Method Validation and Verification
Method Validation and Verification
ISO/IEC 17025
Non-standard methods;
Laboratory-designed /developed methods;
Standard methods used outside their intended scope;
Amplifications and modifications of standard methods.
Eurachem
Validation is also required when it is necessary to demonstrate
the equivalence of results obtained by two methods, e.g. a
newly developed method and an existing standard / regulatory
method.
Method Verification
There is no need to estimate the LOD or LOQ for methods that will always be
applied to measure analyte concentrations much greater than the LOQ.
1 1
𝑠𝑜′ = 𝑠𝑜 +
2 1
LOQ = SD x 10; or
= LOD x 3; or
= Blank value + 10 x SD; or
= 50% above the lowest
fortification level used to
validate the method; or
= LOD x 10
NATA Technical Note 17 - October 2013
Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Linearity and Range
Instrument Working Range
Method Working Range
Residuals
0
0.4 0.00 0.20 0.40 0.60 0.80
0.3 -0.005
0.2
-0.01
0.1
0 -0.015
X Variable 1
0.00 0.20 0.40 0.60 0.80
Linearity and Range
Instrument Working Range
Method Working Range
i) Samples with known concentrations and sample blank should be available
ii) The samples used should be taken the entire measurement procedure
iii) The concentrations of the different samples should preferably cover the whole
range of interest
iv) The instrument should have been calibrated according to the suggested calibration
procedure.
The variability of ion ratios should preferably be determined from calibration standards during the
initial method validation and subsequently during routine analysis. In certain cases, these data may
be used to set performance-based criterial, for individual analyte, rater than applying the fixed,
generic criteria given in Table 5.
SANCO/12571/2013 - 01/01/ 2014
Selectivity
Clenbuterol in Pork LC-MS/MS analysis
The inherent problem with spike samples is that analyte introduced in such
a way will probably not be bound as strongly as that which is naturally
present in the test portion matrix and so the technique will give an
unrealistically high impression of the extraction efficiency.
Wet overnight
Wet overnight
w/o wetting
DWM Sin, YL Wong. Analytical and Bioanalytical Chemistry To be published.
Comparison with results with another method
Intermediate Precision
Single laboratory, different analysts,
different equipment, extended timescale
Reproducibility
Variability by different laboratories,
Different analyst, different equipment, long time scale
Standard Deviation
sr , s i
One-way ANOVA
BS EN 12857:1999
Statistics Tools for Verification
F-test by comparing the precision of the two methods
Should use repeatability standard deviation
BS EN 12857:1999
Ruggedness
A measure of its capacity to remain unaffected by small,
but deliberate variations in method parameters.
Ruggedness provides an indication of the method’s
reliability during normal usage.
ISO/IEC 17025