The document provides instructions for integrating a Topcon CV5000s/KB50s with a PC running CrystalPM software to export examination data. It describes changing the machine's print switch function and serial configuration settings to enable outputting data to the PC. It then explains using the Crystal Machine Integration Tool to receive readings from the CV5000 and import the machine as an EHR setting in CrystalPM to view imported examination records.

Uploaded by

José Meza
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
122 views4 pages

cv5000s Guide

The document provides instructions for integrating a Topcon CV5000s/KB50s with a PC running CrystalPM software to export examination data. It describes changing the machine's print switch function and serial configuration settings to enable outputting data to the PC. It then explains using the Crystal Machine Integration Tool to receive readings from the CV5000 and import the machine as an EHR setting in CrystalPM to view imported examination records.

Uploaded by

José Meza
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Topcon CV5000s/KB50s

Integration Guide

If the CV5000 has not been setup by Topcon support to export data to a pc, the initial settings will need to be
changed.
Holding down the shift button and pressing the menu button will bring up the menu settings.

Go to Function Settings:

Change [Print] Switch function from Printout to: Output Data or Output Data and Print.
Next go to Data I/O Settings:

Click on System Configuration. Uncheck Enable network connection if checked off and go to Serial Config at
bottom of page.
In Serial Configuration there will be several options. If they have an AR or LM already setup here leave that
alone. Choose the next available connection and select [I/O] Database connection. You will need to select the
COM port the machine is using. You can find that on the PC by right clicking on Computer and selecting
Manage, and choosing Device Manager from the popup window. In Device Manager select Ports to see what
COM ports are currently being used. Leave Baud Rate as is.

All setting on the CV5000s should now be correct for exporting data to a PC.

 Next move the Crystal Machine Integration Tool to the PC and save it in CrystalPM Folder. Launch the
tool and select CV5000 and make sure the Baud rate matches what was listed in the machine settings
on the Serial Config page.

 Send a reading from the CV5000 to the PC using the print button, a popup in the lower right hand
corner should confirm that data was received.

 In Crystal go to Records, select EHR Settings from the top of the page. Choose Import Machines > select
first available slot.

 A window will pop up with a list of machines (if the window is blank you will need to click on the
“Search online for additional machines” link located on the bottom left of the pop up window).
 Select Topcon CV5000 from the list of machines. It will ask you to navigate to the saved file. The file
should be saved in the CrystalPM folder > Mach > Mach1, select the .txt file.

 Click out of the records section and then go back in, select EHR Settings > Import Machines. You should
now see the Topcon in the list, clicking on that and following the prompts will allow you to view the
imported machine file and verify that it was imported correctly.
In order to have the machine values imported into your records buttons will need to be created by Crystal
Technical Support.