myQA Daily Users Guide
myQA Daily Users Guide
SW Version: 1.2
Notice
This User’s Guide is an integral part of the myQA Daily, and should always be kept nearby. Observance
of the User’s Guide is required for proper performance and correct operation of the Detector. The myQA
Daily must not be used for any other purpose than described in the accompanying documentation
(intended use). Violation will result in loss of warranty.
The use of the myQA Daily and any other equipment described in this document, the generated data,
as well as the validation of the data prior to clinical use remain the sole responsibility of the user. IBA
Dosimetry and its distributors disclaim all warranties, either expressed or implied, including (but not
limited to) any implied warranties of fitness for a particular purpose and noninfringement. Additional
statements such as presentations, whether oral or written, do not constitute warranties by IBA
Dosimetry and should not be relied upon.
IBA Dosimetry does not accept liability for injury to personnel or damage to equipment that may result
from misuse of this equipment, failure to observe the hazard notices contained in this manual or failure
to observe local health and safety regulations.
No part of the accompanying documentation may be translated or reproduced without written
permission of IBA Dosimetry, unless the reproduction or translation is carried out for the sole purpose to
be included in the training materials used within the organization of the user.
If part of the accompanying documentation is provided in electronic form, these files shall not be
modified in any way. IBA Dosimetry and its suppliers retain title and all ownership rights to the
accompanying documentation (either in electronic or in printed form).
www.iba-dosimetry.com
Doc-ID: P-19-013-510-001 01
Last update: 2020-05-12
© 2020 IBA Dosimetry GmbH, Schwarzenbruck
4. HW Description .............................................................................. 13
4.1. 2D Detector Array .................................................................................................. 13
4.2. Labels of Connectors and LEDs ............................................................................. 16
4.3. Power Supply ......................................................................................................... 17
4.3.1. Battery Mode ......................................................................................................................... 17
4.3.2. External Power Supply Mode ................................................................................................ 17
4.4. Uniformity Correction ............................................................................................. 17
4.5. kTP Measurement ................................................................................................... 18
7. SW Application 2 - Operation....................................................... 42
7.1. Test Run Page....................................................................................................... 42
7.1.1. Test Run – Start-up Page: Task Overview............................................................................ 42
7.1.2. Test Run – Task Execution Page .......................................................................................... 45
7.1.3. Test Run – Summary Page ................................................................................................... 50
7.2. Trends Page .......................................................................................................... 52
7.2.1. Trend Chart Navigation ......................................................................................................... 52
7.2.2. Trend Chart Display .............................................................................................................. 53
7.2.3. Accessing Previous Task Executions ................................................................................... 54
7.3. Setup Page ............................................................................................................ 55
7.3.1. Setup - Machines - Welcome Page ....................................................................................... 55
7.3.2. Setup – Machines – Overview Page ..................................................................................... 55
7.3.3. Setup - Machines - All Beams Page ..................................................................................... 59
Experts: Users authorized by the product license holder and have read and understood myQA Daily
User’s Guide.
Trained personnel: Any personnel who have been trained by an expert or a trained user and have
read and understood myQA Daily User’s Guide.
Administrator: Two roles are required under the administration function: the site IT-administrator to set
up the network and product-administrator to set up user accounts and define measurement settings.
1.6. Warranty
The General Terms and Conditions of IBA Dosimetry GmbH apply. The user was provided with these
by the time the contract agreed.
IBA Dosimetry GmbH accepts no warranty or liability claims in the event of personal injury or damage to
equipment if this was caused by one or several of the following:
■ improper use of myQA Daily and the measurement device
■ improper installation, start-up, operation, and maintenance of myQA Daily
■ failure to observe the instructions in this manual
■ changes to the hardware carried out by the user
■ incorrect maintenance, repairs, and servicing work
■ damage caused by accidents or contamination
Manufacturer
This symbol is accompanied by the name and the address of the manufacturer.
Manufacturing date
This symbol is accompanied by a date to indicate the date of manufacture, as four
digits for the year, two digits for the month, and two digits for the day.
Caution: US Federal law restricts this device to sale by or on the order of a
physician.
IM PO RT ANT N O T IC E
ALWAYS KEEP THIS USER’S GUIDE AVAILABLE
Always store this User’s Guide near the system.
Always ensure this User’s Guide is easily accessible.
A copy of this User’s Guide is always available from IBA Dosimetry GmbH.
Conventions
The functions of the device, dialog captions and dialog text are indicated by bold font.
Examples: ON/OFF, DOSE.
Referrals to chapter and section headings in this manual are indicated by italic font.
Examples: Hardware Description, Technical Specifications.
Throughout this manual, hazardous situations or operations are identified by DANGER, WARNING,
CAUTION and NOTICE. They are intended to avoid accidents, to increase the lifetime and reliability of
the equipment, and to reduce production outage costs. They are indicated by specific signs and colors,
described below:
Sign Meaning
CAUTION, without the safety alert symbol, used to address issues related
to possible hardware damage.
Images
All numbers and selections displayed in dialog images are examples. Values and settings cannot be
regarded as reference.
IM PO RT ANT N O T IC E
ALL PERSONNEL MUST READ THIS CHAPTER
This user manual must be readily accessible to all personnel.
All personnel must read Chapter 2 and be fully aware of its contents before
installation, operating, or servicing the myQA Daily Device.
