Method Validation and Verification

Method validation is required to demonstrate that an analytical method is suitable for its intended use. It establishes the method's performance characteristics including accuracy, precision, linearity, range, selectivity, limit of detection, and limit of quantitation. Validation is important to have confidence in test results and ensure the method is fit for its intended purpose. It should be performed when developing new methods or using standard methods outside their intended scope. Method verification also confirms that a laboratory can correctly perform standard methods.

Uploaded by

fadli
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
418 views

Method Validation and Verification

Method validation is required to demonstrate that an analytical method is suitable for its intended use. It establishes the method's performance characteristics including accuracy, precision, linearity, range, selectivity, limit of detection, and limit of quantitation. Validation is important to have confidence in test results and ensure the method is fit for its intended purpose. It should be performed when developing new methods or using standard methods outside their intended scope. Method verification also confirms that a laboratory can correctly perform standard methods.

Uploaded by

fadli
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 54

What is method validation??

Confirmation, through the provision of objective ISO 9000


evidence, that the requirements for a specific
intended use or application have been fulfilled

Confirmation by examination and provision of ISO/IEC 17025


objective evidence that the particular requirements
for a specific intended use are fulfilled

Verification, where the specified requirements are VIM


adequate for an intended use

Confirm to Fit for your Purpose??


What do we get from Method Validation??
To have better understanding on the method performance
such as limitation, scope, working range……

How well of your method?

Demonstrate whether your method


is fit for a particular purpose

Confidence on your results


When I need to have Method Validation??

ISO/IEC 17025
Non-standard methods;
Laboratory-designed /developed methods;
Standard methods used outside their intended scope;
Amplifications and modifications of standard methods.

NATA Technical Note 17


For facilities involved in medical testing, elements of methods
endorsed “research use only” or “not for diagnostic”.

Eurachem
Validation is also required when it is necessary to demonstrate
the equivalence of results obtained by two methods, e.g. a
newly developed method and an existing standard / regulatory
method.
Method Verification

Standard Methods published by international


organization which already validate the
method by collaborative studies.
A laboratory using standard methods has to confirm that it has ability to carry
out those methods.
Verification is usually carried out by comparing the performance data obtained
by the laboratory when performing a standard method with those claimed by the
same method.
HOKLAS SC No. 20 - 8 July 2014

To demonstrate by meeting system suitability specifications established for the


method, as well as demonstration of accuracy and precision or other method
parameters for the type of method.
Verification of methods by the facility must include statistical correlation with
existing validated methods prior to use.

NATA Technical Note 17 - October 2013


What should I consider before Method Validation??
a) Purpose of measurement (what is to be identified and
why??)
b) What are the likely sample matrices??
c) Are there any interferences expected, and, if so,
should they be determined??
d) What is the scope (what are the expected
concentration levels or ranges)??
e) Are there any specific legislative or regulatory
requirements??
f) Are there any specific equipment accommodation and
environmental conditions that need to be
considered??
g) What type of equipment is to be used?? Is the method
for one specific instrument, or should it be used by all
instruments of the same type??
h) Method used for the preparation, sub-sampling,
procedure and including equipment to be used??

NATA Technical Note 17 - October 2013


Method Performance Characteristics

 Detection Limit / Quantitation Limit


 Linearity and Range
 Sensitivity
 Selectivity
 Accuracy
 Precision
 Ruggedness
 Measurement Uncertainty
Method Validation vs Method Verification

NATA Technical Note 17 - October 2013


Limit of Detection (LOD)

Method Detection Limit (MDL)


Instrumental Detection Limit (IDL)
• LOD based on visual evaluation
7 independent replicates, randomised

• LOD based on the standard deviation of the blank


Blanks n ≥ 20 or independent sample blank with n ≥ 10 each
LOD = Blank value + 3s

• LOD based on the range in which the calibration equation applies


yLOD = a+3sy/x= a +bxLOD  xLOD = 3 sy/x/b

• LOD based on signal-to-noise

There is no need to estimate the LOD or LOQ for methods that will always be
applied to measure analyte concentrations much greater than the LOQ.

NATA Technical Note 17 - October 2013


Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Limit of Detection (LOD)

If the test sample measured duplicate


(n=2), sample blank sample measured
once (nb=1)

1 1
𝑠𝑜′ = 𝑠𝑜 +
2 1

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Limit of Detection (LOD) for qualitative test
False Positive Rate and False Negative Rate

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Limit of Quantitation (LOQ)
The lowest level of analyte that can be determined
with acceptable performance.

