Catalog #EVV2K

ENDOTOXIN INDICATORS

INTENDED USE:

Endotoxin Indicators are designed for validation or monitoring depyrogenation processes such as dry-heat
sterilization and washing procedures. The extent of depyrogenation may be measured by comparing the levels
of endotoxin before and after a depyrogenation cycle using Charles River Endosafe® LAL Reagent. The U. S.
Pharmacopeia suggests that a depyrogenation cycle should reduce an Endotoxin Indicator by at least 1000 fold
(3-log reduction) in endotoxic activity as measured by LAL (Limulus Amebocyte Lysate) methods.1

y
EXPLANATION OF TEST:

l
Dry-heat sterilization is the preferred method for rendering glassware and other heat-stable materials free of

n
endotoxin. The U. S. Pharmacopeia, the Parenteral Drug Association’s Technical Bulletin No. 7, the LAL
Users Group and others have described approaches for preparation and use of Endotoxin Indicators.1-6

O
Charles River Endosafe ’s Endotoxin Indicators have been carefully prepared and assayed for endotoxin content
®

so that the endotoxicity in the vials may be recovered within a two-fold dilution of the labeled potency of the
reference standard lysate. The magnitude of depyrogenation may be assessed by comparing the amount of

n
endotoxin in exposed or processed Indicators to the amount in unmodified Indicators. The most
convenient method for Indicator quantitation is to select a kinetic LAL method and use either Charles

it o
River Endosafe® KTA (Product Code R150), a kinetic turbidimetric assay reagent, or Charles River Endosafe®
Endochrome-K (Product Code R170), a kinetic chromogenic assay reagent.

COMPOSITION:

a
Each vial contains approximately 2,000 EU (Endotoxin Units) of lipopolysaccharide from E.coli 055:B5, as
certified by Charles River Endosafe® with reference materials. The Indicator vial, made of Type I borosilicate glass,

m
and the dried endotoxin is not mixed with fillers or stabilizers.

r
Note: The biological activity for labeling of the Endotoxin Indicators was determined using the Gel-clot reference
lysate. However, an RSE/CSE analysis has not been performed with individual LAL batches. Prior to use, the end user

o
may wish to either verify the sensitivity of the Endotoxin Indicators to be within a two-fold of label claim, or perform

f
testing to assign a different biological activity with a specific LAL batch.

n
STORAGE: Store vials at 2 to 25 degrees C with closures and stoppers intact.

I
WARNING: Endotoxin Indicators contain pyrogenic amounts of endotoxin should they be administered by
parenteral routes. Endotoxin Indicators are for in-vitro use only, and not to be used in humans or animals.

PROCEDURES:

Exposure Procedure. For oven validation, expose Endotoxin Indicators at the sites specified. Immediately
before oven exposure, remove the vial closure and stopper and replace with aluminum foil and expose the
Indicator vial to the depyrogenation process.

Note: Vials labels are heat resistant and do not need to be removed prior to depyrogenation.

Preparation for Analysis. After the Endotoxin Indicators have been baked or processed, the vials from the
depyrogenation process and one or two unexposed Indicators (positive controls) should be prepared for
analysis. The Indicators should be assayed by a validated LAL-test procedure whereby the reagent sensitivity is
verified by a certified CSE (Control Standard Endotoxin, Product Code E110 or E120) or RSE. All vials
should be prepared for LAL testing by rehydrating each Indicator vial with 1 mL of LAL
Reagent Water and vortexing for two minutes initially, and then for one minute every ten minutes for
one-half hour. The LAL analysis should begin promptly for best recovery.
Catalog #EVV2K

ENDOTOXIN INDICATORS

Assay Procedure by Kinetic LAL Methods. A convenient method for analysis of Endotoxin Indicators is the use
of kinetic LAL methods from Charles River Endosafe®. The analysis is conducted by setting up a properly
controlled kinetic chromogenic or turbidimetric analysis using a 2-log standard curve of 0.05 to 5 EU/mL, and
testing a dilution of the positive control and undiluted samples of the baked or exposed
vials. Both the positive control(s) (PPC) and exposed vials should have PPC(s) run to show absence at
interference. Percent recovery for the PPC should be within 50% - 200% to be valid. The results of the
analysis more than satisfies the conditions for >3-log cycle reduction of endotoxin if log (EU/vial of control
vial) - log (EU/vial of exposed vial) is >3.

l y
An example calculation is provided wherein 2,000 EU per vial was recovered in the positive controls and
endotoxin levels recovered in exposed vials were less than lambda, the lowest point on the standard curve:

n
log reduction = log 2000 EU/mL - log 0.05 EU/mL
= 3.301 - (-1.301)

O
= 4.6

REFERENCES:

n
1. U.S. Pharmacopeia 33 - Official Monograph - Endotoxin Indicator for Depyrogenation.

it o
2. Parenteral Drug Association, Inc., Depyrogenation, Technical Report No. 7, 1985.

3. LAL Users Group. Preparation and use of endotoxin indicators for depyrogenation process studies, J. Parenteral Drug Assoc.,

a
43:109,1989.

4. Weary, M and F Pearson. A manufacturer’s guide to depyrogenation. Biopharm. 1, 1988.

m
5. Nakata, T, Destruction of challenged endotoxin in a dry heat oven, PDA J. Pharm. Sci. Technol., 48:59, 1994.

r
6. Hecker, W, D Witthauer and A Staerk. Validation of dry heat inactivation of bacterial endotoxins, PDA J. Pharm. Sci.

o
Technol., 48:197, 1994.

f
Charles River Endosafe

n
Division of Charles River Laboratories, Inc.

I
1023 Wappoo Road, Suite 43B
Charleston, SC 29407
Phone: (800) - 762-7016
Fax: (843) - 766-7576

PIEVV2K-02