The general safety notices and the specific safety notes are also provided in
other chapters.
If the myQA Daily is used in a way not specified in this User’s Guide, the
protection provided by the equipment may be reduced.
IM PO RT ANT N O T IC E
ENSURE PROPER USE OF THE DEVICE
Read these instructions carefully before using the equipment for the first time.
Ensure the necessary safety precautions have been observed before using the
equipment for the first time.
2.3. Labels
Device label
Safety labels
C AUT IO N
USE THE SPECIFIED POWER SUPPLY ONLY!
Only use the power supply provided with myQA Daily: CINCON, TR30RAM090.
2.4. Operator
The myQA Daily should only be used by authorized personnel who are
■ aware of and understand the limitations of the device in measurements of radiation output,
■ knowledgeable about safety procedures to be observed when working with radiation sources,
■ aware of safety precautions required to avoid possible injury when using equipment.
W AR NI N G
USE BY INSUFFICIENTLY QUALIFIED PERSONNEL IS DANGEROUS
There is a risk of danger to staff and patients, and a risk of loss of operation if the
equipment is used by insufficiently qualified staff.
The license holder must clearly define the areas of responsibility, competence
and monitoring of the personnel.
The personnel for the above-mentioned areas of expertise must have the
necessary qualifications for this work (training, instruction). If necessary, this can
be conducted by the manufacturer on behalf of the license holder.
Failure to observe these conditions will result in a loss of warranty.
W AR NI NG
FAILURE TO OBSERVE THE ACCESS RIGHTS COULD BE DANGEROUS
The system must only be used according to the correct authorization.
No restrictions apply to physicists when using the system. Restrictions apply to
dosimetric assistants when using the system.
2.6.1. Measurement
Before using the myQA Daily, the operator must ensure that the equipment is correctly connected and
in proper working condition. The user must verify the general functionality, safety, and duly condition of
the device and data cables
Inside the myQA Daily device, sensors for pressure and ionization chamber temperature are used to
compensate the effect of pressure and temperature on the ionization chamber current (kTP correction).
W AR NI NG
ENSURE THE TEMPERATURE AND PRESSURE FOR CONSISTENCY WITH
AN EXTERNAL REFERENCE
Inside the myQA Daily device, sensors for air pressure an ionization chamber
temperature are used to compensate the effect of pressure and temperature on
the ionization chamber current [kTP correction]. To meet the specified
measurement accuracy, the temperature measurement must be correct within
±1°C and ambient air pressure measurement must be correct within ±3 mbar.
Please verify the correctness of the displayed temperature and air pressure with
respect to a calibrated measurement device.
W AR NI NG
TEST RUN IS PERFORMED WITH INCORRECT CONFIGURATION
Ensure that the correct settings are defined in Test Setup and the correct
configuration (detector orientation, build up, SDD, baseline measurements, etc.)
is used in Test Run. Otherwise, it may happen that the machine apparently
passed the QA tests may be actually not in conformity with the QA protocol.
2.6.2. Handling
Handle the myQA Daily and the connecting equipment with care – avoid impact, excessive weight on
the surface, and strong vibrations. Maximum load of build-up is 22.5 kg (25 g/cm2).
Do not handle the myQA Daily with wet hands.
If condensation is visible on the device, do not turn the device on. If already powered switch the device
immediately off. Condensed water on the device can lead to short circuits.
If an improper handling incident occurs, always perform a device check before using the device again.
C AUT IO N
CLEANING OF THE DEVICE
If necessary, clean the myQA Daily only with a dry soft tissue. The use of any
organic solvents is not recommended. Detailed instructions see Section 9.2.
C AUT IO N
DO NOT COVCER OR BLOCK THE DEVICE VENTILATION WINDOWS
During operation, do not set up the device such that the ventilation windows are
covered or blocked.
C AUT IO N
CHARGE THE BATTERY ONLY WITH THE DESIGNATED CHARGER
The delivered charger is specially designed for this type of battery. It may pose a
fire hazard if it is used for other kinds of batteries.
W AR NI NG
DO NOT USE THE DAMAGED BATTERY CHARGER
Do not use a damaged battery charger to avoid electrical shock.
C AUT IO N
PROTECT THE BATTERY FROM HEAT
Protect the battery from heat, for example, against constant sunlight and fire. An
overheating battery may catch fire or even explode.
IM PO RT ANT N O T IC E
STORAGE & DISPOSAL
Store in a cool, dry, and well-ventilated area. The battery can be stored at
temperature range -20°C to 60°C with relative humidity 10% to 100% including
condensation.
Regulations vary for different countries. Dispose of in accordance with local
regulations.
C AUT IO N
DO NOT STORE THE DEVICE IN THE ACCELERATOR ROOM
Due to the sensitivity of the electronics against radiation, never store the myQA
Daily device in the accelerator room. This would considerably shorten its lifetime.
Move the device into the accelerator room only if measurements are performed.
C AUT IO N
DO NOT EXPOSE THE DEVICE TO IONIZING RADIATION WHILE IN
STORAGE
Exposition to a high dose of ionizing radiation (e.g. stray radiation when stored in
the accelerator room) may cause degradation of electronic components
The myQA Daily device complies with requirements for electrical safety when operated in an
environment classified as “Pollution Degree 2”, i.e., only nonconductive pollution occurs and temporary
conductivity caused by condensation is to be expected (IEC 61010-1).