Common practices are

LOQ = SD x 10; or
= LOD x 3; or
= Blank value + 10 x SD; or
= 50% above the lowest
fortification level used to
validate the method; or
= LOD x 10
NATA Technical Note 17 - October 2013
Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Linearity and Range
Instrument Working Range
Method Working Range

i) Confirm the relationship;


ii) Demonstrate the instrument working range is
compatible with the interval stated in the method
scope;
iii) Verify that the proposed instrument calibration
procedure (single point, bracketing, or multiple
points) is adequate.

NATA Technical Note 17 - October 2013


Calibration Curve
- Six calibration solution including blank or close to blank
- Evenly spaced over the range of interest
- The range should encompress 0-150% or 50-150% of the
concentration likely to be encountered (or concentration span
that exceeds the expected concentration range by ± 10% or
even ± 20%)
- The calibration standards should be run at least in duplicate
and preferably triplicate or more, in random order.

The following parameters should be considered:


Correlation coefficient (r)
Residual plot
0.9
X Variable 1 Residual Plot
0.8
y = 1.112x + 0.0435 0.01
0.7
0.6 R² = 0.9995 0.005
0.5

Residuals
0
0.4 0.00 0.20 0.40 0.60 0.80
0.3 -0.005
0.2
-0.01
0.1
0 -0.015
X Variable 1
0.00 0.20 0.40 0.60 0.80
Linearity and Range
Instrument Working Range
Method Working Range
i) Samples with known concentrations and sample blank should be available
ii) The samples used should be taken the entire measurement procedure
iii) The concentrations of the different samples should preferably cover the whole
range of interest
iv) The instrument should have been calibrated according to the suggested calibration
procedure.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Linearity and Range
Instrument Working Range
Method Working Range
i) Samples with known concentrations and sample blank should be available
ii) The samples used should be taken the entire measurement procedure
iii) The concentrations of the different samples should preferably cover the whole
range of interest
iv) The instrument should have been calibrated according to the suggested calibration
procedure.

The method working range needs to be established for each matrix


covered in the method scope.
This is because interferences can cause non-linear responses, and the
ability of the method to extract/recover the analyte may vary with the
sample matrix.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Sensitivity
• Change in instrument response which corresponds to a change
in the measured quantity
• Slope of calibration curve
• Not a particular important performance
• But useful to check of instrument performance and standards as
a quality assurance and quality control procedures.

NATA Technical Note 17 - October 2013


Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Selectivity
Interferences Poor Extraction

Wrong Identification Positive / Negative Bias

Matrix effect Background


Noise
The selectivity of a procedure must be established for in-house
developed methods, methods adopted from the scientific literature and
methods published by standardisation bodies used outside the scope
specified in the standard method.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Selectivity

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Checking Matrix Effect
• Pure standard vs standards taken through the analysis is
indicative of any losses of analyte which are related to the
method, while enhanced results may indicate reagent
contamination.

• Pure standard taken through the analysis compared with pure


standards added to extracted or digested extracts provides an
indication of matrix enhancement or suppression effects on the
detection system.

• Pure standards added to blank matrix after extraction or


digestion, compared to pure standards fortified in matrix prior to
extraction or digestion, provides an indication of losses of analyte
during processing.
Peak Area Ratio
(Matrix)

NATA Technical Note 17 - October 2013


Peak Area Ratio (Solution)
Selectivity
Interferences

Wrong Identification Matrix Effect / Background Noise

Confirmatory -Isotopic Internal Standard


Techniques / Method Calibration
-Matrix Calibration
-Standard Addition
.
Use highly specific procedures /
.
techniques
.
Overspiking with analyte suspected SANCO/12571/2013 - 01/01/ 2014
Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Selectivity

“Confirmatory method / technique can be applied to


both the identity and concentration.”

“Confirmation of analyte identity and concentration for positive


samples may be achieved using a different detection system or
column or using a specific detection system or using an alternative
analytical technique. In such cases validation of the confirmatory
technique must also be performed.”

NATA Technical Note 17 - October 2013

“Inevitably there is a trade-off between costs and time taken for


analyte identification, and the confidence with which one can decide
if the identification has been made correctly.”

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Selectivity
• EU (2002/657/EC)
• Requirement
• ID point (Identification Point) and Ion Ratio.

The variability of ion ratios should preferably be determined from calibration standards during the
initial method validation and subsequently during routine analysis. In certain cases, these data may
be used to set performance-based criterial, for individual analyte, rater than applying the fixed,
generic criteria given in Table 5.
SANCO/12571/2013 - 01/01/ 2014
Selectivity
Clenbuterol in Pork LC-MS/MS analysis

Ion ratio determined by


277203 (q) and 277132
Standards Ion ration 0.328
with %RSD 2.95%

Method A by enzymatic digestion Ion ratio 0.380


with column A (within 30% of tolerance)
Results by quantifying 277203 7.94 ppb

Method B by perchlorate digestion Ion ratio 0.331


with column A
Results by quantifying 227203 5.49 ppb
Accuracy

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Bias
Three general approaches:
i) Analysis of certified reference materials /
reference materials
ii) Recovery experiments using spiked samples
iii) Comparison with results obtained with another
method
Eurachem -The Fitness for Purpose of Analytical Methods, 2014

To take account any variation between runs, bias must be determined


over several days and preferably throughout the measuring range
and, if applicable, through the use of a suitable combination of
different specimens.

