Module 6
Module 6
As a light-based technique, a steady source able to emit light across a wide range of
wavelengths is essential. A single xenon lamp is commonly used as a high intensity
light source for both UV and visible ranges. Xenon lamps are, however, associated
with higher costs and are less stable in comparison to tungsten and halogen lamps.
For instruments employing two lamps, a tungsten or halogen lamp is commonly used
for visible light,2 whilst a deuterium lamp is the common source of UV light.2 As two
different light sources are needed to scan both the UV and visible wavelengths, the
light source in the instrument must switch during measurement. In practice, this
switchover typically occurs during the scan between 300 and 350 nm where the light
emission is similar from both light sources and the transition can be made more
smoothly.
Wavelength selection
In the next step, certain wavelengths of light suited to the sample type and analyte
for detection must be selected for sample examination from the broad wavelengths
emitted by the light source. Available methods for this include:
- Monochromators
A monochromator separates light into a narrow band of wavelengths. It is most often
based on diffraction gratings that can be rotated to choose incoming and reflected
angles to select the desired wavelength of light.1,2 The diffraction grating's groove
frequency is often measured as the number of grooves per mm. A higher groove
frequency provides a better optical resolution but a narrower usable wavelength
range. A lower groove frequency provides a larger usable wavelength range but a
worse optical resolution. 300 to 2000 grooves per mm is usable for UV Vis
spectroscopy purposes but a minimum of 1200 grooves per mm is typical. The
quality of the spectroscopic measurements is sensitive to physical imperfections in
the diffraction grating and in the optical setup. As a consequence, ruled diffraction
gratings tend to have more defects than blazed holographic diffraction gratings.3
Blazed holographic diffraction gratings tend to provide significantly better quality
measurements.3
- Absorption filters
Absorption filters are commonly made of colored glass or plastic designed to absorb
particular wavelengths of light.2
- Interference filters
Also called dichroic filters, these commonly used filters are made of many layers of
dielectric material where interference occurs between the thin layers of materials.
These filters can be used to eliminate undesirable wavelengths by destructive
interference, thus acting as a wavelength selector.1,2
- Cutoff filters
Cutoff filters allow light either below (shortpass) or above (longpass) a certain
wavelength to pass through. These are commonly implemented using interference
filters.
- Bandpass filters
Bandpass filters allow a range of wavelengths to pass through that can be
implemented by combining shortpass and longpass filters together.
Monochromators are most commonly used for this process due to their versatility.
However, filters are often used together with monochromators to narrow the
wavelengths of light selected further for more precise measurements and to improve
the signal-to-noise ratio.
Sample analysis
XRD TECHNIQUE
Max von Laue, in 1912, discovered that crystalline substances act as three-
dimensional diffraction gratings for X-ray wavelengths similar to the spacing of
planes in a crystal lattice. X-ray diffraction is now a common technique for the study
of crystal structures and atomic spacing.
X-ray diffraction is based on constructive interference of monochromatic X-rays and
a crystalline sample. These X-rays are generated by a cathode ray tube, filtered to
produce monochromatic radiation, collimated to concentrate, and directed toward the
sample. The interaction of the incident rays with the sample produces constructive
interference (and a diffracted ray) when conditions satisfy Bragg's Law (nλ=2d sin θ).
This law relates the wavelength of electromagnetic radiation to the diffraction angle
and the lattice spacing in a crystalline sample. These diffracted X-rays are then
detected, processed and counted. By scanning the sample through a range of
2θangles, all possible diffraction directions of the lattice should be attained due to the
random orientation of the powdered material. Conversion of the diffraction peaks to
d-spacings allows identification of the mineral because each mineral has a set of
unique d-spacings. Typically, this is achieved by comparison of d-spacings with
standard reference patterns.
Numerical: https://www.youtube.com/watch?v=6IzrOWIw3zQ
AAS
Atomic absorption spectrometry (AAS) detects elements in either liquid or solid
samples through the application of characteristic wavelengths of electromagnetic
radiation from a light source. Individual elements will absorb wavelengths differently,
and these absorbances are measured against standards. In effect, AAS takes
advantage of the different radiation wavelengths that are absorbed by different
atoms.
In AAS, analytes are first atomized so that their characteristic wavelengths are
emitted and recorded. Then, during excitation, electrons move up one energy level in
their respective atoms (figure 1) when those atoms absorb a specific energy. This
energy corresponds to a specific wavelength that is characteristic of the element.
Depending on the light wavelength and its intensity, specific elements can be
detected and their concentrations measured.
As electrons return to their original energy state, they emit energy in the form of light
(figure 2).
AAS has an unlimited number of applications and is still a popular choice for
uncomplicated trace elemental analysis. Flame atomic absorption spectrometry
(FAAS) is widely accepted in many industries, which continue to utilize the unique
and specific benefits of this technology. Graphite furnace atomic absorption
spectrometry (GFAAS) is an established technology for measuring elements at parts
per billion (ppb or µg/l) concentrations with incredibly low sample volumes.
IR
Infrared spectroscopy exploits the fact that molecules absorb specific frequencies that are
characteristic of their structure. These absorptions are resonant frequencies, i.e. the frequency
of the absorbed radiation matches the frequency of the bond or group that vibrates. The
energies are determined by the shape of the molecular potential energy surfaces, the masses
of the atoms, and the associated vibronic coupling.
In order for a vibrational mode in a molecule to be "IR active," it must be associated with
changes in the permanent dipole.
A molecule can vibrate in many ways, and each way is called a vibrational mode. Linear
molecules have 3N - 5 degrees of vibrational modes whereas nonlinear molecules have
3N - 6 degrees of vibrational modes (also called vibrational degrees of freedom). As an
example H2O, a non-linear molecule, will have 3 × 3 - 6 = 3 degrees of vibrational freedom,
or modes.
Simple diatomic molecules have only one bond and only one vibrational band. If the
molecule is symmetrical, e.g. N2, the band is not observed in the IR spectrum, but only in the
Raman spectrum. Unsymmetrical diatomic molecules, e.g. CO, absorb in the IR spectrum.
More complex molecules have many bonds, and their vibrational spectra are correspondingly
more complex, i.e. big molecules have many peaks in their IR spectra.
The atoms in a CH2 group, commonly found in organic compounds, can vibrate in six
different ways: symmetric and antisymmetric stretching, scissoring, rocking, wagging and
twisting.
NMR
SEM
TEM