The device is to be used in altitude up to 2000 m.
0123
Note: The above regulatory approvals are only valid with the original antennas provided with the myQA
Daily device.
RoHS Compliance: The myQA Daily device is compliant with the European RoHS Directive (Restriction
of Hazardous Substances in electrical and electronic equipment), EC Directive 2011/65/EU
RED Directive: The myQA Daily device is in compliance with the essential requirements and other
relevant provisions of the European Radio Equipment Directive 2014/53/EU
IM PO RT ANT N O T IC E
PACKING LIST
There is a Packing List inside the delivered package. If any item is missing,
please report in the packing list, sign the list and send via postal mail or fax to
IBA dosimetry.
IM PO RT ANT N O T IC E
RETAIN ALL ORIGINAL PACKING MATERIAL
Retain all original packing material (shipping carton box, packing materials,
suitcase, etc.) in case of the future reshipments. When shipping the equipment,
use the original packing materials to avoid damages. The packing should be
done in the reverse order as given for unpacking.
C AUT IO N
DO NOT USE DAMAGED DEVICES
A device which was damaged during transport or on site shall not be used.
Damaged devices should be returned to IBA dosimetry.
Quantity Item
1 myQA Daily device
1 Battery, NH2057QE34, Inspired Energy, 7.2V 13.6Ah (97.9 Wh) Lithium Ion
Battery (8.0A capable)
1 Battery charger, Inspired Energy, CH5000X, one bay standard smart charger
1 Plug in power supply: CINCON TR30RAM090, 9V
Chamber area
10 cm × 10 cm
20 cm × 20 cm
Electronics
Fig. 4.1. myQA Daily (drawing) - top side and left side. The printed markers are:
X and Y-axes and two diagonal axes. The point of intersection is the iso-center
Locations of part of chambers (on the X and Y-axes and the center area) are indicated with the “+”
sign.
Printed field size marks, “10 cm × 10 cm” and “20 cm × 20 cm”, are scaled on the surface to take
into account of the projection from the source to the surface.
When the myQA Daily device is set up at isocenter SSD = 100 cm (aligned to the markers on the
sides), on the device surface:
“10 cm × 10 cm” is 9.93 cm × 9.93 cm
“20 cm × 20 cm” is 19.88 cm × 19.88 cm
Alignment markers on the sides (7 mm below the surface)
20 cm × 20 cm
3a 10 cm × 10 cm
3b 2
1
3d
3c
Fig. 4.3. Illustration of the chambers used for symmetry and flatness tests (highlighted with a purple
box), left: for 10 cm × 10 cm field; right: for 20 cm × 20 cm.
Test parameter 10 cm × 10 cm 20 cm × 20 cm
Field size 28 28
Symmetry / flatness 25 57
Energy 4 4
Output 9 9
Beam Symmetry
Beam symmetry Sx and Sy is calculated from the uniformity- and kTP -corrected measurement signals
Mi,j of the chambers marked in Fig 4.3 for the two reference fields. Indices i and j denote the chamber
position in X and Y direction, with i = j = 0 at the center position.
For crossline symmetry, the ratio RX,i,j = Mi,j / M−i,j is calculated from the measurement signals of each
chamber and its mirrored counterpart. The maximum ratio in percent is the measurement value for field
symmetry in crossline direction: SX = 100% · max{ RX,i,j }. Similarly, the field-symmetry in inline direction
SY = 100% · max{ RY,i,j }, where RY,i,j = Mi,j / Mi,−j.
Beam Flatness
Beam flatness F is calculated from the uniformity- and kTP -corrected measurement signals Mi,j of the
chambers marked in Fig 4.3 for the two reference fields. Indices i and j denote the chamber position in
X and Y direction, with i = j = 0 at the center position.
F = 100.0% ▪ max{ Mi,j } / min{ Mi,j }
Note: Beam flatness is also calculated for FFF beams. In the case of FFF beams, it is normal to obtain
unusually large values for the flatness parameters that also may be very sensitive to variations in
beam center offsets. If this raises a problem, please increase the acceptance tolerance for the
flatness parameter in the test-setup.
Battery
chamber
Table 5 Connectors and LEDs on the front side of the myQA Daily device
on – System booted
Green LED (lower) Activity Blinking: web-interface activated, signaling that the
device may be configured via the web-interface after
briefly pressing the Config button (Section 5.2.1).
Red LED Error Indicates a System Error (see more information below)
After turning on the power of the myQA Daily, ensure that the red error LED is off. If it turns on, please
try the following:
First turn device off and then on again.
If the red LED goes on again, connect to the web-interface, download the logs by clicking
Download all logs button (See Section 5.2.6) and send the logs to the Service Dept. of IBA
Dosimetry.
After that, or if the web-interface cannot be accessed, reset the device to default values via the
Config button by a long press (about 10 s), and then reconfigure the network settings (Section
5.2).
If red LED still is on, contact an IBA Dosimetry representative. For contact information, see
Section 13.1.
Exchange battery
To remove the battery, push onto the battery chamber hatch (see Fig. 4.4), and then move the latches
on both sides of the battery chamber to the outside position so that a red indicator is visible. You can
then open the hatch of the battery chamber and remove the battery by pulling on the plastic tab.
