1st Slides
1st Slides
1st Slides
INSTRUMENTAL ANALYSIS
Dr Aref Zayed
COURSE OUTLINE
1. Introduction to Instrumental Methods of Analysis.
2. Spectrometry:
Introduction to Electromagnetic Radiation
Quantum-Mechanical Properties of Radiation.
3. Ultra-Violet (UV) & Visible Spectrometry
4. Infrared Spectroscopy (IR).
5. Nuclear Magnetic Resonance (NMR) Spectroscopy :
Uses in Identification of Organic Compounds
6. Mass Spectrometry (MS): Uses in Quantitative and
Qualitative
7. Chromatography:
1. Liquid Chromatography (LC)
2. Gas Chromatography (GC).
PHARMACEUTICAL
INSTRUMENTAL ANALYSIS
Selecting an Analytical Method
INTRODUCTION
Types of Analyses:
1. Quantitative
2. Qualitative
2. Instrumental Methods:
Reliable and modern instruments were introduced.
Measurement of physical properties like conductivity,
electrode potential, mass to charge ratio, light
absorption/ emission.
Highly efficient separation techniques:
Chromatography, Electrophoresis.
INSTRUMENTS FOR ANALYSIS
An instrument for chemical analysis converts
information stored in the physical or chemical
characteristics of the analyte to information that
can be manipulated and interpreted by a human.
Stimulus Response
Energy Analytical
Source Information
Sample
TYPES OF INSTRUMENTAL METHODS
According to characteristic properties employed
to generate analytical signal:
2. Electrical Properties:
Electrical potential, charge, current or resistance.
3. Miscellaneous:
Mass-to-charge ratio, mass, radioactivity.
e.g. Mass Spectrometry (MS)
ANALYTICAL METHOD SELECTION
Sample
Instrument
ANALYTICAL METHOD SELECTION
bias= µ - xt
Concentration
ANALYTICAL METHOD SELECTION
Instrument Performance Characteristics
3. Sensitivity
Ability of method to distinguish between small
differences in analyte concentration.
Depends on :
Slope of calibration curve
Precision (reproducibility)
y = mx + b
Instrument
Instrument
Response
Response
50 100 50 100
Concentration Concentration
ANALYTICAL METHOD SELECTION
Instrument Performance Characteristics
4. Detection limit (DL): The minimum concentration or mass
of the analyte the can be detected at a known confidence
level.
LLOQ
Dynamic Range
Concentration
ANALYTICAL METHOD SELECTION
Instrument Performance Characteristics
Calibration curves
Standard addition methods
Internal standard method
ANALYTICAL PROCESS
•Obtain a representative bulk sample from the lot.
Sampling •Obtain a homogenous laboratory sample from the bulk sample
Report and
Evaluation
INTRODUCTION TO
SPECTROMETRIC METHODS
OUTLINE
Spectroscopy: Science the deals with interaction
of various types of radiation with matter.
Originally, only electromagnetic radiation with matter
Now include interaction of matter with other forms of
energy e.g. beams of electrons and ions
Amplitude (A)
Y: Electric field
X: Time or distance.
WAVE PARAMETERS
Amplitude (A): Length of electric vector at a maximum
in the wave
Frequency
c= v 𝛌
Proportionality
constant
(medium dependent)
QUANTUM MECHANICAL
PROPERTIES OF RADIATION
(Chapter 6)
COMMON SPECTROSCOPIC METHODS
Type Spectroscopy Usual Wavelength Range
NMR 0.6-10 m
IR absorption and Raman 0.78-300 µm
Scattering
UV visible absorption, emission 180-780 nm
and fluorescence
Questions:
Which color has a longer wavelength, red or
blue?
Which color has a higher frequency, yellow or
violet?
Which radiation has a higher energy, UV or IR?
PHOTOELECTRIC EFFECT
When electromagnetic radiation is emitted or
absorbed , a permanent transfer of energy occurs.
This cannot be explained by describing radiation
as waves but rather as a stream of discrete
particles called photons.
Discovery of photoelectric effect in the 19th
century made it necessary to adapt the particle
model.
PHOTOELECTRIC EFFECT
Hertz (1887): Spark jumped more readily
between 2 charged spheres when
illuminated with light.
Einstein (1905): Theoretical explanation of
photoelectric effect.
Millikan (1916): Systematic studies
confirmed Einstein’s theory.
hv
-
- - -
A
V
PHOTOELECTRIC EFFECT
Alkali metal-coated photocathode is irradiated
with monochromatic radiation (𝛌)
Electrons emitted from its surface with a range of
kinetic energies.
