Sample For Solution Manual Measurement and Instrumentation 3rd Edition by Morris & Langari
Sample For Solution Manual Measurement and Instrumentation 3rd Edition by Morris & Langari
Sample For Solution Manual Measurement and Instrumentation 3rd Edition by Morris & Langari
hand, the foot and the cubit, and limitations due to variation in size of hands etc). Moving on to standard
units and improvements in these over time – e.g. platinum bar for length initially, then wavelengths of
radiation, then distance travelled by light in an interval of 1/299792458 seconds. Give examples of
standard units for other physical quantities. Mention Imperial, metric and SI units.
1.2 Discussion should cover primary sensor, secondary sensors, conversion elements, signal processing
elements, signal transmission element, signal presentation units and signal recording units.
Sample answer: Discussion should cover the primary sensor, explaining its function and giving
examples of primary sensors. Mention that some primary sensors like a liquid-in-glass thermometer are a
complete measurement system in themselves. Intelligent instruments also have one or more secondary
sensors. These measure the environmental conditions, particularly temperature and pressure, surrounding
a measurement system in order to correct the output of primary sensors affected by the environment
conditions.
Conversion elements are needed in a measurement system where the output variable of a primary
transducer is in an inconvenient form and has to be converted to a more convenient form. For instance,
the displacement-measuring strain gauge has an output in the form of a varying resistance. The resistance
change cannot be easily measured and so it is converted to a change in voltage by a bridge circuit, which
is a typical example of a conversion element. In some cases, the primary sensor and variable conversion
element are combined, and the combination is known as a transducer. Signal processing elements exist
to improve the quality of the output of a measurement system in some way. A very common type of
signal processing element is the electronic amplifier, which amplifies the output of the primary transducer
or variable conversion element, thus improving the sensitivity and resolution of measurement. This
element of a measuring system is particularly important where the primary transducer has a low output.
For example, thermocouples have a typical output of only a few millivolts. Other types of signal
processing element are those that filter out induced noise and remove mean levels etc. In some devices,
signal processing is incorporated into a transducer, which is then known as a transmitter. In addition to a
primary sensor, variable conversion element and signal processing element, some measurement systems
have one or two other components, firstly to transmit the signal to some remote point and secondly to
display or record the signal if it is not fed automatically into a feedback control system. Signal
transmission is needed when the observation or application point of the output of a measurement system
-5-
is some distance away from the site of the primary transducer. Sometimes, this separation is made solely
for purposes of convenience, but more often, it follows from the physical inaccessibility or environmental
unsuitability of the site of the primary transducer for mounting the signal presentation/recording unit.
The signal transmission element has traditionally consisted of single or multi-cored cable, which is often
screened to minimize signal corruption by induced electrical noise. However, fiber-optic cables are being
used in ever increasing numbers in modern installations, in part because of their low transmission loss
and imperviousness to the effects of electrical and magnetic fields. The final optional element in a
measurement system is the point where the measured signal is utilized. In some cases, this element is
omitted altogether because the measurement is used as part of an automatic control scheme, and the
transmitted signal is fed directly into the control system. In other cases, this element in the measurement
system takes the form either of a signal presentation unit or of a signal-recording unit. Give examples
Sample answer: The starting point in choosing the most suitable instrument to use for measurement of a
particular quantity in a manufacturing plant or other system is the specification of the instrument
characteristics required, especially parameters like the desired measurement accuracy, resolution,
sensitivity and dynamic performance. It is also essential to know the environmental conditions that the
instrument will be subjected to, as some conditions will immediately either eliminate the possibility of
using certain types of instrument or else will create a requirement for expensive protection of the
instrument. It should also be noted that protection reduces the performance of some instruments,
especially in terms of their dynamic characteristics (for example, sheaths protecting thermocouples and
resistance thermometers reduce their speed of response). Provision of this type of information usually
requires the expert knowledge of personnel who are intimately acquainted with the operation of the
manufacturing plant or system in question. Then, a skilled instrument engineer, having knowledge of all
the instruments that are available for measuring the quantity in question, will be able to evaluate the
possible list of instruments in terms of their accuracy, cost and suitability for the environmental
conditions and thus choose the most appropriate instrument. As far as possible, measurement systems and
instruments should be chosen that are as insensitive as possible to the operating environment, although
this requirement is often difficult to meet because of cost and other performance considerations. The
extent to which the measured system will be disturbed during the measuring process is another important
-6-
factor in instrument choice. For example, significant pressure loss can be caused to the measured system
Instrument cost is an important consideration. Generally, the better the characteristics, the higher
the cost. However, in comparing the cost and relative suitability of different instruments for a particular
are also very important because the instrument chosen will often have to be capable of operating for
long periods without performance degradation and a requirement for costly maintenance. In consequence
of this, the initial cost of an instrument often has a low weighting in the evaluation exercise.