To insert a battery, open the battery chamber and insert the battery, with the contacts first facing
upwards into the battery chamber. Close the battery chamber hatch and move the locking latches to the
inside position. A green indicator should be visible next to both latches.
Note: Please recharge the battery when the battery charge indicator shows only one bar left, which
indicates a remaining capacity of 0% to 20%. The device will switch off when the remaining charge level
drops below 20%.
Note: Plug-in power supply does not charge the battery. Please charge the battery with the provided
battery charger (Section 4.3.1).
(1) Clients
(3)
WLAN Hospital User’s PC / mobile
or network (Web-browser)
Ethernet
myQA Daily device User interface
in a treatment room
(2)
Measurement data
myQA Daily server
Server / PC
(Hosting myQA Daily SW)
Fig, 5.1. Overview of myQA Daily measurement system - device connection. In this chapter, only
communication (1) is described while communications (2) and (3) are normal communication between the
device and your local area network.
The data communication also can be via Ethernet connection. It is a wired communication and the devices
are connected with Ethernet cables. Ethernet refers to the family of local area network (LAN) products
covered by the IEEE 802.3 standard. The myQA Daily device supports the data transfer rate up to 1
GBit/s.
By default, the hostname of myQA Daily device is “myqadaily-[S/N]”. The device serial number (S/N) is a 5-
digit number and can be found on the device information label on bottom of the device. This label also
contains two URLs for initial configuration http://myqadaily-[S/N].local/ or http://169.254.22.11/.
Note: If the device is connected to a network without a DHCP-Server, no static IP-Address is configured.
A link-local address (LLADR) from the 169.254.xx.yy Pool is used (Subnetmask 255.255.0.0). The
first address that is tried is 169.254.22.11.
Fig, 5.2. Illustration of direct connection between the myQA Daily device and a PC
Switch the device on and wait until the power button LED lights green.
Press briefly (< 1 s) the Config button on the device (Section 4.2). The Activity LED starts to blink.
There are three options to open configuration web interface:
1. via URL http://myqadaily-[s/n].local/ ([s/n] to be replaced by the device serial number)
2. via URL http://169.254.22.11/ (valid if the device is connected to the PC with a direct cable
connection or if no DHCP server exists on the local network)
3. via Windows Explorer
- Click to expand the Network browser, and then look for the device hostname (Fig. 5.3)
- Double-click the device name.
myqadaily-12345
Fig. 5.3. myQA Daily device name in the Network node in Windows Explorer
Note: If the device name cannot be found, check whether the Windows Network Discovery is enabled.
myqadaily-12345
Enter username, “admin”, password, (default: “admin”), and click LOGIN. The Device Info page opens.
Fig. 5.5. myQA Daily device configuration web-interface - Device Info page (example)
Among the device information, Hostname (or device name), Friendly Name, and Password can be
changed on the right side of the page. Click the Apply or the Change button after entering the new
information.
Note: If the hostname is changed, the new hostname is taken into account only after restarting the device
(powered off and back powered on).
Note: For security reasons, please change the password on the first login to the web-interface.
Click the Navigation icon in the upper-left corner to expand the Navigation panel:
myQA Daily User’s Guide | Setting up for Measurements | 21
Fig. 5.6. myQA Daily device configuration web-interface - Navigation panel
Different pages can be accessed by clicking the page name on the Navigation panel for different
configuration tasks. The item, Licenses, is about open source licenses and described in Section 11.2.
Fig. 5.7. myQA Daily device configuration web-interface - Configure Wireless page (example)
Fig. 5.8. myQA Daily device configuration web-interface - Configure Ethernet (wired) page (example)
Section Ethernet IPv4 respectively Ethernet IPv6 shows the current Ethernet IP settings.
Section Set Configuration allows selecting DHCP (selected by default) or static IP configuration.
If the configuration was changed, click on Save.
System Configuration
Click the Reset button to reset the device to the default settings.
Backup Configuration
Click the download Configuration button and then give a filename to the configuration and select a
desired location and save the file.
Restore configuration
Click the Select File button and then browse and select the saved configuration file to be restored.
Then click Restore configuration to restore the configuration.
Fig. 5.10. myQA Daily device configuration web-interface - System Update page
Fig. 5.11. myQA Daily device configuration web-interface - Download logs page
(Gun) (Target)
Chamber area
Electronics
Fig. 5.12. myQA Daily device under a LINAC and illustration of dose curve coordinates
Position the device in the center of the beam. Then align the effective measurement point, as indicated
by the markings on the sides of the device, to the isocenter. Use the room-lasers, if available. (See Fig.
5.12 and Fig. 5.13)
Room lasers
Turn on the light field and check the size of the light field with respect to the print on the top surface of
the device. The print is scaled for a source-axis distance (radiation source to isocenter) of 1000 mm.
At the LINAC console, choose the first field to be irradiated.
The myQA Daily device does not need warm-up to reach its specified measurement accuracy.
Pre-irradiation
Ionization chambers need a minimal amount of pre-irradiation to reach a stable response and their
maximum measurement accuracy.
When using the myQA Daily device with radiation fields of less than 100 MU and less than 1 Gy dose to
the ionization chamber array, a minimum amount of pre-irradiation of 0.25 Gy (field size: 25 cm × 25 cm).