Current produced.
Main Results:
1. Light with constant frequency : intensity of
radiation 𝛼 current
2. Stoppage voltage V𝛐 depends on frequency (v)
3. Stoppage voltage V𝛐 depends on coating metal.
4. Stoppage voltage V𝛐 is independent of the intensity of
radiation.
PHOTOELECTRIC EFFECT
Frequency and maximum kinetic energy
Max kinetic energy,
Slope = h eV𝛐 = hv + ω
ω E = hv
Frequency v
𝒄
E = h = eV𝛐 - ω
𝝀
EXAMPLE
Calculate the energy of
(a) 5.3- Aሶ X-ray photon
(b) 530-nm photon of visible radiation
Solution:
𝒄
E = hv=h
𝝀
−
6.63 x 10 34 J.s x 3.00 𝑥 108 m/s
(a) E= − = 3.75 x 10-16 J
5.3 A x (10 10 m/A)
−
6.63 x 10 34 J.s x 3.00 𝑥 108 m/s =
(b) E= − 3.75 x 10-19 J
530 nm x (10 9 m/nm)
ENERGY STATES OF CHEMICAL SPECIES
Quantum theory:
1. Atoms, ions and molecules can exist only in
certain discrete states, characterized by definite
amount of energy. When a species changes its
state, it absorbs or emits an amount of energy
EXACTLY equal to the energy difference
between the states.
𝒄
Energy Level Energy
E1 - E0 = hv = h 1 -13.6 eV
𝝀
E2-E1= (-3.4) - (-13.6) = 10.2 eV 2 -3.4 eV
3 -1.51 eV
→ 𝛌 =1.21 x 10-7 m 4 -.85 eV
5 -.54 eV
3s
Fig. : Energy level diagram for Na atom the source of line spectrum
Rotational
Vibrational
CONTINUUM SPECTRUM
Very broad spectrum in emission from solids.
Produced by blackbody radiation- Thermal
excitation and relaxation of many vibrational and
rotational levels.
Molecules:
Electronic, vibrational and rotational energy levels –broad
band spectra with many features
Non-resonance Fluorescence:
Produced by irradiation molecules in
solution or gaseous state.
1. Molecules are excited into a vibrational
state in an excited electronic level.
2. Lifetime vibrational (10-15 s)<< excited
electronic state (10-8 s).
3. Non-radiative relaxation occur before
emission.
4. Energy (frequency) of emitted light <
energy of absorbed light.
5. The shift to lower frequency is called
stokes shift.
PHOSPHORESCENCE
Produced when excited molecule relaxes to
metastable excited electronic state (triplet state),
which has an average lifetime > 10-5 s.
Examples of drugs that have phosphorescence
property : Aspirin, benzoic acid and morphine,
QUANTITATIVE ASPECTS OF
SPECTROCHEMICAL
MEASUREMENTS
SPECTROCHEMICAL METHODS
Class Radiation Measured Type of Methods
Emission Emitted Radiation • Atomic Emission
Luminescence Luminescent Radiation • Atomic and Molecular
Fluorescence
• Phosphorescence
Scattering Scattered Radiation • Raman Scattering
• Turbidmetry
Absorption Incident and transmitted • Atomic and Molecular
Radiation (Absorbed) Absorption.
S=kc
Electrical Signal Concentration
Constant
ABSORPTION METHODS
Quantitative absorption methods require two
power measurements:
1. Before the beam pass through the sample (P0)
2. After the beam pass through the sample (P).
P0 c P
Absorbing solution of
concentration c
A = abc
A = 𝜖 bc
P0 P
S=kc
Electrical Signal Concentration
Constant
ABSORPTION METHODS
Quantitative absorption methods require two
power measurements:
1. Before the beam pass through the sample (P0)
2. After the beam pass through the sample (P).
P0 c P
Absorbing solution of
concentration c
A = abc
A = 𝜖 bc
P0 P
A = abc
A = 𝜖 bc
A = 𝑨𝟏%
𝟏𝒄𝒎 bc
Concentration
Specific absorbance
(g. 100 ml-1)
(absorbance of 1% w/v
or (w/v%)
solution in 1 cm cell)
MWT of Analyte
10𝜖
M=
(g. mol-1 ) 𝑨𝟏%
𝟏𝒄𝒎
PRACTICE EXAMPLES
Q1 : The molar absorptivity of a particular
chemical is 1.5 M-1 cm-1. What is the
concentration of a solution made from this
chemical that has an absorbance of 0.72 with a
cell path length of 1.1cm? (www.chem.ucla.edu)