Cost is very strongly correlated with the performance of an instrument, as measured by its
static characteristics. Increasing the accuracy or resolution of an instrument, for example, can only be
done at a penalty of increasing its manufacturing cost. Instrument choice therefore proceeds by specifying
the minimum characteristics required by a measurement situation and then searching manufacturers'
catalogues to find an instrument whose characteristics match those required. To select an instrument with
characteristics superior to those required would only mean paying more than necessary for a level of
As well as purchase cost, other important factors in the assessment exercise are instrument
durability and the maintenance requirements. Assuming that one had $20,000 to spend, one would not
spend $15,000 on a new motor car whose projected life was five years if a car of equivalent specification
with a projected life of ten years was available for $20,000. Likewise, durability is an important
consideration in the choice of instruments. The projected life of instruments often depends on the
conditions in that the instrument will have to operate. Maintenance requirements must also be taken into
As a general rule, a good assessment criterion is obtained if the total purchase cost and
estimated maintenance costs of an instrument over its life are divided by the period of its expected life.
The figure obtained is thus a cost per year. However, this rule becomes modified where instruments are
being installed on a process whose life is expected to be limited, perhaps in the manufacture of a
particular model of car. Then, the total costs can only be divided by the period of time that an instrument
is expected to be used for, unless an alternative use for the instrument is envisaged at the end of this
period.
characteristics, ruggedness and durability, maintenance requirements and purchase cost. To carry out such
an evaluation properly, the instrument engineer must have a wide knowledge of the range of instruments
available for measuring particular physical quantities, and he/she must also have a deep understanding of
-7-
how instrument characteristics are affected by particular measurement situations and operating
conditions.
1.4 Discussion should cover regulating trade, monitoring functions and control functions.
Sample answer: Present-day applications of measuring instruments can be classified into three major
areas.
The first of these is their use in regulating trade, applying instruments that measure physical
quantities such as length, volume and mass in terms of standard units. Give examples.
provide information that enables human beings to take some prescribed action accordingly. The gardener
uses a thermometer to determine whether he should turn the heat on in his greenhouse or open the
windows if it is too hot. Regular study of a barometer allows us to decide whether we should take our
umbrellas if we are planning to go out for a few hours. Whilst there are thus many uses of instrumentation
in our normal domestic lives, the majority of monitoring functions exist to provide the information
necessary to allow a human being to control some industrial operation or process. In a chemical process
for instance, the progress of chemical reactions is indicated by the measurement of temperatures and
pressures at various points, and such measurements allow the operator to take correct decisions regarding
the electrical supply to heaters, cooling water flows, valve positions etc. One other important use of
monitoring instruments is in calibrating the instruments used in the automatic process control systems
described below.
Use as part of automatic feedback control systems forms the third application area of
measurement systems. Give an example – e.g. a heating control system. Explain that the characteristics of
the measuring instruments used in any feedback control system are of fundamental importance to the
quality of control achieved. The accuracy and resolution with which an output variable of a process is
controlled can never be better than the accuracy and resolution of the measuring instruments used.