If the dose of the irradiation is higher than or equal to 100 MU and the dose to the ionization chamber array
is higher than or equal to 1 Gy, no pre-irradiation is necessary.
For simplifying the procedure, it is recommended to pre-irradiate with 500 MU during the LINAC's warm up
phase in the morning.
Operating system
■ Windows 10, 64bit
IM PO RT ANT N O T IC E
COMPUTER REQUIREMENT
The myQA software may only be used on computers that are in compliance with the
applicable FCC and CE regulations.
SW installation procedure
Insert the installation disc into the DVD drive of the hosting PC or copy the installation files to the
server.
Double click myQA Daily.exe to open the installation wizard.
Select I agree to the License terms and conditions box to enable the INSTALL button.
SW uninstallation procedure
Click Start > Control Panel > Programs > Uninstall a program button.
Right-click myQA Daily and then click Uninstall.
To remove the user related application data including the database (dqa.db), delete the myQA Daily folder
the directory c:\ProgramData\IBA Dosimetry.
In the Start menu:
Type in “run” in the search box and then click the program Run.
Enter “%programdata%, press Enter.
Alternatively, enter “%programdata% in the Search area of an Explorer window and then press Enter.
Note: To keep the option of restoring the measurement data in the future, please keep a backup of dqa.db.
Fig. 6.2. Diagram of device setup for activating the myQA SW license
IM PO RT ANT N O T IC E
DATABASE BACKUP
The SW does not automatically back up the database. It is the user‘s responsibility
to back up the database regularly. Please ask your administrator to setup a backup
routine for you.
IM PO RT ANT N O T IC E
REFRESHING THE PAGE
Occasionally the user can be logged in only after refreshing the page. If you cannot
log in after entering correct credentials, please press the F5 key on the keyboard to
refresh your browser.
The main page display can be changed by selecting the page tab in the navigation bar. When the window
is reduced to certain size, the three main page tabs collapse into a small icon located at far-left of the bar
(Fig. 6.8). Click it to access the page tab again.
Opens a dropdown menu for setting up user accounts and view the
Settings icon information about the application (Section 6.8.3 )
The User Profile page is opened by clicking the Profile icon . It has two tabs:
General settings - Displays the current user’s account information that can be modified in this tab page by
the user (Fig. 6.9, left).
Change password – here user can change his own password (Fig. 6.9, right). See password
requirements in Section 0.
Once a change is made, the SAVE button is enable. Click SAVE to confirm the change.
Fig. 6.9. User Profile page – General settings tab (left) and Change password tab (right)
SW displayed language
By default, the English language is selected in the SW. However, the myQA Daily provides 11 languages.
On the User Profile - General settings page, the user can select a preferred language to be displayed in
the application in the Language box.
By default, the English language is displayed in the SW. However, the myQA Daily provides 10 languages.
Theme
There are options to select from different color themes for displaying myQA Daily.
Temperature unit
The myQA Daily application can only be accessed by the users with login credentials. When myQA Daily is
installed, an administrator account is automatically created. Additional administrative accounts can be
created. With an administrator account, new users with different roles can be created (Section 6.8.3.2).
Since User and General in the Settings menu require administrative privileges, these menu options will
only be visible to users with the role “Administrator”
Displays a User List containing all user accounts created for myQA Daily. For each
user account, its names, role (user rights), and email address are displayed.
Users Action commands are also available to edit or delete the account and change the
account password (Section 0). To create a new user, click the ADD USER button.
The User List can be filtered by clicking the Filter button on the top.
The About page displays general information about myQA Daily including SOUP
About
license information.
User List
Click the ADD USER button on the upper-right area of the Settings page. The Create User page
opens
Enter / select the required information in the Username, Display Name, Role (see user rights in Table
9), Email box, and then click the CREATE USER button. A temporary password is automatically
generated (Fig. 6.14). Copy or write down the password and click OK.
About password
The temporary password will be required to be changed after the first-time login.
myQA Daily requires complex passwords containing at least:
8 characters
one capital letter
one lower case letter
one number
one special characters ($, *, & …).
User rights
There are three roles in myQA Daily. The table below shows the granted user rights for each role.
Each user has three action command icons in the Action column of the User List (Fig. 6.13). See the
command description in Table 11.
Edit Account
Opens the Edit user page for editing the account (display name,
role, and email address)
Delete Account
Displays an overlay to confirm deleting the user account. Click Yes
to delete or click CANCEL to cancel the delete action.
The bookmark name can be customized by right-clicking the bookmark. Select Edit… and then change
the bookmark name in the Bookmark dialog.
■ When one user starts executing a task for a specific machine, no other user can access that task. Only
after finishing the task, it can be viewed by other users.
■ If a task is in execution, it is displayed on the Test Run - Task Execution page (Section 7.1.2).
■ Task results are displayed only for the current day on the Test Run – Task Execution - Summary
page (Section 7.1.2.4) and Test Run - Summary page (Section 7.1.3). The results of previously
executed tasks can be viewed only in the Trends page (Section 0).
■ Task executions can be repeated.
There are two ways to execute a task run: (1) measuring a sequence of beams (Section 7.1.2.1) - that
expects all beams to be executed in the defined order and (2) measuring individual beams (Section
7.1.2.2).