A = 𝜖 bc
PRACTICE EXAMPLES
Q2 : A solution with a concentration of 0.14M is
measured to have an absorbance of 0.43. Another
solution of the same chemical is measured under
the same conditions and has an absorbance of
0.37. What is its concentration?
(www.chem.ucla.edu)
𝑨 𝟏 𝒄𝟏
= Same analyte, same conditions
𝑨 𝟐 𝒄𝟐
C1 = 0.12M
PRACTICE EXAMPLES
Q3: A compound had a molar absorptivity of
2.17 x 103 L cm-1 mol-1 . What concentration of the
compound would be required to produce a solution that
has a transmittance of 8.42% in 2.50-cm cell?
(Problem 6-19 Skoog 5th & 6th Ed, but different Numbers!)
A = 𝜖bc
A= -log T
c = 1.98 x 10-4 M
(c = 1.35 x 10-4 M in 6th Ed)
PRACTICE EXAMPLES
Q4: Chloroaniline (MWT =127.6) in a sample is
determined as the amine picrate. A 0.0265 g sample is
reacted with picric acid and diluted to 1L. The
absorbance of this solution in 1 cm cell is equal to 0.368.
What is the % chloroaniline in the sample? (Molar
absorptivity for chloroaniline = 1.25 x 104 L mol-1. cm-1)
A=𝜖bc
c chloroaniline= 2.94 x 10-5 mole/L
So there is 2.94 x 10-5 mole/L x 127.6 g/mole= 3.75 x 10-3 g in 1 L
solution.
3.75 x 10−3
x 100%= 15.0 % choloraniline in the sample
2.65 x10−2
PRACTICE EXAMPLES
Q5: A compound of formula weight 280 absorbed
65% of the radiation at a certain 𝛌 in a 2-cm cell at
concentration of 15.0 g/ml. Calculate its molar
absorptivity at the same 𝛌. (Past exam paper )
%T= 35%
T=0.35
A=-log 0.35= 0.456
𝜖 bc
M=
𝑨
Concentration
𝑴𝑨 (g. L-1)
𝜖=
bc
PRACTICE EXAMPLES
Q6: The absorbance of pure compounds X and Y each in 5x10-5 M
concentration are as follows:
Compound A 264 A 378
X 0.510 0.192
Y 0.335 0.150
𝑨 𝟏 𝒄𝟏
= Same analyte, same conditions
𝑨 𝟐 𝒄𝟐
c2 should be the same for the same compound at any wavelength
So calculate c2 for each compound at the 2 wavelengths
𝟎.𝟓𝟏𝟎 𝟓𝐱𝟏𝟎−𝟓 𝟎.𝟏𝟗𝟐 𝟓𝐱𝟏𝟎−𝟓
For X at 264 nm = . At 378 nm =
𝟎.𝟑𝟗𝟓 𝒄𝟐 𝟎.𝟏𝟒𝟕 𝒄𝟐
𝟎.𝟑𝟑𝟓 𝟓𝐱𝟏𝟎−𝟓 𝟎.𝟏𝟓𝟎 𝟓𝐱𝟏𝟎−𝟓
For Y at 378 nm = . At 378 nm =
𝟎.𝟑𝟗𝟓 𝒄𝟐 𝟎.𝟏𝟒𝟕 𝒄𝟐
The unknown is compound X (c2 calculated at 264 & 378 nm is the same)
QUANTITATIVE ANALYSIS OF MIXTURES
Beer’s law also applies to samples containing
more than one kind of absorbing species. As long
as there is no interaction between the species.
= 𝜖M b cM + 𝜖N b cN
QUANTITATIVE ANALYSIS OF MIXTURES
PRACTICE EXAMPLE
A 0.0450 M solution of para-aminobenzoic acid (PABA) had an
absorbance of 0.844 at 267 nm in a 1.00 cm cuvette, and an
absorbance of 0.034 at 240 nm. A 0.0366 M solution of nicotinic acid
had absorbances of 0.010 and 0.755 at 267 and 240 nm, respectively.
A MIXTURE of PABA and nicotinic acid had absorbances of 0.552
and 0.403 at 267 and 240 nm, respectively. Calculate the
concentration of PABA and nicotinic acid in the mixture.
