1. Open case n° 1 removing the cover and the side indicated as "side to open".
2. Remove all packing materials, all singled packed parts and the column assembly,
fixed to the bottom side of the case by means of tie rods, in order to freely access the
tilting assembly.
3. Remove the tie rods that fix the pallet with the tilting assembly to the bottom side of
the case. Remove all lateral covers of the case.
4. Adjust the height of the transportation dollies p/n SXAA02 acting on the wheeled
jackscrews as shown in the following figure:
5. Attach the transportation dollies p/n SXAA02 to the CLISIS base using M10 screws A
and to the main frame using M8 screws B as shown in the following figure.
7. Raise the base up using the wheeled jackscrews taking care to keep it relatively level
and slide it into the room.
The transportation dollies SXAA02 are not shipped with each unit.
They must be purchased as separate item by the installer and re used for several
units.
8. Position the tilt assembly on the base plate (over or under floor) interposing the base
spacer p/n SXAA014 as detailed in section 5 of the manual.
9. Verify for perfect level alignment with the floor using, if necessary, correction spacers.
10. Lock the tilt assembly to the base plate using, if necessary, the additional spacer and
the fixation screws kit as detailed in section 5.5.
2. Fix the tube stand to the sliding base using the eight M8 screws 3 and tighten them
with a force not exceeding 3,5 Kgm as shown in the following figure.
3. Mount the end switches assembly 1 and 2 using the four M4 screws 3 as shown in
the following figure.
4. Mount the tube stand control board ref. 006050.167 and the FFD driver on the right
side of the tube stand as shown in the following figure.
1. Remove the tie rods that fix the SFD pallet to the bottom side of the case.
2. Remove the fixation screws mounted on the SFD supporting plate of the CLISIS
table.
3. Load the SFD assembly on a transpallet as shown in the attached figure and position
it on the front side of the table.
4. Lift the assembly up until the SFD fixation holes are aligned with the CLISIS table
mounting plate as shown in the following figure:
5. Insert the eight M8 fixation screws 2 and tighten them with a force not exceeding 3,5
Kgm as shown in the following figure:
6. Insert the four M6 fixation screws 3 and tighten them with a force not exceeding 2
Kgm as shown in the following figure:
6. Mount the four fixation nuts 1 and tighten them slightly to prevent a damage of the
grains as shown in the following figure.
7. Drive the transpallet away form the unit and remove the SFD shipping frame 4 taking
out screws 5 as shown in the following figure.
9. Fix all cables connectors to the related mating plug on the SFD control board and the
ground cable to SFD central ground terminal as shown in the following figure.
10. Mount the end switches assemblies as shown in the following figure.
The compression cone assembly bar must be coupled with the tomography tower, the SFD
parallax correction and the tube arm. The operation is delicate and requires an assistant in
addition to the installer. During this procedure, it is very important to be careful carefully
couple the bearing of the tomography tower with the running slot locate at the back side of
the compression bar without damaging the bearing itself.
1. Align the SFD with the tomography tower in order to be able to insert vertically the
compressor bar as shown in the following figure.
2. Gently insert the compressor bar into the tomography tower bearing and the SFD
support, taking care not to force the guidance bearing as shown in the following
figure.
4. Remove screws 4 to completely free the tube stand from the chain attachment as
shown in the following figure.
6. Remove the safety lock 4. Fix the compressor bar to the tube stand using the three
M6 screws 1 and tighten them with a force not exceeding 2 Kgm. Remove the
compressor bar and SFD transportation locks 2 and 3 as shown in the following
figure.
Once removed the safety lock 4 the compressor bar assembly will be free to
vertically move along the tomography bar.
7. Mount the compressor assembly cables chain and route the interconnection cables
as shown in the following figure.
Fix all cables connectors to the related mating plug on the tube stand control board
and the ground cable to the stand central ground terminal as shown in the following
figure.
The CLISIS table has been designed in order to accept only the x-ray tube assemblies
listed in section 2.6. For each tube a specific attachment has been designed and tested.
It is installer responsibility to contact Mecall support in case the tube assembly that
is supposed to be installed on the unit is not present in the compatibility list.