1.5 Mention and explanation is expected about specifying characteristics required in sensor,
environmental conditions around measurement system, choice of primary and possibly secondary
sensors, durability (lifetime) and cost, choice of conversion element in case of non-voltage sensor
Sample answer:
-8-
Specifying characteristics required: mention of parameters like the desired measurement accuracy, resolution,
Environmental conditions that the instrument will be subjected to: mention of temperature, humidity, dirt,
gases, flames or explosive environment. Some conditions will eliminate the possibility of using certain types
of instrument or else create a requirement for expensive protection of the instrument. Mention should be noted
that protection reduces the performance of some instruments, especially in terms of their dynamic
characteristics (for example, sheaths protecting thermocouples and resistance thermometers reduce their speed
of response).
Primary sensor: need to choose a sensor that satisfies characteristics required (accuracy etc) and will work in
expected environmental conditions (with protection if necessary). Also need to balance cost of sensor against
Secondary sensor: if characteristics of primary sensor will be affected by operating conditions, secondary
sensors need to be specified that will measure environmental conditions and allow output of primary sensor to
be corrected.
Conversion element: this will be needed if output variable of primary transducer is not in a voltage form and
has to be converted. Give example, e.g. displacement-measuring strain gauge has an output in the form of a
Signal processing: This improves the quality of the output of a measurement system in some way. Give
examples, e.g. electronic amplifier that amplifies the output of the primary transducer or conversion element,
thus improving the sensitivity and resolution of measurement. Mention that this is particularly important
where the primary transducer has a low output like a thermocouple with a typical output of only a few
millivolts. Other types of signal processing element are those that filter out induced noise and remove mean
levels etc.
Signal transmission: some measurement systems need a component to transmit the signal to some remote
point when the observation or application point of the output of a measurement system is some distance away
from the site of the primary transducer. Mention that this is needed where the environmental conditions (e.g.
heat) are not suitable for signal display/ signal recording equipment or a controller that the measurement system
feeds into. Briefly mention electrical (via copper wires), fiber optic, open air path, radio and pneumatic
transmission.
Signal display/recording: Some measurement systems include a component to display or record the
measured signal if it is not fed automatically into a feedback control system. Briefly discuss chart recorders,
(a) Imperial units: An early system of measurement units, characterized by having varying and cumbersome
multiplication factors relating fundamental units to subdivisions. Examples: length measurement units and
conversion factors (1760 for miles to yards, 3 for yards to feet and 12 for feet to inches). Other examples are
volume (cubic inch, cubic feet, gallon), mass (pound, hundredweight, ton), pressure (pounds per square inch,
atmosphere) etc.
(b) SI units: an internationally-agreed system of metric units where all multiples and sub-divisions of basic
metric units are related to the base unit by factors of ten, and such units are therefore much easier to use.
Examples are meter for length, cubic meter for volume, gram for mass, bar for pressure etc.
(c) Primary sensor: A sensor that measures the value of some measured quantity. Some examples are
(d) Secondary sensor: In some measurement systems, the output of the primary sensor is affected by the
environmental conditions (especially the temperature) of the environment that the sensor is operating in. One or
more secondary sensors are used to measure each aspect of the environment that is affecting the measurement
reading of the primary sensor). For example, if a pressure sensor is affected by the temperature of the
environment it is working in, a temperature sensor is used as a secondary sensor to measure the temperature of
the environment and allow the reading from the pressure sensor to be corrected.
(e) Signal processing: Signal processing elements exist to improve the quality of the output of a measurement
system in some way. A very common type of signal processing element is the electronic amplifier, which
amplifies the output of the primary transducer or conversion element, thus improving the sensitivity and
resolution of measurement. Other types of signal processing element are those that filter out induced noise and
2.1 Describe active and passive, null-type and deflection type, analog and digital, indicating and signal-
output types. Explain difference between these. Give examples of each. (See section 2.2 in book for
fuller details).
(a) quantity being measured modulates the magnitude of some external power source.
Relative merits:
Active: better measurement resolution (but limited by heating effect due to power source and also by
Explain difference by means of an example, e.g. null and deflection types of pressure gauge. (See section
Null-type normally reserved for calibration duties where best accuracy is needed.