%T= 35%
T=0.35
A=-log 0.35= 0.456
𝜖 bc
M=
𝑨
Concentration
𝑴𝑨 (g. L-1)
𝜖=
bc
PRACTICE EXAMPLES
Q6: The absorbance of pure compounds X and Y each in 5x10-5 M
concentration are as follows:
Compound A 264 A 378
X 0.510 0.192
Y 0.335 0.150
𝑨 𝟏 𝒄𝟏
= Same analyte, same conditions
𝑨 𝟐 𝒄𝟐
c2 should be the same for the same compound at any wavelength
So calculate c2 for each compound at the 2 wavelengths
𝟎.𝟓𝟏𝟎 𝟓𝐱𝟏𝟎−𝟓 𝟎.𝟏𝟗𝟐 𝟓𝐱𝟏𝟎−𝟓
For X at 264 nm = . At 378 nm =
𝟎.𝟑𝟗𝟓 𝒄𝟐 𝟎.𝟏𝟒𝟕 𝒄𝟐
𝟎.𝟑𝟑𝟓 𝟓𝐱𝟏𝟎−𝟓 𝟎.𝟏𝟓𝟎 𝟓𝐱𝟏𝟎−𝟓
For Y at 378 nm = . At 378 nm =
𝟎.𝟑𝟗𝟓 𝒄𝟐 𝟎.𝟏𝟒𝟕 𝒄𝟐
The unknown is compound X (c2 calculated at 264 & 378 nm is the same)
BEER’S LAW
Quantitative analysis of mixtures and deviations
from Beer’s law.
QUANTITATIVE ANALYSIS OF MIXTURES
Beer’s law also applies to samples containing
more than one kind of absorbing species. As long
as there is no interaction between the species.
= 𝜖M b cM + 𝜖N b cN
QUANTITATIVE ANALYSIS OF MIXTURES
PRACTICE EXAMPLE
A 0.0450 M solution of para-aminobenzoic acid (PABA) had an
absorbance of 0.844 at 267 nm in a 1.00 cm cuvette, and an
absorbance of 0.034 at 240 nm. A 0.0366 M solution of nicotinic acid
had absorbances of 0.010 and 0.755 at 267 and 240 nm, respectively.
A MIXTURE of PABA and nicotinic acid had absorbances of 0.552
and 0.403 at 267 and 240 nm, respectively. Calculate the
concentration of PABA and nicotinic acid in the mixture.
114
ABSORBING SPECIES
116
TYPES OF ELECTRONIC TRANSITIONS
Three types:
1. 𝝅, 𝜹 and n electrons. (mainly organic species)
117
ABSORPTION BY ORGANIC SPECIES (𝝅, 𝜹 and
n electrons)
All organic species absorbs electromagnetic radiation
because they contain valence electrons that can be
excited.
Single bonds: electrons have high excitation energy
so absorption is restricted to 𝛌 < 185 (vacuum UV
region)- experimentally difficult region.
Spectrophotometric investigation of organic
compounds is thus done at 𝛌 > 185 (long UV and
visible).
Absorption of long UV and visible is restricted to a
limited number of functional groups called
chromophores which contains electrons with low
excitation energy. 118
TYPES OF ABSORBING ELECTRONS
1. Electrons that participate directly in bonding
between atoms.
2. Nonbonding (unshared) outer electrons (O, S, N,
halogen)
Overlap of atomic orbitals results in molecular
orbitals which are occupied by bonding electrons.
This overlap results in low-energy bonding
orbital and high-energy antibonding orbital.
In the ground state, electrons of a molecule
occupy low-energy bonding orbital.
119
TYPES OF ABSORBING ELECTRONS
n
𝝅
𝜹
4 possible electronic transitions :
𝜹 → 𝜹*
𝛥 Energy
n → 𝜹*
121
𝝅 → 𝝅*
n → 𝝅*
TYPES OF MOLECULAR ORBITALS IN
FORMALDEHYDE
=𝜹
H =𝝅
C O =n
122
ELECTRONIC TRANSITIONS
𝜹 → 𝜹*
Requires large energy. (vacuum UV region).
Single bonds in organic molecules e.g C-C, C-H
e.g. CH4 exhibits 𝛌max 125 nm, C2H6 exhibits 𝛌max 135 nm
Absorption maxima due to 𝜹 → 𝜹 * are never observed in
ordinarily accessible region.
n → 𝜹*
Saturated compounds containing atoms with unshared
electron pairs (non-bonding electrons). E.g. CH3OH
Requires less energy than 𝜹 → 𝜹 * .
