perform only the maintenance procedures described in this chapter. If additional
maintenance is required, contact a PerkinElmer Service Engineer.
Daily Checks
The instrument is constructed with high quality components and requires little maintenance
other than to keep it clean and free of dust.
To protect the optical system from dust and fumes, you should keep the sample
compartment cover closed except for when you are carrying out work in the compartment.
You should observe the following care routine to maintain your instrument in good condition:
• Immediately clean all spilled materials from the affected area and wipe it dry with lint
free paper or cloth.
If you have to wipe the sample compartment windows, make sure you do not introduce
scratches. The windows are optical components and you should handle them in the
sampe way as high quality cells.
• Do not leave samples, particularly those given to fuming or evaporation, in the sample
compartment for longer than necessary.
• If any type of sample handling system is installed and portions of it are left in the
sample compartment (such as a sipper and flowcell), make certain that the system is
cleaned at the end of the working day.
Generally, such systems should be filled with deionized water when left overnight.
Spills
Take care not to spill liquids onto the spectrometer.
CAUTION Expensive damage can result to the optics or electronics if liquids are
spilled and run inside the instrument.
Drain Holes
1. First remove the cell holder or other sample handling accessory from the sample
compartment.
2. Using a soft cloth and mild laboratory detergent solution, lightly scrub away all foreign
material.
3. Using a clean cloth dampened with water, rinse the cleaned surfaces thoroughly.
The windows seal the sample compartment and thus protect the instrument’s optics from
dust and fuming or aggressive samples.
• The windows are an optical component and require the same care and handling as cells.
• The windows can be removed to clean them. Each window has a magnetic frame and
can be carefully removed by hand.
Windows are most suitably cleaned by wiping them with a soft cloth moistened with
ethanol.
Replace the window in its original position after cleaning.
any optical component. Optical faults of a minor nature, scratches, lint, finger marks, etc. on
the optical surfaces can easily introduce substantial analytical errors.
You should observe the following list of cell handling rules to prevent analytical errors and to
achieve utmost precision:
• Only hold cells by non-optical surfaces, such as the matt finish surfaces.
• Protect cells from scratches, and never permit them to rub against one another or
against other hard surfaces.
• Avoid abrasive, corrosive or stain-producing cleaning agents, and make certain that the
exposed surfaces of cells are optically clean.
• Always wipe the optical surfaces of cells dry and free of finger marks, using a soft cloth
or cleaning tissue, just before placing them in the cell holder.
• When measuring cold solutions, always bear in mind that condensation can form on the
optical surfaces.
• Make certain no bubbles cling to the inner surfaces of the cell, particularly when
handling cold solutions.
For maximum precision and accuracy, calibrate and test with cells of the same type, and
always insert cells into the holders with the same orientation.
60 . Maintenance
Replacing a Lamp
If a lamp failure warning message is displayed, or you are seeing excessive noise on the
baseline, you can replace the lamps as follows.
Electrical Hazard
High voltages are present at the lamp connectors in the lamp
compartment – severe electric shock hazard.
Never unplug or plug in a lamp cable while power is ON.
WARNING Switch off the spectrometer and remove the plug from the electrical
supply before starting with the replacement.
UV Radiation
The lamps emit intense UV radiation which can damage your eyes.
Do not open the lamp compartment when the lamps are on.
Do not gaze into a lighted lamp.
The source lamps are located in the lamp compartment at the top left rear of the
spectrometer, as shown in Figure 18.
Lamp compartment
1. Switch off the spectrometer and disconnect the line power cord, if necessary wait until
the lamps and lamp compartment have cooled to room temperature.
2. Remove the lamp compartment cover by lifting vertically from the handhold on the rear
of the instrument.
3. Remove the lamp cover by undoing the two screws and lifting vertically.
4. Carefully pull the halogen lamp connector from the rear of the halogen lamp, as shown
in Figure 21.
62 . Maintenance
Halogen lamp
connector
Halogen lamp
assembly
5. Remove the halogen lamp assembly from the bracket by slackening the thumbscrew
and pulling the lamp mount vertically upward.
Save the thumbscrew for use with the new lamp assembly.
6. Unpack the new lamp assembly, taking care to hold it only by the metal mount to
prevent finger marks on the bulb.
7. Slip the slot at the base of the lamp mount over the stud on the bracket in the lamp
compartment and then secure with the thumbscrew.
8. Carefully push the halogen lamp connector firmly onto the pins on the base of the
halogen lamp.
9. Wipe the halogen lamp bulb with a soft cloth moistened with alcohol to remove dirt,
since this would otherwise be burned in when the lamp is hot.
NOTE: Due to the pre-aligned mounts, the alignment of lamps after installation is generally so good
that further alignment is not required.
NOTE: An operating hours counter is incorporated in the red deuterium lamp lead. By means of a
gap between the two display bars it is possible to read off the number of hours that the lamp
has been in operation. One scale division corresponds to approximately 100 hours.