Several concentration levels must be incorporated in the


determination of bias. Otherwise the facility must be able to prove
that the method of analysis employed has the same trueness
throughout the measuring range.
NATA Technical Note 17 - October 2013
Analysis of certified reference materials
A strong evidence to demonstrate the trueness and the realization of
traceability.

NIST Special Publication 260-181, June 2014


Analysis of certified reference materials

NIST Special Publication 260-181, June 2014


Analysis of certified reference materials

NIST Special Publication 260-181, June 2014


Analysis of certified reference materials

NIST Special Publication 260-181, June 2014


RM should only be used for one purpose
during a validation study.
For example, an RM used for calibration
shall not also be used to evaluate bias.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Recovery experiments using spiked samples
Matrix selection

NIST Special Publication 260-181, June 2014

At least five different common food matrixes (protein, carbohydrate,


oil, dietary fibre and water), and at least three food types
representative of each food matrix.
HOKLAS SC-37- 8/07/ 2014
SANCO/12571/2013 - 01/01/ 2014
Recovery

HOKLAS SC-37- 8/07/ 2014

A practical default range of 60-140% may be used for individual


recoveries in routine multi-residue analysis.

SANCO/12571/2013 - 01/01/ 2014


Limitation by using spiked samples

The inherent problem with spike samples is that analyte introduced in such
a way will probably not be bound as strongly as that which is naturally
present in the test portion matrix and so the technique will give an
unrealistically high impression of the extraction efficiency.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Incurred tea leaves
Extraction
efficiency of
incurred sample

Wet overnight

Wet overnight

w/o wetting
DWM Sin, YL Wong. Analytical and Bioanalytical Chemistry To be published.
Comparison with results with another method

Reference method, such as standard method or


primary method, with smaller uncertainty than
candidate method

Method currently in routine use in the laboratory.


Eurachem -The Fitness for Purpose of Analytical Methods, 2014

The significance of the bias may be estimated by


statistical analysis (t-test) of the results obtained.

NATA Technical Note 17 - October 2013


Precision
Repeatability
Variability by same analyst, same equipment,
short timescale

Intermediate Precision
Single laboratory, different analysts,
different equipment, extended timescale

Reproducibility
Variability by different laboratories,
Different analyst, different equipment, long time scale

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Precision is generally dependent on analyte concentration,
and so should be determined at a number of concentrations
across the range of interest.

6 – 15 replicates for each material

Standard Deviation
sr , s i

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Simultaneous Determination of Repeatability
and Intermediate Precision

6 – 15 groups of duplicate measurement


on different analyte/days/equipment for
each material

One-way ANOVA

Repeatability Intermediate Precision


=Within group precision = Square root of the sum of
squares of the within-group and
between group
Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Eurachem -The Fitness for Purpose of Analytical Methods, 2014
Repeatability standard deviation obtained
with at least 6 degrees of freedom
e.g. 7 times in a series with one test item,
4 times in a series with two test items,
3 times in a series with three test items.

NATA Technical Note 17 - October 2013


Precision Limits
enables the analyst to decide whether there is a significant
difference, at a specified level of confidence, between results
from duplicate analyses of a sample obtained under specified
conditions.

Repeatability Limit (r) = 2.8 x sr

Reproducibility Limit (R) = 2.8 x sR

Eurachem -The Fitness for Purpose of Analytical Methods, 2014


Statistics Tools for Verification

Y1-Y2 < 16.7 mg/L for lemonade@~430 mg/L

BS EN 12857:1999
Statistics Tools for Verification
F-test by comparing the precision of the two methods
Should use repeatability standard deviation

F = (Repeatability SD of lab)2 / (Repeatability SD of Std Method)2


If F<F(95%): statistically no significant difference

BS EN 12857:1999
Ruggedness
A measure of its capacity to remain unaffected by small,
but deliberate variations in method parameters.
Ruggedness provides an indication of the method’s
reliability during normal usage.

To know which procedures / factors affect the


analytical results significantly.

In some cases, information may be available during in-house


method development. Intermediate precision, by their
nature, take into account some aspects of a method’s
ruggedness.

NATA Technical Note 17 - October 2013


Validation is always a balance
between costs, risks and technical
possibilities.

ISO/IEC 17025