Do not install non approved tube assembly if not authorized by the Mecall support.
The tube arm support is delivered with the X-ray tube adaptation block already installed.
2. Open the trunnion ring of the x-ray tube removing the screws 1 that connect the two
half-ring 2 and insert the X-ray tube.
3. Insert the x-ray tube into the trunnion ring and mount the two half rings 2.
4. Verify that the rotation of the x-ray tube inside the trunnion ring is regular and lock
screws 1.
1. Mount the tube lower cover that is fixed to the tube support by means of screws 11
and to the tube arm cover by means of screws 5.
WARNING:
During the installation of the tube lower cover the installer must pay attention that wings 19 of
the tube support pass trough the slots of the cover and that the cover itself is correctly fixed
to its support and to the x-ray tube.
2. Mount the collimator cone 15 to the tube output port using screws 14.
IMPORTANT NOTICE
THE COLLIMATOR CONE HAS BEEN DESIGNED TO PERFECTLY MATCH THE X-RAY
TUBES APPROVED FOR THE CLISIS TABLES. THEREFORE NO ADDITIONAL
SPACERS MUST BE INSERTED BETWEEN THE TUBE OUTPUT PORT AND THE
MOUNTING CONE FOR ANY REASON.
3. Unscrew the 4 socket head screws 13, which control the adapter seal tongues using a
3-mm hexagon key 12.
4. Place the collimator on the cone and lock screws 13 in the way to fix the collimator in
a position mechanically centered with the cone.
5. Mount the tube rotation handle 18 taking care that cables 17 is routed through the air
slot of the tube lower cover. Using screws 16 connect the handle to the x-ray tube
support wings 19.
7. Route the tube cables set (H.V, rotor, DAP) as shown in the following figure:
WARNING:
Check carefully that all cables routed from the column to the x-ray tube have been
displaced in the way to allow the 180° rotation of the x-ray tube. Check also that the
cables are long enough to set the FFD to its maximum position (150 cm).
The CLISIS table has been designed to accept different models of Image
Intensifiers. For each of them has been a specific attachment kit has been designed.
The installer must makes sure that the required adaptation kit is available at the
installation side before starting the installation of the TV system.
1. Mount the Image Intensifier adaptation support 1 to the Image intensifier itself. The
mounting procedure depends on the specific I.I. model and it is shown in the following
figure:
2. Til the unit to the +90° position in order to have an easy access to the SFD bottom
side.
3. Remove the six fixation nuts from the Image Intensifier support.
4. Insert the number of spacers specified in the previous drawings and necessary to
align the Image Intensifier center with the SFD center.
5. Fix the I. I. assembly to the SFD support using the six nut M5 and tighten them with a
force not exceeding 1,5 Kgm as shown in the following figure.
1. Remove the rear cover 6. Mount the tabletop left and right support 1 using the M8
fixations screws 2 and tighten them with a force not exceeding 3,5 Kgm as shown in
the following figure. The correct position is indicated on both sides by a factory
alignment screw 3.
2. Mount the tabletop movement linear actuator 3 into the right support arm and connect
CN313 and CN321 with its matching plugs 9 and 10 as described in the following
figures
3. Position the tabletop assembly 2 over the supporting arms 1 as shown in the following
figure. Fix it to the supporting arms 1 using the four M8 screws 3 and tighten them
with a force not exceeding 3,5 Kgm
1. Remove the rear cover 6. Mount the tabletop right support 1 using the eight M8
fixations screws 2 and tighten them with a force not exceeding 3,5 Kgm as shown in
the following figure. The correct position is indicated on both sides by a factory
alignment screw 2.
2. Mount the tabletop movement linear actuator 3 into the right support arm and connect
CN313 and CN321 with its matching plugs 9 and 10 as described in the following
figures
4. Slide the linear bearing 1 along the tabletop rails 2 and fix them to the SFD
support 3 by means of screws 4 as shown in the following figure. Tighten the
screws with a force not exceeding 1,5 Kgm.
2. Route the console cable, the I/O safety cable, the ground cable and the power
supply/emergency cable up to the control desk.
4. At the control desk connect the fluoro foot switch cable to plug CN1001.