1. Switch off the spectrometer and disconnect the line power cord, if necessary wait until
the lamps and lamp compartment have cooled to room temperature.
2. Remove the lamp compartment cover by lifting vertically from the handhold on the rear
of the instrument.
3. Remove the lamp cover by undoing the two screws and lifting vertically.
4. Unplug the deuterium lamp connector by squeezing in the two lugs at each side of the
connector and carefully pulling the connector vertically upward.
5. Remove the lamp assembly from the bracket by slackening the thumbscrew and pulling
the lamp mount vertically upward.
Save the thumbscrew for use with the new lamp assembly.
6. Unpack the new lamp assembly, taking care to hold it only by the metal mount to
prevent finger marks on the lamp window.
7. Slip the slot at the base of the lamp mount over the stud on the bracket in the lamp
compartment and then secure with the thumbscrew.
NOTE: The socket in the lamp compartment is asymmetric; the deuterium lamp connector can be
inserted in one direction only. Make certain that the connector is the right way round before
inserting it. Never attempt to insert the connector by force.
9. Wipe the lamp window with a soft cloth moistened with alcohol to remove dirt, since
this would otherwise be burned in when the lamp is hot.
NOTE: Due to the pre-aligned mounts, the alignment of lamps after installation is generally so good
that further alignment is not required.
64 . Maintenance
Replacement Parts
Supplies, accessories, and replacement parts can be ordered directly from PerkinElmer.
e-ssentials, PerkinElmer’s catalog service, offers a full selection of high-quality ultraviolet
supplies, see essentials.perkinelmer.com.
Accessory Installation
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument,
switch OFF all instruments in the system and disconnect them from the
line power supply before you alter, or make any new, electrical
WARNING connections.
NOTE: To operate the spectrometer with some accessories, for example Peltier cell changers, you
need an accessory printed circuit board (PCB) fitted in the connector panel. The accessory
PCB will be installed by a PerkinElmer service engineer.
This section of the manual details changes that may need to be made to the basic
spectrometer before installing an accessory and an overview of the installation required for
some of the specific accessories. Full procedures for installing the accessories in the
spectrometer are described in the directions provided with the respective accessories.
2. Push back on the base of the lid to release the three lugs holding it in place.
NOTE: To re-fit the lid, make sure the lugs are properly inserted into the holes and then holding the
lid vertically, pull it towards the front the instrument making sure the lugs are fully locked
into place.
66 . Maintenance
¾ Each window has a magnetic frame and can be carefully removed by hand.
Replace the window in its original position.
1. With the sample compartment cover open or removed, undo the four thumbscrews
shown in Figure 24.
Thumbscrews Thumbscrews
2. Lift the baseplate up off the locating pins and remove it from the sample compartment.
1. After removing the sample compartment cover and baseplate as described above, undo
the four retaining screws shown in Figure 25.
Lambda 650/850/950 Hardware Guide . 67
Retaining Screw
Locating Slot
Retaining Screws
Retaining Screw
Locating Hole
2. Lift up the sample compartment off the locating pins and remove it from the
spectrometer.
Electrical Hazard
The connector for the detector unit carries high voltage (130 V dc).
Switch off and disconnect the instrument before removing or fitting a
detector/accessory unit.
WARNING
1. Switch off the spectrometer and disconnect the line power cord.
3. Undo the two retaining screws located under the blanking caps on the top of the
detector (Figure 26) and are accessible using the long posidrive screwdriver supplied
with the instrument.
4. Remove the detector compartment cover by carefully lifting vertically using the two
hand holds, as shown in Figure 27.
Hand holds
5. Move the detector unit carefully to the right and remove from the spectrometer.
The detector unit is installed by performing this procedure in reverse, tightening the
retaining screws and making sure the blanking caps are in place.
Make sure you screw down the detector unit as this creates the earth
bonding for elctrical safety.
WARNING
Details of the installation procedure for a sphere or Universal Reflectance Accessory (URA)
are given in the relevant manual on the Lambda 650/850/950 User Manuals CD.
Lambda 650/850/950 Hardware Guide . 69
NOTE: The insulation of externally connected accessories should meet the requirements of IEC
61010-1 or IEC 60950.
External PC
Printer Temperature Sensor
Sipper
Temperature Sensor
NOTE: When connecting the temperature sensor, align the red mark on the plug with the red mark
on the socket. Pull back the collar on the plug to connect/disconnect the plug. Release the
collar to secure the plug.
Sippers
Sipper
Spectrometer
Connector Panel
Pin Configuration
There are two 15-pin connectors fitted to the connector panel of the Lambda 650/850/950
spectrometer to connect accessories. The connectors are identically configured so that the
accessories can be connected to any of them. The pin numbering is shown in Figure 31, and
the configuration is given in the following table.