2.4 Analog instrument gives an output that varies continuously as the quantity being measured changes.
The output can have an infinite number of values within the range that the instrument is designed to
measure but the number of different positions that the eye can discriminate between is strictly limited,
A digital instrument has an output that varies in discrete steps and so can only have a finite number of
values.
Digital instrument is best for computer control systems since analog to digital conversion is needed
for an analog instrument. Explain problems with A-D conversion – cost, conversion time (see section
2.2 in book).
-2-
2.5 Static characteristics: These are their steady-state attributes (when the output measurement value has
settled to a constant reading after any initial varying output) such as accuracy, measurement
Define and explain the various static characteristics (see section 2.3 in book).
Dynamic characteristics: describe behavior of a measuring instrument between the time a measured
quantity changes value and the time when the instrument output attains a steady value in response.
Depending on time allowed, student may be expected to sketch the main types of dynamic response
2.6 Definition and explanation expected for accuracy (or inaccuracy or measurement uncertainty),
precision (or repeatability or reproducibility), tolerance, range (or span), linearity (or non-linearity),
sensitivity of measurement, threshold, resolution, sensitivity to disturbance – explaining zero drift (or
bias) and sensitivity drift, hysteresis, dead space. Student should draw sketches to illustrate as
appropriate.
2.7 Accuracy more usually expressed as inaccuracy or measurement uncertainty. The latter quantify the
extent to which a measurement may be incorrect and are usually expressed as a percentage of the full
Precision is a term that describes an instrument's degree of freedom from random errors. If a large
number of readings are taken of the same quantity by a high precision instrument, then the spread of
Precision is often, though incorrectly, confused with accuracy. High precision does not imply
anything about measurement accuracy. A high precision instrument may have a low accuracy. Low
accuracy measurements from a high precision instrument are normally caused by a bias in the
Depending on time allowed, student should be expected to illustrate with an example of high precision
2.9 Dynamic characteristics describe the behavior of an instrument following the time that the measured
quantity changes value up until the time when the output reading attains a steady value. Various kinds
of dynamic behavior can be observed in different instruments ranging from an output that varies
slowly until it reaches a final constant value to an output that oscillates about the final value until a
steady reading is obtained. The dynamic characteristics are a very important factor in deciding on the
A zero-order instrument responds instantaneously (or effective so) to a change in measured quantity
A large number of instruments have a first order characteristic. This limits their use in control systems
because it is necessary to take account of the time lag that occurs between a measured quantity
changing in value and the measuring instrument indicating the change. Fortunately, the time constant
of many first order instruments is small relative to the dynamics of the process being measured, and so
no serious problems are created in such cases. However, if there is a need to sample the output of a
measurement system at a high frequency, the time lag before the instrument responds to a change in
the value of the measured quantity may preclude the use of instruments with a first-order
characteristic.
A second-order instrument has an oscillatory output unless damping is applied. When damped, the
output response resembles that of a first-order instrument and the above arguments apply.
(See discussion on first and second order instruments in section 2.4 of book for fuller explanation).
2.10 The accuracy of an instrument is a measure of how close the output reading of the instrument is to the
correct value. In practice, it is more usual to quote the inaccuracy or measurement uncertainty value
rather than the accuracy value for an instrument. Inaccuracy or measurement uncertainty is the extent
to which a reading might be wrong, and is often quoted as a percentage of the full-scale reading of an
instrument.
Likely error is 1.5% of 1100 ºC (the full scale reading), i.e. 16.5 ºC.
2.11 Tolerance is a term that is closely related to accuracy. It describes the maximum deviation of a
manufactured component from some specified value. For instance, crankshafts are machined with a
diameter tolerance quoted as so many microns (10 -6 m), and electric circuit components such as
Expected shortest rod is 5000 mm ‒ 2% = 4900 mm. Expected longest rod is 5000 mm + 2% = 5100
mm.