When creating a task, a physicist can define multiple beams for a task in the order that they are
programmed in the LINAC so that the measurement sequence (beam order) matches that in the LINAC
program. In this way, the user interaction in myQA Daily is minimized when the task is performed as a
sequence for all beams.
If single measurements already exist when a measurement sequence is started, all beams will be re-
measured. A measurement sequence can only be executed for all defined beams. There is no partial
execution.
It is possible to halt a task execution and continue the same task execution within two hours. After two
hours, any measurement data will be discarded, and the task will show as not in execution. This means, in
case of a network interruption or a browser crash, an existing task execution can be resumed just by
opening the task execution again. No data will be lost if it happens within two hours of the start.
If a task execution has been started and the user reopens it, a dialog with the option to resume (keep
already done measurements), restart or discard is displayed.
Sequence measurements can be stopped. A dialog asks the user if measured results should be retained or
discarded.
Note: When the device is disconnecting, after measurement is aborted, the browser might not respond.
Please refresh the browser by pressing F5 key on the keyboard to return to the test execution page.
Start button
1 3
Fig. 7.1. Test Run start-up page: Test Run – Task overview page
1. Machine list in the navigation panel – All Machines and individual machines are listed. By selecting
All Machines or an individual machine, the corresponding tasks of selected machine(s) are displayed
in blocks.
2. Execute date filter – it includes all days and all days in a week. All tasks in a week (by selecting All) or
the tasks of a particular day (by selecting this day) are displayed. By default, the current weekday is
selected. However, a different day can be selected so that the tasks not initially planned for the current
day can be executed.
When tasks are executed, they are always executed for the actual current day. The execution day
cannot be modified to an earlier or later day, i.e., the task results show up on the current executing day
but not its initially planned day. On the initially planned day, this task is still listed as not executed.
3. Main display (task blocks): displays the overview of tasks for the selected machine on the current day
(by default, or a user-selected day) in blocks. If a weekday is selected: The tasks are sorted by status
first, then machine name. If ‘All’ days are selected: The tasks are sorted by machine name (status is
ignored).
Fig. 7.2. Popup message when start a task that had been started
Click the CONTINUE button to keep the already measured data and continue measurement for
unmeasured beams.
Click the RESTART button to discard the measured data and start the measurement from the
beginning.
Click the CANCEL button to switch back to the Test Run – Task Overview page.
2a
1
2
Fig. 7.4. Test Run – Task Execution page with Start selected in navigation
The navigation (left) navigates the main display to Start (Section 7.1.2.1), individual assigned beams of the
task (Section 7.1.2.2), and Summary (Section 7.1.2.4) sub-pages. The main display displays the task
details of selected item in the navigation (Fig. 7.4):
■ Start page for execute the measurement sequence - to test all beams assigned to the task continuously
in sequence (Section 7.1.2.1).
■ Individual assigned beams for execute single beam measurements
By scrolling down the Start page or selecting the beam in navigation, this beam is displayed, where it
can be run individually (Section 7.1.2.2).
■ Summary page: once the task execution finishes, the page will jump to this page automatically (Section
7.1.2.4).
■ Discard command
By clicking Discard command on the upper-right corner, a confirmation dialog appears and asks to
confirm cancellation of the task execution. Click OK to confirm.
Note: The administrator / physicist can define security checklist or not in the Setup page. If there are no
checks defined, nothing will be displayed here.
Instructions (3)
Shows general instructions for performing measurements.
On the Measurement Sequence dialog (Fig. 7.5), the beams are represented by circles with status
indicated in the beam order from left to right (1):
Dash circle: waiting for the beam or it is currently being measured,
Solid circle with a green checkmark: measurement finished and status: passed,
Solid circle with a red exclamation mark: measurement finished and status: failed.
The information of the current beam is shown in the box below the (2) current beam circle: energy field
size, and monitor units; waiting for it or currently measuring.
ABORT SEQUENCE button: To stop running the sequence, click this button and then click YES to confirm.
The dialog shows that it is waiting for the first beam with certain energy and field size to be delivered.
Turn on the beam. Once a beam is detected, the measurement starts.
Once all monitor units are delivered and the beam is detected off, the dialog shows that it is waiting for
the next beam.
After the sequence has been completed or the sequence is aborted by clicking the ABORT
SEQUENCE button, the Measurement Sequence dialog is closed and the results are displayed in Test
Run – Task Execution – Summary (Section 7.1.2.4), where the user can check the overall results.
Click the FINISH TASK button to finish the task.
Fig. 7.6. Test Run – Start page – single beam section (example)
1 5
2
4
If the task is approved, the Approval sign will appear (2). If the secondary approval has not been
performed yet, the Approval pending sign will appear (3). To carry out this approval, a physicist or an
administrator can click the View button (1) and then click the Approve button in Fig. 7.8.
Optionally, a comment can be added before clicking the Approve button. To do so, an approval comment
tooltip will appear next to the approver:
Fig. 7.9. Test Run – Summary page – task secondary approval display
1 1a
Note: Users with physicist role can overwrite a task result on the Summary page.
The resulting QA status can be overwritten by clicking Overwrite Status and then selecting a new
status in the status dropdown list.
After entering comments in the Please provide a reason for status change box, red frame becomes
gray and the FINISH TASK button (2) becomes enabled.
Note: Only physicists can overwrite the status (dosimetric assistants cannot overwrite status).
