Egg Analyzer - Height Calibration Procedure ALTURA

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

EGG ANALYZERTM

HEIGHT CALIBRATION

PROCEDURE

08-HCP-001 Version A

A 1
NOTICE

This manual contains confidential information that is proprietary to Orka Technology Ltd. and is
subject to contractual restrictions on the use and disclosure hereof. No part of this publication may be
reproduced without prior written approval from Orka Technology Ltd..
The information contained herein is intended for the sole and exclusive use of customers of Orka
Technology Ltd. in connection with the device. Any other unauthorized use of this manual or any of
the information it contains is prohibited.
No representations for fitness for any purpose, other than what is specifically mentioned in this
manual, are made either by Orka Technology Ltd. or its agents.
Orka Technology Ltd. reserves the right to revise this publication without undertaking any obligation
to notify anyone of such revisions.
Should you have any questions about this device or require assistance, contact:

Manufacturer:
Orka Technology Ltd.
Email: [email protected]

A 2
The calibration procedure includes the following steps:

• Press the O/C+PRINT buttons simultaneously and turn ON the unit in order to
enter the calibration mode.
• Follow the guiding instructions as appears on the LCD screen.

Note The height calibration below applicable to unit with software version 3.00 and above.

The unit will present the following message:


Sample 0mm level,
Adjust best location
PRINT=IN MEMORY=OUT
Press START
Empty tray (remove the plate from tray) and use the MEMORY and PRINT buttons to slide the tray
IN and OUT until you place the tray bellow the Ultrasonic sensor to measure the zero level.
Tray position should be exactly as shown in the following figure.

Then press START.


The unit will read the zero level and present the message:
Sample 15mm level,
Adjust best location
PRINT=IN MEMORY=OUT
Press START

A 3
Place the calibration Jig so the high part is headed into the unit and use the MEMORY and PRINT
buttons to slide the tray IN and OUT until you place the tray bellow the Ultrasonic sensor to measure
the 15mm level.

Tray and Jig position should be exactly as shown in the following figure (Most of the JIG is located in
the unit and you can only see the 3mm section of the test JIG position on the center of the tray round
holder hole).

Then press START.


The unit will present the following message:
Calibration Test
Adjust best location
For object to test
High = 15.00 mm
The last line of “High = 15.00 mm” is dynamic and will change according the actual high it is
measuring. It is important to press the START button as soon as the message shows 15.00 mm to set
the limit properly.
The longer you wait the lower the value will go due to the shifting of the sensor’s accuracy (If the
value goes lower than 14.80, it is recommended to perform the high test again).
When you see the High = 15.00 mm - Press the START button
Egg Analyzer
Adjust best location unit
Press START

THE UNIT WILL MOVE TO THE MAIN SCREEN- END OF CALIBRATION PROCEDURE.
Turn the Unit OFF and ON again.