2.12 Maximum error is 1.5% of the full scale reading, i.e. 1.5% × 20,000 = 100 bar.
2.13 Tolerance is a term that is closely related to accuracy. It describes the maximum deviation of a
manufactured component from some specified value. Examples might be expected, such as crankshafts
are machined with a diameter tolerance quoted as so many microns (10 -6 m), and electric circuit
components such as resistors have tolerances of perhaps 5%.
Maximum deviation in length (given by the tolerance) is 1.5% of the nominal length, i.e. 1.5% × 250
mm = 3.75 mm. Thus shortest and longest bricks likely are 250 – 3.75 and 250 + 3.75, i.e. 246.25 and
253.75
2.16
2.17 (a) The maximum error expected in any measurement reading is 0.5% of the full scale reading, which
is 30 bar for this particular instrument. Hence, the maximum likely error is 0.5% x 30 bar = 0.15 bar
[30%]
(b) The maximum measurement error is a constant value related to the full scale reading of the instrument,
irrespective of the magnitude of the quantity that the instrument is actually measuring. In this case, as
worked out above, the magnitude of the error is 0.15 bar. Thus, when measuring a pressure of 5 bar, the
maximum possible error of 0.15 bar is 3% of the measurement value. [30%]
(c) If the measurement error is deemed to be too high, you could either use a higher quality instrument with
better accuracy (usually more expensive) or use a pressure sensor with a similar quality (and price) but
that has a smaller range, for example 0-10 bar. The recommended option is the latter (because it is
cheaper). [40%]
-5-
2.18
(a) The mean (average) value of the ten measurements made with the ultrasonic rule is 4.293 meters.
The maximum deviation below this mean value is -0.003 meters and the maximum deviation above the
mean value is +0.003 meters. Thus the precision of the ultrasonic rule can be expressed as ±0.003 meters
(±3 mm).
[50%]
(b) The correct value of the room width has been measured as 4.276 meters by the calibrated steel rule. All
ultrasonic rule measurements are above this, with the largest value being 4.296 meters. This last
measurement is the one that exhibits the largest measurement error. This maximum measurement error
can be calculated as : 4.296 – 4.276 = 0.020 meters (20 mm). Thus the maximum measurement
inaccuracy can be expressed as +20 mm.
[50%]
2.19
(a)
Precision is a term that describes an instrument's degree of freedom from random errors. If a large
number of readings are taken of the same quantity by a high precision instrument, then the spread of
Accuracy more usually expressed as inaccuracy or measurement uncertainty. The latter quantify the
extent to which a measurement may be incorrect and are usually expressed as a percentage of the full
Precision is often, though incorrectly, confused with accuracy. High precision does not imply
anything about measurement accuracy. A high precision instrument may have a low accuracy. Low
accuracy measurements from a high precision instrument are normally caused by a bias in the
An illustration showing high precision but low accuracy should be given (such as figure 2.5 in book).
[60%]
(b)
Precision expressed as the spread of values from minimum to maximum = 70.7 ˗ 69.4 = 1.3 Kg
Mean reading = 69.9 Kg; correct reading = 70.5 Kg; Inaccuracy = 0.6 Kg = 0.85% of correct value
[40%]
-6-
2.22 (a)
(b) From graph, output is 4.75 Volts for input of 16 bar. Thus, sensitivity is 4.75/16 = 0.297 V/bar
(c) Maximum non-linearity: On graph drawn, this is apparently the data point for input of 10 bar.
Measured output for data point at 10 bar input is 3.15V (taken from table of measured input-output data).
2.23 Sensitivity Drift (also known as scale factor drift) defines the amount by which an instrument's
sensitivity of measurement varies as environmental conditions (e.g. temperature and pressure) change.
Zero drift (also known as bias) describes the effect where the zero reading of an instrument is modified by a
change in environmental conditions. This causes a constant error that exists over the full range of measurement
of the instrument.
Sensitivity drift and Zero drift are collectively known as the Sensitivity to Disturbance of an instrument. As
variations occur in the temperature, pressure etc. in the environment surrounding a measurement system, certain
static instrument characteristics change, and the sensitivity to disturbance is a measure of the magnitude of this
change. This change can cause either zero drift or sensitivity drift, or sometimes both of these.