The result overview of all beams is displayed by selecting Summary in navigation. Detailed results of
individual beams are shown by scrolling down the page or selecting the specific beam in navigation.
For description of task status and Overwrite Status, see section above.
View trends command:
Opens the Trends page (Section 0) and displays the dose output trends of the measured beams.
Print command:
The finished task execution page allows printing of the task. By clicking the Print command in the upper-
right corner in this Summary page or in the task block in the Test Run – Task Overview page, the
browsers default Print dialog opens. Please refer to the browsers print help for details on the available
print settings.
View command:
Clicking View opens the measurement and reference data graphical view, including the 2D data and 1D
inline and crossline profiles. By placing the cursor on a data point, a tooltip will appear displaying the
values of counts and dose in cGy of this data point.
Download command
By clicking Download, the measurement data of this beam will be downloaded in the .json format.
Fig. 7.15. Trends page (example). 1: Beams, 2: Parameters, 3: Trend Chart (Section 7.2.2), 4: Selected
task details (Section 7.2.3)
Machine/Task
■ A dropdown list of all registered machines.
■ A dropdown list of all tasks for the selected machine. By default the last measured task is displayed.
Beams
All beams assigned to the selected task are listed. Toggle the checkboxes to display (selected) or hide (un-
selected) a beam. Different beams are visualized by different colors in the chart (1).
■ Parameters
Check mark the boxes of the parameters of which the trends will be displayed. The different parameters
are displayed with different symbols (2).
52 | SW Application 2 - Operation | myQA Daily User’s Guide
7.2.2. Trend Chart Display
The chart shows deviation in length and percentage on separate axes. The axes for the deviation of Center
inline/crossline and Field Width inline/crossline in mm is on the left side of the chart. The axes for the
deviation of Symmetry inline/crossline, Flatness, Dose Output, and Energy Factor in percent is on the right.
The actual values of a data point can be viewed via a tooltip directly in the chart. If the reference dose (see
Section 7.3.3.3) is available, the dose in cGy is also displayed (see left figure below).
Fig. 7.17. Viewing the actual value of a measured data point on a trend chart (example) by placing the
cursor at the data point (examples)
By default, the trends are displayed for a period of the last 30 days. Other different lengths of periods can
be also selected with the Date filter dropdown menu.
Fig. 7.18. Date filter dropdown menu Fig. 7.19. Tolerance dropdown menu
In this dropdown menu, the tolerance of a beam / parameter can be selected (Fig. 7.19) to be displayed as
a green area around the trend curve (Fig. 7.15).
Individual curves in the chart can be hidden / displayed by clicking on the chart legend. They can be
removed/added to the chart by selecting or deselecting beams and parameters in the filters on the left side
of the page (Fig. 7.15).
Secondary approval
Fig. 7.20. Viewing a selected executed task (examples, with (left) and without (right) secondary approval)
In the task information on the left, click the View icon (see Fig. 7.20); the Test Run – Summary page
will open displaying the detailed task executions of the selected task and date.
Setup page
The Overview page allows to create, view, edit, or delete machines, tasks and beams. The All tasks and
All beams pages show more information and functions on tasks and beams, respectively. The reference
values are measured for individual beams in All beams page.
Only users with administrator or physicist rights can define measurement settings in the Setup page.
A machine must be created in myQA Daily QA for each LINAC that will be tested with the myQA Daily. It is
defined with a name and source axis distance (SAD).
The Add machine command and a list of added machines are displayed in the Machine Navigation
panel.
■ + Add machine command (1): Opens the Add machine dialog (Section 7.3.2.1).
■ List of machines (2):
i) Select a machine for displaying its beams and tasks in the main display.
ii) When a machine is selected (e.g., Linac 1), its SAD is displayed, and the Edit machine button
(7.3.2.1), Duplicate machine (Section 7.3.2.2) and Delete machine (7.3.2.3) button become
available.
Main display
The beams and tasks for the machine selected in the Machine navigation panel are displayed on the
main area of the page in blocks under Beams and Tasks, respectively. A beam stands for a specific
energy of a LINAC, and a task defines beams that will be measured during executing the task.
Beams blocks
1
7
2 3
Under Beams (Fig. 7.24), all beams created for the selected machine are displayed in blocks containing
1. Name of the task
2. Energy
3. Field size
4. Monitor Units (MU) to be delivered
56 | SW Application 2 - Operation | myQA Daily User’s Guide
5. Dosimetric tests that will be executed including, by default,
- Profile Constancy,
- Dose Output Constancy,
- Energy Constancy.
6. By clicking Beam Details the Setup - Machines - All beams page is displayed (Section 7.3.3),
showing the beam properties and parameters tolerances. The reference measurement can be
performed and any of the three default dosimetric tests can be enabled or disabled.
7. The Edit beam and Delete beam icons of the beam
Tasks blocks
1
6
2
3
Under Tasks (Fig. 7.25), all tasks created by the user are displayed in blocks containing:
1. Name of the task
2. Number of beams to be tested
3. Name of the measurement device
4. Planned test days
5. Task Details - By clicking Task Details, the Setup - Machines - All tasks page (Section 7.3.4) is
displayed with the task information (test dates, measurement device, assigned beams, and check list)
6. The Edit task and Delete task icons of the task.
By clicking Delete machine, a message pops up asking whether this machine should be deleted. Click
OK to confirm the delete action or CANCEL to cancel the delete action.