Absorption due to n → 𝜹 * is in the region 150-250 nm
with most peaks <200 nm.
𝛜 are low to intermediate and range between 100-3000 L cm-1
mol-1
Number of organic functional groups with n → 𝜹 * is small.
123
Absorption maxima shift to shorter 𝛌 in polar solvents.
EXAMPLES OF ABSORPTION DUE TO n → 𝜹 *
TRANSITIONS
124
ELECTRONIC TRANSITIONS
n → 𝝅 * and 𝝅 → 𝝅 *
Most absorption spectroscopy applications are based
upon these transitions.
Absorption peaks are in the region 200- 700 nm
(experimentally convenient spectral region).
Both require the presence of an unsaturated functional
group (double bond) to provide 𝝅 orbitals.
Unsaturated absorbing centres are called
chromosphores.
𝛜 for n → 𝝅 * 10 -100 L cm-1 mol-1, 𝛜 for 𝝅 → 𝝅 * 1000-
10000.
125
EXAMPLE OF ABSORPTION SPECTRUM
126
SOLVENT EFFECTS
127
SOLVENT EFFECTS
A small (5 nm) bathochromic (red) shift with
increasing solvent polarity affects both n → 𝝅 *
and 𝝅 → 𝝅 * transition.
Polarisation forces between the solvent and the
absorber lower energy level of both unexcited
and excited states, but effect on excited is
greater. Thus small effect is observed.
Effect is overshadowed in n → 𝝅 * due to the larger
hypsochromic effect.
In 𝝅 → 𝝅 * small bathochromic or red shift is observed
with increasing solvent polarity.
128
ORGANIC CHROMOPHORES
129
Absorption maxima depends on chromophore and is affected by
solvent and structural details of the molecule.
EFFECTS OF CONJUGATION OF CHROMOPHORES
In molecular orbital, electrons are considered to
be delocalised by conjugation (at least 4 atomic
centres involved)
Conjugation lowers 𝝅 * energy and thus
absorption maxima are shifted to longer 𝛌
(bathochromic effect).
e.g. 1,3-butadiene CH2=CHCH=CH2 has longer 𝛌 than
unconjugated diene.
Bathochormic effect is even larger when 3 double bonds
are conjugated and so on.
130
EFFECT OF MULTICHROMOPHORE ON
ABSORPTION
131
EFFECT ON MOLAR ABSORPTIVITY
Conjugation also results in higher absorption signal
(higher 𝛜) . This is called hyperchromic effect.
e.g. At 185 nm,1-hexene has a molar absorptivity of
about 10,000 L mol-1 cm-1 but hexa-1,4-diene has a
molar absorptivity of twice as much as 1-hexene.
However, when the double bonds are conjugated as
in hexa-1,3-diene the molar absorptivity is about
21,000 L mol-1 cm-1
132
EFFECTS OF CONJUGATION OF CHROMOPHORES
If conjugated
- shifts to higher l’s (red shift)
H2C=CHCH=CH2 lmax=217 emax = ~21,000 133
SPECTRAL NOMENCLATURE OF SHIFTS
134
SUMMARY
s → s* transition in vacuum UV (single bonds)
O=
e.g. CH2=CHCH2CH2CCH3 𝛌max = 282 nm (n→𝝅*)
O=
142
AUXOCHROMES
Auxochrome is a functional group that does not
itself absorb on the UV-Vis region but has the
effect of shifting chromophore peaks to longer 𝛌
as well as increasing their intensities.
Auxochromic substituents have at least one pair of n
pair of electron capable of interacting with 𝝅
electrons of the ring.
e.g. –OH –NH2 have auxochromic effects , particularly
with respect to the B band.
Interaction stabilises the 𝝅* state lowering its energy
thus resulting in bathochromic shift.
AUXOCHROMIC EFFECT
SOLVENTS
In choosing a solvent to perform absorption
measurement, solvent should be transparent in
the UV-Vis region.
Solvent transparency minimum: The wavelength
below which solvent cannot be used because of
absorption.
The type of solvent affects the absorption
maxima of the analyte.
Common solvents:
UV spectrophotometry : water, 95% ethanol,
cyclohexane and 1,4-dioxane.
Visible region: any colourless solvent.
UV-VIS ABSORPTION
INSTRUMENTATION
INSTRUMENT COMPONENTS
1. Sources
2. Wavelength selector
3. Sample container
4. Radiation transducer
5. Signal processors and readout devices
SOURCES
A continuum source is required whose power does
not change sharply over a considerable range of
wavelengths.