2.24 (a)
(b) The three measurement characteristics evident in the data plotted on the graph are non-linearity, zero drift
(c) Maximum non-linearity: On graph drawn, this is apparently the data point for input of 60 gram.
Measured output for data point at 60g input is 61g (taken from table of measured input-output data).
Thus non-linearity is 63 - 61 = 2g
Zero drift: This is the output when the input is zero. From the graph, zero drift = 4.5g.
Sensitivity calculated from graph (gradient of straight line fitted to the data
82.0 4.5 77.5 0.97 ) is
80 80
Hence, sensitivity drift 0.97 1.00 0.03
[Marking: 40% for part (a) , 20% for part (b) and 40% for part (c)]
- 10 -
2.25 Zero drift (also known as bias) describes the effect where the zero reading of an instrument is modified
by a change in environmental conditions (e.g. environmental temperature and pressure). This causes a
constant error that exists over the full range of measurement of the instrument.
The zero drift coefficient is the amount of change in the instrument output for a given change in an
sensitive to several environmental parameters, then it will have several zero drift coefficients, one for
If an instrument has a voltage output, the zero drift due to environmental temperature change would be
The zero drift at the temperature of 30ºC is the constant difference between the pairs of output readings,
The zero drift coefficient is the magnitude of drift (0.19 pascal) divided by the magnitude of the
temperature change causing the drift (10ºC). Thus the zero drift coefficient is 0.19/10 = 0.019 pascal/ºC.
[30%]
x 5 10 15 20 25 30
Students may make a graph of these data points and fit a straight line in order to determine the sensitivity.
This is acceptable. However, there is a quicker analytical solution in this case since the value of y
increments by exactly 13.1 for each increment of 5 in the value of x.
The sensitivity, expressed as the ratio y/x can therefore be calculated as 13.1/5 = 2.62
[40%]
(b) When the instrument is subsequently used in an environment at a temperature of 50°C, the table of
data values changes to the following:
x 5 10 15 20 25 30
As before, graphical analysis of the data is unnecessary since this time value of y increments by exactly
14.7 for each increment of 5 in the value of x. The new sensitivity, expressed as the ratio y/x can therefore
be calculated as 14.7/5 = 2.94
Hence, the sensitivity drift (change in sensitivity per °C) can be expressed as 0.32/30 = 0.01067/°C
- 11 -
[60%]
(°C)
20 21.5
35 36.5
50 51.5
65 66.5
The bias is the difference between the correct value and the measured value. This difference is 1.5°C for
each pair of measurements.
Hence, the bias due to the instrument being out of calibration is 1.5°C
2.30
x: 0 20 40 60 80 100
y: 0 31.5 63.0 94.5 126.0 157.5
(a) At 21°C, the deflection increases by 1.0 mm for each 50 Kg increase in load. Therefore, the sensitivity is
1.0/50 = 0.020 mm/Kg = 20 µm/Kg
At 35°C, the deflection increases by 1.1 mm for each 50 Kg increase in load. Therefore, the sensitivity is
1.1/50 = 0.022 mm/Kg = 22 µm/Kg
[40%]
(b) The total zero drift due to the increase in temperature is the change in deflection when the load is zero,
i.e. 0.2 mm.
The total sensitivity drift due to the increase in temperature is the change in sensitivity, i.e. (22 20) =
2.0 µm/Kg.
[30%]
(c) The zero drift of 0.2 mm is caused by a temperature increase of 14°C. Thus, the zero drift coefficient can
be expressed as 0.2/14 = 0.0143 mm/°C = 14.3 µm/°C
The sensitivity drift caused by the temperature increase of 14°C is 2.0 µm/Kg. Thus, the sensitivity drift
coefficient can be expressed as 2/14 = 0.143 (µm per Kg)/(°C).