1 7
2
4 3
Beam navigation 6
Beam sections
Collapse
Expan
d
If changes have been made, click the SAVE CHANGES button to conform or CANCEL button to cancel
the changes.
2 8
8
5
9 6 7
Fig. 7.32. Expanded Reference Measurement block, left: no reference values, right: with reference values
Fig. 7.33. Current reference values displayed in the Reference Measurement dialog.
1
5
2 3
The reference Measurement dialog displays the information of the measurement device (1) and the beam
(2), reference dose (if entered) (3), measurement instructions (4) are on the left side and the Device /
Measurement Status box (5) and reference value (6, if exists) is the right side.
Turn on the beam. When a beam is detected, the measurement starts automatically.
When the beam is off, the measurement stops. The measured parameters are displayed.
The Measurement Status box informs whether the measurement is completed and the measurement
for the next beam can be started. The measurement can be repeated a few times as desired. The
parameters are the average of all measurements.
To calibrate for absolute dose, enter the absolute dose for a specific setup of the machine as the reference
dose in the Reference dose box in the Add beam dialog (see the Section 7.3.3.4). To do so, the entered
reference dose and the factor of dose per count, calculated as,
𝑑𝑜𝑠𝑒 𝑜𝑢𝑡𝑝𝑢𝑡
𝐹𝑎𝑐𝑡𝑜𝑟 = ,
𝑟𝑒𝑓𝑒𝑛𝑐𝑒 𝑑𝑜𝑠𝑒
will be displayed in the reference measurement results in the reference Measurement dialog.
When the reference dose is updated, the most recent value is used and displayed in the reference
measurements.
Fig. 7.36. Add beam dialog. Information for boxes with “*” is mandatory.
Task 1
1 2
4
Task 2
Fig. 7.38. Add task Dialog. Information for boxes with “*” is mandatory.
Minimum
information
Connect the device to the network and turn on the power so that the device becomes online.
On the right-side of the dialog, all online devices are displayed under Online Devices. Click on the
device description box to copy the parameters to the form on the right-side.
If the device is not online or the device is online but the network is configured such that the online
devices are not displayed, enter the information in the text fields on the left side manually. The minimum
requirement is the serial number and hostname/IP address of the device.
About the port number
If the device is online, the port number broadcasted by the (selected) device will be automatically taken.
If the device is not online or cannot be detected automatically, and the information needs to be entered
manually, please check what port number is used by the device and enter this number accordingly. By
default, it is set to “1338” in the device.
If there is a communication problem with the device, please see Table 6 in Section 5.3 for
troubleshooting.
Click SAVE to confirm adding task or CANCEL to cancel the action.
This a routine procedure for repeating the daily tests under the conditions that the measurement setup
(HW and SW) has been completely done.
1. Log in and go to the Test Run page of the machine to be checked. Today’s tasks will be displayed
2. Click the Start button in the task block which task will be executed.
3. If a security checklist exists, go through all check-items and check the corresponding box if it is passed.
4. Click the START MEASUREMENT SEQUENCE button and than turn on the beam to start the first
beam measurement.
5. Follow the instruction in the Measurement Sequence dialog to finish the measurements for the rest of
beams.
6. 3. Finish task by clicking the FINISH button.
7. For users of the role physicist, a task result can be overwritten by clicking Overwrite Status. An
overwrite reason is required to fill into the box below.
8. Check trends
Go to the Trends page to view trends for selected parameters and export trend charts.
9.6. Battery
Manufacturer: Inspired Energy, Order code: NH2057QE34
Lithium Ion Battery
7.2V 13.6Ah (97.9.Wh) Lithium Ion Battery (8.0A capable)
9.7. Mechanics
Item Characteristic
Outer dimensions (l × h × w) 560 mm × 60 mm × 320 mm
Weight 6.5 kg
10.2. Cleaning
To clean the housing of the myQA Daily, use a soft, dry duster, moistened with a cleaning agent. Do not
use organic solvents.
Use dry compressed air to clean the connectors on the rear side of the device, and the cable connectors.
C AUT IO N
CLEANING OF THE DEVICE
If necessary, clean the myQA Daily only with a dry soft tissue. The use of any
organic solvents is not recommended.
10.3. Repair
Reparation of extremely low-current devices such as ionization chamber detectors require special
knowledge. Do not try to the repair the myQA Daily yourself.
Table 12 RF characteristics
802.11b/g/n: 5 MHz
Channel spacing BT: 1 MHz 20/40/80/160 MHz
BLE: 2 MHz
Type of
DSSS, OFDM, FHSS and GFSK OFDM
modulation
64-bit and 128-bit WEP, TKIP, 128-bit AES-CCMP, 128-bit and 256-bit AES-
Encryption
GCMP
Note: For security reason, the FDA recommends to use WPA2 authentication with AES encryption.
IM PO RT ANT N O T IC E
Interfere with other wireless technologies
The wireless LAN is always enabled.
Be aware that other wireless technologies using the same frequency as the
myQA Daily may be interfered by or interfere with the data transmission of the
myQA Daily device.
Customer information
Address:
Product information
Appendices:
Sent to:
IBA Dosimetry
Bahnhofstraße 5
D-90592 Schwarzenbruck
Germany
Fax.: (49) 9128-607-26