[30%]
- 13 -
2.32
Let the temperature reported by the balloon at some general time t be Tr . Then Th is related to Tr by
Th T 0.012h 20 0.012h
the relation: Tr o
1 D 1 D 1 10D
20 0.072t
It is given that h = 6t (velocity is 6 m/s), thus: Tr
1 10D
The transient or complementary function part of the solution (Th = 0) is given by: Trcf Ce t / 10
The particular integral part of the solution is given by: Tr pi 20 0.072(t 10)
Thus, the whole solution is given by: Tr Trcf Trpi Ce t / 10 20 0.072(t 10)
Thus C = 0.72 and the solution can be written as: Tr 20 0.72e t / 10 0.072(t 10)
Using the above expression to calculate Tr for various values of t, the following table is constructed:
0 0 0.00 0.00
10 60 19.28 0.46
[50%]
(b) At 8000m, t = 1333.3 seconds. Calculating Tr from the above expression:
(c) Since the temperature falls according to the relation: Th = T0 - 0.012h, the true temperature at an
This result might have been inferred from the table above where it can be seen that the error has
converged to a value of 0.72. For large values of t, the transducer reading lags the true temperature
value by a period of time equal to the time constant of 10 seconds. In this time, the balloon travels a
distance of 60 metres and the temperature falls by 0.72°. Thus for large values of t, the output reading is
2.33
[50%]
The time constant is the time taken for the output reading to rise to 63% of its final value. Since the
output reading is rising from 0ºC to 100ºC, this means the time when the output has risen to 63 ºC.
Using the graph of temperature readings, this point is reached after 27 seconds. Thus the time constant
2.34 Note: it is assumed that students are familiar with the solution of a first order differential equation, as
commonly covered in mathematics courses.
The water temperature on the sea surface, T 0, is 20°C.
The temperature-measuring instrument characteristic is approximately first order with a time constant () of 50s.
It is also given that the submarine is diving at a rate of 0.5 meters/second, i.e. x 0.5 t
Part (a)
Let the temperature reported by the temperature sensor at some general time (t) be T r .
Using the above expression to calculate T r , for various values of t, the following table can be constructed:
Time (t) Depth (x) in Actual temperature Temperature Temperature error
meters (Tx) in C reading (Tr) in C in C
0 0 20.0 20.0 0.0
100 50 19.5 19.716 0.216
200 100 19.0 19.245 0.245
300 150 18.5 18.749 0.249
400 200 18.0 18.250 0.250
500 250 17.5 17.750 0.250
[70%]
2.35 For a step input, the general differential equation describing the behavior of a second order system can
be written as:
d 2qo dqo
a2 a1 a0qo b0qi (1)
dt 2 dt
where qi is the measured quantity, qo is the instrument output reading, and ao , a1 , a2 and bo are
constants.
b0qi
qo (2)
a0 a1D a2 D2
It is convenient to re-express the variables a0, a1, a2 and b0 in equation (2) in terms of three parameters
K b0 a 0 ; ω a 0 a 2 ; ξ a1 2 a 0 a 2
ξ can be written as ξ a1 a ω
1
2 a0 a 2 a0 2 a0
qo
b0 a 0qi (3)
1 a1 a 0D a 2 a 0D
2
b0 a1 2ξ D a 2 2 D2
K ; D ; D
a ω a
a0 0 0 ω2
Hence, dividing equation (3) through by q i and substituting for a 0 , a1 and a 2 gives:
q0 K
(4)
2 2
qi D 2 D 1
See Figure 2.12 in section 2.4(c) for sketches of instrument response in heavy damped, critically camped
Student should identify the critically damped case ( = 0.707) as the usual design target.
- 18 -
2.36
Let 𝑇𝑟 be the temperature indicated by the thermocouple at some time t and 𝑇𝑥 be the actual temperature.
𝑡
For a first order system with time constant 𝜏, 𝑇𝑟 = 𝑇𝑥 − (𝑇𝑥 − 𝑇0 )𝑒 − ⁄𝜏
Hence 𝑇𝑟 = 100 − [(100 − 20) × 𝑒 −8⁄4 ] = 100 − 80/𝑒 2 = 100 - 10.83 = 89.17C.